K Number
K213908

Validate with FDA (Live)

Device Name
SKR 3000
Date Cleared
2022-01-31

(48 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SKR 3000 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpose diagnostic procedures.

The SKR 3000 is not indicated for use in mammography, fluoroscopy and angiography applications.

Device Description

The digital radiography SKR 3000 performs X-ray imaging of the human body using an Xray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.

  • This device is not intended for use in mammography
  • This device is also used for carrying out exposures on children.

The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation - correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465.

This submission is to add new flat-panel x-ray detectors (FPDs), P-82 and P-85, into the SKR 3000. The P-82 and P-85 employ the same surface material infused with Silver ions (antibacterial properties) as the predicate device. The only difference between the P-82 and P-85 is the number of Li-ion capacitors. The P-85 has two Li-ion capacitors and the P-82 has one. These new P-82 and P-85 are not applicable to the serial radiography which acquires multiple frames of radiography image serially.

The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.

The SKR 3000 is distributed under a commercial name AeroDR 3.

AI/ML Overview

The provided text describes the Konica Minolta SKR 3000, a digital radiography system, and seeks 510(k) clearance by demonstrating substantial equivalence to a predicate device (K210619), which is also an SKR 3000 model. The submission focuses on adding new flat-panel x-ray detectors (FPDs), P-82 and P-85, to the existing system.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly defines "acceptance criteria" by comparing the specifications and performance of the subject device (SKR 3000 with P-82/P-85 FPDs) against its predicate device (SKR 3000 with P-65 FPD). The acceptance criteria are essentially the performance levels of the predicate device, which the new FPDs must meet or exceed.

Feature / Performance MetricAcceptance Criteria (Predicate P-65)Reported Device Performance (Subject P-82/P-85)Meets Criteria?
Indications for UseSame as SubjectGenerates radiographic images of human anatomy, replaces film/screen in general diagnostic procedures, not for mammography, fluoroscopy, angiography.Yes
Detection methodIndirect conversion methodIndirect conversion methodYes
ScintillatorCsI (Cesium Iodide)CsI (Cesium Iodide)Yes
TFT sensor substrateGlass-based TFT substrateFilm-based TFT substrateN/A (difference accepted, no new safety/effectiveness issues)
Image area sizeP-65: 348.8×425.6mm (3,488×4,256 pixels)P-82/P-85: 348.8×425.6mm (3,488×4,256 pixels)Yes
Pixel size100 µm / 200 µm / 400 µm100 µm / 200 µmYes (smaller range still includes acceptable sizes)
A/D conversion16 bit (65,536 gradients)16 bit (65,536 gradients)Yes
Max. ResolutionP-65: 4.0 lp/mmP-82/P-85: 4.0 lp/mmYes
MTF (1.0 lp/mm)(Non-binning) 0.62, (2x2 binning) 0.58(Non-binning) 0.62, (2x2 binning) 0.58Yes
DQE (1.0 lp/mm)56% @ 1mR59% @ 1mRYes (exceeds)
External dimensionsP-65: 384(W)×460(D)×15(H)mmP-82/P-85: 384(W)×460(D)×15(H)mmYes
IP Code (IEC 60529)IPX6IP56N/A (minor difference, presumed acceptable)
Battery TypeLithium-ion capacitorLithium-ion capacitorYes
Number of batteriesP-65: TwoP-82: One, P-85: TwoN/A (difference in configuration, performance evaluated)
Battery duration in standbyP-65: Approx. 13.2 hoursP-82: Approx. 6.0 hours, P-85: Approx. 13.2 hoursYes (P-85 meets, P-82 is different but acceptable for its configuration)
Surface MaterialSurface infused with Silver ions (antibacterial properties)Surface infused with Silver ions (antibacterial properties)Yes
Communication I/FWired and WirelessWired and WirelessYes
Operator console (Software)CS-7, AeroDR3 interface for P-65 (CTDS)CS-7, AeroDR3 interface for P-82 and P-85 (CTDS)Yes
Image ProcessingSame complex image processing algorithmsSame complex image processing algorithmsYes
Serial radiographyApplicableNot applicableN/A (difference in feature, not an "acceptance criterion" in this context as new FPDs don't support it)

Note: The acceptance criteria are largely implied by the claim of substantial equivalence. The document primarily focuses on demonstrating that new FPDs (P-82 and P-85) either match or improve upon the predicate's performance for critical imaging parameters. Differences in the TFT substrate material, pixel size options, number of batteries, and serial radiography capability are noted but explained as not raising new safety or effectiveness concerns.

2. Sample size used for the test set and the data provenance:

The document states: "The performance tests according to the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' and the other verification and validation including the items required by the risk analysis for the SKR 3000 were performed and the results demonstrated that the predetermined acceptance criteria were met."

This indicates that specific performance tests were conducted. However, the document does not explicitly state the sample size used for the test sets (e.g., number of images, number of phantom studies, number of human subjects, if any) nor the data provenance (e.g., country of origin, retrospective or prospective nature of clinical data if used). Given the type of device (X-ray system component) and the nature of the submission (adding new FPDs to an existing cleared system), the "performance data" presented is primarily technical specifications and phantom-based measurements, not typically large-scale clinical trials with human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not mention the use of experts to establish ground truth. As this is a technical performance comparison of imaging hardware (FPDs), ground truth would likely be established through objective physical measurements and established technical standards (e.g., imaging phantoms, dosimeters) rather than expert human interpretation of medical images for diagnostic accuracy.

4. Adjudication method for the test set:

Since there is no mention of human experts or clinical image interpretation studies, there is no adjudication method described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

The document does not mention an MRMC study, nor does it refer to AI or AI-assisted improvements for human readers. This device is a digital radiography system (hardware), and the submission focuses on its technical performance compared to a predicate, not on AI algorithms or their impact on reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device described is an X-ray imaging system, not an algorithm. Therefore, a standalone algorithm-only performance study is not applicable in this context. The performance evaluated is that of the hardware components (FPDs) within the system.

7. The type of ground truth used:

The ground truth for the performance parameters (e.g., Max. Resolution, MTF, DQE) would be established through objective physical measurements using standardized phantoms and test procedures as per industry standards (e.g., "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices"). For other specifications like battery life or dimensions, ground truth is based on engineering measurements and design specifications.

8. The sample size for the training set:

The document does not refer to a training set. This is because the submission is for hardware components (FPDs) for an X-ray system, not for a machine learning or AI-based diagnostic algorithm that would require training data.

9. How the ground truth for the training set was established:

As there is no mention of a training set, there is no information on how its ground truth would be established.

{0}------------------------------------------------

January 31, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Konica Minolta, Inc. % Jan Maniscalco Director of QA/RA Konica Minolta Healthcare Americas, Inc. 411 Newark-Pompton Turnpike WAYNE NJ 07470

Re: K213908

Trade/Device Name: SKR 3000 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, LLZ Dated: December 10, 2021 Received: December 14, 2021

Dear Jan Maniscalco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk Assistant Director Diagnostic X-ray Systems Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213908

Device Name SKR 3000

Indications for Use (Describe)

The SKR 3000 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpose diagnostic procedures.

The SKR 3000 is not indicated for use in mammography, fluoroscopy and angiography applications.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the Konica Minolta logo. The logo consists of a blue sphere with several horizontal white lines across it. Below the sphere, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

510(k) Summary K213908

Company:KONICA MINOLTA, INC.1 Sakura-machi, Hino-shi, Tokyo, 191-8511, Japan
Contact:Tsutomu FukuiSenior Manager of Regulatory & QMS DivisionQuality Assurance Operations, Healthcare Business UnitTelephone: +81 42 589 8429Email: tsutomu.fukui1@konicaminolta.com
Date Prepared:January 24, 2022
Device Name:SKR 3000
Common Name:Digital Radiography
Regulation Number:21 CFR 892.1680
Regulation Name:Stationary x-ray system
Regulatory Class:Class II
Product Code(s):MQB, LLZ
Predicate Device:K210619 - SKR 3000 (KONICA MINOLTA, INC.)Regulation Number: 21 CFR 892.1680Regulation Name: Stationary x-ray systemRegulatory Class: Class IIProduct Codes: MQB, LLZ

Device Description

The digital radiography SKR 3000 performs X-ray imaging of the human body using an Xray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.

  • This device is not intended for use in mammography

  • This device is also used for carrying out exposures on children.

  • Page 1 of 4 -

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Konica Minolta logo. The logo consists of a blue circle at the top with several white horizontal lines running across it. Below the circle is the text "KONICA MINOLTA" in a bold, sans-serif font.

The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation - correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465.

This submission is to add new flat-panel x-ray detectors (FPDs), P-82 and P-85, into the SKR 3000. The P-82 and P-85 employ the same surface material infused with Silver ions (antibacterial properties) as the predicate device. The only difference between the P-82 and P-85 is the number of Li-ion capacitors. The P-85 has two Li-ion capacitors and the P-82 has one. These new P-82 and P-85 are not applicable to the serial radiography which acquires multiple frames of radiography image serially.

The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.

The SKR 3000 is distributed under a commercial name AeroDR 3.

Indications for Use

This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.

This device is not indicated for use in mammography, fluoroscopy, and angiography applications.

Comparison Table

The comparison to the predicate device was summarized in the table blow.

Subject DevicePredicate Device
Device NameSKR 3000SKR 3000
510(K) NumberK213908K210619
Indications for UseThis device is indicated for use ingenerating radiographic images ofhuman anatomy. It is intended to areplace radiographic film/screenThis device is indicated for use ingenerating radiographic images ofhuman anatomy. It is intended to areplace radiographic film/screen
Subject DevicePredicate Device
system in general-purpose diagnosticprocedures.This device is not indicated for use inmammography, fluoroscopy, andangiography applications.system in general-purpose diagnosticprocedures.This device is not indicated for use inmammography, fluoroscopy, andangiography applications.
SpecificationDetection methodScintillatorTFT sensor substrateImage area sizeIndirect conversion methodCsl (Cesium Iodide)Film-based TFT substrateP-82: 348.8×425.6mm(3,488×4,256 pixels)P-85: 348.8×425.6mm(3,488×4,256 pixels)Indirect conversion methodCsl (Cesium Iodide)Glass-based TFT substrateP-65: 348.8×425.6mm(3,488×4,256 pixels)
Pixel size100 μm / 200 μm100 μm / 200 μm / 400 μm
A/D conversion16 bit (65,536 gradients)16 bit (65,536 gradients)
Max. ResolutionP-82: 4.0 lp/mmP-85: 4.0 lp/mmP-65: 4.0 lp/mm
MTF (1.0 lp/mm)(Non-binning) 0.62(2x2 binning) 0.58(Non-binning) 0.62(2x2 binning) 0.58
DQE (1.0 lp/mm)59% @ 1mR56% @ 1mR
MechanicalExternal dimensionsP-82: 384(W)×460(D)×15(H)mmP-85: 384(W)×460(D)×15(H)mmP-65: 384(W)×460(D)×15(H)mm
IP Code (IEC 60529)IP56IPX6
BatteryTypeNumber of batteriesBattery duration instandby statusLithium-ion capacitorP-82: OneP-85: TwoP-82: Approx. 6.0 hoursP-85: Approx. 13.2 hoursLithium-ion capacitorP-65: TwoP-65: Approx. 13.2 hours
Surface MaterialSurface infused with Silver ions(antibacterial properties)Surface infused with Silver ions(antibacterial properties)
Communication I/FWired and WirelessWired and Wireless
Peripherals,Cables/minorcomponentsAeroDR Interface Units, DetectorInterface Units, Power Supply Unit,Generator Interface Units, BatteryAeroDR Interface Units, DetectorInterface Units, Power Supply Unit,Generator Interface Units, Battery
Subject DevicePredicate Device
Charger, etc.Charger, etc.
Operator console(Software)CS-7- AeroDR3 interface for P-82 and P-85(CTDS)CS-7- AeroDR3 interface for P-65 (CTDS)
Image ProcessingAuto-gradation processFrequency processing (F process)Equalization processing (E process)Hybrid processing (HF process - HEprocess)Hybrid smoothing process (HS process)REALISM processing (RE process - RFprocess)Realism smoothing process (RSprocess)Grid removal process/ScatteredRadiation Correction (Intelligent-Grid)Automatic exposure field recognitionprocessAuto-gradation processFrequency processing (F process)Equalization processing (E process)Hybrid processing (HF process - HEprocess)Hybrid smoothing process (HS process)REALISM processing (RE process - RFprocess)Realism smoothing process (RSprocess)Grid removal process/ScatteredRadiation Correction (Intelligent-Grid)Automatic exposure field recognitionprocess
Serial radiographyNot applicableApplicable

Page 2 of 4 -

Giving Shape to Ideas

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circle with several horizontal white lines running across it. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

ー Page 3 of 4 -

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circular emblem at the top with several horizontal white lines running across it. Below the emblem, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

Performance Data

The SKR 3000 is designed to comply with the following standard; AAMI/ANSI ES 60601-1 (Ed.3.1), IEC 60601-1-2 (Ed.4.0), and ISO 10993-1 (2018). The performance tests according to the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" and the other verification and validation including the items required by the risk analysis for the SKR 3000 were performed and the results demonstrated that the predetermined acceptance criteria were met. The results of risk management did not require clinical studies to demonstrate the substantial equivalency of the proposed device.

Conclusion

The SKR 3000 has the same intended use and indications for use, technological characteristics, and principal operations. The technological differences raised no new issues of safety or effectiveness as compared to its predicate device (K210619). Performance tests demonstrate that the SKR 3000 performs according to specifications and functions as intended. All the information to demonstrate assurance of our evaluation is attached to relevant sections of this submission.

Therefore, as for our conclusion, the SKR 3000 is substantially equivalent to the predicate devices and presents no new questions of safety or effectiveness.

  • Page 4 of 4

Giving Shape to Ideas

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.