K Number
K213013
Date Cleared
2022-09-19

(364 days)

Product Code
Regulation Number
880.5860
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

STERILE SINGLE USE LDV SYRINGE/CHOICARE STERILE SINGLE USE LDV SYRINGE is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.

Device Description

A Disposable Syringe with Needle is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. This is a single-use device. This product is packed by sterile paper and sterilized by E.O gas.

AI/ML Overview

This document is an FDA 510(k) Premarket Notification for a medical device (Sterile Single-use LDV Syringe, CHOICARE Sterile Single-use LDV Syringe). The provided text describes the device, its intended use, performance data, and a comparison to a predicate device to demonstrate substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

Important Note: The provided document is a 510(k) submission for a syringe, which is a relatively low-risk medical device. The types of acceptance criteria and studies are primarily focused on bench testing and adherence to existing international standards, rather than complex clinical studies or AI/software performance evaluations. Therefore, many of the requested points related to AI, MRMC studies, and expert ground truth establishment for complex image analysis are not applicable or detailed in this document.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by adherence to various ISO standards and specific quantitative thresholds for certain tests. The reported device performance is consistently "Pass" for all tested criteria, indicating the device met the acceptance criteria.

Requirement – Test (Standard)Acceptance Criteria (Implicit for "Pass")Reported Device Performance
ISO 7886-1:2017 (Syringe Performance)Conformance to standard requirements for manual use syringes (e.g., proper dimensions, tightness, piston/plunger assembly)Pass
Visual checkN/A (Visual inspection, implied to be defect-free)Pass
Dimension (out diameter of the needle tube)N/A (Within specified tolerance per standard)Pass
Length of the needle tubeN/A (Within specified tolerance per standard)Pass
NozzleN/A (Conforms to design and standard)Pass
TightnessN/A (No leakage at specified pressures)Pass
Piston/Plunger AssemblyN/A (Smooth movement, no binding)Pass
ISO 7864:2016 (Hypodermic Needle Performance)Conformance to standard requirements (e.g., elasticity, flexural strength, pullout)Pass
ElasticityN/A (Meets specified flexibility)Pass
Flexural strengthN/A (Withstands specified bending force)Pass
PulloutN/A (Withstands specified pullout force)Pass
ISO 9626:2016 (Stainless Steel Needle Tubing)Conformance to standard requirements (e.g., stiffness, resistance to breakage/corrosion)Pass
StiffnessN/A (Meets specified rigidity)Pass
Resistance to breakageN/A (Withstands specified forces without breaking)Pass
Resistance to corrosionN/A (Shows no significant corrosion after specified tests)Pass
Dead Volume)Satisfy the requirements for bacterial endotoxins
Particulate Matter Injection (USP )Satisfy the requirements for particulate matterPass
Sterility & EO ResidualsConformance to standard requirements (e.g., SAL of 10^-6, low EO residuals)Pass (for all listed tests)
LAL test (USP39 )Satisfy the requirements for bacterial endotoxinsPass
E.O sterilization validation (ISO 11135:2014)Process validated to achieve sterilityPass
Sterility test (ISO 11737-2)No microbial growthPass
E.O Residual test (ISO 10993-7:2008)Meets test requirements for ethylene oxide sterilization residualsPass
Needle Injury Test (ISO 23908:2011 Sharps Injury Protection)Conformance to standard requirements for sharps injury protection (e.g., acceptable penetration, pull-out, cap removal, activation/locking, unlocking forces)Pass (for all listed tests)
Needle penetration forceN/A (Within specified range)Pass
Pull-out forceN/A (Within specified range)Pass
Needle cap removal forceN/A (Within specified range)Pass
Activation (locking) force (for safety feature)N/A (Within specified range)Pass
Unlocking force (for safety feature)N/A (Within specified range)Pass

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the numerical sample sizes for each bench test conducted. However, for a 510(k) submission, testing is typically performed on a statistically relevant number of units as required by the referenced ISO standards.
  • Data Provenance: The document indicates that the tests were performed by Jeil Tech Co., Ltd. (Republic of Korea). The studies are "bench tests," which are laboratory tests, not clinical studies involving human patients. The nature is prospective in the sense that the tests were performed specifically for this submission to verify the device's adherence to standards and safety.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device and study. Bench testing for a syringe relies on objective measurements against engineering specifications and international standards (e.g., ISO, USP). There isn't "ground truth" to be established by human experts in the context of diagnostic interpretation or clinical outcomes. The "experts" involved would be qualified lab technicians and engineers performing and verifying the tests.


4. Adjudication Method for the Test Set

This is not applicable. Since the tests are objective bench measurements against quantitative or qualitative (Pass/Fail based on standard adherence) criteria, there is no need for an adjudication method. The results are either within specification or not.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

This is not applicable. This submission is for a physical medical device (syringe), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC comparative effectiveness study was performed or required.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This is not applicable. As stated above, this is not an AI/algorithm device.


7. The Type of Ground Truth Used

The "ground truth" for this device's performance is the adherence to established international standards (ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ISO 10993 series, ISO 11135, ISO 11737-2) and specific quantitative requirements (e.g., dead volume, endotoxin limits). This is an engineering specification and regulatory standard conformance type of ground truth.


8. The Sample Size for the Training Set

This is not applicable. There is no "training set" as this is not an AI or machine learning device. The device is manufactured based on design specifications, and tested to ensure it meets those specifications and relevant standards.


9. How the Ground Truth for the Training Set was Established

This is not applicable as there is no training set for this device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).