K Number
K213211
Date Cleared
2022-01-27

(120 days)

Product Code
Regulation Number
862.3250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Carolina Liquid Chemistries Cocaine Metabolite Enzyme Immunoassay (COCM) Test System is intended for the qualitative determination of benzoylecgonine (cocaine metabolite) in human urine at a cutoff value of 300 ng/mL. The assay is designed for professional use with a Carolina Liquid Chemistries CLC6410 automated clinical chemistry analyzer. For in vitro diagnostic use only. The assay provides a rapid screening procedure for determining the presence of benzoylecgonine in urine. The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Device Description

The Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) Test System is a ready-to-use, liquid reagent homogeneous enzyme immunoassay for qualitatively determining the presence of cocaine metabolite (benzoylecgonine) in human urine. The assay uses specific antibody that can detect benzoylecgonine in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs. The assay is based on competition between benzoylecgonine labeled with the enzyme glucose-6phosphate dehydrogenase (G6PDH) and free drug from the urine sample, for a fixed amount of antibody. In the absence of free drug from the urine sample, the specific antibody binds to the drug labeled with G6PDH causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD+) to NADH.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionSpecifics from Study (Reported Device Performance)
PrecisionWithin Run:- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 4/4 Negative results.- 300 ng/mL (Cutoff): 3/1 Neg/Pos (3 Negative, 1 Positive).- 375 ng/mL, 450 ng/mL, 525 ng/mL, 600 ng/mL: 4/4 Positive results.Run-to-Run:- 0 ng/mL, 75 ng/mL, 150 ng/mL, 225 ng/mL: 88/88 Negative results.- 300 ng/mL (Cutoff): 53/35 Neg/Pos (53 Negative, 35 Positive).- 375 ng/mL, 450 ng/mL, 525 ng/mL: 88/88 Positive results.
SpecificitySupported by cross-reactivity studies that supported the predicate device (K020763). (No specific numerical data provided in this document, but stated as acceptable by reference to predicate)
Interference (pH)No substantial interference noted. - 225 ng/mL Target: All samples from pH 3.1 to 11.1 tested Negative.- 375 ng/mL Target: All samples from pH 3.1 to 11.1 tested Positive.
Interference (Specific Gravity)No substantial interference noted.- 225 ng/mL Target: All samples from SG 1.000 to 1.029 tested Negative.- 375 ng/mL Target: All samples from SG 1.001 to 1.028 tested Positive.
CarryoverNo carryover was noted during testing with alternating high (1000 ng/mL) and low (0 ng/mL) samples.
Method Comparison and AccuracyAgreement among positives: 100% (40/40 samples correctly identified). Agreement among negatives: 100% (41/41 samples correctly identified). No discordant samples identified.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision:
    • Within Run: 4 replicates per concentration.
    • Run-to-Run: 88 replicates per concentration (2 samples/day for 22 non-consecutive days).
  • Interference (pH): 18 real human urine samples (9 pH pools, each split and spiked at two concentrations).
  • Interference (Specific Gravity): 20 real human urine samples (10 SG pools, each split and spiked at two concentrations).
  • Carryover: 21 samples (10 "High" at 1000 ng/mL, 11 "Low" at 0 ng/mL).
  • Method Comparison and Accuracy: 81 samples total.
    • 41 LC/MS Confirmed Negative Samples (20 drug-free, 21 between 0-300 ng/mL, with 8 between 150-300 ng/mL).
    • 40 LC/MS Confirmed Positive Samples (at least 8 between 300-450 ng/mL, remaining >450 ng/mL).

Data Provenance: The document states "real human urine samples" for interference testing, and for precision and accuracy, samples were "drug free urine" or "spiked with benzoylecgonine." The document does not specify the country of origin of the data or whether the data was retrospective or prospective. It implies a controlled laboratory setting (spiking, adjusting pH/SG).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of experts to establish a ground truth. The ground truth for the test set (Method Comparison and Accuracy, Precision) was established through instrumental analysis: LC/MS (Liquid Chromatography/Mass Spectrometry), which is explicitly called the "confirmatory method" and "reference method."

4. Adjudication Method for the Test Set

No adjudication method (e.g., 2+1, 3+1) is mentioned, as the ground truth was established by instrumental methods (LC/MS) rather than human experts requiring consensus.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No MRMC study was done, as this device is an in vitro diagnostic (IVD) assay for laboratory use, not an AI-assisted diagnostic tool for human readers. Therefore, there is no mention of effect size for human readers improving with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

This device is a standalone algorithm/assay in the sense that it performs the qualitative determination of benzoylecgonine without a human-in-the-loop for the primary result interpretation. The "professional use" mentioned refers to laboratory professionals operating the automated analyzer and interpreting the results within a clinical context, not interacting with AI for improved diagnostic performance. The performance data presented (precision, accuracy, etc.) represents the standalone performance of the assay.

7. The Type of Ground Truth Used

The primary ground truth used for performance evaluation (especially for accuracy and precision) is Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS). For the precision study, it states: "All sample concentrations were verified by a confirmatory method (LCMS)." For method comparison and accuracy, it states: "Qualitative Mode Accuracy study with LC-MS/MS as reference method." This is an instrumental, objective laboratory method.

8. The Sample Size for the Training Set

The document describes performance studies for a developed immunoassay. It does not mention a "training set" in the context of machine learning, as this is a chemical assay, not an AI/ML algorithm that requires training data in that sense. The reagents and parameters are developed through R&D, not trained on a dataset like a deep learning model.

9. How the Ground Truth for the Training Set Was Established

As no training set (in the ML sense) is relevant for this type of immunoassay, this question is not applicable. The assay's analytical characteristics are determined by its chemical and biological components and detection principles, which are optimized during development and validated through the studies described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

January 27, 2022

Carolina Liquid Chemistries, Corp. Philip Shugart Chief Executive Officer 313 Gallimore Dairy Road Greensboro, North Carolina 27409

Re: K213211

Trade/Device Name: Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) Test System Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: September 27, 2021 Received: September 29, 2021

Dear Philip Shugart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213211

Device Name

Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) Test System

Indications for Use (Describe)

The Carolina Liquid Chemistries Cocaine Metabolite Enzyme Immunoassay (COCM) Test System is intended for the qualitative determination of benzoylecgonine (cocaine metabolite) in human urine at a cutoff value of 300 ng/mL. The assay is designed for professional use with a Carolina Liquid Chemistries CLC6410 automated clinical chemistry analyzer. For in vitro diagnostic use only. The assay provides a rapid screening procedure for determining the presence of benzoylecgonine in urine. The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k)#: K213211

A) Device Information:

Sponsor/Company Name:Carolina Liquid Chemistries, Corp.313 Gallimore Dairy RD.Greensboro, NC 27409Phone: 1-877-722-8910Fax: 1-336-722-8910
Correspondent Contact Information:Philip G ShugartCEOEmail: pshugart@carolinachemistries.comPhone: 1-877-722-8910Fax: 1-336-722-8910
Common Name of Device:Enzyme Immunoassay Cocaine and CocaineMetabolites
Trade Name of Device:Carolina Liquid Chemistries Cocaine andCocaine Metabolite Enzyme Immunoassay(COCM) Test System.
Device Product Code, Classification,Classification Name & PanelDIO, Class II, 21 CFR 862.3250 Opiate TestSystem 91-Toxicology

Predicate Device Information

Predicate Device:Cocaine Metabolite Enzyme Immunoassay
Predicate Device Manufacturer:Lin-Zhi International
Predicate Device CommonCocaine Metabolite Enzyme Immunoassay
Predicate Device Premarket Notification #K020763
Predicate Device Product Code,Classification, Classification Name &PanelDIO, Class II, 21 CFR 862.3250 Opiate TestSystem 91-Toxicology

{4}------------------------------------------------

B) Device Description:

The Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) Test System is a ready-to-use, liquid reagent homogeneous enzyme immunoassay for qualitatively determining the presence of cocaine metabolite (benzoylecgonine) in human urine. The assay uses specific antibody that can detect benzoylecgonine in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs. The assay is based on competition between benzoylecgonine labeled with the enzyme glucose-6phosphate dehydrogenase (G6PDH) and free drug from the urine sample, for a fixed amount of antibody. In the absence of free drug from the urine sample, the specific antibody binds to the drug labeled with G6PDH causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD+) to NADH.

C) Indications for Use:

The Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) Test System is intended for the qualitative determination of benzoylecgonine (cocaine metabolite) in human urine at a cutoff value of 300 ng/mL. The assay is designed for professional use with a Carolina Liquid Chemistries CLC6410 automated clinical chemistry analyzer. For in vitro diagnostic use only. The assay provides a rapid screening procedure for determining the presence of benzoylecgonine in urine. The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

{5}------------------------------------------------

D) Predicate Product Comparison Chart:

Characteristic:Candidate Device:Predicate Device:
Device Identifier and SubmitterCarolina Liquid Chemistries Corporation CocaineMetabolite Enzyme Immunoassay (COCM) Test SystemCocaine Metabolite Enzyme Immunoassay Lin-ZhiInternational, Inc (K020763)
Indication for UseThe Carolina Liquid Chemistries Cocaine and CocaineMetabolite Enzyme Immunoassay (COCM) Test System isintended for the qualitative determination ofbenzoylecgonine (cocaine metabolite) in human urine at acut off value of 300 ng/mL. The assay is designed forprofessional use on the CLC6410 automated clinicalchemistry analyzer. For in vitro diagnostic use only.The Lin-Zhi Cocaine Metabolite Enzyme Immunoassay isintended for the qualitative and semi-quantitativedetermination of benzoylecgonine (cocaine metabolite) inhuman urine at a cut off value of 300 ng/mL. The assay isdesigned for professional use on a number of automatedclinical chemistry analyzer. For in vitro diagnostic use only.
Operating Principle TechnologyThe assay is based on competitionbetween drug in the sample and drug labeled with theenzyme glucose-6- phosphatedehydrogenase (G6PDH) for a fixed amount of antibody inthe reagent Enzyme ImmunoassaySame
Measurand AnalyteCocaine Metabolite (benzoylecgonine)Same
Test MatrixUrineSame
MethodologyHomogeneous Enzyme ImmunoassaySame
Reagent FormLiquid Ready to UseSame
Antibodymouse monoclonalanti-benzoylecgonine antibodySame
Storage Temp2-8-CSame
Principal OperatorTrained ProfessionalsSame
Calibrator Levels2Same
Reference InstrumentCLC6410Syncron CX4CE

E) Test Principle:

The Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme immunoassay readyto-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6- phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, benzoylecgonine-labeled G6PDH conjugate is bound to the antibody, and

the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody binds the free drug; the unbound benzoylecgonine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

{6}------------------------------------------------

F) Summary of Supporting Data: Analytical Performance

  • Precision- Precision was determined by spiking benzoylecgonine into drug free urine a. at the following concentrations (0, -75%, -50%, -25%, cutoff (300 ng/mL), +125%, +150%, +175% and 200%). Testing for within run was performed by running two replicates of each sample twice in one day. The between run was performed by running two replicates of each sample twice a day for 22 non-consecutive days. All sample concentrations were verified by a confirmatory method (LCMS). The results for the 300 ng/mL cutoff are summarized in the table below:
Within Run
% CutoffConcentrationNNeg/Pos
0%044/0
-75%7544/0
-50%15044/0
-25%22544/0
Cutoff30043/1
125%37540/4
150%45040/4
175%52540/4
200%60040/4
Run-to-Run
% CutoffConcentrationNNeg/Pos
0%08888/0
-75%758888/0
-50%1508888/0
-25%2258888/0
Cutoff3008853/35
125%375880/88
150%450880/88
175%525880/88
  • Specificity- Specificity of the assay is supported by cross reactivity studies that b. supported the predicate device, K020763

Interference Testing-C.

  1. The Effect of pH: To investigate the effect of urine pH, negative human urine samples was divided into nine pools. Those pools were adjusted using sodium hydroxide (NaOH) and/or hydrochloric acid (HCL) to various pH conditions. After the pools were adjusted for pH, they were measured and the actual pH of each recorded. They were then divided in half so that there were two at each pH

{7}------------------------------------------------

level. Each half was then spiked with benzoylecgonine: one to 225 ng/mL and the other to 375 ng/mL (±25% of the 300 ng/mL cutoff). These 18-real human urine samples were then tested using the COCM Reagent Kit and results recorded in a table formatted similarly below.

InterferingSubstances-Actual pHTest Result-Actual pHTest Result-
pHpH Target225ng/mLTargetpH Target375ng/mLTarget
33.1Neg33.1Pos
44Neg43.9Pos
54.9Neg55.1Pos
66.1Neg66Pos
76.9Neg76.8Pos
88Neg88Pos
98.8Neg99.1Pos
1010.1Neg1010Pos
1111Neg1111.1Pos
  1. The Effect of Specific Gravity: To investigate the effect of urine specific gravity, negative human urine samples were divided into ten pools. These pools were adjusted using DI water and sodium chloride to the following target specific gravity conditions. After the pools were adjusted for specific gravity, they were measured, and the actual specific gravity was recorded. Each pool was then divided in half so there were two at each level of specific gravity. Each half was then spiked with benzoylecgonine: one to 225 ng/mL and the other to 375 ng/mL ±25% of the 300 ng/mL cutoff). Those 20-real human urine samples were then tested using the COCM Reagent Kit. The results were recorded in the table below.
InterferingSubstancesTest ResultTest Result
SGSG TargetActual SG225ng/mLTargetSG TargetActual SG375ng/mLTarget
1.0001.000Neg1.0001.001Pos
1.0031.002Neg1.0031.004Pos
1.0071.007Neg1.0071.008Pos
1.0111.010Neg1.0111.010Pos
1.0121.013Neg1.0121.013Pos
1.0171.017Neg1.0171.017Pos
1.0181.019Neg1.0181.019Pos
1.0211.021Neg1.0211.021Pos
1.0241.025Neg1.0241.025Pos
1.0281.029Neg1.0281.028Pos

No Substantial Interference was noted.

{8}------------------------------------------------

d. Carryover Testing- To determine carryover 21 samples, 10 "High" samples were spiked with benzoylecgonine at (1000 ng/mL) and 11 "Low" (0 ng/mL). They were assayed in the following order:

CarryoverSample 1 (0)NegSample 8(1000)PosSample 15(0)Neg
Sample 2 (0)NegSample 9 (0)NegSample 16(1000)Pos
Sample 3 (0)NegSample 10(0)NegSample 17(1000)Pos
Sample 4(1000)PosSample 11(0)NegSample 18(0)Neg
Sample 5(1000)PosSample 12(0)NegSample 19(1000)Pos
Sample 6 (0)NegSample 13(1000)PosSample 20(1000)Pos
Sample 7(1000)PosSample 14(1000)PosSample 21(0)Neg

No Carryover was noted during testing.

e. Method Comparison and Accuracy- Using 81 samples across the range of the assay should be tested according to the following distribution:

  • · 41 LC/MS Confirmed Negative Samples
    • 20 drug-free samples.
    • Remaining 21 samples between 0 ng/mL to 300 ng/mL.
      • 8 of the above 21 negative samples were within -50% of cutoff (150 ng/mL to 300 ng/mL)
  • · 40 LC/MS Confirmed Positives Samples
    • At least 8 samples within +50% of cutoff (300 ng/mL to 450 ng/mL)

All remaining samples greater than 300 ng/mL

Qualitative Mode Accuracy study with LC-MS/MS as reference method for 300 ng/mL cutoff shown below:

ResultsN=81Cutoff=300ng/mLDrug Free0 ng/mLLow Negative<50% off Cutoff(150ng/mL)Near CutoffNegativeBetween 50%and 99% ofCutoff(150-299 ng/mL)Near CutoffPositiveBetween 100%and 150% ofCutoff(300-450 ng/mL)High Positive>150% of Cutoff(>450 ng/mL)% Agreement
Positive832100%
Negative201110100%

Agreement among positives: 100% /Agreement among negatives: 100%

No discordant samples were identified

{9}------------------------------------------------

Conclusion:

The Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) was evaluated for several performance characteristics, including precision, sensitivity, accuracy, analytical recovery, specificity, carryover, and interference. All studies show acceptable results when compared to the predicate device.

In conclusion, the Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Enzyme Immunoassay (COCM) is substantially equivalent to the predicate.

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).