K Number
K210753
Device Name
KardiaMobile 6L
Manufacturer
Date Cleared
2021-06-30

(107 days)

Product Code
Regulation Number
870.2920
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KardiaMobile 6L is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVE. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.

Device Description

KardiaMobile 6L is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two channel ECG rhythm recordings. KardiaMobile 6L provides output of one or six ECG leads, including Lead I, Lead II, Lead III, aVL, aVR and aVF. The device utilizes the computing power of iOS-based or Android-based devices (referred to as "Mobile Computing Platforms" (MCP) within this submission) to obtain and analyze ECG signals. KardiaMobile 6L consists of KardiaMobile 6L Hardware (portable small wireless hardware with electrodes) and the Kardia App, which is installed on an MCP (i.e., iOS or Android devices). KardiaMobile 6L hardware uses Bluetooth to transmit the ECG signal from the electrodes to the Kardia App on the MCP. KardiaMobile 6L displays thus record their ECG and additionally provides ECG analysis using the KardiaAI (K181823, K201985) ECG analysis suite, which includes the determinations of Atrial Fibrillation, Normal Sinus Rhythm, Bradycardia, Tachycardia, or Unclassified. The device is intended to be used by patients with known or suspected heart conditions and health conscious individuals as well as by healthcare professionals (HCPs) who want to remotely monitor their patient's heart health. Patients can forward the recorded ECG to their HCP, who can review the ECG for rhythm and to measure the QT interval. The device is available for Over-the-Counter (OTC) purchase and for purchase with a prescription. The Kardia App is also comes in an alternate variant, called the KardiaStation App that is exclusively used within hospitals and clinics; this app is identical to the Kardia App with the exception of incorporating patient administration workflows.

AI/ML Overview

Here's a summary of the acceptance criteria and study details based on the provided text, focusing on the expanded use case of QT interval measurement for healthcare professionals:


1. Table of Acceptance Criteria and Reported Device Performance

The provided document specifically addresses the expanded use case for healthcare professionals to measure the QT interval. It compares the KardiaMobile 6L's ability to measure QT intervals against a gold standard 12-lead diagnostic ECG device.

Acceptance Criterion (Implicit)Reported Device Performance
QT interval measured by KardiaMobile 6L is equivalent to a gold-standard 12-lead diagnostic ECG device.The comparative statistical analysis of this assessment determined that the QT interval measured using the subject device is equivalent to the interval measured from a commercial gold-standard diagnostic 12-lead ECG.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 313 subjects.
  • Data Provenance: Not explicitly stated regarding country of origin, but the study design suggests prospective data collection as ECG recordings were "taken comparison" and included "both healthy volunteers and patients suspected of long QT syndrome or other genetic heart disease."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • The document states that a "core lab used to provide precise QT measurements for Thorough QT studies was utilized to measure the QT and heart-rate corrected QTc in each ECGs."
  • Number of Experts: Not specified.
  • Qualifications of Experts: Implied expertise in QT measurement for Thorough QT studies, indicating highly specialized cardiac electrophysiology expertise.

4. Adjudication Method for the Test Set

  • The document mentions that the core lab measured QT and QTc "in a randomized and blinded order, using standard measurement techniques." This suggests a robust, independent assessment, but a specific "adjudication method" (e.g., 2+1, 3+1) is not explicitly stated.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not explicitly described for the expanded QT interval measurement use. The study focused on the equivalence of QT measurements from the device compared to a gold standard, not on the improvement of human readers with AI assistance. The "human readers" in this context were the core lab experts establishing ground truth, not users of the device whose performance was being evaluated for improvement with AI.

6. Standalone (Algorithm Only) Performance

  • No, an explicit standalone (algorithm only without human-in-the-loop performance) study for QT interval measurement was not described. The study validated that "a healthcare professional can review and measure the QT interval reliably using an ECG recorded using KardiaMobile 6L." While the device records the ECG, the measurement itself is performed by a human expert/core lab. The device's role here is data acquisition for human interpretation.
  • It's important to note that the KardiaAI platform (K181823, K201985) does run algorithms for detecting Atrial Fibrillation, Normal Sinus Rhythm, Bradycardia, Tachycardia, or Unclassified rhythms, and these would have had standalone performance studies in their original clearances. However, for the expanded QT interval measurement claim, the study focuses on the reliability of the ECG recording for human measurement.

7. Type of Ground Truth Used

  • Expert Consensus / Highly Specialized Core Lab Measurement: The QT and heart-rate corrected QTc measurements were performed by a "core lab used to provide precise QT measurements for Thorough QT studies." This represents a high-level, specialized form of expert consensus and measurement standard.

8. Sample Size for the Training Set

  • Not provided. The document focuses on the clinical validation study for the QT interval measurement claim, which is a test set evaluation. Information about the training set for any underlying algorithms is not included in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established

  • Not provided. As the training set size is not mentioned, neither is the method for establishing its ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 30, 2021

AliveCor, Inc. % Prabhu Raghavan Principal MDQR, LLC 444 Castro Street, Suite 600 Mountain View, California 94041

Re: K210753

Trade/Device Name: KardiaMobile 6L Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH, DPS, QDA Dated: March 12, 2021 Received: March 15, 2021

Dear Prabhu Raghavan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K210753

Device Name KardiaMobile 6L

Indications for Use (Describe)

KardiaMobile 6L is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVE. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210753

Image /page/3/Picture/1 description: The image shows the AliveCor logo in a teal color. The logo is a stylized wordmark with a rounded, modern font. Above the logo, the text "Page 1 of 4" is visible, indicating that this image is part of a multi-page document. The logo is the main focus of the image.

510(k) Summary for K210753

Prepared in accordance with the requirements of 21 CFR 807.92

Submitter Information [807.92(a)(1)]

Submitter/ApplicantAliveCor, Inc.444 Castro Street, Suite 600Mountain View, CA 94041Phone: 650-396-8553Fax: 650-282-7932
Primary Contact PersonPrabhu RaghavanRegulatory Consultant for AliveCorMDQR, LLC444 Castro Street, Suite 600Mountain View, CA 94041Phone: 408-316-5707Fax: 650-282-7932Email: prabhu@mdqr.solutions
Submitter andSecondary Contact PersonSaket Bhatt
VP of Regulatory and Quality
444 Castro Street, Suite 600
Mountain View, CA 94041
Phone: 408-701-7319
Fax: 650-282-7932
Email: ra@alivecor.com

Date Prepared

May 27, 2021

Device Information [807.92(a)(2)]

Trade NameKardiaMobile 6L
Common NameTransmitters and Receivers, Electrocardiograph, Telephone,
Classification21 CFR§870.2920
Device ClassClass II
Product CodeDXH
Subsequent Product CodeDPS, QDA

Predicate Information [807.92(a)(3)|

AliveCor, Inc., K183319, Triangle System (marketed as Predicate(s) KardiaMobile 6L)

Device Description [807.92(a)(4)]

KardiaMobile 6L is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two channel ECG rhythm recordings. KardiaMobile 6L provides output of one or six ECG leads, including Lead I,

{4}------------------------------------------------

K210753

510(k) Summary for KardiaMobile 6L AliveCor, Inc.

Lead II, Lead III, aVL, aVR and aVF. The device utilizes the computing power of iOS-based or Android-based devices (referred to as "Mobile Computing Platforms" (MCP) within this submission) to obtain and analyze ECG signals. KardiaMobile 6L consists of KardiaMobile 6L Hardware (portable small wireless hardware with electrodes) and the Kardia App, which is installed on an MCP (i.e., iOS or Android devices). KardiaMobile 6L hardware uses Bluetooth to transmit the ECG signal from the electrodes to the Kardia App on the MCP. KardiaMobile 6L displays thus record their ECG and additionally provides ECG analysis using the KardiaAI (K181823, K201985) ECG analysis suite, which includes the determinations of Atrial Fibrillation, Normal Sinus Rhythm, Bradycardia, Tachycardia, or Unclassified. The device is intended to be used by patients with known or suspected heart conditions and health conscious individuals as well as by healthcare professionals (HCPs) who want to remotely monitor their patient's heart health. Patients can forward the recorded ECG to their HCP, who can review the ECG for rhythm and to measure the QT interval. The device is available for Over-the-Counter (OTC) purchase and for purchase with a prescription. The Kardia App is also comes in an alternate variant, called the KardiaStation App that is exclusively used within hospitals and clinics; this app is identical to the Kardia App with the exception of incorporating patient administration workflows.

Indications for use [807.92(a)(5)]

KardiaMobile 6L is intended to record, store and transfer one- and two-channel electrocardiogram (ECG) rhythms. In single channel mode. KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.

Substantial Equivalence

The KardiaMobile 6L subject device has the same intended use, physical characteristics, technological characteristics and software as the K183319, Triangle System (marketed as "KardiaMobile 6L") predicate device. The difference between the two devices is labeling only, in that the subject device expands the use by HCPs to measure the QT interval in addition to the rhythm analysis application in the predicate. This expanded use has been analyzed and addressed through performance testing. The testing results showed that differences between the subject and predicate device do not raise different questions of safety or effectiveness. OT interval measurement is prescription use.

{5}------------------------------------------------

FeatureAliveCor KardiaMobile 6L(Subject Device)AliveCor KardiaMobile 6L(K183319)(Predicate Device)
Product CodeDXH, DPS, QDADXH, DPS, QDA
Mechanism of ActionUser completes circuit with skincontact and hardware transmits ECGsignal to MCP to convert and displayECG waveformNo difference
Where used (intended use)Mobile/active users at rest (ambulatory)No difference
Use cases for HCPsRhythm analysisQT Interval MeasurementRhythm analysis only
Anatomical sitesLeft hand fingers to right hand fingersLeft hand fingers to right hand fingersand to left leg/kneeNo difference
Available Algorithms(through KardiaAIK181823)Atrial FibrillationNoise AlgorithmNormal Sinus RhythmTachycardiaBradycardiaNo difference
Data Acquisition:
Frequency Response0.5Hz - 40HzNo difference
No. of ECG electrodesThree (3) dry electrodesNo difference
No. of ECG channelsSingle-channel and two-channelNo difference
Resolution16 bitsNo difference
Sample Rate300 Samples/SecondNo difference
Power Supply:
Battery1 Lithium Manganese Dioxide CoinCellsNo difference
Battery Life (typical)100 hours operationalNo difference

Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)]

{6}------------------------------------------------

K210753

510(k) Summary for KardiaMobile 6L AliveCor, Inc.

Performance Data [807.92(b)]

All necessary testing was conducted on KardiaMobile 6L to support a determination of substantial equivalence to the predicate device.

Nonclinical Testing Summary [807.92(b)(1)]

Since both the subject and the predicate have the same hardware and software, and the changes are to device labeling only, no bench testing was required.

Clinical Testing Summary [807.92(b)(2)]

Clinical performance testing was conducted to validate that a healthcare professional can review and measure the QT interval reliably using an ECG recorded using KardiaMobile 6L. The clinical evaluation compared KardiaMobile 6L 6-lead measurements (recorded in a sitting position) to an FDA-cleared gold standard 12-lead diagnostic ECG device (recorded supine) in 313 subjects. Subjects included both healthy volunteers and patients suspected of long QT syndrome or other genetic heart disease. ECG recordings from both the subject device and gold standard device were taken comparison. A core lab used to provide precise OT measurements for Thorough QT studies was utilized to measure the QT and heart-rate corrected QTc in each ECGs, in a randomized and blinded order, using standard measurement techniques. The comparative statistical analysis of this assessment determined that the QT interval measured using the subject device is equivalent to the interval measured from a commercial gold-standard diagnostic 12-lead ECG. The results of the testing demonstrated that the modifications to the KardiaMobile 6L labeling do not raise different questions of safety or effectiveness.

Conclusions [807.92(b)(3)]

KardiaMobile 6L has the same intended use as the predicate device, and the labeling differences do not raise different questions of safety or effectiveness. The labeling differences between the subject device and the predicate device have been clinically validated. Therefore, KardiaMobile 6L is substantially equivalent to the predicate device.

§ 870.2920 Telephone electrocardiograph transmitter and receiver.

(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).