(127 days)
The Non-Contact Infrared Thermometer, Models JXB-315, JXB-320, JXB-311, is a non-sterile, reusable, handheld device. It can be used by consumers in household environment and doctor in clinic as reference. It is intended for non-contact measuring of human body temperature of people of all ages (neonate, infant, pediatric, adult) by detecting infrared heat from the center of the forehead.
The Non-contact Infrared Thermometer (include models JXB-315, JXB-320, JXB-311) is a hand-held forehead thermometer, and it is intended for the non-contact intermittent measurement and monitoring of forehead temperature of human body which measures the body temperature based on the infrared energy emitted from the forehead. It is indicated for use by people of all ages. The thermometer is also intended to measure the object temperature. After measurement, the temperature is directly displayed on the LED screen. These thermometers have the following features: 1) Body temperature mode 2) Fahrenheit and Celsius temperature unit setting LED display screen with automatic range selection, resolution is 0.1 degree Celsius (0.1 degree Fahrenheit) 3) 4) Memorize the latest 32 measurement data LED display screen 5) 6) Low battery indication 7) Measurement distance indication 8) Over range prompt (HI/LO) 9) Auto shutdown when the device is idle for 30 seconds 10) High temperature indication
The provided document describes the acceptance criteria and a clinical study conducted for the Non-contact Infrared Thermometer, Models JXB-315, JXB-319, JXB-320, JXB-311.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document references compliance with specific standards for measurement accuracy rather than explicitly stating numerical acceptance criteria. However, it does compare the subject device's declared accuracy with the predicate device and relevant standards.
| Acceptance Criteria (Standard Reference) | Reported Device Performance (Subject Device) |
|---|---|
| Measurement Accuracy (Body Temperature Mode): ASTM E1965-98 and ISO 80601-2-56 standards (range-specific accuracies) | Within the range of $34.0°C$ to $43.0°C$, the accuracy is the same as the predicate device ($±0.2°C$ for $35.0°C-42.0°C$, $±0.3°C$ for edges). For $32.0°C-34.9°C$, it is $±0.3°C$. Verified to comply with ASTM E1965-98 and ISO 80601-2-56. |
| Clinical Accuracy (Clinical Bias and Repeatability): Acceptance criteria of the clinical validation protocol based on ISO 80601-2-56:2017+AMD2018. | Clinical bias and clinical repeatability met the acceptance criteria of the clinical validation protocol. |
| Electromagnetic Compatibility (EMC): IEC 60601-1-2:2014 | Passed |
| Electrical Safety: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, IEC 60601-1-11:2015 | Passed |
| Biocompatibility: ISO 10993-5:2009 (in vitro cytotoxicity), ISO 10993-10:2010 (irritation and skin sensitization) | Passed (demonstrated no potential toxicity, skin sensitization, and skin irritation) |
| Software Verification and Validation: Consistent with moderate level of concern. All software requirement specifications met, and hazards mitigated. | Passed |
2. Sample size used for the test set and the data provenance
- Sample Size: 150 subjects for the clinical accuracy validation.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "clinical investigation," which implies prospective data collection for the purpose of the validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical study.
4. Adjudication method for the test set
The document does not specify an adjudication method for the test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a standalone thermometer, and the study focused on its accuracy in measuring body temperature, not on assisting human readers with interpreting medical images or data.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance study was done for the device itself to measure its accuracy against established standards. The section "Clinical Accuracy Validation" specifically details how the device's readings were validated against clinical ground truth.
7. The type of ground truth used
The ground truth for the clinical accuracy validation was based on clinical measurements performed on subjects, implicitly alongside a reference method (though the reference method is not explicitly named, it's standard practice in such studies to compare against a highly accurate clinical thermometer). The study assessed "clinical bias and clinical repeatability" against this ground truth.
8. The sample size for the training set
The document does not mention a training set because this is a hardware medical device (a thermometer), not an AI/machine learning algorithm that typically requires a distinct training phase.
9. How the ground truth for the training set was established
Not applicable, as there is no mention of a training set for this hardware device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 28, 2022
Guangzhou Berrcom Medical Device Co., Ltd. % Yoyo Chen Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518000 China
Re: K213082
Trade/Device Name: Non-contact Infrared Thermometer, Models JXB-315, JXB-319, JXB-311 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 14, 2021 Received: December 27, 2021
Dear Yoyo Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213082
Device Name
Non-contact Infrared Thermometer, Models JXB-315, JXB-319, JXB-320, JXB-311
Indications for Use (Describe)
The Non-Contact Infrared Thermometer, Models JXB-315, JXB-320, JXB-311, is a non-sterile, reusable, handheld device. It can be used by consumers in household environment and doctor in clinic as reference. It is intended for non-contact measuring of human body temperature of people of all ages (neonate, infant, pediatric, adult) by detecting infrared heat from the center of the forehead.
Type of Use (Select one or both, as applicable)
Registration Under Part 107. FAA Rules & Regulations
Drone Flight Operation Under 107. FAA Rules & Regulations
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Berrcom" in blue font. The first letter "B" is stylized to look like a heart. The rest of the letters are in a simple, sans-serif font. The word is written in lowercase letters.
510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
1. Administrative Information
| Submission Date | December 14, 2021 |
|---|---|
| Manufacturerinformation | Guangzhou Berrcom Medical Device Co., Ltd.Address:No.38 Huanzhen Xi Road, Dagang Town, Nansha, 511470,Guangzhou, Guangdong, PEOPLE'S REPUBLIC OF CHINAContact person: Zhigang DuTEL: +86(20)34938449E-Mail: dube888@berrcom.com |
| SubmissionCorrespondent | Shenzhen Joyantech Consulting Co., Ltd.1713A, 17th Floor, Block A, Zhongguan Times Square, LiuxianAvenue, Xili Town, Nanshan District, Shenzhen, GuangdongProvince, China.Contact person: Ms.Yoyo ChenE-Mail: yoyo@cefda.com; field@cefda.com |
| Image: Logo | |
| Establishmentregistration number | 3008395508 |
Device Information 2.
| Type of 510(k)Submission: | Traditional |
|---|---|
| Device Name: | Non-contact Infrared Thermometer |
| Models: | JXB-315, JXB-319, JXB-320, JXB-311 |
| Classification Name: | Thermometer, Electronic, Clinical |
| Review Panel: | General Hospital |
| Device Class: | 2 |
| Regulation Number: | 880.2910 |
| Product Code: | FLL |
3. Predicate Device
| Manufacturer: | BroadMaster Biotech, Corp. |
|---|---|
| Device Name: | Advocate Non-Contact Infrared Thermometer |
| Model: | EF001A |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Berrcom. The logo is in blue and features a stylized "B" that resembles a heart shape. The text "errcom" is written in a sans-serif font and is connected to the "B".
| 510(K) Number: | K180355 |
|---|---|
| Product Code: | FLL |
| Regulation Number: | 880.2910 |
| Device Class: | 2 |
Device Description 4.
The Non-contact Infrared Thermometer (include models JXB-315, JXB-320, JXB-311) is a hand-held forehead thermometer, and it is intended for the non-contact intermittent measurement and monitoring of forehead temperature of human body which measures the body temperature based on the infrared energy emitted from the forehead. It is indicated for use by people of all ages.
The thermometer is also intended to measure the object temperature. After measurement, the temperature is directly displayed on the LED screen.
These thermometers have the following features:
-
- Body temperature mode
-
- Fahrenheit and Celsius temperature unit setting
- LED display screen with automatic range selection, resolution is 0.1℃ (0.1℃) 3)
-
- Memorize the latest 32 measurement data
- LED display screen 5)
-
- Low battery indication
-
- Measurement distance indication
-
- Over range prompt (HI/LO)
-
- Auto shutdown when the device is idle for 30 seconds
-
- High temperature indication
5. Intended Use
The Non-Contact Infrared Thermometer, Models JXB-315, JXB-319, JXB-320, JXB-311, is a non-sterile, reusable, handheld device. It can be used by consumers in household environment and doctor in clinic as reference. It is intended for non-contact measuring of human body temperature of people of all ages (neonate, infant, pediatric, adult) by detecting infrared heat from the center of the forehead.
| Items | Subject Devices(K213082) | Predicate Device(K180355) | Comments |
|---|---|---|---|
| Product Code | FLL | FLL | Same |
| Regulation number | 880.2910 | 880.2910 | Same |
| Manufacturer | Guangzhou Berrcom MedicalDevice Co., Ltd. | BroadMaster Biotech, Corp. | / |
| Indications for use | The Non-Contact InfraredThermometer, Models JXB-315,JXB-319, JXB-320, JXB-311, is anon-sterile, reusable, handheld | Advocate Non-ContactInfrared Thermometer is anon-sterile, reusable,handheld device. It can be | Same |
| Items | Subject Devices(K213082) | Predicate Device(K180355) | Comments |
| device. It can be used by consumersin household environment and doctorin clinic as reference. It is intendedfor non-contact measuring of humanbody temperature of people of allages (neonate, infant, pediatric,adult) by detecting infrared heat fromthe center of the forehead. | used by consumers inhomecare environment anddoctors in clinic as reference.It is intended for measuringhuman body temperature ofall ranges of people bydetecting infrared heat fromthe forehead. | ||
| Prescription/OTC | OTC use | OTC use | Same |
| Intended user | Lay user and professional | Lay user and professional | Same |
| Intended patientpopulation | Neonate, infant, pediatric, adult | Neonate, infant, pediatric,adult | Same |
| Sensor | Thermopile | Thermopile | Same |
| Operationalprinciple | Infrared radiation detection | Infrared radiation detection | Same |
| Contact type | Non-contact use | Non-contact use | Same |
| Display type | LED display | LCD display | SE(Note 1) |
| Measurement site | Forehead | Forehead | Same |
| MeasurementRange | $32.0°C ~ 43.0°C(89.6°F ~ 109.4°F)$ | $32.0°C | Same |
| Measurementaccuracy | $32.0°C~ 34.9°C(89.6°F~94.8°F) :±0.3°C(±0.6°F)$ | $34.0°C | SE(Note 2) |
| $35.0°C~ 42.0°C(95°F~107.6°F):±0.2°C(±0.4°F);$ | $35.0°C | ||
| $42.1°C~ 43.0°C(107.8°F~109.4°F ):±0.3°C(±0.6°F)$ | $42.1°C | ||
| Measurementdistance | ≤3cm (1.2in) | 5-10 cm (2-3.94in) | SE(Note 3) |
| Display resolution | 0.1°C (0.1°F) | 0.1°C (0.1°F) | Same |
| Power supply | DC 3V (2pcs AAA batteries) | DC 3V (2pcs AAA batteries) | Same |
| Measurement time | 1 second | 1 second | Same |
| Memory capacity | 32sets | 12 sets | SE(Note 4) |
| Low battery | Yes | Yes | Same |
| Items | Subject Devices(K213082) | Predicate Device(K180355) | Comments |
| indication | |||
| Degree of protection | IP22 | IP20 | SE(Note 5) |
| Auto-off time | 30s | 60s | SE(Note 6) |
| Operation Condition | Temperature: 10°C | Temperature: 10°C | SE(Note 7) |
| Storage andtransportationcondition | Temperature: -20°C | Temperature: -20°C | Same |
| Materials of skin-contactingcomponents | ABS, PC | Enclosure of red &black ABS,LCD Lens of PMMA andProbe of Metals | SE(Note 8) |
| Biocompatibility | Comply withISO10993-5:2009 &ISO10993-10:2010 | Comply withISO10993-5:2009 &ISO10993-10:2010 | Same |
| Electrical safety | IEC 60601-1 | IEC 60601-1 | Same |
| EMC | IEC60601-1-2 | IEC60601-1-2 | Same |
| Performance | ASTM E1965-98ISO80601-2-56 | ASTM E1965-98ISO80601-2-56 | Same |
6. Comparison with predicate device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Derrcom" in a stylized, sans-serif font. The word is rendered in a blue color. The "D" in "Derrcom" is stylized with a heart shape.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Berrcom" in blue font. The "B" is stylized with a heart shape on the left side. The rest of the letters are in lowercase and connected in a cursive style. The overall design is simple and clean.
Note 1: Display type
The subject device meets the basic safety requirement of IEC 60601-1:2005+AMD 1: 2012, and IEC 60601-1-11. The difference does not raise any issues on the device safety and effectiveness.
Note 2: Measurement Accuracy
The measurement accuracy of subject devices in the range of 34°C to 43°C is same with predicate device, and the measurement accuracy of subject device still compiles with the requirement of ASTM E1965-98 and ISO 80601-2-56 standard. This difference does not raise any issues on the device safety and effectiveness.
Note 3: Measurement distance
The performance of subject device within the specified measurement distance compiles with the requirement of ASTM E1965-98 and ISO 80601-2-56 standard. This difference does not raise any issues on the device safety and effectiveness.
Note 4: Memory capacity
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Berrcom" in blue font. The first letter "B" is stylized with a heart shape on the left side. The rest of the letters are in lowercase and connected to each other. The image appears to be a logo or brand name.
urpose of function of measurement data memories is intended to store and view the previous readings. This function has been verified during software verification, and the performance testing shows that the subject device complies with performance standard ISO 80601-2-56 and ASTM E1965-98. The difference does not raise any issues on the device safety and effectiveness.
Note 5: Degree of protection
The subject device complies with the requirement of IEC60601-1, ASTM E1965-98 and ISO 80601-2-56 standard. The difference does not raise any issues on the device safety and effectiveness.
Note 6: Auto-off time
Although the time of auto-off time is shorter than predicate device, but both of subject devices and predicate device are meet the basic safety requirement of IEC 60601-1, and IEC 60601-111. The difference does not raise any issues on the device safety and effectiveness.
Note 7: Operation Condition
The operation condition of the subject device is tested and validated to meet the requirement of ASTM E1965-98 and ISO80601-2-56. The difference does not raise any issues on the device safety and effectiveness.
Note 8: Materials of skin-contacting components
The biocompatibility tests have conducted on the finished device to demonstrate that the subject device did not have any potential toxicity, skin sensitization and skin irritation during normal use. The difference does not raise any issues on the device safety and effectiveness.
7. Non-Clinical Test Summary
7.1. Electromagnetic Compatibility and Electrical Safety Test
The subject device meets the following standards:
-
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
-
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
-
- IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Berrcom" in a stylized, sans-serif font. The word is rendered in a dark blue color. The "B" in "Berrcom" is designed with a rounded shape that resembles a heart or a stylized leaf, adding a unique visual element to the logo. The rest of the letters are more conventional, but still maintain a rounded and modern appearance.
patibility Test
The subject device has passed biocompatibility tests in according to following standards.
-
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
-
- ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
7.3. Performance Test-Bench
The subject device has passed performance tests in according to following standards.
- ISO 80601-2-56:2017+AMD2018 Medical electrical equipment Part 2-56: Particular 1) requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
-
- ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
7.4. Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met, and all software hazards have been mitigated to acceptable risk levels.
8. Clinical Accuracy Validation
Clinical accuracy validation was conducted in according to ISO 80601-2-56:2017+AMD2018. This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consisting of 150 subjects, of which 50 subjects are infants, 50 subjects are children, and the rest 50 subjects are adults (NOTE: Infants---newborn to one year; Children-- greater than one to five years; Adults---greater than five years old.). The clinical validation results demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical validation protocol.
9. Conclusion
The subject devices Non-contact Infrared Thermometer (Models: JXB-315, JXB-319, JXB-320, JXB-311) is substantially equivalent to the predicate device (K180355). This conclusion is based upon comparison on intended use, technological characteristics and applicable safety standards. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.