K Number
K140422

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-06-05

(107 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HET™ Bipolar Electrocautery Forceps and Monitor is intended to be used for grasping, manipulating and coagulating soft tissue during general surgery.

The HET™ Bipolar Electrocautery Forceps and Monitor may be used for the treatment of symptomatic Grade I and Grade II internal hemorrhoids.

Device Description

The HET™ Bipolar Forceps is a sterile, single-use bipolar forceps having a tapered tubular configuration. The device is connected via an integrated bipolar cable to the bipolar output of an electrosurgical generator.

The accessory monitor provides power for a temperature sensor and an LED light source mounted on the disposable forceps. The HET Systems HET™ Bipolar Electrocautery Forceps and Monitor may be used with any Bipolar Electrosurgical generator in the coagulation mode with an output power set at 10 W and a maximum voltage of 1250v.

The monitor displays the temperature at the forceps-tissue interface. The HET" Monitor does not generate RF energy.

AI/ML Overview

The provided document describes the HET™ Bipolar Electrocautery Forceps and Monitor (K140422). However, it is a 510(k) summary for a submission aiming to demonstrate substantial equivalence to a predicate device (K121085). This document primarily discusses performance testing related to equivalence with the predicate and does not detail specific acceptance criteria or an independent study to "prove the device meets acceptance criteria" for a novel device approval.

Instead, the submission focuses on demonstrating that the new device, when used with various electrosurgical generators, performs comparably to the predicate device.

Here's an analysis based on the information provided, highlighting what is (and isn't) present:

1. Table of acceptance criteria and reported device performance:

The document does not explicitly state formal acceptance criteria with numerical targets. Instead, it refers to performance testing done to show equivalence to the predicate device. The key performance evaluations were:

Performance AspectReported Device Performance
Zone of thermal injuryStudies were conducted in an ex vivo tissue model to evaluate the zone of thermal injury. The implication is that the results were comparable to the predicate device to support substantial equivalence. (No specific numerical values or acceptance ranges are provided).
Treatment timeStudies in an ex vivo tissue model confirmed that the treatment time was "acceptable to allow safe treatment." (No specific numerical values or acceptance ranges are provided).
BiocompatibilityPreviously conducted and reported in the 510(k) Notice for the predicate (K121085). Because the devices are the same, the information is applicable. Included cytotoxicity, intracutaneous irritation, sensitization, acute systemic toxicity, and hemocompatibility, all conducted in accordance with ISO 10993 and GLP requirements. The implication is that these met acceptance criteria at the time of the predicate's clearance.
Software ValidationPreviously conducted and reported for the predicate (K121085). Applicable to the current device due to identicality.
Electrical SafetyPreviously conducted and reported for the predicate (K121085). Applicable to the current device due to identicality.
Clinical Performance (Predicate Device)Human clinical evaluations for the predicate device (K121085) demonstrated outcomes substantially equivalent to clinical outcomes reported in the literature for an additional predicate device. This is referenced as supporting the predicate device's clinical performance, which the current device is equivalent to.

Since this is a 510(k) for substantial equivalence of an unchanged device but with an expanded compatibility claim (use with any bipolar electrosurgical generator), the "study that proves the device meets the acceptance criteria" is implicitly the suite of tests performed to show that the device performs equivalently to the already cleared predicate device across various compatible generators.


2. Sample size used for the test set and the data provenance:

  • Test set sample size: Not specified. The document states "a series of studies (in an ex vivo tissue model)" but does not provide details on the number of samples or repetitions.
  • Data provenance: Ex vivo tissue model. No country of origin is specified for the ex vivo studies. It's a combination of "retrospective" (referencing prior predicate studies) and "prospective" (new ex vivo studies specifically for this submission to test compatibility with various generators). The human clinical data mentioned refers to the predicate device and is therefore also retrospective for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable / not specified. The studies cited are primarily engineering/performance tests on ex vivo tissue models (thermal injury, treatment time) and biocompatibility tests. These types of tests do not typically involve human experts establishing a "ground truth" in the way clinical studies or diagnostic AI studies do. The clinical outcomes mentioned refer to the predicate device and are not detailed here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable / not specified. Adjudication methods are relevant for subjective assessments, typically in clinical or diagnostic studies. The performance tests described (thermal injury, treatment time, biocompatibility, electrical safety) are objective measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a surgical tool, not a diagnostic AI system with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a medical device, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the ex vivo performance testing (thermal injury, treatment time): The "ground truth" is based on direct physical measurements (e.g., assessing tissue damage, timing procedures).
  • For biocompatibility, software, and electrical safety: These are based on established standards (ISO 10993 for biocompatibility) and engineering validation.
  • For clinical performance (referenced for the predicate device): The ground truth was clinical outcomes, compared to literature.

8. The sample size for the training set:

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. As above, this is not an AI/machine learning device.

{0}------------------------------------------------

JUN 0 5 2014

K140422

Page 1 of 3

510(k) SUMMARY (Per 21 CFR 807.92)

General Company Information

Name:HET Systems, LLC
Contact:Howard SchrayerRegulatory Affairs Consultant
Address:200 Middlesex Turnpike - Suite 304Iselin, NJ 08830
Telephone:(609) 924 - 9510
Fax:(732) 683 - 9476
Date PreparedApril 29, 2014
General Device Information
Product Name:HETTM Bipolar Electrocautery Forceps and Monitor
Classification:"Electrosurgical cutting and coagulation device and accessories"Product code: GEI - Class II

Predicate Device

HET Systems, LLC - HET™ Bipolar Electrocautery Forceps and Monitor - 510(k) Number K121085

Description

The HET™ Bipolar Forceps is a sterile, single-use bipolar forceps having a tapered tubular configuration. The device is connected via an integrated bipolar cable to the bipolar output of an electrosurgical generator.

The accessory monitor provides power for a temperature sensor and an LED light source mounted on the disposable forceps. The HET Systems HET™ Bipolar Electrocautery Forceps and Monitor may be used with any Bipolar Electrosurgical generator in the coagulation mode with an output power set at 10 W and a maximum voltage of 1250v.

The monitor displays the temperature at the forceps-tissue interface. The HET" Monitor does not generate RF energy.

{1}------------------------------------------------

Intended Use (Indications)

The HET™ Bipolar Electrocautery Forceps and Monitor is intended to be used for grasping, manipulating and coagulating soft tissue during general surgery.

The HET™ Bipolar Electrocautery Forceps and Monitor may be used for the treatment of symptomatic Grade I and Grade II internal hemorrhoids.

Substantial Equivalence

This submission supports the position that the HET™ Bipolar Electrocautery Forceps and Monitor is substantially equivalent to a previously cleared device, specifically, the HET Bipolar Electrocautery Forceps and Monitor (K121085).

The technological characteristics of the device are unchanged (exactly the same as) the characteristics of the predicate HET Systems device that was cleared under 510(k) K121085. The purpose of this submission is to demonstrate that the HET Bipolar Electrocautery Forceps and Monitor may be used with any bipolar electrosurgical generator that can be operated at 10W output power and a maximum voltage of 1250. The materials, software, and treatment mechanism (i.e., bipolar energy used to produce heat) are unchanged. Both the predicate system and the subject device use thermal energy to achieve the treatment effect (tissue coagulation). The predicate device is indicated for the treatment of internal hemorrhoids.

The 510(k) Notice contains reports of a series of studies (in an ex vivo tissue model model) that were conducted to evaluate the performance and thermal effects of the device as compared to the predicate device.

Biocompatibility studies, software validations, and electrical safety studies were conducted and reported in the 510(k) Notice for the predicate HET Systems device. Because the two devices are the same, the information reported for the predicate HET device is applicable.

Performance Testing

Testing was conducted to characterize the device's performance when used in conjunction with eight FDA cleared, commercially available, bipolar electrosurgical generators. The studies evaluated to zone of thermal injury and confirmed that the treatment time was acceptable to allow safe treatment.

{2}------------------------------------------------

Biocompatibility Testing

Testing was previously conducted, and reported in the 510(k) Notice for the predicate, to evaluate the device's biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiation sterilized device and in accordance with the GLP requirements: cytotoxicity, intracutaneous irritation, sensitization, acute systemic toxicity and hemocompatibility.

Clinical Testing

The 510(k) Notice for the predicate also contains reports of human clinical evaluations that were conducted to evaluate the clinical performance of the device. These studies demonstrated outcomes substantially equivalent to clinical outcomes reported in the literature for an additional predicate device.

Conclusions

HET Systems, LLC believes that the information provided establishes that similar leqally marketed devices have been used for the same clinical applications as the indications for the HET™ Bipolar Electrocautery Forceps and Monitor. The materials from which the HET Systems device is fabricated have an established history of use in medical applications; and devices produced by HET Systems have been tested in accordance with applicable guidelines and standards.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized lines representing the wings and body.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 5, 2014

HET Systems, LLC % Mr. Howard L. Schrayer Regulatory Affairs Consultant 200 Middlesex Turnpike, Suite 304 Iselin, New Jersey 08830

Re: K140422

Trade/Device Name: HET" Bipolar Electrocautery Forceps and Monitor Regulation Number: 21 CFR 878.4000 Regulation Name: Electrosurgical, cutting & coagulation device & accessories Regulatory Class: Class II Product Code: GEI Dated: April 30, 2014 Received: May 1, 2014

Dear Mr. Schrayer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Mr. Howard L. Schrayer

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.

Sincerely yours,

David Krause -S

l'or

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K140422

INDICATIONS FOR USE

510(k) Number (if known): K140422

Device Name: HET™ Bipolar Electrocautery Forceps and Monitor

Indications For Use:

The HET™ Bipolar Electrocautery Forceps and Monitor is intended to be used for grasping, manipulating and coagulating soft tissue during general surgery.

The HET™ Bipolar Electrocautery Forceps and Monitor may be used for the treatment of symptomatic Grade I and Grade II internal hemorrhoids.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

lan P. Broverman -S 5. cn=lan P. Broverman -S

Digitally signed by lan P. Broverman -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=200079294 Date: 2014.06.04 15:38:11 -04'00'

5

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.