K Number
K200121

Validate with FDA (Live)

Date Cleared
2020-04-18

(88 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Route 92 Medical 088 Access System, 110 cm, is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

Device Description

The Route 92 Medical 088 Access System, 110 cm, is comprised of a Support Catheter and a Delivery Catheter. The distal portion of the Support Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, single-lumen variable stiffness catheter. Both catheters are hydrophilically coated. The Route 92 Medical Access System, 110 cm, is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Route 92 Medical 088 Access System, 110 cm:

1. Table of Acceptance Criteria and Reported Device Performance:

The document outlines performance criteria in two main sections: Biocompatibility Testing and Performance Testing. For both, the acceptance criterion is implicitly "PASS" meaning the device met pre-determined criteria for each test. Specific quantitative acceptance criteria are not detailed in this summary.

Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
BiocompatibilityCytotoxicity – ISO MEM ElutionNon-cytotoxicNon-cytotoxic
Sensitization – ISO Guinea Pig MaximizationNo sensitization responseNo sensitization response
Irritation – ISO Intracutaneous ReactivityRequirements metRequirements met
Acute Systemic Toxicity – ISO Acute Systemic InjectionRequirements metRequirements met
Pyrogen – Material Mediated PyrogenNon-pyrogenicNon-pyrogenic
Hemocompatibility – Complement ActivationNo adverse effects expectedNo adverse effects expected
Hemocompatibility – Partial Thromboplastin TimeNon-activatorNon-activator
Hemocompatibility – ASTM HemolysisNon-hemolyticNon-hemolytic
Hemocompatibility – ThromboresistanceSimilar to control devicesSimilar to control devices
Performance TestingDimensional VerificationConformance to specificationsPASS
Luer IntegrityMet pre-determined criteriaPASS
RHV SealingMet pre-determined criteriaPASS
Tensile StrengthMet pre-determined criteriaPASS
Kink ResistanceMet pre-determined criteriaPASS
Torsion ResistanceMet pre-determined criteriaPASS
Tip FlexibilityMet pre-determined criteriaPASS
Air LeakageMet pre-determined criteriaPASS
Liquid Leakage / Static BurstMet pre-determined criteriaPASS
Dynamic BurstMaintained mechanical integrityPASS
Hydrophilic Coating IntegrityMet pre-determined criteriaPASS
Particulate RecoveryMet pre-determined criteriaPASS
Simulated Use TestingMet pre-determined criteriaPASS
RadiopacityMet pre-determined criteriaPASS
Contrast DeliveryMet pre-determined criteriaPASS
In Vivo ThromboresistanceMet pre-determined criteriaPASS
Packaging IntegrityMet pre-determined criteriaPASS

2. Sample Sizes Used for the Test Set and Data Provenance:

The document states "All samples met the pre-determined acceptance criteria" for various performance tests, but does not specify the exact sample size for any of the individual tests. The provenance of the data (e.g., country of origin, retrospective or prospective) is also not provided. Many tests mention adherence to ISO standards, implying controlled laboratory conditions, but details about the test setting are not available. The Biocompatibility and Performance tests are "Non-Clinical Testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided as the tests conducted are primarily non-clinical, mechanical, and material-based, not involving human interpretation of medical images or patient outcomes that would typically require expert ground truth establishment in that manner. For the "Simulated Use Testing," "Radiopacity," and "Contrast Delivery," and "In Vivo Thromboresistance" which involve models and animals, the experts involved in conducting and evaluating these tests are not specified.

4. Adjudication Method for the Test Set:

An adjudication method is not applicable and therefore not mentioned, as the tests are primarily objective physical and chemical characterizations, rather than assessments requiring human judgment or consensus from multiple reviewers for a "ground truth" determination.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not conducted. This document describes a 510(k) submission for a medical device (catheter system), not an AI or imaging diagnostic tool. Therefore, human reader performance with or without AI assistance is not relevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm or software.

7. The Type of Ground Truth Used:

For the non-clinical tests, the "ground truth" is defined by the specifications and acceptable ranges established by industry standards (e.g., ISO) and the manufacturer's pre-determined acceptance criteria. For "Simulated Use Testing," "Radiopacity," "Contrast Delivery," and "In Vivo Thromboresistance," the ground truth would be based on direct observation and measurement within the respective models (neurovascular model, animal model), demonstrating the device's functional integrity and performance as intended. No human pathology, outcomes data, or expert consensus in the medical diagnostic sense is mentioned as ground truth.

8. The Sample Size for the Training Set:

This information is not applicable as this is a physical medical device, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as this is a physical medical device, not an AI model requiring a training set and associated ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 18, 2020

Route 92 Medical Inc. Kathy Tansey Senior Director of Regulatory Affairs and Quality Assurance 1700 S. El Camino Real, Suite 206 San Mateo, California 94402

Re: K200121

Trade/Device Name: Route 92 Medical 088 Access System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: January 17, 2020 Received: January 21, 2020

Dear Kathy Tansey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200121

Device Name Route 92 Medical 088 Access System

Indications for Use (Describe)

The Route 92 Medical 088 Access System, 110 cm, is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

510(k) Summary

Sponsor:Route 92 Medical1700 S. El Camino Real, Suite 206San Mateo, CA 94022Phone: 650-581-1179
Contact:Kathy Tansey
Date Prepared:January 17, 2019
Device Name:Route 92 Medical 088 Access System
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter (Product Code OJP, 21 CFR870.1250)
Primary Predicate Device:Secondary Predicate Device:Route 92 Medical Access System, K180201Imperative Care Inc. EagleRay Long Sheath, K180169

Device Description

The Route 92 Medical 088 Access System, 110 cm, is comprised of a Support Catheter and a Delivery Catheter. The distal portion of the Support Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, single-lumen variable stiffness catheter. Both catheters are hydrophilically coated. The Route 92 Medical Access System, 110 cm, is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

Indications for Use

The Route 92 Medical 088 Access System, 110 cm, is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

Comparison to the Predicate Device

The method of action, design, and materials of the Route 92 Medical 088 Access System, 110 cm, are equivalent to the Predicate Devices as shown in the following table.

{4}------------------------------------------------

AttributePrimary PredicateRoute 92 Medical AccessSystem (K180201)Secondary PredicateImperative Care Inc.EagleRay Long Sheath(K180169)SubjectRoute 92 Medical 088Access System, 110 cm
Indicationsfor UseThe Route 92 MedicalAccess System is indicatedfor use with compatibleguide catheters in facilitatingthe insertion and guidance ofmicrocatheters into aselected blood vessel in theneurovascular system.The EagleRay LongSheath and EagleRayAccess Catheter areindicated for theintroduction ofinterventional devices intothe peripheral, coronary,and neuro vasculature.The Route 92 Medical 088Access System, 110 cm, isindicated for use withcompatible guide cathetersin facilitating the insertionand guidance ofmicrocatheters into aselected blood vessel in theneurovascular system.
DeviceDescriptionSterile, single-use, variablestiffness, coil-reinforcedcatheter with proximalcontrol wireSterile, single-use,variable stiffness, wire-reinforced catheterSterile, single-use, variablestiffness, coil-reinforcedcatheter with proximalcontrol wire
TargetedpopulationPatients requiring use of amicrocatheter in theneurovascular systemPatients requiring use of amicrocatheter in theperipheral, coronary orneurovascular systemsPatients requiring use of amicrocatheter in theneurovascular system
UserPhysicians trained inneurovascular interventionaltechniquesPhysicians trained ininterventional techniquesPhysicians trained inneurovascular interventionaltechniques
AnatomicalSitesNeurovasculature onlyPeripheral, coronary orneurovascular systemsNeurovasculature only
MaterialsPolymers and metalscommonly used in themanufacture of medicaldevicesPolymers and metalscommonly used in themanufacture of medicaldevicesPolymers and metalscommonly used in themanufacture of medicaldevices
Support Catheter
InnerDiameter0.070".088"0.088"
OuterDiameterDistal: 0.082"Proximal: 0.087"Distal: 0.106"Proximal: 0.108"Distal: 0.101"Proximal: 0.105"
Length136cm80cm -110cm110cm
InnerDiameter0.019"N/A0.019"
OuterDiameter0.062"N/ADistal: 0.080"Proximal: 0.062"
Length143cmN/A151cm

{5}------------------------------------------------

Non-Clinical Testing

Biocompatibility Testing

The Route 92 Medical 088 Access System, 110 cm, is constructed using materials that are commonly used in the medical device industry. All patient contacting components have been evaluated for biocompatibility in accordance with ISO 10993-1 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. The Route 92 Medical Access System is classified per ISO 10993-1 as externally communicating with limited circulating blood contact (< 24 hours). A summary of the biocompatibility testing is provided below.

TestConclusions
Cytotoxicity – ISO MEM ElutionThe test article is non-cytotoxic.
Sensitization – ISO Guinea PigMaximization Sensitization Test(Normal Saline and Sesame Oil)The test article did not elicit asensitization response.
Irritation – ISO IntracutaneousReactivity (Normal Saline andSesame Oil)Requirements of the ISOintracutaneous reactivity test have beenmet for the test article.
Acute Systemic Toxicity – ISOAcute Systemic Injection (NormalSaline and Sesame Oil)Requirements of the ISO acutesystemic injection test have been metfor the test article.
Pyrogen – Material MediatedPyrogen (Normal Saline)The test article is non-pyrogenic.
Hemocompatibility – ComplementActivation (SC5b-9)The test article would not be expectedto result in adverse effects in vivo.
Hemocompatibility – PartialThromboplastin TimeThe test article is considered to be anon-activator of the intrinsiccoagulation pathway.
Hemocompatibility – ASTMHemolysisThe test article is considered non-hemolytic.
Hemocompatibility –ThromboresistanceThe test articles have similarthromboresistance characteristics asthe control devices.

{6}------------------------------------------------

Performance Testing

The successful completion of the performance testing listed in the following table demonstrates that the Route 92 Medical 088 Access System, 110 cm, is suitable for its intended use.

TestTest MethodResults
DimensionalVerificationDevice dimensions were measuredto confirm conformance to thespecificationsPASSAll samples met the pre-determinedacceptance criteria
Luer IntegrityTested per ISO 80369-7:2016PASSAll samples met the pre-determinedacceptance criteria
RHV SealingRHV sealing around the cathetershafts was testedPASSAll samples met the pre-determinedacceptance criteria
Tensile StrengthThe tensile strength of the cathetersections and bonds was testedPASSAll samples met the pre-determinedacceptance criteria
Kink ResistanceTest specimen segments wereformed into a defined benddiameter to evaluate kinkresistancePASSAll samples met the pre-determinedacceptance criteria
Torsion ResistanceThe test specimens were rotated toevaluate integrity after rotationPASSAll samples met the pre-determinedacceptance criteria
Tip FlexibilityTest specimens were tested for tipflexibilityPASSAll samples met the pre-determinedacceptance criteria
Air LeakageTested per ISO 10555-1:2013Annex D.PASSAll samples met the pre-determinedacceptance criteria
Liquid Leakage /Static BurstTested per ISO 10555-1:2013Annex C.PASSAll samples met the pre-determinedacceptance criteria
Dynamic BurstMechanical integrity wasmaintained up to the specifiedpressuresPASSAll samples met the pre-determinedacceptance criteria
TestTest MethodResults
HydrophilicCoating IntegrityThe integrity of the hydrophiliccoating was evaluated aftermultiple insertion and withdrawalcycles.PASSAll samples met the pre-determinedacceptance criteria
ParticulateRecoveryAfter multiple insertion andwithdrawal cycles, the effluentwater rinsed and flushed from thedevices and model was tested perUSP <788>.PASSAll samples met the pre-determinedacceptance criteria
Simulated UseTestingDeliverability and compatibilitywith accessory devices wasevaluated in a neurovascular modelPASSAll samples met the pre-determinedacceptance criteria
RadiopacityRadiopacity of the device wasevaluated in an animal modelunder fluoroscopyPASSAll samples met the pre-determinedacceptance criteria
Contrast DeliveryContrast delivery using thedevice was evaluated in ananimal model under fluoroscopyPASSAll samples met the pre-determinedacceptance criteria
In VivoThromboresistanceThromboresistance was evaluatedunder clinical conditions in theswine modelPASSAll samples met the pre-determinedacceptance criteria
PackagingIntegrityISO 11607-1 Part 1ISO 11607-2 Part 2PASSAll samples met the pre-determinedacceptance criteria

{7}------------------------------------------------

Substantial Equivalence

The Route 92 Medical 088 Access System, 110 cm, has the same intended use, the same technological characteristics and same method of action as the Predicate Devices. The successful completion of biocompatibility testing and performance testing demonstrates that the Route 92 Medical 088 Access System, 110 cm, is substantially equivalent to the Predicate Devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).