(27 days)
The Spartan MC 0165™ is intended for the introduction of interventional devices or infusion of diagnostic agents into the neuro, peripheral and coronary vasculatures.
The Spartan MC 0165™ is a sterile single-use microcatheter device. On the proximal end is a Luer connector for infusion of diagnostic agents. The microcatheter has a single lumen of 0.0165" ID, has a flexible and variable stiffness composite body, and 2 radiopaque marker bands on the distal tip for visualization under fluoroscopy. It comes with a sterile stainless steel shaping mandrel to shape the distal portion of the catheter.
The Spartan MC 0165™ is a microcatheter device. The provided text describes several performance studies conducted to demonstrate its substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Coating Lubricity (Friction Force) | Acceptable friction force under simulated use conditions similar to the predicate device. | The Spartan MC 0165™ was found to have acceptable friction force under simulated use conditions similar to the predicate device. |
Hub Functional & Dimensional | Meet ISO 594-1:1986-06-15 and ISO 594-2:1998-09-01 requirements. | The Spartan MC 0165™ met the acceptance criteria for hub functional and dimensional requirements. |
Torque Strength | Acceptable torsional strength similar to the predicate device during simulated use. | The Spartan MC 0165™ exhibited acceptable torsional strength similar to the predicate device. |
Tensile | Meet ISO 10555-1:2013 Annex B requirements. | The Spartan MC 0165™ met the acceptance criteria for tensile strength. |
Air Aspiration | Meet ISO 10555-1:2013(E) Annex D requirements for air leakage. | The Spartan MC 0165™ met the acceptance criteria for air aspiration. |
Liquid Leak | Meet ISO 10555-1:2013(E) Annex C requirements for liquid leakage. | The Spartan MC 0165™ met the acceptance criteria for liquid leakage. |
Particulate and Coating Integrity | Meet acceptance criteria for particulate generation and coating integrity, and be substantially equivalent to the predicate. | The Spartan MC 0165™ met the acceptance criteria for particulate generation and coating integrity, and was found substantially equivalent to the predicate. |
Tip Shape | Retain steam-shaped tip shape and be substantially equivalent to the predicate device. | The Spartan MC 0165™ met the acceptance criteria for tip shapeability and was found substantially equivalent to the predicate. |
Stiffness | Meet acceptance criteria for catheter stiffness and be substantially equivalent to the predicate. | The Spartan MC 0165™ met the acceptance criteria for catheter stiffness and was found substantially equivalent to the predicate. |
Kink Resistance | Withstand bends and be substantially equivalent to the predicate. | The Spartan MC 0165™ met the acceptance criteria for kink resistance and was found substantially equivalent to the predicate. |
Radiopacity | Be visible under fluoroscopy and be substantially equivalent to the predicate device. | The Spartan MC 0165™ met the acceptance criteria for radiopacity and was found substantially equivalent to the predicate. |
Corrosion | Meet ISO 10555-1:2013(E) Annex A requirements. | The Spartan MC 0165™ met the acceptance criteria for corrosion resistance. |
Static Burst Pressure | Meet ISO 10555-1:2013(E) Annex F requirements. | The Spartan MC 0165™ met the acceptance criteria for static burst pressure. |
Torque Response | Meet acceptance criteria for torque response and be substantially equivalent to the predicate. | The Spartan MC 0165™ met the acceptance criteria for torque response and was found substantially equivalent to the predicate. |
Buckling | Meet acceptance criteria for tip deflection and buckling, and be substantially equivalent to the predicate. | The Spartan MC 0165™ met the acceptance criteria for tip deflection and buckling, and was found substantially equivalent to the predicate. |
Design Validation | Meet requirements for which it was designed and tested in a challenging use setting. | The Spartan MC 0165™ met the requirements for which it was designed and tested. |
Dimensional Testing | Meet device specifications. | The Spartan MC 0165™ met the dimensional requirements. |
Visual Inspection | Meet visual inspection requirements. | The Spartan MC 0165™ met the visual inspection requirements. |
Dead Space Volume | Dead space volume to be measured and reported in labeling. | The Spartan MC 0165™ dead space volume was measured and is reported in the labeling. |
Pressure Flow Testing with Contrast Media | Flow rates and pressures for various solutions to be evaluated and reported in labeling. | The Spartan MC 0165™TM pressure-flow performance was evaluated, and flow rates and pressures of various saline and contrast media solutions are reported in the labeling. |
Sensitization (Kligman Maximization Test) | Classified as a non-sensitizer (0% sensitization). | Spartan MC 0165 elicited no reaction at the challenge (0% sensitization), classified as a non-sensitizer. |
Genotoxicity (Mouse Lymphoma Mutagenesis Assay) | Non-mutagenic (IMF of Test Article |
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).