K Number
K213843
Date Cleared
2022-03-31

(112 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

The subject device is a single use, powder free nitrile examination glove. The subject device is black. It can be available in four specifications: S,M,L,XL. The subject device is non-sterile, ambidextrous.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for Nitrile Examination Gloves (K213843). This submission is for a medical device (gloves), not an AI/Software device. Therefore, the questions related to AI/Software performance, such as MRMC studies, ground truth establishment for training sets, and expert adjudication, are not applicable.

The document assesses the substantial equivalence of the Nitrile Examination Gloves to a predicate device (K171422) based on non-clinical performance testing.

Here's the information extracted from the document concerning the acceptance criteria and study proving the device meets them:

1. A table of acceptance criteria and the reported device performance

The document provides a "Summary of non-clinical performance testing" in Table 2 (pages 6-7), comparing acceptance criteria with test results.

Test MethodPurposeAcceptance CriteriaResults
ASTM D6319Physical Dimensions TestLength (mm): S: ≥220; M/L/XL: ≥230Length (mm): S: ≥220; Pass
M/L/XL: ≥230; Pass
Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10Width (mm): S: 83-88 /Pass; M: 95-97/ Pass; L: 106-109/ Pass; XL:111-116/ Pass
Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Thickness (mm): Finger: 0.16-0.18/Pass; Palm: 0.12-0.14/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.5S: 0/125/Pass; M: 0/125/Pass; L: 0/125/Pass; XL: 0/125/Pass*
*Note: The numerical result (e.g., 0/125) likely indicates 0 failures out of 125 samples tested, meeting the AQL of 2.5 (meaning acceptance on a 125 sample with 0 defects).
ASTM D6124Powder ContentMeet the requirements of ASTM D6124

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.