K Number
K202366
Manufacturer
Date Cleared
2021-01-29

(163 days)

Product Code
Regulation Number
870.1210
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MAGIC Infusion Catheter is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The MAGIC Infusion Catheter may be used for controlled, regional infusion into selected vessels and is not intended for use in the coronary vasculature.

Device Description

The MAGIC Infusion Catheters are microcatheters designed with progressive suppleness and a rigid proximal shaft to allow control and navigability in the vascular system. By their diameter and progressive suppleness features, these catheters are specifically designed for catheterization of small diameter, sinuous distal vessels. The catheter body and its distal tip (ring) are radiopaque to provide visibility under fluoroscopy. The MAGIC Infusion Catheter has an external hydrophilic coating which provides a lubricious surface during use. A PTFE coated mandrel is included inside the MAGIC Infusion Catheter to provide support during product preparation and insertion through the guide catheter. The MAGIC Infusion Catheters and accompanying support mandrels are provided sterile, non-pyrogenic, and intended for single use only.

AI/ML Overview

The provided document describes the MAGIC Infusion Catheter and its 510(k) summary. The study is a bench and biocompatibility testing designed to demonstrate substantial equivalence to a predicate device.

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Bench TestAcceptance CriteriaResults
Dimensional and Physical AttributesShall meet the required dimensional measurements and physical attributes per established specifications.PASS
Visual InspectionShall meet the established acceptance criteria for surface defects and contamination.PASS
Torque Strength TestCatheters shall withstand the established acceptance criteria for torque strength without breaking.PASS (Device integrity suitable for intended clinical use)
Catheter Tensile TestShall meet ISO 105551-1, Annex BPASS
Corrosion ResistanceShall meet ISO 105551-1, Annex APASS
Flow RateReference DataN/A (Note: Not an acceptance criterion for passing/failing)
Dynamic PressureCatheters shall withstand the maximum infusion pressure (as indicated on the product labeling) under dynamic conditions. (Adopted from ISO 105551-1)PASS (Device met maximum labeled infusion pressure)
Static Burst PressureCatheters shall withstand the maximum infusion pressure (as indicated on the product labeling) under static conditions. (Adopted from ISO 105551-1)PASS (Device integrity suitable for intended clinical use)
Pressure and Hub Test (Fluid Leakage)Catheters shall not leak when injected with water at the maximum infusion pressure (as indicated on the product labeling) for a 30 second duration.PASS (Device integrity suitable for intended clinical use)
Simulated Use / Coating Lubricity and DurabilityCatheters shall be tracked using a benchtop model under simulated use conditions per product IFU, and meet established acceptance criteria for product performance, trackability, coating lubricity and durability.PASS (Results comparable to predicate device)
Kink ResistanceCatheter body shall be wrapped around different pin gauges at various locations and meet the expected acceptance criteria for kink resistance. No kinks shall be noted during Simulated Use.PASS (Results comparable to predicate device)
RadiopacityMarker band shall be detectable under radiographic techniques. (Per ASTM F640-12)PASS
ParticulatesShall meet USP criteria. No particulates shall be noted pre and post Simulated Use.PASS (Results comparable to predicate device)
Biocompatibility: Cytotoxicity – MEM Elution TestISO 10993-5 (implicitly: PASSED)PASSED
Biocompatibility: Cytotoxicity – Agar Overlay AssayISO 10993-5 (implicitly: PASSED)PASSED
Biocompatibility: Sensitization – Guinea Pig Maximization TestISO 10993-10 (implicitly: PASSED)PASSED
Biocompatibility: Irritation – Intracutaneous ReactivityISO 10993-10 (implicitly: PASSED)PASSED
Biocompatibility: Acute Systemic Toxicity – Systemic Injection TestISO 10993-11 (implicitly: PASSED)PASSED
Biocompatibility: Systemic Toxicity – Rabbit Pyrogen TestISO 10993-11 (implicitly: PASSED)PASSED
Biocompatibility: Hemocompatibility – HemolysisISO 10993-4 / ASTM F756 (implicitly: PASSED)PASSED
Biocompatibility: Hemocompatibility – Complement ActivationISO 10993-4 (implicitly: PASSED)PASSED
Biocompatibility: Hemocompatibility – Thrombogenicity: Canine ModelISO 10993-4 (implicitly: PASSED)PASSED
Biocompatibility: Hemocompatibility – Thrombogenicity: Partial Thromboplastin Time AssayISO 10993-4 / ASTM F2382-18 (implicitly: PASSED)PASSED
Biocompatibility: Hemocompatibility – Thrombogenicity: Platelet/Leukocyte Count AssayISO 10993-4 / ASTM F2888-19 (implicitly: PASSED)PASSED

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample sizes for each specific bench test. It generally refers to "the 1.2F MAGIC Infusion Catheter (subject device)" as being tested. For biocompatibility, it states "The 1.2F MAGIC Infusion Catheter was assessed for biocompatibility."

The provenance of the data is retrospective, as it refers to testing conducted on the subject device to demonstrate substantial equivalence to a previously cleared predicate device. The document does not specify the country of origin of the data beyond the applicant "Balt USA, LLC" being located in Irvine, CA, USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device submission. The study describes physical and biological bench testing of a medical device, not a diagnostic or AI-powered device that requires expert-established ground truth from a test set of clinical images or data. The "ground truth" for these tests is defined by established engineering standards and biological testing protocols (e.g., ISO standards, ASTM standards, USP criteria).

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving expert readers for diagnostic accuracy. This submission relies on objective physical and biological measurements against predefined acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is not applicable. This submission is for a physical medical device (infusion catheter) and does not involve AI or human readers for diagnostic interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. This is not an AI-powered device.

7. The Type of Ground Truth Used

The ground truth for the bench tests is based on established engineering specifications, international standards (ISO 105551-1, ASTM F640-12, USP ), and biological testing protocols (ISO 10993-series, ASTM F756, ASTM F2382-18, ASTM F2888-19). These standards define the acceptable performance characteristics and biological response for such medical devices.

8. The Sample Size for the Training Set

This information is not applicable. This is not an AI/machine learning device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for this type of device.

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).