K Number
K213449

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-11-15

(20 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.

MultiBase Abutments EV:
DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures.

Device Description

The proposed DS (Dentsply Sirona) Implants abutments with EV connection are used in conjunction with an endosseous dental implant with EV connection to aid in prosthetic rehabilitation in fully edentulous or partially edentulous maxillary and/or mandibular arches.
The proposed devices in this bundled 510(k), which are included under the same master brand - DS Implants abutments with EV connection, include:
(A) Cover Screw EV, Healing Abutment EV, HealDesign EV, Temp Abutment EV, TiDesign EV, CastDesign EV, Abutment Screw EV, and
(B) MultiBase Abutment EV and its accessory components including the Multibase heal cap. MultiBase temporary cylinder, MultiBase semi-burnout cylinder, and MultiBase bridge screw.

AI/ML Overview

This document is a 510(k) summary for Dentsply Sirona's DS Implants abutments with EV connection. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing new acceptance criteria or conducting studies to prove performance against specific criteria for a novel device. Therefore, much of the requested information about acceptance criteria and a study proving a device meets them is not directly available in this document.

However, based on the provided text, we can infer some aspects related to performance and substantial equivalence:

1. A table of acceptance criteria and the reported device performance

The document does not present explicit acceptance criteria in a table format. Instead, it relies on demonstrating that the proposed devices are substantially equivalent to existing predicate devices, meaning they perform as intended and similarly to the cleared devices. The key performance aspect discussed is fatigue performance.

  • Implied Acceptance Criteria (Fatigue Performance): The proposed DS Implants abutments must exhibit fatigue performance equivalent to or better than the identified worst-case predicate abutment (MultiBase Abutment EV, 30°).
  • Reported Device Performance (Fatigue Performance): The document states: "The fatigue test results obtained with the existing worst-case abutment together with the compatible implants with EV connection... indicate that the proposed DS Implants abutments with EV connection (A) and (B) perform as intended." This implies that the performance meets the established safety and effectiveness levels of the predicate devices.

Other performance aspects implicitly covered by substantial equivalence include biocompatibility and sterilization efficacy, which are assumed to be met if the proposed device's characteristics are similar or validated against existing standards and predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "Therefore, no new fatigue performance data is provided in this submission in support of substantial equivalence." This means there was no new test set for fatigue performance for this submission. The data provenance for the original fatigue testing of the predicate device (MultiBase Abutment EV, 30°) is not specified, nor is whether it was prospective or retrospective.

For biocompatibility, a "review of existing cytotoxicity testing on file for reference device Friadent Aurobase Abutment D3.0 (K030639)" was performed. No sample size for this review is mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided as the submission relies on previously conducted tests and established equivalency, not on a new ground truth assessment by experts for the current device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as there was no new test set involving human assessment or adjudication for this 510(k) submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This aspect is not applicable. The device is an endosseous dental implant abutment, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This aspect is not applicable for the same reason as point 5.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For fatigue performance, the ground truth would be established through mechanical testing compliant with recognized standards (e.g., ISO standards mentioned for sterilization, implying similar rigorous standards for mechanical tests like fatigue). For biocompatibility, it's based on cytotoxicity testing and compliance with ISO 10993-1:2018. The document implicitly relies on these established standards as "ground truth" for safe and effective performance.

8. The sample size for the training set

This information is not applicable as this is a medical device clearance, not an AI/algorithm submission requiring a training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Dentsply Sirona Courtney Clark Senior Director of Regulatory Affairs, Corporate 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401

Re: K213449

Trade/Device Name: DS Implants abutments provided with the EV connection Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 25, 2021 Received: October 26, 2021

Dear Courtney Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213449

Device Name

DS Implants abutments provided with the EV connection

Indications for Use (Describe)

DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213449

Device Name

DS Implants abutments provided with the EV connection

Indications for Use (Describe)

MultiBase Abutments EV:

DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401

Image /page/4/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a leaf or a curved form in gray. To the right of the shape are the words "Dentsply" stacked on top of "Sirona", both in a simple, sans-serif font and in a light gray color.

510(k) SUMMARY DS Implants abutments with EV connection K213449

1. Submitter Information:

Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401

Primary Contact Person: Courtney Clark Telephone Number: 248-895-4379 Fax Number: 717-849-4343

Date Prepared: November 10, 2021

2. Device Name:

  • Proprietary Name: DS Implants abutments provided with the EV connection ●
    • Classification Name: Endosseous dental implant abutment
  • CFR Number: 872.3630 ●
  • Device Class: II
  • Product Code: NHA

3. Predicate Devices:

The proposed devices in this bundled 510(k), which are included under the same master brand - DS (Denstply Sirona) Implants abutments with EV connection, include:

  • (A) Cover Screw EV, Healing Abutment EV, HealDesign EV, Temp Abutment EV, TiDesign EV, CastDesign EV, Abutment Screw EV, and
  • (B) MultiBase Abutment EV and its accessory components including the Multibase heal cap, MultiBase temporary cylinder, MultiBase semi-burnout cylinder, and MultiBase bridge screw.

{5}------------------------------------------------

There are two primary predicate devices identified relating to the substantial equivalence for the DS Implants abutments with EV connection, as referred to above as bundles (A) and (B):

Primary Predicate (A)Device NameFDA ProductCode510(k)Company Name
OsseoSpeed PlusNHAK120414Dentsply Sirona(formerly: Astra Tech AB)
Primary Predicate (B)Device NameFDA ProductCode510(k)Company Name
Multibase Abutments EVand ATLANTISSuprastructuresNHAK163350Dentsply Sirona

NOTE: The OsseoSpeed Plus K120414 clearance included the Astra Tech Implant System Plus abutments, which are currently marketed as Astra Tech Implant EV abutments. Therefore, this section will hereafter refer to the abutments cleared under OsseoSpeed Plus (K120414) as Astra Tech Implant EV abutments and the implants cleared under the same 510(k) as Astra Tech EV Implant.

This Special 510(k) bundled premarket notification includes minor modifications to both the Primary Predicate (A) Astra Tech Implant EV abutments (K120414) and Primary Predicate (B) MultiBase Abutments EV (K163350) devices. All proposed abutments are merged under the same master brand - DS Implants abutments with EV connection. In addition, Primary Predicate (B) MultiBase Abutments EV (K163350) continues to represent the worst-case abutments for fatigue performance testing for all proposed devices.

In addition to the predicate devices listed above, reference device Friadent Aurobase Abutment D3.0 (K030639) is included to demonstrate substantial equivalence in support of biocompatibility.

ReferenceNameDeviceFDA Product Code510(k)Company Name
Xive 3.0Implant systemDentalDZEK030639Dentsply Sirona(formerly: FriadentGmbH)

4. Description of Device:

The proposed DS (Dentsply Sirona) Implants abutments with EV connection are used in conjunction with an endosseous dental implant with EV connection to aid in prosthetic rehabilitation in fully edentulous or partially edentulous maxillary and/or mandibular arches.

The proposed devices in this bundled 510(k), which are included under the same master brand - DS Implants abutments with EV connection, include:

  • (A) Cover Screw EV, Healing Abutment EV, HealDesign EV, Temp Abutment EV, TiDesign EV, CastDesign EV, Abutment Screw EV, and

{6}------------------------------------------------

  • (B) MultiBase Abutment EV and its accessory components including the Multibase heal cap. MultiBase temporary cylinder, MultiBase semi-burnout cylinder, and MultiBase bridge screw.
    These device groupings will be referenced as (A) and (B) throughout this summary.

The proposed devices in this bundled 510(k) are available in the following size ranges.

Table 5.4.1.: Size ranges for proposed DS Implants abutments with EV connection (A) and (B)

ItemDiameterGingiva HeightAngulation
Healing Abutment EV4.0 and 4.3 mm2.0 - 6.0 mm
HealDesign EV4.0 - 7.5 mm2.5 - 6.5 mm
Temp Abutment EV4.0 - 5.0 mm12.0 - 12.3 mm
TiDesign EV4.5 - 7.0 mm1.5 - 4.5 mm0° and 15°
CastDesign EV3.9 and 5.1 mm7.0 - 7.3 mm
MultiBase Abutment EV4.5 - 4.9 mm1.5 - 4.5 mm0°, 17° and 30°

The proposed (A) CastDesign EV abutment has a metal base and a plastic cylinder, which can be modified by wax-up and cast-to technique. The proposed (A) CastDesign EV abutment is a straight abutment with a minimum post height of 4 mm for single-unit loading applications (i.e. crowns). It is not used to fabricate an angled abutment.

The proposed (B) MultiBase EV accessory MultiBase Semi-Burnout Cvlinder has a metal base and plastic cylinder, which can be modified by wax-up and cast-to technique. It is used in the fabrication of a prosthetic restoration in the laboratory. The proposed (B) MultiBase Semi-Burnout Cylinder accessory does not add any additional angular correction than already present in the underlying MultiBase Abutment EV (which allows for angulation of 179 and 30°).

The proposed DS Implants abutments with EV connection (A) and (B), are compatible with

    1. Primary Predicate (A) Astra Tech EV Implant (K120414),
    1. Astra Tech Implant EV Profile, cleared as OsseoSpeed Profile EV Implant (K130999), and
    1. PrimeTaper EV Implant (K210610), which is the new DS Implants brand with EV connection.

The proposed (A) CastDesign EV abutments are not compatible with the Astra Tech Implant EV Profile, cleared as OsseoSpeed Profile EV Implant (K130999).

The purpose of this submission is to expand the assortments by including additional sizes and variants (diameter and gingiva height), modifications to the shape of the abutments in the lower gingival forming area in some models to facilitate removal of healing abutments or to accommodate to different gingiva heights, slightly deeper grooves on temporary abutments to improve retention towards composite/acrylic of the restoration, and material change in burn-out sleeve (part of the lab fabrication process) for the proposed accessories (A) CastDesign EV and (B) MultiBase Semi-burnout Cylinder models. In addition,

{7}------------------------------------------------

to facilitate product identification, color coding via anodization was added to some abutments which previously did not have such identification. All proposed devices (A) and (B) are merged under the same master brand - DS Implants abutments with EV connection.

5. Indications for Use:

The proposed Indications for Use for DS Implants abutments with EV connection (A) and (B) are:

(A):

DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.

(B):

DS Implants abutments provided with the EV connection are intended to be used in conjunction with implants with the EV connection in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures.

The modifications to the Indications for Use for the proposed devices are limited to the brand name change to consolidate the abutments under the new master DS Implants (short for Dentsply Sirona Implants) name and do not alter the intended use of the proposed devices as compared to Primary Predicate (A) Astra Tech Implant EV abutments (K120414) and Primary Predicate (B) Multibase Abutments EV (K163350). The modified Indications for Use state that the abutments are compatible with implants with an EV connection.

  1. Substantial Equivalence:

The design of the proposed DS Implants Abutments with EV connection (A) and (B), is based on the previously cleared abutments in the Primary Predicate (A) Astra Tech Implant System Plus (K120414) and Primary Predicate (B) Multibase Abutment EV (K163350) respectively. For some of the proposed abutments, the design remains unchanged and only the indications were updated, whereas some devices underwent minor design changes or additional variations are introduced to the abutment family.

An overview of the similarities and differences between the proposed and predicate abutment devices is given in Table 5.6.1. and Table 5.6.2 below. A discussion of the similarities and differences follows the tables.

{8}------------------------------------------------

Primary Predicate (A) AbutmentsPrimary Predicate (A)Primary Predicate (B) Abutments
Proposed devices (A)Proposed devices (B)Primary Predicate (B)
Dentsply SironaDS Implants Abutments with EVconnectionDentsply SironaAstra Tech Implant EV AbutmentsK120414Dentsply SironaDS Implants Abutments with EVconnectionDentsply SironaMultibase Abutment EVK163350
Indications for UseDS Implants abutments provided with theEV connection are intended to be used inconjunction with implants with the EVconnection in fully edentulous or partiallyedentulous maxillary and/or mandibulararches to provide support for crowns,bridges or overdentures.Astra Tech Implant System Plus abutmentsare intended to be used in conjunction withAstra Tech Implant System Plus in fullyedentulous or partially edentulousmaxillary and/or mandibular arches toprovide support for crowns, bridges oroverdentures.Indications for UseDS Implants abutments provided with theEV connection are intended to be used inconjunction with implants with the EVconnection in fully edentulous or partiallyedentulous maxillary and/or mandibulararches to provide support for bridges oroverdentures.The Multibase Abutments EV are intendedto be used in conjunction with Astra TechImplant System EV in fully edentulous orpartially edentulous maxillary and/ormandibular arches to provide support forbridges or overdentures.
Connection sizesS, M, L3.0, 3.6, 4.2, 4.8, 5.4Connection sizesS, M, L3.6, 4.2, 4.8
ProsthesisAttachmentCement-retainedScrew-retainedCement-retainedScrew-retainedProsthesisAttachmentScrew-retainedScrew-retained
Abutment angulationTiDesign EV:0°, 15°TiDesign Plus:0°, 15°, 20°AbutmentangulationMultiBase Abutment EV:0°, 17°, 30°Multibase Abutment EV:0°, 17°, 30°
Abutment designOne-piece (0°),Two-piece (0°, 15°)One-piece (0°),Two-piece (0°, 15°, 20°)Abutment designOne-piece (0°),Two-piece (17°, 30°)One-piece (0°),Two-piece (17°, 30°)
Gingiva HeightHealing Abutment EV:2, 3, 4, 6 mmHealDesign EV:2.5, 3.5, 4.5, 6.5 mmTiDesign EV:1.5, 2.5, 3.5, 4.5 mmHealing Uni Plus:2, 3, 4, 6 mmHealDesign Plus:3.5, 4.5 mmTiDesign Plus:1.5, 2.5 mmGingiva HeightMultiBase Abutment EV:1.5, 2.5, 3.5, 4.5 mmMultiBase Abutment EV:1.5, 2.5, 3.5 mm
Materials,AbutmentsScrewBurnout sleeveAbutments:Titanium Alloy or Gold AlloyAbutment screws:Titanium AlloyBurn-out sleeve:PlasticAbutments:Titanium Alloy or Gold AlloyAbutment screws:Titanium AlloyBurn-out sleeve:PlasticMaterials,AbutmentsAbutments:Titanium Alloy or Gold AlloyAbutments:Titanium Alloy
ManufacturingprocessMillingMillingScrewAbutment screws:Titanium AlloyAbutment screws:Titanium Alloy
Surface treatmentAnodization(abutment + abutment screw)Anodization(abutment + abutment screw)Burnout sleeveBurn-out sleeve:Titanium AlloyMultiBase holder:Titanium Alloy
Sterility stateSterile or non-sterileSterile or non-sterileHolderBurn-out sleeve:PlasticMultiBase holder:PlasticMultiBase holder:Plastic
Sterilization methodfor sterile productsE-beam irradiationE-beam irradiationManufacturingprocessMillingMilling
Sterilization for non-sterile productsMoist heat (steam) sterilizationMoist heat (steam) sterilizationSurface treatmentAnodization(abutment + abutment screw)Anodization(abutment screw)
Sterility stateSterile or non-sterileSterile
Sterilization methodfor sterile productsE-beam irradiationE-beam irradiation
Sterilization fornon-sterile productsMoist heat (steam) sterilizationMoist heat (steam) sterilization

Table 5.6.1. Similarities and Differences between the Proposed (A) and

{9}------------------------------------------------

Table 5.6.2. Similarities and Differences between the Proposed (B) and Primary Predicate (R) Abutments

The proposed DS Implants abutments with EV connection (A) and (B) have the same intended use, similar indications for use, same prosthesis attachment method (screw-retained or cement retained), and same manufacturing, packaging, and sterilization processes as the Primary Predicate (A) Astra Tech Implant EV Abutments (K120414) and Primary Predicate (B) Multibase Abutments EV (K163350) respectively. The proposed and primary predicate abutments and accessories (K120414, K163350) are either one-piece or two-piece abutment designs and made of the same materials with the exception of the burn-out sleeve (part of the lab fabrication process) for the proposed accessories (A) CastDesign EV and (B) MultiBase Semi-burnout Cylinder models, which are made of the same materials as cleared in Dentsply Sirona devices (K030639 and K945847, respectively).

{10}------------------------------------------------

For the proposed DS Implants abutments with EV connection (A) and (B), designators S, M and L are used instead of a reference to implant diameter size. Size S has identical interface dimensions as 3.6 connection, size M has identical interface dimensions as 4.2 connection and size L has identical interface dimensions as 4.8 connection. Color coding and size legend are located in the compatible Implants Instructions for Use. The proposed devices have EV connection sizes within the range of the primary predicate devices (A) and (B) and the available sizes have identical interface dimensions as primary predicate devices (A) and (B).

The proposed (A) HealDesign EV, TiDesign EV, and (B) MultiBase Abutments are available in additional gingiva heights, but those changes do not introduce a new worstcase abutment-implant configuration. In addition, the angulation for the proposed devices (A) and (B) is within the range of primary predicate devices (A) K120414 and (B) K163350. Angulation of 30° continues to be the worst-case for both proposed and predicate devices.

7. Non-Clinical Performance Data

FDA document "Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments Class II Special Controls Guidance Document Guidance for Industry and FDA Staff' was used for determining the performance testing required for the proposed DS Implants Abutments with EV connection.

The addition of the proposed abutment variants does not change the previously defined worstcase abutment-implant combination scenario. It was determined that the abutment with the highest angulation (MultiBase Abutment EV, 30°) continues to be the worst-case for fatigue testing, representing the entire assortment of DS Implants with EV connection. Furthermore, the Primary Predicate (B) 30° MultiBase Abutment EV (K163350) and the proposed (B) 30° MultiBase Abutment EV are identical and therefore fatigue testing performed on Primary Predicate (B) remains valid for the worst-case combinations of proposed abutments (A) and (B) and compatible implants. Therefore, no new fatigue performance data is provided in this submission in support of substantial equivalence.

The fatigue test results obtained with the existing worst-case abutment together with the compatible implants with EV connection, PrimeTaper EV (K210610), AstraTech EV Implant (K120414) and OsseoSpeed Profile EV (K130999), indicate that the proposed DS Implants abutments with EV connection (A) and (B) perform as intended.

Biocompatibility evaluation in accordance with EN ISO 10993-1:2018 (Biological evaluation of medical devices – Part 1: Evaluation and testing with a risk management process) was performed. This included a review of existing cytotoxicity testing on file for reference device Friadent Aurobase Abutment D3.0 (K030639). It was determined that no additional testing was required to demonstrate biological safety of proposed devices.

The proposed DS Implants abutments with EV connection (A) and (B) sterile device modifications do not introduce a new worst-case scenario to sterilize and therefore a new electronic beam (E-beam) irradiation sterilization was not performed. The existing

{11}------------------------------------------------

validation for primary predicate devices (A) K120414 and (B) K163350, which meets the requirements according to ISO 11137-1:2006 (Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]) and ISO 11137-2:2013 (Sterilization of health care products - Radiation - Part 2:Establishing the sterilization dose), remains valid.

No new steam heat sterilization validation was required on the proposed DS Implants abutments with EV connection (A) and (B) non-sterile devices as the modifications do not introduce a new worst-case scenario to sterilize. The existing validation for primary predicate devices (A) K120414 and (B) K163350 according to ISO 17665-1:2006 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices) remains valid.

The shelf life of the proposed sterile DS Implants Abutments with EV Connection (A) and (B) is five (5) years. The packaging materials and configuration for the proposed sterile DS Implants Abutments with EV Connection (A) and (B) are the same as that of the primary predicate devices (A) K120414 and (B) K163350. Therefore, the same shelf life is applicable for the proposed sterile devices (A) and (B), including predicate packaging validation which meets the requirements of ISO 11607-1: 2019 (Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems).

    1. Clinical Performance Data
      No human clinical data was included to support substantial equivalence.

9. Conclusion Regarding Substantial Equivalence

The information included in this bundled premarket notification supports the substantial equivalence of the proposed DS Implants abutments with EV connection (A) and (B) with the Primary Predicate (A) AstraTech Implant EV Abutment (K120414) and Primary Predicate (B) MultiBase Abutment EV (K163350). The proposed devices have the same intended use, similar indications for use and performance characteristics as the legally marketed primary predicate devices. As a result of the minor modifications for use and minor modifications in design and materials, no additional performance testing is included in this premarket notification to demonstrate substantial equivalence.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)