K Number
K213187
Date Cleared
2022-01-03

(96 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

AI/ML Overview

This document is a 510(k) Premarket Notification for a Disposable Nitrile Medical Powder-free Glove. It does not describe an AI/ML-driven medical device. Therefore, the request to describe acceptance criteria and a study proving an AI/ML device meets those criteria cannot be fully answered from this document.

However, I can extract information related to the non-clinical performance testing conducted for this medical device (the glove), which serves a similar purpose to acceptance criteria and study results for verifying performance against established benchmarks.

Here's an adaptation of the requested format focusing on the non-clinical performance of the medical glove, as an AI/ML study was not conducted:


Device: Disposable Nitrile Medical Powder-free Glove (non-AI/ML device)

This document does not pertain to an AI/ML medical device. Therefore, information regarding "Multi-Reader Multi-Case (MRMC) comparative effectiveness study," "standalone (algorithm only) performance," "effect size of human readers improving with AI," "sample size for training set," and "how ground truth for training set was established" is not applicable.

The following tables and information describe the acceptance criteria and the non-clinical testing proving the device meets these criteria.

1. Table of Acceptance Criteria and the Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm): S: ≥220; M/L/XL: ≥230
Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10
Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Length (mm): > 230/Pass
Width (mm): S: 83-86 /Pass; M: 94-98/ Pass; L: 101-103/ Pass; XL:113-115/ Pass
Thickness (mm): Finger: 0.13-0.15/Pass; Palm: 0.09/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.