(251 days)
Not Found
No
The summary describes image processing and quantitative analysis techniques but does not mention AI, ML, or related concepts like neural networks or deep learning. The validation focuses on traditional quantitative methods and clinical studies comparing results to established standards like FFR and IVUS.
No.
The device is described as software for quantitative analysis of X-ray angiographic images to assess coronary vessels and lesions, primarily for diagnostic or assessment purposes rather than direct treatment or therapy.
Yes
The device is described as providing "validated and reproducible quantified results" for the "assessment of coronary vessels" and "to obtain quantifications of lesions in coronary vessels, to determine the functional significance of the individual or consecutive multiple lesions". These actions, such as assessment and quantification for determining functional significance, are core components of arriving at a diagnosis.
Yes
The device description explicitly states "QANGIO XA 3D is the Medis software solution" and "has been developed as a standalone application to run on a Windows based operating system." It processes existing X-ray angiographic images and does not include any hardware components.
Based on the provided information, this device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze biological samples: IVDs are designed to examine specimens taken from the human body, such as blood, urine, tissue, etc., to provide information about a person's health.
- This device analyzes medical images: QANGIO XA 3D analyzes X-ray angiographic images, which are visual representations of the coronary vessels, not biological samples.
- The intended use is image analysis for diagnosis and treatment planning: The device is used to quantify features within the images to support the assessment of coronary vessels and aid in treatment decisions. This is a function of medical image analysis software, not an IVD.
The device falls under the category of medical image processing software used for diagnostic and procedural planning purposes, which is distinct from IVDs.
N/A
Intended Use / Indications for Use
QANGIO XA 3D is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. QANGIO XA 3D enables interventional cardiologists and researchers to obtain quantifications of one or more lesions in the analyzed coronary vessel segment. In particular, QANGIO XA 3D provides:
- . Quantitative results of coronary vessel segments based on a 3D reconstructed model;
- Dimensions of the cardiovascular vessels and lesions;
- Quantification of the pressure drop in coronary vessels.
QANGIO XA 3D is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by QANGIO XA 3D are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Product codes (comma separated list FDA assigned to the subject device)
QHA, LLZ
Device Description
QANGIO XA 3D is the Medis software solution for the quantitative analysis of cardiac X-ray Angiography (XA) studies. QANGIO XA 3D features:
- . Acquisition guide and series selection guide: A visualization method is provided to find suitable rotation and angulation angles for a second XA image if the first XA image is selected (acquisition guide). Another visualization method is provided to support the selection of matching XA images in a completed patient study (selection guide). Input: XA image. Output: XA image with required angulation compared to input image. Method: Comparing angulation information from DICOM.
- . Annotations and basic measurements: User tools are provided for basic textual annotations onto an XA image, to perform a simple distance or area measurement in a 2D XA image, and to create snapshot images of the screen. Input: XA image. Output: simple measurement results. Method: manual delineation.
- . 2D and 3D OCA vessel detection: 2D vessel boundaries are (automatically) detected of a user-selected coronary vessel segment in an XA image. The 2D vessel boundaries from two XA images of the same vessel under different angulation are used to reconstruct a 3D model of the user-selected vessel segment. Input: 2 XA images. Output: 3D model of targeted coronary vessel.
- 2D and 3D OCA vessel quantification: From the 3D model of the vessel segment, each ellipse is described by its two diameters, which are stored as a minimum and maximum diameter. The total set of diameters is called the "arterial diameter function". From the arterial diameter function, a so-called "reference diameter" function is calculated using a recursive method that fits a linear approximation through the diameters, excluding the areas that are considered narrowed portions of the vessel segment. Analogue to 2D X-ray angiographic analysis provided in our predicate device X-RAY VVA, all quantitative parameters provided by QAngio XA 3D with respect to a lesion are calculated from these arterial and reference diameter functions. By default the software will indicate the lesion with the most severe narrowing (minimal lumen diameter / area), which is defined to be the narrowing with the largest percent of area-stenosis. Further, the start and end point of the lesion will be provided, resulting in the calculation of the lesion. Other quantitative results derived from the diameter functions are: maximum diameter / area, average diameter over the lesion or the whole vessel segment.
- QFR calculations:
- Ouantitative Flow Ratio (OFR) analysis: The OFR value (ratio of the pressure drop) is calculated based o on computer calculations of the functional significance of a coronary vessel narrowing. The computation is based on computational fluid dynamics (CFD) principles applied to a 3D reconstruction of the coronary vessel segment (as described above) and one of the flow models. Input: 3D model of targeted coronary vessel and a flow model. Output: QFR curve describing the QFR value at each location along the centerline of the vessel segment. From this QFR curve function the parameters "Vessel QFR", "Index QFR", and "Lesion QFR" can be derived. Two different flow models are available:
- The so-called fixed flow model, which uses a fixed empiric hyperemic flow velocity, derived o from literature. Using this model will yield fOFR values.
- The contrast flow models the hyperemic flow velocity derived from coronary o angiograms without pharmacologically induced hyperemia (i.e the contrast flow was converted into the virtual hyperemic flow based on data derived from previous studies). Using this model will yield so-called cQFR values.
- The agreement between the fQFR and cQFR model has been shown in a comparative study. о
- о The QFR value is validated against the FFR value in several validation studies. The WIFI II study, FAVOR II Europe-Japan, FAVOR II China, OFR and Non Culprit lesions in STEMI Patients (and many others). For FFR, only the value at a specific point along a coronary segment is known. i.e. at the place of the pressure sensor, while the calculated OFR curve describes its value along the entire segment. For the comparison of FFR and OFR at a specific location, the OFR value is sampled from the OFR curve at the location of the FFR wire. Since this sample is at a different position in each analysis, it is admissible to see this as a validation of the entire QFR curve. From the QFR curve, three different QFR values are derived (i.e. samples of the QFR curve) and reported as output:
- The vessel OFR is the value at the end of the OFR curve; o
- The index QFR is a caliper at a user-defined position in the QFR curve; O
- o The lesion QFR is 1 minus the QFR drop between the lesion markers;
- Ouantitative Flow Ratio (OFR) analysis: The OFR value (ratio of the pressure drop) is calculated based o on computer calculations of the functional significance of a coronary vessel narrowing. The computation is based on computational fluid dynamics (CFD) principles applied to a 3D reconstruction of the coronary vessel segment (as described above) and one of the flow models. Input: 3D model of targeted coronary vessel and a flow model. Output: QFR curve describing the QFR value at each location along the centerline of the vessel segment. From this QFR curve function the parameters "Vessel QFR", "Index QFR", and "Lesion QFR" can be derived. Two different flow models are available:
- Single vessel report: A report is generated by the software that shows patient information, image acquisition . information (both obtained from the DICOM input), analysis results (vessel sizing and OFR value) and snapshot images showing the vessel boundaries.
QANGIO XA 3D will be used by interventional cardiologists and researchers to obtain quantifications of lesions in coronary vessels, to determine the functional significance of the individual or consecutive multiple lesions, prior, during or after percutaneous coronary intervention treatment. OANGIO XA 3D has been developed as a standalone application to run on a Windows based operating system. The import of images and the export of analysis results are via CD / DVD, a PACS or network environment.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray Angiography (XA)
Anatomical Site
Coronary vessels / coronary arteries
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Interventional cardiologists and researchers, clinical settings.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
- 3D vessel reconstruction: The reconstruction of a 3D model from the 2D vessel boundaries detected on two XA images of the same vessel under different angulation have been validated clinically in studies using OCA parameters and IVUS pullbacks in patients.
- Quantitative Flow Ratio (OFR) analysis: The calculation of the OFR value is both technically validated. Estimates for precision of OFR are based on a strong body of evidence derived from multiple prospective studies. Two of the multicenter studies (FAVOR II Europe-Japan and FAVOR II China) show excellent results of true in-procedure QFR analyzed by end-users. In these prospective, observational, investigator-initiated multi-center studies QFR was evaluated for feasibility and diagnostic performance and agreement with pressure-wire derived fractional flow reserve (FFR) as reference standard. QFR and FFR were analyzed separately by different observers. The primary endpoints of sensitivity of QFR were compared to two-dimensional QCA (2D-QCA) with FFR as reference standard.
- The QFR value is validated against the FFR value in several validation studies, including The WIFI II study, FAVOR II Europe-Japan, FAVOR II China, OFR and Non Culprit lesions in STEMI Patients.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1600 Angiographic x-ray system.
(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
Medis Medical Imaging Systems bv % B. Goedhart, Ph.D. VP Corporate R&D Schuttersveld 9 Leiden, 2316 XG THE NETHERLANDS
Re: K182611
Trade/Device Name: QANGIO XA 3D Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: Class II Product Code: QHA, LLZ Dated: May 1, 2019 Received: May 3, 2019
Dear Dr. Goedhart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
May 30, 2019
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K182611
Device Name QANGIO XA 3D
Indications for Use (Describe)
QANGIO XA 3D is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by QAngio XA 3D are used in a clinical setting on X-ray images of an individual patient. The results are only intended for use by the responsible clinicians.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
---|---|
---------------------------------------------------------------------------------------------------- | --------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
5. 510(k) Summary of Safety and Effectiveness
Submission in accordance with the requirements of 21 CFR Part 807.87(h).
Submitter | : Medis medical imaging systems bv |
---|---|
Address | : Schuttersveld 9 |
: 2316 XG Leiden | |
: The Netherlands | |
Contact Person | : B. Goedhart, PhD, Vice President R&D |
: bgoedhart@medis.nl | |
Telephone | : +31 71 5241 700 |
Fax | : +31 71 521 5617 |
Prepared | : May 29, 2019 |
Trade / Device Name | : QANGIO XA 3D |
Common Name | : Radiological Image Processing Software |
Regulatory Class | : II |
Regulation Description | : Angiographic X-ray system |
Regulation Number | : 21 CFR 892.1600 |
Classification Product Code | : QHA |
Subsequent Product Code | : LLZ |
Predicate Device
- The Pie Medical: CAAS™ Workstation (K180019), Regulation Number : 21 CFR 892.1600. ●
Device Description
QANGIO XA 3D is the Medis software solution for the quantitative analysis of cardiac X-ray Angiography (XA) studies. QANGIO XA 3D features:
- . Acquisition guide and series selection guide: A visualization method is provided to find suitable rotation and angulation angles for a second XA image if the first XA image is selected (acquisition guide). Another visualization method is provided to support the selection of matching XA images in a completed patient study (selection guide). Input: XA image. Output: XA image with required angulation compared to input image. Method: Comparing angulation information from DICOM.
- . Annotations and basic measurements: User tools are provided for basic textual annotations onto an XA image, to perform a simple distance or area measurement in a 2D XA image, and to create snapshot images of the screen. Input: XA image. Output: simple measurement results. Method: manual delineation.
- . 2D and 3D OCA vessel detection: 2D vessel boundaries are (automatically) detected of a user-selected coronary vessel segment in an XA image. The 2D vessel boundaries from two XA images of the same vessel under different angulation are used to reconstruct a 3D model of the user-selected vessel segment. Input: 2 XA images. Output: 3D model of targeted coronary vessel.
- 2D and 3D OCA vessel quantification: From the 3D model of the vessel segment, each ellipse is described by its two diameters, which are stored as a minimum and maximum diameter. The total set of diameters is called the "arterial diameter function". From the arterial diameter function, a so-called "reference diameter" function is calculated using a recursive method that fits a linear approximation through the diameters, excluding the areas that are considered narrowed portions of the vessel segment. Analogue to 2D X-ray angiographic analysis provided in our predicate device X-RAY VVA, all quantitative parameters provided by QAngio XA 3D with respect to a lesion are calculated from these arterial and reference diameter functions. By default the software will indicate the lesion with the most severe narrowing (minimal lumen diameter / area), which is defined to be the narrowing with the largest percent of area-stenosis. Further, the start and end point of the lesion will be provided, resulting in the calculation of the lesion. Other quantitative results derived from the diameter functions are: maximum diameter / area, average diameter over the lesion or the whole vessel segment.
4
- QFR calculations:
- Ouantitative Flow Ratio (OFR) analysis: The OFR value (ratio of the pressure drop) is calculated based o on computer calculations of the functional significance of a coronary vessel narrowing. The computation is based on computational fluid dynamics (CFD) principles applied to a 3D reconstruction of the coronary vessel segment (as described above) and one of the flow models. Input: 3D model of targeted coronary vessel and a flow model. Output: QFR curve describing the QFR value at each location along the centerline of the vessel segment. From this QFR curve function the parameters "Vessel QFR", "Index QFR", and "Lesion QFR" can be derived. Two different flow models are available:
- The so-called fixed flow model, which uses a fixed empiric hyperemic flow velocity, derived o from literature. Using this model will yield fOFR values.
- The contrast flow models the hyperemic flow velocity derived from coronary o angiograms without pharmacologically induced hyperemia (i.e the contrast flow was converted into the virtual hyperemic flow based on data derived from previous studies). Using this model will yield so-called cQFR values.
- The agreement between the fQFR and cQFR model has been shown in a comparative study. о
- о The QFR value is validated against the FFR value in several validation studies. The WIFI II study, FAVOR II Europe-Japan, FAVOR II China, OFR and Non Culprit lesions in STEMI Patients (and many others). For FFR, only the value at a specific point along a coronary segment is known. i.e. at the place of the pressure sensor, while the calculated OFR curve describes its value along the entire segment. For the comparison of FFR and OFR at a specific location, the OFR value is sampled from the OFR curve at the location of the FFR wire. Since this sample is at a different position in each analysis, it is admissible to see this as a validation of the entire QFR curve. From the QFR curve, three different QFR values are derived (i.e. samples of the QFR curve) and reported as output:
- The vessel OFR is the value at the end of the OFR curve; o
- The index QFR is a caliper at a user-defined position in the QFR curve; O
- o The lesion QFR is 1 minus the QFR drop between the lesion markers;
- Ouantitative Flow Ratio (OFR) analysis: The OFR value (ratio of the pressure drop) is calculated based o on computer calculations of the functional significance of a coronary vessel narrowing. The computation is based on computational fluid dynamics (CFD) principles applied to a 3D reconstruction of the coronary vessel segment (as described above) and one of the flow models. Input: 3D model of targeted coronary vessel and a flow model. Output: QFR curve describing the QFR value at each location along the centerline of the vessel segment. From this QFR curve function the parameters "Vessel QFR", "Index QFR", and "Lesion QFR" can be derived. Two different flow models are available:
- Single vessel report: A report is generated by the software that shows patient information, image acquisition . information (both obtained from the DICOM input), analysis results (vessel sizing and OFR value) and snapshot images showing the vessel boundaries.
QANGIO XA 3D will be used by interventional cardiologists and researchers to obtain quantifications of lesions in coronary vessels, to determine the functional significance of the individual or consecutive multiple lesions, prior, during or after percutaneous coronary intervention treatment. OANGIO XA 3D has been developed as a standalone application to run on a Windows based operating system. The import of images and the export of analysis results are via CD / DVD, a PACS or network environment.
Intended Use
QANGIO XA 3D is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. QANGIO XA 3D enables interventional cardiologists and researchers to obtain quantifications of one or more lesions in the analyzed coronary vessel segment. In particular, QANGIO XA 3D provides:
- . Quantitative results of coronary vessel segments based on a 3D reconstructed model;
- Dimensions of the cardiovascular vessels and lesions;
- Quantification of the pressure drop in coronary vessels.
Indications for Use
OANGIO XA 3D is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by QANGIO XA 3D are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Substantial Equivalence Information
The predicate device for QANGIO XA 3D is the CAASTM Workstation.
5
Especially this 2D vessel analysis (for the 2D XA project imaging) forms the basis for the 3D vessel analysis in QANGIO XA 3D. This analysis produces a 3D model of a coronary vessel from two 2D vessels detected in regular X-ray projection images. Similar functionality is provided by the CAAS™ Workstation (where it is called "QCA3D"). Further, QANGIO XA 3D provides visualization functionalities for 3D review of the model, which are also supported by CAASTM Workstation.
OANGIO XA 3D provides (see Intended Use description above) a quantification parameter called OFR (Quantitative Flow Ratio), for which a similar parameter is also provided by the CAAS™ Workstation (where it is called "vFFR").
From the Intended Use comparison it is concluded that the CAAS Workstation offers comparable (and more) functionality to the QANGIO XA 3D device. The main functionality, being 3D vessel reconstruction and the calculation of the pressure drop, are explicitly mentioned in the intended use and indications for use.
All functionality (including 3D vessel reconstruction and the calculation of the QFR pressure drop) is specified in the OANGIO XA 3D system requirements. All requirements are tested and all results of the tests performed are summarized in the software test statement and especially the requirements coverage matrix of QANGIO XA 3D proving traceability between requirements and tests successfully executed.
Extensive clinical evaluation has been performed based on both scientific literature review and clinical studies executed with the device. The quantified results provided by QANGIO XA 3D for the main functionality (including 3D vessel reconstruction and the calculation of the QFR pressure drop) have all been validated, both technically and clinically:
- . 3D vessel reconstruction: The reconstruction of a 3D model from the 2D vessel boundaries detected on two XA images of the same vessel under different angulation have been validated clinically in studies using OCA parameters and IVUS pullbacks in patients.
- . Quantitative Flow Ratio (OFR) analysis: The calculation of the OFR value is both technically validated. Estimates for precision of OFR are based on a strong body of evidence derived from multiple prospective studies. Two of the multicenter studies (FAVOR II Europe-Japan and FAVOR II China) show excellent results of true in-procedure QFR analyzed by end-users. In these prospective, observational, investigator-initiated multi-center studies QFR was evaluated for feasibility and diagnostic performance and agreement with pressure-wire derived fractional flow reserve (FFR) as reference standard. QFR and FFR were analyzed separately by different observers. The primary endpoints of sensitivity of QFR were compared to two-dimensional QCA (2D-QCA) with FFR as reference standard.
Conclusions
QANGIO XA 3D has the same intended use as the predicate device.
OANGIO XA 3D complies with international process standards (ISO 13485, ISO 14971, IEC 62366 and ISO 15223).
Testing, validation and clinical evaluation have produced results consistent with design input requirements. QANGIO XA 3D is a software-only device for which there no applicable mandatory performance standards.
During the development, potential hazards were controlled by a risk management plan, including risk analysis, risk mitigation, verification and evaluation.
Medis concludes that QANGIO XA 3D is a safe and effective medical device, and is at least as safe and effective as its predicate devices. The use of QANGIO XA 3D does not change the intended use of X-ray Angiography acquisition, nor does the use of this software result in any new potential hazards.