(29 days)
The Affinity NT Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
The Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with Trillium™ Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity NT™ Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
The Affinity NT™ Integrated Uncoated CVR/Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
Not Found
This appears to be a 510(k) clearance letter and related "Indications for Use" forms for various models of Medtronic Affinity NT Oxygenators. This type of document from the FDA does not contain the information requested about acceptance criteria and device performance studies (specifically, those related to AI/algorithm or human reader performance).
The purpose of a 510(k) clearance is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to present detailed performance data in the context of controlled clinical trials or AI validation studies with specific acceptance criteria and ground truth methodologies.
Therefore, I cannot fulfill the request using only the provided text. The document is about:
- Device Identification: Medtronic's Affinity NT Oxygenator series.
- Regulatory Clearance: Confirmation of 510(k) clearance (K191029).
- Indications for Use: The medical use (extracorporeal perfusion circuit for oxygenation, CO2 removal, and temperature control during CPB procedures up to 6 hours).
- Regulatory Class: Class II device.
- Prescription Use: All listed devices are prescription-use only.
- General Controls: Reminders about compliance with FDA general controls (registration, listing, GMP, labeling, adverse event reporting, etc.).
There is no information in this document about:
- Acceptance criteria table and reported device performance related to a study.
- Sample size and data provenance for a test set.
- Number or qualifications of experts for ground truth establishment.
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study, or effect size with AI assistance.
- Standalone (algorithm-only) performance.
- Type of ground truth used (e.g., expert consensus, pathology, outcomes).
- Sample size for training set.
- How ground truth for the training set was established.
To provide the requested information, a different type of document would be needed, such as a detailed study report, a clinical trial summary (often found in PMA applications or supplement data, but not typically in a 510(k) summary for a Class II device like this oxygenator), or a specific AI/software validation report.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
May 17, 2019
Medtronic, Inc. Sammie Joseph-Fredericks Sr. Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428
Re: K191029
Trade/Device Name: Affinity NT™ Oxygenator (Model 511), Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with Trillium™ Biosurface (Model 511T), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541-R), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T-R), Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator (Model 541TT), Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator (Model 541TT-R), Affinity NT™ Oxygenator with Cortiva™ BioActive Surface (Model CB511), Affinity NT™ Integrated Uncoated CVR/Oxygenator with Cortiva™ BioActive Surface (Model CB541) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: April 17, 2019 Received: April 18, 2019
Dear Sammie Joseph-Fredericks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Nicole Ibrahim Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K191029
Device Name Affinity NT™ Oxygenator (Model 511)
Indications for Use (Describe)
The Affinity NT Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with Trillium™ Biosurface (Model 511T)
Indications for Use (Describe)
The Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with Trillium™ Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541)
Indications for Use (Describe)
The Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable)
| ✖ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ❍ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541-R)
Indications for Use (Describe)
The Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T)
Indications for Use (Describe)
The Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T-R)
Indications for Use (Describe)
The Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known)
K191029
Device Name Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator (Model 541TT)
Indications for Use (Describe)
The Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
510(k) Number (if known)
K191029
Device Name Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator (Model 541TT-R)
Indications for Use (Describe)
The Affinity® NT Integrated Trillium™ CVR/Membrane Oxygenator is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable)
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
510(k) Number (if known)
K191029
Device Name Affinity NTTM Oxygenator with CortivaTM BioActive Surface (Model CB511)
Indications for Use (Describe)
The Affinity NT™ Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable):
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{11}------------------------------------------------
510(k) Number (if known)
K191029
Device Name
Affinity NTTM Integrated Uncoated CVR/Oxygenator with Cortiva™ BioActive Surface (Model CB541)
Indications for Use (Describe)
The Affinity NT™ Integrated Uncoated CVR/Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood, cool or warm the blood and oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
Type of Use (Select one or both, as applicable):
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
§ 870.4350 Cardiopulmonary bypass oxygenator.
(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”