K Number
K213802
Date Cleared
2022-08-26

(263 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

StimOn™ Pain Relief System (GM2439) is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain.

Device Description

StimOn™ Pain Relief System (GM2439) is a hand-held stimulator designed to pass electrical signals via surface electrodes through the skin to the underlying nerves. It may be used as a TENS device to aid the blocking of pain signals traveling to the brain.

AI/ML Overview

The document describes the StimOn™ Pain Relief System (GM2439) and its substantial equivalence to predicate devices, but it does not contain specific acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) for an AI/algorithm-based device, nor does it detail a study proving the device meets such criteria in terms of AI performance.

Instead, the document focuses on the regulatory clearance of a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief, primarily through comparison with existing predicate devices and compliance with relevant electrical safety and performance standards. The "performance testing" mentioned refers to these safety and electrical performance standards, not to clinical efficacy measured by AI performance metrics.

Therefore, many of the requested items regarding acceptance criteria and study details for an AI-based device's performance cannot be extracted from this document.

Here's the information that can be extracted, and where the document does not provide the requested information:

1. A table of acceptance criteria and the reported device performance:

The document does not specify quantified acceptance criteria for AI performance (e.g., sensitivity > X%, specificity > Y%). Instead, it discusses the device's technical characteristics and how they compare to predicate devices. The "performance testing" section refers to compliance with electrical safety and performance standards, not AI accuracy.

CharacteristicAcceptance Criteria (Not explicitly stated for AI performance)Reported Device Performance (from "Table 1. Comparison table" and subsequent descriptions)
Intended UseSame as predicate devicesSymptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain.
Product CodeGZJGZJ
Power SupplyNot explicitly stated as acceptance criteria, but compared to predicatesOne 3.7V, 125mAh rechargeable Lithium polymer battery
Treatment Timer MaximumNot explicitly stated as acceptance criteria, but compared to predicates15 minutes
Number of Output ModesNot explicitly stated as acceptance criteria1 (RF Mode)
Number of WaveformsNot explicitly stated as acceptance criteria1
WaveformNot explicitly stated as acceptance criteriaBiphasic (Sine Wave)
RF Stimulation ModeNot explicitly stated as acceptance criteriaTranscutaneous Pulsed Radio-frequency (PRF) mode
RF FrequencyNot explicitly stated as acceptance criteria500 kHz fixed
Burst TimeNot explicitly stated as acceptance criteria25ms
Burst FrequencyNot explicitly stated as acceptance criteria2Hz
Electrode ContactNot explicitly stated as acceptance criteriaHydrogel pad
Load ImpedanceNot explicitly stated as acceptance criteria, but stated as a shutdown threshold1 k Ohm, Over this load, the system will automatically shut off.
Maximum Output VoltageNot explicitly stated as acceptance criteria7.4 V+/-10% (6.6 V +/- 20% @@500Ω with Electrode Pad)
Maximum Output CurrentNot explicitly stated as acceptance criteria14.8 mA (Peak)@500Ω (13.2 mA (Peak)@500Ω with Electrode Pad)
Maximum Average Current (Over primary phase)Not explicitly stated as acceptance criteria9.43 mA (Peak)@500Ω (8.41mA(Peak)@500Ω with Electrode Pad)
Maximum Average Power Density (Smallest electrode area) “P”Must be within value range of predicates112.81 µW/mm² @500Ohm with Electrode
Maximum Current (RMS) Density (mA/mm²)Not explicitly stated as acceptance criteria0.0242 mA /mm² @500Ω with Electrode
Net ChargeNot explicitly stated as acceptance criteriaThe theoretical value is 0.
Maximum average phase chargeNot explicitly stated as acceptance criteria0.00942μC
Electrical Safety & PerformanceCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ANSI/AAMI NS4Device underwent testing and demonstrated compliance (implied by substantial equivalence).
BiocompatibilityMeet standards for electrodes leveraged from K180865Leveraged from cleared electrodes (K180865).
UsabilityCompliance with IEC 62366: Medical devices Part 1: Application of usability engineering to medical devicesDevice complies with IEC 62366 and is similar to K180943 (BioWaveGO) cleared for OTC use.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable / Not provided. The document states "No Human Clinical Performance Testing have been required" for this application. The performance testing refers to electrical and safety standards, not clinical performance data from a test set of patients/cases.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. No clinical test set requiring expert ground truth was used for assessing the device's efficacy, as per the submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided. No clinical test set requiring adjudication was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable / Not provided. This device is a Transcutaneous Electrical Nerve Stimulator, not an AI-assisted diagnostic or interpretative device that augments human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable / Not provided. This device is hardware for pain relief, not an algorithm/AI system in the sense of a diagnostic or predictive tool.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable / Not provided. No clinical ground truth was established for "performance" as it relates to AI or clinical efficacy. The "performance testing" was for electrical safety, electromagnetic compatibility, and biocompatibility, which rely on technical standards and laboratory measurements.

8. The sample size for the training set

  • Not applicable / Not provided. Not an AI/algorithm-based device requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. Not an AI/algorithm-based device requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

August 26, 2022

Gimer Medical Co., Ltd. % Anita Chen RA ZhengCheng Limited Company 238 No.19. 335 Lane, Fu-Xi Road, Shulin District New Taipei City, Taiwan 238 Taiwan

Re: K213802

Trade/Device Name: StimOn™ Pain Relief System (GM2439) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: July 5, 2022 Received: July 25, 2022

Dear Anita Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213802

Device Name StimOn™ Pain Relief System (GM2439)

Indications for Use (Describe)

StimOn™ Pain Relief System (GM2439) is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92.

The assigned 510(k) Number: K213802

1.ManufacturerGimer Medical Co., Ltd.
Mail AddressAddress: 9F.-5, No. 97, Sec. 1, Xintai 5th Road,Xizhi District, New Taipei City, TaiwanTel. +886-2-2697-2680Fax. +886-2-2697-2670Website: http://www.gimermed.com/
Manufacturer addressAddress: 9F.-7 and 9F.-8, No. 97, Sec. 1, Xintai 5thRoad, Xizhi District, New Taipei City, Taiwan
Contact PersonMrs. Anita Chen/ Regulatory Advisor of GimerMedical Co., Ltd.
Phone:+886(0) 939-855-759
E-mail:Anita9104303@gmail.com
Date Prepared8/26/22
2Device Name
Trade Name:StimOn™ Pain Relief System (GM2439)
Common or usual nameNerve stimulator
Product CodeGZJ
Regulation NameTranscutaneous electrical nerve stimulator for painrelief
CFR ClassificationCFR Part 882.5890
Device ClassII
Classification PanelNeurology
3Primary Predicate Device
510(k) number:K081835
Trade or proprietary orBioinduction Acticare HFT/HFI

{4}------------------------------------------------

model name:

4

Manufacturer:Bioinduction Limited(178-180 Hotwell Road, BRISTOL, BS8 4RP,United Kingdom, Telephone number: +44 117 375275, Fax number: +44 117 377 5405)
Secondary Predicate Device
510(k) number:K161091
Trade or proprietary ormodel name:STIMPOD NMS460 Nerve Stimulator
Manufacturer:XAVANT TECHNOLOGY (PTY) LTDStreet Address: Unit 102 The Tannery IndustrialPark309 Derdepoort RoadSilvertonCity: PretoriaState/Province: GautengCountry: South Africa
Reference Device ofElectrode Pad (Accessary)
510(k) number:K180865
Trade or proprietary or modelname:ZMI Self-Adhesive Electrodes
Manufacturer:ZMI Electronics, Ltd.6F-1,286-4,Shin Ya RoadKaohsiung, TW 806
Reference Device for OTCusability
510(k) number:K180943
Trade or proprietary or modelname:BioWaveGO
Manufacturer:Biowave Corporation8 Knight Street, Suite 201Norwalk, CT 06851

{5}------------------------------------------------

5Device Description:StimOn™ Pain Relief System (GM2439) is a hand-held stimulator designed to pass electrical signals via surface electrodes through the skin to the underlying nerves. It may be used as a TENS device to aid the blocking of pain signals traveling to the brain.
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{6}------------------------------------------------

    1. Indications for Use: StimOn™ Pain Relief System (GM2439) is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain. 7. Technological A comparison of the device features, intended use, and
    • Characteristics and Substantial Equivalence Comparison with Predicates:

other information demonstrates that the Product name is substantially equivalent to the predicate device as summarized in Table 1. The differences raise no new question of safety and effectiveness.

Table 1. Comparison table

DescriptionSubject DevicePredicate DevicePredicate Device
510(k) NumberK213802K081835K161091
ManufactureGimer MedicalCo., LtdBioinduction LimitedXAVANTTECHNOLOGY(PTY) LTD
Product codeGZJGZJ, IPFGZJ
Device NameStimOn™Pain Relief System(GM2439)BioinductionActicareHFT/HFISTIMPODNMS460NerveStimulator
Indications for useStimOn™ PainRelief System(GM2439) is usedfor the symptomaticrelief andmanagement ofchronic intractablepain and/or as anadjunctive treatmentin the managementof post-surgical orAs a TranscutaneousElectronic NerveStimulation (TENS)device for: thesymptomatic relief ofchronic intractablepain, and as anadjunctive treatmentThe STIMPODNMS460 NerveStimulator is aTranscutaneousElectrical NerveStimulation(TENS)device used forsymptomatic relief
post-traumatic painin the management ofand management of
post-surgical orchronic intractable
post-traumatic pain.pain and/or as an
As a Neuromuscularadjunctive
Electrostimulationtreatment in the
(NMS) device for:management of
the relaxation ofpost-surgical pain,
musclepost traumatic acute
spasms, prevention orpain problems, as
retardation of disusewell as an adjunct
atrophy, increasinglocal bloodfor pain control dueto rehabilitation.
circulation, muscle
re-education,
immediate post-
surgical stimulation
of the calf muscles to
prevent venous
thrombosis and to
maintain or increase
the range of motion.
Prescription/over-the counter useOver-the-counterusePrescriptionPrescription
Power SupplyOne 3.7V, 125mAhrechargeableLithium polymerbatteryNot publiclyavailable4 x AAA alkalinebatteries
Treatment TimerMaximum15 minutesNot publiclyavailable99 minutes

{7}------------------------------------------------

{8}------------------------------------------------

Number of OutputModes1 (RF Mode)Not publicly available1
Number ofWaveforms1Not publicly available2 (Monophasic,Biphasic with offset)
WaveformBiphasic(SineWave)Not publicly availableMonophasic squarewave/Hybrid RFwave
RFStimulationModeTranscutaneousPulsed Radio-frequency(PRF)modeNot publicly availableHybrid RFwaveform whichconsists of aMonophasic SquareWave with asuperimposed RadioFrequency waveform
RF Frequency500 kHz fixedNot publicly available160kHz fixed
Burst Time25msNot publicly available100μs, 200μs
Burst Frequency2HzNot publicly available1,2,5,10 Hz
Electrode ContactHydrogel padNot publicly availableProbe electrode
Load Impedance1 k Ohm, Over thisload, the systemwill automaticallyshut off.Not publicly available7 K Ohm

{9}------------------------------------------------

Maximum OutputVoltage7.4 V+/-10%6.6 V +/- 20% @@500Ω withElectrode PadNot publiclyavailable220 V
Maximum OutputCurrent14.8 mA(Peak)@500Ω13.2 mA(Peak)@500Ω withElectrode PadNot publiclyavailable43.7mA @500Ohm
MaximumAverage Current(Over primaryphase)9.43 mA (Peak)@500Ω8.41mA(Peak)@500Ω withElectrodeNot publiclyavailable29.85mA@500Ohm
MaximumAverage PowerDensity (Smallestelectrode area) “P”112.81 µW/mm²@500Ohm withElectrodeNotpubliclyavailable71.5µW/mm²@500Ohm
MaximumCurrent (RMS)Density(mA/mm²)0.0242 mA /mm²@500Ωwith ElectrodeNot publiclyavailable93.28µA/mm²@500Ohm
Net ChargeThe theoretical valueis 0.Not publiclyavailable6.06μC
Maximumaveragephasecharge [µC]0.00942μCNotpubliclyavailable6.06µC

{10}------------------------------------------------

Summary of the technological characteristics of subject and predicate devices:

The StimOn™ Pain Relief System (GM2439) and Bioinduction Acticare are the same in use for body contact with the hydrogel pad. The other predicate STMPOD NMS460 is using the probe electrode instead. The subject and predicate devices have available RF stimulation modes. The slight differences between these three devices are the burst time. burst frequency, and RF frequency and the differences between the three devices are treatment timer maximum, power source, load impedance and maximum output.

The StimOn™ Pain Relief System (G2439) only has a single stimulation treatment mode at 500KHz, and impedance ranges from 100 to 1000 ohms to simplify the operation of device. Comparison of RF Stimulating Options (RF mode) of StimOn™ Pain Relief System (GM2439), and two predicates of Bioinduction Acticare and STMPOD NMS460 in waveform, RF Frequency, Burst time, Burst Frequency in Table1.

It indicated that from a safety perspective, the StimOn™ Pain Relief System (GM2439) its maximum average power density is within the value range of the two predicates (above STIMPOD and below Acticare). Thus, in the RF mode of energy output mechanism, it could infer that the treatment efficacy of StimOn™ Pain Relief System (GM2439) is sufficient to compare with the two predicate device and the differences between the subject and predicate(s) do not impact safety and effectiveness.

8. Performance Testing

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device.

  • IEC 60601-1. Medical Electrical Equipment Part 1: General Requirements for Safety,
  • •IEC 60601-1-2: Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
  • •IEC 60601-1-11: Medical Electrical Equipment Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
    • · IEC 60601-2-10: Medical electrical equipment Part 2: Particular requirements for the

{11}------------------------------------------------

basic safety and essential performance of nerve and muscle stimulators

· ASI/AAMI NS4: Medical Electrical Equipment Transcutaneous electrical nerve stimulators

Biocompatibility and Performance of electrodes:

The subject device utilizes electrodes cleared under K180865. Performance data (e.g. impedance, current dispersion, adhesion and electrode stability) and biocompatibility testing for the electrodes was leveraged from K180865.

Animal testing

No animal studies have been required for these devices.

Clinical testing

No Human Clinical Performance Testing have been required for StimOn™ Pain Relief System (GM2439) device for this application.

Usabilitv

The StimOn™ Pain Relief System (GM2439) is similar in intended use and technical specifications to the BioWaveGO (K180943), which was cleared for over-the-counter (OTC) use. The subject device also complies with IEC 62366: Medical devices Part 1: Application of usability engineering to medical devices.

9. Conclusion

StimOn™ Pain Relief System (GM2439) has the same intended use as the predicate device. Although there is a slightly different technological design, as compared to the predicate, the conclusions drawn from the test data demonstrate that the device is as safe and as effective as the legally marketed device identified in the submission. Therefore, the subject device is "Substantially Equivalent" to the predicate devices.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).