K Number
K200287
Device Name
BP-BT Kiosk
Manufacturer
Date Cleared
2020-05-15

(100 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BP-BT Kiosk is designed to measure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 6.7 inches (17.0 cm) to 16.5 inches (42.0 cm). Indication: Prescription use in medical or clinic environment only.Rx only.

Device Description

BP-BT Kiosk has the same design as the predicated device with an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is first adapted to the arm size and then, inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300 mmHg. BP-BT Kiosk will not inflate the cuff higher than 300 mmHg. BP-BT Kiosk will turn on an irregular heartbeat if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and transmitted via Bluetooth wireless module. The cuff is also deflated automatically to 0 mmHg at the same time.

AI/ML Overview

The provided text describes the BP-BT Kiosk, a non-invasive blood pressure measurement system, and its clinical validation for substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the BP-BT Kiosk are based on the standard IEC 80601-2-30, which pertains to automated non-invasive sphygmomanometers. The reported device performance demonstrates compliance with these standards for blood pressure and pulse rate measurement accuracy.

Acceptance Criteria (Based on IEC 80601-2-30)Reported Device Performance (BP-BT Kiosk)
Blood Pressure Accuracy:
Average deviation of systolic pressure0.19 mmHg
Standard deviation of systolic pressure1.56 mmHg
Average deviation of diastolic pressure0.25 mmHg
Standard deviation of diastolic pressure1.65 mmHg
Pulse Rate Accuracy:
Error of pulse rateWithin ±5%

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: 267 sets of data were measured for the clinical test.
  • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only mentions a "Clinical Test" was performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the "number of experts" or their specific qualifications for establishing the ground truth. However, it indicates that the comparison was made against a "Mercury Blood pressure monitor," which is generally considered a gold standard (ground truth) for non-invasive blood pressure measurement in clinical settings, typically operated by trained medical professionals.

4. Adjudication Method for the Test Set

The document does not describe a specific adjudication method like "2+1" or "3+1." The evaluation appears to be a direct comparison between the BP-BT Kiosk and a Mercury Blood Pressure Monitor.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study was not done. The BP-BT Kiosk is a standalone medical device for measuring blood pressure and pulse rate, not an AI-assisted diagnostic tool that would involve human readers. Therefore, the concept of "human readers improve with AI vs. without AI assistance" does not apply to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance evaluation was done. The "Clinical Test" directly assessed the accuracy of the BP-BT Kiosk's measurements (algorithm only) against a Mercury Blood Pressure Monitor. The reported performance metrics (average and standard deviations for systolic and diastolic pressure, and pulse rate error) reflect the device's standalone accuracy without human intervention in the measurement process.

7. The Type of Ground Truth Used

The type of ground truth used for the clinical test was comparison with a Mercury Blood Pressure Monitor. This is considered a gold standard method for accurate non-invasive blood pressure measurement.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" or its sample size. This suggests that the device's capabilities were likely developed and validated using existing methodologies for non-invasive blood pressure measurement, and the clinical test served as the primary validation of its performance against a known standard.

9. How the Ground Truth for the Training Set Was Established

Since a dedicated "training set" is not mentioned in the context of machine learning or AI, the concept of establishing ground truth for a training set as typically understood in AI/ML development does not apply here. The device's underlying technology and algorithms for oscillometric blood pressure measurement would have been developed based on established physiological principles and engineering practices, with validation against reference methods during its design and development phases. The "Clinical Test" served as the final-stage validation for regulatory submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 15, 2020

LD Technology LLC Albert Maarek Quality Manager 100 N. Biscayne Blvd Suite 502 Miami, Florida 33132

Re: K200287

Trade/Device Name: BP-BT Kiosk Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 10, 2020 Received: April 14, 2020

Dear Albert Maarek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200287

Device Name BP-BT Kiosk

Indications for Use (Describe)

BP-BT Kiosk is designed to measure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 6.7 inches (17.0 cm) to 16.5 inches (42.0 cm). Indication: Prescription use in medical or clinic environment only.Rx only.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K200287

Image /page/3/Picture/1 description: The image shows the logo for LD Technology. The logo consists of a circular emblem on the left and the text "LD Technology" on the right. The emblem features a green sphere with plant-like structures extending from the bottom, all enclosed within a white circle. The text "LD Technology" is in a light green color, with "LD" in a larger, bolder font and "Technology" in a smaller, italicized font.

510(k) Premarket Notification Number: date: January, 29,2020

510(k) Summary BP-BT Kiosk

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92

  1. Submitter's Identification: Manufacturer: L.D TECHNOLOGY Address: 100 N. Biscayne Blvd, Suite 502 Miami, FL, 33132, USA Tel: 305-379-9900 E mail: albert.ldteck@gmail.com

2. Device Name / Classification

Trade name: Non-Invasive Blood Pressure measurement system Device Name and Model: BP-BT Kiosk Regulation number: 21CFR 870.1130 Product Codes: DXN Device Class: Class II Classification Name: Blood pressure monitor Classification Panel: Cardiology

3. Predicate legally marketed device

A&D Medical TM-2657 Family of Digital Blood Pressure Monitors K151953. Applicant: A&D Engineering, Inc. Product Code: DXN

4. Device Description

BP-BT Kiosk has the same design as the predicated device with an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is first adapted to the arm size and then, inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300 mmHg. BP-BT Kiosk will not inflate the cuff higher than 300 mmHg. BP-BT Kiosk will turn on an irregular heartbeat if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and transmitted via Bluetooth wireless module. The cuff is also deflated automatically to 0 mmHg at the same time. The detail of summary of substantial equivalence is listed below.

{4}------------------------------------------------

5. Intended use and indications for use

BP-BT Kiosk is designed to measure blood pressure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 6.7 inches (17.0 cm) to 16.5 inches (42.0 cm). Indication : Prescription use in medical or clinic environment only. Rx only.

6. Performances, technical specifications and materials Performances

The main purpose of the device is to measure the subject blood pressure and Pulse rate.

Technical specifications i.e. table of comparison with the predicate device Patient contact materials:

The material in contact with the patient is the cuff. The Cuff material is nylon and latex free.

Image /page/4/Picture/7 description: The image shows a black warning symbol on a white background. The symbol is an equilateral triangle with a thick black border. Inside the triangle is a large black exclamation point, indicating a warning or caution. The exclamation point is centered within the triangle, with a small space between the point and the top of the triangle.

7. Contra-indications

  • . Patients with any implanted electronic device (i.e. Pacemaker.).
  • Patients with arterial catheters
  • Patients with venous pulsations may cause erroneous reading in blood pressure (e.g. ● tricuspid valve regurgitation).
  • Patients that have low blood perfusion. Using the blood pressure device may cause skin erosion and/or pressure necrosis.
  • Patients that have double mastectomy procedure

8. Undesirable side effects:

No side effects or adverse reactions are known to date.

9. Substantial equivalence

Predicate legally marketed device:

A&D Medical TM-2657 Family of Digital Blood Pressure Monitors K151953. Applicant: A&D Engineering, Inc. Product Code: DXN

Table of comparison

InformationTM-2657 FamilyBP-7000 Kiosk
Intended useTM-2657, TM-2657P, TM-2657PBT, andTM-2657PRS are designed to measure bloodpressure (systolic and diastolic) and pulse ratein adult patients with arm circumference rangebetween 7.1 inches (18.0 cm) to 13.8 inches(35.0 cm).BP-BT Kiosk is designed to measure bloodpressure (systolic and diastolic) and pulse ratein adult patients with arm circumferencerange between 6.7 inches (17.0 cm) to 16.5inches (42.0 cm).
Power Supply100-240V AC 50/60HzAC 100-240V, 50-60Hz ; D.C.12V, 3.5A

{5}------------------------------------------------

Cuff designWinding mechanism operated bygeared motorWinding mechanism operated bygeared motor
Display TypeDisplay by LEDDisplay by LED
CommunicationWired - RS232C standard (TM-2655Family). Wireless - Bluetooth 2.1 Standard(UA-767PBT)Bluetooth serial port and USB interface.
Testing BenchType of protection against electric shock:Class I. BFCompliant with standards:60601-1,60601-1-280601-2-30.Clinical TestType of protection against electric shock:Class I. BFCompliant with standards:60601-1,60601-1-260601-2-30Clinical Test
Measurement typesSystolic and diastolic Blood pressurePulse rateMethod : oscillometricSystolic and diastolic Blood pressurePulse RateMethod: oscillometric
Operating and storageenvironmentOperating: 10°C to 40°C with 15 % to 85%RHStorage: -20°C to 60°C with 10 -95% RHOperating: 5°C to 40°C with15% to 80% RHStorage: -20 °C to 55 °C with ≤93%RH
External dimensions andweight241 (W) x 330 (H) x 390 (D) mmApprox. 5.5 kg387.7mm (L) × 244.1mm (W) × 340.2mm (H)Approx. 6 kg
Measurement ranges andAccuracyPressure: 0- 299 mmHgBP : +/- 3%, Pulse : +/- 5 %Pressure: 0- 300 mmHgBP: ± 2 mmHg Pulse rate: ± 5 %
Target populationAdult general publicAdult general public
Where usedOver the counterPrescription use in medical or clinicenvironment only.

10. Performances and Effectiveness

Performances and effectiveness are demonstrated by:

  • CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware's.
  • IEC 80601-2-30: Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. 10/31/2010 Second Edition
  • Testing comprises the Software verification and validation (SRS/SDS/STR). -
  • Risk analysis and -

{6}------------------------------------------------

Clinical Test.: 267 sets of data were measured by the blood pressure monitor and Mercury Blood pressure monitor

  • The average deviation of systolic pressure: 0.19mmHg .
  • The standard deviation of systolic pressure: 1.56 mmHg
  • The average deviation of diastolic pressure: 0.25mmHg ●
  • The standard deviation of diastolic pressure: 1.65mmHg ●
  • The error of pulse rate measured by blood pressure monitor was within ±5%. ●

The Following differences:

  • The BP-BT Kiosk has a) a larger amplitude in arm circumference range and b) a D.C. converter 12V, 3.5A comparing to the predicate device.
    Do not affect the performances and the effectiveness of the BP-BT s kiosk.

11. General Safety Concerns

The laboratory test reports, and clinical study of the BP-BT Kiosk have demonstrated the general safety of the device compared to the legally marketed predicate device.

12. Standards

  • IEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety > and essential performance. Third Edition December 2006
  • IEC60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests. Fourth Edition 2014

  • IEC 80601-2-30: Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. 10/31/2010 Second Edition

  • Guidance for: Industry; FDA staff; and Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005

  • ISO 14971: Medical devices - Application of risk management to medical devices. March 01, 2007

Conclusion

BP-BT Kiosk is equivalent in performance, technology, safety and efficacy to the legally marketed predicate device.

Premarket notification [510K] Number: K200287

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).