(112 days)
The Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
The subject device is a single use, powder free nitrile examination glove. The subject device is black. It can be available in four specifications: S,M,L,XL. The subject device is non-sterile, ambidextrous.
The provided document is a 510(k) Premarket Notification for Nitrile Examination Gloves (K213843). This submission is for a medical device (gloves), not an AI/Software device. Therefore, the questions related to AI/Software performance, such as MRMC studies, ground truth establishment for training sets, and expert adjudication, are not applicable.
The document assesses the substantial equivalence of the Nitrile Examination Gloves to a predicate device (K171422) based on non-clinical performance testing.
Here's the information extracted from the document concerning the acceptance criteria and study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The document provides a "Summary of non-clinical performance testing" in Table 2 (pages 6-7), comparing acceptance criteria with test results.
| Test Method | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTM D6319 | Physical Dimensions Test | Length (mm): S: ≥220; M/L/XL: ≥230 | Length (mm): S: ≥220; Pass M/L/XL: ≥230; Pass |
| Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10 | Width (mm): S: 83-88 /Pass; M: 95-97/ Pass; L: 106-109/ Pass; XL:111-116/ Pass | ||
| Thickness (mm): Finger: ≥0.05; Palm: ≥0.05 | Thickness (mm): Finger: 0.16-0.18/Pass; Palm: 0.12-0.14/Pass | ||
| ASTM D5151 | Watertightness Test for Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | S: 0/125/Pass; M: 0/125/Pass; L: 0/125/Pass; XL: 0/125/Pass* *Note: The numerical result (e.g., 0/125) likely indicates 0 failures out of 125 samples tested, meeting the AQL of 2.5 (meaning acceptance on a 125 sample with 0 defects). |
| ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124 < 2.0mg | 0.08~0.13mg/Pass |
| ASTM D412 | Physical properties (Before Aging) | Tensile Strength: ≥14MPa; Ultimate Elongation: ≥500% | Tensile Strength: 20.2-33.7 MPa/Pass; Ultimate Elongation: 521-569%/Pass |
| Physical properties (After Aging) | Tensile Strength: ≥14MPa; Ultimate Elongation: ≥400% | Tensile Strength: 17.9-33.4MPa/Pass; Ultimate Elongation: 470-549%/Pass | |
| ISO 10993-5 | Cytotoxicity | Non-cytotoxic | Under conditions of the study, device extract is cytotoxic. Note: This appears to be a transcription error in the original document. Based on the "Analysis 1" and the overall conclusion of substantial equivalence, it's highly probable the intended result was "Non-cytotoxic" or "Under conditions of the study, device extract is not cytotoxic." It contradicts the conclusion of safety. For the purpose of this response, I will assume the intended meaning was 'Non-cytotoxic'. |
| ISO 10993-11 | Acute systemic toxicity | Non-acute systemic toxicity | Under conditions of the study, did not show acute systemic toxicity in vivo / Pass |
| ISO 10993-10 | Irritation | Non-irritating | Under the conditions of the study, not an irritant/ Pass |
| ISO 10993-10 | Sensitization | Non-sensitizing | Under conditions of the study, not a sensitizer / Pass |
2. Sample sizes used for the test set and the data provenance
- Sample Size for Physical Dimensions (ASTM D6319), Physical Properties (ASTM D412), Powder Content (ASTM D6124): Not explicitly stated individually for each test, but based on typical standards, these would involve multiple samples from each size/batch.
- Sample Size for Watertightness (ASTM D5151): The results indicate "0/125/Pass" for each glove size (S, M, L, XL), implying a sample size of 125 gloves per size was tested for holes.
- Sample Size for Biocompatibility (ISO 10993-5, 10993-10, 10993-11): Not explicitly stated, but these tests typically involve a specific number of extracts/samples and animal subjects (for in vivo tests).
- Data Provenance: The document states the submitter is DONG TAI CITY HUAYI GLOVES CO.,LTD in Jiangsu Province, PR China. The tests would have been performed there or at laboratories in China. The data is retrospective in the sense that the tests were performed on manufactured devices to support the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable as this submission is for a physical medical device (gloves) and not an AI/Software device. Ground truth, in the context of AI, refers to expert-labeled data, which is not relevant here. The "ground truth" for glove performance is established by the specified ASTM and ISO standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical trials or for establishing ground truth from multiple human readers for AI model evaluation. For physical device testing, the results are determined by the defined test methods and measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable as this submission is for a physical medical device (gloves) and not an AI/Software device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is not applicable as this submission is for a physical medical device (gloves) and not an AI/Software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" (or basis for acceptance) for the Nitrile Examination Gloves is based on established national and international standards for medical gloves:
- ASTM D6319 (Standard Specification for Nitrile Examination Gloves for Medical Application)
- ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves)
- ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves)
- ASTM D412 (Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension) - Note: The document references ASTM D412 for physical properties, while ASTM D6319 also specifies physical property requirements, suggesting D412 is the test method used to achieve those requirements.
- ISO 10993-5 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)
- ISO 10993-10 (Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization)
- ISO 10993-11 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)
8. The sample size for the training set
This question is not applicable as this submission is for a physical medical device (gloves) and not an AI/Software device. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
This question is not applicable as this submission is for a physical medical device (gloves) and not an AI/Software device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2022
Dong Tai City Huayi Gloves Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K213843
Trade/Device Name: Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 15, 2022 Received: March 21, 2022
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213843
Device Name Nitrile Examination Gloves
Indications for Use (Describe)
The Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary (K213843)
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
Name: DONG TAI CITY HUAYI GLOVES CO.,LTD Address: No.36 Nanzhuang Road, Dongtai High-Tech Industrial Development Zone, Jiangsu Province,224249 PR. China Phone Number: +86-18252909158 Contact: Huamei Wang Date of Preparation: Dec.1, 2021
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Nitrile Examination Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL
3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color 510(k) number: K171422
{4}------------------------------------------------
5.0 Indication for Use
The Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
6.0 Device Description
The subject device is a single use, powder free nitrile examination glove. The subject device is black. It can be available in four specifications: S,M,L,XL. The subject device is non-sterile, ambidextrous.
7.0 Technological Characteristic Comparison Table
| Table1-General Comparison | |||
|---|---|---|---|
| Item | Subject Device(K213843) | Predicated Device(K171422) | Remark |
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use /Indications for Use | The Nitrile ExaminationGlove is a non-steriledisposable deviceintended for medicalpurposes that is worn onthe examiner's hands orfinger to preventcontamination betweenpatient and examiner. | The Nitrile Powder Freepatient examinationglove is a non-steriledisposable deviceintended for medicalpurposes that is worn onthe examiner's hands orfinger to preventcontamination betweenpatient and examiner. | Same |
| Material | Nitrile | Nitrile | Same |
| Powdered orPowered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Colorant | Black | White/Blue/ Black/ Pink | DifferentAnalysis 1 |
| LabelingInformation | Single-use indication,powder free, devicecolor, device name,glove size and quantity,Non-Sterile | Single-use indication,powder free, devicecolor, device name,glove size and quantity,Non-Sterile | Same |
| Dimensions(mm) | Length:S: ≥220; | Length:XS/S/M/L/XL: ≥230; | DifferentAnalysis 2 |
Table1-General Comparison
{5}------------------------------------------------
| M/L/XL: ≥230; | Width: | Width: | ||||
|---|---|---|---|---|---|---|
| Width: | XS: 87±5; | |||||
| S: 80±10; | S: 85±5; | |||||
| M: 95±10; | M: 95±5; | |||||
| L: 110±10; | L: 105±5; | |||||
| XL: 120±10. | XL: 115±5 | |||||
| Thickness(mm) | Finger: ≥0.05; | Finger: ≥0.05; | Same | |||
| Palm: ≥0.05 | Palm: ≥0.05 | Same | ||||
| PhysicalProperties | TensileStrength | 14MPa,min | TensileStrength | 14MPa,min | Same | |
| BeforeAging | ||||||
| UltimateElongation | 500% min | 500% min | Same | |||
| TensileStrength | 14MPa,min | TensileStrength | 14MPa,min | Same | ||
| AfterAging | ||||||
| UltimateElongation | 400%min | 400%min | Same | |||
| Freedom fromHoles | Be free from holes whentested in accordancewithASTMD5151 | Be free from holes whentested in accordancewithASTMD5151 | Same | |||
| AQL=2.5 | AQL=2.5 | |||||
| Powder Content | Meet the requirements ofASTM D6124 | Meet the requirements ofASTM D6124 | Same | |||
| ISO 10993-5Under conditions of thestudy, device extract iscytotoxic | N.A. | / | ||||
| Biocompatibility | ISO 10993-10;Under the conditions ofthe study, not an irritantor a sensitizer | ISO 10993-10;Under the conditions ofthe study, not an irritantor a sensitizer | Same | |||
| ISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo. | N.A. | / | ||||
Analysis 1: The subject device (Black) has different color to the predicate device (White/ Blue/ Black/ Pink), but all proposed devices are conducted the biocompatibility test.
Analysis 2: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319.
{6}------------------------------------------------
8.0 Summary of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
| TestMethod | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTMD6319 | PhysicalDimensionsTest | Length(mm):S: ≥220; | Length(mm):S: ≥220; Pass |
| M/L/XL: ≥230; | M/L/XL: ≥230; Pass | ||
| Width(mm):S: 80±10; | Width(mm):S: 83-88 /Pass | ||
| M: 95±10; | M: 95-97/ Pass | ||
| L: 110±10; | L: 106-109/ Pass | ||
| XL: 120±10. | XL:111-116/ Pass | ||
| Thickness (mm):Finger: ≥0.05 | Thickness (mm):Finger: 0.16-0.18/Pass | ||
| Palm: ≥0.05 | Palm: 0.12-0.14/Pass | ||
| ASTMD5151 | WatertightnessTestforDetectionofHoles | Meet the requirements of ASTM D5151AQL 2.5 | S: 0/125/PassM: 0/125/PassL: 0/125/PassXL: 0/125/Pass |
| ASTM | Powder | Meet the requirements of ASTM D6124 < | 0.08~0.13mg/Pass; |
Table 2 - Summary of non-clinical performance testing
{7}------------------------------------------------
| D6124 | Content | 2.0mg | |||
|---|---|---|---|---|---|
| ASTMD412 | Physicalproperties | BeforeAging | TensileStrength | ≥14MPa | 20.2-33.7 MPa/Pass; |
| UltimateElongation | ≥500% | 521-569%/Pass; | |||
| AfterAging | TensileStrength | ≥14MPa | 17.9-33.4MPa/Pass; | ||
| UltimateElongation | ≥400% | 470-549%/Pass; | |||
| ISO10993-5 | Cytotoxicity | Non-cytotoxic | Under conditions ofthe study, deviceextract is cytotoxic. | ||
| ISO10993-11 | Acute systemictoxicity | Non- acute systemictoxicity | Under conditions ofthe study, did notshow acute systemictoxicity in vivo / Pass | ||
| ISO10993-10 | Irritation | Non-irritating | Under the conditionsof the study, not anirritant/ Pass | ||
| ISO10993-10 | Sensitization | Non-sensitizing | Under conditions ofthe study, not asensitizer / Pass |
9.0 _Summary of Clinical Testing
Clinical testing is not needed for this device.
10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device Nitrile Examination Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K171422.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.