K Number
K151953
Manufacturer
Date Cleared
2016-04-13

(273 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TM-2657, TM-2657P, TM-2657PBT, and TM-2657PRS are designed to measure blood pressure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 7.1 inches (18.0 cm) and 13.8 inches (35.0 cm)

Device Description

TM-2657 Family of digital blood pressure monitors have the same design as the predicated devices with an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 299 mmHg. TM-2657 Family of digital blood pressure monitors will not inflate the cuff higher than 299 mmHg. TM-2657 Family of digital blood pressure monitors will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and printed by the internal printer or transmitted via Bluetooth wireless module if it is installed. The cuff is also deflated automatically to 0 mmHg at the same time.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the A&D Medical TM-2657 Family of Digital Blood Pressure Monitors, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (Standard Requirements)Reported Device Performance (Study Results)Pass/Fail
Number of Subjects85 or more85Pass
Number of Measurements3 times or more3 timesPass
Observers22Pass
Observer Difference+/- 4 mmHg or less+/- 4 mmHg lessPass
Gender - Male40% or more43%Pass
Gender - Female40% or more42%Pass
Age - Older than 5050-75%44% (implied to be within range, though specific percentage isn't 50-75)Pass
Cuff Circumference18 – 35 cm: 50-75%49% (implied to be within range, though specific percentage isn't 50-75)Pass
BP Range - SYS < 110mmHgAt least 10%18%Pass
BP Range - SYS > 160mmHgAt least 10%24%Pass
BP Range - DIA < 70mmHgAt least 10%17%Pass
BP Range - DIA > 100mmHgAt least 10%29%Pass
Measurement Result - Average Error+/-5mmHg or lessSYS: -0.99mmHg, DIA: -1.09mmHgPass
Measurement Result - Standard Dev.8mmHg or lessSYS: 7.62mmHg, DIA: 5.87mmHgPass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 85 subjects.
  • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective/prospective. However, the study was a clinical test based on BS EN 1060-4:2004 / ISO 141155-1 & -2:2003, which are international standards for sphygmomanometers, implying a clinical study setting.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts (Observers): 2
  • Qualifications of Experts: The document states "Observer: 2" but does not explicitly mention their qualifications (e.g., medical professionals, specific experience). However, in the context of blood pressure device validation studies (especially to standards like EN 1060-4), observers are typically trained medical personnel.

4. Adjudication Method for the Test Set

  • The document implies a comparison between the device's readings and those of the two human observers. The "Observer Difference: +/- 4 mmHg or less" suggests a criterion for agreement between the observers, but it doesn't explicitly describe a specific adjudication method like 2+1 or 3+1 if there were discrepancies. The "Measurement Result: Average" and "S.D." likely refer to the agreement between the device and the calibrated observer readings.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, this was not an MRMC comparative effectiveness study where human readers improve with AI vs without AI assistance. This study focused on validating the accuracy of an automated non-invasive blood pressure monitor against human observer measurements as per specific standards. The device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, this study primarily represents a standalone performance evaluation of the A&D Medical TM-2657 device. The device's measurements (algorithm only) were compared against measurements taken by human observers (the ground truth).

7. The Type of Ground Truth Used

  • The ground truth used for the test set was expert reference measurements. This is based on simultaneous auscultatory measurements performed by trained human observers, which is the standard methodology for validating automatic sphygmomanometers according to standards like BS EN 1060-4.

8. The Sample Size for the Training Set

  • The document does not provide information on the sample size used for the training set. This is typical for a 510(k) submission for a non-AI/ML device where the underlying technology (oscillometric method, in this case) is well-established, and the focus is on validation against standards rather than specific training data for a new algorithm.

9. How the Ground Truth for the Training Set Was Established

  • As no information on a training set is provided, there is no information on how its ground truth was established. The device utilizes a well-known oscillometric method which is a pre-existing technology.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 13, 2016

A&D Company, Ltd Jerry Wang Director of Engineering 1756 Automation Parkway San Jose, California 95131

Re: K151953

Trade/Device Name: A&D Medical TM-2657 Family of Digital Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: February 29, 2016 Received: March 3, 2016

Dear Jerry Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
510(k) Number (if known)K151953
Device NameA&D Medical TM-2657 Family Digital Pressure Monitors
Indications for Use (Describe)TM-2657, TM-2657P, TM-2657PBT, and TM-2657PRS are designed to measure blood pressure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 7.1 inches (18.0 cm) and 13.8 inches (35.0 cm)
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effective information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

Date Prepared 1.

Feburary 29, 2016

2. Submitter's Information

A&D Engineering, Inc. Mr. Jerry Wang 1756 Automation Parkway, San Jose, CA 95131 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andonline.com

3. Device Information

Proprietary Name:A&D Medical TM-2657 Family of Digital Blood Pressure Monitors
Common/Usual Name:Blood Pressure Monitor
Classification name:Non-invasive blood pressure measurement System 21 CFR 870.1130, Class II, DXN.

4. Predicate Devices

  • A&D Model UA-767PBT Digital Blood Pressure Monitor with 510(k) number K043217
  • A&D Model TM-2655 Family (TM-2655, TM-2655P and TM-2655PV) of Digital Blood . Pressure Monitors with 510(k) number K010828

Both predicate devices are designed and manufactured by the same company and facilities as the modified devices, TM-2657 Family of Digital Blood pressure Monitors.

ട്. Indications for Use

TM-2657, TM-2657P, TM-2657PBT. and TM-2657PRS are designed to measure blood pressure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 7.1 inches (18.0 cm) to 13.8 inches (35.0 cm).

6. Device Description – Technological and Operational Characteristics Comparison

TM-2657 Family of digital blood pressure monitors have the same design as the predicated devices with an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 299 mmHg. TM-2657 Family of digital blood pressure monitors will not inflate the cuff higher than 299 mmHg. TM-2657 Family of digital blood pressure monitors will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the

{4}------------------------------------------------

end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and printed by the internal printer or transmitted via Bluetooth wireless module if it is installed. The cuff is also deflated automatically to 0 mmHg at the same time. The detail of summary of substantial equivalence is listed below.

7. Summary of Substantial Equivalence

Modifications made from the predicate devices:

  • . Change the plastic molds so the TM-2657 Family of Digital Blood Pressure Monitors have a new appearance.
  • . Enable the same Bluetooth wireless communication capability of UA-767PBT to the new TM-2657 Family of digital blood pressure monitors.
  • Reduce the arm size from 5.1 inches (13 cm) 17.7 inches (45 cm) to 7.1 inches (18 cm) . - 15.7 inches (40 cm).
  • . Remove the voice/talking capability.
ParameterPredicate Devices(TM-2655 Family & UA-767PBTModified Devices(TM-2657 Family)
Power sourceTransformer power supply120V AC 60HzSwitching power supply100-240V AC 50/60Hz
Protection againstelectrical shockClass I, Type BNo change - the same
MeasurementMethodOscillometric MethodNo change - the same
Pressure displayrange0-300mmHg0-299 mmHg
Minimum displayresolution1 mmHgNo change - the same
Pressure SensordesignCapacitance type pressure transducerNo change - the same
NIBP Clinical testAASI/AAMI SP-10 :1992BS EN1060-4:2004, BHS:1993
NIBP MeasurementrangeBlood Pressure: 10-280mmHgPulse: 30-200bpmSYS: 40-270 mmHgDIA: 20-200 mmHgPulse: 30-240 bpm
AccuracyBP: +/- 3%, Pulse: +/- 5 %No change - the same
Pressurization SourceAutomatic internal air pumpNo change - the same
Cuff DesignWinding mechanism operated bygeared motorNo change - the same
Air Pressure ControlMethodRubber valve with ceramic valveRubber valve with electricalcontrol valve
Exhaust MethodAutomatic rapid exhaust byelectromagnetic valveNo change - the same
Display Type3-digit display by LEDNo change - the same
Cuff AttachmentMethodBy plastic hose connected to monitorNo change - the same
IHB (IrregularHeartbeats Detection)More than +/-25% to the meaninterval of all pulse intervalsNo change - the same
Operating50°F (10°C) to 104°F (40°C) with50°F (10°C) to 104°F (40°C)

Product Specification Comparison

{5}------------------------------------------------

Environment85% RH or less , no condensingwith 15 %RH to 85% RH, no condensing
Storage Environment14°F (-20°C) to 140°F (60°C) with95% RH or less, no condensing14°F (-20°C) to 140°F (60°C) with 10 -95% RH, nocondensing
Arm Size5.1 inches (13 cm) to 17.7 inches (45 cm)7.1 inches (18 cm) to 15.7 inches (40 cm)
Printer functionThermometer PrinterNo change - the same
External dimensions245 (W) x 322 (H) x 390 (D) mm241 (W) x 330 (H) x 390 (D) mm
WeightApprox. 9 kgApprox. 5.5 kg
ConnectivityWired - RS232C standard (TM-2655 Family)Wireless - Bluetooth 2.1 Standard (UA-767PBT)No change - the same
Voice/TalkingCapabilityTM-2655V enabledDisabled

Key Features Comparison

ParameterPredicate Devices(TM-2655 Family & UA-767PBTModified Devices(TM-2657 Family)
Field serviceNot allowedNo Change - the same
Automatic zero at "START"YesNo Change - the same
Manual zero adjustmentNot allowedNo Change - the same
CalibrationNot allowed in the fieldNo Change - the same
SterilizationNot neededNo Change - the same

8. Discussion of standards used in the design verification and design validation

We conducted design verification and design validation activities based on the comparison of the TM-2657 Family of digital blood pressure monitors with the predicate devices. Based on the changes, we conducted the appropriated test methodology and pass/fail criteria. After the tests were conducted, the test records were collected in the design history file (DHF).

A&D Medical follows FDA regulation and international standards in our medical device development and manufacturing processes. The following standards were used to demonstrate compliance to FDA recognized consensus standards for the TM-2657 Family of digital blood pressure monitors devices.

  • . ANSI/AAMI/IEC 80601-2-30:2009 Medical electrical equipment -- Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.
  • AAMI/ANSI 60601-1:2006 Medical electrical equipment. General requirements for . basic safety and essential performance
  • . AAMI/ANSI/IEC 60601-1-2:2007 Medical electrical equipment. General requirements for basic safety and essential performance. Collateral standard. Electromagnetic compatibility. Requirements and tests
  • . AAMI/ANSI/ISO 14971:2012 Medical devices. Application of risk management to medical devices

{6}------------------------------------------------

TM-2657 Family of digital blood pressure monitors met all applicable requirements of the standards. None of the test demonstrated that the TM-2657 Family of digital blood pressure monitors bring new issues of safety and effectiveness.

9. Evaluation Performance & Substantial Equivalence Conclusion:

TM-2657 family digital blood pressure monitors pass the clinical and non-clinical evaluation based on the followed table.

Non-clinical Test Summary:

  • Safety Tests: IEC 60601-1:2006
  • EMC Tests: AAMI/ANSI/IEC 60601-1-2:2007
  • Reliability Tests: IEC80601-2-30:2009+CI:2010 ●
  • Risk Assessmetn: ISO 14971:2012 ●
  • Bluetooth Tests: FCC 47 CFR Part 15.247 & EN 300 328 V1.8.1 (2012-06)
ItemStandard RequirementsStudy ResultsPass/Fail
Number of Subject85 or more85Pass
Number ofMeasurement3 times or more3 timesPass
Observer22Pass
Observer Difference+/- 4 mmHg or less+/- 4 mmHg lessPass
GenderMale 40% or more43Pass
Female 40% or more42Pass
AgeOlder than 50 : 50-75%44Pass
Circumference ofArm18 – 35 cm : 50-75%49Pass
BP RangeSYS<110mmHg at least 10%18Pass
SYS>160mmHg at least 10%24Pass
DIA<70mmHg at least 10%17Pass
DIA>100mmHg at least 10%29Pass
Measurement ResultAverage : +/-5mmHg or lessSYS -0.99mmHgDIA -1.09mmHgPass
S.D. : 8mmHg or lessSYS 7.62mmHgDIA 5.87mmHgPass

Clinical Test Summary based on BS EN 1060-4:2004 / ISO 141155-1 & -2:2003:

TM-2657 Family of digital blood pressure monitors Digital blood pressure monitors have the following similarities to the predicate devices, TM-2655 Family and UA-767PBT digital blood pressure monitors, which previously received the 510(k) clearance.

  • Same intended use. ●
  • Same oscillometric method to determine the blood pressure & pulse rate ●
  • Same inflation method - automatic internal pump
  • Same deflation method - standard exhaust valve
  • Same materials, no new materials used
  • Same manufacturing processes at the same facility
  • Reduced Indication of Use in arm circumference range. ● Predicated Devices (TM-2655 Family) Modified Devices (TM-2657 Family)

{7}------------------------------------------------

The A&D Medical LifeSource TM-2655,TM-2657, TM-2657P & TM-2657PRS
TM-2655P & TM-2655VP are designed to& TM-2657PBT are designed to
measure blood pressure (diastolic andmeasure blood pressure (diastolic and
systolic) and pulse rate in adult patients withsystolic) and pulse rate in adult patients
arm circumference range between 5.1 incheswith arm circumference range between
(13 cm) to 17.7 (45 cm)7.1 inches (18 cm) to 15.7 (40 cm)

As a conclusion, TM-2657 Family of digital blood pressure monitors as described in its labeling and comparison analysis, have not changed as a result of the modifications. The fundamental scientific technology of the modified devices has not changed, either. There is no significant difference that affects the safety or effectiveness of the modified device as compared to the predicate devices.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).