K Number
K212514

Validate with FDA (Live)

Date Cleared
2022-04-29

(262 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin.

The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.

Device Description

The devices are provided in two types of configurations: Safety Pen Needle for Single Use and Insulin Pen Needle.

The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of a needle cap, needle hub, safety protective cover, self-destruction seat and sealed paper. The safety protective cover is manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of protective cap, needle tube, needle protective cover and sealed paper.

AI/ML Overview

The provided document describes a 510(k) premarket notification for a medical device, specifically Safety Pen Needles for Single Use and Insulin Pen Needles. It focuses on demonstrating substantial equivalence to a predicate device (K181447) rather than presenting a study for general acceptance criteria alone.

Therefore, the response will be structured to address the acceptance criteria as derived from the comparative and standalone testing performed to establish substantial equivalence.

Here's an analysis based on the provided text:


1. Table of acceptance criteria and the reported device performance

The document lists numerous non-clinical tests and standards that the proposed device was tested against. The acceptance criteria are implicit in compliance with these standards and the comparisons to the predicate device. The performance is stated as "complied with" or "met the requirements".

Acceptance Criteria (Standard / Test)Reference Clause/StandardReported Device Performance
Biocompatibility
CytotoxicityISO 10993-5: 2009No cytotoxicity
Irritation/Intracutaneous ReactivityISO 10993-10: 2010No intracutaneous reactivity
SensitizationISO 10993-10: 2010No skin sensitization
Systemic ToxicityISO 10993-11: 2017No systemic toxicity
HemolysisASTM F756-17No Hemolysis
PyrogenUSP <151>No Pyrogen
Subacute ToxicityNot explicitly listed in table, but mentioned in textNo adverse effect
Sterilization & Endotoxin
EO Residue (EO, ECH)ISO 10993-7:2008Did not exceed limit
Bacterial Endotoxins LimitUSP <85>Did not exceed 20 EU/device
Sterility Assurance Level (SAL)Not explicitly listed in table, but mentioned in text10^-6
Packaging & Shelf Life
Seal Integrity (Visual)ASTM F1886 / F1886M-16Deemed acceptable
Seal StrengthASTM F88/F88M-15Complied with requirements
Dye Penetration (Seal Leaks)ASTM F1929-15Complied with requirements
Shelf Life (5 years)Physical, Mechanical, Chemical, Package TestsMaintain its performance
Physical & Mechanical (General Needles)
DimensionsISO 11608-2:2012 Cl. 4.2, ISO 9626:2016 Cl. 5.6Complied with requirements
Flow RateISO 11608-2:2012 Cl. 4.3Complied with requirements
Bond (Hub & Needle Tube)ISO 11608-2:2012 Cl. 4.4, ISO 7864:2016 Cl. 4.12Complied with requirements
Needle PointsISO 11608-2:2012 Cl. 4.5, ISO 7864:2016 Cl. 4.11Complied with requirements
Freedom from DefectsISO 11608-2:2012 Cl. 4.6Complied with requirements
LubricationISO 11608-2:2012 Cl. 4.7Complied with requirements
Dislocation of Measuring PointISO 11608-2:2012 Cl. 4.8Complied with requirements
Functional Compatibility with Pen SystemsISO 11608-2:2012 Cl. 4.9Complied with requirements
Ease of Assembly/DisassemblyISO 11608-2:2012 Cl. 4.10Complied with requirements
CleanlinessISO 7864:2016 Cl. 4.3, ISO 9626:2016 Cl. 5.3Complied with requirements
Acidity/AlkalinityISO 7864:2016 Cl. 4.4, ISO 9626:2016 Cl. 5.4Complied with requirements
Extractable MetalsISO 7864:2016 Cl. 4.5Complied with requirements
Size DesignationISO 7864:2016 Cl. 4.6, ISO 9626:2016 Cl. 5.5Complied with requirements
Color CodingISO 7864:2016 Cl. 4.7Complied with requirements
Needle HubISO 7864:2016 Cl. 4.8Complied with requirements
Needle CapISO 7864:2016 Cl. 4.9Complied with requirements
Needle TubeISO 7864:2016 Cl. 4.10Complied with requirements
Lumen PatencyISO 7864:2016 Cl. 4.13Complied with requirements
Surface Finish & AppearanceISO 9626:2016 Cl. 5.2Complied with requirements
StiffnessISO 9626:2016 Cl. 5.8Complied with requirements
Resistance to BreakageISO 9626:2016 Cl. 5.9Complied with requirements
Resistance to CorrosionISO 9626:2016 Cl. 5.10Complied with requirements
Particulate TestingUSP <788>Complied with requirements
Specific Safety Pen Needle Features
Safety Feature PerformanceISO 23908:2011 & FDA GuidanceMet pre-established criteria
Safety Feature Activation Forces(Compared to Predicate)Average at 3.71N (Predicate: 3.73N)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes for each test in the acceptance criteria. It broadly states "Non-clinical tests were conducted..." and "Physical, Mechanical, and Chemical testing listed in following table were performed on the proposed device."
Regarding data provenance, the manufacturer is Berpu Medical Technology Co., Ltd. in Wenzhou, Zhejiang, China. Therefore, the data provenance is China, and the data is prospective as it involves direct testing of the proposed devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the document describes non-clinical laboratory and performance testing for a medical device (pen needles), not a diagnostic algorithm requiring expert ground truth for interpretation of images or other data. The "ground truth" here is defined by compliance with established international and ASTM standards.

4. Adjudication method for the test set

This section is not applicable for the same reasons as point 3. The evaluation is based on objective measurements against specified standard criteria, not subjective expert judgment requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is a physical medical device (pen needle), not an AI-powered diagnostic tool, and therefore, no MRMC study or AI assistance evaluation was performed or is relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable as the device is a physical medical instrument, not an algorithm or software. The non-clinical tests described are "standalone" in the sense that they evaluate the device's intrinsic properties against standards, without human interaction influencing the core performance metrics being measured (e.g., flow rate, bond strength).

7. The type of ground truth used

The "ground truth" for the non-clinical tests and the product's performance is established by international and national standards (e.g., ISO, ASTM, USP) and the specifications derived from these standards. For example, the criteria for cytotoxicity or sterility assurance level (SAL) are defined by the respective ISO and USP standards. The "pre-established criteria" mentioned for the simulated clinical study and safety feature test would also fall under this category, likely derived from the relevant standards (ISO 23908:2011) and FDA guidance.

8. The sample size for the training set

This section is not applicable. The document describes the testing of a physical medical device. There is no "training set" in the context of an AI/ML algorithm. The training in device manufacturing typically refers to process validation and quality control, not data training.

9. How the ground truth for the training set was established

This section is not applicable for the same reason as point 8; there is no training set mentioned in the context of this device's approval process.


{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Berpu Medical Technology Co., Ltd. % Diana Hong General Management Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 China

Re: K212514

Trade/Device Name: Safety Pen Needle for Single Use, Insulin Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: March 23, 2022 Received: March 30, 2022

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K212514

Device Name Safety Pen Needle for Single Use Insulin Pen Needle

Indications for Use (Describe)

The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin.

The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K212514

    1. Date of Preparation: 04/28/2022
    1. Sponsor Identification

Berpu Medical Technology Co., Ltd. NO.14 Xingji Road, Yongxing Street, Longwan District, Wenzhou, Zhejiang, China 325000

Establishment Registration Number: 3004496829

Contact Person: Buxin Yu Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: ybx@berpu.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Safety Pen Needle for Single Use Insulin Pen Needle Common Name: Antistick needle and needle

Regulatory Information Classification Name: Needle, Hypodermic, Single Lumen; Classification: II; Product Code: FMI; Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital

Indication for use:

The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin.

The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.

Device Description

The devices are provided in two types of configurations: Safety Pen Needle for Single Use and Insulin Pen Needle.

The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of a needle cap, needle hub, safety protective cover, self-destruction seat and sealed paper. The safety protective cover is manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

GaugeLengthWall Type
34G4mmRegular wall, Thin wall, Extra-thin wall
34G5mmRegular wall, Thin wall, Extra-thin wall
34G6mmRegular wall, Thin wall, Extra-thin wall
34G8mmRegular wall, Thin wall, Extra-thin wall
33G4mmRegular wall, Thin wall, Extra-thin wall
33G5mmRegular wall, Thin wall, Extra-thin wall
33G6mmRegular wall, Thin wall, Extra-thin wall
33G8mmRegular wall, Thin wall, Extra-thin wall

{5}------------------------------------------------

32G4mmRegular wall, Thin wall, Extra-thin wall,Ultra-thin wall
5mmRegular wall, Thin wall, Extra-thin wall
6mmRegular wall, Thin wall, Extra-thin wall
8mmRegular wall, Thin wall, Extra-thin wall
31G4mmRegular wall, Thin wall, Extra-thin wall
5mmRegular wall, Thin wall, Extra-thin wall
6mmRegular wall, Thin wall, Extra-thin wall
8mmRegular wall, Thin wall, Extra-thin wall
30G4mmRegular wall, Thin wall, Extra-thin wall
5mmRegular wall, Thin wall, Extra-thin wall
6mmRegular wall, Thin wall, Extra-thin wall
8mmRegular wall, Thin wall, Extra-thin wall
29G4mmRegular wall, Thin wall, Extra-thin wall
5mmRegular wall, Thin wall, Extra-thin wall
6mmRegular wall, Thin wall, Extra-thin wall
8mmRegular wall, Thin wall, Extra-thin wall

The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of protective cap, needle tube, needle protective cover and sealed paper.

GaugeLengthWall Type
34G4mmRegular wall, Thin wall, Extra-thin wall
34G5mmRegular wall, Thin wall, Extra-thin wall
34G6mmRegular wall, Thin wall, Extra-thin wall
34G8mmRegular wall, Thin wall, Extra-thin wall
33G4mmRegular wall, Thin wall, Extra-thin wall
33G4mmRegular wall, Thin wall, Extra-thin wall
33G5mmRegular wall, Thin wall, Extra-thin wall
33G6mmRegular wall, Thin wall, Extra-thin wall
33G8mmRegular wall, Thin wall, Extra-thin wall
28G4mmRegular wall, Thin wall
28G5mmRegular wall, Thin wall
28G6mmRegular wall, Thin wall

{6}------------------------------------------------

8mmRegular wall, Thin wall
10mmRegular wall, Thin wall
13mmRegular wall, Thin wall

The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 106.

  • ડ. Identification of Predicate Device
    510(k) Number: K181447 Product Name: Safety insulin needle for single use

    1. Non-Clinical Test Conclusion
      Non-clinical tests were conducted to verify that the proposed device met all design specifications in order to demonstrate that the subject device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • A ISO 10993-5: 2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity

  • ISO 10993-10: 2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization.

  • ISO 10993-11: 2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity

  • ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

  • ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection

  • ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)

  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration

  • ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods

  • ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices

  • A ISO 11608-2 Second edition 2012-04-01, Needle-based injection systems for medical use -Requirements and test methods - Part 2: Needles

  • ISO 23908:2011 Sharps injury protection - Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling

  • A ISO 10993-7:2008 Biological evaluation of medical devices-Part 7: Test of Ethylene Oxide Residues.

  • USP<85> Bacterial Endotoxins Test A

  • USP<151> Pyrogen Test A

{7}------------------------------------------------

Physical, Mechanical, and Chemical testing listed in following table were performed on the proposed device. The test results show that the device conforms with the requirements of related standards.

DimensionsClause 4.2 of ISO 11608-2:2012
Determination of flow rate through the needleClause 4.3 of ISO 11608-2:2012
Bond between hub and needle tubeClause 4.4 of ISO 11608-2:2012
Needle pointsClause 4.5 of ISO 11608-2:2012
Freedom from defectsClause 4.6 of ISO 11608-2:2012
LubricationClause 4.7 of ISO 11608-2:2012
Dislocation of measuring point at patient endClause 4.8 of ISO 11608-2:2012
Determination of functional compatibility with needle-based injection systemsClause 4.9 of ISO 11608-2:2012
Ease of assembly and disassemblyClause 4.10 of ISO 11608-2:2012
CleanlinessClause 4.3 of ISO 7864:2016
Limits for acidity or alkalinityClause 4.4 of ISO 7864:2016
Limits for extractable metalsClause 4.5 of ISO 7864:2016
Size designationClause 4.6 of ISO 7864:2016
Colour codingClause 4.7 of ISO 7864:2016
Needle hubClause 4.8 of ISO 7864:2016
Needle CapClause 4.9 of ISO 7864:2016
Needle tubeClause 4.10 of ISO 7864:2016
Needle pointClause 4.11 of ISO 7864:2016
Bond between hub and needle tubeClause 4.12 of ISO 7864:2016
Patency of lumenClause 4.13 of ISO 7864:2016
Surface finish and appearanceClause 5.2 of ISO 9626:2016
CleanlinessClause 5.3 of ISO 9626:2016
Limits for acidity and alkalinityClause 5.4 of ISO 9626:2016
Size designationClause 5.5 of ISO 9626:2016
DimensionsClause 5.6 of ISO 9626:2016
StiffnessClause 5.8 of ISO 9626:2016
Resistance to breakageClause 5.9 of ISO 9626:2016
Resistance to corrosionClause 5.10 of ISO 9626:2016
EO residueISO 10993-7:2008
ECH residueISO 10993-7:2008
Bacteria Endotoxin LimitUSP <85>
Shelf Life EvaluationPhysical, Mechanical, Chemical, Package Testswere performed on aging samples to verify theclaimed shelf life of the device

Particulate testing

USP <788>

{8}------------------------------------------------

Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection and sterility maintenance.

Sterile barrier package integrity testing was performed on the proposed devices which includes visual inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15), and dye penetration testing (ASTM F1929-15).

Sterilization and shelf life testing, listed in following table, were performed on the proposed devices. EO ECH residue did not exceed the limit of ISO 10993-7. Endotoxin limit did not exceed 20EU/device. Shelf life testing results showed that the device can maintain its performance during the claimed shelf life of 5 years.

Biocompatibility testing

The contact classification of the proposed devices are Externally Communicating - Blood path, indirect for a Prolonged contact duration. The proposed devices were evaluated for the following tests. The results for the biocompatibility testing showed that there are no negative impacts from the materials that are used in the proposed devices.

  • Cytotoxicity,

  • Sensitization,

  • A Intracutaneous reactivity,
  • Acute Systemic Toxicity,

  • Hemolysis,

  • Pyrogen,

  • Subacute Toxicity

Simulated Clinical Study

A simulated clinical study was performed on proposed device according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005, and ISO 23908:2011 to evaluate the safety mechanism of the proposed device. The results

{9}------------------------------------------------

demonstrated that the proposed device met the pre-established criteria.

Safety Feature Test

The safety feature test was performed on the Safety Pen Needle for Single Use and predicate device to evaluate safety feature. The results demonstrated that both the proposed device and predicate device met the acceptance criteria.

    1. Clinical Test Conclusion
      No clinical study is included in this submission.

{10}------------------------------------------------

8. Summary of Technology Characteristics

Table 1 Comparison of Safety Pen Needle for Single Use
ITEMProposed DevicePredicate DeviceRemark
K212514K181447
ProductSafety Pen Needle for Single UseSafety insulin needle for single use/
Product CodeFMIFMISame
Regulation No.21 CRF 880.557021 CRF 880.5570Same
ClassClass IIClass IISame
Indication forUseThe Safety Pen Needle for Single Useis intended for use with pen injectordevices for the subcutaneous injectionof insulin.The Safety insulin needle for singleuse is intended for use with peninjector devices for the subcutaneousinjection of insulin.Same
ConfigurationNeedle tubeNeedle TubeDifferent
Needle hubHub
Safety protective coverSafety protective cover
Self-destruction seatSelf-destruction seat
Hub sheathHub sheath
Sealed paperSealed paper
/Safety seat
/Spring
Operation ModeFor manual use onlyFor manual use onlySame
Environment ofuseIn hospital or in the homeenvironmentIn hospital or in the homeenvironmentSame
Method ofattachment topen injectorThrough threaded connectionThrough threaded connectionSame
Safety FeaturePrevent from needlestickPrevent from needlestickSame
Safety featureactivation forcesAverage at 3.71NAverage at 3.73NSimilar
Single UseSingle UseSingle UseSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Needle Gauge29G, 30G, 31G, 32G, 33G, 34G29G, 30G, 31G, 32G, 33G, 34GSame
Needle Length4mm, 5mm, 6mm, 8mm4mm, 5mm, 6mm, 8mmSame
NeedlePerformanceComplied withISO 7864,ISO 9626,ISO 11608-2Complied withISO 7864,ISO 9626,ISO 11608-2Same
Material
Table 1Comparison of Safety Pen Needle for Single Use
---------------------------------------------------------

{11}------------------------------------------------

Safety
protective coverPolypropylene (PP)PolypropyleneDifferent
Needle hubPolypropylene (PP)Polypropylene
Self-destructionPolypropylene (PP)MABS
seat
Hub sheathPolypropylene (PP)
Needle tubeStainless Steel (SUS304)
Sealed paperMedical paper (Blister Paper)
AdhesiveUV adhesive
LubricantPolydimethylsiloxane
Biocompatibility
CytotoxicityNo cytotoxicityNo cytotoxicitySame
IrritationNo intracutaneous reactivityNo intracutaneous reactivity
SensitizationNo sensitizationNo skin sensitization
SystemicNo systemic toxicityNo systemic toxicity
Toxicity
HemolysisNo HemolysisNo Hemolysis
PyrogenNo PyrogenNo Pyrogen
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame

Different Analysis - Configuration

The configuration of the Safety Pen Needle for Single Use is similar to the configurations of predicate device. Although the components of the proposed device and the predicate device are not identical, this difference does not affect the indication for use of the device. In addition, a series of performance tests have been conducted on the proposed device to demonstrate that the proposed device meets the requirements of the relevant standards. Therefore, it can be considered that the minor differences in the components will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Similar Analysis - Safety feature activation forces

The activation force of the Safety Pen Needle for Single Use is similar to the predicate device. The activation force difference between the two devices is 0.02N. In addition, activation force met the acceptance criteria. Therefore, the minor difference in activation force will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Different Analysis -Material

The materials for Safety Pen Needle for Single Use is different from the predicate device. However, biocompatibility testing has been performed on the proposed device and the results do not show any

{12}------------------------------------------------

adverse effect. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

ITEMProposed DevicePredicate DeviceRemark
K212514K181447
ProductInsulin Pen NeedleSafety insulin needle for single use/
Product CodeFMIFMISame
Regulation No.21 CRF 880.557021 CRF 880.5570Same
ClassClass IIClass IISame
Indication for UseThe Insulin Pen Needle is intendedfor use with pen injector devices forthe subcutaneous injection of insulin.The Safety insulin needle for singleuse is intended for use with peninjector devices for the subcutaneousinjection of insulin.Same
ConfigurationNeedle tubeNeedle hubHub sheathNeedle protective coverSealed paper///Needle TubeHubSafety protective coverSelf-destruction seatHub sheathSealed paperSafety seatSpringDifferent
Operation ModeFor manual use onlyFor manual use onlySame
Environment of useIn hospital or in the homeenvironmentIn hospital or in the homeenvironmentSame
Method ofattachment topen injectorThrough threaded connectionThrough threaded connectionSame
Single UseSingle UseSingle UseSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Needle Gauge28, 33G, 34G29G, 30G, 31G, 32G, 33G, 34GDifferent
Needle Length4mm, 5mm, 6mm, 8mm, 10mm,13mm4mm, 5mm, 6mm, 8mmDifferent
NeedlePerformanceComplied withISO 7864,ISO 9626,ISO 11608-2Complied withISO 7864,ISO 9626,ISO 11608-2Same
Material
Needle hubPolypropylene (PP)PolypropyleneDifferent
Hub sheathPolypropylene (PP)304 Stainless Steel
Sealed paperMedical paper (Blister Paper)Polypropylene
Needleprotective coverPolypropylene (PP)MABS
Needle tubeStainless Steel (SUS304)
AdhesiveUV adhesive
LubricantPolydimethylsiloxane
Biocompatibility
CytotoxicityNo cytotoxicityNo cytotoxicitySame
IrritationNo intracutaneous reactivityNo intracutaneous reactivitySame
SensitizationNo sensitizationNo skin sensitizationSame
SystemicToxicityNo systemic toxicityNo systemic toxicitySame
HemolysisNo HemolysisNo HemolysisSame
PyrogenNo PyrogenNo PyrogenSame
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame

Table 2 Comparison of Insulin Pen Needle

{13}------------------------------------------------

Different Analysis -Configuration

The components of Insulin Pen Needle are different to that of predicate device. The proposed device does not have a safety mechanism; however, the presence or absence of a safety mechanism does not affect the indication for use and the performance of the device. In addition, a series of performance tests have been conducted on the proposed device and the results showed that the product meets the requirements of the relevant standards. Therefore, this difference does not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Different Analysis – Needle Gauge

The needle gauge of Insulin Pen Needle is different from predicate devices. The proposed 33G and 34G can be covered by the predicate device, while the proposed 28G is out of the specification range of predicate device. However, all the needle size of proposed device has been tested. The test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference does not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Different Analysis - Needle length

Compared with the predicate device, the Insulin Pen Needle has the additional 10mm and 13mm length specifications. However, these two additional length specifications have been tested. The test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference does not affect the Substantial Equivalence (SE) between the proposed and predicate device.

{14}------------------------------------------------

Different Analysis -Material

The material for Safety Pen Needle for Single Use is different from the predicate device. However, biocompatibility testing has been performed on the proposed device and the results do not show any adverse effect. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

    1. Substantially Equivalent (SE) Conclusion
      The non-clinical tests demonstrate that the Safety Pen Needle for Single Use and Insulin Pen Needle are Substantially Equivalent (SE) to the legally marketed predicate device cleared under K181447.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).