(26 days)
The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
The BOSS Balloon Guide Catheter system is a sterile, single-use intravascular catheter. The BOSS Balloon Guide Catheter is a 90 cm long, variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and a radiopaque marker band at the tip distal to the balloon. The catheter shaft has an annular inflation lumen and a coaxial central lumen with stainless steel coil and is braid reinforced. The hub central port leads to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port, connecting to the annular inflation lumen, and is used to facilitate inflating and deflating the balloon with a syringe. The BOSS Balloon Guide Catheter uses a distal hydrophilic coating to provide lubricity and reduce friction between the catheter shaft and the vessel wall. The BOSS Balloon Guide Catheter is packaged with the 6F dilator to facilitate the option of direct access into the blood vessel without the use of an introducer sheath.
This document is a 510(k) Summary for the BOSS™ Balloon Guide Catheter, a medical device. It describes the non-clinical performance testing conducted to demonstrate substantial equivalence to legally marketed predicate devices, which means no clinical studies were required. Therefore, the information typically associated with AI/ML device studies (e.g., training set, test set, expert adjudication, MRMC studies, standalone performance) is not applicable here as this is a medical device, not an AI/ML diagnostic tool.
The "acceptance criteria" for this device are the pre-determined thresholds for various physical, functional, and biocompatibility tests, as detailed in the "NONCLINICAL PERFORMANCE TESTING SUMMARY" and "Biocompatibility Testing Summary" sections. The "study that proves the device meets the acceptance criteria" refers to the successful completion of all these non-clinical (bench and animal) tests.
Here's a breakdown of the requested information based on the provided text, reinterpreting "acceptance criteria" and "study" in the context of a medical device 510(k) submission:
1. Table of Acceptance Criteria and Reported Device Performance
| Study Name | Description | Reference Standard | Acceptance Criteria (Implied / Stated) | Reported Device Performance |
|---|---|---|---|---|
| Visual Inspection and Dimensional Verification | To demonstrate that the product meets the dimensional specifications | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting pre-determined specifications | Pass: All samples met the pre-determined acceptance criteria |
| Surface Inspection | To demonstrate the product satisfies the visual surface requirements | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting pre-determined requirements | Pass: All samples met the pre-determined acceptance criteria |
| Torque Testing | To demonstrate that the product is capable of 360 degrees of rotation about the central lumen axis without failure. | FDA guidance PTCA Catheters:2010 §VIII.A.10 | 360 degrees rotation without failure, meeting pre-determined criteria | Pass: All samples met the pre-determined acceptance criteria |
| Tensile | To demonstrate the product satisfies the tensile strength requirements for bonds and tip pull test | ISO 10555-1:2013, FDA guidance PTCA Catheters:2010 §VIII.A.7, 8 | Meeting tensile strength requirements | Pass: All samples met the pre-determined acceptance criteria |
| Kink Resistance | To demonstrate that the product has acceptable kink resistance | FDA Guidance PTCA:2010 §VIII.A.9 Kink Test | Acceptable kink resistance, meeting pre-determined criteria | Pass: All samples met the pre-determined acceptance criteria |
| Catheter Lubricity | Pad friction test to compare coated to uncoated samples | Characterization only | Characterization: Implied reduction in friction desired | Results show a 97% reduction in friction compared to uncoated samples |
| Particulates, Coating Integrity | This study was conducted to determine the quantity and size of particles generated during simulated use | AAMI TIR42:10 Evaluation of particulates associated with vascular medical devices, USP <788> Particulate Matter in Injections | Meeting pre-determined acceptance criteria for particulate quantity and size | Pass: All samples met the pre-determined acceptance criteria |
| Radiopacity | To determine the radiopaque characteristics of the device. | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting pre-determined acceptable radiopaque characteristics | Pass: All samples met the pre-determined acceptance criteria |
| Push/Track, Device Compatibility | To demonstrate that the device is deliverable through tortuous path model without kinking or buckling, is able inflate, deflate balloon at target, is compatible with treatment devices and removed without damage. | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Deliverability, proper inflation/deflation, compatibility, removal without damage, meeting pre-determined criteria | Pass: All samples met the pre-determined acceptance criteria |
| System Leak - Liquid Leak | To demonstrate that the product meets the liquid leakage under pressure requirements | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting liquid leakage requirements | Pass: All samples met the pre-determined acceptance criteria |
| System Leak - Aspiration | To demonstrate that the product meets the hub aspiration air leakage requirements | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting hub aspiration air leakage requirements | Pass: All samples met the pre-determined acceptance criteria |
| Tip Stiffness | To demonstrate that the stiffness of the distal end of the product is similar to other marketed devices. | N/A (Comparative, not a standard test) | Stiffness similar to other marketed devices, meeting pre-determined criteria | Pass: All samples met the pre-determined acceptance criteria |
| Corrosion Resistance | To demonstrate the product satisfies the corrosion resistance requirements | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | Meeting corrosion resistance requirements | Pass: All samples met the pre-determined acceptance criteria |
| Catheter Burst Pressure Under Static Conditions | To demonstrate the catheter does not leak or rupture up to 300 psi of internal pressure. | ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements | No leak or rupture at up to 300 psi | Pass: All samples met the pre-determined acceptance criteria |
| Balloon Inflation / Deflation | To demonstrate balloon meets the inflation and deflation time specifications | ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation Catheters | Meeting inflation and deflation time specifications | Pass: All samples met the pre-determined acceptance criteria |
| Balloon Fatigue | To demonstrate that there is no degradation of the balloon after 10 inflation cycles. | ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation Catheters | No degradation after 10 inflation cycles | Pass: All samples met the pre-determined acceptance criteria |
| Balloon Diameter / Volume | To demonstrate that the product meets the inflation volume vs balloon diameter specifications | ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation Catheters | Meeting inflation volume vs. balloon diameter specifications | Pass: All samples met the pre-determined acceptance criteria |
| Balloon Rated Burst Volume | To demonstrate that the balloon is capable of withstanding an injection volume of 1.5 ml. | ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation Catheters | Withstanding 1.5 ml injection volume | Pass: All samples met the pre-determined acceptance criteria |
| Small Bore Connector Compliance with Standard | To demonstrate that the product meets the requirements for small bore connectors | ISO 80369-7 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7, Connectors for intravascular or hypodermic applications. | Meeting requirements for small bore connectors | Pass: All samples met the pre-determined acceptance criteria |
| Usability Study / Simulated Use | Evaluation of device performance to meet the user needs under simulated use conditions with accessories and treatment devices | FDA Guidance Human Factors and Usability Engineering 2016, IEC 62366:2015 | Meeting user needs under simulated use conditions | Pass: All samples met the pre-determined acceptance criteria |
| Conditioning, Distribution, and Shelf life Aging | To demonstrate the device met all specifications at both baseline (T=0) and following accelerated aging to a 12-month shelf life equivalent (T=12) | Evaluation of device following accelerated aging to a 12-month shelf life equivalent; ISTA procedure 3A (2018), ASTM D4169-16, ASTM F1980-16, ASTM F88-15, ASTM F2096-11 (for packaging) | Meeting all specifications at T=0 and T=12-month AA equivalent | Pass: All samples met the pre-determined acceptance criteria |
| Visual Inspection Packaging and Labeling | To demonstrate that the product meets the packaging and labeling visual and adherence requirements | N/A | Meeting visual and adherence requirements for packaging and labeling | Pass: All samples met the pre-determined acceptance criteria |
| Biocompatibility Testing | ||||
| Cytotoxicity | To demonstrate non-cytotoxicity | ISO 10993-5:2009 | Non-cytotoxic | Pass: Noncytotoxic according to the predetermined acceptance criteria |
| Intracutaneous Irritation | To demonstrate acceptable intracutaneous reactivity | ISO 10993-10:2010 | Meeting test requirements for intracutaneous reactivity | Pass: Test requirements for intracutaneous reactivity were met according to the predetermined acceptance criteria |
| Sensitization | To demonstrate no sensitization response | ISO 10993-10:2010, Kligman Maximization Test | No sensitization response | Pass: Did not elicit a sensitization response according to the predetermined acceptance criteria |
| Systemic Toxicity | To demonstrate acceptable systemic toxicity | ISO 10993-11:2017 | Meeting test requirements for systemic toxicity | Pass: Test requirements for systemic toxicity were met according to the predetermined acceptance criteria |
| Material Mediated Pyrogenicity | To demonstrate non-pyrogenicity | ISO 10993-11:2017, USP 40 <151> Pyrogen Test | Non-pyrogenic | Pass: Nonpyrogenic, met the predetermined acceptance criteria |
| Hemolysis | To demonstrate non-hemolytic properties | ASTM F756-17, ISO 10993-4 | Non-hemolytic | Pass: Non-hemolytic, met the predetermined acceptance criteria |
| Complement Activation | To demonstrate similar complement activation characteristics as control device | ISO 10993-4:2017, SC5b-9 Complement Activation | Similar complement activation characteristics to control device | Pass: Demonstrates similar complement activation characteristics as the control device, met the predetermined acceptance criteria |
| In vivo Thromboresistance | To demonstrate similar thromboresistance characteristics as control device | ISO 10993-4:2017 | Similar thromboresistance characteristics to control device | Pass: Demonstrates similar thromboresistance characteristics as the control device, met the predetermined acceptance criteria |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size: The document repeatedly states "All samples met the pre-determined acceptance criteria" or similar for the non-clinical bench tests. While the exact numerical sample size for each test is not explicitly listed, it implies that a sufficient number of devices were tested to meet the requirements of the referenced standards (e.g., ISO 10555-1, FDA guidances). For the animal study, the sample size is not specified but it states "an acute porcine model."
- Data Provenance:
- Country of Origin: Not specified for the bench tests. The animal study used an "acute porcine model."
- Retrospective or Prospective: All testing was prospective, designed specifically to evaluate the device against established standards and criteria for regulatory submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not applicable. The "ground truth" for this medical device 510(k) submission is established by the performance standards (e.g., ISO, ASTM, FDA guidances) and the physical properties of the device itself (e.g., dimensions, strength, material characteristics). Experts were involved in setting these standards and conducting the tests, but not in the "ground truth" establishment in the sense of labeling data for an AI/ML model for diagnostic purposes.
4. Adjudication Method for the Test Set
Not applicable. This is not a study requiring human adjudication of results in the context of an AI/ML model. The "adjudication" is inherent in whether the device "passes" or "fails" the pre-defined engineering and performance criteria.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (catheter) and the submission is based on non-clinical performance and substantial equivalence to predicate devices, not an AI/ML diagnostic tool requiring MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" in this context is based on:
- Established international and national standards for medical devices (e.g., ISO 10555 series, ISO 10993 series, AAMI, ASTM).
- FDA guidance documents for specific device types (e.g., PTCA Catheters).
- Pre-determined engineering specifications and performance criteria for the device.
- Results from a non-GLP (Good Laboratory Practice) animal study.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. There is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 2, 2020
Regulatory Technology Services LLC Prithul Bom Responsible Third Party Official 1000 Westgate Drive, Suite #510k Saint Paul, Minnesota 55114
Re: K200910
Device Name: BOSS™ Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OJP, DOY Dated: April 4, 2020 Received: April 6, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200910
Device Name BOSS Balloon Guide Catheter
Indications for Use (Describe)
The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
BOSS™ Balloon Guide Catheter 510(k) Summary
SUBMITTER [807.92(a)(1)]
| Submitter's Name: | Marblehead Medical, LLC |
|---|---|
| Address: | 11600 96th Ave N Maple Grove, MN 55369 |
| Contact Person: | Kristin Mortenson |
| Telephone: | (612) 202-1142 |
| Mortenson.Kristin@MarbleheadMedical.com | |
| Date Prepared: | April 24, 2020 |
| DEVICE [807.92(a)(2)] | |
| Device Trade Name: | BOSS™ Balloon Guide Catheter |
| Common Name: | Balloon Guide Catheter |
| Classification Name: | Percutaneous Catheter |
| Product Code: | QJP, DQY |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 870.1250 |
| PREDICATE [807.92(a)(3)] | |
| Primary Predicate Device: | 8F FlowGate2™ Balloon Guide Catheter, K153729 |
| Secondary Predicate Device: | Neuron™ MAX System, K111380 |
DEVICE DESCRIPTION [807.92(a)(4)]
The BOSS Balloon Guide Catheter system is a sterile, single-use intravascular catheter. The BOSS Balloon Guide Catheter is a 90 cm long, variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and a radiopaque marker band at the tip distal to the balloon. The catheter shaft has an annular inflation lumen and a coaxial central lumen with stainless steel coil and is braid reinforced. The hub central port leads to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port, connecting to the annular inflation lumen, and is used to facilitate inflating and deflating the balloon with a syringe. The BOSS Balloon Guide Catheter uses a distal hydrophilic coating to provide lubricity and reduce friction between the catheter shaft and the vessel wall. The BOSS Balloon Guide Catheter is packaged with the 6F dilator to facilitate the option of direct access into the blood vessel without the use of an introducer sheath.
INDICATION FOR USE [807.92(a)(5)]
The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular
{4}------------------------------------------------
systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
PREDICATE DEVICE COMPARISON [807.92(a)(6)]
The following table provides a comparison of the key characteristics of the BOSS Balloon Guide Catheter System to the predicate devices.
| Primary Predicate device | Secondary Predicate device | Subject Device | |
|---|---|---|---|
| Feature | FlowGate2TM Balloon Guide Catheter (K153720) | NeuronTM MAX System (K111380) | BOSS Balloon Guide Catheter System |
| FDA Classification | Class II | Class II | Same |
| Product Code(s) | DQY | DQY | QJP, DQY |
| Regulation Number | 870.1250 | 870.1250 | Same |
| Regulation Name | Percutaneous Catheter | Percutaneous Catheter | Same |
| Indications for Use | FlowGateTM Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices. | The NeuronTM MAX System is indicated for the introduction of interventional devices into the peripheral, coronary and neuro vasculature. | The primary predicate and subject device have the same indications for use, except the primary predicate includes "and other angiographic" and the subject device does not include angiographic procedures. The primary predicate and subject device do not include coronary vasculature like the secondary predicate. The secondary predicate is included to support the direct vascular access and the hydrophilic coating. The secondary predicate does not include a balloon, therefore does not include temporary vascular occlusion. In addition, the secondary predicate does not indicate use as conduit for retrieval devices, a more specific use within the intended use. |
| Primary Predicate device | Secondary Predicate device | Subject Device | |
| Feature | FlowGate2TM Balloon Guide Catheter (K153720) | NeuronTM MAX System (K111380) | BOSS Balloon Guide Catheter System |
| These differences in wording between the subject device and the predicates are not critical to the intended therapeutic or surgical use of the device, the introduction and guidance of intravascular devices into selected vasculature. The differences in wording do not affect the safety and effectiveness of the device when the device is used as labeled. | |||
| Anatomical Locations | Peripheral and neuro vasculature | Peripheral, coronary and neuro vascular systems | Same as primary predicate |
| Principle of Operation | The Balloon Guide Catheter is used to facilitate the selective placement of interventional devices. After needle access and insertion of up to a 0.038" guidewire into the femoral artery, the device is inserted into the artery through an introducer sheath over the guidewire. The lumen of device is used for insertion and guidance of an intravascular catheter and / or retrieval devices into a selected blood vessel for treatment. | Same as primary predicate except includes the direct access use option without an introducer sheath | Same as secondary predicate. Subject device includes a direct access use option whereas the primary predicate does not. |
| Material | Commonly used medical grade plastics and stainless steel | Not listed | Same as primary predicate |
| Reinforced Catheter Shaft Reinforcement | Stainless Steel coil and braid reinforced | Stainless steel braid reinforced | Same as primary predicate |
| Marker Band / | 0.02-in (0.5 mm) wide | Radiopaque marker is | 0.08-in (2 mm) wide |
| Primary Predicate device | Secondary Predicatedevice | Subject Device | |
| Feature | FlowGate2™ BalloonGuide Catheter(K153720) | Neuron™ MAX System(K111380) | BOSS Balloon GuideCatheter System |
| Location | 0.03-in (0.75 mm) from thedistal tip edge | included on the distalend, width andlocation not listed | 0.02-in (0.5mm) from thedistal tip edge distal tothe balloon. Similar,difference does not raisenew questions regardingsafety and efficacy, alldevices utilizeradiopaque markerbands |
| CompliantBalloon | Yes, silicone | N/A | Yes, polyurethane.Similar to primarypredicate, materialdifferences do not raisenew questions of safetyand efficacy, bothmaterials are used forintravascular use. |
| Labeled ShaftOuterDiameter | 0.108-in (max)(2.7mm) 8Fr | 0.110-in (max)(2.8mm) 8F | 0.123-in (max)(3.1mm) 9.4FrSimilar, differences donot raise new questionsof safety and efficacy. |
| Labeled ShaftInnerDimension | 0.084-in (2.21mm) 6.4Fr | 0.088-in (2.24 mm)6.7F | Same as secondarypredicate |
| Effectivelength | 85 and 95 cm | 80 and 90 cm | 90 cm, same assecondary predicate |
| Tip Shape | Straight | Straight | Same |
| MaximumBalloonVolume | 0.6ml | N/A | Same as primarypredicate |
| Radiopaque | Distal Tip has radiopaquemarker bands, stainlesssteel reinforcement in thecatheter shaft renders theshaft visible onfluoroscopy | Distal Tip hasradiopaque zone | Same as primarypredicate |
| Coating | N/A | Hydrophilic Coating –distal shaft | Same as the secondarypredicate. Coating is on |
| Feature | Primary Predicate device | Secondary Predicate device | Subject Device |
| FlowGate2TM BalloonGuide Catheter(K153720) | NeuronTM MAX System(K111380) | BOSS Balloon GuideCatheter System | |
| InternalConstruction | Coaxial Lumen | Single lumen | the distal portionproximal to the balloonSame as primarypredicate |
| Supplied items/ accessories | Dilator, RotatingHemostasis Valve, TuohyBorst Valve with Sideport,Peel Away Sheaths, Luer-Activated Valves | Dilator, RotatingHemostasis Valve(RHV), Extensiontubing | Dilator is provided, sameas both predicates. Theadditional accessoriesare commerciallyavailable devices.Differences do not raisenew questions of safetyand efficacy. |
| How Supplied | Sterile, single use | Sterile, single use | Same |
| SterilizationMethod | EtO | EtO | Same |
| SterilityAssuranceLevel | 10-6 | 10-6 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
The differences between the subject device and predicates described in the comparison table above are not critical to the intended therapeutic or surgical use of the device, do not raise questions of safety and effectiveness, and as shown through testing and analysis, do not affect the safety and effectiveness of the device when used as labeled.
NONCLINICAL PERFORMANCE TESTING SUMMARY [807.92(b)]
Determination of substantial equivalence is based on an assessment of non-clinical performance bench test data. The BOSS Balloon Guide Catheter System has successfully completed the following relevant performance testing to demonstrate substantial equivalence. Testing was performed to evaluate physical integrity, functionality, and performance of the BOSS Balloon Guide Catheter. A summary of the tests performed is provided in the table below:
{8}------------------------------------------------
| Performance Bench Testing Summary | |||
|---|---|---|---|
| Study Name | Description | Reference Standard | Results |
| VisualInspection andDimensionalVerification | To demonstrate that theproduct meets thedimensionalspecifications | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| SurfaceInspection | To demonstrate theproduct satisfies thevisual surfacerequirements | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| Torque Testing | To demonstrate that theproduct is capable of360 degrees of rotationabout the central lumenaxis without failure. | FDA guidance PTCACatheters:2010 §VIII.A.10 | PassAll samples metthe pre-determinedacceptance criteria |
| Tensile | To demonstrate theproduct satisfies thetensile strengthrequirements for bondsand tip pull test | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements, FDAguidance PTCACatheters:2010 §VIII.A.7,8 | PassAll samples met thepre-determinedacceptance criteria |
| Kink Resistance | To demonstrate that theproduct has acceptablekink resistance | FDA Guidance PTCA:2010§VIII.A.9 Kink Test | PassAll samples met thepre-determinedacceptance criteria |
| CatheterLubricity | Pad friction test tocompare coated touncoated samples | Characterization only | Results show a 97%reduction in frictioncompared touncoated samples |
| Particulates,CoatingIntegrity | This study wasconducted to determinethe quantity and size ofparticles generatedduring simulated use | AAMI TIR42:10 Evaluationof particulates associatedwith vascular medicaldevices, USP <788>Particulate Matter inInjections | PassAll samples met thepre-determinedacceptance criteria |
| Radiopacity | To determine theradiopaquecharacteristics of thedevice. | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| Performance Bench Testing Summary | |||
| Study Name | Description | Reference Standard | Results |
| Push/Track,DeviceCompatibility | To demonstrate that thedevice is deliverablethrough tortuous pathmodel without kinkingor buckling, is ableinflate, deflate balloonat target, is compatiblewith treatment devicesand removed withoutdamage. | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| System Leak -Liquid Leak | To demonstrate that theproduct meets the liquidleakage under pressurerequirements | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| System Leak -Aspiration | To demonstrate that theproduct meets the hubaspiration air leakagerequirements | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| Tip Stiffness | To demonstrate that thestiffness of the distalend of the product issimilar to othermarketed devices. | N/A | PassAll samples metthe pre-determinedacceptance criteria |
| CorrosionResistance | To demonstrate theproduct satisfies thecorrosion resistancerequirements | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| Catheter BurstPressure UnderStatic Conditions | To demonstrate thecatheter does not leak orrupture up to 300 psi ofinternal pressure. | ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements | PassAll samples met thepre-determinedacceptance criteria |
| Balloon Inflation /Deflation | To demonstrate balloonmeets the inflation anddeflation timespecifications | ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation Catheters | PassAll samples met thepre-determinedacceptance criteria |
| Performance Bench Testing Summary | |||
| Study Name | Description | Reference Standard | Results |
| Balloon Fatigue | To demonstrate thatthere is no degradationof the balloon after 10inflation cycles. | ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation Catheters | PassAll samples met thepre-determinedacceptance criteria |
| BalloonDiameter /Volume | To demonstrate that theproduct meets theinflation volume vsballoon diameterspecifications | ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation Catheters | PassAll samples met thepre-determinedacceptance criteria |
| Balloon RatedBurst Volume | To demonstrate that theballoon is capable ofwithstanding aninjection volume of 1.5ml. | ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation Catheters | PassAll samples met thepre-determinedacceptance criteria |
| Small BoreConnectorCompliancewith Standard | To demonstrate that theproduct meets therequirements for smallbore connectors | ISO 80369-7 2016 Small-bore connectors forliquids and gases inhealthcare applications -Part 7, Connectors forintravascular orhypodermic applications. | PassAll samples met thepre-determinedacceptance criteria |
| Usability Study /Simulated Use | Evaluation of deviceperformance to meetthe user needs undersimulated useconditions withaccessories andtreatment devices | FDA Guidance HumanFactors and UsabilityEngineering 2016, IEC62366:2015 | PassAll samples met thepre-determinedacceptance criteria |
| Conditioning,Distribution,and Shelf lifeAging | To demonstrate thedevice met allspecifications at bothbaseline (T=0) andfollowing acceleratedaging to a 12-monthshelf life equivalent(T=12) | Evaluation of devicefollowing acceleratedaging to a 12-month shelflife equivalent | PassAll samples met thepre-determinedacceptance criteria |
| Performance Bench Testing Summary | |||
| Study Name | Description | Reference Standard | Results |
| Package Testing | |||
| Conditioning,Distribution,and Shelf LifeAging | Evaluation of packagingstrength and integrity atboth baseline (T=0) andfollowing acceleratedaging to a 12-monthshelf life equivalent(T=12) | ISTA procedure 3A (2018),ASTM D4169-16, ASTMF1980-16,ASTM F88-15, ASTMF2096-11 | PassAll samples met thepre-determinedacceptance criteria |
| VisualInspectionPackaging andLabeling | To demonstrate that theproduct meets thepackaging and labelingvisual and adherencerequirements | N/A | PassAll samples met thepre-determinedacceptance criteria |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
The results of these tests provide reasonable assurance that the BOSS Balloon Guide Catheter has been designed and tested to assure conformance to the requirements for its intended use and indications for use. No new safety or performance issues were raised during the testing; therefore, this device is considered to be substantially equivalent to the predicate devices.
Biocompatibility Testing Summary
The device is categorized as Externally Communicating Device, Circulating Blood, Limited Contact (≤ 24 hours), per ISO 10993-1, the following testing was conducted:
| Test Name | Test Method | Results |
|---|---|---|
| Cytotoxicity | Tested in accordance with ISO 10993-5:2009, Biological Evaluation of MedicalDevices – Part 5: Tests for in vitro toxicity,Neutral Red Uptake Method | PassNoncytotoxic according tothepredeterminedacceptance criteria |
| IntracutaneousIrritation | Tested in accordance with ISO 10993-10:2010, Biological Evaluation of MedicalDevices – Part 10: Tests for Irritation andSkin Sensitization | PassTest requirements forintracutaneous reactivitywere met according to thepredeterminedacceptance criteria |
| Sensitization | Tested in accordance with ISO 10993-10,Biological Evaluation of Medical Devices –Part 10 Tests forIrritation and Skin Sensitization, KligmanMaximization Test | PassDid not elicit asensitizationresponse according to thepredeterminedacceptance criteria |
| Test Name | Test Method | Results |
| Systemic Toxicity | Tested in accordance with ISO 10993-11:2017, Biological Evaluation of MedicalDevices - Part 11: Tests for SystemicToxicity | PassTest requirements forsystemic toxicity weremet according to thepredeterminedacceptance criteria |
| Material MediatedPyrogenicity | Tested in accordance with ISO 10993-11:2017, BiologicalEvaluation of Medical Devices - Part 11:Tests for Systemic Toxicity and USP 40<151> Pyrogen Test | PassNonpyrogenic, met thepredeterminedacceptance criteria |
| Hemolysis | Tested in accordance with ASTM F756-17,Standard Practice for Assessment ofHemolytic Properties of Materials and ISO10993-4, Biological Evaluation of MedicalDevices - Part 4: Selection of Tests forInteractions with Blood, Tests forHemolytic Properties, Direct and IndirectMethods | PassNon-hemolytic, met thepredeterminedacceptance criteria |
| ComplementActivation | Tested in accordance with ISO 10993-4,Biological Evaluation of Medical Devices -Part 4:2017: Selection of Tests forInteractions with Blood, SC5b-9Complement Activation | PassDemonstrates similarcomplement activationcharacteristics as thecontrol device, met thepredeterminedacceptance criteria |
| In vivoThromboresistance | Tested in accordance with ISO 10993-4:2017, Biological Evaluation of MedicalDevices - Part 4: Selection of Tests forInteractions with Blood | PassDemonstrates similarthromboresistancecharacteristics as thecontrol device, met thepredeterminedacceptance criteria |
{12}------------------------------------------------
Sterilization and Shelf Life
The BOSS Balloon Guide Catheter sterilization process using 100% Ethylene Oxide (EO) has been validated in accordance with ISO 11135-1:2014 to achieve a SAL of 10 °. EO and ECH residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels (BET) were below the level of 2.15 EU/device in accordance with ANSI AAMI ST72:2011/(R)2016. Both baseline and accelerated shelf life testing were conducted (T = 0 and T = 12-month AA (accelerated aging)) demonstrating the device will perform as intended to support the proposed 12-month shelf-life.
{13}------------------------------------------------
Animal Testing
A non-GLP animal test was conducted to evaluate the in vivo performance of the device in an acute porcine model. Trackability and handling of the guide catheter, balloon inflation / deflation, temporary vascular occlusion, radiopacity, and catheter compatibility were assessed. The results for the subject device were comparable to the primary predicate device, demonstrating acceptable results.
Clinical Testing
No clinical studies were required to demonstrate substantial equivalence.
Conclusion
Marblehead Medical LLC concludes through a review of the benchtop assessments, the comparison of the device classification, intended use, operating principle, technological characteristics, sterility, and biocompatibility that the BOSS Balloon Guide Catheter System is substantially equivalent to the predicate devices.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).