K Number
K201910
Device Name
EZTrack
Manufacturer
Date Cleared
2020-12-22

(166 days)

Product Code
Regulation Number
882.1400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EZTrack is intended for use by a trained/qualified EEG technologist or physician on both adult and pediatric subjects with focal or multifocal epilepsy at least 3 years of age for the visualization of human brain from analysis of electroencephalographic (EEG) signals produced by electrically active tissue of the brain. EZTrack calculates and displays the Fragility Index, a quantitative index based on an analysis of spatiotemporal EEG patterns that is intended for interpretation by trained physicians to aid in the evaluation of patients with focal or multifocal epilepsy.

The device does not provide any diagnostic conclusion about the patient's condition to the user and should be interpreted along with other clinical data, including the original EEG, medical imaging, and other standard neurological and neuropsychological assessments.

Device Description

EZTrack is a web-based software-only device that allows visualization of human brain function based on the analysis of electroencephalographic (EEG) signals. The EZTrack algorithm produces a fragility score for each EEG recording node. The EZTrack fragility values are shown to correlate with regions that clinicians have annotated as seizure onset zones (SOZ) prior to resective surgery, and may be used in conjunction with other clinical data such as EEG, medial imaging, neuropsychological testing, and other neurologic assessments in order to aid in the evaluation of patients with focal or multifocal epilepsy. The device does not provide any diagnostic conclusion about the patient's condition. EZtrack displays the fragility of each EEG channel in a heatmap to aid in interpretation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the EZTrack device:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state acceptance criteria in terms of specific performance thresholds (e.g., sensitivity, specificity, AUC). Instead, the performance is described in terms of a statistically significant correlation and an effect size.

Acceptance Criterion (Implicit)Reported Device Performance
Fragility data correlates with clinically annotated Seizure Onset Zone (SOZ) and differentiates treatment success from failure.EZTrack demonstrated a statistically significant difference (p-value=0.02) between the successful and failed Confidence Statistic distributions. An average effect size difference between the two groups of 0.627 was observed, meaning fragility had a 0.627 higher standardized confidence in the clinically annotated SOZ in successful outcomes compared to failed outcomes.
Software meets verification and validation requirements.Software verification and validation testing were conducted and documentation provided as recommended by FDA guidance. The software was considered a Moderate Level of Concern.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 91 patients (comprising 462 seizures).
    • 44 patients had successful outcomes (seizure-free).
    • 47 patients had failed outcomes (seizure recurrence).
  • Data Provenance: Retrospective study. The country of origin is not explicitly stated in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not explicitly stated, but it mentions "clinicians" and "Consensus agreement of the spatial distribution of visual EEG signatures together with pre-implantation data were used to construct the clinically annotated SOZ." This implies multiple clinicians were involved in a consensus process.
  • Qualifications of Experts: The text refers to "clinicians" attempting to identify visual EEG signatures to isolate the SOZ during invasive monitoring. It also mentions "trained physicians" who would interpret the Fragility Index. Without further detail, it's difficult to specify exact qualifications (e.g., "Radiologist with 10 years of experience"), but they are clearly medical professionals with expertise in EEG and epilepsy.

4. Adjudication Method for the Test Set

The adjudication method for establishing the clinically annotated SOZ appears to be based on clinician consensus using "visual EEG signatures (e.g. HFOs, spikes, or burst activity) together with pre-implantation data." The text does not specify a numerical adjudication method like "2+1" or "3+1."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or described in the provided text. The study focused on the EZTrack algorithm's correlation with SOZ and outcome, not on how human readers perform with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study was conducted. The clinical study described is a retrospective study that demonstrates the EZTrack algorithm's output (fragility data) correlates with the clinically annotated Seizure Onset Zone (SOZ) and differentiates between successful and failed patient outcomes. The text explicitly states, "EZTrack demonstrated a statistically significant difference (p-value=0.02) between the successful and failed Confidence Statistic distributions, and an average effect size difference between the two groups of 0.627." This is a direct measurement of the algorithm's performance without human intervention in the analysis.

7. Type of Ground Truth Used

The ground truth used was expert consensus / clinical outcome data.

  • Clinically annotated SOZ: Established by "clinicians" via "consensus agreement of the spatial distribution of visual EEG signatures together with pre-implantation data." This implies a form of expert consensus derived from clinical evaluation.
  • Patient Outcome: Categorized as "seizure free (success), or having seizure recurrence (failure) at their 6-12 months post-op evaluations." This is objective outcomes data used to assess the clinical relevance of the SOZ identification.

8. Sample Size for the Training Set

The document does not provide information regarding the sample size used for the training set. The clinical study described is a retrospective study used for performance evaluation, not necessarily for training the algorithm.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any potential training set was established. The clinical study described focuses on evaluating the device's performance against established clinical SOZ and patient outcomes.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Neurologic LLC % Joshua Crist Consultant Biologics Consulting 1555 King Street Alexandria, Virginia 22314

Re: K201910

Trade/Device Name: EZTrack Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT Dated: November 19, 2020 Received: November 20, 2020

Dear Joshua Crist:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201910

Device Name EZTrack

Indications for Use (Describe)

EZTrack is intended for use by a trained/qualified EEG technologist or physician on both adult and pediatric subjects with focal or multifocal epilepsy at least 3 years of age for the visualization of human brain from analysis of electroencephalographic (EEG) signals produced by electrically active tissue of the brain. EZTrack calculates and displays the Fragility Index, a quantitative index based on an analysis of spatiotemporal EEG patterns that is intended for interpretation by trained physicians to aid in the evaluation of patients with focal or multifocal epilepsy.

The device does not provide any diagnostic conclusion about the patient's condition to the user and should be interpreted along with other clinical data, including the original EEG, medical imaging, and other standard neurological and neuropsychological assessments.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

SUBMITTER 1.

Submitter:Neurologic LLC6510 Chesterfield AvenueMcLean, Virginia 22101617-875-9380
Contact Person:Sridevi SarmaPresident and Co-FounderNeurologic LLC6510 Chesterfield AvenueMcLean, Virginia 22101617-875-9380
Submission Correspondent:Joshua CristConsultantBiologics Consulting1555 King StreetAlexandria, VA 22314216-387-0199jcrist@biologicsconsulting.com
Date Prepared:November 13, 2020

2. DEVICE

Name of Device:EZTrack
Common or Usual Name:Electroencephalograph Software
Classification Name:882.1400 - Electroencephalograph
Regulatory Class:II
Product Code:OLT

3. PREDICATE DEVICE

Predicate Device Name:Persyst 14 EEG Review and Analysis Software
---------------------------------------------------------------------

{4}------------------------------------------------

Manufacturer:Persyst Development Corporation
510(k) Number:K182181
Reference Devices:GeoSourceElectrical Geodesics, Inc.K092844

DEVICE DESCRIPTION 4.

EZTrack is a web-based software-only device that allows visualization of human brain function based on the analysis of electroencephalographic (EEG) signals. The EZTrack algorithm produces a fragility score for each EEG recording node. The EZTrack fragility values are shown to correlate with regions that clinicians have annotated as seizure onset zones (SOZ) prior to resective surgery, and may be used in conjunction with other clinical data such as EEG, medial imaging, neuropsychological testing, and other neurologic assessments in order to aid in the evaluation of patients with focal or multifocal epilepsy. The device does not provide any diagnostic conclusion about the patient's condition. EZtrack displays the fragility of each EEG channel in a heatmap to aid in interpretation.

ട്. INDICATION FOR USE

EZTrack is intended for use by a trained/qualified EEG technologist or physician on both adult and pediatric subjects with focal or multifocal epilepsy at least 3 years of age for the visualization of human brain function from analysis of electroencephalographic (EEG) signals produced by electrically active tissue of the brain. EZTrack calculates and displays the Fragility Index, a quantitative index based on an analysis of spatiotemporal EEG patterns that is intended for interpretation by trained physicians to aid in the evaluation of patients with focal or multifocal epilepsy.

The device does not provide any diagnostic conclusion about the patient's condition to the user and should be interpreted along with other clinical data, including the original EEG, medical imaging, and other standard neurological and neuropsychological assessments.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6. WITH THE PREDICATE DEVICE

Comparison of Indications

The predicate device, Persyst 14 (K182181) has the following indications for use:

  1. Persyst 14 EEG Review and Analysis Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. The device is intended to be used by qualified

{5}------------------------------------------------

medical practitioners who will exercise professional judgment in using the information.

  1. The Seizure Detection component of Persyst 14 is intended to mark previously acquired sections of adult (greater than or equal to 18 vears) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20 svstem.

  2. The Spike Detection component of Persyst 14 is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in patients at least one month old. Persyst 14 Spike Detection performance has not been assessed for intracranial recordings.

Persyst 14 includes the calculation and display of a set of qualitative measures intended to 4. monitor and analyze the EEG waveform. These include FFT, Rhythmicity, Peak Envelope, Artifact Intensity, Amplitude, Relative Symmetry and Suppression Ratio. Automatic event marking is not applicable to the quantitative measures. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

  1. Persyst 14 displays physiological signals, including the calculation and display of a heart rate measurement based on the ECG channel in the EEG recording, which are intended to aid in the analysis of an EEG. Heart rate measurement of Persyst 14 is not applicable to patients with pacemaker and/or active implantable devices.

  2. The aEEG functionality included in Persyst 14 is intended to monitor the state of the brain. The automated event marking function of Persyst 14 is not applicable to aEEG.

  3. Persyst 14 provides notifications for seizure detection, quantitative EEG and aEEG that can be used when processing arecord during acquisition. These include an on screen display and the optional sending of an email message. Delays of up to several minutes can occur between the beginning of a seizure and when the Persyst 14 notifications will be shown to a user. Persyst 14 notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

  4. Persyst AR (Artifact Reduction) is intended to reduce EMG, eye movement, and electrode artifacts in a standard 10-20 EEG recording. AR does not remove the entifact signal, and is not effective for other types of artifacts. AR may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.

9. This device does not provide any diagnostic conclusion about the patient's condition to the user.

The predicate device has indications for use to output various quantitative and qualitative EEG marks, measurements, and indices. Similarly, EZTrack has indications for use to output and

{6}------------------------------------------------

display a novel, quantitative EEG index (the Fragility Index) that is intended to represent a measurement of node instability to aid in the evaluation of patients with focal epilepsy. Both devices analyze EEG data with software-only algorithms and display indices to the end user. EZTrack differs from the Persyst 14 indications for use in the specific analyses conducted, the display method, and in the target patient population (patients with focal or multi-focal epilepsy). These differences do not raise new types of safety and effectiveness questions and are supported by performance testing discussed below.

Comparison of Technology

EZTrack and the Persyst 14 (K182181) are both software-only devices that provide insights to the user about electrical activity in the brain. Both devices rely on algorithms that process raw EEG data in order to provide this information to the user. Neither device provides any diagnostic conclusion about the patient's condition to the user

The main technological differences between the two devices are the algorithms and resulting data visualization methods. The predicate Persyst 14 (K182181) performs a wide variety of calculations on EEG to provide various quantitative and qualitative EEG marks, measurements, and indices, such as seizure detection, spike detection, event marking, artifact reduction, and trends/qualitative measures. These data may be visualized with markups and notifications over EEG data, as well as various output visualizations for trends, such as FFT spectrogram, FFT power ratio, FFT spectral edge, relative asymmetry index, suppression ratio, spike density, and seizure probability trends.

EZTrack provides a different analysis than the predicate, i.e. a fragility analysis, which is a measure of node instability to evaluate whether the activity for each electrode is representative of epileptogenic activity. EZTrack can be used on scalp EEG. electrocorticography (ECoG), or stereoelectroencephalography (SEEG) data to aid in evaluation of patients with focal or multifocal epilepsy. EZTrack's resulting analyses are displayed in a heatmap.

EZTrack differs in the technological features of the algorithm and the method by which the results are displayed to the user. The differences in indications and technological features for EZTrack do not raise new types of safety and effectiveness questions. The same questions for each device include whether appropriate software verification and validation data have been provided to support the software, and whether clinical validation testing has confirmed the software meets the users' needs for the intended use of the device. This performance testing has been provided for EZTrack and therefore supports substantial equivalence of the proposed device to the predicate.

A comparison of technological features is provided in Table 1 below.

Proposed DevicePrimary Predicate DeviceReference Device
510(k) NumberTBDK182181K092844
ApplicantNeurologic LLCPersyst DevelopmentCorporationElectrical Geodesics, Inc.
Proposed DevicePrimary Predicate DeviceReference Device
Device NameEZTrackPersyst 14 EEG Review andAnalysis SoftwareGeoSource
ClassificationRegulation882.1400Electroencephalograph882.1400Electroencephalograph882.1400Electroencephalograph
Product CodeOLTOMB, OLT, OMAOLX
Indications forUseEZTrack is intended for use by atrained/qualified EEG technologistor physician on both adult andpediatric subjects with focal ormultifocal epilepsy at least 3 yearsof age for the visualization of humanbrain function from analysis ofelectroencephalographic (EEG)signals produced by electricallyactive tissue of the brain. EZTrackcalculates and displays the FragilityIndex, a quantitative index based onan analysis of spatiotemporal EEGpatterns that is intended forinterpretation by trained physiciansto aid in the evaluation of patientswith focal or multifocal epilepsy.The device does not provide anydiagnostic conclusion about thepatient's condition to the user andshould be interpreted along withother clinical data, including theoriginal EEG, medical imaging, andother standard neurological andneuropsychological assessments.See Section 6 above.GeoSource is intended for use bya trained/qualified EEGtechnologist or physician on bothadult and pediatric subjects atleast 3 years of age for thevisualization of human brainfunction by fusing a variety ofEEG information with renderedimages of an idealized head modeland an idealized MRI image
Software-OnlyYesYesYes
EEG DataSourceAny EEG, electrocorticography(ECOG), orstereoelectroencephalography(SEEG) data in European DataFormat or Brain Vision formatAny EEG,electrocorticography(ECOG), orstereoelectroencephalography (SEEG) data.Geodesic EEG System usingNet Station
Methods ofCalculationNetwork Fragility AnalysisVarious analyses, includingseizure detection, spikedetection, event marking,artifact reduction, andtrends/qualitative measures.Idealized Head Model(average)LORETA, LAURA,sLORETA
Proposed DevicePrimary Predicate DeviceReference Device
Method ofDisplayFragility HeatmapMarkups and notificationsover EEG data, as well asvarious outputvisualizations for trends,such as FFT spectrogram,FFT power ratio, FFTspectral edge, relativeasymmetry index,suppression ratio, spikedensity, and seizureprobability trend.Idealized MRI (average)
MultipleTimepointsAs x-axis in heatmapAs x-axis in live monitoringor analysesFrame-by-Frame
ComputationLocationPerformed on EZTrack ServersPerformed Locally onSystem, some remoteanalyses availablePerformed Locally on System
OperatingSystemWeb-Based Platform, accessibleon Windows, Mac OS, LinuxWindows 10Mac OS

Table 1: Device Comparison Table

{7}------------------------------------------------

{8}------------------------------------------------

7. DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility

There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.

Electrical Safety and electromagnetic compatibility (EMC)

Not applicable. The subject device is a software-only device. It contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a Moderate Level of Concern. This performance testing was also provided for the predicate, and this testing therefore supports substantial equivalence of the EZTrack device.

Bench Testing

{9}------------------------------------------------

Not Applicable. Bench testing was not necessary to establish the substantial equivalence of this device.

Animal Study

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

Clinical Study

A retrospective study of 91 patients (462 seizures, 44 successes, 47 failures, ages 3-65) demonstrates that the EZTrack algorithm provides fragility data that correlates with the location of the clinically annotated Seizure Onset Zone (SOZ). Note this fragility data is used to aid in evaluation of patients with focal or multifocal epilepsy and does not provide any diagnostic conclusions about a patient's condition.

During invasive monitoring, clinicians attempted to identify visual EEG signatures (e.g. HFOs, spikes, or burst activity) to isolate the SOZ. Consensus agreement of the spatial distribution of visual EEG signatures together with pre-implantation data were used to construct the clinically annotated SOZ, which was an estimate of the true Epileptogenic Zone. The patients underwent a large resection, or targeted laser ablation, resulting in a resected region, which was generally a super set of the SOZ. Patients were categorized as either seizure free (success), or having seizure recurrence (failure) at their 6-12 months post-op evaluations.

EZTrack demonstrated a statistically significant difference (p-value=0.02) between the successful and failed Confidence Statistic distributions, and an average effect size difference between the two groups of 0.627. On average, fragility had a 0.627 higher standardized confidence in the clinically annotated SOZ in success outcomes, than in failed outcomes.

The Geodesics device (K092844) serves as a reference device to support the acceptability of these performance data collection methods. That is, both EZTrack and the reference device are supported by clinical data that demonstrate correlation of the algorithm output with the zone of tissue that physicians ultimately decide to resect, demonstrating the algorithms' validity as a source of information that may be used within then intended use of the respective devices.

The results of this study demonstrate that the EZTrack outputs a fragility index that can be used to aid in evaluation of patients with focal or multifocal epilepsy. Like the predicate device. clinical data has been provided for EZTrack to demonstrate the device meets the users' needs for the intended use of the device, therefore supporting substantial equivalence.

8. CONCLUSIONS

Both devices are software-only applications that provide algorithmic analyses of EEG data in order to provide insights into the electrical activity in the brain through indices displayed to the end user. The EZTrack device has different technological characteristics when compared to Persyst 14. EZ track differs in the way the EEG data are analyzed and in the way the data are displayed to the user. Neither device provides diagnostic conclusion about the patient's condition.

In summary, EZTrack has the same intended use as the predicate Persyst, has slight differences in the patient population and technological features, reflected in differences in the indications for use language. The differences in technological features/indications do not raise new questions of

{10}------------------------------------------------

safety and effectiveness and are supported by Software Verification/Validation Testing as well as Clinical Testing. The proposed EZTrack device is therefore substantially equivalent to the predicate Persyst 14 EEG Review and Analysis Software (K182181).

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).