(96 days)
The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.
The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.
This document is a 510(k) Premarket Notification for a Disposable Nitrile Medical Powder-free Glove. It does not describe an AI/ML-driven medical device. Therefore, the request to describe acceptance criteria and a study proving an AI/ML device meets those criteria cannot be fully answered from this document.
However, I can extract information related to the non-clinical performance testing conducted for this medical device (the glove), which serves a similar purpose to acceptance criteria and study results for verifying performance against established benchmarks.
Here's an adaptation of the requested format focusing on the non-clinical performance of the medical glove, as an AI/ML study was not conducted:
Device: Disposable Nitrile Medical Powder-free Glove (non-AI/ML device)
This document does not pertain to an AI/ML medical device. Therefore, information regarding "Multi-Reader Multi-Case (MRMC) comparative effectiveness study," "standalone (algorithm only) performance," "effect size of human readers improving with AI," "sample size for training set," and "how ground truth for training set was established" is not applicable.
The following tables and information describe the acceptance criteria and the non-clinical testing proving the device meets these criteria.
1. Table of Acceptance Criteria and the Reported Device Performance
| Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| ASTM D6319 | Physical Dimensions Test | Length (mm): S: ≥220; M/L/XL: ≥230Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10Thickness (mm): Finger: ≥0.05; Palm: ≥0.05 | Length (mm): > 230/PassWidth (mm): S: 83-86 /Pass; M: 94-98/ Pass; L: 101-103/ Pass; XL:113-115/ PassThickness (mm): Finger: 0.13-0.15/Pass; Palm: 0.09/Pass |
| ASTM D5151 | Watertightness Test for Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | 0/125/Pass |
| ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124 << 2.0mg | 0.21-0.24mg/Pass |
| ASTM D412 | Physical Properties | Before Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥500%After Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥400% | Before Aging: 14.3-20.7MPa/Pass; 500-699%/PassAfter Aging: 14.0-17.8MPa/Pass; 490-569%/Pass |
| ISO 10993-11 | Cytotoxicity (Systemic Toxicity) | Non-acute systemic toxicity | Under conditions of the study, did not show acute systemic toxicity in vivo / Pass |
| ISO 10993-10 | Irritation | Non-irritating | Under the conditions of the study, not an irritant/ Pass |
| ISO 10993-10 | Sensitization | Non-sensitizing | Under conditions of the study, not a sensitizer./ Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size for each test, but it provides specific values for some tests:
- Watertightness Test (ASTM D5151): Tested 125 units (0/125/Pass implies a sample size of 125).
- For other tests like dimensions, powder content, and physical properties, specific sample sizes are not detailed but are assumed to be in compliance with the respective ASTM and ISO standards.
- Data Provenance: Not specified, but generally, these non-clinical tests are conducted by the manufacturer or a contracted lab in the country of origin (China, in this case). The data is retrospective in the sense that it's gathered specifically for the submission.
3. Number of Experts and Qualifications for Ground Truth
- Not Applicable. This is a physical medical device (glove), not an AI/ML device requiring expert interpretation for ground truth. Ground truth is established through standardized physical, chemical, and biological testing methods.
4. Adjudication Method for the Test Set
- Not Applicable. As this is not an AI/ML device involving human interpretation, no adjudication method is relevant. Test results are objective measurements against established standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. This is a physical medical device, not an AI/ML diagnostic or assistive tool.
6. Standalone Performance (Algorithm Only)
- Not Applicable. This is a physical medical device, not an algorithm.
7. Type of Ground Truth Used
- Objective Measurement against Industry Standards: The ground truth for this device's performance is established by standardized test methods from organizations like ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). These standards define the physical, chemical, and biological properties expected of such a device. For example, "Tensile Strength ≥14MPa" is an objective ground truth defined by ASTM D412.
8. The Sample Size for the Training Set
- Not Applicable. This is a physical medical device, not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. This is a physical medical device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 3, 2022
Fujian Chunhui Medical Supply Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K213187
Trade/Device Name: Disposable Nitrile Medical Powder-free Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 18, 2021 Received: September 29, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213187
Device Name Disposable Nitrile Medical Powder-free Glove
Indications for Use (Describe)
The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K213187
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
1.0 Submitter's Information
Name: Fujian Chunhui Medical Supply Co., Ltd. Address: No.179, Shanggeteng Ping, Shifang Village, Shifang County, Wuping Town, Longyan City, Fujian Province, China. Phone Number: +86-18060003170 Contact: Han Cai Date of Preparation: Sept 18, 2021
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Disposable Nitrile Medical Powder-free Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S. M. L. XL
3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate Device Information
Manufacturer: GUANG DONG KINGFA SCI. & TECH.CO., LTD. Device: Nitrile examination gloves 510(k) number: K203593
5.0 Indication for Use
{4}------------------------------------------------
The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.
6.0 Device Description
The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.
7.0 Technological Characteristic Comparison Table
| Item | Subject Device | Predicated Device(K203593) | Comparison |
|---|---|---|---|
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use | The Disposable NitrileMedical Powder-free Gloveis intended to be worn onthe hands of examiners toprevent contaminationbetween patient andexaminer. This is asingle-use, powder-free,non-sterile device. | The nitrile examinationglove is intended to beworn on the hands ofexaminers toprevent contaminationbetween patient andexaminer. This is asingle-use, powder-free,non-sterile device. | Same |
| Material | Nitrile | Nitrile | Same |
| Powdered orPowered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Colorant | Blue | Blue | Same |
| LabelingInformation | Single-use indication,powder free, device color,device name, glove size andquantity, Non-Sterile | Single-use indication,powder free, device color,device name, glove sizeand quantity, Non-Sterile | Same |
| Dimensions(mm) | Length:S:≥220;M/L/XL: ≥230;Width:S: 80±10;M: 95±10;L: 110±10;XL: 120±10 | Length:S:≥220;M/L/XL: ≥230;Width:S: 80±10;M: 95±10;L: 110±10;XL: 120±10 | Same |
Table1-General Comparison
{5}------------------------------------------------
| Thickness(mm) | Finger: ≥0.05;Palm: ≥0.05 | Finger: ≥0.05;Palm: ≥0.05 | Same | |||
|---|---|---|---|---|---|---|
| PhysicalProperties | BeforeAging | TensileStrength | 14MPa, min | TensileStrength | 14MPa, min | Same |
| UltimateElongation | 500% min | UltimateElongation | 500% min | Same | ||
| AfterAging | TensileStrength | 14MPa, min | TensileStrength | 14MPa, min | Same | |
| UltimateElongation | 400%min | UltimateElongation | 400%min | Same | ||
| Freedom fromHoles | Be free from holes whentested in accordance withASTMD5151 AQL=2.5 | Be free from holes whentested in accordance withASTMD5151 AQL=2.5 | Same | |||
| Powder Content | Meet the requirements ofASTM D6124 | Meet the requirements ofASTM D6124 | Same | |||
| Biocompatibility | ISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizer | ISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizer | Same | |||
| ISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.ISO 10993-5Under conditions of thestudy, device extract is | ISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.ISO 10993-5Under conditions of thestudy, device extract is | Same |
8.0 Summary of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for
{6}------------------------------------------------
systemic toxicity.
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
| TestMethod | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTMD6319 | PhysicalDimensionsTest | Length(mm):S:≥220;M/L/XL:≥230;Width(mm):S: 80±10;M: 95±10;L: 110±10;XL: 120±10Thickness (mm):Finger: ≥0.05Palm: ≥0.05 | Length(mm):> 230/Pass;Width(mm):S: 83-86 /PassM: 94-98/ PassL: 101-103/ PassXL:113-115/ PassThickness (mm):Finger: 0.13-0.15/PassPalm: 0.09/Pass |
| ASTMD5151 | WatertightnessTest forDetection ofHoles | Meet the requirements of ASTM D5151AQL 2.5 | 0/125/Pass |
| ASTMD6124 | PowderContent | Meet the requirements of ASTM D6124 <2.0mg | 0.21-0.24mg/Pass; |
| ASTMD412 | Physicalproperties | Before AgingTensile Strength ≥14MPaUltimate Elongation ≥500%After AgingTensile Strength ≥14MPaUltimate Elongation ≥400% | 14.3-20.7MPa/Pass;500-699%/Pass;14.0-17.8MPa/Pass;490-569%/Pass; |
| ISO10993-11 | Cytotoxicity | Non- acute systemictoxicity | Under conditions ofthe study, did notshow acute systemic |
| toxicity in vivo / Pass | |||
| ISO10993-10 | Irritation | Non-irritating | Under the conditionsof the study, not anirritant/ Pass |
| ISO10993-10 | Sensitization | Non-sensitizing | Under conditions ofthe study, not asensitizer./ Pass |
Table 2 - Summary of non-clinical performance testing
{7}------------------------------------------------
9.0 _Summary of Clinical Testing
Clinical testing is not needed for this device.
10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device Nitrile Examination Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicated device K203593.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.