K Number
K201311
Device Name
PF4 IgG assay
Date Cleared
2020-06-18

(31 days)

Product Code
Regulation Number
864.7695
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PF4 IgG assay is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum or plasma. The presence of heparin-associated antibodies are commonly found in patients with Heparin Induced Thrombocytopenia (HIT).

Device Description

PF4 IgG assay is an enzyme linked immunosorbent assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum or plasma that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.

AI/ML Overview

The Immucor GTI Diagnostics, Inc. PF4 IgG assay is a qualitative screening assay for the detection of heparin-associated IgG antibodies in human serum or plasma, commonly found in patients with Heparin Induced Thrombocytopenia (HIT). The device is a Class II medical device, product code LCO.

This 510(k) submission is for modifications to a previously cleared device (K071781) and includes: a change in the Positive Serum Control raw material, the addition of plasma as a sample type, and a change in the Stopping Solution.

1. Table of acceptance criteria and reported device performance:

Feature/CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (Candidate)Comments
Modification: Positive Serum Control Raw Material Change
Process ValidationProduct meets established acceptance criteria.All acceptance criteria were met. Product consistently delivers product meeting established acceptance criteria.Validation batches of PF4 Positive Serum Control manufactured and tested.
Precision Study100% agreement within run and between run. All pre-defined Acceptance Criteria met.100% positive, demonstrating 100% agreement within run and between run. All pre-defined Acceptance Criteria met.Tested by 3 operators, 5 assays each, using a single PF4 IgG kit lot and 3 validation lots of Positive Serum Control.
Modification: Addition of Plasma as a Sample Type
Agreement between Serum and PlasmaNot explicitly stated for predicate, but implies equivalent results for both sample types.98.85% Agreement (172/174) between qualitative results of serum and plasma.Internal study.
Correlation between OD values (Plasma vs. Serum)Not explicitly stated for predicate, but implies strong correlation.R² = 99.3%. Slope = 0.9476, Intercept = -0.009914 (very good agreement).Linear regression analysis of OD values.
Equivalence of ACD and Sodium Citrate Plasma to SerumEquivalent results.All samples tested gave equivalent results to serum samples.Internal study.
Modification: Stopping Solution Change
Lot to Lot Comparison (New vs. Predicate Stopping Solution)New Stopping Solution performs equivalently, meeting established acceptance criteria.All acceptance criteria were met, demonstrating equivalence.Three lots of new Stopping Solution (ESS) compared to predicate (3M NaOH).
Process Validation (New Stopping Solution)Well-to-well variation meets acceptance criteria. Qualitative results meet expected results for QC panel.All wells met acceptance criteria for well-to-well variation. All 21 positive and 24 negative samples met expected qualitative results on all three lots of new Stopping Solution and one lot of 3M NaOH.Three lots of new Stopping Solution (ESS) tested with PF4 IgG assay.
Shelf LifeNot explicitly detailed but assumed to be acceptable.24 months. All acceptance criteria were met for stability studies.Supported by stability studies.

2. Sample size used for the test set and the data provenance:

  • Positive Serum Control Raw Material Change (Precision Study):
    • No specific "test set" sample size for patient samples is provided. The study focused on the performance of the new positive serum control material within the assay.
    • The study used a single lot of the PF4 IgG kit and three validation lots of the Positive Serum Control.
    • Data provenance is internal, generated as part of process validation.
  • Addition of Plasma as a Sample Type:
    • Test set size: 174 frozen serum and plasma samples.
    • Data provenance: Samples obtained from Florida Hospital for an internal study. Retrospective, as samples were frozen.
    • Sub-study for ACD/Sodium Citrate vs. Serum: Blood collected from 24 individuals. Origin not specified beyond "internal study." Retrospective.
  • Stopping Solution Change:
    • Test set size: No patient samples explicitly mentioned in the summary for the Stopping Solution evaluation itself. The testing involved comparing the performance of the new Stopping Solution with the predicate using various assay components and a QC panel.
    • QC Panel: 21 positive samples and 24 negative samples.
    • Data provenance: Internal, generated as part of lot-to-lot comparison and process validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not provide information on the number or qualifications of experts used to establish ground truth for the test sets. For the plasma sample type study, it states 174 frozen serum and plasma samples were used, and for the ACD/Sodium Citrate comparison, blood was collected from 24 individuals "known to be PF4: heparin antibody negative." How this "known" status was established (e.g., by experts, by a predicate device, by clinical diagnosis) is not detailed. The ground truth for the QC panel used in the stopping solution study is based on "expected qualitative result," implying pre-characterized samples, not expert consensus at the time of the study.

4. Adjudication method for the test set:

The document does not describe any adjudication method for establishing ground truth for the test sets. The studies appear to rely on the inherent reactivity of the samples (e.g., "known to be PF4: heparin antibody negative," "high level reacting anti:PF4 heparin antibody sample," "expected qualitative result" for QC panels).

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an in vitro diagnostic (IVD) assay, specifically an ELISA. It is not an AI-assisted diagnostic tool that involves human readers interpreting images or data. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device is an IVD assay, not an algorithm. Its performance is evaluated biochemically based on optical density measurements, which are then interpreted qualitatively (positive/negative) according to the assay's cut-offs. The studies described are standalone performance evaluations of the modified assay components and sample compatibility.

7. The type of ground truth used:

  • For the plasma sample type comparison, the ground truth appears to be based on the qualitative results obtained from the serum sample, as the goal was to demonstrate equivalence between serum and plasma samples. For the sub-study on ACD/Sodium Citrate, it refers to individuals "known to be PF4: heparin antibody negative" and "high level reacting anti:PF4 heparin antibody sample," indicating pre-characterized samples.
  • For the stopping solution change, the ground truth for the QC panel was based on "expected qualitative result," implying pre-defined and characterized positive and negative samples.
  • The primary ground truth for the assay itself (predicate and candidate) would typically be established against either clinical diagnosis of HIT (outcome data) or a well-established reference method, but this is not explicitly detailed for these specific modification studies. The studies focus on demonstrating that the modifications do not adversely affect the device's performance compared to its previous state.

8. The sample size for the training set:

The document does not describe a training set as this is not an algorithm requiring machine learning. The studies are performance and validation studies for changes to an IVD assay.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

June 18, 2020

Immucor GTI Diagnostics, Inc. Allison Stray Sr. Director QA/RA 20925 Crossroads Circle Waukesha, Wisconsin 53185

Re: K201311

Trade/Device Name: PF4 IgG assay Regulation Number: 21 CFR 864.7695 Regulation Name: Platelet Factor 4 Radioimmunoassay Regulatory Class: Class II Product Code: LCO Dated: May 15, 2020 Received: May 18, 2020

Dear Allison Stray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Takeesha Taylor-Bell Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name PF4 IgG

Indications for Use (Describe)

PF4 IgG assay is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum or plasma. The presence of heparin-associated antibodies are commonly found in patients with Heparin Induced Thrombocytopenia (HIT).

IVD

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUCOR." in large purple font. Above the word is a curved line that is colored with a gradient from red to blue. The word is the logo for the company Immucor, which provides transfusion and transplantation diagnostic products.

Section 5.0 510(k) Summary

1. Submitter
Owner's Name:Immucor GTI Diagnostics, Inc.
Address:20925 Crossroads Circle, Waukesha, WI 53186
Phone:(262) 754-1000
Fax:(262) 754-9831
Establishment Registration Number:2183608
Name of Contact Person:Allison Stray (astray@immucor.com)
Date Summary Prepared:May 15, 2020
2. Name of Device
Device Name:PF4/PVS Complex IgG Antibody Detection Assay
Proprietary Name:PF4 IgG
Common Name:Platelet Factor 4 Assay
Regulation Number:21CFR 864.7695
Classification:Class II
Product Code:LCO
Establishment Registration Number:2183608
    1. Name of Predicate Device for Claiming Equivalence PF4 IgG (K071781)

4. Description of Device

PF4 IgG assay is an enzyme linked immunosorbent assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum or plasma that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.

Antibodies that react with PF4 when it is complexed to heparin are found in some patients undergoing heparin therapy. The presence of these antibodies has been shown to be associated with heparin induced thrombocytopenia Type II (Type II HIT). Heparin Induced Thrombocytopenia (HIT) is an adverse reaction that can occur in patients that are undergoing heparin therapy. HIT occurs in up to 3% of patients that are receiving heparin and can result from treatment with either low molecular weight heparin (LMWH) or unfractionated heparin (UFH). HIT is usually associated with thrombocytopenia and in some cases the more severe complications of arterial or venous thrombosis. The pathophysiology of HIT is linked to the formation of antibodies which are specific for epitopes formed when heparin binds to PF4. These antibodies can subsequently bind to platelets via the FcyRIIa receptor resulting in platelet activation.

It is now known that antibodies which bind to PF4/heparin complexes also bind to PF4 when it is complexed with other polyanionic compounds such as polyvinyl sulfonate (PVS). In the PF4 IgG assay, a complex of PF4/PVS, which has been immobilized in the microwells serve as a target for the binding of antibodies associated with Type II HIT.

Test Procedure

Patient sample is added to microwells coated with platelet factor 4 (PF4) complexed to polyvinyl sulfonate (PVS). If an antibody recognizing a site on PF4:PVS is present, binding will occur. Unbound antibodies are then washed away. An alkaline phosphatase labeled anti-human globulin reagent (Anti-IgG) is added to the wells and incubated. The unbound Anti-IgG is washed away and the substrate PNPP (p-nitrophenyl phosphate) is added. After a 30-minute incubation period, the reaction is stopped with Stopping Solution. The optical density of the color that develops is measured in a spectrophotometer.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Immucor. The logo consists of the word "IMMUCOR" in purple, with a multicolored arc above it. The arc is red on the left, transitioning to orange, yellow, green, and blue on the right. The logo is simple and modern.

Engineering drawings or schematics are not applicable to the PF4 IgG Assay. The device consists of reagents and drawings are not pertinent to describe the device.

5. Intended Use

PF4 IgG is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum or plasma.

6. Substantial Equivalence

This Special 510(k) is for a Positive Serum Control raw material change and other changes the prior company did not file. The Positive Serum Control raw material modification in this submission is from human serum containing Bovine Albumin and 0.1% NaN3 to human serum containing Bovine Albumin and 0.1% NaN3 plus a chimeric (mouse Fab/Human Fc) monoclonal antibody in the PF4 IgG positive control reagent.

The modification of the PF4 IgG and need for this submission is due to supply of the predicate PF4 IgG positive control material. The current supply of the PF4 IgG positive control material is depleting and will not be available in the future.

Two (2) additional changes, the prior company, did not file are being included. Those changes include: 1) Addition of plasma as a sample type, 2) Stopping Solution changed from 3M NaOH to a buffered EDTA solution. The prior company did document a 510(k) Letter to File for both the sample type change and Stopping Solution change. Their rationale for not filing with the Food and Drug Administration included the changes did not have any significant impact to the safety or effectiveness of the device.

There were no design or functionality changes. All elements of the Design History File (DHF) for the proposed device were used as possible and design outputs were created for the proposed device.

The Similarities between the Predicate and Candidate Device:

  • The intended use has not changed ●
  • The indications for use have not changed .
  • The technology (ELISA) and assay steps have not changed ●
  • . Reportable results have not changed
  • . The packaging configuration has not changed
  • . There are no design or functionality changes between the predicate and candidate device
  • All reagents are identical with the exception of the Positive Serum Control and the Stopping ● Solution

The Differences between the Predicate and Candidate Device:

  • The predicate device only allowed for Serum as the specimen collection type. The candidate ● device allows for without anticoagulant (Serum) and ACD or Sodium Citrate (Plasma) as the specimen collection type.
  • The formulation of the Stopping Solution for PF4 IgG is different than the predicate device. . The PF4 IgG Stopping Solution is a buffered EDTA solution which is ready for use, whereas the predicate device utilizes 3M NaOH.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUCOR" in large, bold, purple letters. Above the word is a curved line that transitions from purple to red to orange to yellow to green to blue. The word "IMMUCOR" is the main focus of the image.

  • . The candidate device contains Positive Serum Control as Human serum containing Bovine Albumin and 0.1% NaN3. The predicate device contains Positive Serum Control as Human serum containing Bovine Albumin and 0.1% NaNy plus chimeric (mouse Fab/Human Fc) monoclonal antibody.
  • The candidate device has changes to the IFU as a result of the specimen collection type change and Stopping Solution type change. ●
The table below provides the comparison between the current version of PF4 IgG and the proposed device.
#Features /CharacteristicsPredicate DeviceCandidate DeviceComments
1Intended UsePF4 IgG is a qualitative screening assay for thedetection of heparin associated IgG antibodies inhuman serum.PF4 IgG is a qualitative screening assay forthe detection of heparin associated IgGantibodies in human serum or plasma.The intended use ofthe qualitative assay todetect IgG antibodieshas not changed.The revised verbiageadds plasma as aspecimen type.
2Indications for UsePF4 IgG is a qualitative screening assay for thedetection of heparin associated IgG antibodies inhuman serum. The presence of heparin-associatedantibodies are commonly found in patients withHeparin Induced Thrombocytopenia (HIT).PF4 IgG is a qualitative screening assay forthe detection of heparin associated IgGantibodies in human serum or plasma. Thepresence of heparin-associated antibodies arecommonly found in patients with HeparinInduced Thrombocytopenia (HIT).N/A
3TechnologyELISA with a colorimetric measurement systemSameN/A
4Specimen CollectionSerumWithout anticoagulant (Serum) and ACD orSodium Citrate (Plasma)Prior company madethis change in 2010
5Reportable ResultsQualitative assay; results are reported as positive ornegativeSameN/A
6PackagingConfiguration13 and 45 Test kitsSameN/A
#Features /CharacteristicsPredicate DeviceCandidate DeviceComments
7Reagents
Microwell StripsImmobilized PF4/PVS ComplexSameN/A
Concentrated WashSolution10X Tris Buffer, NaCl, Tween 20, 1% NaN3SameN/A
Specimen DiluentPhosphate Buffered Saline, 0.05% NaN3SameN/A
Substrate BufferDiethanolamine and magnesium chloride, 0.02% NaN3SameN/A
SubstratePNPP (crystalline powder)SameN/A
Stopping Solution3 M NaOHBuffered EDTA SolutionPrior company made this change in 2010
Positive SerumControlHuman serum containing Bovine Albumin and 0.1% NaN3Human serum containing Bovine Albumin and 0.1% NaN3 plusChimeric (mouse Fab/Human Fc) monoclonal antibodyN/A
Negative SerumControlHuman serum containing 0.1% NaN3SameN/A
ConjugateGoat anti-human IgG conjugated to alkaline phosphatase enzymeSameN/A
8Instructions for Use(IFU)PF4 IgG Direction InsertNo changes to the IFU for the Positive Control Serum raw material modification.IFU changes for the addition of plasma, new Stopping Solution and minor changes are included in Section 13.0, Proposed Labeling.N/A

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Immucor. The logo is purple and has a rainbow-colored arc above it. The text is in all caps and has a period after the R.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for IMMUCOR. The logo consists of the word "IMMUCOR" in purple, with a rainbow-colored arc above it. The arc starts with red on the left and transitions to blue on the right. The logo is simple and modern, and the colors are eye-catching.

7. Company Name Change

Genetic Testing Institute Inc. was purchased by Gen-Probe Inc. in December 2010 and the company name was changed to Gen-Probe GTI Diagnostics, Inc. Immucor purchased Gen-Probe GTI Diagnostics Inc. in March 2013. The company name was subsequently changed to Immucor GTI Diagnostics, Inc.

8. Non-Clinical Studies Supporting Safety and Effectiveness of the Candidate Device

The details of each of the following studies are covered in Sections 14.0 (Sterilization and Shelf Life) and 18.0 (Performance Testing Bench). Only a brief summary of these studies is provided in this section. The manufacturing facility is in conformance with the design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.

8.1 Modification, Positive Serum Control Raw Material Change

Process Validation

Three validation batches of PF4 Positive Serum Control were manufactured for this process validation.

The three validation batches were tested per the PF4 IgG Instructions For Use (IFU-303290.IFUEN, Rev. E) and all acceptance criteria were met.

Based upon the data generated, the PF4 Positive Serum Control manufacturing process consistently delivers product meeting established acceptance criteria and is acceptable for use in the production of the PF4 IgG assay.

Precision Study

Testing was performed by 3 operators and five assays were conducted by each operator.

Testing was completed using a single lot of the PF4 IgG kit. Each of the three validation lots of the Positive Serum Control were tested concurrently on each of the 5 assays (days).

All results met the IFU acceptance criteria and were 100% positive, demonstrating 100% agreement within run and between run. All pre-defined Acceptance Criteria were met.

8.2 Modification, Addition of Plasma as a Sample Type

Comparison of Serum and Plasma (ACD or Sodium Citrate) Sample Types

Samples were obtained from Florida Hospital to perform an internal study. 174 frozen serum and plasma samples were used for the study and testing was performed according to the PF4 IgG Instructions For Use.

98.85% Agreement (172/174) was met between the qualitative results obtained with serum and plasma. A linear regression was performed comparing the OD values obtained from serum and ACD or Sodium Citrate (plasma) samples for PF4 IgG. The regression analysis of plasma versus serum for the PF4 IgG assay demonstrated very good correlation between OD values (R2 = 99.3%). In addition, there was very good agreement between the OD values (slope =0.9476 and intercept = -0.009914).

Equivalent results using serum or plasma (ACD or Sodium Citrate) were obtained in the PF4 IgG assay. Both sample types are acceptable for use in the PF4 IgG assay.

Comparison of ACD and Sodium Citrate (plasma) to Serum

The objective of the internal study was to demonstrate plasma samples collected in ACD or Sodium Citrate give equivalent results to serum.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUCOR" in large, purple, sans-serif font. Above the word is a curved line that transitions from red to orange to yellow to green to blue to purple. The word "IMMUCOR" is a company name.

Blood was collected from 24 individuals known to be PF4: heparin antibody negative in ACD (yellow top), Serum (red top) or Sodium Citrate (blue top) tubes. The serum or plasma were separated and equal volume aliquots of each were prepared.

A high level reacting anti:PF4 heparin antibody sample was used for the spiking study. A dilution (1:100) intended to give a medium positive result and a dilution (1:1000) intended to give a low positive result were chosen. The samples were also run without spiking to ensure each sample showed negative reactivity.

The results for all samples tested gave equivalent results to the serum samples. The use of ACD or Sodium Citrate (plasma) samples in the PF4 IgG assay gives equivalent results to those of serum samples.

8.3 Modification, Stopping Solution Change

Lot to Lot Comparison

A lot to lot comparison study was performed to demonstrate the new Stopping Solution (ESS) performs equivalent to the predicate Stopping Solution (3 M NaOH) meeting established acceptance criteria.

Three lots of ESS (referred to as the test solution) were manufactured according to SOP-292 and tested per QAP-017A. The predicate Stopping Solution, 3 M NaOH, was used as a reference in the study.

All acceptance criteria were met demonstrating the new Stopping Solution (ESS) is acceptable to use in the PF4 IgG assay.

Process Validation

Three lots of the new Stopping Solution (ESS) were manufactured according to SOP-292 and tested with the PF4 IgG assay according to the IFU.

Well-to-well variation and expected qualitative results were evaluated to determine if the new Stopping Solution (ESS) was equivalent to the predicate 3 M NaOH Stopping Solution. The predicate Stopping Solution, 3 M NaOH, was used as the reference in the study.

For Well-to-well variation, all wells met acceptance criteria. For qualitative results, the finished product QC panel was tested with all four lots of Stopping Solution. All 21 positive samples and 24 negative samples met their expected qualitative result on all three lots of the new Stopping Solution and the one lot of 3M NaOH Stopping Solution. All assays met the QC requirements in the IFU.

The data demonstrate the new Stopping Solution (ESS) can be used in the PF4 IgG assay.

8.4 Shelf Life

The PF4 IgG assay has a shelf life of 24 months. See Section 14.0, Sterilization and Shelf Life for stability studies performed to support this premarket submission. All acceptance criteria were met.

Conclusion

Based on the non-clinical performance and stability studies, the data demonstrate the modifications to the PF4 IgG device does not present new issues of safety and effectiveness.

§ 864.7695 Platelet factor 4 radioimmunoassay.

(a)
Identification. A platelet factor 4 radioimmunoassay is a device used to measure the level of platelet factor 4, a protein released during platelet activation by radioimmunoassay. This device measures platelet activiation, which may indicate a coagulation disorder, such as myocardial infarction or coronary artery disease.(b)
Classification. Class II (performance standards).