K Number
K200403

Validate with FDA (Live)

Date Cleared
2020-04-14

(55 days)

Product Code
Regulation Number
862.1665
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stat Profile® Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of Sodium, Potassium, Chloride, Ionized Calcium, and Ionized Magnesium in heparinized arterial and venous whole blood.

Device Description

The Stat Profile Prime Plus Analyzer System is a low cost, low maintenance analyzer for hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that insures the analyzer is working properly at all times.

The Stat Profile Prime Plus Analyzer accepts samples from syringes and open tubes. The minimum sample size for analysis is 135 µL.

Sample collection, preparation and application to the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Stat Profile® Prime Plus Analyzer System, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a quantitative table format. Instead, it presents performance data (method comparison, imprecision) that would be evaluated against established clinical or analytical goals to demonstrate substantial equivalence. The predicate device's performance often serves as the de facto acceptance benchmark.

I will interpret the "Measurement Range" from Table 4 as an implicit acceptance criterion for the device's operational range, and the "Method Comparison Studies" and "Total Imprecision Performance" tables as reported device performance demonstrating equivalence.

ParameterAcceptance Criteria (Measurement Range)Reported Device Performance (Method Comparison: Slope, Intercept, r)Reported Device Performance (Total Imprecision: Total %CV - Level 4/5/Linearity)Reported Device Performance (Within-Run Precision: %CV)
Sodium (Na)80-200 mmol/LSlope: 0.9964, Intercept: 0.4488, r: 0.9949Level 4: 0.6%, Level 5: 0.8%, Linearity: 1.2%Samples 1-7: 0.22% - 0.67%
Potassium (K)1.0-20.0 mmol/LSlope: 1.0158, Intercept: -0.0678, r: 0.9993Level 4: 2.2%, Level 5: 1.9%, Linearity: 1.5%Samples 1-7: 0.00% - 2.66%
Chloride (Cl)50-200 mmol/LSlope: 0.9963, Intercept: 0.4416, r: 0.9971Level 4: 0.6%, Level 5: 0.6%, Linearity: 0.9%Samples 1-7: 0.00% - 0.67%
Ionized Calcium (iCa)0.4-10.8 mg/dLSlope: 0.9820, Intercept: 0.0239, r: 0.9871Level 4: 2.1%, Level 5: 2.7%, Linearity: 1.9%Samples 1-7: 0.40% - 1.13%
Ionized Magnesium (iMg)0.24-3.65 mg/dLSlope: 1.0020, Intercept: -0.0021, r: 0.9910Level 4: 4.4%, Level 5: 6.5%, Linearity: 5.4%Samples 1-7: 0.52% - 2.63%

2. Sample Size Used for the Test Set and Data Provenance

  • Method Comparison Studies (Test Set):

    • Sample Sizes:
      • Na: 432 samples (18 altered samples)
      • K: 435 samples (21 altered samples)
      • Cl: 434 samples (20 altered samples)
      • iCa: 434 samples (20 altered samples)
      • iMg: 426 samples (13 altered samples)
    • Data Provenance: The study was a Point-of-Care (POC) study conducted at 3 POC sites including a Cardiothoracic Intensive Care Unit (CTICU), an Emergency Department (ED), and a Respiratory Therapy Lab (RT). The data provenance is described as comparing "results obtained by trained Healthcare Professionals to results obtained by POC personnel on the same specimens using the same analyzer." The specimens were either quality control materials or discarded blood gas specimens.
    • Retrospective or Prospective: Not explicitly stated, but the description of "comparing results obtained by trained Healthcare Professionals to results obtained by POC personnel" suggests a prospective collection or at least a controlled, concurrent comparison for the purpose of the study. It's not described as a retrospective analysis of existing patient data.
  • Total Imprecision Performance (Test Set):

    • The estimates were obtained from different POC personnel running 3 levels of Quality Control/Linearity Materials in duplicate each day for a total of 20 runs on 3 analyzers. (Implies a substantial number of measurements, but the exact N for each measurement type is not detailed beyond "20 runs").
  • Within-Run Whole Blood Precision (Test Set):

    • Each precision run consisted of ten (10) replicate measurements. A total of five (5) different native samples and two (2) altered samples were evaluated at each site. This was done by a minimum of two (2) point-of-care operators at each of the three (3) POC sites, for a total of nine (9) operators.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The study compares the performance of the Stat Profile® Prime Plus Analyzer System against existing laboratory methods (referred to as "Lab" in Table 1) and its own predicate device.
For the Method Comparison Studies, the "ground truth" or reference method is the "Lab" result, which would typically be generated by a validated, high-accuracy laboratory analyzer. The document does not specify individual human experts or their qualifications for establishing this ground truth; rather, it refers to the "Lab" as the reference. For clinical laboratory devices, the "ground truth" is usually the result from a recognized reference standard method or a highly accurate laboratory instrument rather than individual expert consensus.

For the Imprecision studies, the "ground truth" is typically the measured mean of repeated measurements, and the accuracy is relative to a known value for Quality Control materials. Again, this doesn't involve human experts establishing ground truth in the way it might for imaging studies.

4. Adjudication Method for the Test Set

This type of diagnostic device (blood analyzer) does not typically involve human adjudication in the same way as, for example, image-based diagnostic systems. The method comparison studies compare the device's quantitative output to that of a reference laboratory method. No specific adjudication method like "2+1" or "3+1" is mentioned or applicable here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-based diagnostic systems where human readers interpret medical images or data. The Stat Profile® Prime Plus Analyzer System is a quantitative in vitro diagnostic device, not an AI-assisted diagnostic tool that supports human interpretation in that manner. The study focused on demonstrating the analytical performance and equivalence of the device to a predicate and laboratory methods.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

The device itself is a standalone analyzer that measures parameters and produces results. The performance studies (Method Comparison, Imprecision) evaluate the device's standalone analytical performance. Although POC personnel operate the device, the data presented (slopes, intercepts, r-values, %CV) reflects the intrinsic performance of the algorithm/hardware combination (the "device") in generating quantitative measurements, rather than human interpretation. So, yes, the performance metrics reported are for the device operating in a standalone capacity (as a measurement instrument).

7. The Type of Ground Truth Used

  • Method Comparison Studies: The ground truth for the method comparison studies was obtained from a reference laboratory method or a "Lab" analyzer. This implies comparison to established, presumably accurate, laboratory instrumentation with well-defined performance characteristics.
  • Imprecision and Within-Run Precision Studies: The ground truth for these studies relies on the known concentration values of quality control materials and the statistical analysis of repeated measurements to determine the reproducibility and variability of the device.

8. The Sample Size for the Training Set

The document does not describe the device as employing a machine learning or AI algorithm that requires a distinct "training set" in the conventional sense of AI/ML development. The device uses established "Ion-Selective Electrode (ISE)" technology and measurement algorithms. Therefore, there is no mention of a "training set" for an AI model.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" for an AI model, this question is not applicable. The measurement principles are based on known electrochemical properties rather than learned patterns from a training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

April 14, 2020

Nova Biomedical Corporation Rachel Gilbert Regulatory Affairs Specialist II 200 Prospect Street Waltham, MA 02454

Re: K200403

Trade/Device Name: Stat Profile® Prime Plus Analyzer System Regulation Number: 21 CFR 862.1665 Regulation Name: Sodium test system Regulatory Class: Class II Product Code: JGS, CEM, CGZ, JFP, CFA Dated: February 12, 2020 Received: February 19, 2020

Dear Rachel Gilbert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200403

Device Name

Stat Profile® Prime Plus Analyzer System

Indications for Use (Describe)

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of Sodium, Chloride, Ionized Calcium, and Ionized Magnesium in heparinized arterial and venous whole blood.

Sodium (Na)Sodium measurements are used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseasesinvolving electrolyte imbalance.
Potassium (K)Potassium measurements are used in the diagnosis and treatment of disease conditionscharacterized by low or high potassium levels.
Chloride (Cl)Chloride measurements are used in the diagnosis and treatment of electrolyte andmetabolic disorders such as cystic fibrosis and diabetic acidosis.
Ionized Calcium(iCa)Ionized calcium measurements are used in the diagnosis and treatment of parathyroiddisease, a variety of bone diseases, chronic renal disease and tetany (intermittentmuscular contractions or spasms)
IonizedMagnesium (iMg)Ionized magnesium measurements are used in the diagnosis and treatment ofhypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia(abnormally high levels of magnesium).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K200403

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Rachel Gilbert, Regulatory Affairs Specialist
Date Prepared:February 12, 2020

Proprietary Name: Stat Profile® Prime Plus Analyzer System

Common or Usual Name: Blood Analyzer

Classification Name: Multiple

Regulation SectionClassificationProduct Code
21 CFR § 862.1665 Sodium Test SystemClass IIJGS
21 CFR § 862.1600 Potassium Test SystemClass IICEM
21 CFR § 862.1170 Chloride Test SystemClass IICGZ
21 CFR § 862.1145 Calcium Test SystemClass IIJFP
21 CFR § 862.1495 Magnesium Test SystemClass I, reservedCFA

Product Code:

ിട്ടു

Predicate Device:

K180428 – Stat Profile Prime Plus Analyzer System

Device Description:

The Stat Profile Prime Plus Analyzer System is a low cost, low maintenance analyzer for hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

Primary Sensor Card Port:

There are two options for the primary sensor card:

  • Primary Sensor Card 1 shall enable and report the following listed analytes: . o Sodium, Potassium, Chloride, Calcium, Ionized Magnesium
  • . Primary Sensor Card 2 shall enable and report the following listed analytes:
    • o Sodium, Potassium, Chloride, Calcium, Ionized Magnesium

{4}------------------------------------------------

As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that insures the analyzer is working properly at all times.

The Stat Profile Prime Plus Analyzer accepts samples from syringes and open tubes. The minimum sample size for analysis is 135 µL.

Sample collection, preparation and application to the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

Stat Profile Prime Plus Analyzer System Components:

The Stat Profile Prime Plus Analyzer System is comprised of the following components.

  • Stat Profile Prime Plus Analyzer System ●
  • . Primary Sensor Cartridge
  • Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
  • Stat Profile Prime Plus Calibrator Cartridge ●
  • Stat Profile Prime Plus External Ampuled Control
  • . IFU/Labeling

Sample Types:

The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial and venous whole blood.

Measured Parameters:

The Stat Profile Prime Plus Analyzer measures:

  • Sodium (Na) ●
  • Potassium (K)
  • . Chloride (CI)
  • Ionized Calcium (iCa) ●
  • lonized Magnesium (iMg) ●

Calculated Parameters:

The Prime Plus Analyzer also provides the following parameter results calculated based on results of the directly measured parameters:

  • Anion Gap (Gap) ●
  • Blood Osmolality (OSM) ●
  • nCA to nMg Ratio (nCa/nMg)
  • Normalized Calcium (nCa)
  • Normalized Magnesium (nMq)

Intended Use:

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of Sodium, Potassium, Chloride, lonized Calcium, and lonized Magnesium in heparinized arterial and venous whole blood.

{5}------------------------------------------------

Indications for Use:

Sodium (Na)Sodium measurements are used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, ordiseases involving electrolyte imbalance.
Potassium (K)Potassium measurements are used in the diagnosis and treatment of diseaseconditions characterized by low or high potassium levels.
Chloride (CI)Chloride measurements are used in the diagnosis and treatment of electrolyteand metabolic disorders such as cystic fibrosis and diabetic acidosis.
Ionized Calcium (iCa)Ionized calcium measurements are used in the diagnosis and treatment ofparathyroid disease, a variety of bone diseases, chronic renal disease and tetany(intermittent muscular contractions or spasms)
Ionized Magnesium (iMg)Ionized magnesium measurements are used in the diagnosis and treatment ofhypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia(abnormally high levels of magnesium).

Summary of the Technological Characteristics:

The Stat Profile Prime Plus Analyzer is substantially equivalent to the previously cleared for market Stat Profile Prime Plus Analyzer in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters.

Principle of Measurement:

Sodium, Potassium, Chloride, Ionized Calcium, Ionized Magnesium:

The parameters are measured by an lon-Selective Electrode (ISE) that selectively measures the activity of ionic species. When the ISE is contacted with a sample, potential is developed. The potential is proportional to the logarithm of the ionic activity and is measured versus a reference electrode.

Summary of Performance Testing:

Bench testing was previously completed and summarized in K180428 to show that the Stat Profile Prime Plus Analyzer demonstrates substantial equivalence to the predicate submission.

The bench testing included:

  • Method Comparison Studies
  • Precision/Reproducibility Within Run and Run to Run Studies ●
  • Linearity Testing ●
  • . Specificity/Interference Testing
  • Detection Limit .

Summary of Point-of-Care Testing:

A Point-of-Care (POC) study was conducted to show that the Stat Profile Prime Plus Analyzer demonstrates substantial equivalence to the predicate submission. The testing compared results obtained by trained Healthcare Professionals to results obtained by POC personnel on the same specimens using the same analyzer. The Stat Profile Prime Plus Analyzer was evaluated by point-of-care (POC) personnel in 3 POC sites including a Cardiothoracic Intensive Care Unit (CTICU), an Emergency Department (ED) and a Respiratory Therapy Lab (RT). A total of 61 Respiratory Care, 12 Nursing, and 1 Exercise Physiology POC personnel participated from the 3 POC settings over the study. The personnel represent trained, qualified staff found in typical POC sites where blood gas analyzers are utilized. All testing was performed using quality control materials or discarded blood gas specimens.

{6}------------------------------------------------

Method Comparison Studies:

Method Comparison studies on venous and arterial whole blood specimens were conducted using methods described in CLSI "Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline - Second Edition", CLSI EP9-A2. Combined method comparison data from all 3 POC settings is summarized in Table 1.

analyteN# alteredsamplesrangeSlopeInterceptrMDL95% ConfidenceInterval of Bias
LowerLimitUpperLimit
Na4321890.0 - 187.00.99640.44880.9949120160119.8120.2
159.6160.1
K435211.10 - 17.601.0158-0.06780.99932.86.02.772.81
5.986.04
Cl4342056.0 - 173.00.99630.44160.99718013079.980.4
129.8130.1
iCa434200.51 - 2.480.98200.02390.98710.801.540.800.82
1.531.54
iMg426130.24 - 1.361.0020-0.00210.99100.30.80.290.30
0.800.80
Table 1: Venous and Aterial Whole Blood Method Comparison Results – Point of Care vs Lab (ED
RT and CTICU)

Total Imprecision Performance:

The estimates for total impression were obtained from different POC personnel running 3 levels of Stat Profile Prime Plus Quality Control/Linearity Materials in duplicate each day for a total of 20 runs on 3 Stat Profile Prime Plus analyzers. The protocol was based upon methods described in CLSI Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition, CLSI EP5-A2T. The total imprecision data from one representative POC site is shown in Table 2 is representative of the expected total imprecision performance obtainable by POC personnel using the Stat Profile Prime Plus analyzer using external quality control and linearity materials.

MeanWithin RunSDWithin Run%CVTotal SDTotal %CV
Total Imprecision Data-Level 4
Na (mmol/L)136.20.80.60.90.6
K (mmol/L)3.880.051.40.082.2
Cl (mmol/L)123.20.60.50.70.6
Ca (mmol/L)1.070.021.70.022.1
Mg (mmol/L)0.600.023.60.034.4
Total Imprecision Data-Level 5
Na (mmol/L)109.81.00.90.90.8
K (mmol/L)6.240.061.00.121.9
Cl (mmol/L)95.10.40.40.60.6
Ca (mmol/L)1.560.042.80.042.7
Mg (mmol/L)1.070.032.70.076.5

Table 2: Total Imprecision From ED Site

{7}------------------------------------------------

MeanWithin RunSDWithin Run%CVTotal SDTotal %CV
Total Imprecision Data-Linearity Level 4
Na (mmol/L)168.61.50.92.01.2
K (mmol/L)11.640.121.00.181.5
Cl (mmol/L)132.71.00.81.20.9
Ca (mmol/L)0.480.011.60.011.9
Mg (mmol/L)0.240.012.30.015.4

Within-Run Whole Blood Precision:

Whole blood with-run precision of the Stat Profile Prime Plus Analyzer System in the hands of point-ofcare operators was assessed by a minimum of two (2) point-of-care operators at each of the three (3) POC sites for a total of nine (9) operators across the 3 testing locations. Each precision run consisted of ten (10) replicate measurements using both fresh, native and altered whole blood samples. A total of five (5) different native samples and two (2) altered samples were evaluated at each site. Each whole blood specimen was maintained in a syringe. The POC operator performed all sample analysis steps including sample analysis, removal of resultant air bubble(s) from the syringe, recapping of the syringe and mixing prior to the next sample analysis. The whole blood within-run precision data from one representative POC site is shown in Table 3 and is representative of the expected within-run precision obtainable by POC personnel using the Stat Profile Prime Plus analyzer using whole blood samples.

Table 3: Within Run Precision with Whole Blood Samples (n=10) – ED

MeanSD%CV95% CI
Sample 1
Na (mmol/L)123.70.820.67122 - 125
K (mmol/L)4.710.030.674.6 - 4.8
Cl (mmol/L)88.60.520.5888 - 90
Ca (mmol/L)1.240.010.431.22 - 1.25
Mg (mmol/L)0.560.011.500.55 - 0.58
Sample 2
Na (mmol/L)138.20.420.31137 - 139
K (mmol/L)1.940.052.661.8 - 2.0
Cl (mmol/L)116.80.630.54116 - 118
Ca (mmol/L)0.750.011.130.73 - 0.76
Mg (mmol/L)0.310.012.630.30 - 0.33
Sample 3
Na (mmol/L)141.10.320.22140 - 142
K (mmol/L)4.020.041.053.9 - 4.1
Cl (mmol/L)103.80.420.41103 - 105
Ca (mmol/L)1.280.010.401.27 - 1.29
Mg (mmol/L)0.610.000.520.60 - 0.62
Sample 4
Na (mmol/L)142.90.570.40142 - 144
K (mmol/L)4.110.030.774.0 - 4.2
Cl (mmol/L)103.00.000.00103 - 103
Ca (mmol/L)1.310.010.541.29 - 1.32
Mg (mmol/L)0.620.011.020.61 - 0.63

{8}------------------------------------------------

MeanSD%CV95% CI
Sample 5
Na (mmol/L)137.90.320.23137 - 139
K (mmol/L)3.300.000.003.3 - 3.3
Cl (mmol/L)109.10.320.29108 - 110
Ca (mmol/L)0.980.010.520.97 - 0.99
Mg (mmol/L)0.460.000.910.45 - 0.47
Sample 6 (Altered)
Na (mmol/L)133.60.700.52132 - 135
K (mmol/L)3.730.051.303.6 - 3.8
Cl (mmol/L)104.40.700.67103 - 106
Ca (mmol/L)1.130.010.951.11 - 1.16
Mg (mmol/L)0.480.012.290.46 - 0.50
Sample 7 (Altered)
Na (mmol/L)138.30.670.49137 - 140
K (mmol/L)7.910.091.117.7 - 8.1
Cl (mmol/L)109.60.520.47109 - 111
Ca (mmol/L)1.220.010.611.20 - 1.23
Mg (mmol/L)0.580.010.980.57 - 0.59

Conclusion:

The results of the POC clinical performance verification testing confirmed that the Stat Profile Prime Plus
Analyzer is substantially equivalent to the predicate Stat Profile

{9}------------------------------------------------

CharacteristicPredicate:Proposed:
Indication For UseThe Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of sodium, potassium, chloride, ionized calcium, and ionized magnesium in heparinized arterial and venous whole blood.The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of care usage for quantitative determination of sodium, potassium, chloride ionized calcium, and ionized magnesium in heparinized arterial and venous whole blood.
Acceptable Samples
Sample TypesLithium heparin whole blood from syringes and open tubesSame
Sample Volumes135µLSame
Measurement Range
Sodium80-200 mmol/LSame
Potassium1.0-20.0 mmol/LSame
Chloride50-200 mmol/LSame
Ionized Calcium0.4-10.8 mg/dLSame
Ionized Magnesium0.24-3.65 mg/dLSame
Principles ofMeasurement
SodiumIon-Selective ElectrodeSame
PotassiumIon-Selective ElectrodeSame
ChlorideIon-Selective ElectrodeSame
Ionized CalciumIon-Selective ElectrodeSame
Ionized MagnesiumIon-Selective ElectrodeSame
Touch Screen10.1" WXGA 1280 x 800 color touch screenSame
MenuFully configurable test menu based on available sensorsSame
Bar Code ScannerInternal Integrated 1D/2DSame
Printer2" Roll, Thermal TransferSame
PumpPeristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT)Same
Analog BoardPrecision low level analog front end w/ amperometric and potentiometric amplifiers, air detector circuitry and temperature control circuitrySame
Table 4: Comparison of Predicate and Proposed Devices
---------------------------------------------------------------

§ 862.1665 Sodium test system.

(a)
Identification. A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.(b)
Classification. Class II.