(98 days)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy.
The provided text describes the acceptance criteria and the study that proves Ansell Healthcare Products LLC's Microflex® Nitrile Patient Examination Gloves (Blue, Green, and Pink colored) meet these criteria for use with chemotherapy drugs and, for the pink glove, Fentanyl Citrate.
Here's the breakdown as requested:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the gloves are based on various ASTM standards and ISO standards for biocompatibility, in addition to performance against specific chemotherapy drugs. The reported device performance is directly compared against these standards and against a legally marketed predicate device (Micro-Touch NitraTex Sterile Nitrile Powder-Free Blue Examination Gloves, K082457).
Acceptance Criteria & Device Performance – Summary Table
| Characteristic | Standard/Acceptance Criteria | Reported Device Performance (Microflex® Nitrile Patient Examination Gloves) |
|---|---|---|
| Physical Properties: | ASTM D6319-10 (Nitrile Examination Gloves for Medical Application) | Meets ASTM D6319-10 requirements |
| Dimensions: | Length: Minimum 230mm; Palm width (XS-XL): 70±10 to 120±10 mm; Thickness (Finger, Palm): minimum 0.05 mm | Length: Minimum 240mm; Palm width (XS-XL): 75±5 to 115±5 mm; Thickness (Finger, Palm, Cuff): Finger - min 0.09mm, Palm - min 0.06mm, Cuff - min 0.05mm. |
| Tensile Strength: | Before Aging: minimum 14 MPa; After Aging: minimum 14 MPa | Before Aging: minimum 16 MPa; After Aging: minimum 14 MPa |
| Ultimate Elongation: | Before Aging: minimum 500%; After Aging: minimum 400% | Before Aging: minimum 500%; After Aging: minimum 400% |
| Freedom from holes: | ASTM D6319-10 & ASTM D5151-06 requirements of AQL 2.5 | Meets or exceeds ASTM D6319-10 and ASTM D5151-06 requirements of AQL 2.5 |
| Powder Residual: | ASTM D6319-10 & ASTM D6124-06: ≤ 2 mg per glove | Meets applicable requirement for powder free; ≤ 2 mg per glove |
| Biocompatibility: | ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Skin Irritation, Sensitization), ISO 10993-11: 2006 (Acute Systemic Toxicity) | Cytotoxicity: Under the conditions of the study, undiluted, 1:2, 1:4, 1:8, 1:16 dilution was cytotoxic. 1:32 and 1:64 are not cytotoxic. Skin Irritation: Under the conditions of the study, not an irritant.Sensitization: Under the conditions of the study, not a sensitizer.Acute Systemic Toxicity: Under the conditions of the study, no evidence of systemic toxicity. |
| Chemotherapy Drug Permeation (ASTM D6978-05): | Predicate Device Permeation Times (for comparison) | Blue Gloves: |
| Carmustine (BCNU) - 3.3 mg/ml | Predicate: 32.1 min | Blue: 17.4 min; Green: 27.9 min; Pink: 23.4 min |
| Cyclophosphamide - 20.0 mg/ml | Predicate: >240 min | All: >240 min |
| Doxorubicin HCl - 2.0 mg/ml | Predicate: >240 min | All: >240 min |
| Etoposide - 20.0 mg/ml | Predicate: >240 min | All: >240 min |
| Fentanyl Citrate - 100 mcg/2mL | N/A (Tested for Pink only) | Pink: >240 min |
| Fluorouracil - 50.0 mg/ml | Predicate: >240 min | All: >240 min |
| Methotrexate - 25.0 mg/ml | Predicate: >240 min | All: >240 min |
| Paclitaxel - 6.0 mg/ml | Predicate: >240 min | All: >240 min |
| Thiotepa (THT) - 10.0 mg/ml | Predicate: 140.7 min | Blue: 67.1 min; Green: 48.6 min; Pink: 64.9 min |
| Vincristine Sulfate - 1.0 mg/ml | Predicate: >240 min | All: >240 min |
| Warning: Carmustine and Thiotepa have low permeation times. The warning "Do not use with Carmustine" is consistently included for all glove types. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify the exact sample sizes (number of gloves or test replicates) used for each physical, chemical (chemotherapy permeation), and biocompatibility test. It only provides the summarized results (e.g., minimum values, average breakthrough times, or "meets requirements").
- Data Provenance: The data comes from the testing performed by the manufacturer, Ansell Healthcare Products LLC, for their 510(k) premarket notification. The document itself does not specify the country of origin where the testing was physically conducted. The data is retrospective in the sense that it was generated for the purpose of this submission, rather than a prospective clinical trial.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
- This submission is for medical gloves and relies on laboratory-based performance testing against established ASTM and ISO standards, not human expert interpretation of medical images or patient data. Therefore, the concept of "ground truth" established by a panel of medical experts (like radiologists for imaging devices) is not applicable here. The "ground truth" is defined by the objective measurement protocols outlined in the referenced standards (e.g., ASTM D6319-10 for physical characteristics, ASTM D6978-05 for chemotherapy drug permeation, ISO 10993 series for biocompatibility). The testing would have been conducted by trained laboratory personnel in accordance with these validated methods.
4. Adjudication Method for the Test Set
- As the "ground truth" is based on objective laboratory measurements against defined standards, there is no adjudication method by multiple human reviewers in the context of medical image interpretation. The testing results are quantitative and directly compared to the numerical criteria specified in the standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. This device is a medical glove, not an AI-assisted diagnostic tool. Therefore, the concept of assessing human reader improvement with or without AI assistance is not relevant to this submission. The device's performance is measured by its physical properties, barrier protection, and biocompatibility.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done
- No standalone algorithm performance study was done. This device is a physical product (a glove), not a software algorithm. Its performance is assessed through laboratory tests as described above.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
- The "ground truth" for this device is based on objective, standardized laboratory test methods and their defined acceptance criteria. This includes:
- Physical Properties: Measurements (dimensions, tensile strength, elongation) compared against ASTM D6319-10.
- Barrier Integrity: Freedom from holes assessed against ASTM D6319-10 and ASTM D5151-06 (AQL 2.5).
- Chemical Permeation: Breakthrough detection times for chemotherapy drugs and Fentanyl Citrate determined per ASTM D6978-05.
- Biocompatibility: In vitro and in vivo testing for cytotoxicity, skin irritation, sensitization, and acute systemic toxicity evaluated against ISO 10993 series.
8. The Sample Size for the Training Set
- This submission describes a medical device (gloves), not an AI/ML algorithm. Therefore, the concept of a "training set" for model development is not applicable. The gloves were manufactured and then tested according to established quality control and regulatory requirements.
9. How the Ground Truth for the Training Set was Established
- As there is no training set for an AI/ML model, there is no ground truth established for a training set. The device's performance is demonstrated through direct testing against predefined physical, chemical, and biological standards.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 19, 2020
Ansell Healthcare Products LLC Donald Cronk Associate Director, Regulatory Affairs - Americas 2301 Robb Drive Reno, Nevada 89523
Re: K200671
Trade/Device Name: Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl CitrateRegulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO (for pink glove only) Dated: May 21, 2020 Received: May 22, 2020
Dear Donald Cronk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200671
Device Name
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
|---|---|
| Carmustine (BCNU) - 3.3 mg/ml | 17.4 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 67.1 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmusine (BCNU) : 17.4 minutes and Thiotepa: 67.1 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K200671
Device Name
Microllex® Nitrile Patient Examination Gloves with Aloe and Chamonile Green Colored Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Tested chemotherapy drugs are as follows: | |
|---|---|
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
| Carmustine (BCNU) - 3.3 mg/ml | 27.9 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 48.6 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 27.9 minutes and Thiotepa: 48.6 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Indications for Use
510(k) Number (if known) K200671
Device Name
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Druss and Fentanyl Citrate
Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Tested chemotherapy drugs are as follows: | |
|---|---|
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
| Carmustine (BCNU) - 3.3 mg/ml | 23.4 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fentanyl Citrate Injection - 100 mcg/2mL | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 64.9 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation (BCNU): 23.4 minutes and Thiotepa: 64.9 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
510K Summary
510(k) Number:
Submitter:
Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523
Contact Person:
Don Cronk Associate Director, Regulatory Affairs - Americas Phone: (775) 470-7106 Email: don.cronk@ansell.com
Date Prepared:
6/19/2020
Name of the Device:
| Trade Names: | Microflex® Nitrile Patient Examination Gloves with Aloe and ChamomileBlue Colored Tested for Use with Chemotherapy DrugsMicroflex® Nitrile Patient Examination Gloves with Aloe and ChamomileGreen Colored Tested for Use with Chemotherapy DrugsMicroflex® Nitrile Patient Examination Gloves with Aloe and ChamomilePink Colored Tested for Use with Chemotherapy Drugs and FentanylCitrate |
|---|---|
| Common Name: | Patient Examination Glove |
| Classification Name: | Patient Examination Glove |
| Classification Regulation: | 21 CFR 880.6250 |
| Device Class: | I |
| Product Code: | LZA, LZC, QDO |
| Classification Panel: | Non-powdered patient examination glove |
Legally Marketed Predicate Device:
| Company: | Ansell Healthcare Products LLC |
|---|---|
| Trade Name: | Micro-Touch NitraTex Sterile Nitrile Powder-Free Blue ExaminationGloves (Chemotherapy Use) |
| 510(k) Number: | K082457 |
| Device Class: | Class I |
| Product Code: | LZA (Nitrile) |
| Device Name: | Patient Examination Glove (21 CFR 880.6250) |
Device Description:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy.
{6}------------------------------------------------
Characteristic:
- Ambidextrous with beaded cuff and straight fingers
- Finger-textured, ●
- Blue, green or pink colored
- Containing aloe and chamomile.
- Five (5) sizes – extra-small, small, medium, large, and extra-large.
- . Tested against chemotherapy drugs and fentanyl citrate (pink only).
High levels of ozone will degrade rubber material of the glove, therefore the glove should be protected from ozone in particular.
The gloves are designed to meets the specifications of ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
Indications for Use Statements:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|---|---|
| Carmustine (BCNU) – 3.3 mg/ml | 23.4 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fentanyl Citrate Injection – 100 mcg/2mL | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 64.9 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 23.4 minutes and Thiotepa: 64.9 minutes. Warning: Do not use with Carmustine.
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
{7}------------------------------------------------
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|---|---|
| Carmustine (BCNU) – 3.3 mg/ml | 17.4 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 67.1 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 17.4 minutes and Thiotepa: 67.1 minutes. Warning: Do not use with Carmustine.
Microflex® Nitrile Patient Examination Glove with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|---|---|
| Carmustine (BCNU) – 3.3 mg/ml | 27.9 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 48.6 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 27.9 minutes and Thiotepa: 48.6 minutes. Warning: Do not use with Carmustine.
Technological Characteristics:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate have the following technological characteristics as compared to ASTM or equivalent standards:
{8}------------------------------------------------
| Characteristics | Standard/Test/FDA Guidance | Result Summary |
|---|---|---|
| Physical Characteristics: | ||
| Dimensions: | ASTM D6319-10 | Meets ASTM D6319-10 requirements forlength, width and thickness |
| Length | Minimum 230mm | Minimum 240mm |
| Palm width (mm) | ||
| Size - XS | 70 ± 10 | 75 ± 5 |
| Size - S | 80 ± 10 | 85 ± 5 |
| Size - M | 95 ± 10 | 95 ± 5 |
| Size - L | 110± 10 | 105 ± 5 |
| Size - XL | 120 ± 10 | 115 ± 5 |
| Thickness (mm) - single-wall | ||
| Finger | minimum 0.05 | Finger - min 0.09 |
| Palm | minimum 0.05 | Palm - min 0.06 |
| Cuff | - | Cuff – min 0.05 |
| Physical Properties: | ASTM D6319-10 | Meets ASTM D6319-10 requirements for tensilestrength and ultimate elongation before andafter accelerated aging: |
| Tensile Strength | ||
| Before Aging | minimum 14 MPa | minimum 16 MPa |
| After Aging | minimum 14 MPa | minimum 14 MPa |
| Ultimate Elongation | ||
| Before Aging | minimum 500% | minimum 500% |
| After Aging | minimum 400% | minimum 400% |
| Freedom from holes | ASTM D6319-10ASTM D5151-06 | Meets or exceeds ASTM D6319-10 and ASTMD5151-06 requirements of AQL 2.5 |
| Powder Residual | ASTM D6319-10 | Meets applicable requirement for powder free;≤ 2 mg per glove |
| ASTM D6124-06 | ||
| Biocompatibility: | ||
| ISO In Vitro Cytotoxicity | ISO 10993-5:2009 | Under the conditions of the study, undiluted, 1:2, 1:4, 1:8, 1:16 dilution was cytotoxic., 1:32 and 1:64 are not cytotoxic |
| ISO Skin Irritation Study | ISO10993-10:2010 | Under the conditions of the study, not an irritant |
| ISO Maximization Sensitization Study | ISO 10993-10:2010 | Under the conditions of the study, not a sensitizer |
| ISO acute systemic toxicity | ISO 10993-11: 2006 | Under the conditions of the study, no evidence of systemic toxicity |
{9}------------------------------------------------
Substantial Equivalence:
| Predicate Device | Proposed Subject Device | SubstantialEquivalence toPredicate | |
|---|---|---|---|
| Trade name | Micro-Touch NitraTex SterileNitrile Powder-Free BlueExamination Gloves(Chemotherapy Use) | Microflex® Nitrile PatientExamination Gloves with Aloeand Chamomile Blue ColoredTested for Use withChemotherapy DrugsMicroflex® Nitrile PatientExamination Gloves with Aloeand Chamomile Green ColoredTested for Use withChemotherapy DrugsMicroflex® Nitrile PatientExamination Gloves with Aloeand Chamomile Pink ColoredTested for Use withChemotherapy Drugs andFentanyl Citrate | Not applicable |
| 510k Number | K082457 | K200671 | Not Applicable |
| Product Owner | Ansell Healthcare | Ansell Healthcare | Ansell Healthcare |
| Product Code | LZA | LZA, LZC, QDO | Similar |
{10}------------------------------------------------
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Yes |
|---|---|---|---|
| Regulatory Class | I | I | Yes |
| Regulation Name | Patient Examination Glove | Patient Examination Glove | Yes |
| Indications for use | The patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | Yes |
| MaterialComposition | Synthetic nitrile rubber | Synthetic nitrile rubber | Yes |
| Coating | Polyacrylic polymer inner coating to aid donning | Polyacrylic polymer inner coating to aid donning | Yes |
| Aloe andChamomile | N/A | Aloe and Chamomile coated on the donning surface | As noted |
| Design | Non-sterile | Non-sterile | Yes |
| Single use | Single use | Yes | |
| Powder-free | Powder-free | Yes | |
| Ambidextrous | Ambidextrous | Yes | |
| Beaded cuff | Beaded cuff | Yes | |
| Color | Blue | Green, Blue, and Pink | As noted |
| Performance | Meets ASTM D6319-10requirements | Meets ASTM D6319-10requirements | Yes |
| a. Dimensions | |||
| b. PhysicalProperties | Meets ASTM D6319-10requirements | Meets ASTM D6319-10requirements | Yes |
| c. Freedom fromholes | Meets ASTM D6319-10requirements of GI, AQL 2.5 | Meets ASTM D6319-10requirements of GI, AQL 2.5 | Yes |
| d. Powder Residual | Meets ASTM D6319-10requirements; Not more than2.0mg/glove | Meets ASTM D6319-10requirements; Not more than2.0mg/glove | Yes |
{11}------------------------------------------------
| e. Sterility | Non-sterile | Non-sterile | Yes | ||
|---|---|---|---|---|---|
| Biocompatibility | Passes Primary Skin Irritation Testand Dermal Sensitization Test | Passes Primary Skin IrritationTest and Dermal SensitizationTest and Acute Systemic ToxicityTest | Yes | ||
| ChemotherapyClaim | Predicate Device | Pink Device | As Noted (FentanylCitrate) | ||
| Carmustine (BCNU) | 32.1 | Carmustine (BCNU) | 23.4 | ||
| Cyclophosphamide | >240 | Cyclophosphamide | >240 | ||
| Doxorubicin HCl(Adriamycin) | >240 | Doxorubicin HCl(Adriamycin) | >240 | ||
| Etoposide (Toposar) | >240 | Etoposide (Toposar) | >240 | ||
| Fluorouracil (Adrucil) | >240 | Fluorouracil (Adrucil) | >240 | ||
| Methotrexate | >240 | Methotrexate | >240 | ||
| Paclitaxel (Taxol) | >240 | Paclitaxel (Taxol) | >240 | ||
| Thiotepa (THT) | 140.7 | Thiotepa (THT) | 64.9 | ||
| Vincristine Sulfate | >240 | Vincristine Sulfate | >240 | ||
| Fentanyl CitrateInjection | >240 | ||||
| ChemotherapyClaim | Predicate Device | Blue Device | Yes | ||
| Carmustine (BCNU) | 32.1 | Carmustine (BCNU) | 17.4 | ||
| Cyclophosphamide | >240 | Cyclophosphamide | >240 | ||
| Doxorubicin HCl(Adriamycin) | >240 | Doxorubicin HCl(Adriamycin) | >240 | ||
| Etoposide (Toposar) | >240 | Etoposide (Toposar) | >240 | ||
| Fluorouracil (Adrucil) | >240 | Fluorouracil (Adrucil) | >240 | ||
| Methotrexate | >240 | Methotrexate | >240 | ||
| Paclitaxel (Taxol) | >240 | Paclitaxel (Taxol) | >240 | ||
| Thiotepa (THT) | 140.7 | Thiotepa (THT) | 67.1 | ||
| Vincristine Sulfate | >240 | Vincristine Sulfate | >240 |
{12}------------------------------------------------
| ChemotherapyClaim | Predicate Device | Green Device | Yes |
|---|---|---|---|
| Carmustine (BCNU) 32.1 | Carmustine (BCNU) 27.9 | ||
| Cyclophosphamide >240 | Cyclophosphamide >240 | ||
| Doxorubicin HCl(Adriamycin) >240 | Doxorubicin HCl(Adriamycin) >240 | ||
| Etoposide (Toposar) >240 | Etoposide (Toposar) >240 | ||
| Fluorouracil (Adrucil) >240 | Fluorouracil (Adrucil) >240 | ||
| Methotrexate >240 | Methotrexate >240 | ||
| Paclitaxel (Taxol) >240 | Paclitaxel (Taxol) >240 | ||
| Thiotepa (THT) 140.7 | Thiotepa (THT) 48.6 | ||
| Vincristine Sulfate >240 | Vincristine Sulfate >240 |
The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6319, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization, irritation and acute systemic toxicity, in accord with the ISO 10993 series of standards.
A clinical study was not required for the subject or predicate devices.
Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.