K Number
K213688
Date Cleared
2022-02-18

(87 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Nitrile Examination Gloves are made from nitrile butadiene rubber that covers the hand up to the wrist. They are cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. The device is available in four sizes, which are S, M, L, XL. It could be selected by the user depending on the size of hand. All sizes share the same blue color. The gloves are non-sterile, powder free and are for single use only.

AI/ML Overview

The provided document is a 510(k) summary for Nitrile Examination Gloves (Model: ZMG1351). It details the device's characteristics, comparison to a predicate device, and results from non-clinical testing. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
Dimension - ASTM D6319-19Evaluate the glove palm widthS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmS: 8486mmM: 9698mmL: 104106mmXL: 113115mm
Evaluate the glove lengthS: Min 220mmM: Min 230mmL: Min 230mmXL: Min 230mmS: 241251mmM: 236249mmL: 237245mmXL: 240249mm
Evaluate the glove finger thicknessS, M, L, XL: Min 0.05mmS: 0.1110.137mmM: 0.1240.150mmL: 0.1120.133mmXL: 0.1170.137mm
Evaluate the glove palm thicknessS, M, L, XL: Min 0.05mmS: 0.0800.097mmM: 0.0870.098mmL: 0.0740.086mmXL: 0.0770.094mm
Physical Properties - ASTM D6319-19Evaluate the glove physical properties before and after agingBefore Aging:- Tensile Strength: Min 14Mpa- Ultimate Elongation: Min 500%After Aging:- Tensile Strength: Min 14Mpa- Ultimate Elongation: Min 400%Before Aging:- Tensile Strength: 19.927.4Mpa- Ultimate Elongation: 509553%After Aging:- Tensile Strength: 20.135.1Mpa- Ultimate Elongation: 454504%
Freedom from Holes - ASTM D6319-19Detect the holes that allow water leakageDo not show droplet, stream, or other type of water leakage (Pass at AQL 2.5, as per comparison table)Performance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
Powder Residue - ASTM D6319-19Evaluate the residue powder≤2mg/glovePerformance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
Skin Irritation - ISO 10993-10:2010Evaluated for the potential to cause skin irritationUnder the condition of the test, not an irritant.Performance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
Sensitization - ISO 10993-10:2010Evaluated for the potential to cause skin sensitizationUnder the condition of the test, not a sensitizer.No-sensitization (Performance reported as meeting the acceptance criteria as per "No- sensitization" in "Test Results" column and confirmation of compliance for all tests)
Acute Systemic Toxicity - ISO 10993-11:2017Evaluated for acute systemic toxicityUnder the condition of the test, not a systemic toxicity.No-acute systemic toxicity (Performance reported as meeting the acceptance criteria as per "No- acute systemic toxicity" in "Test Results" column and confirmation of compliance for all tests)

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify the exact sample sizes used for each test. It simply states that "Bench tests were conducted to verify that the proposed device met all design specifications." The data provenance is not explicitly mentioned, but given the manufacturer (Wuhan Zonsen Medical Products Co.,Ltd) and contactor (Shanghai CV Technology Co., Ltd.) locations, it is highly likely the testing and data originated from China. The study appears to be retrospective in the sense that these are design verification tests performed on manufactured gloves.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable as the document describes physical and biocompatibility testing of medical gloves, not an AI or diagnostic device that requires expert interpretation to establish a ground truth. The "ground truth" for these tests is based on established international standards (ASTM and ISO).

4. Adjudication Method for the Test Set

This information is not applicable for the same reason as above. The tests performed are objective measurements against predefined criteria in established standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, not for basic medical devices like examination gloves.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical product (nitrile examination gloves), not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the acceptance criteria and testing of the Nitrile Examination Gloves is based on established industry standards and protocols:

  • ASTM D6319-19: Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ISO 10993-10: Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
  • ISO 10993-11: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
    The tests measure physical properties and biological responses against scientifically defined and accepted limits.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" as this is not an AI or machine learning device. The tests are for product verification and validation.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

February 18, 2022

Wuhan Zonsen Medical Products Co.,Ltd % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No.19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China

Re: K213688

Trade/Device Name: Nitrile Examination Gloves (Model: ZMG1351) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: November 10, 2021 Received: November 23, 2021

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213688

Device Name

Nitrile Examination Gloves (Model: ZMG1351)

Indications for Use (Describe)

Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information

510(k) Number:K213688
Date:February 3rd, 2022
Type of 510(k) Submission:Traditional 510(k)
Basis for 510(k) Submission:New device
Submitter/Manufacturer:Wuhan Zonsen Medical Products Co.,LtdNo 8 Jinchao Road, Zhucheng Street, Xinzhou District, Wuhan City,Hubei Province, ChinaTel: +86-27-82737771E-mail: Cynthia@zonsenmed.com
Contactor:Doris DongShanghai CV Technology Co., Ltd.Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 ChinaE-mail: doris.d@ceve.org.cnTel: 86 21-31261348 / Fax: 86 21-57712250

2. Device Description

Proprietary Name:Nitrile Examination Gloves
Model:ZMG1351
Size:S, M, L, XL
Classification Name:Non-powdered patient examination glove
Product Code:LZA
Device Class:1
Regulation Number:21 CFR 880.6250
Review Panel:General Hospital
Indications for use:Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.
Device Description:Nitrile Examination Gloves are made from nitrile butadiene rubber that covers the hand up to the wrist. They are cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. The device is available in four sizes, which are S, M, L, XL. It could be selected by the user depending on the size of hand. All sizes share the same blue color. The gloves are non-sterile, powder free and are for single use only.

3. Predicate Device Identification

510(k) Number:K211319
Product Name:Purism Non-Sterile Powder Free Nitrile Examination Gloves
Submitter/Manufacturer:Dezhou Purism Medical Technology Co., Ltd.

{4}------------------------------------------------

4. Technological Characteristics Comparison

ParametersNew DevicePredicate DeviceComparison
510(k) NumberK213688K211319---
510(k) OwnerWuhan Zonsen Medical ProductsCo.,LtdDezhou Purism MedicalTechnology Co., Ltd.---
Device NameNitrile Examination GlovesPurism Non-Sterile Powder FreeNitrile Examination Gloves---
ModelZMG1351/---
Product CodeLZALZASame
Regulation No.21 CFR 880.625021 CFR 880.6250Same
Class11Same
Intended useNitrile Examination Gloves aredisposable devices intended formedical purposes that are wornon the examiner's hand to preventcontamination between patientand examiner.The Purism Non-Sterile PowderFree Nitrile Examination Glovesis disposable devices intended formedical purposes that is worn onthe examiner's hand to preventcontamination between patientand examiner.Same
MaterialNitrile Butadiene Rubber LatexNitrile Butadiene Rubber LatexSame
Environment ofuseOTCOTCSame
DesignDisposable (Single use)Disposable (Single use)Same
Non-sterileNon-sterileSame
Powder-freePowder-freeSame
AmbidextrousAmbidextrousSame
CuffedCuffedSame
SizeS, M, L, XLS, M, L, XLSame
ColorBlueBlueSame
SpecificationsPowder-free Nitrile ExaminationGloves Meet ASTM D6319-19Powder-free Nitrile ExaminationGloves Meet ASTM D6319-19Same
Dimensions- Overall LengthMeets ASTM D6319-19S: Min 220mmM: Min 230mmL: Min 230mmXL: Min 230mmMeets ASTM D6319-19S: Min 230mmM: Min 230mmL: Min 230mmXL: Min 230mmSimilar
Dimensions- Palm WidthMeets ASTM D6319-19S: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmMeets ASTM D6319-19S: 80±10mmM: 95±10mmL: 110±10mmXL: ≥110mm
Dimensions- Finger ThicknessMeets ASTM D6319-19Min 0.05mmMeets ASTM D6319-19Min 0.10±0.02mm
Dimensions- Palm ThicknessMeets ASTM D6319-19Min 0.05mmMeets ASTM D6319-19Min 0.06±0.02mm

{5}------------------------------------------------

Physical PropertiesMeets ASTM D6319-19Before Aging:- Tensile Strength: min 14Mpa- Ultimate Elongation: min 500%Meets ASTM D6319-19After aging:- Tensile Strength: min 14Mpa- Ultimate Elongation: min 400%Meets ASTM D6319-19Before Aging:- Tensile Strength: min 14Mpa- Ultimate Elongation: min 500%Meets ASTM D6319-19After aging:- Tensile Strength: min 14Mpa- Ultimate Elongation: min 400%Same
Freedom fromHolesMeets ASTM D6319-19Pass at AQL 2.5ASTM D5151-06Pass at AQL 2.5Similar
Powder residueMeets ASTM D6319-19Below 2mg/glove of residualpowderMeets ASTM D6124-06Below 2mg/glove of residualpowder
BiocompatibilityMeets ISO 10993-10:2010;Skin Irritation: Under thecondition of the test, not anirritant.Skin Sensitization: Under thecondition of the test, not asensitizer.Meets ISO 10993-11:2017;Acute Systemic Toxicity: Underthe condition of the test, not asystemic toxicity.Meets ISO 10993-10:2010;Skin Irritation: Under thecondition of the test, not anirritant.Skin Sensitization: Under thecondition of the test, not asensitizer.Meets ISO 10993-11:2017;Acute Systemic Toxicity: Underthe condition of the test, not asystemic toxicity.Meets ISO 10993-5:2009;In Vitro Cytotoxicity: Under thecondition of the test, cytotoxicSimilar

5. Non-clinical Testing Summary

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

  • ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.

  • ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

Test MethodPurposeAcceptance CriteriaTest Results
Dimension-ASTM D6319-19Evaluate theglove palm widthSize(mm)(mm)
S$80\pm10$84~86
M$95\pm10$96~98
L$110\pm10$104~106
XL$120\pm10$113~115
Evaluate theSize(mm)(mm)
glove lengthSMin 220241~251
MMin 230236~249
LMin 230237~245
XLMin 230240~249
Evaluate theglove fingerthicknessSize(mm)(mm)
SMin 0.050.111~0.137
MMin 0.050.124~0.150
LMin 0.050.112~0.133
XLMin 0.050.117~0.137
Evaluate theglove palmthicknessSize(mm)(mm)
SMin 0.050.080~0.097
MMin 0.050.087~0.098
LMin 0.050.074~0.086
XLMin 0.050.077~0.094
Physical Properties-ASTM D6319-19Evaluate theglove physicalpropertiesbeforeand after agingBeforeAgingAfter agingBefore AgingAfter aging at
Tensilestrength:Tensilestrength:Tensilestrength:Tensilestrength:
min 14Mpamin 14Mpa19.9~27.4Mpa20.1~35.1Mpa
Ultimateelongation:Ultimateelongation:Ultimateelongation:Ultimateelongation:
min 500%min 400%509~553%454~504%
Freedom fromHoles-ASTM D6319-19Detect the holesthat allow waterleakageDo not show droplet, stream or other type ofwater leakage
Powder residue-ASTM D6319-19Evaluate theresidue powder≤2mg/glove
Skin Irritation -ISO 10993-10:2010Evaluated for thepotential to causeskin irritation
Sensitization -ISO 10993-10:2010Evaluated for thepotential to causeskin sensitizationUnder the condition of thetest, not a sensitizer.No- sensitization
Acute SystemicToxicity-ISO 10993-11:2017Evaluated foracute systemictoxicityUnder the condition of thetest, not a systemictoxicity.No- acute systemic toxicity

{6}------------------------------------------------

6. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission, the Nitrile Examination Gloves, are as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K211319.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.