K Number
K201378
Device Name
3D Accuitomo 150N
Date Cleared
2021-02-18

(268 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The 3D Accuitomo 150N Model X800N is intended to be used for panoramic tomography including linear tomography and scanogram, cephalometric radiography, and cone beam computed tomography. The 3D Accuitomo 150N Model X800N is an extraoral source X-ray unit that is used for dental and head radiographic examination and diagnosis of teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dentomaxillofacial areas, and hand for maturity assessment, by exposing an X-ray image receptor to ionizing radiation. The device uses a cone shaped X-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 2D or 3D viewing stations. The device is to be operated and used by medical doctors, dentists, and other legally qualified professionals for pediatric and adult patients.
Device Description
The X800N consists of the following components: x-ray arm with extraoral x-ray source assembly faced with x-ray detector, cephalometric support, high-voltage generator, control assembly, patient positioning device, and software containing communication and image construction. The X800N has three main radiographic modes as follows: - Panoramic tomography including linear tomography and scanogram - - Cephalometric radiography - - -Cone beam computed tomography
More Information

K#171012

K#073696, K#073695

No
The document does not mention AI, ML, or related terms, and the description of the device and performance studies focuses on standard imaging and reconstruction techniques.

No

This device is intended for radiographic examination and diagnosis, not for treatment or therapy.

Yes

The device is explicitly stated to be used for "radiographic examination and diagnosis of teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dentomaxillofacial areas, and hand for maturity assessment." This clearly indicates its role in providing diagnostic information.

No

The device description explicitly lists multiple hardware components, including an x-ray arm, detector, generator, and control assembly, in addition to software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The 3D Accuitomo 150N Model X800N is an X-ray imaging device. It uses ionizing radiation to create images of internal structures of the body for diagnostic purposes. It does not perform tests on biological samples.
  • Intended Use: The intended use clearly states it's for "radiographic examination and diagnosis" by exposing an X-ray image receptor to ionizing radiation. This is consistent with an imaging device, not an IVD.

Therefore, the device described is a medical imaging device, specifically an X-ray unit, and not an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

The 3D Accuitomo 150N Model X800N is intended to be used for panoramic tomography including linear tomography and scanogram, cephalometric radiography, and cone beam computed tomography.

The 3D Accuitomo 150N Model X800N is an extraoral source X-ray unit that is used for dental and head radiographic examination and diagnosis of teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dentomaxillofacial areas, and hand for maturity assessment, by exposing an X-ray image receptor to ionizing radiation. The device uses a cone shaped X-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 2D or 3D viewing stations.

The device is to be operated and used by medical doctors, dentists, and other legally qualified professionals for pediatric and adult patients.

Product codes

OAS, MUH

Device Description

The X800N consists of the following components: x-ray arm with extraoral x-ray source assembly faced with x-ray detector, cephalometric support, high-voltage generator, control assembly, patient positioning device, and software containing communication and image construction.

The X800N has three main radiographic modes as follows:

  • Panoramic tomography including linear tomography and scanogram
  • Cephalometric radiography
  • Cone beam computed tomography

Mentions image processing

The examined volume image is reconstructed to be viewed in 2D or 3D viewing stations.

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray

Anatomical Site

teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dentomaxillofacial areas, and hand

Indicated Patient Age Range

pediatric and adult patients

Intended User / Care Setting

medical doctors, dentists, and other legally qualified professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

The images evaluated were obtained in J. Morita's facility from adult and child human skull phantoms made from resin molding on real human dry skulls. The use of phantom image for evaluation was judged to be appropriate in the majority of cases and no evaluator judged it is not suitable to use the phantom image. The images were then ranked by the evaluators according to five categories: "Excellent Quality for Diagnosis," "Acceptable Quality for Diagnosis," Not Acceptable Quality for Diagnosis," "Not reviewed," and "Other."

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Clinical evaluation was performed to evaluate if the CT image obtained by the X800N for general examination and diagnosis of the ENT area is substantially equivalent to that obtained by J. Morita's reference devices, the X550 (K#073696) and MCT-1 (K#073695). Experienced doctors who are using CBCT for their ENT medical practice subjectively evaluated images taken by the X800N, X550, and MCT-1. All the evaluators ranked the images of the X800N as "Excellent Quality for Diagnosis" or "Acceptable Quality for Diagnosis." The overall results were better than that of the X550. And the results of the X800N were comparable to the MCT-1.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

Veraview X800 (K#171012)

Reference Device(s)

Veraviewepocs X550 (K#073696), MCT-1 (K#073695)

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

February 18, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

J. Morita USA, Inc. % Mr. Keith Barritt Attorney Fish & Richardson P.C. 1000 Maine Avenue. S.W., Suite 1000 WASHINGTON DC 20024

Re: K201378

Trade/Device Name: 3D Accuitomo 150N Model X800N Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS, MUH Dated: December 30, 2020 Received: December 31, 2020

Dear Mr. Barritt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

1

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K201378

Device Name 3D Accuitomo 150N Model X800N

Indications for Use (Describe)

The 3D Accuitomo 150N Model X800N is intended to be used for panoramic tomography including linear tomography and scanogram, cephalometric radiography, and cone beam computed tomography.

The 3D Accuitomo 150N Model X800N is an extraoral source X-ray unit that is used for dental and head radiographic examination and diagnosis of teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dentomaxillofacial areas, and hand for maturity assessment, by exposing an X-ray image receptor to ionizing radiation. The device uses a cone shaped X-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 2D or 3D viewing stations.

The device is to be operated and used by medical doctors, dentists, and other legally qualified professionals for pediatric and adult patients.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

K#201378

510(k) Summary J. Morita USA Inc. 3D Accuitomo 150N Model X800N X-ray system for Panoramic, Cephalometric and CBCT imaging

The following information is provided pursuant to 21 CFR 807.92.

807.92(a)(1): Submitter's Name/Address, Contact, and Preparation Date

(i) 510(k) Submitter

Registration No. 2081055 J. Morita USA, Inc. 9 Mason, Irvine, CA 92618, USA Phone: 949-581-9600 Fax: 949-581-9688

(ii) 510(k) Submitter Contact

Keith A. Barritt Fish & Richardson P.C. 1000 Maine Avenue, S.W., Suite 1000 Washington DC 20024, USA Phone: (202) 626-6433 Fax: (202) 783-2331 Email: barritt@fr.com

(iii) Preparation Date

February 17, 2021

807.92(a)(2): Name of Device

Trade or Proprietary Name:3D Accuitomo 150N Model X800N
Common Name:Dental Computed Tomography X-ray System imaging
Device Classification Name:System, X-Ray, Tomography, Computed
Review Panel:892 Radiology device
Product Code:OAS, MUH
Regulation:21 CFR 892.1750 Class II
Device Class:Traditional
510(k) Type:

4

807.92(a)(3): Predicate Device

This device is substantially equivalent for purposes of FDA medical device regulations to the following predicate device:

Trade or Proprietary Name:Veraview X800 (K#171012)
Common Name:Dental Computed Tomography X-ray System imaging
Classification Name:Computed tomography x-ray system
Device Classification Panel:Class II, Radiology
Product Code:OAS
Regulation:21 CFR 892.1750

As discussed further below, the following J. Morita reference devices were used to confirm the suitability of the images of the 3D Accuitomo 150N Model X800N (the "X800N") for the general examination and diagnostic ENT indication being added: Veraviewepocs X550 (K#073696) and MCT-1 (K#073695).

807.92(a)(4): Device Description

The purpose of this 510(k) application is to obtain clearance from FDA for a new ENT indication for J. Morita's already cleared device, the Veraview X800 (K#171012). The new indication is for general examination and diagnosis of the skull including the ENT.

The X800N consists of the following components: x-ray arm with extraoral x-ray source assembly faced with x-ray detector, cephalometric support, high-voltage generator, control assembly, patient positioning device, and software containing communication and image construction.

The X800N has three main radiographic modes as follows:

  • Panoramic tomography including linear tomography and scanogram -
  • Cephalometric radiography -
  • -Cone beam computed tomography

807.92(a)(5): Intended Use

The intended use of the X800N is:

  • Panoramic tomography including linear tomography and scanogram -
  • Cephalometric radiography -
  • Cone beam computed tomography -
  • General examination and diagnosis of the skull including the ENT. -

5

Indication for Use

X800N is intended to be used for panoramic tomography including linear tomography and scanogram, cephalometric radiography, and cone beam computed tomography. X800N is an extraoral source X-ray unit that is used for dental and head radiographic examination and diagnosis of teeth, jaw, oral structure, temporomandibular joint, skull including the ENT, dento-maxillofacial areas, and hand for maturity assessment, by exposing an X-ray image receptor to ionizing radiation.

The device uses a cone shaped X-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 2D or 3D viewing stations.

The device is to be operated and used by medical doctors, dentists, and other legally qualified professionals for pediatric and adult patients.

807.92(a)(6): Technological Characteristics

The predicate device Veraview X800 (K#171012) is identical in regards to mechanical performance, imaging area, and the other safety and efficacy related parameters. For the X800N, exactly the same image as the dental application for the X800 is taken and utilized for the ENT application.

ItemDevice submitted in this 510(k)Predicate device
Product name3D Accuitomo 150NVeraview X800
ModelX800NX800
ManufacturerJ. MORITA MFG. CORP.Same as left
Indications for useX800N is intended to be used for
panoramic tomography including
linear tomography and scanogram,
cephalometric radiography, and
cone beam computed tomography.
X800N is an extraoral source X-
ray unit that is used for dental and
head radiographic examination
and diagnosis of teeth, jaw, oral
structure, temporomandibular
joint, skull including the ENT,
dento-maxillofacial areas, and
hand for maturity assessment, by
exposing an X-ray image receptor
to ionizing radiation.
The device uses a cone shaped X-Veraview X800 is intended to be
used for panoramic tomography
including linear tomography and
scanogram, cephalometric
radiography, and cone beam
computed tomography.
Veraview X800 is an extraoral
source X-ray unit that is used for
dental and head radiographic
examination and diagnosis of
teeth, jaw, oral structure,
temporomandibular joint, skull
including the dento-maxillofacial
areas, and hand for maturity
assessment, by exposing an X-
ray image receptor to ionizing
ItemDevice submitted in this 510(k)Predicate device
ray beam projected on to a flat
panel detector, and the examined
volume image is reconstructed to
be viewed in 2D or 3D viewing
stations.
The device is to be operated and
used by medical doctors, dentists,
and other legally qualified
professionals for pediatric and
adult patients.radiation.
The device uses a cone shaped
X-ray beam projected on to a flat
panel detector, and the examined
volume image is reconstructed to
be viewed in 2D or 3D viewing
stations.
The device is to be operated and
used by dentists and other
legally qualified professionals
for pediatric and adult patients.
Energy used and/or
deliveredAC 120 V, 60 HzSame as left
DesignLoading factorsTube Potential: 60-100kV
Tube Current: 2-10mA
Loading time: max. 18.5sSame as left
X-ray tube
assemblyNominal focal spot: 0.5 at target
angle
Inherent filtration min. 2.5mmALSame as left
X-ray detector 1Internal parts code: D001-15038-
50*
See column in predicate device.Same as left
X-ray detector 2Internal parts code: D001-16044-
50*
CMOS flat panel
Scintillator: CsI
Pixel size for Pan: 0.1 mm
Pixel size for CBCT: 0.1 mm,
0.2 mmSame as left
ItemDevice submitted in this 510(k)Predicate device
X-ray detector 3Internal parts code: D001-04188-
50*
See column in predicate device.Same as left-
Soft-
wareViewer
(general purpose
viewer: not
included in this
submission of
the device)Viewer software:
510K number: K073704
Device name: i-DixelSame as left
Perfor
mancePanoramic
  • spatial
    resolution | Line pair resolution on IEC
    61223-3-4: min. 2.5 LP/mm | Same as left |
    | | | Panoramic
  • noise | Low contrast resolution on IEC
    61223-3-4: min. diameter 2.0 mm
    at High speed mode, min.
    diameter 1.0 mm at High
    resolution mode | Same as left |
    | Cephalometric
  • spatial
    resolution | | Line pair resolution on IEC
    61223-3-4: min. 2.5 LP/mm | Same as left | |
    | Cephalometric
  • noise | | Low contrast resolution on IEC
    61223-3-4: min. diameter 2.5 mm | Same as left | |
    | CBCT
  • spatial
    resolution | | min. 10% MTF at 2.0 LP/mm at
    standard mode
    min. 10% MTF at 2.5 LP/mm at
    high resolution mode | Same as left | |
    | CBCT
  • noise | | The standard deviation of the gray
    scale of the center region of the
    Contrast phantom shall be less
    than 12.5 (10% of the full scale). | Same as left | |

A comparison table of the X800N to the predicate device is shown below:

6

7

8

807.92(b)(1): Non-clinical Testing

The new X800N has been tested for compliance and developed in accordance with the following international standards:

IEC/ISO StandardFDA Recognition number
IEC 60601-1:2005+A.MD1:201219-4
IEC 60601-1-2:201419-8
IEC 60601-1-3:2008+AMD1:201312-269
IEC 60601-1-6:2010+AMD1:20135-89
IEC 60601-2-54:2009+AMD1:201512-296
IEC 60601-2-63:201712-310
IEC 61223-3-4:200012-221
IEC 62304:2006+AMD1:201513-79
IEC 62366:20155-114
ISO 10993-1:2009/Cor1:20102-220
ISO 14971:20075-40

The software changes were validated pursuant to the FDA's 2005 "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

807.92(b)(2): Clinical Testing

Clinical evaluation was performed to evaluate if the CT image obtained by the X800N for general examination and diagnosis of the ENT area is substantially equivalent to that obtained by J. Morita's reference devices, the X550 (K#073696) and MCT-1 (K#073695), which are FDAauthorized for "dental radiographic examination and diagnosis of teeth, jaw, oral structure, TMjoints and skull, including the ENT and dento-maxillofacial areas" and for taking x-ray images "of the head and neck areas, including the ENT and dento-maxillofacial areas, for use in diagnostic support," respectively.

Methods:

Experienced doctors who are using CBCT for their ENT medical practice subjectively evaluated images taken by the X800N, X550, and MCT-1.

The images they evaluated were obtained in J. Morita's facility from adult and child human skull phantoms made from resin molding on real human dry skulls. The use of phantom image for evaluation was judged to be appropriate in the majority of cases and no evaluator judged it is not suitable to use the phantom image.

The images were then ranked by the evaluators according to five categories: "Excellent Quality for Diagnosis," "Acceptable Quality for Diagnosis," Not Acceptable Quality for Diagnosis," "Not reviewed," and "Other."

9

Results:

All the evaluators ranked the images of the X800N as "Excellent Quality for Diagnosis" or "Acceptable Quality for Diagnosis." The overall results were better than that of the X550. Thus, the images of the X800N are comparable to the images from the X550. And the results of the X800N were comparable to the MCT-1.

807.92(b)(3): Conclusions from Testing

The clinical evaluation results show that the image quality of the X800N is comparable to the images quality of both the X550 and MCT-1. Thus, the images taken by X800N are sufficient to use for general examination and diagnosis of the ENT area.

As the only changes from the previous device are indication for use, change of display icon from dental figure, and the description of operation instructions for the ENT application, the X800N is substantially equivalent to the predicate device.