K Number
K211752

Validate with FDA (Live)

Date Cleared
2022-01-21

(228 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Infrared Thermometer is a reusable thermometer intended for the intermittent determination of human body temperature in a non-contact mode using the center of the forehead as the measurement site on people of all ages. It can be used by consumers in the household environment and by healthcare provider.

Device Description

The infrared thermometer uses infrared temperature sensor to detect infrared energy radiated from the forehead. The intensity of the emitted energy depends on the temperature of the object. The infrared probe can recognize the emitted energy and transfer it to an electronic signal. The electronic signal can be processed in the Infrared Thermometer to convert to a temperature reading, which is displayed on the screen. This is a continuous non-invasive method of measuring body temperature. After measuring the forehead temperature with an infrared temperature sensor, the calculated temperature value is obtained. The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization. The device is for over-the-counter use.

The Infrared Thermometer, Model CFT-308, consists of the following parts:
a) Infrared sensor
b) Integrated circuit
c) Microprocessor (with Bluetooth function) (Bluetooth transmission is optional)
d) OLED display
e) Buzzer
f) 2 Buttons ("Start" button, "Measuring" button)
g) Alkaline batteries; size AAA, 2 x 1.5 V
h) Lens

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Infrared Thermometer (K211752):

Note: The provided document is a 510(k) summary, which often provides a high-level overview rather than granular detail for all study aspects. Some specific information (like the exact number and qualifications of experts for ground truth establishment within the clinical study, or specific effect sizes for MRMC studies if applicable) are not explicitly stated in this type of submission.


Acceptance Criteria and Reported Device Performance

The clinical study for the Infrared Thermometer Model CFT-308 aimed to demonstrate accuracy compared to a reference thermometer. The accuracy criteria are derived from the referenced standards, specifically ISO 80601-2-56:2017 and ASTM E1965-98:2016.

Acceptance Criteria (from referenced standards)Reported Device Performance (from "Accuracy" row in Table II-1 and clinical study)
ISO 80601-2-56:2017 & ASTM E1965-98:2016 Accuracy Requirements:
Laboratory Accuracy:
±0.2°C (for 35.0°C ~ 42.0°C)±0.2°C: 35.0 ~ 42.0°C
±0.3°C (for 32.0-34.9°C and 42.1-43.0°C)±0.3°C: 32.0~34.9°C, ±0.3°C: 42.1 ~ 43.0°C
±0.4°F (for 95°F~107.6°F)±0.4°F for 95°F~107.6°F
±0.5°F (for ranges outside 95°F~107.6°F)Outside that range ±0.5°F
Clinical Accuracy: (Demonstrated through comparison with a reference thermometer, ensuring the subject device meets its declared accuracy range)The clinical study results showed that the measured temperature of the subject device (CFT-308) met its declared accuracy range when compared to the reference armpit thermometer.

Important Note on "Reported Device Performance": The table above directly extracts the claimed accuracy from the comparison table (Table II-1) within the 510(k) summary, which represents the device's declared accuracy. The clinical study was performed to validate this declared accuracy. The document states, "Through clinical trials, the measured temperature of subject device and armpit thermometers are compared to demonstrate the measurement accuracy of subject device meets its declared accuracy range."


Study Details

  1. Sample Size used for the test set and the data provenance:

    • Sample Size: 150 subjects completed the clinical trial.
      • 15 subjects aged 0 to 3 months.
      • 35 subjects aged 3 months to 1-year-old.
      • 50 subjects aged from 1 to 5 years old.
      • 50 subjects aged 5 years and above.
    • Data Provenance: Prospective clinical study conducted in China ("second hospital of Tianjin medical university").
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document implies that the ground truth was established by a "reference thermometer (Model Number: CRW-23)" which measured "armpit test temperature." It does not explicitly state that human experts were used to establish this ground truth beyond operating the reference thermometer. For an infrared thermometer, the reference standard is typically another calibrated thermometer (often an invasive one like an oral or rectal thermometer in some studies, or an axillary/armpit thermometer as used here) rather than expert consensus on a subjective measure. Therefore, "experts" in the traditional sense for ground truth labeling (e.g., radiologists interpreting images) are not directly applicable or specified for this type of device. The accuracy of the reference thermometer itself would be implicitly relied upon.
  3. Adjudication method for the test set:

    • This is not applicable or specified for this type of device where the ground truth is established by a reference thermometer, not by subjective human interpretation requiring consensus.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This device is an infrared thermometer, not an AI-assisted diagnostic tool interpreted by human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, for clinical accuracy, the study evaluates the performance of the device (CFT-308) itself in measuring temperature compared to a reference standard (CRW-23). While a human operates the device, the performance being assessed is the device's ability to accurately measure temperature relative to the ground truth, effectively a standalone performance assessment of the measurement component.
  6. The type of ground truth used:

    • The ground truth was established using a reference thermometer (Model Number: CRW-23) measuring armpit temperature. This represents a form of empirical measurement/reference standard, aiming to provide an accurate physiological temperature reading against which the non-contact forehead thermometer's readings are compared.
  7. The sample size for the training set:

    • The document does not specify a separate "training set" sample size. The clinical study described is for validation/testing. Infrared thermometers are typically calibrated during manufacturing, and their algorithms are based on physics and laboratory calibration rather than a machine learning "training set" in the common sense for AI algorithms.
  8. How the ground truth for the training set was established:

    • As no explicit "training set" for an AI algorithm is mentioned, this question is not directly applicable. The device's underlying physics and pre-determined calibration (which leverages established temperature standards) would form its operational basis.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

January 21, 2022

Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Official Correspondent 2nd Floor 3rd Floor and Room 410-412 4th Floor No. 2 Building No. 9 Shuangyuan Road Shijingshan District, 100041 Beijing, China

Re: K211752

Trade/Device Name: Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 14, 2021 Received: December 22, 2021

Dear Haiying Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211752

Device Name Infrared Thermometer

Indications for Use (Describe)

Infrared Thermometer is a reusable thermometer intentittent determination of human body temperature in a non-contact mode using the center of the forehead as the measurement site on people of all ages. It can be used by consumers in the household environment and by healthcare provider.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section II 510(k) Summary

This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.

There is no prior submission for the device.

2.1 Submitter Information

Manufacturer Name: ●

Establishment Registration Number: 3005569927

Beijing Choice Electronic Technology Co., Ltd.

2nd Floor 3rd Floor and Room 410-412 4th Floor No. 2 Building, No. 9 Shuangyuan Road Shijingshan District 100041 Beijing PEOPLE'S REPUBLIC OF CHINA

● Contact Person:

Haiying Zhao

Beijing Choice Electronic Technology Co., Ltd.

2nd Floor 3rd Floor and Room 410-412 4th Floor No. 2 Building, No. 9 Shuangyuan Road Shijingshan District 100041 Beijing PEOPLE'S REPUBLIC OF CHINA

Phone: +86-10-88204631

Fax: +86-10-88204632

Email: cc@choicemmed.com

  • Date prepared: May 25, 2021

2.2 Subject Device Information

Regulation Name: Clinical Electronic Thermometer

Device Trade/Proprietary Name: Infrared Thermometer

Model: CFT-308

Purpose of submission: 510 (k)

{4}------------------------------------------------

Premarket Notification 510(k) Submission—Section II 510(k) Summary Regulation Number: 21 CFR 880.2910

Product Code: FLL

Class: II

Panel: General Hospital

2.3 Predicate Device

510(k) Number: K191829

Regulation Name: Clinical Electronic Thermometer

Device Trade/Proprietary Name: Microlife Non-Contact Infrared Forehead Thermometer

Model: FR1DG1 (NC200)

Product Code: FLL

Regulation Number: 21 CFR 880.2910

Device Class: II

Panel: General Hospital

Manufacturer: Microlife Intellectual Property GmbH, Switzerland

2.4 Device Description

The infrared thermometer uses infrared temperature sensor to detect infrared energy radiated from the forehead. The intensity of the emitted energy depends on the temperature of the object. The infrared probe can recognize the emitted energy and transfer it to an electronic signal. The electronic signal can be processed in the Infrared Thermometer to convert to a temperature reading, which is displayed on the screen.

This is a continuous non-invasive method of measuring body temperature. After measuring the forehead temperature with an infrared temperature sensor, the calculated temperature value is obtained.

The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.

The device is for over-the-counter use.

{5}------------------------------------------------

Premarket Notification 510(k) Submission—Section II 510(k) Summary

The Infrared Thermometer, Model CFT-308, consists of the following parts:

  • a) Infrared sensor
  • b) Integrated circuit
  • c) Microprocessor (with Bluetooth function) (Bluetooth transmission is optional)
  • d) OLED display
  • e) Buzzer
  • f) 2 Buttons ("Start" button, "Measuring" button)
  • g) Alkaline batteries; size AAA, 2 x 1.5 V
  • h) Lens

{6}------------------------------------------------

2.5 Comparison list of the technological characteristics

Table II-1 Performance Specification Comparison Table between the Subject Device and Predicate Device

Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
ItemInfrared ThermometerMicrolife Non-Contact Infrared ForeheadThermometer-
ModelCFT-308FR1DG1 (NC200)-
Regulation No.21 CFR 880.291021 CFR 880.2910
ClassificationIIII
Regulation NameClinical Electronic ThermometerClinical Electronic Thermometer
Product CodeFLLFLL
Indications for useInfrared Thermometer is a reusablethermometer intended for the intermittentdetermination of human body temperature ina non-contact mode using the center of theforehead as the measurement site on people ofall ages. It can be used by consumers in thehousehold environment and by healthcareprovider.The Microlife Non-Contact InfraredForehead Thermometer, Model FR1DG1(NC200) is intended for the intermittentmeasurement and monitoring of humanbody temperature. The device is indicatedfor use by people of all ages in the home.Difference1
Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
Componentsa) Infrared sensorb) Integrated circuitc) Microprocessor (with Bluetooth function)(Bluetooth transmission is optional)d) OLED displaye) Buzzerf) 2 Buttons (“Start” button, “Measuring”button)g) Alkaline batteries; size AAA, 2 x 1.5 Vh) Lensa) Thermopile Sensorb) Application-Specific IntegratedCircuitryc) Erasable Programmable Read-OnlyMemory Integrated Circuitd) Capacitance-touch Integrated Circuite) LCD and Backlightf) 3 buttons (“START/POWER” button,“M” button, “MODE” button)g) Alkaline batteries; size AAA, 2 x 1.5 Vh) LensDifference2
Performance SpecificationThermometertypeInfrared thermometerInfrared thermometer
DeviceMeasurementTechnologyNon-ContactNon-Contact
InfraredInfrared
TemperatureMeasurementdistanceAppropriate within 1 cmAppropriate within 5 cmDifference3
Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
Measuringlocation(human)ForeheadForehead
Physicaldimension1513850.2 mm156.7 * 43 * 47 mm-
Power supplyAlkaline batteries; size AAA, 2 x 1.5 VAlkaline batteries; size AAA, 2 x 1.5 V
Display resolution0.1°C or 0.1°F0.1°C or 0.1°F
Measuring range32.0°C ~ 43.0°C (89.6°F~109.4°F)32.0 ~ 43.0°C (89.6-109.4 °F)
Accuracy±0.2°C: 35.0 ~ 42.0°C±0.3°C: 32.034.9°C,±0.3°C: 42.1 ~ 43.0°C0.4°F for 95°F107.6°FOutside that range ±0.5°F±0.2°C: 35.0 ~ 42.0°C±0.3°C: 34.0 ~ 34.9°C±0.3°C: 42.1 ~ 43.0°C±0.4 °F: 95.0 ~ 107.6 °F±0.5 °F: 93.2 94.8 °F,107.8109.4 °F
Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
Response Time1 second3 secondDifference4
Shelf life5 years5 years
OperatingconditionsBody mode:59°F104°F (15°C40°C),Relative humidity: 15%~93%(non-condensing)Body mode:1540°C (59°F104°F),15-95% relative maximum humidityDifference5
Storageconditions13°F~158°F (-25℃~70℃)Relative humidity ≤93% (non-condensing) instorage/transport-2555°C(-13°F131°F)15-95% relative maximumhumidityDifference5
ReferencesiteArmpitOralDifference6
DisplayOLED displayLCD displayDifference7
Date, time, andbeeper settingYesYes
Memory10 sets memories30 sets memoriesDifference8

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
AutomaticSwitch-offThe thermometer shall automatically shut offwithout manipulating in 30 seconds.Approx. 1 minute after last measurementhas been takenDifference9
Beeper indicationYesYes
Age settingSelectable between three age ranges:0 ~ 3 months (Screen display 0-3 Months)3 ~ 36 months (Screen display 3-36 Months)>36 months (Screen display 36+ Months)None
Fever remindIf the individual temperature being measuredis in the fever range, four “beeps” will beheard and upset face icon will appear onthe screenDifferent fever ranges at different ages asbelow:Age range Above Normal 0 ~ 1 months >37.4°C(>99.4°F) 1 ~ 3 months >37.4°C(>99.4°F) 3 ~ 36 months >37.6°C(>99.6°F)10 short beeps and a red LCD backlightalerts that the temperature equal to orhigher than 37.5°CDifference10

{11}------------------------------------------------

Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
>36 months>37.7°C(>99.9°F)
Clinical StudySupportYes. Clinical test reportYes. Clinical test report
ContactingMaterialBattery CoverEnclosureABSABS
Performance TestingLaboratoryTestingISO80601-2-56:2017ASTM E1965-98:2016ISO80601-2-56:2017ASTM E1965-98:2016
Electrical SafetyIEC60601-1:2020IEC60601-1-11:2020IEC60601-1:2012,IEC60601-1-11:2015Difference11
EMC andElectricalSafeElectromagneticCompatibilityIEC60601-1-2:2014FCC certificationANSI C63.27-2017IEC60601-1-2:2014Difference12
BiocompatibilityBiocompatibilityTestingISO 10993-1:2018ISO 10993-5:2009ISO 10993-10:2010ISO 10993-1:2018ISO 10993-5:2009ISO 10993-10:2010
Comparison ElementsSubject DevicePredicate DeviceSimilar orDifferent
Software ValidationFDA November 2005 document “Guidancefor the Content of Premarket Submissions forSoftware Contained in Medical Devices”.FDA November 2005 document“Guidance for the Content of PremarketSubmissions for Software Contained inMedical Devices”.
Risk ManagementRisk Management in Compliance withISO14971:2019Risk Management in Compliance withISO14971:2007Difference13

{12}------------------------------------------------

{13}------------------------------------------------

Premarket Notification 510(k) Submission—Section II 510(k) Summary

  • Difference 1: The intended uses of subject device and predicate device are similar. They are both used for intermittent measurement on the forehead in a no touch mode, and are intended to be used by all ages. The use environment of subject device and predicate device are home and medical setting. While there are some minor differences, these do not raise different questions of safety and effectiveness. The device has been tested in accordance with the appropriate standards.
  • . Difference 2: The subject device and the predicate device have the similar components. The subject device has the function of Bluetooth transmission (Bluetooth transmission is optional). The safety and effectiveness of the Bluetooth function was tested per FCC Part 15B and Part 15C and ANSI C63.27-2017 American National standard for evaluation of wireless coexistence.
  • Difference 3: Subject device and predicate device have similar measuring requirements. Both require that the probe be aimed at the center of the forehead during the test and keep a certain distance. Subject device 's distance is ≤ 1cm, predicate device 's distance is ≤ 5cm. It does not affect performance and accuracy which was evaluated in the performance testing. The accuracy of the subject device under this algorithm is demonstrated through the Clinical study. The Clinical Study Report was present in Performance Testing - Clinical. The difference in measurement distance has no impact on the safety of the device and is compliant to IEC60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
  • . Difference 4: Subject device response time is 1s. Tested to be compliant to the Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers.
  • . Difference 5: Subject device and predicate device have similar operating conditions and storage conditions. The storage and operating environment conditions of subject device meet the IEC60601-1: 2005, AMD1:2012, AMD2:2020 and IEC60601-1-11: 2015, AMD1:2020 standards. The test report is presented in the Electromagnetic Compatibility and Electrical Safety.
  • . Difference 6: Subject device and predicate device have different reference sites. The subject device is developed with reference to armpit test temperature. Through clinical trials, the measured temperature of subject device and armpit thermometers are compared to demonstrate the measurement accuracy of subject device meets its declared accuracy range. The test report is presented in the Performance Testing-Clinical.

{14}------------------------------------------------

Premarket Notification 510(k) Submission-Section II 510(k) Summary

  • Difference 7: Subject device and predicate device have different display type. The subject device display meets the requirement of IEC60601-1: 2005, AMD1:2012, AMD2:2020 and IEC60601-1-2:2014. The test report is presented in the Electromagnetic Compatibility and Electrical Safety.
  • Difference 8: Subject device and predicate device have same memory type, but the amount of storage is different. This does not affect the safety and effectiveness of the device.
  • Difference 9: Subject device and predicate device have the same function of automatic switch-off, the difference is the time for automatic shutdown. This does not affect the safety and effectiveness of the device.
  • . Difference 10: Subject device and predicate device have same function of above normal temperature reminder; the difference is subject device can set age range but predicate device not include this function. Because different ages have different definitions of above normal temperatures, this subject device has the function of selecting different age ranges (0-1months, 3-36months, and +36months). This does not affect the safety and effectiveness of the device.
  • . Difference 11: Subject device and predicate device utilize the same standard, but the subject device uses the latest standards.
  • Difference 12: The subject device has Bluetooth function (Bluetooth transmission is optional), so we have conducted testing for the subject device according Bluetooth Wireless Test according to FCC Part 15B and Part 15C and wireless coexistence testing. The test report is presented in the Electromagnetic Compatibility and Electrical Safety.
  • . Difference 13: Subject device and predicate device have same standard, but the subject device uses the latest standards.

2.6 Indications for use

Infrared Thermometer is a reusable thermometer intended for the intermittent determination of human body temperature in a non-contact mode using the center of the forehead as the measurement site on people of all ages. It can be used by consumers in the household environment and by healthcare provider.

{15}------------------------------------------------

2.7 Testing

The Infrared Thermometer CFT-308 was supported by both laboratory and clinical accuracy testing in order to ensure that they were appropriate performance and functional features to fully comply with recognized standards and is substantially equivalent to the predicate device.

Non-clinical Test

The Infrared Thermometer CFT-308 is designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:

  • IEC60601-1: 2005, AMD1:2012, AMD2:2020 Medical electrical equipment Part 1: ● General requirements for basic safety and essential performance
  • . IEC60601-1-11: 2015, AMD1:2020 Medical electrical equipment—Part 1-11: General requirements for basic safety and essential performance- Collateral Standard : Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • . IEC60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • . ISO80601-2-56:2017 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • . ASTM E1965-98:2016 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
  • ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • . ISO10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

We have performance tests per FDA guidance "Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers''.

The Software Validation is in compliance with FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

{16}------------------------------------------------

Premarket Notification 510(k) Submission-Section II 510(k) Summary

Compliant to FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".

No.Test Name
1System Performance Test
2Performance Test according to ISO 80601-2-56: 2017
3Electromagnetic Compatibility Test According to IEC60601-1-2:2014
4Electrical Safety Test According to IEC60601-1: 2005, AMD1:2012,AMD2:2020
5Used in the home healthcare environment test according toIEC60601-1-11: 2015, AMD1:2020
6Bluetooth Wireless Test according to FCC Part 15B and Part 15C
7Wireless coexistence testing
8Biological evaluation of medical devices - Part 1: Evaluation and testing within arisk management process ISO 10993-1:2018
9Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityISO10993-5:2009
10Biological evaluation of medical devices - Part 10: Tests for irritation and skinsensitization ISO10993-10:2010

Table II-2 The list of non-clinical test performed on the subject devices.

The test results indicate that the safety and effectiveness of the subject device is substantially equivalent to that of the predicate device.

Clinical Test

Clinical testing was conducted according to ASTM E1965-98(2016) and ISO 80601-2-56:2017. The Clinical Study of CFT-308 was conducted May 29th, 2020 to March 19th, 2021 in the second hospital of Tianjin medical university. The subject device was compared to a reference thermometer (Model Number: CRW-23) for this study.152 subjects were screened. 2 subjects were eliminated, and 150 subjects completed the clinical trial (15 subjects aged 0 to 3 months, 35 subjects aged 3 months to 1-year-old, 50 subjects aged from 1 to 5 years old, 50 subjects aged 5 years and above).

{17}------------------------------------------------

2.8 Determination of substantial equivalence

The subject device of Infrared Thermometer CFT-308 has the same classification information, same intended use, similar components, similar performance effectiveness as the predicated device. The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.