(58 days)
The Sublime™ Radial Access 014 RX PTA Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) dilation of peripheral vasculature stenoses in the iliac, femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
The 014 RX PTA Dilatation Catheter is a co-axial rapid exchange (RX) catheter system with a semi compliant balloon at the distal end designed for use with a 0.014" guidewire. The guidewire will pass through a lumen from an RX access bond. The balloon has two radiopaque markers that aid in the placement of the balloon within the stenosis. The balloon and catheter shaft are coated with a hydrophilic coating. The proximal end of the catheter has a single standard luer hub connector for connection of an inflation device. The inflation device is used to inflate and deflate the balloon with a contrast medium. The clearance between the inner and outer shafts acts as the passage for the inflation medium for balloon expansion. The device is used by positioning the balloon catheter over a guidewire. The balloon is aligned under fluoroscopy in the diseased vessel at the area to be treated. The balloon is then inflated with inflation media to pressures ranging between the nominal and the rated burst pressure to dilate the occluded area. On completion the balloon is then deflated under vacuum and removed from the patient. The 014 RX PTA Dilatation Catheter is to be provided sterile (via ethylene oxide, EtO) and is intended for single use only.
The provided text describes a 510(k) premarket notification for a medical device, the Surmodics™ Sublime™ Radial Access 014 RX PTA Dilatation Catheter. This document asserts substantial equivalence to a predicate device, meaning it's claimed to be as safe and effective as a legally marketed device, and therefore does not typically involve the extensive clinical trials and efficacy studies seen with novel devices.
Here's an analysis of the acceptance criteria and supporting studies based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't present a formalized table of acceptance criteria with quantified performance targets and corresponding results. Instead, it states that "All test results met documented acceptance criteria and did not raise new questions of safety or effectiveness." for the biocompatibility tests and that performance bench testing results "demonstrate that the technological differences identified do not raise new questions of safety or effectiveness compared to the predicate device."
Here's a summary of the types of tests performed and the implicit "acceptance criteria" met:
| Acceptance Criteria Category | Reported Device Performance | Implicit Acceptance Criteria |
|---|---|---|
| Performance Bench Testing | ||
| Rated burst pressure (RBP) | Met criteria | RBP is sufficient for intended use and comparable to predicate. |
| Inflation & deflation time | Met criteria | Inflation/deflation times are within acceptable ranges for clinical use. |
| Balloon diameters at nominal pressure | Met criteria | Balloon diameters at nominal pressure are within specifications. |
| Simulated use pushability & trackability | Met criteria | Device can be successfully maneuvered and positioned in simulated vessels. |
| Coating lubricity | Met criteria | Coating reduces friction sufficiently. |
| Coating durability | Met criteria | Coating integrity is maintained during use. |
| Coating length | Met criteria | Coating covers the specified length of the device. |
| Ancillary tool compatibility (guidewire) | Met criteria | Device is compatible with specified guidewires. |
| Tip profile | Met criteria | Tip profile is within dimensions for safe navigation. |
| Multiple inflation/fatigue & leak test | Met criteria | Balloon can withstand multiple inflations and does not leak. |
| Tensile strength | Met criteria | Catheter components have sufficient tensile strength. |
| Flexibility & kink | Met criteria | Device is flexible enough for navigation and resists kinking. |
| Particulate | Met criteria | Particulate matter is within acceptable limits. |
| Torque strength | Met criteria | Device can be torqued without structural failure. |
| Biocompatibility Testing | ||
| MEM Elution Cytotoxicity | Met documented acceptance criteria | No cytotoxic effects observed. |
| Kligman Maximization Sensitization | Met documented acceptance criteria | No significant sensitization response observed. |
| Irritation by Intracutaneous Injection | Met documented acceptance criteria | No significant irritation observed. |
| Acute Systemic Toxicity by Systemic Injection | Met documented acceptance criteria | No acute systemic toxic effects observed. |
| Rabbit Pyrogen Test (Material Mediated) | Met documented acceptance criteria | No pyrogenic response observed. |
| Hemolysis ASTM Method | Met documented acceptance criteria | No significant hemolysis observed. |
| SC5b Complement Activation Assay | Met documented acceptance criteria | No significant complement activation observed. |
| Thrombogenicity Testing | Met documented acceptance criteria | No significant thrombogenic potential observed. |
| Sterilization Testing | ||
| Sterilization Product Testing | Demonstrated suitability | Sterilization cycle effectively sterilizes the device. |
| Product Bioburden | Demonstrated suitability | Bioburden level is acceptable for sterilization. |
| LAL/Endotoxin Testing | Demonstrated suitability | Endotoxin levels are below acceptable limits. |
| Residual Degas Assessment | Demonstrated suitability | EO residuals are within acceptable limits after degassing. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify the sample sizes for the performance bench tests, biocompatibility tests, or sterilization tests. It only lists the types of tests performed.
Regarding data provenance:
- The tests were conducted by the manufacturer, Creagh Medical, Ltd., dba Surmodics, Inc.
- The sterilization site is Synergy Health Ireland.
- The device is intended for the U.S. market, regulated by the FDA.
- The data would be considered retrospective in the sense that these are studies conducted on manufactured devices as part of the development and regulatory submission process, not ongoing clinical data collection.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not provided in the document. The type of device (dilatation catheter) and the studies performed (bench, biocompatibility, sterilization) do not inherently require "ground truth" to be established by experts in the same way an AI diagnostic device would. For these types of tests, ground truth is typically established by physical measurement against engineering specifications and industry standards.
4. Adjudication Method for the Test Set:
This information is not applicable and therefore not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or expert panel reviews for diagnostic devices to resolve discrepancies in interpretations of medical images or patient data. The current document describes bench, biocompatibility, and sterilization testing, which rely on standardized laboratory procedures and measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a physical medical instrument (a dilatation catheter), not an AI algorithm assisting human readers. Therefore, the concept of human readers improving with or without AI assistance is not applicable to this device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Study was done:
No, a standalone algorithm performance study was not done. This device is a physical medical instrument and does not involve any artificial intelligence algorithms.
7. The Type of Ground Truth Used:
For the performance bench testing, the "ground truth" would be the engineering specifications, industry standards (e.g., ISO), and design requirements for a PTA balloon catheter.
For biocompatibility, the ground truth is established by internationally recognized standards and guidelines (ISO 10993-1), with testing against biological responses (e.g., cytotoxicity, irritation, hemolysis).
For sterilization, the ground truth is established by sterilization standards (ISO 11135) and validated procedures to ensure sterility assurance levels.
8. The Sample Size for the Training Set:
This information is not applicable and therefore not provided. "Training set" refers to data used to train machine learning models. This device is a physical catheter, not an AI algorithm.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable for the same reason as above (not an AI device).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 14, 2020
Creagh Medical Ltd. Amy Yanta Regulatory Affairs 9924 W 74th St Eden Prairie, MN 55344
Re: K200700
Trade/Device Name: Surmodics™ Sublime™ Radial Access 014 RX PTA (Percutaneous Transluminal Angioplasty) Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: March 16, 2020 Received: March 17, 2020
Dear Ms. Yanta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200700
Device Name
Surmodics™ Sublime™ Radial Access 014 RX PTA (Percutaneous Transluminal Angioplasty) Dilatation Catheter
Indications for Use (Describe)
The Sublime™ Radial Access 014 RX PTA Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) dilation of peripheral vasculature stenoses in the iliac, femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
Image /page/3/Picture/2 description: The image shows the logo for "SURMODICS". The logo consists of a stylized graphic to the left of the company name. The graphic is a gray, folded or layered shape, possibly representing a ribbon or a stylized "S". The text "SURMODICS" is in a bold, sans-serif font, with each letter clearly legible.
Date Prepared: March 16, 2020
Submitters Name / Contact Person 510k Submitter Address
Creagh Medical, Ltd., dba Surmodics, Inc. IDA Business Park, Ballinasloe, Co. Galway, H53 K8P4 Ireland
Contact for Official/Routine Correspondence
Amy Yanta Regulatory Affairs 9924 W 74th St Eden Prairie, MN 55344 Phone: (952) 500-7562 Email: ayanta@surmodics.com
510k Submitter Establishment
Registration Number
3005994106
| General Information | |
|---|---|
| Trade Name: | SurmodicsTM SublimeTM Radial Access 014 RX PTA (PercutaneousTransluminal Angioplasty) Dilatation Catheter |
| Common / Usual Name: | 014 RX PTA Dilatation Catheter |
| Classification Name | Catheter, Angioplasty, Peripheral, Transluminal |
| Regulation/Product Code | 21 CFR 870.1250 |
| Device Panel | Cardiovascular |
| Regulatory Classification: | Class II |
| Product Code: | LIT |
| Predicate Device: | 014 Hydrophilic Coated PTA Balloon Dilatation Catheter510(k)#: K171251 |
Device Description
The 014 RX PTA Dilatation Catheter is a co-axial rapid exchange (RX) catheter system with a semi compliant balloon at the distal end designed for use with a 0.014" guidewire. The guidewire will pass through a lumen from an RX access bond. The balloon has two radiopaque markers that aid in the placement of the balloon within the stenosis. The balloon and catheter shaft are coated with a hydrophilic coating. The proximal end of the catheter has a single standard luer hub connector for connection of an inflation device. The inflation device is used to inflate and deflate the balloon with a contrast medium. The clearance between the inner and outer shafts acts as the passage for the inflation medium for balloon expansion. The device is used by positioning the balloon catheter over a guidewire. The balloon is aligned under fluoroscopy in the diseased vessel at the area to be treated. The balloon is then inflated with inflation media to pressures ranging between the nominal and the rated burst pressure to dilate the occluded area. On completion the balloon is then deflated under vacuum and removed from the patient. The 014 RX PTA Dilatation Catheter is to be provided sterile (via ethylene oxide, EtO) and is intended for single use only.
{4}------------------------------------------------
Indications for Use
The Sublime™ Radial Access 014 RX PTA Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) dilation of peripheral vasculature stenoses in the iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Comparison of Technological Characteristics
The 014 RX PTA Dilatation Catheter incorporates substantially equivalent device design, materials, packaging, manufacturing processes & sterilization method as the predicate 014 Hydrophilic Coated PTA Balloon Dilatation Catheter. The 014 RX PTA Dilatation Catheter incorporates substantially equivalent device design, materials, and hydrophilic coating as the predicate 014 Hydrophilic Coated PTA Balloon Dilatation Catheter. The 014 RX PTA Dilatation Catheter has the same indications for use and classification as the predicate device. Where substantial equivalence is not directly demonstrated from the perspective of technology and performance, design verification testing provides evidence of the substantial equivalence of the 014 RX PTA Dilatation Catheter with the 014 Hydrophilic Coated PTA Balloon Dilatation Catheter.
Substantial Equivalence and Summary of Studies
The following non-clinical testing was performed:
- Performance Bench Testing
- Biocompatibility
- Sterilization
Performance Bench Testing
Results of design verification testing demonstrate that the technological differences identified do not raise new questions of safety or effectiveness compared to the predicate device. The 014 RX PTA Dilatation Catheter has been evaluated through the following tests:
- Rated burst pressure (RBP) -
- Inflation & deflation time -
- -Balloon diameters at nominal pressure
- Simulated use pushability & trackability -
- -Coating lubricity
- Coating durability -
- -Coating length
- -Ancillary tool compatibility (guidewire)
- -Tip profile
- -Multiple inflation/fatigue & leak test
- -Tensile strength
- Flexibility & kink -
- -Particulate
- -Torque strength
Biocompatibility
Biocompatibility of the 014 RX PTA Dilatation Catheter has been evaluated in accordance with ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, Guidance for Industry and Food and Drug Administration Staff". Per the requirements of ISO 10993
{5}------------------------------------------------
1 the 014 RX PTA Dilatation Catheter is classified as an externally communicating device in contact with circulating blood for limited exposure duration. Biocompatibility tests appropriate for the device classification were selected, and testing was completed in accordance with FDA Good Laboratory Practice (GLP) regulations (21 CFR, Part 58). The following biocompatibility tests were performed:
- -MEM Elution Cytotoxicity Testing
- Kligman Maximization Sensitization Test -
- -Irritation by Intracutaneous Injection
- Acute Systemic Toxicity by Systemic Injection -
- Rabbit Pyrogen Test (Material Mediated) -
- -Hemolysis ASTM Method
- -SC5b Complement Activation Assay
- -Thrombogenicity Testing
All test results met documented acceptance criteria and did not raise new questions of safety or effectiveness.
Sterilization
The Ethylene Oxide (EtO) sterilization cycle used for the predicate 014 Hydrophilic Coated PTA Balloon Dilatation Catheter device will be adapted to include the 014 RX PTA Dilatation Catheter. To confirm the suitability of the sterilization cycle for the device the following sterilization product testing has been completed:
- Sterilization Product Testing at the sub-lethal cycle and full cycle of the validated sterilization cycle.
- Product Bioburden (Bioburden Validation)
- LAL/Endotoxin Testing (LAL Validation)
- Residual Degas Assessment ●
The results of the sterilization product testing have demonstrated that the Ethylene Oxide (EtO) sterilization method for the 014 RX PTA Dilatation Catheter meets the requirements of ISO 11135, and that the sterility of the device will be maintained.
Creagh Medical utilized the following sterilization site:
Synergy Health Ireland, Sragh Industrial Estate, Tullamore. Co. Offaly, Ireland.
Clinical Data
No clinical data is being submitted for the 014 Rx PTA Balloon Catheter.
Conclusions
Based upon the device description, indications for use, technological characteristics & performance data it can be concluded that the 014 RX PTA Dilatation Catheter is substantially equivalent to the predicate devices and is appropriate for the intended use.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).