K Number
K213084
Date Cleared
2022-02-25

(155 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ear Thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It is a non-contact method for forehead measurement, the recommended measurement distance is 0cm. The Ear Thermometer can be used by consumers in the household environment and by healthcare providers.

Device Description

The Ear Thermometers are a hand-held, battery powered infrared thermometer. It is intended to measure the temperature of human body from ear or forehead. The infrared sensor contained in the device can monitor the energy radiated from the human ear tympanum and forehead. The device can analyze and processes the signal transmitted by the infrared sensor, and finally displays it on the screen in digital form to achieve the measurement of human body temperature. People of all ages in household environment and doctors in clinic as reference are suitable for using the device.

AI/ML Overview

The Berrcom Ear Thermometer (models ET005, ET008, ET009) underwent a clinical accuracy validation to demonstrate its performance. The study followed ISO 80601-2-56:2017+AMD2018.

1. Acceptance Criteria and Reported Device Performance:

The document explicitly states that the clinical validation results demonstrated that "the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical validation protocol." However, the specific numerical acceptance criteria (e.g., maximum allowable bias or repeatability) are not provided in the submitted text. Therefore, the table below reflects what is stated in the document regarding meeting the criteria, but the exact values for the criteria cannot be extracted.

Acceptance Criteria CategoryAcceptance Criteria (Not explicitly stated in document)Reported Device Performance
Clinical Bias(Not explicitly stated, but met)Met acceptance criteria
Clinical Repeatability(Not explicitly stated, but met)Met acceptance criteria

The document also provides measurement accuracy specifications which act as performance criteria:

Temperature RangePerformance Criteria
Ear & Forehead: 32.0°C ~ 34.9°C (89.6°F ~ 94.8°F)± 0.3°C (± 0.5°F)
Ear & Forehead: 35.0°C ~ 42.0°C (95°F ~ 107.6°F)± 0.2°C (± 0.4°F)
Ear & Forehead: 42.1°C ~ 43.0°C (107.8°F ~ 109.4°F)± 0.3°C (± 0.5°F)
The document does not explicitly state that these criteria were tested during the clinical accuracy validation for individual subjects. Instead, it states the overall clinical bias and clinical repeatability met the acceptance criteria.

2. Sample Size and Data Provenance:

  • Sample Size: 150 subjects. This included:
    • 50 infants (newborn to one year)
    • 50 children (greater than one to five years)
    • 50 adults (greater than five years old)
  • Data Provenance: The document does not specify the country of origin of the data. It mentions "clinical investigation," which implies a prospective study.

3. Number of Experts and Qualifications:

The document does not provide information regarding the involvement of experts to establish ground truth, nor their qualifications.

4. Adjudication Method:

The document does not provide information regarding any adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. This study focuses on the accuracy of a single device against a ground truth, not on the improvement of human readers with AI assistance.

6. Standalone Performance:

Yes, a standalone performance study was done. The "Clinical accuracy validation" described is a test of the device's accuracy in measuring body temperature.

7. Type of Ground Truth Used:

The document refers to "clinical validation results demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical validation protocol." While it implies a comparison to a referent temperature measurement standard (which would constitute the ground truth), the specific method for establishing this ground truth (e.g., rectal thermometer, oral thermometer, etc.) is not explicitly stated. However, given the context of clinical thermometry standards, it is highly likely a validated reference thermometer was used.

8. Sample Size for Training Set:

The document does not provide information about a training set since this is a medical device validation (Ear Thermometer), not an AI/machine learning model that typically has a training set.

9. How Ground Truth for Training Set was Established:

Not applicable, as no training set is mentioned for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

February 25, 2022

Guangzhou Berrcom Medical Device Co., Ltd. % Yoyo Chen Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518000 China

Re: K213084

Trade/Device Name: Ear Thermometer, Model ET005, ET008, ET009 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 15, 2021 Received: September 23, 2021

Dear Yoyo Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213084

Device Name

Ear Thermometer, Model: ET005, ET008, ET009

Indications for Use (Describe)

The Ear Thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It is a non-contact method for forehead measurement, the recommended measurement distance is 0cm. The Ear Thermometer can be used by consumers in the household environment and by healthcare providers.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Berrcom" in blue font. The first letter "B" is stylized with a heart-shaped curve on the left side. The rest of the letters are in a simple, sans-serif font. The word appears to be a logo or brand name.

510(k) Summary

This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.

1. Administrative Information

I. Administrative Information
Submission DateDecember 3, 2021
ManufacturerinformationGuangzhou Berrcom Medical Device Co., Ltd.
Address:
No.38 Huanzhen Xi Road, Dagang Town, Nansha, 511470,Guangzhou, Guangdong, PEOPLE'S REPUBLIC OF CHINA
Contact person: Zhigang Du
TEL: +86(20)34938449
E-Mail: dube888@berrcom.com
SubmissionCorrespondentShenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square, LiuxianAvenue, Xili Town, Nanshan District, Shenzhen, GuangdongProvince, China.
Contact person: Ms.Yoyo Chen
E-Mail: yoyo@cefda.com; field@cefda.com
Image: logo
Establishmentregistration number3008395508

2. Device Information

Type of 510(k) Submission:Traditional
Device Name:Ear Thermometer
Model:ET005,ET008,ET009
Classification Name:Clinical Electronic Thermometer (Infrared Thermometer)
Review Panel:General Hospital
Device Class:2
Regulation Number:880.2910
Product Code:FLL

3. Predicate Device

Manufacturer:Guangzhou Berrcom Medical Device Co., Ltd.
Device Name:Infrared Thermometer Model MD-H30
Model:MD-H30

{4}------------------------------------------------

4. Device Description

The Ear Thermometers are a hand-held, battery powered infrared thermometer. It is intended to measure the temperature of human body from ear or forehead. The infrared sensor contained in the device can monitor the energy radiated from the human ear tympanum and forehead. The device can analyze and processes the signal transmitted by the infrared sensor, and finally displays it on the screen in digital form to achieve the measurement of human body temperature. People of all ages in household environment and doctors in clinic as reference are suitable for using the device.

It has the following features:

  • Forehead and ear temperature measurement. 1)
    1. Fahrenheit and Celsius temperature unit setting.
  • Display resolution is 0.1°C (0.1°F). 3)
    1. White color LED digital display screen.
    1. Automatically shut down if there is no operation within 30 seconds to conserve energy.
    1. Mute or unmute setting.
    1. High temperature indication.
    1. Exceeds measurement range prompt.

5. Intended Use/Indications for Use

The Ear Thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It is a noncontact measurement method for forehead measurement, the recommended measurement distance is 0cm to 3cm. The Ear Thermometer can be used by consumers in the household environment and by healthcare providers.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Berrcom" in a stylized, sans-serif font. The word is rendered in a blue color. The "B" in "Berrcom" is stylized with a heart-shaped curve on the left side. The letters are connected to each other, giving the word a flowing appearance.

Comparison with predicate device 6.

ItemsSubject Devices(K213084)Predicate Device(K191570)Comments
Product CodeFLLFLLSame
Regulation number880.2910880.2910Same
ManufacturerGuangzhou Berrcom MedicalDevice Co., Ltd.Guangzhou Berrcom MedicalDevice Co., Ltd.Same
Indications for useThe Ear Thermometer is intended forthe intermittent measurement of bodytemperature from the auditory canal orcentral forehead skin surface onpeople of all ages. It is a non-contactmeasurement method for foreheadmeasurement, the recommendedmeasurement distance is 0cm to 3cm.The Ear Thermometer can be used byconsumers in the householdenvironment and by healthcareproviders.The infrared thermometer isintended for the intermittentmeasurement of bodytemperature from the auditorycanal or central forehead skinsurface on people of all ages. Itcan be used by consumers inthe household environment andby healthcare providers.Same
Thermometer typeDigital thermometerDigital thermometerSame
SensorThermopileThermopileSame
Operational principleInfrared radiation detectionInfrared radiation detectionSame
Measurement referencebody siteAxillaAxillaSame
Measurement modeAdjusted modeAdjusted modeSame
Measurement methodNon-contact typeNon-contact typeSame
Measurement distanceof foreheadmeasurement0cm to 3cm0cm to 3cmSame
Patient populationPeople of all agesPeople of all agesSame
RX/OTC UseOTCOTCSame
Display typeLED screenLCD screenDifferent(Note 1)
Measurement siteForehead and earForehead and earSame
Measurement RangeEar & Forehead:32.0°C ~ 43.0°C(89.6°F ~ 109.4°F)Ear & Forehead: 32.0°C43°C(89.6°F109.4°F)Same
Measurement accuracy32.0°C~ 34.9°C(89.6°F94.8°F) :$\pm$ 0.3°C( $\pm$ 0.5°F)35.0°C 42.0°C(95°F107.6°F):$\pm$ 0.2°C( $\pm$ 0.4°F);42.1°C 43.0°C(107.8°F~109.4°F ):$\pm$ 0.4°F (0.2°C) within95107.6°F (3542°C),$\pm$ 0.5°F (0.3°C) for other range.Same
ItemsSubject Devices(K213084)Predicate Device(K191570)Comments
$\pm 0.3°C(\pm 0.5°F)$
Display resolution0.1°C (0.1°F)0.1°C (0.1°F)Same
Power supplyDC 3V (2pcs AAA batteries)DC 3V (2pcs AAA batteries)Same
Measurement time1 second1 secondSame
Auto-off time30s30sSame
Audible AlertYesYesSame
Use of Probe CoversNoNoSame
Operation ConditionTemperature: 10°C40°CRelative Humidity: ≤85%,noncondensingAtmospheric Pressure: 70kPa106kPaTemperature: 10°C40°CRelative Humidity: ≤85%Atmospheric Pressure:70kPa106kPaSame
Storage andtransportation conditionTemperature: -20°C55°CRelative Humidity: ≤95%RH,noncondensingAtmospheric Pressure: 70kpa106kpaTemperature: -20°C55°CRelative Humidity: ≤95%RH,Atmospheric Pressure:70kpa106kpaSame
IP ClassIP22IP22Same
Materials of skin-contacting componentsABS, PCABSDifferent(Note 2)
Physical Dimensions144×34×43.5mm (L x W x H)156mm46mm48mm (L x W x H)Different(Note 2)
Weight80g77gDifferent(Note 2)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Berrcom" in a stylized, sans-serif font. The word is rendered in a blue color. The "B" in "Berrcom" is designed with a unique, rounded shape, resembling a stylized heart or a curved shape. The letters are connected with rounded edges, giving the word a smooth and flowing appearance.

Note 1: Display type

The subject device meets the basic safety requirement of IEC 60601-1:2005+AMD 1: 2012, and IEC 60601-1-11. The difference does not raise any issues on the device safety and effectiveness.

Note 2: Materials of skin-contacting components, Physical Dimensions, and weight The biocompatibility tests have conducted on the subject device to demonstrate that the subject device did not have any potential toxicity, skin sensitization and skin irritation during normal use. Otherwise, the subject device has passed the electrical safety tests in according to IEC 60601-1-2, IEC 60601-1, and IEC 60601-1-11 to demonstration that the subject device could be as safe as the predicate device in the intended use environment, and the performance testing shows that the subject device complies with performance standard ISO 80601-2-56 and ASTM E1965-98. The difference does not raise any issues on the device safety and effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Berrcom" in a stylized, sans-serif font. The word is rendered in a dark blue color. The "B" in "Berrcom" is stylized with a heart-shaped curve on the left side of the letter. The rest of the letters are in lowercase.

nical Test Summary

7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject device meets the following standards:

  • ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and 1) A2:2010/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    1. IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

7.2. Biocompatibility Test

The subject device has passed biocompatibility tests in according to following standards.

  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro 1) cytotoxicity
    1. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

7.3. Performance Test-Bench

The subject device has passed performance tests in according to following standards.

    1. ISO 80601-2-56:2017+AMD2018 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
    1. ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature

7.4. Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met, and all software hazards have been mitigated to acceptable risk levels.

8. Clinical Accuracy Validation

Clinical accuracy validation was conducted in according to ISO 80601-2-56:2017+AMD2018. This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consisting of 150 subjects, of which 50 subjects are infants, 50 subjects are children, and the rest 50 subjects are adults (NOTE: Infants---newborn to one year; Children-- greater than one to five years; Adults---greater than five years old.). The clinical validation results demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical validation protocol.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Berrcom" in a stylized, sans-serif font. The word is rendered in a dark blue color. The "B" in "Berrcom" is stylized to resemble a heart shape, with the left side of the "B" forming the curve of the heart.

9. Conclusion

The Ear Thermometers (Model: ET005, ET008, ET009) are substantially equivalent to the infrared Thermometer Model MD-H30 cleared under K191570. This conclusion is based upon comparison on intended use, technological characteristics and applicable safety standards. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.