K Number
K214024

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-09-12

(264 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.

Device Description

The SOFIA 88 Catheter is a non-tapered, single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal seqment is designed to facilitate vessel selection with 60 cm of distal-shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (SOFIA 88 Catheter), demonstrating its substantial equivalence to a predicate device. This type of regulatory submission primarily relies on performance (benchtop) and pre-clinical (animal) testing rather than clinical studies with human subjects or AI algorithm performance.

Therefore, many of the requested criteria (e.g., acceptance criteria for AI performance, sample size for test sets (human data), number of experts for ground truth, MRMC studies, standalone AI performance, training set details) are not applicable to this specific document as it pertains to a physical medical device (catheter) and not an AI/ML powered device.

However, I can extract the relevant information regarding the performance and equivalency testing conducted for the SOFIA 88 Catheter.


Device: SOFIA 88 Catheter

Type of Device: Percutaneous Catheter (physical medical device)


1. A table of acceptance criteria and the reported device performance

The acceptance criteria are generally implied by the "Pass" result for each test, indicating that the device met the pre-defined standards for each physical and functional characteristic. The document doesn't explicitly state quantitative acceptance limits for each test but rather confirms successful completion.

Test TitleTest Method SummaryReported Device Performance
Dimensional VerificationMeasured catheter length, proximal and distal outer diameters, and distal inner diameter.Pass
Coating Lubricity and DurabilityEvaluated hydrophilic coating for frictional force and durability.Pass
Simulated UseEvaluated performance under simulated use in a tortuous anatomical model to assess preparation, introduction, tracking, and support of the device.Pass
Dynamic and Static Burst PressureSubjected device to rated burst pressures to test catheter integrity.Pass
Air LeakageTested device for air leakage requirements.Pass
Liquid LeakageTested device for liquid leakage requirements.Pass
Tensile StrengthTensile tested device to failure, recorded the force at breakage.Pass
Flexural FatigueSubjected device to flexural fatigue from repeated bending during simulated use testing and from repeated hoop stress on the catheter from pressure testing.Pass
Particulate TestingMeasured number and size of particulates generated during simulated use in a tortuous anatomical model and compared results with the predicate device.Pass
Kink ResistanceEvaluated device for kink resistance after simulated use testing.Pass
Torque StrengthEvaluated device for torque strength by measuring the number of catheter rotations until failure after tracking through a tortuous anatomical model.Pass
Radio DetectabilityEvaluated device visibility under fluoroscopy.Pass
Corrosion ResistanceEvaluated device corrosion after immersion in saline.Pass
Small-bore ConnectorEvaluated luer connector for dimensional and performance requirements per reference standard.Pass
Stiffness ProfileCompared catheter stiffness profile along proximal and distal sections to the reference device.Pass
Biocompatibility Tests
CytotoxicityL-929 mouse fibroblast cells (MEM) with 6.0 cm²/mL at 37°C for 24 hrs.Non-cytotoxic
Irritation ReactivityNormal saline and sesame seed oil (SSO) on New Zealand White Rabbits with 6.0 cm²/mL at 50°C for 72 hrs.Non-irritant
MaximizationNormal saline and sesame seed oil (SSO) on Guinea Pigs with 6.0 cm²/mL at 50°C for 72 hrs.Non-sensitizing
Systemic ToxicityNormal saline and sesame seed oil (SSO) on Albino outbred strain (ND4) mice with 6.0 cm²/mL at 50°C for 72 hrs.Non-acute systemically toxic
PyrogenicityNormal saline on New Zealand White Rabbits with 6.0 cm²/mL at 50°C for 72 hrs.Non-pyrogenic
Hemocompatibility (Blood Loop)Loop system circulated with freshly drawn sheep blood; test article exposed to circulating blood at 37°C for 4 hrs.Thromboresistant
Hemolysis AssayPBS – Phosphate Buffered Saline / Blood from Three New Zealand White Rabbits; 6.0 cm²/mL at 50°C for 72 hrs.Non-hemolytic
Complement Activation AssayNormal Human Serum (NHS); 6.0 cm²/mL at 37°C for 60 min.Non-activator of complement system
Partial Thromboplastin Time (PTT) AssayHuman Plasma / Freshly Drawn Human Plasma; 6.0 cm²/mL at 37°C for 15 min.No effect on the PTT
Heparinized Blood Platelet and Leukocyte Count AssayHuman Blood / Freshly Drawn Human Blood; 12 cm²/mL at 37°C for 60 min.Pass
Animal Study
Acute Animal TestingComparison of SOFIA 88 Catheter to SOFIA 6F DAC for catheter tracking and tip stability in a porcine model. GLP-compliant. Examined for dissection/perforation, vasospasm, hemorrhage, necrosis, edema.Performed equally (SOFIA 88 Catheter and SOFIA 6F DAC). Found comparable. No issues noted.

2. Sample size used for the test set and the data provenance

  • Benchtop Testing: While specific sample sizes for each benchtop test are not provided in this summary, these tests generally involve a statistically significant number of device units (e.g., n=3, 5, or more depending on the test and standard requirements) to ensure reproducibility and reliability. The provenance is internal laboratory testing by MicroVention, Inc. and its contracted labs.
  • Biocompatibility Testing: Sample sizes are mentioned for some tests (e.g., New Zealand White Rabbits for irritation/pyrogenicity, Guinea Pigs for maximization, ND4 mice for systemic toxicity, sheep blood for blood loop assay, three rabbits for hemolysis, human serum/plasma for others). The provenance is laboratory testing conducted according to ISO and USP standards.
  • Animal Study: The document refers to "a porcine model" without specifying the exact number of animals. It's an acute animal study. The provenance is internal or contracted laboratory animal testing (conducted in accordance with FDA GLP Regulation 21 CFR Part 58).

All data provenance is internal R&D and pre-clinical testing, not human clinical retrospective or prospective data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the ground truth for a physical medical device like a catheter is established through objective physical and chemical measurements (benchtop tests) and observed physiological responses in animal models, not through expert human interpretation of medical images or data requiring "ground truth" establishment in the context of an AI algorithm.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as there is no human interpretation or subjective judgment that would require an adjudication method in the context of this device's testing. Measurements are objective and pass/fail criteria are pre-defined.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This device is a physical catheter, not an AI application, so no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is a physical device, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established via objective engineering and material science measurements (e.g., dimensions, pressure, force), validated laboratory biological assays (e.g., cytotoxicity, hemolysis), and observed physiological responses and pathological findings in an animal model. It is not derived from expert consensus, pathology, or outcomes data in the sense of clinical diagnostic accuracy.

8. The sample size for the training set

This is not applicable as this is a physical medical device, not an AI/ML powered device that requires a training set.

9. How the ground truth for the training set was established

This is not applicable for the reason stated above.

{0}------------------------------------------------

September 12, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

MicroVention, Inc. Alick Tan, Ph.D. Senior Regulatory Affairs Specialist 35 Enterprise Aliso Viejo, California 92656

Re: K214024

Trade/Device Name: SOFIA 88 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: August 12, 2022 Received: August 12, 2022

Dear Alick Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214024

Device Name SOFIA 88 Catheter

Indications for Use (Describe)

The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic on the left, resembling a series of concentric arcs in shades of blue and gray. To the right of the graphic is the word "MicroVention" in a blue sans-serif font, with "TERUMO" in a smaller, bolder blue font directly underneath.

K214024 510(k) Summary

510(k) OwnerMicroVention, Inc.35 EnterpriseAliso Viejo, CA 92656Establishment Registration No: 3013556777
Contact PersonAlick Tan, PhDSenior Regulatory Affairs SpecialistTelephone: 949-615-2603Email: alick.tan@microvention.com
Date summaryPreparedSeptember 12, 2022
Trade NameSOFIA 88 Catheter
Common NameCatheter, Percutaneous, Neurovasculature
ClassificationClass II, DQY, QJP
Regulation21 CFR 870.1250
Predicate DeviceSofia 6F PLUS/Distal Access Catheters (K150366)
Reference DeviceTracStar™ Large Distal Platform; ZOOM™ 88 Large DistalPlatform; ZOOM™ 88-T Large Distal Platform (K203764)

Device Description

The SOFIA 88 Catheter is a non-tapered, single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal seqment is designed to facilitate vessel selection with 60 cm of distal-shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.

Indications for Use

The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.

Comparison of Indications for Use and Technological Characteristics

Substantial equivalence of the SOFIA 88 Catheter to the predicate Sofia 6F PLUS/Distal Access Catheters (SOFIA 6F DAC) and to the reference device TracStar Large Distal Platform; ZOOM 88 Large Distal Platform; ZOOM 88-T Large Distal Platform (ZOOM 88) was established through an evaluation of the intended use, indications for use, principles of operation, device design, materials of construction, and an assessment of usability, safety, and effectiveness via bench and animal studies.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic to the left, resembling a circular or orbital design with varying shades of gray and blue. To the right of the graphic, the word "MicroVention" is written in a combination of blue and gray lettering, with "Micro" in blue and "Vention" in gray. Below "MicroVention", the word "TERUMO" is written in blue, in a slightly smaller and bolder font.

The data presented in this submission demonstrates the technological similarity and equivalency of the SOFIA 88 Catheter compared with the predicate device SOFIA 6F DAC (K150366) and the reference device ZOOM 88 (K203764).

The devices:

  • . have the same intended use,
  • . use the same principle of operation,
  • incorporate the same basic design,
  • use similar construction and material, and ●
  • are ethylene oxide (EtO)-sterilized and packaged using the same processes. ●

Device Comparison Table

ZOOM 88 (K203764)SOFIA 6F DAC(K150366)SOFIA 88 Catheter
Reference DevicePredicate DeviceSubject Device
Indications for UseThe ZOOM 88 andZOOM 88-T LargeDistal Platform areindicated for theintroduction ofinterventional devicesinto the peripheral,coronary, and neurovasculature.The SOFIAPLUS/Distal AccessCatheters areindicated for generalintravascular use,including the neuroand peripheralvasculature. It can beused to facilitateintroduction ofdiagnostic andtherapeutic agents. Itis not intended for usein coronary arteries.The SOFIA 88Catheter is indicatedfor generalintravascular use,including the neuroand peripheralvasculature. TheSOFIA 88 Cathetercan be used tofacilitate introductionof diagnostic agents ortherapeutic devices.The SOFIA 88Catheter is notintended for use incoronary arteries.
DeviceClassificationClass IIQJP, DQY21 CFR 870.1250Class IIDQY, DQO21 CFR 870.125021 CFR 870.1200Class IIDQY, QJP21 CFR 870.1250
ZOOM 88 (K203764)SOFIA 6F DAC(K150366)SOFIA 88 Catheter
Reference DevicePredicate DeviceSubject Device
Catheter BodyCommonly usedmedical grade plastics& metals withhydrophilic coatingOuter layer ofpolyurethaneelastomer, polyetherblock amide, andpolyamide; inner layerof metal braid/coil,PTFE, and polyolefinelastomerOuter layer ofpolyurethaneelastomer (differentpolyurethaneformulation), polyetherblock amide, andpolyamide; inner layerof metal braid/coil,PTFE, and polyolefinelastomer
Marker-Platinum/IridiumSame as SOFIA 6FDAC (K150366)
Hub-NylonSame as SOFIA 6FDAC (K150366)
Strain Relief-PolyurethaneSame as SOFIA 6FDAC (K150366)
IntroducerNonePebax®Fluorinated ethylenepropylene
Shaping MandrelNoneStainless steelNone
Catheter size8F6F8F, Same as ZOOM88 (K203764)
ID0.088 in (2.2 mm)0.070 in (1.8 mm)0.0875 in (2.22 mm)min
OD0.110 in (2.79 mm)max0.0825 in (2.10 mm)Proximal: 0.108 in(2.74 mm) max
Distal: 0.102 in (2.59mm) max
Effective Length80–110 cm115, 125, 131, 135 cm115 cm
CoatingHydrophilic coatingHydrophilic coatingSame as SOFIA 6FDAC (K150366)
Tip ConfigurationBeveled distal edge,soft, flexible,atraumatic tipStraightSame as SOFIA 6FDAC (K150366)
GuidewireCompatibility≤ 0.038 in0.035 in≤ 0.038 in
AccessoriesRotating hemostasisvalve (RHV)Introducer sheath andshaping mandrelIntroducer sheath
ZOOM 88 (K203764)SOFIA 6F DAC(K150366)SOFIA 88 Catheter
Reference DevicePredicate DeviceSubject Device
ConditionSuppliedSterile and single useSterile and single useSame
SterilizationMethodEtOEtOSame
PackagingConfigurationThe catheters areplaced in a protectivepolyethylene tube,mounted withaccessory RHV onto apolyethylenepackaging card,placed into a pouch,sealed and labeled.The sealed pouch andIFU are placed in alabeled shelf cartonbox.Catheter placed into aHDPE dispenser tube.Dispenser tube,introducer andshaping mandrelplaced on apolyethylenepackaging card that isinserted into a Tyvekpouch. Pouch and IFUplaced in bleachedsulfate carton box.Catheter placed into aHDPE dispenser tube.Dispenser tube andintroducer placed on apolyethylenepackaging card that isinserted into a Tyvekpouch. Pouch and IFUplaced in bleachedsulfate carton box.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for MicroVention, a company that specializes in neurovascular products. The logo features a stylized graphic of interconnected blue and gray circles on the left. To the right of the graphic is the company name "MicroVention" in blue, with the registered trademark symbol. Below the company name is the word "TERUMO" in blue, which is the parent company of MicroVention.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a circular graphic on the left side, composed of overlapping curved lines in shades of gray and blue, creating a sense of motion or rotation. To the right of the graphic, the word "MicroVention" is written in a blue sans-serif font, with a small registered trademark symbol next to it. Below "MicroVention", the word "TERUMO" is written in a larger, bolder blue sans-serif font.

Performance Testing

Performance testing was conducted to demonstrate the substantial equivalence of the SOFIA 88 Catheter to the predicate device.

The following testing data were provided to evaluate the performance of the SOFIA 88 Catheter.

TestTest Method SummaryResults
DimensionalVerificationDimensional specifications were verified bymeasuring the catheter length, proximal and distalouter diameters, and distal inner diameter.Pass
Coating Lubricityand DurabilityThe hydrophilic coating was evaluated for frictionalforce and durability.Pass
Simulated UsePerformance was evaluated under simulated use ina tortuous anatomical model to assess preparation,introduction, tracking, and support of the device.Pass
Dynamic and StaticBurst PressureDevice was subjected to rated burst pressures totest catheter integrity.Pass

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for MicroVention, a company that is part of Terumo. The logo features a stylized graphic of three curved lines that resemble a swirl or a vortex. To the right of the graphic is the word "MicroVention" in a blue sans-serif font. Below "MicroVention" is the word "TERUMO" in a smaller, bolder, blue sans-serif font.

TestTest Method SummaryResults
Air LeakageDevice was tested for air leakage requirements.Pass
Liquid LeakageDevice was tested for liquid leakage requirements.Pass
Tensile StrengthDevice was tensile tested to failure, and recordedthe force at breakage.Pass
Flexural FatigueDevice was subjected to flexural fatigue fromrepeated bending during simulated use testing andflexural fatigue from repeated hoop stress on thecatheter from pressure testing.Pass
Particulate TestingThe number and size of particulates generatedduring simulated use in a tortuous anatomical modelwere measured and results were compared with thepredicate device.Pass
Kink ResistanceDevice was evaluated for kink resistance aftersimulated use testing.Pass
Torque StrengthDevice was evaluated for torque strength bymeasuring the number of catheter rotations untilfailure after tracking through a tortuous anatomicalmodel.Pass
Radio DetectabilityDevice visibility was evaluated under fluoroscopy.Pass
Corrosion ResistanceDevice corrosion was evaluated after immersion insaline.Pass
Small-bore ConnectorThe luer connector was evaluated for dimensionaland performance requirements per referencestandard.Pass
Stiffness ProfileThe catheter stiffness profile along proximal anddistal sections was compared to the referencedevice.Pass

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic to the left, resembling interconnected circles or orbits in shades of gray and blue. To the right of the graphic is the word "MicroVention" in a blue sans-serif font, with "TERUMO" in a bolder, blue font positioned directly below it.

Biocompatibility Results

TestTest Method SummaryResults
Extract(s) & TestSystemsExtract Conditions
Cytotoxicity(ISO MEM ElutionTest)L-929 mousefibroblast cellsprepared using MEMMaintenance GrowthMedia with 5% FBS6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 37 °C for24 hrsNon-cytotoxic
Irritation Reactivity(ISO IntracutaneousReactivity Test)Normal saline andsesame seed oil(SSO) (testedseparately)New Zealand WhiteRabbits6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 50 °C for72 hrsNon-irritant
Maximization(ISO Guinea PigMaximization Test)Normal saline andsesame seed oil(SSO) (testedseparately)Guinea Pig6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 50 °C for72 hrsNon-sensitizing
Systemic Toxicity(ISO AcuteSystemic ToxicityTest)Normal saline andsesame seed oil(SSO) (testedseparately)Albino outbred strain(ND4) mice6.0 cm2/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 50 °C for72 hrsNon-acutesystemically toxic
Pyrogenicity(ISO/USP MaterialMediatedPyrogencity Test)Normal salineNew Zealand WhiteRabbits6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 50 °C for72 hrsNon-pyrogenic
TestTest Method SummaryResults
Extract(s) & Test SystemsExtract Conditions
HemocompatibilityIn-Vitro Blood LoopAssay(ISO In-vitro BloodLoop Assay withComparison Article)A loop systemcirculated with freshlydrawn sheep bloodThe test article isexposed to thecirculating blood thatfills to the capacity ofthe loop at 37 °C for4 hrsThromboresistant
HemocompatibilityHemolysis Assay(ISO ASTMHemolysis Assay)PBS – PhosphateBuffered SalineBlood - Three NewZealand WhiteRabbits6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 50 °C for72 hrsNon-hemolytic
HemocompatibilityComplementActivation Assay(ISOHemocompatibility:ComplementActivation Assay(SC5b-9))Normal HumanSerum (NHS)6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 37 °C for60 minNon-activator ofcomplement system
HemocompatibilityPartialThromboplastin(ISOHemocompatibility:PartialThromboplastinTime (PTT) Assay)Human PlasmaFreshly DrawnHuman Plasma6.0 cm²/mL (exposedsurface area toextraction mediumvolume ratio),extracted at 37 °C for15 minNo effect on the PTT
TestTest Method SummaryResults
Extract(s) & Test SystemsExtract Conditions
HemocompatibilityHeparinized BloodPlatelet andLeukocyteHuman Blood12 cm²/mL (exposed surface area to extraction medium volume ratio), extracted at 37 °C for 60 minPass
Freshly DrawnHuman Blood
(Hemocompatibility:Heparinized BloodPlatelet andLeukocyte CountAssay)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for MicroVention, a Terumo company. The logo features a stylized, abstract graphic to the left of the text "MicroVention". Below "MicroVention" is the word "TERUMO", indicating that MicroVention is associated with Terumo Corporation. The text is in a blue font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MicroVention TERUMO. The word "MicroVention" is written in a combination of blue and gray letters, with "Micro" in blue and "Vention" in gray. To the left of the word is a circular graphic made of curved lines in shades of gray and blue. Below "MicroVention" is the word "TERUMO" in blue, in a smaller, bolder font.

Animal Study

Acute animal testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulation (21 CFR Part 58) comparing the SOFIA 88 Catheter to the SOFIA 6F DAC. The testing was intended to assess preclinical safety and efficacy for catheter tracking and tip stability in a porcine model. The porcine model was chosen since the vessel sizes of the pig model allow for insertion and navigation of standard-sized devices used in humans; porcine vessel diameters are comparable with human vasculature. The tracking results demonstrated that the SOFIA 88 Catheter and SOFIA 6F DAC devices performed equally. Catheter tip stability was found to be comparable for both devices. No dissection/perforation was noted for the test-article- and controltreated vessels. Similar degrees of vasospasm were noted in the vessels instrumented with both devices. During postmortem examination there were no remarkable gross findings for any of the vessel samples for both the test and control devices. There was no hemorrhage, necrosis, or edema in test article-treated and control article-treated arterial sections. Overall, the histologic findings were consistent with routine catheterization procedures, which are commonly observed after interventional procedures in porcine safety models. The results of the present study did not raise any safety issues with either the test SOFIA 88 Catheter or control SOFIA 6F DAC. The devices are deemed equivalent.

Conclusion

MicroVention concludes through a review of the benchtop testing, non-clinical animal study assessments, the comparison of the device classification, indications for use, operating principles, technological characteristics, sterility, and biocompatibility testing that the SOFIA 88 Catheter is substantially equivalent to the predicate SOFIA 6F DAC. Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).