K Number
K213716
Device Name
InnovaBurn
Date Cleared
2022-09-29

(309 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

InnovaBurn™ is indicated for the management of wounds including; partial- and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled / undermined wounds (donor sites/ grafts, post Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, and skin tears), partial-thickness second degree burns, and draining wounds.

The device is intended for one-time use.

Device Description

InnovaBurn™ is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound covering. InnovaBurn" is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The wound dressing is provided in sheets that are approximately 40-100 microns thick in sizes greater than 5 x 5cm (25cm2) up to 20 x 20cm (400cm2). They are provided as single-use, sterile wound coverings.

AI/ML Overview

The provided document is a 510(k) summary for the InnovaBurn™ device and does not contain detailed acceptance criteria or a study proving that the device meets specific performance criteria through a clinical trial or a formal statistical study with a specific sample size. Instead, it relies on substantial equivalence to a predicate device.

Here's a breakdown of the information available based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria or performance metrics for InnovaBurn™ that would typically be found in a clinical study report (e.g., sensitivity, specificity, accuracy).

Instead, it relies on the following:

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility: Acceptable results for various tests.All test results were acceptable.
Clinical Immunogenicity: Acceptable results for various tests.All results were acceptable.
Viral Inactivation: Acceptable results.Results were acceptable.
Compositional Testing: Acceptable results for various components.All results were acceptable.
Heavy Metal Residual Testing: Acceptable results per FDA guidance.All results were acceptable.
Other Tests (Implicit for a medical device):Biological Risk Assessment, Toxicological Evaluation, Transportation Testing, Packaging Stability, Product Stability, Sterilization Validation Testing, Endotoxin, Physician Usability were conducted. (Specific "performance" results are not detailed for these beyond their completion).
Substantial Equivalence: To predicate device.Achieved substantial equivalence based on indications for use, technological characteristics, and performance test results.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable. The document describes a bench testing and substantial equivalence comparison rather than a clinical study with a "test set" of patient data. Performance data from the predicate device (InnovaMatrix® K193552) was leveraged.
  • Data Provenance: Not applicable in the context of a clinical test set. The data presented is from laboratory testing and analytical comparisons, primarily performed on the predicate device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not applicable. This information is relevant for studies involving human interpretation or clinical data for ground truth establishment. This submission relies on analytical and bench testing performed in a laboratory setting.

4. Adjudication Method for the Test Set

Not applicable. No clinical test set with human adjudication is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document focuses on regulatory approval based on substantial equivalence, not on comparing human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. InnovaBurn™ is a physical medical device (collagen wound dressing), not an AI algorithm. Therefore, "standalone algorithm performance" is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for the performance claims is derived from analytical testing results against established biocompatibility, safety, and compositional standards, and the documented characteristics and performance of the predicate device. For instance, biocompatibility involves comparing test results to predetermined acceptable limits.

8. The Sample Size for the Training Set

Not applicable. InnovaBurn™ is a physical medical device, not an AI model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 29, 2022

Triad Life Sciences, Inc. Donna Best Chief Operating Officer 1770 Moriah Woods Blvd. Suite 18 Memphis. Tennessee 38117

Re: K213716 Trade/Device Name: InnovaBurn™ Regulatory Class: Unclassified Product Code: KGN Dated: December 16, 2021 Received: December 17, 2021

Dear Donna Best:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

K213716 - Donna Best

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213716

Device Name InnovaBurn™

Indications for Use (Describe)

InnovaBurn™ is indicated for the management of wounds including; partial- and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled / undermined wounds (donor sites/ grafts, post Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, and skin tears), partial-thickness second degree burns, and draining wounds.

The device is intended for one-time use.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summarv – K213716

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary is provided below.

1. SUBMITTER

Triad Life Sciences, Inc. 1770 Moriah Woods Blvd., Suite 18 Memphis, TN 38117 Registration Number: 3017660750

Contact Person: Donna Best Phone: 901-333-6000 Email: dbest(@triadls.com

2. DEVICE

Name of Device: InnovaBurn™ Common Name: Collagen Wound Dressing Classification Regulation/Class: Unclassified Product Code: KGN Panel: General and Plastic Surgery

3. PREDICATE AND REFERENCE DEVICE

Predicate Device: InnovaMatrix® (K193552) Reference Device: Cytal® Wound Matrix 3-Layer (K192725)

4. DEVICE DESCRIPTION

Description

InnovaBurn™ is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound covering. InnovaBurn" is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The wound dressing is provided in sheets that are approximately 40-100 microns thick in sizes greater than 5 x 5cm (25cm2) up to 20 x 20cm (400cm2). They are provided as single-use, sterile wound coverings.

{4}------------------------------------------------

510(k) Summarv – K213716

5. INDICATION FOR USE

InnovaBurn™ is indicated for the management of wounds including: partial- and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled / undermined wounds, surgical wounds (donor sites/grafts, post Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, and skin tears), partialthickness second-degree burns and draining wounds. The device is intended for one-time use.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

InnovaBurn™ is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound coverings. InnovaBurn" is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The biodegradable wound matrix provides a protective cover to the wound.

The only modification made to the device since its previous clearance in K193552 is the addition of sizes.

Table 1 provides a summary comparison between the subject, predicate, and reference devices.

Subject DevicePredicate DeviceReference Device
510(k) NumberK213716K193552K192725
ApplicantTriad Life Sciences, Inc.Triad Life Sciences, Inc.ACell, Inc.
Device NameInnovaBurn™InnovaMatrix®Cytal® Wound Matrix3-Layer
Regulation/ClassUnclassified Collagen WoundDressingUnclassified CollagenWound DressingUnclassified CollagenWound Dressing
Product CodeKGNKGNKGN
Prescription orOTC?PrescriptionPrescriptionPrescription
Indications for UseInnovaBurn™ is indicated forthe management of woundsincluding: partial- and full-thickness wounds, pressureulcers, venous ulcers, diabeticulcers, chronic vascular ulcers,tunneled/undermined wounds,surgical wounds (donorsites/grafts, post-Mohs surgery,post-laser surgery, podiatric,wound dehiscence), traumawounds (abrasions, lacerations,and skin tears) partial thicknesssecond-degree burns anddraining wounds.Same, except indicationsincluded second-degreeburns and not partialthickness second-degreeburns.Same, exceptindications includedsecond-degree burnsand not partialthickness second-degree burns.
DesignSingle layer extracellularmatrix, Porcine PlacentaSameMulti-Layerextracellular matrix,Porcine UrinaryBladder Matrix(UBM)

Table 1: Device Comparison Table

{5}------------------------------------------------

510(k) Summarv – K213716

Subject DevicePredicate DeviceReference Device
SizesGreater than 5 x 5 cm (25 cm2) to20 x 20 cm (400 cm2)1 x 1 cm (1 cm2) to5 x 5 cm (25 cm2)16 cm x 25 cm (400 cm2) &16 cm x 35 cm (560 cm2)
Single Use?YesSameSame
SterileE-Beam-sterilizedSameSame

Overall, the difference in technological characteristics of the subject and predicate device, i.e., the addition of sizes, does not raise any questions of safety and effectiveness.

7. PERFORMANCE DATA

The subject device, InnovaBurn™, is comprised of identical materials and is processed and sterilized using identical methods as the predicate device, InnovaMatrix® (K193552).

Biocompatibility testing. including In Vitro Cvtotoxicity. Skin Sensitization. Intracutaneous Reactivity, Acute Systemic Toxicity, Subacute Toxicity, Implantation (1-week, 2-week, and 13-week), In Vitro Bacterial Reverse Mutation, Mouse Lymphoma Assay, Sub-Chronic Systemic Toxicity, and Material-Mediated Pyrogenicity was conducted on InnovaMatrix® (K193552). All test results were acceptable. Clinical immunogenicity testing, including Human Repeat Insult Patch Testing and Human Skin Prick Testing, was conducted on InnovaMatrix® (K193552). All results were acceptable. Viral inactivation testing was performed on InnovaMatrix® (K193552), and results were acceptable. Compositional testing, including collagen analysis, elastin, hyaluronic acid, laminin, fibronectin, nucleic acid, and sulfated glycosaminoglycan analyses, and cell debris testing was performed on InnovaMatrix® (K193552). All results were acceptable. Heavy metal residual testing per FDA Q3D(R1) Elemental Impurities Guidance for Industry (March 2020), including Class 1 Elements; Class 2A Elements; and Applicable Class 3 Elements [Li (lithium), Sb (antimony), and Cu (copper)] was conducted on InnovaMatrix® (K193552), and all results were acceptable. Due to the identical nature of the device composition, the biocompatibility testing, clinical immunogenicity testing, viral inactivation testing, compositional testing, and heavy metals residual testing completed on InnovaMatrix® (K193552) were leveraged to support the subject device InnovaBurn™.

The following testing was conducted in support of InnovaBurn™:

  • Biological Risk Assessment and Toxicological Evaluation ●
  • Transportation Testing ●
  • Packaging Stability ●
  • Product Stability
  • Sterilization Validation Testing ●
  • Endotoxin ●
  • Physician Usability .

{6}------------------------------------------------

8. CONCLUSIONS

InnovaBurn™ has the identical indications for use as the predicate device InnovaMatrix®. The technical characteristics are the same as the technological characteristics of the predicate wound dressing. The only modification for the subject device is the addition of larger sizes to the predicate device. The sizes of the subject device are withing the size ranges of the reference device, Cytal® Wound Matrix 3-Layer. The reference device has the same indications for use as the subject device. The predicate, reference, and subject devices are porcine-derived, decellularized dressings that are comprised primarily of collagen and are intended for the management of wounds. Based on the indications for use, technological characteristics, and performance test results, InnovaBurn" is substantially equivalent to the predicate device InnovaMatrix® (K193552).

N/A