(199 days)
The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow.
- Patient orientation identification
- Surview range recommendation
- Automatic centering the patient anatomy
- Provide visual images of patient on the table
Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.
Patient population limitation: Patient younger than 16 years are not supported.
Precise Position is an optional feature to assist user for position the patient before the body examination such as CT scan. The purpose of this feature is to reduce the patient position time via the camera detection and calculation result. It includes automatic detect patient orientation, patient anatomy scan range and center of patient anatomy.
Precise Position including a camera with both color and depth function is installed in the ceiling of the scan room, in such a way to cover the entire patient on the patient table. The camera control and image data transmit via the high speed fiber and copper hybrid USB cable. The power supply of the camera is from the gantry. Precise position adopts the AI algorithm (Convolution Neural Network) to detect the joints of the patient body, and then identify surview start/end position and patient orientation. The algorithm can also support detect center of patient anatomy.
Limitation for Precise Position
There is no limitation for Precise Position except below items:
• Patients below the age of 16 are not supported.
• Decubitus orientations are not supported.
The Precise Position display results may get affected by the following conditions:
• When the patient is covered by sheet, blanket etc.,
• When the patient is not completely covered by the ceiling camera view, e.g. blocked by the gantry or out of camera's FOV etc.
• When the patient is wearing clothes that reflects light, e.g. plastic-like clothes.
• When the patient is wearing black clothes.
• When the patient is wearing thick clothes.
• When there are other people around the patient.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Time Reduction in Patient Positioning (Efficiency) | Up to 23% time reduction in patient positioning achieved with "Precise Positioning workflow" compared to without Precise Position. |
| Accuracy of Vertical (Iso)center Positioning | With "Precise Position," the vertical position accuracy is increased up to 50%. (This implies a reduction in the average offset for vertical isocenter). |
| Consistency in Vertical (Iso)center Positioning | Up to 70% increase in Vertical position consistency with Precise Position. (This implies a reduction in the standard deviation for vertical isocenter positioning). |
| Consistency in Surview (Horizontal) Start Position | Up to 70% increase in horizontal position consistency with Precise Position. (This implies a reduction in the standard deviation for surview horizontal start position). |
| Intended Performance | The device performs as intended, is safe for its intended use, and has a favorable benefit-risk ratio. (This is a general acceptance, demonstrated by meeting the specific quantitative metrics above and by showing no clinical risks identified by the evaluated clinical data and compliance with various standards.) |
| Safety and Effectiveness | Demonstrated to be substantially equivalent to the primary currently marketed and predicate device (K180015) in terms of safety and effectiveness, based on non-clinical performance tests meeting international and FDA-recognized consensus standards. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Total 80 clinical scan positions.
- 40 cases used with Precise Position.
- 40 cases without the usage of Precise Position.
- Data Provenance: The document does not explicitly state the country of origin for the data or if it was retrospective or prospective. However, it mentions a "clinical evaluation… done by 5 Clinical experts" and volunteers not receiving radiation, implying a prospective study conducted for the purpose of this evaluation. The manufacturer is Philips Healthcare (Suzhou) Co., Ltd., which suggests the study likely occurred in China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: 5 Clinical experts.
- Qualifications: The specific qualifications (e.g., number of years of experience, specific specialty like "radiologist") are not explicitly stated beyond "Clinical experts."
4. Adjudication Method for the Test Set
- The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). It states that the "thorough clinical evaluation of this feature is done by 5 Clinical experts," implying they collectively contributed to the evaluation, but the exact consensus or adjudication process is not detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- This was not a typical MRMC comparative effectiveness study in the sense of multiple readers interpreting cases with and without AI.
- Instead, it was a comparative study on operational efficiency and positioning accuracy. It compared:
- Time taken for patient positioning by users with and without the Precise Position feature.
- Accuracy of vertical (iso)center positioning by users with and without the Precise Position feature.
- Consistency (standard deviation) of positioning by users with and without the Precise Position feature.
- Effect Size (Human Improvement with AI):
- Time Reduction: Users achieved up to 23% time reduction in patient positioning with the "Precise Positioning workflow" (which includes the AI).
- Vertical Position Accuracy: Users achieved an increase of up to 50% in vertical position accuracy with "Precise Position."
- Positioning Consistency (Vertical and Horizontal): Users achieved an increase of up to 70% in consistency with "Precise Position."
These metrics indicate the improvement in human operators' performance when assisted by the AI-powered Precise Position device.
6. If a Standalone (Algorithm Only) Performance Study Was Done
- The document does not explicitly state whether a standalone (algorithm only, without human-in-the-loop) performance study was conducted for the AI component of the Precise Position device. The clinical evaluation focuses on the human-with-AI system performance. The AI algorithm (Convolution Neural Network) is described as being used to detect joints and then determine positioning parameters, suggesting its performance is evaluated as part of the overall integrated system.
7. The Type of Ground Truth Used
- For the clinical evaluation, the ground truth for measuring time, accuracy, and consistency appears to be based on:
- Direct measurements of time taken for positioning.
- Measurements of offset in mm for vertical (iso) center position.
- Standard deviation calculations for vertical (iso)center positioning and surview (horizontal) start position.
- These measurements were likely compared against an ideal or intended positioning, which would be implicitly defined by the CT system's requirements and presumably verified by the clinical experts. It's not "pathology," "outcomes data," or a direct "expert consensus" on disease presence/absence, but rather a consensus on the correctness and optimal nature of the patient positioning parameters established by the device.
8. The Sample Size for the Training Set
- The document does not provide the sample size for the training set used for the AI algorithm (Convolution Neural Network). It only discusses the test set used for validating the combined human-AI system.
9. How the Ground Truth for the Training Set Was Established
- The document does not describe how the ground truth for the training set of the AI algorithm was established. It only mentions that the AI algorithm (Convolution Neural Network) is used to detect "joints of the patient body" to then identify surview start/end position and patient orientation, and support detection of the center of patient anatomy. This implies annotation of patient body parts and anatomical landmarks in imaging data for training purposes, but specific details on its establishment are not provided.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Healthcare (Suzhou) Co., Ltd. % Shiguang An Advanced Regulatory Engineer No. 258, Zhong Yuan Road, Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA
Re: K203514
Trade/Device Name: Precise Position Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: May 19, 2021 Received: May 19, 2021
Dear Shiguang An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
June 17, 2021
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K203514
Device Name
Precise Position
Indications for Use (Describe)
The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow.
- Patient orientation identification
- · Surview range recommendation
- · Automatic centering the patient anatomy
- · Provide visual images of patient on the table
Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.
Patient population limitation: Patient younger than 16 years are not supported.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
collection of information is estimated to average 79 hours p time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | November 18, 2020 |
|---|---|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.No. 258, ZhongYuan Road, Suzhou Industrial Park, 215024Suzhou, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINAEstablishment Registration Number: 3009529630 |
| Primary ContactPerson: | Shiguang AnAdvanced Regulatory EngineerPhone: +86-13940106467E-mail: shiguang.an@philips.com |
| Secondary ContactPerson: | Erhong WangSenior Regulatory Affairs ManagerPhone: +86-13021019589E-mail: erhong.wang@philips.com |
| Device Name: | Precise position |
| Classification: | Classification NameComputed tomography x-ray systemClassification Regulation:21CFR §892.1750Classification Panel:RadiologyDevice Class:Class IIPrimary product code:JAK |
| Predicate Device: | Trade Name:Philips Incisive CTManufacturer:Philips Healthcare (Suzhou) Co., Ltd.510(k) Clearance:K180015-March 20, 2018Classification Regulation:21 CFR, Part 892.1750Classification Name:Computed tomography x-ray systemClassification Panel:RadiologyDevice Class:Class IIProduct CodeJAK |
| Reference Device: | Manufacturer:Auto Positioning |
| GE Hangwei Medical SystemCo., Ltd. | |
| 510(k) Clearance: | K192956 (January 16, 2020) |
| Classification Regulation: | 21 CFR, Part 892.1750 |
| Classification Name: | Computed tomography x-raysystem |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Product Code | JAK |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
| Devicedescription: | Precise Position is an optional feature to assist user for position thepatient before the body examination such as CT scan. The purposeof this feature is to reduce the patient position time via the cameradetection and calculation result. It includes automatic detect patientorientation, patient anatomy scan range and center of patientanatomy. |
|---|---|
| Precise Position including a camera with both color and depthfunction is installed in the ceiling of the scan room, in such a wayto cover the entire patient on the patient table. The camera controland image data transmit via the high speed fiber and copper hybridUSB cable. The power supply of the camera is from the gantry.Precise position adopts the AI algorithm (Convolution NeuralNetwork) to detect the joints of the patient body, and then identifysurview start/end position and patient orientation. The algorithmcan also support detect center of patient anatomy. | |
| Limitation for Precise PositionThere is no limitation for Precise Position except below items:• Patients below the age of 16 are not supported.• Decubitus orientations are not supported. | |
| The Precise Position display results may get affected by thefollowing conditions:• When the patient is covered by sheet, blanket etc.,• When the patient is not completely covered by the ceiling cameraview, e.g. blocked by the gantry or out of camera's FOV etc.• When the patient is wearing clothes that reflects light, e.g. plastic-like clothes.• When the patient is wearing black clothes.• When the patient is wearing thick clothes.• When there are other people around the patient. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the words "Traditional 510(k)". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Traditional" appearing first, followed by "510(k)".
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
| Indications for Use: | The Precise Position is intended for use with Philips Incisive CT systems. The device provides the following guided workflow.Patient orientation identification Surview range recommendation Automatic centering the patient anatomy Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.Patient population limitation: Patient younger than 16 years are not supported. |
|---|---|
| ---------------------- | ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the primary focus of the image and dominates the visual space.
| Fundamentalscientifictechnology: | Based on the information provided above, the precise position isconsidered substantially equivalent to the primary currentlymarketed and predicate device Philips Incisive CT (K180015,20/March/2018) in terms of fundamental scientific technology. | ||
|---|---|---|---|
| Attribute | Predicate DevicePhilips Incisive CT(K180015) | Proposed Device PrecisePosition | |
| Patientpositioningworkflow | Manually positionthe patient via thecouch motion buttonand laser. Normally,need several roundadjustments by user. | Precise position providesthe auto workflow to setthe surview start/endposition, center of patientanatomy, and patientorientation. | |
| Detectionalgorithm | The manualworkflow does notneed specialalgorithm. | Precise position adopt theAI algorithm (ConvolutionNeural Network) to detectthe joints of the patientbody automatically, andthen automatic detectsurview start/end position,center of patient anatomyand patient orientation. | |
| Hardwareneed tosupport thepatientpositioning | Manual patientpositioning does notneed any uniquehardware excepttraditional exitinghardware on the CTor other system.Exiting hardwareincludes tablemotion button(in/out, up/down) ongantry panel orCTBOX, lasers andetc. | Precision position featureneed color and depthcamera, high speed USBtransmission cable, andpower cable, and relativethe enhanced workflow.The tradition exitingpatient positioninghardware still be availablefor user all the time, it isconvenient to switchbetween the cameraautomatically detectingmode and manual mode. | |
| Patientpopulation | All ages | More than 16 years old. | |
| Environmentof use | Hospitals, outpatientclinics, researchinstitutions, andother clinicalfacilities. | Hospitals, outpatientclinics, researchinstitutions, and otherclinical facilities. | |
| Summary of Non-Clinical | The Precise Position complies with the following international andFDA-recognized consensus standards: |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
| Performance data: | • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements. For Basic Safety and Essential Performance (IEC 60601-1:2012, MOD).FDA/CDRH recognition number 19-4• IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests.FDA/CDRH recognition number 19-8• ISO 14971 Medical devices – Application of risk management to medical devices. FDA/CDRH recognition number 5-40.• IEC 62304:2015, Medical device software -- Software life cycle processesFDA/CDRH recognition number 13-79There are no risks identified in risk management documentation that require clinical data for the purpose of clinical evaluation; Risk Management Plan as Appendix 001, Risk Management Report as Appendix 002, and Risk Management Matrix as Appendix 003 of Precise Position.There are no clinical risks identified by the evaluated clinical data.Sufficient evidence is available to demonstrate the ability of Precise Position achieve the intended performances during normal condition of use.Full consistency exists between the state-of-the-art, the evaluated data, the risk management documentation and the information materials supplied.Therefore, the Precise Position is substantially equivalent to the primary currently marketed and predicate device (K180015, 20/March/2018) in terms of safety and effectiveness. |
|---|---|
| Summary ofClinical Data: | Precise Position is evaluation covered total 80 clinical scan positions in which 40 cases used with Precise Position and another 40 cases without the usage of Precise Position in order to meet the sample size calculation of 40 cases performed average on a CT scanner per day.The thorough clinical evaluation of this feature is done by 5 Clinical experts.The testing did not deliver radiation to volunteers as Precise Position is only for supporting the positioning of the patient for the localization radiograph and therefore “radiation" was not required for those volunteers. |
| The clinical evaluation was done with 3 major objectives as follows, | |
| To calculate the time saved per surview planning with and without | |
| Precise Position. | |
| To calculate the accuracy of vertical (iso)center positioning and | |
| Surview (horizontal) start /end position with and without PrecisePosition | |
| Intra operator consistency in positioning the patient and surview scanrange | |
| The summary of clinical evaluation testing was clearly demonstratingthe objectives in terms of user benefits as shown belowAverage Time taken to position without Precise Position and withPrecise Position is recorded. Time at user select the patient fromGantry Panel, to time when user pressed Go button from Gantry Panelis measured and concluded that up to 23% time reduction in patientpositioning achieved with “Precise Positioning workflow”. | |
| The average offset in mm for vertical (iso) center position among 5operators without Precise Position and with Precise Position isrecorded and results shown that with "Precise Position" the verticalposition accuracy is increased up to 50%. | |
| Standard deviation in mm for vertical (iso)center positioning &Surview (horizontal) start position among 5 operators without PrecisePosition and with Precise Position is recorded and results shown up to70% increase in Vertical and horizontal position consistency withPrecise Position. | |
| Overall, the Precise Position Clinical Review Report. (Appendix_007)concluded that Philips Incisive CT systems with Precise Position,under normal condition of use, perform as intended, are safe for itsintended use and have a favorable benefit-risk ratio. Further clinicalinvestigations are not necessary, as sufficient evidence exists tosupport these conclusions. | |
| Substantial | The Precise Position is substantially equivalent to the primary |
| Equivalence | currently marketed and predicate device (K180015, 20/March/2018) in |
| Conclusion: | terms of design features, fundamental scientific technology, |
| indications for use, and safety and effectiveness. Additionally, | |
| substantial equivalence was demonstrated with non-clinical | |
| performance tests, which complied with the requirements specified in | |
| the international and FDA-recognized consensus standards, IEC 62304 | |
| and ISO 14971. The results of these tests demonstrate that Precise | |
| Position met the acceptance criteria and is adequate for this intended | |
| use. |
{8}------------------------------------------------
Traditional 510(k)
Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.