K Number
K200741

Validate with FDA (Live)

Date Cleared
2020-03-31

(8 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home.

Device Description

The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute or chronic care facility. This premarket notification is an indication expansion to include home use of the device. The product described in this submission includes:

  • . Tablo Console, a single module consisting of multiple fluidic systems that perform the activities of a water purification system (WPS) and a conventional dialysis delivery system (DDS), and
  • . The following accessories/components
    • Tablo Script (optional accessory software) o
    • Tablo Straws O
    • O Outset Acid Concentrate 1K, 2K and 3K
    • Outset Bicarbonate Concentrate O
    • Patient USB Stick O
  • Hand Crank O
  • O Locking Power Cord
  • Drain Line O
  • O Water Line
  • Disinfectant Straw O
AI/ML Overview

Here's an analysis of the acceptance criteria and supporting studies for the Tablo Hemodialysis System, based on the provided FDA 510(k) summary:

The document describes the acceptance criteria and performance for several bench performance tests and mentions a clinical study to support the device's substantial equivalence and expanded indication for home use.

1. Table of Acceptance Criteria and Reported Device Performance

Test PerformedAcceptance CriteriaReported Device Performance
Bench Performance Testing
AAMI Water and AAMI Dialysate Quality VerificationThe system shall meet ANSI/AAMI 13959:2014 and ANSI/AAMI 11663:2014 requirements for water and dialysate under worst-case input test conditions.Pass
1K Concentrate Compatibility VerificationThe system shall be compatible with the Outset branded 1K acid concentrate.Pass
Drip Chamber Level Sensor VerificationThe modified sensors shall meet pre-specified functional and performance requirements.Pass
Suitability Test - USP<61> Microbial Enumeration TestThe system shall not inhibit detection and/or the recovery of potential organisms.Pass
Reprocessing Disinfection ValidationThe system shall be labeled with cleaning instructions in accordance with FDA guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," issued March 17, 2015.Pass
Clinical Study (for home use indication)
Safety (Adverse Events)(Implicitly: low adverse event rate) The Tablo Hemodialysis System is associated with a low adverse event rate and can safely be used at home.Low adverse event rate
Adequacy (Weekly standardized Kt/V)(Implicitly: consistently and effectively deliver a weekly standardized Kt/V of 2.1) The recommended adequacy target of home hemodialysis (a weekly standardized Kt/V of 2.1) can consistently and effectively be delivered.Achieved
Ultrafiltration Goals(Implicitly: reliably achieve ultrafiltration goals) Patients reliably achieve ultrafiltration goals using the System at home.Reliably achieved

2. Sample Size Used for the Test Set and Data Provenance

  • Bench Performance Tests: No specific sample sizes (e.g., number of units tested) are provided for these tests within the summary.
  • Clinical Study: No specific sample size (number of patients) is explicitly stated in the summary.
  • Data Provenance: The clinical study is described as a "prospective, multicenter, open label, non-randomized, cross-over study." This indicates the data was collected prospectively from multiple clinical sites. The country of origin is not specified but is implicitly the US, given the FDA submission context.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Bench Performance Tests: Not applicable. Ground truth for these tests is based on established engineering and regulatory standards (e.g., AAMI, USP, FDA guidance).
  • Clinical Study: Not applicable in the traditional sense of expert consensus on image interpretation. The "ground truth" for the clinical study involved direct measurement of patient outcomes (adverse events, Kt/V, ultrafiltration) which are objective physiological parameters.

4. Adjudication Method for the Test Set

  • Bench Performance Tests: Not applicable. These are objective tests against defined physical or chemical standards.
  • Clinical Study: Not explicitly stated. For clinical trial outcome measures like adverse events and adequacy, typically events are recorded as they occur and then reviewed by study investigators and/or an independent data safety monitoring board.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices where human readers interpret medical images. The Tablo Hemodialysis System is a medical device for therapy, not diagnosis.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

A "standalone" performance evaluation in the context of an AI algorithm is not applicable here. The Tablo Hemodialysis System is a physical medical device that interacts directly with patients, albeit with software control. Its performance is evaluated through its direct physical and physiological effects, not through an algorithm's interpretive ability. The software reliability improvements mentioned refer to the device's operational software, not an AI diagnostic algorithm.

7. Type of Ground Truth Used

  • Bench Performance Tests: Regulatory and engineering standards (e.g., ANSI/AAMI, USP, FDA guidance for reprocessing).
  • Clinical Study: Objective physiological measurements of patient outcomes:
    • Safety: Adverse event rates.
    • Adequacy: Weekly standardized Kt/V (a measure of dialysis efficacy).
    • Ultrafiltration: Achievement of ultrafiltration goals.

8. Sample Size for the Training Set

  • Software Verification and Validation: No specific sample size (e.g., lines of code, test cases) is provided for the "training set" of the software. Instead, the summary states that "Software verification and validation testing were conducted and passed for the incremental software changes," following FDA guidance for "major" level of concern software. This implies a comprehensive testing process, but not a "training set" in the machine learning sense.
  • Clinical Study: Not applicable; the clinical study evaluated the device's performance, it did not "train" a model.

9. How the Ground Truth for the Training Set Was Established

  • Software Verification and Validation: Not applicable in the context of a machine learning "training set." The "ground truth" for software validation would be the defined functional and non-functional requirements and specifications against which the software's behavior is tested.
  • Clinical Study: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 31, 2020

Outset Medical, Inc. Jennifer Mascioli-Tudor Vice President of Quality Assurance and Regulatory Affairs 1830 Bering Drive San Jose, CA 95112

Re: K200741

Trade/Device Name: Tablo Hemodialysis System Regulation Number: 21 CFR 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI, FIP Dated: March 20, 2020 Received: March 24, 2020

Dear Jennifer Mascioli-Tudor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200741

Device Name Tablo Hemodialysis System

Indications for Use (Describe)

The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 4 Outset Medical, Inc.

K200741

510(k) Summary

I. SUBMITTER

Name:

Outset Medical, Inc. 1830 Bering Drive San Jose, CA 95112

Phone: Primary Contact: Secondary Contact: Date Prepared:

(669) 231-8200 Jennifer Mascioli-Tudor Mara Marshak March 20, 2020

II. DEVICE

Name of Device:

Tablo® Hemodialysis System

Common or Usual Name: Hemodialysis delivery system and water purification system

Classification Name: High Permeability Hemodialysis System (21 CFR § 876.5860) Product Code: KDI Water Purification System for Hemodialysis - Product Code: FIP

Regulatory Class: II

III. PREDICATE DEVICE

Tablo Console, K160881 The predicate device has not been subject to a design related recall. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute or chronic care facility. This premarket notification is an indication expansion to include home use of the device. The product described in this submission includes:

  • . Tablo Console, a single module consisting of multiple fluidic systems that perform the activities of a water purification system (WPS) and a conventional dialysis delivery system (DDS), and
  • . The following accessories/components
    • Tablo Script (optional accessory software) o
    • Tablo Straws O
    • O Outset Acid Concentrate 1K, 2K and 3K
    • Outset Bicarbonate Concentrate O
    • Patient USB Stick O

{4}------------------------------------------------

  • Hand Crank O
  • O Locking Power Cord
  • Drain Line O
  • O Water Line
  • Disinfectant Straw O

V. INDICATIONS FOR USE

The indications for use statement was modified to include home use:

The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The modified Tablo Hemodialysis System has the same fundamental technology, principle of operation, and principle functionality as the predicate. The device is a high permeability hemodialysis system, which combines a water purification system and dialysate delivery system as one device system.

The following technological differences exist between the predicate and subject devices:

  • Added Outset 1K acid concentrate
  • Venous drip chamber level sensor change
  • 1-way data transmission from the device to the cloud for home users ●
  • . Software reliability improvements

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

No additional testing was conducted to support the modified device.

Electrical safety and electromagnetic compatibility (EMC)

Complies with ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-10, IEC 60601-1-11, and IEC 60601-2-16. RFID testing per AIM 7351731.

Software Verification and Validation Testing

Software verification and validation testing were conducted and passed for the incremental software changes. Documentation provided is per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is considered a "major" level of concern, since a failure or latent flaw in the software could

{5}------------------------------------------------

directly result in serious injury or death to the patient or operator. Completed software testing supports safety and effectiveness of the device.

Bench Performance Testing

The following test were completed to support the subject device.

Test PerformedAcceptance CriteriaResult
AAMI Water andAAMI DialysateQualityVerificationThe system shall meet ANSI/AAMI 13959:2014 andANSI/AAMI 11663:2014 requirements for water anddialysate under worst case input test conditions.Pass
1K ConcentrateCompatibilityVerificationThe system shall be compatible with the Outsetbranded 1K acid concentratePass
Drip ChamberLevel SensorVerificationThe modified sensors shall meet pre-specifiedfunctional and performance requirements.Pass
Suitability Test -USP<61>MicrobialEnumeration TestThe system shall not inhibit detection and/or therecovery of potential organisms.Pass
ReprocessingDisinfectionValidationThe system shall be labeled with cleaninginstructions in accordance with FDA guidance,Reprocessing Medical Devices in Health Care Settings:Validation Methods and Labeling, issued March 17,2015.Pass

Human Factors

Human factors validation conducted in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices", issued February 3, 2016.

Animal Study

No animal studies were conducted to support the modified device.

Clinical Studies

Outset Medical conducted the Tablo clinical study, a prospective, multicenter, open label, nonrandomized, cross-over study evaluating the use of the Tablo Hemodialysis System In-Center and In-Home by trained subjects with end stage renal disease (ESRD) who are on stable dialysis regimen. Use of the Tablo Hemodialysis System is associated with a low adverse event rate and can safely be used at home by ESRD patients. The recommended adequacy target of home hemodialysis (a weekly standardized Kt/V of 2.1) can consistently and effectively be delivered with the Tablo Hemodialysis System when used by End Stage Renal Disease patients in the home environment. Additionally, patients reliably achieve ultrafiltration goals using the System at home. The Tablo Hemodialysis System is safe and effective for home use.

{6}------------------------------------------------

VIII. CONCLUSION

The clinical and non-clinical data support that the modified Tablo Hemodialysis System, when used in the home environment, is as safe and effective and therefore, is substantially equivalent to the predicate device, K160881.

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”