(30 days)
SpaceFlex Knee disposable cement spacer molds are indicated for use to mold a temporary Total Knee Replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (G3A Bone Cement), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
G21 SpaceFlex Knee system are disposable cement spacer molds made of polypropylene (PP) intended to be filled with low-viscosity polymethyl methacrylate bone cement directly in the operating room. Once the bone cement has hardened, the SpaceFlex Knee system creates a polymethyl methacrylate based bone cement spacer for patients undergoing a two-stage procedure following a periprosthetic joint infection. The device can be used in either the left or right knee joint. The SpaceFlex Knee molds are single-use and cannot be re-used or re-sterilized.
This document is a 510(k) summary for a medical device called "SpaceFlex Knee - 80mm Size." It aims to demonstrate that this new device is substantially equivalent to existing predicate devices.
1. Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Performance Specification) | Reported Device Performance |
|---|---|
| Device functions as intended post-testing. | Functioned as intended. |
| All results are satisfactory. | All results were satisfactory. |
| Meets all performance specifications. | Met all performance specifications. |
| Passed Dimensional and Visual Characterization | Test passed. |
| Passed Rupture Test | Test passed. |
| Passed Fatigue Test | Test passed. |
| Passed Elution Test | Test passed. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample size for the test set. It mentions "The mechanical properties of the SpaceFlex knee System 80mm size were tested in accordance with applicable international standards," implying that a sufficient number of samples were tested to meet the requirements of those standards.
The provenance of the data is directly from the manufacturer, G21 S.r.l., whose address is in Italy. The testing described appears to be prospective performance testing conducted on the device itself, rather than retrospective analysis of patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:
This type of study does not involve establishing ground truth through expert review in the way a diagnostic AI study would. The performance data presented here relates to mechanical and material properties of the device, not clinical diagnostic accuracy. Therefore, the concept of "experts" establishing ground truth in this context is not applicable. The "ground truth" for these tests would be the established international standards and engineering specifications.
4. Adjudication Method for the Test Set:
Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or patient outcomes, where there might be disagreements among expert reviewers. This document describes laboratory-based mechanical and material performance testing of a physical device. Therefore, adjudication methods are not applicable here. The results of these tests are objective measurements against predefined standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. This study is focused on the mechanical and material performance of a physical medical device (a cement spacer mold for knee replacement), not on the diagnostic accuracy or effectiveness of an AI algorithm interpreted by human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No, a standalone (algorithm only) performance study was not done. This document is about the physical characteristics and performance of a medical device, not an AI algorithm.
7. The Type of Ground Truth Used:
The "ground truth" for this study is based on predefined international standards and engineering specifications for the mechanical properties and material characteristics of medical devices, specifically those for knee joint prostheses and bone cement. The tests (Dimensional and Visual Characterization, Rupture Test, Fatigue Test, Elution Test) verify that the device meets these established physical and material requirements.
8. The Sample Size for the Training Set:
This document describes the performance testing of a physical medical device, not the development or training of an AI algorithm. Therefore, the concept of a "training set" and its sample size is not applicable.
9. How the Ground Truth for the Training Set was Established:
As this study does not involve an AI algorithm with a training set, the question of how ground truth for a training set was established is not applicable.
{0}------------------------------------------------
August 13, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
G21 S.r.l % Barry Sands President and Founder RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913
Re: K201960
Trade/Device Name: SpaceFlex Knee - 80mm Size Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBB Dated: July 10, 2020 Received: July 14, 2020
Dear Barry Sands:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201960
Device Name
SpaceFlex Knee - 80mm Size
Indications for Use (Describe)
SpaceFlex Knee disposable cement spacer molds are indicated for use to mold a temporary Total Knee Replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (G3A Bone Cement), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
__ Over The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Papervork Reduction Act of 1995
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this ourden, to
Department of Health and Human Services Food and Drug Administration Office of Chief Information Office r Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently val d OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
510(k) SUMMARY
G21 's SpaceFlex 80mm size 510k Submission (K201960)
l. SUBMITTER
G21 S.r.l. Via S. Pertini, 8 41039 S. Possidonio (MO) Italy
Phone: +39 0535 30312 Fax: +39 0535 417332
Contact Person: Barry E. Sands RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, MA 01913 Phone: (978) 358-7307 Fax: +1 (978) 477-0206
Date Prepared: February 12, 2020
ll. DEVICE
| Trade/Device Name: | SpaceFlex Knee – 80mm Size |
|---|---|
| Common or Usual Name: | Temporary Knee Prosthesis |
| Classification Name: | Polymethylmethacrylate (PMMA) bone cement, antibioticknee joint patellofemorotibial polymer/metal/polymersemi-constrained cemented prosthesis |
| Regulation Number: | 21 CFR 888.3027, 21 CFR 888.3560 |
| Regulatory Class: | Class II |
| Product codes | MBB, JWH |
III. PREDICATE DEVICES
Primary Predicate: SpaceFlex Knee, G21 S.R.L. (K190216)
Additional Predicate: StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis, Biomet, Inc. (K161273)
{4}------------------------------------------------
IV. DEVICE DESCRIPTION
G21 SpaceFlex Knee system are disposable cement spacer molds made of polypropylene (PP) intended to be filled with low-viscosity polymethyl methacrylate bone cement directly in the operating room. Once the bone cement has hardened, the SpaceFlex Knee system creates a polymethyl methacrylate based bone cement spacer for patients undergoing a two-stage procedure following a periprosthetic joint infection. The device can be used in either the left or right knee joint. The SpaceFlex Knee molds are singleuse and cannot be re-used or re-sterilized.
Intended Use:
SpaceFlex Knee is a mold intended to assist in the creation of a temporary disposable cement spacer for a total knee replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process.
Indications for Use:
SpaceFlex Knee disposable cement spacer molds are indicated for use to mold a temporary total knee replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (G3A Bone Cement), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the subject device and the predicate device are disposable cement spacer molds with two separate molds - femoral and tibial. The subject devices possess identical technical characteristics as the primary predicate. The only difference between subject and predicate devices is a larger dimensional footprint.
The technological features of the SpaceFlex Knee are the same as the primary predicate device.
VI. PERFORMANCE DATA
The mechanical properties of the SpaceFlex knee System 80mm size were tested in accordance with applicable international standards. In all instances the device
{5}------------------------------------------------
functioned as intended and all results were satisfactory and met all performance specifications. The testing performed includes:
-
- Dimensional and Visual Characterization
-
- Rupture Test
-
- Fatigue Test
-
- Elution Test
VII. CONCLUSION
The SpaceFlex Knee – 80mm and its primary predicate, G21 SpaceFlex Knee system, have the same intended use and indications for use, technological characteristics and principle of operation. The material differences between the molds in the subject and the predicate device do not raise any significant new risks. Through performance and mechanical testing and material information it has been established that these differences do not present any new issues of safety or effectiveness. The SpaceFlex Knee – 80mm is substantially equivalent to the primary predicate, G21 SpaceFlex Knee System, and additional predicate StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis.
§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.
(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”