K Number
K181447
Date Cleared
2019-03-12

(284 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.

Device Description

The proposed device, Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of needle tube, hub, safety protective cover, self-destruction seat, spring, hub sheath, safety seat and sealed paper.

The Safety insulin needle for single use is offered in various gauge sized and length.

The proposed device is available in EO sterilized to achieve a Sterility Assurance Level (SAL) of 10°.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (Safety insulin needle for single use) seeking substantial equivalence to a predicate device. It details non-clinical tests conducted to support this claim, but it does not contain information about acceptance criteria or a study proving the device meets those criteria in the way typically associated with performance metrics like sensitivity, specificity, or accuracy for an AI/diagnostic device.

The "acceptance criteria" discussed are compliance with various ISO and ASTM standards and the "study" is a compilation of non-clinical tests demonstrating adherence to these standards and the substantial equivalence to a predicate device. There is no AI component, nor is there a multi-reader multi-case study, stand-alone algorithm performance, or human-in-the-loop analysis mentioned.

Let's break down what is available in the text according to your requested categories:

1. A table of acceptance criteria and the reported device performance

The provided document doesn't present acceptance criteria in the form of performance metrics (like sensitivity/specificity) for a diagnostic device. Instead, it refers to compliance with established international standards for medical devices and biocompatibility.

Acceptance Criteria CategorySpecific Criteria/Standard (Implicit Acceptance)Reported Device Performance
Mechanical PerformanceISO 7864:2016 (Sterile hypodermic needles for single use)Complied with (implicitly, tested and met requirements)
ISO 9626:2016 (Stainless steel needle tubing)Complied with (implicitly, tested and met requirements)
ISO 11608-2:2012 (Needle-based injection systems)Complied with (implicitly, tested and met requirements)
Material PerformanceISO 10993-1, -4, -5, -10, -11 (Biological evaluation of medical devices)Complied with (implicitly, patient-contact materials conform to series of standards)
ASTM F88/F88M-15 (Seal Strength of Flexible Barrier Materials)Complied with (implicitly, tested and met requirements)
ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Packaging)Complied with (implicitly, tested and met requirements)
USP <85> Bacterial Endotoxins TestComplied with (No pyrogen, Endotoxin Limit 20EU per device)
BiocompatibilityCytotoxicity standards (e.g., ISO 10993-5)No cytotoxicity
Irritation standards (e.g., ISO 10993-10)No intracutaneous reactivity
Sensitization standards (e.g., ISO 10993-10)No skin sensitization
Systemic Toxicity standards (e.g., ISO 10993-11)No systemic toxicity
Hemolysis standards (e.g., ISO 10993-4)No Hemolysis
Pyrogen standards (e.g., USP <85>)No Pyrogen (Endotoxin Limit 20EU per device)
SterilizationISO 11135 (Ethylene Oxide Sterilization)Sterilization parameter established to achieve SAL of 10^-6
ISO 10993-7 (Ethylene Oxide Sterilization Residuals)Complied with (implicitly, tested and met requirements)
SAL (Sterility Assurance Level)10^-6
OtherSingle Use confirmationYes
Labeling compliance (21 CFR part 801)Complied with

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for individual tests. It states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications" and that "all the needle sizes of proposed device have been tested" and "all the needle lengths of the proposed device have been validated and tested". The data provenance is not explicitly stated beyond the manufacturer being Berpu Medical Technology Co., Ltd in China. These are laboratory/physical tests, not clinical data sets in the typical sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device is a physical medical instrument (safety insulin needle), not an AI/diagnostic software. Ground truth in this context would refer to objective measurements of physical properties and biological responses according to established standards, not expert interpretation of data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The device is a physical safety insulin needle. There is no AI component or human reader interaction in the way an MRMC study would assess.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" is defined by the requirements and test methods outlined in the cited ISO and ASTM standards. This involves objective measurements of physical properties (e.g., needle dimensions, seal strength), chemical residues (e.g., EO residuals), and biological responses (e.g., cytotoxicity tests in cell cultures, systemic toxicity in animal models per ISO 10993 series).

8. The sample size for the training set

Not applicable. There is no mention of a training set as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Berpu Medical Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 Cn

Re: K181447

Trade/Device Name: Safety insulin needle for single use Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: February 7, 2019 Received: February 12, 2019

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sarah B. Mollo -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181447

Device Name Safety insulin needle for single use

Indications for Use (Describe)

The safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K181447

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

    1. Date of Preparation: 01/23/2019
    1. Sponsor Identification

BERPU MEDICAL TECHNOLOGY CO., LTD NO.14 Xingji Road, Yongxing Street, Longwan District, 325000, Wenzhou, Zhejiang Province

Establishment Registration Number: 3004496829

Contact Person: Jundong Tan Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: BERPU@BERPU.COM

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Safety insulin needle for single use

Regulatory Information: Classification Name: Hypodermic single lumen needle Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital

Intended Use Statement:

The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.

Indications for Use:

The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.

Device Description

The proposed device, Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of needle tube, hub, safety protective cover, self-destruction seat, spring, hub sheath, safety seat and sealed paper.

The Safety insulin needle for single use is offered in various gauge sized and length.

The proposed device is available in EO sterilized to achieve a Sterility Assurance Level (SAL) of 10°.

    1. Identification of Predicate Device
      510(k) Number: K152514 Product Name: Clickfine AutoProtect Pen Needle

Non-Clinical Test Conclusion 6.

Non-clinical tests were conducted to verify that the proposed device met all design specifications in order to demonstrate that the subject device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{5}------------------------------------------------

  • ISO 7864:2016, Sterile hypodermic needles for single use.

  • ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices.

  • ISO 11608-2 Second edition 2012-04-01, Needle-based injection systems for medical use -Requirements and test methods - Part 2: Needles

  • ISO 10993-1:2009 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process

  • ISO 10993-4:2017, Biological Evaluation of Medical Device, Part 4: Selection of test for Interactions with Blood.

  • ISO 10993-5: 2009(R), Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.

  • ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

  • ISO 10993-11:2006 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity

  • ISO 10993-7:2008 Biological Evaluation of Medical Devices-Part 7: Ethylene Oxide Sterilization Residuals.

  • ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials. >
  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

  • ♪ USP <85> Bacterial Endotoxins Test
    1. Clinical Test Conclusion

No clinical study is included in this submission.

Substantially Equivalent (SE) Comparison 8.

Table 1 General Comparison

ItemProposed DevicePredicate DeviceK152514Remark
Product CodeFMIFMISE
Regulation Number21 CFR 880.557021 CFR 880.5570SE
ClassIIIISE
Indications for UseThe Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.The Clickfine AutoProtect Pen Needle is intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needleSEAnalysis 1

{6}------------------------------------------------

sticks from the patient end ofthe needle. The shield alsoserves to hide the needle beforeand after injection.
ConfigurationNeedle TubeNeedle TubeSE
HubHubAnalysis 2
Safety protective coverNeedle shield
Self-destruction seatHousing
SpringSpring
Hub sheathOuter protective container
Safety seatSafety lock indicator
Sealed paperPeel tab
Operation ModeFor manual use onlyFor manual use onlySE
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801SE

SE Analysis 1 - Indications for Use

The Indications for Use of the proposed device are different to that of predicate device. The difference is that the Indications for Use of predicate device include subcutaneous injection of exenatide while the proposed device does not. However, the indications for use of proposed device are completely included by those of the predicate. Additionally, the attached safety shield has the same role during the clinical application as that of the predicate device, although it is not expressed in the indications for use of the proposed device. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

SE Analysis 2 - Configuration

The components name of proposed device are different to that of the predicate, however, the components share the same configuration and function. Therefore, this difference in name will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

ItemProposed DevicePredicate DeviceK152514Remark
Needle Gauge29G, 30G, 31G, 32G,33G, 34G29G, 30G, 31GAnalysis 3
Needle Length4mm, 5mm, 6mm,8mm5mm, 6mm, 8mmAnalysis 4
Wall typeThin-walledExtra-thin-walledRegular walledThin-walledAnalysis 5
Needle PerformanceComplied withISO 7864:2016,Complied withISO 7864: 1993SE
Table 2 Performance Comparison

{7}------------------------------------------------

ISO 9626:2016,ISO 9626: 1991
ISO 11608-2ISO 11608-2

Analysis 3 - Needle Gauge

The subject device includes higher needle gauges than that of predicate devices. This difference in needle size will not raise new questions of safety and effectiveness. In addition, all the needle sizes of proposed device have been tested and met ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Analysis 4 - Needle length

The proposed device has an additional needle length than the of predicate device. This difference in needle size will not raise new questions of safety and effectiveness. In addition, all the needle lengths of the proposed device have been validated and tested and met the ISO 9626 standards requirements. Therefore, this difference will not affect the Substantially Equivalency (SE) between the proposed and predicate device.

Analysis 5 - Wall type

The wall thickness of the proposed device is different than the predicate device. the needles including both wall thickness (i.e. thin-walled and extra-thin-walled) of proposed device have been tested and the test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

ItemProposed DevicePredicate DeviceK152514Remark
Patient-contact material
HubPolypropyleneStainless SteelSE
Needle tube304 Stainless SteelPolypropyleneAnalysis 6
Safety seatPolypropylenePolypropylene
Safety protective coverMABSMBS
Biocompatibility
CytotoxicityNo cytotoxicityConforms to ISOSE
IrritationNo intracutaneousreactivity10993series standardAnalysis 6
SensitizationNo skin sensitization
Systemic ToxicityNo systemic toxicity
HemolysisNo Hemolysis
PyrogenNo Pyrogen
SterilizationEO SterilizationGamma irradiationSE Analysis 7

Table 3 Safety Comparison

{8}------------------------------------------------

SAL10-610-6SE
Single UseYesYesSE
Endotoxin Limit20EU per device20EU per deviceSE

Analysis 6 - Patient-contact Material and Biocompatibility

Although the material of proposed and the predicate device is different, the patient-contact material of the proposed device material conforms to the ISO 10993 series of standards. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

Analysis 7 - Sterilization

The proposed and predicate devices are sterilized by different method. However, the sterilization parameter for proposed device was established per ISO11135 to achieve the SAL of 10-6. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.

9. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis of differences above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).