K Number
K211405
Date Cleared
2021-10-04

(151 days)

Product Code
Regulation Number
870.1340
Panel
CV
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Device Description

The Prelude Guide Sheath Introducer is a sterile, disposable device consisting of a (a) a coil reinforced shaft with an atraumatic tip and the distal end: (b) a hemostasis valve with a side port and color-coded stopcock; and (c) a tapered tip dilator with snap-fit hub at the proximal end.

AI/ML Overview

This is a 510(k) summary for a medical device (Prelude Guide Sheath). The document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance data and material comparisons, not on evaluating an AI/ML algorithm. Therefore, many of the requested criteria related to AI/ML device studies (such as sample size for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable or readily extractable from this specific document.

However, I can provide the acceptance criteria and performance as described for the non-clinical tests.

Acceptance Criteria and Reported Device Performance

The document states that "The results of the testing demonstrated that the subject Prelude Guide Sheath met the predetermined acceptance criteria applicable to the performance of the device." The specific numerical acceptance criteria for each test are not provided in this summary, but the general categories of tests performed are listed. The "Reported Device Performance" for each test is simply that the device met those unstated criteria.

Here's a table based on the provided "Performance Tests" section:

No.Verification/Validation TestApplicable Standard(s)Acceptance Criteria (Implied)Reported Device Performance
1Insertion Force/Tip Insertion, Peel BackInternal RequirementMet internal requirementsMet predetermined criteria
2Peak Tensile Force Shaft/Hub & TipISO 10555-1:2013, ISO 10555-3:2013, ISO 11070:2014Met standards' requirementsMet predetermined criteria
3Visual (Extraneous/Surface Defects)ISO 11070:2014Met standards' requirementsMet predetermined criteria
4Sheath StiffnessInternal RequirementMet internal requirementsMet predetermined criteria
5Sheath KinkTested through ValidationMet validation requirementsMet predetermined criteria
6Radiopacity marker band & tipISO 11070:2014, ASTM F640-12, ISO 10555-1:2013Met standards' requirementsMet predetermined criteria
7Hydrophilic coating durability & coverage/coating lengthInternal RequirementMet internal requirementsMet predetermined criteria
8Hydrophilic Coating LubricityTested through ValidationMet validation requirementsMet predetermined criteria
9Coating integrity test (particulate evaluation)ASTM F1877-16, Class II Special Controls Guidance for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) CathetersMet standards' requirementsMet predetermined criteria
10Corrosion resistanceISO 11070:2014Met standards' requirementsMet predetermined criteria
11Sheath assembly leakISO 11070:2014Met standards' requirementsMet predetermined criteria
12Hydrogen Peroxide ResistanceISO 10555-1:2013, ISO 10555-3:2013Met standards' requirementsMet predetermined criteria
13Radiopacity DilatorISO 11070:2014, ISO 10555-1:2013Met standards' requirementsMet predetermined criteria
14Dilator tip fractureInternal RequirementMet internal requirementsMet predetermined criteria
15Dilator-to-Hub TensileISO 10555-1:2013Met standards' requirementsMet predetermined criteria
16Sheath tip to dilator taper lengthInternal RequirementMet internal requirementsMet predetermined criteria
17Dilator Stiffness & Hub Snap ForceNo Standard - Tested through ValidationMet validation requirementsMet predetermined criteria
18Dilator Insertion/Removal from SheathNo Standard - Testing through ValidationMet validation requirementsMet predetermined criteria
19Dimensional tests for OD & ID of sheath & dilatorISO 11070:2014Met standards' requirementsMet predetermined criteria
20Curve shape & visual for sheath & dilatorInternal RequirementMet internal requirementsMet predetermined criteria
21Curve RetentionNo standard - Testing through ValidationMet validation requirementsMet predetermined criteria
22Sheath Effective LengthISO 10555-1:2013Met standards' requirementsMet predetermined criteria
23Luer Design TestsISO 11070:2014, ISO 594-1: 1986, ISO 594-2:1998Met standards' requirementsMet predetermined criteria
24Hemostasis Valve LeakageISO 11070:2014Met standards' requirementsMet predetermined criteria
25EO Sterilization ValidationISO 11135:2014, ISO 10993-7:2008, AAMI TIR28:2016Met standards' requirementsMet predetermined criteria
26Package Integrity TestingISO 11607-1:2019, ISO 11607-2:2019, ASTM D4169-16Met standards' requirementsMet predetermined criteria
27Biocompatibility Testing (Cytotoxicity, Sensitization, etc.)ISO 10993-1:2009, ISO 10993-4: 2017, ISO 10993-5:2009, ISO 10993-6:2009, ISO 10993-7:2008, ISO 10993-10:2010, ISO 10993-11:2017, ISO 10993-12:2012Met standards' requirementsMet predetermined criteria
28Chemical Characterization – Particulate AnalysisASTM F1877-16Met standards' requirementsMet predetermined criteria

The remaining requested information is largely irrelevant to this 510(k) submission, as it pertains to AI/ML algorithm validation for which this device is not. This submission focuses on the physical and material characteristics of a medical device (a guide sheath introducer) and its substantial equivalence to existing predicate devices, rather than the performance of an AI/ML system.

Therefore, the following information is not applicable or extractable from the provided document:

  1. Sample sized used for the test set and the data provenance: Not an AI/ML study, so "test sets" in that context don't apply. The non-clinical tests would have their own sample sizes for a variety of physical or chemical tests, but these are not detailed as a single "test set" for an algorithm.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not an AI/ML study; no "ground truth" in the diagnostic/interpretive sense is established by experts here.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
  7. The sample size for the training set: Not applicable.
  8. How the ground truth for the training set was established: Not applicable.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).