(106 days)
The HET Bipolar System is intended to be used for the treatment of symptomatic grade I and grade II internal hemorrhoids.
The HET™ Bipolar Forceps is a sterile, single-use bipolar forceps having a tapered tubular configuration. The device is connected via an integrated bipolar cable to the bipolar output of an electrosurgical generator. The accessory monitor displays the temperature at the forceps-tissue interface and provides power for a temperature sensor and an LED light source mounted on the disposable forceps. The HET™ Bipolar Electrocautery Forceps and Monitor do not generate RF energy themselves and may be used with any Bipolar Electrosurgical generator in the coagulation power setting with an output power set at 10 W and a maximum voltage of 1250V.
This document is a 510(k) Pre-Market Notification for the HET™ Bipolar Electrocautery Forceps and Monitor. The submission states that the device is identical to its predicate device (K140422) in terms of design, materials, software, and treatment mechanism. The purpose of this 510(k) is to limit the indications for use to symptomatic grade I and grade II internal hemorrhoids and clarify that the action is tissue ligation. Therefore, the performance data presented refers to the predicate device.
Here's an analysis based on the provided document:
Acceptance Criteria and Device Performance
Since the subject device is identical to the predicate device and the 510(k) is primarily for clarifying indications and the mechanism of action (tissue ligation), there are no new acceptance criteria or device performance results reported for the subject device in this document. The performance data refers to the predicate device (K140422).
Based on the information for the predicate device (K140422) as referenced in this submission:
| Acceptance Criteria (Implied) | Reported Device Performance (for Predicate K140422) |
|---|---|
| Acceptable Zone of Thermal Injury | Performance testing evaluated the zone of thermal injury for the predicate device (K140422) when used with eight FDA-cleared, commercially available bipolar electrosurgical generators. The testing "confirmed that the treatment time was acceptable to allow safe treatment." |
| Acceptable Treatment Time for Safe Treatment | The testing confirmed that the treatment time was acceptable to allow safe treatment. |
| Conformance to Electromedical Safety Standards | Additional testing was provided to support conformance to the latest revision of IEC 60601-1-2:2014. |
| Biocompatibility in accordance with ISO 10993 recommendations | Biocompatibility testing was previously conducted and reported in 510(k) K121085, including cytotoxicity, intracutaneous irritation, sensitization, acute systemic toxicity, and hemocompatibility, performed on the final, finished, gamma-irradiation sterilized device and in accordance with GLP. |
| Sterility Assurance Level (SAL) | The HET™ Bipolar Forceps are sterilized with EtO to a sterility level of 10-6. |
Study Details for Predicate Device (K140422) as referenced:
-
Sample size used for the test set and the data provenance:
- Sample size: Not explicitly stated in this document. The document mentions "Performance testing was conducted" to characterize the device's performance when used with "eight FDA cleared, commercially available, bipolar electrosurgical generators." The specific number of cases, tissue samples, or treatments is not provided.
- Data provenance: Not specified in this document. It's likely from laboratory or bench testing given the nature of evaluating thermal injury and treatment time. It's not stated whether it's retrospective or prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as the reported performance testing (zone of thermal injury, treatment time) appears to be objective, measurable parameters typically assessed through engineering and scientific methods rather than expert interpretation of images or clinical outcomes in the context of ground truth establishment.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. The performance tests described (thermal injury, treatment time, electrical safety, biocompatibility, sterilization) are objective measurements and engineering assessments, not requiring adjudication in the context of expert case review.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC or AI-assisted study was performed or is relevant to this device. This is a medical device for electrosurgical treatment, not an AI diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is not an AI algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For performance testing related to thermal injury and treatment time, the "ground truth" would be objective measurements (e.g., histological analysis of tissue for thermal damage depth, precise timing measurements).
- For biocompatibility, the ground truth is established by standardized biological assays (e.g., cell viability for cytotoxicity, skin reaction for irritation/sensitization).
- For sterilization, the ground truth is the measured Sterility Assurance Level (SAL).
- For electrical safety, the ground truth is conformance to the specified standard (IEC 60601-1-2:2014).
-
The sample size for the training set:
- Not applicable. This is not an AI or machine learning device requiring a training set.
-
How the ground truth for the training set was established:
- Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue.
May 7, 2020
Covidien, LLC Liron Bar Yaakov Senior Manager Regulatory Affairs 6135 Gunbarrel Avenue Boulder, Colorado 80301
Re: K200146
Trade/Device Name: HET Bipolar Electrocautery Forceps, HET Bipolar Electrocautery Monitor Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 2, 2020 Received: April 3, 2020
Dear Liron Yaakov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K200146
Device Name
Device Name
HET™ Bipolar Electrocautery Forceps and Monitor
Indications for Use (Describe)
The HET Bipolar System is intended to be used for the treatment of symptomatic grade I and grade II internal hemorrhoids.
| Type of Use (Select one or both, as applicable) | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor and a person is not required to respond to a collection of
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
510(k) Summary - K200146
Date Prepared: May 5th, 2020
Submitter's Name and Address:
Covidien Ilc 15 Hampshire Street Mansfield, MA 02048
Contact Person:
Liron Bar Yaakov Senior Manager, regulatory Affairs Phone: (720) 422-9135 Fax: (408) 328-7310
Name of Device:
Proprietary Name: HET™ Bipolar Electrocautery Forceps and Monitor Common/Usual Name: Electrosurgical Instrument Classification Panel: General & Plastic Surgery Device Regulation: 21 CFR 878.4400, Class II Product Code: GEI
Establishment Registration Number, Owner/Operator Number:
Establishment Registration Number: 3004904811 Owner/Operator Number: 1282497 The HET™ Bipolar System is manufactured by a contract manufacturer, Contract Medical Manufacturing, LLC (CMM). The CMM facility is located at: 1 Jacks Hill Road - Unit 3e Oxford, CT 06478 The CMM facility has been assigned establishment registration number 3006183537 and Owner/Operator number 10042275.
Predicate Device(s):
K140422 HET™ Bipolar Electrocautery Forceps and Monitor
Device Description:
The HET™ Bipolar Forceps is a sterile, single-use bipolar forceps having a tapered tubular configuration. The device is connected via an integrated bipolar cable to the bipolar output of an electrosurgical generator. The accessory monitor displays the temperature at the forceps-tissue interface and provides power for a temperature sensor and an LED light source mounted on the disposable forceps. The HET™ Bipolar Electrocautery Forceps and Monitor do not generate RF energy themselves and may be used with any Bipolar Electrosurgical generator in the coagulation power setting with an output power set at 10 W and a maximum voltage of 1250V.
Indications for Use:
The HET Bipolar System is intended to be used for the treatment of symptomatic grade I and grade II internal hemorrhoids.
{4}------------------------------------------------
Technological Characteristics of the Device Compared to Predicate Device
The technological characteristics of the device are unchanged (i.e. are exactly the same) as compared to the characteristics of the predicate device that was cleared under 510(k) K140422. The purpose of this submission is to limit the indication within the previously cleared indications to treatment of symptomatic grade I and grade II internal hemorrhoids and to clarify that the action performed by the HET Bipolar Electrocautery Forceps and Monitor is considered as tissue ligation. The materials, software, and treatment mechanism (i.e., bipolar energy used to produce heat) are unchanged. Both the predicate device and the subject device use thermal energy delivered during forceps tissue clamping/compression to achieve the treatment effect, on the same target tissue and patient population. Both the subject device and the predicate device are indicated for the treatment of internal hemorrhoids.
As the subject device is identical to the predicate device, biocompatibility studies, software validations, and electrical safety studies previously reported and cleared for the predicate device remain applicable to the subject device.
Performance Data
Performance testing was conducted and reported in the 510(k) Notice for the predicate device K140422. Testing was conducted to characterize the device's performance when used in conjunction with eight FDA cleared, commercially available, bipolar electrosurgical generators. The studies evaluated to zone of thermal injury and confirmed that the treatment time was acceptable to allow safe treatment. Because the two devices are identical, the labeling updates provide clarifications and consistency with the revised indication for use and do not change the intended use of the device, user population or patient population. The information reported for the predicate device is applicable to the subject device. Additional testing was provided to support conformance to the latest revision of IEC 60601-1-2:2014.
Biocompatibility
Testing was previously conducted, and reported in the 510(k) K121085, to evaluate the device's biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiation sterilized device and in accordance with the GLP requirements: cytotoxicity, intracutaneous irritation, sensitization, acute systemic toxicity and hemocompatibility.
Sterilization
The HET™ Bipolar Forceps are sterilized with EtO to a sterility level of 10-6.
Conclusion
The subject device has the same intended use and the identical design as the legally marketed predicate device. The labeling changes made to the subject device do not raise different kinds of safety and effectiveness questions. Therefore, Covidien llc considers the subject device, the HET™ Bipolar Electrocautery Forceps and Monitor, to be substantially equivalent to the legally marketed predicate device, the HET™ Bipolar Electrocautery Forceps and Monitor (K140422).
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.