(206 days)
No
The summary describes a system of dental implants and abutments, including some digitally designed components, but there is no mention of AI or ML being used in the design, manufacturing, or function of the device. The performance studies focus on mechanical and biological safety testing.
Yes.
The device is intended to restore chewing function by providing support for prosthetic devices such as artificial teeth, which is a therapeutic purpose.
No
This device, the NUVO IF Implant System, is intended to provide support for prosthetic devices (like artificial teeth) to restore chewing function. It is a medical device used for treatment (restoration of function), not for diagnosing a condition.
No
The device description explicitly details physical components such as implants, cover screws, and abutments, which are hardware.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Definition of IVD: In Vitro Diagnostics are medical devices intended for use in vitro for the examination of specimens, including blood, tissue, and urine, derived from the human body, to provide information for diagnostic, monitoring, or compatibility purposes.
- Intended Use: The intended use of the NUVO IF Implant System and its associated components is to be surgically placed in the maxilla or mandible to support prosthetic devices for restoring chewing function. This is a surgical and prosthetic device, not a device used for examining specimens outside the body.
- Device Description: The device description details implants, cover screws, and abutments made of titanium and other materials, designed for surgical implantation and prosthetic support. There is no mention of components or functions related to analyzing biological specimens.
- Lack of IVD-related information: The document does not mention any of the typical characteristics of an IVD, such as:
- Analysis of biological samples (blood, tissue, etc.)
- Diagnostic, monitoring, or compatibility testing
- Reagents or assays
- Measurement of biomarkers or other biological parameters
The device described is a dental implant system, which is a type of surgical and prosthetic medical device.
N/A
Intended Use / Indications for Use
Indications for Use for NUVO IF Implants and conventional abutments:
The NUVO IF Implant System is intended to be surgically placed in the maxilla or mandible to provide support for prosthetic devices such as artificial teeth in order to restore chewing function. It may be used with single-stage or twostage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
The Titanium Temporary Abutment is indicated to provide temporary support for prosthesis structure for up to 6 months.
The Attachment Equator and Attachment Removable Prosthesis are indicated for the attachment of full or partial dentures to NUVO implants.
Indications for Use for Rotational and Anti-rotational Titanium Bases abutments:
Titanium Base Abutment is a titanium base placed onto dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement or screw-retained single or multi-unit restorations or screw-retained single restorations.
All digitally designed copings and/or crowns to be used with the Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The Rotational Titanium Base is indicated for cement or screw-retained multi-unit restorations.
Indications for Use for CARES® Abutment IF:
The CARES® Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations (copings or crowns). It is indicated for cement-retained single restorations. All digitally designed abutments for use with the CARES® Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.
Product codes (comma separated list FDA assigned to the subject device)
DZE, NHA
Device Description
Implants
- . Intended for single use;
- Provided sterile via gamma irradiation
- Manufactured of commercially pure titanium (Grade 4) per ASTM F67;
- Bone level design
- Apically tapered thread-form with apical cutting flutes in 3.5, 3.75 & 4.3 mm diameters with lengths of 7, 10, 11.5, 13, 16 & 18 mm and in 5.0 mm diameter with lengths of 7, 10, 11.5, 13 & 16 mm.
- Provided with an internal hexagonal implant-to-abutment interface compatible with the subject Cover Screw and Abutments.
Cover Screws
- Intended for single use;
- Used to protect the internal geometry of the subject Nuvo IF implants during the healing phase when a two-stage protocol is used;
- Placed out of occlusion;
- Provided sterile via gamma irradiation in the same barrier package as the subject implants—also provided individually packaged in sterile condition via ethylene oxide gas;
- Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136;
- Provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the subject Nuvo IF implants.
Abutments
- Intended for single use;
- Provided sterile via ethylene oxide gas or non-sterile
- Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136 or in a combination of POM (Polyoxymethylene) polymer and cobalt-chromium alloy (CoCr), conforming to ASTM F1537;
- Conical format available in different diameters, height of cementable area and gingival height;
- Screw-retained to the implant;
- Provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the subject Nuvo IF implants;
- Provided with coronal geometries in rotational (non-indexed) versions to support multiunit restorations and in anti-rotational (indexed) versions to support single restorations
- Titanium Base abutments are two-piece abutments which are composed of a titanium base and a patient-specific CAD/CAM top-half, that when assembled together form the final finished device.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
maxilla or mandible
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Dynamic fatigue test per ISO 14801 and FDA guidance entitled Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, issued on: May 12, 2004, was performed to determine the fatigue strength for the worst-case constructs assembled using the subject devices.
Sterilization of the subject endosseous dental implant devices via gamma irradiation was validated per ISO 11137-1 and ISO 11137-2. A minimum Sterility Assurance Level (SAL) of 1 x 10 ° has been validated.
Sterilization of the subject abutments via ethylene oxide gas was validated per ISO 11135. A minimum Sterility Assurance Level (SAL) of 1 x 10 ° has been validated.
Ethylene oxide residuals have been assessed per ISO 10993-7. Residuals are within accepted limits.
Sterilization of the subject abutments via moist heat was validated per ISO 17665-2. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Biological Safety Assessment guided by ISO 10993-1 and FDA guidance entitled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" : Guidance for Industry and Food and Drug Administration Staff, issued June 16, 2016. Reference to previous biocompatibility testing is supplied as follows:
-Cytotoxicity testing was performed per ISO 10993-5.
-Chemical characterization was performed per ISO 10993-18.
Package transport integrity testing has been performed per ISTA 2A.
Product and package stability has been validated per ASTM F1980.
Product endotoxin testing (LAL) is performed monthly on products representative of the subject devices per ANSI/AAMI ST72.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
K150203, K150367, K153624, K163194, K173902, K190040, K190718, K191191, K192229
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 872.3640 Endosseous dental implant.
(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it.
June 18, 2020
JJGC Indústria e Comércio de Materiais Dentários S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K193234
Trade/Device Name: NUVO IF Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 19, 2020 Received: May 20, 2020
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K193234
Device Name NUVO IF Implant System
Indications for Use (Describe)
Indications for Use for NUVO IF Implants and conventional abutments:
The NUVO IF Implant System is intended to be surgically placed in the maxilla or mandible to provide support for prosthetic devices such as artificial teeth in order to restore chewing function. It may be used with single-stage or twostage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
The Titanium Temporary Abutment is indicated to provide temporary support for prosthesis structure for up to 6 months.
The Attachment Equator and Attachment Removable Prosthesis are indicated for the attachment of full or partial dentures to NUVO implants.
Indications for Use for Rotational and Anti-rotational Titanium Bases abutments:
Titanium Base Abutment is a titanium base placed onto dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement or screw-retained single or multi-unit restorations or screw-retained single restorations.
All digitally designed copings and/or crowns to be used with the Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The Rotational Titanium Base is indicated for cement or screw-retained multi-unit restorations.
Indications for Use for CARES® Abutment IF:
The CARES® Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations (copings or crowns). It is indicated for cementretained single restorations. All digitally designed abutments for use with the CARES® Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
of Health and Hun Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary
K193234
ADMINISTRATIVE INFORMATION | |
---|---|
Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA |
(dba Neodent) | |
Av. Juscelino Kubitschek de Oliveira, 3291 | |
Curitiba, Parana, Brazil 81270-200 | |
Registration No.: 3008261720 | |
Owner/Operator No.: 10031702 | |
Contact Person | Jennifer M. Jackson, MS |
Director of Regulatory Affairs, | |
Straumann USA | |
E-mail: jennifer.jackson@straumann.com | |
Telephone (978) 747-2509 | |
Date Prepared | 17 June 2020 |
Preparer / Alternate Contact | Luiza Vaccari Toppel |
Regulatory Affairs Coordinator | |
JJGC Indústria e Comércio de Materiais Dentários SA | |
E-mail: luiza.toppel@neodent.com |
DEVICE NAME AND CLASSIFICATION
Trade/ Proprietary Name | NUVO IF Implant System |
---|---|
Common Name | Endosseous dental implant |
Classification Name(s) | Endosseous dental implant |
Classification Regulation(s) | 21 CFR 872.3640, Class II |
Product Code(s) | DZE; NHA |
Classification Panel | Dental Products Panel |
Reviewing Branch | Dental Devices Branch |
4
PREDICATE DEVICE INFORMATION
| Primary Predicate Device | K101945 – Neodent Implant System, JJGC Indústria e
Comércio de Materiais Dentários SA |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reference Predicate Devices | K150203 – Medentika CAD/CAM Abuments, Medentika
GmbH
K150367 – Neodent Implant System, JJGC
K153624 – Neodent Implant System, JJGC
K163194 – Neodent Implant System - GM Line, JJGC
K173902 – Neodent Implant System - GM Line, JJGC
K190040 – Straumann BLX Line Extension - New
Abutments, Institut Straumann AG
K190718 – GM Zygomatic Implants, JJGC
K191191 – Neodent Implant System - Temporary
Abutments, JJGC
K192229 – Neodent Implant System – Neodent Titanium
Base for Bridge, JJGC |
INDICATIONS FOR USE
Indications for Use for NUVO IF Implants and conventional abutments:
The NUVO IF Implant System is intended to be surgically placed in the maxilla or mandible to provide support for prosthetic devices such as artificial teeth in order to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
The Titanium Temporary Abutment is indicated to provide temporary support for prosthesis structure for up to 6 months.
The Attachment Equator and Attachment Removable Prosthesis abutments are indicated for the attachment of full or partial dentures to NUVO implants.
Indications for Use for Rotational and Anti-rotational Titanium Bases abutments:
Titanium Base Abutment is a titanium base placed onto dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations or screw-retained single restorations.
All digitally designed copings and/or crowns to be used with the Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The Rotational Titanium Base is indicated for cement or screw-retained multi-unit restorations.
Indications for Use for CARES® Abutment IF:
The CARES® Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations (copings or crowns). It is indicated for screw-retained or cement-retained single restorations.
All digitally designed abutments for use with the CARES® Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.
5
SUBJECT DEVICE DESCRIPTIONS
Implants
- . Intended for single use;
- Provided sterile via gamma irradiation
- Manufactured of commercially pure titanium (Grade 4) per ASTM F67;
- Bone level design
- Apically tapered thread-form with apical cutting flutes in 3.5, 3.75 & 4.3 mm diameters with lengths of 7, 10, 11.5, 13, 16 & 18 mm and in 5.0 mm diameter with lengths of 7, 10, 11.5, 13 & 16 mm.
- Provided with an internal hexagonal implant-to-abutment interface compatible with the subject Cover Screw and Abutments.
Cover Screws
- Intended for single use;
- Used to protect the internal geometry of the subject Nuvo IF implants during the healing phase when a two-stage protocol is used;
- Placed out of occlusion;
- Provided sterile via gamma irradiation in the same barrier package as the subject implants—also provided individually packaged in sterile condition via ethylene oxide gas;
- Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136;
- Provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the subject Nuvo IF implants.
Abutments
- Intended for single use;
- Provided sterile via ethylene oxide gas or non-sterile
- Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136 or in a combination of POM (Polyoxymethylene) polymer and cobalt-chromium alloy (CoCr), conforming to ASTM F1537;
- Conical format available in different diameters, height of cementable area and gingival height;
- Screw-retained to the implant;
- Provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the subject Nuvo IF implants;
- Provided with coronal geometries in rotational (non-indexed) versions to support multiunit restorations and in anti-rotational (indexed) versions to support single restorations
- Titanium Base abutments are two-piece abutments which are composed of a titanium base and a patient-specific CAD/CAM top-half, that when assembled together form the final finished device.
SUBSTANTIAL EQUIVALENCE COMPARISON TABLES
The Substantial Equivalence Comparison tables are provided on the pages that follow below.
6
Table 1. Substantial Equivalence -- Indication for Use Statements | |||||||
---|---|---|---|---|---|---|---|
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATES | |||||
COMPARISON | K193234 | ||||||
NUVO IF Implant System | |||||||
JJGC Indústria e Comércio de Materiais | |||||||
Dentários S.A. | K101945 | ||||||
Neodent Implant System – | |||||||
CM Alvim Implant System | |||||||
JJGC Indústria e Comércio de | |||||||
Materiais Dentários S.A. | K150367 | ||||||
Neodent Implant System – | |||||||
TiBase & Preface | |||||||
JJGC Indústria e Comércio de | |||||||
Materiais Dentários S.A. | K163194 | ||||||
Neodent Implant System – | |||||||
GM Line | |||||||
JJGC Indústria e Comércio de | |||||||
Materiais Dentários S.A. | K192229 | ||||||
Neodent Implant System – | |||||||
GM Titanium Base for | |||||||
Bridge | |||||||
JJGC Indústria e Comércio de | |||||||
Materiais Dentários S.A. | K191191 | ||||||
Neodent Implant System – | |||||||
Temporary Abutments | |||||||
JJGC Indústria e Comércio de | |||||||
Materiais Dentários S.A. | K190040 | ||||||
Straumann® PUREloc | |||||||
abutments | |||||||
Institut Straumann AG | |||||||
Indications | |||||||
for Use | |||||||
Statement | Indications for Use for NUVO IF Implants | ||||||
and conventional abutments: | |||||||
The NUVO IF Implant System is intended to | |||||||
be surgically placed in the maxilla or | |||||||
mandible to provide support for prosthetic | |||||||
devices such as artificial teeth in order to | |||||||
restore chewing function. It may be used | |||||||
with single-stage or two-stage procedures, | |||||||
for single or multiple unit restorations, and | |||||||
may be loaded immediately when good | |||||||
primary stability is achieved and with | |||||||
appropriate occlusal loading. Multiple | |||||||
tooth applications may be rigidly splinted. |
The Titanium Temporary Abutment is
indicated to provide temporary support for
prosthesis structure for up to 6 months.
The Attachment Equator and Attachment
Removable Prosthesis abutments are
indicated for the attachment of full or
partial dentures to NUVO implants.
Indications for Use for Rotational and Anti-
rotational Titanium Bases abutments:
Titanium Base Abutment is a titanium base
placed onto dental implants to provide
support for customized prosthetic
restorations. It is used with a coping and
crown, or crown alone, and is indicated for
cement-retained single or multi-unit
restorations or screw-retained single
restorations.
All digitally designed copings and/or
crowns to be used with the Titanium Base
Abutment System are intended to be sent | The Neodent Implant System
is intended to be surgically
placed in the bone of the
upper or lower jaw to
provide support for
prosthetic devices such as
artificial teeth, to restore
chewing function. It may be
used with single-stage or
two-stage procedures, for
single or multiple unit
restorations, and may be
loaded immediately when
good primary stability is
achieved and with
appropriate occlusal loading.
Multiple tooth applications
may be rigidly splinted. | Titanium Base Abutment is a
titanium base placed onto
Neodent dental implants to
provide support for
customized prosthetic
restorations. It is used with a
coping and crown, or crown
alone, and is indicated for
cement-retained single or
multi-unit restorations, or
screw-retained single
restorations.
PreFace Abutment is a
titanium abutment to be used
in fabricating a full custom
abutment and placed onto
Neodent dental implants to
provide support for
customized prosthetic
restorations.
PreFace Abutments are indicated for
screw-retained single
restorations or cement-
retained single or multi-unit
restorations. | Indications for Use for GM
implants and conventional
abutments:
The Neodent Implant System
is intended to be surgically
placed in the bone of the
upper or lower jaw to
provide support for
prosthetic devices such as
artificial teeth, to restore
chewing function. It may be
used with single-stage or
two-stage procedures, for
single or multiple unit
restorations, and may be
loaded immediately when
good primary stability is
achieved and with
appropriate occlusal loading.
Indications for Use for GM
Titanium Base abutments:
Titanium Base Abutment is a
titanium base placed onto
Neodent dental implants to
provide support for
customized prosthetic
restorations. It is used with a
coping and crown, or crown
alone, and is indicated for
cement-retained single or
multi-unit restorations, or
screw-retained single
restorations.
All digitally designed copings
and/or crowns for use with
the Neodent Titanium Base | Titanium Base Abutment is a
titanium base placed onto
Neodent dental implants to
provide support for
customized prosthetic
restorations. It is used with a
coping and crown, or crown
alone, and is indicated for
cement-retained single or
multi-unit restorations or
screw-retained single
restorations.
All digitally designed copings
and/or crowns to be used with the
Neodent Titanium Base
Abutment System are
intended to be sent to
Straumann for manufacture
at a validated milling center.
The GM Titanium Base for
Bridge is indicated for
cement or screw-retained
multi-unit restorations. | The Neodent Implant System
is intended to be surgically
placed in the bone of the
upper or lower jaw to
provide support for
prosthetic devices such as
artificial teeth, to restore
chewing function. It may be
used with single-stage or
two-stage procedures, for
single or multiple unit
restorations, and may be
loaded immediately when
good primary stability is
achieved and with
appropriate occlusal loading.
The Neodent Implant System
- Temporary Abutments are
indicated to be used on
Neodent implants to provide
temporary support for
prosthesis structure for up to
6 months. | The Straumann®
Retentive System is
indicated for the
attachment of full or
partial dentures on
Straumann implants |
| | SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATES | | | | |
| COMPARISON | K193234
NUVO IF Implant System
JJGC Indústria e Comércio de Materiais
Dentários S.A. | K101945
Neodent Implant System -
CM Alvim Implant System
JJGC Indústria e Comércio de
Materiais Dentários S.A. | K150367
Neodent Implant System -
TiBase & Preface
JJGC Indústria e Comércio de
Materiais Dentários S.A. | K163194
Neodent Implant System -
GM Line
JJGC Indústria e Comércio de
Materiais Dentários S.A. | K192229
Neodent Implant System -
GM Titanium Base for
Bridge
JJGC Indústria e Comércio de
Materiais Dentários S.A. | K191191
Neodent Implant System -
Temporary Abutments
JJGC Indústria e Comércio de
Materiais Dentários S.A. | K190040
Straumann® PUREloc
abutments
Institut Straumann AG |
| | validated milling center.
The Rotational Titanium Base is indicated
for cement or screw-retained multi-unit
restorations.
Indications for Use for CARES® Abutment
IF:
The CARES® Abutment is a customized
prosthetic abutment, manufactured in
titanium alloy, placed onto dental implants
to provide support for customized
prosthetic restorations (copings or crowns).
It is indicated for screw-retained or
cement-retained single restorations.
All digitally designed abutments for use
with the CARES® Abutment are intended to
be sent to Straumann for manufacturing at | | | intended to be sent to
Straumann for manufacture
at a validated milling center.
Indications for Use for GM
Pro Peek Abutments:
The Pro PEEK Abutments are
indicated to be used on
Neodent implants to provide
temporary support for
prosthesis structure for up to
6 months. They can be used in
one or two stage procedures
and also immediate load when
there is good primary stability. | | | |
Table 1: Substantial Equivalence – Indication for Use Statements
7
8
The scope of the Indications for Use Statement for the subject devices (implants and conventional abutments section) are contained within the corresponding section of the Indications for Use Statement of the primary predicate device K101945. The specific languages for Titanium Temporary Abutment and Attachment Equator/Attachment Removable Prosthesis are the same of the corresponding section of the Indications for Use Statement of K191191 (Neodent Implant System - Temporary Abutments) and K190040 (Straumann® PUREloc abutments), except for the names of the devices.
The Indications for Use Statement for the subject device (Rotational and Anti-rotational Titanium Bases abutments section) is identical to the corresponding section of the Indications for Use Statement of K163194 (GM Titanium Base abutments section), except for the names of the devices. The specific indication for use for Titanium Base Rotational is the same of K192229 (Neodent Implant System - GM Titanium Base for Bridge).
The Indications for Use Statement for the subject device (CARES® Abutment IF section) is similar to the PreFace section of the Indications for Use Statement of the reference device K150367. The slight differences are the names of the devices and, for the subject device statement, the requirement to manufacture at a Straumann milling center. For K150367 the requirement to manufacture at a Straumann milling center was included in the labeling. The slight differences in wording between the Indications for Use Statements for the subject device and the reference device K150367 do not affect the intended use with dental implants for rehabilitation of the edentulous maxilla or mandible.
A comparison of the technological characteristics of the subject device and the predicate and reference devices is provided in the following table.
9
Table 2: Substantial Equivalence Comparison – Technological characteristics - Implants
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATE | ||
---|---|---|---|---|
K193234 | K101945 | K163194 | ||
COMPARISON | NUVO IF Implant System | |||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | Neodent Implant System - CM Line | |||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | Neodent Implant System - GM Line | |||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | EQUIVALENCE DISCUSSION | |||
Implant Type | Bone Level | Bone Level | Same | |
The subject devices have the same external | ||||
design as the primary predicate. | ||||
Implant-to- | ||||
Abutment | ||||
Interface | Internal Hex | Cone Morse | Equivalent | |
The implant interface is equivalent to that of | ||||
the predicate. Dynamic fatigue testing has | ||||
been performed to establish the performance | ||||
of the interface. | ||||
Implant | ||||
Diameter | 3.5 mm, 3.75 mm, 4.3 mm and 5.0 mm | 3.5 mm, 4.3 mm and 5.0 mm | 3.5 to 5.0 mm | Equivalent |
The subject device diameters are in the range | ||||
of diameters of the primary and reference | ||||
predicate devices. | ||||
Implant Length | 3.5 mm: 7, 10, 11.5, 13, 16 & 18 mm | |||
3.75 mm: 7, 10, 11.5, 13, 16 & 18 mm | ||||
4.3 mm: 7, 10, 11.5, 13, 16 & 18 mm | ||||
5.0 mm: 7, 10, 11.5, 13 & 16 mm | Alvim CM Implant | |||
All diameters: 8, 10, 11.5, 13 & 16 mm |
Titamax CM Implant
3.5 mm: 7, 8, 9, 11, 12, 15 & 17mm
3.75 mm: 7, 8, 9, 11, 12, 15 & 17mm
4.0 mm: 7, 8, 9, 11, 12, 15 & 17mm
5.0 mm: 7, 8, 9, 11, 12, 15 & 17mm | 8 to 18 mm | Equivalent
The subject device lengths are within the range
established by the primary and reference
predicate devices. |
| Thread Design | Apically Tapered, Dual Helix | Apically Tapered, Dual Helix
(Trade Name = Alvim) | | Same |
| Surface Finish | Sand blasted and acid etched | Sand blasted and acid etched
(Trade Name = Neoporos) | | Same |
| Material | Commercially Pure Titanium (Grade 4) | Commercially Pure Titanium (Grade 4) | | Same |
| Single Use | Yes | Yes | | Same |
| Sterilization
Method | Gamma Irradiation to an SAL of 1x10-6 | Gamma Irradiation to an SAL of 1x10-6 | | Same |
| | | | EQUIVALENCE DISCUSSION | |
| COMPARISON | SUBJECT DEVICES | PRIMARY PREDICATE | | |
| | K193234 | K101945 | | |
| | NUVO IF Implant System | Neodent Implant System - CM Alvim Implant System | | |
| | JJGC Indústria e Comércio de Materiais Dentários S.A. | JJGC Indústria e Comércio de Materiais Dentários S.A. | | |
| Cover Screw/Healing Abutment | | | | |
| Principal of operation | Screw used to protect internal features of the implant
during submerged healing protocol and to define gingiva
shape during delayed loading protocols. | Conditioning the soft tissues and closing the implant
interface during healing phase. | Equivalent
The principal of operation of subject device is equivalent of
the primary and reference predicate device. | |
| Implant-to-Abutment
Interface | Internal Hex | Cone Morse | Equivalent
The implant interface is equivalent to that of the predicate. | |
| Diameter | Cover Screw Platform Ø:
NP - 3.5 mm SP - 4.5 mm | Conventional abutments:
3.5 to 4.5 mm | Equivalent
Subject device diameters are within the range of diameters
of the predicate devices or larger. Larger diameters do not
represent a worst case in terms of performance. | |
| | Healing Abutment Platform Ø:
NP - 3.5 mm SP - 4.5 mm and 5.5 mm | | | |
| Gingival Height | 1.0, 3.0 and 5.0 mm | 0.8 to 6.5 mm | Equivalent
Subject device heights are included in the range of the
predicate device heights. | |
| Angulation | Straight | Straight | Same | |
| Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same | |
| Surface Treatment | Machined and anodized (only NP platform) | Machined and anodized | Same | |
| Single Use | Yes | Yes | Same | |
| Sterilization Method | Co-packaged implant and cover screw Gamma Irradiation to
an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Equivalent
The subject device when provided co-packaged with the
implant share the same sterilization method and minimum
SAL of the reference predicate device. | |
| | Cover screw and Healing provided individually
Ethylene Oxide to an SAL of 1x10-6 | | The subject devices when provided individually share the
same sterilization method and minimum SAL of the primary
predicate device. | |
| Multi-Unit Abutment | | | | |
| Principal of operation | Abutment designed to accept a screw-retained
prosthetic. | To support final restorations when placed on implants. | Equivalent
The principal of operation of subject device is equivalent of
the primary and reference predicate device. | |
| Implant-to-Abutment
Interface | Internal Hex | Cone Morse | Equivalent
The implant interface is equivalent to that of the predicate. | |
| | SUBJECT DEVICES | PRIMARY PREDICATE | EQUIVALENCE DISCUSSION | |
| COMPARISON | K193234
NUVO IF Implant System
JJGC Indústria e Comércio de Materiais Dentários S.A. | K101945
Neodent Implant System - CM Alvim Implant System
JJGC Indústria e Comércio de Materiais Dentários S.A. | | |
| Indexing | Indexed (angled abutments) and non-indexed (straight abutments) | Indexed (angled abutments) and non-indexed (straight abutments) | Same | |
| Diameter | Platform Ø:
NP – 3.5 & 4.8 mm
SP – 4.8 mm | Platform Ø: 3.5 mm, 4.5 mm | Equivalent
Subject device diameters are within the range of diameters
of the predicate devices or larger. Larger diameters do not
represent a worst case in terms of performance. | |
| Gingival Height | NP – 0.5, 1, 2, 2.3, 3, 4 & 5 mm
SP – 1, 2, 2.3, 3, 4 & 5 mm | Straight: 0.8 to 6.5 mm
Angled: 1.5 to 3.5 mm | Equivalent
Subject devices are included in the range of the predicate
device gingival heights or higher.
Higher gingival height do not represent a worst case in terms
of performance. | |
| Angulation | 0°, 17° & 30° | 0°, 17° & 30° | Same | |
| Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same | |
| Surface Treatment | Machined and anodized (only NP platform) | Machined | Equivalent
Anodized surface treatment is presented in other abutments
of K101945. | |
| Single Use | Yes | Yes | Same | |
| Sterilization Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same | |
| Principal of operation | Abutment designed to accept a cement-retained prosthetic.
Customizable Cement Retained Abutment designed to accept a cement-retained prosthetic. Features to facilitate
trimming height of coronal geometry from 6 mm down to a minimum of 4 mm. | To support final restorations when placed on implants. | Equivalent
The principal of operation of subject device is within of the
primary predicate device.
These devices are equivalent in design to the subject straight
Cement-Retained Abutments described above. The ability to
modify the abutment height does not affect the intended
use. The length of the combination of the abutment and the
cemented prosthesis will be equivalent to the straight
Cement-Retained Abutments. | |
| Implant-to-Abutment
Interface | Internal Hex | Cone Morse | Equivalent
The implant interface is equivalent to that of the predicate. | |
| Indexing | Indexed | Both Indexed and Non-indexed | Same | |
| Diameter | Platform Ø:
NP – 3.5 mm
SP – 4.5 mm | Platform Ø: 3.5 mm, 4.5 mm | Same | |
| | SUBJECT DEVICES | PRIMARY PREDICATE | | |
| COMPARISON | K193234 | K101945 | EQUIVALENCE DISCUSSION | |
| | NUVO IF Implant System | Neodent Implant System - CM Alvim Implant System | | |
| | JJGC Indústria e Comércio de Materiais Dentários S.A. | JJGC Indústria e Comércio de Materiais Dentários S.A. | | |
| Gingival Height | Straight - 0.5, 1. 2, 1.5 & 3 mm
17° - 1.5 & 3 mm
25° - 2 & 3 mm | Straight - 0.8, 1.5, 2.5, 3.5, 4.5, 5.5 & 6.5 mm
Angled - 1.5, 2.5 & 3.5 mm | Equivalent
Subject device are included in the range of the predicate
device gingival heights. | |
| Angulation | 0°, 17° & 25° | 0°, 17° & 30° | Equivalent
The subject devices are included in the range of the predicate
device angulation. | |
| Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same | |
| Surface Treatment | Machined and anodized (only NP platform) | Machined | Equivalent
Anodized surface treatment is presented in other abutments
of K101945. | |
| Single Use | Yes | Yes | Same | |
| Sterilization Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same | |
10
Table 3: Substantial Equivalence Comparison – Traditional Abutments (Cover Screw/Healing Abutment/Multi-Unit Abutnent/Cement Retained Abutment)
11
12
13
Table 4: Substantial Equivalence Comparison – Technological characteristics – Traditional Abutments)
SUBJECT DEVICES | REFERENCE PREDICATE | EQUIVALENCE DISCUSSION | SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATES | ||||
---|---|---|---|---|---|---|---|---|---|
K193234 | K191191 | COMPARISON | K193234 | ||||||
NUVO IF Implant System | |||||||||
JJGC Indústria e Comércio de | |||||||||
Materiais Dentários S.A. | K101945 | ||||||||
Neodent Implant System - CM | |||||||||
Alvim Implant System | |||||||||
JJGC Indústria e Comércio de | |||||||||
Materiais Dentários S.A. | K173902 | ||||||||
Neodent Implant System - GM Line | |||||||||
JJGC Indústria e Comércio de | |||||||||
Materiais Dentários S.A. | K190040 | ||||||||
Straumann BLX Line Extension - | |||||||||
New Abutments | |||||||||
Institut Straumann AG | EQUIVALENCE DISCUSSION | ||||||||
COMPARISON | NUVO IF Implant System | ||||||||
JJGC Indústria e Comércio de Materiais Dentários S.A. | Neodent Implant System - Temporary Abutments | ||||||||
JJGC Indústria e Comércio de Materiais Dentários S.A. | Equator Attachment | Equator Attachment IF | CM Mini Ball Attachment | GM Attachment Equator | Novaloc Abutments | ||||
Titanium Temporary | Titanium Temporary Abutment for Crown and for | ||||||||
Bridge | GM Temporary Abutment | Principal of operation | Abutment to accept detachable | ||||||
over-denture prostheses. | |||||||||
Coronal Geometry: To accept | |||||||||
o-ring style matrices | Abutment to accept detachable | ||||||||
over-denture prostheses. | |||||||||
Coronal Geometry: To accept | |||||||||
o-ring style matrices | Abutment to accept detachable | ||||||||
over-denture prostheses. | |||||||||
Coronal Geometry: To accept o- | |||||||||
ring style matrices | Same | ||||||||
The principal of operation of | |||||||||
subject device is the same of the | |||||||||
primary predicate device. | |||||||||
Principal of operation | Abutment for the creation of a temporary, non-occlusal | ||||||||
restoration to address esthetics during fabrication of final | |||||||||
prosthesis. | Abutment for the creation of a temporary, non-occlusal | ||||||||
restoration to address esthetics during fabrication of final | |||||||||
prosthesis. | Same | ||||||||
The principal of operation of subject device is the same of | |||||||||
the primary predicate device. | Implant-to-Abutment | ||||||||
Interface | Internal Hex | Cone Morse | Cone Morse | Equivalent | |||||
The implant-to-abutment interface | |||||||||
is equivalent to that of the | |||||||||
reference predicate. Dynamic | |||||||||
fatigue testing is provided to | |||||||||
support the performance of the | |||||||||
subject Internal Hex interface. | |||||||||
Implant-to-Abutment | |||||||||
Interface | Internal Hex | Cone Morse | Equivalent | ||||||
The implant interface is equivalent to that of the predicate. | Indexing | Non-indexed | Non-indexed | Non-indexed | Same | ||||
Indexing | Both Indexed and Non-indexed | Both Indexed and Non-indexed | Same | Diameter | Platform Ø: NP - 3.5 mm | ||||
SP - 4.5 | Not Defined | Platform Ø: 3.5 to 5.0 mm | Equivalent | ||||||
Subject devices are within the | |||||||||
range of reference predicate | |||||||||
devices diameters. | |||||||||
Diameter | Platform Ø: | ||||||||
NP – 3.5 mm | |||||||||
SP - 4.5 & 5.5 mm | Platform Ø: 3.5 mm, 4.5 mm | Equivalent | |||||||
Subject device diameters are within the range of diameters | |||||||||
of the predicate devices or larger. Larger diameters do not | |||||||||
represent a worst case in terms of performance. | Gingival Height | NP – 1, 2, 3, 4 & 5 mm | |||||||
SP – 1, 2, 3, 4 & 5 mm | 1.5, 2.5, 3.5, 4.5 & 5.5 mm | 1.5, 2.5, 3.5, 4.5 & 5.5 mm | Equivalent | ||||||
Subject devices are included in the | |||||||||
range of the predicate devices | |||||||||
gingival heights. | |||||||||
Gingival Height | 1.0 mm | 0.8; 1.5; 2.5 and 3.5 mm | Equivalent | ||||||
Subject device are included in the range of the predicate | |||||||||
device gingival heights. | Angulation | Straight | Straight | Straight | Same | ||||
Angulation | Straight | Straight | Same | Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same | |
Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same | Surface Treatment | Machined with titanium nitride | ||||
coating | Machined | Machined with titanium nitride | |||||||
coating | Equivalent | ||||||||
Subject devices and reference | |||||||||
predicate devices have the same | |||||||||
surface treatment | |||||||||
Surface Treatment | Machined (SP platform) | ||||||||
Machined and anodized (NP platform) | Machined | Equivalent | |||||||
Anodized surface treatment is presented in other abutments | |||||||||
of primary predicate device K101945. | Single Use | Yes | Yes | Yes | Same | ||||
Duration of Use | Up to 6 months | Up to 6 months | Same | Sterilization Method | Ethylene Oxide to an SAL of | ||||
1×10-6 | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same | ||||||
Single Use | Yes | Yes | Same | ||||||
Sterilization Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same |
14
Table 5: Substantial Equivalence Comparison – Traditional Abutnents (Equator Attachment and Removable Prosthesis Attachment)
15
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATES | EQUIVALENCE DISCUSSION | ||
---|---|---|---|---|---|
COMPARISON | K193234 | ||||
NUVO IF Implant System | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K101945 | ||||
Neodent Implant System – CM | |||||
Alvim Implant System | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K173902 | ||||
Neodent Implant System - GM Line | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K190040 | ||||
Straumann BLX Line Extension - | |||||
New Abutments | |||||
Institut Straumann AG | |||||
Removable Prosthesis | |||||
Attachment | Removable Prosthesis | ||||
Attachment IF | CM Mini Ball Attachment | GM Attachment Equator | Novaloc Abutments | ||
Principal of operation | Abutment to accept detachable | ||||
over-denture prostheses. | Abutment to accept detachable | ||||
over-denture prostheses. | Abutment to accept detachable | ||||
over-denture prostheses. | Abutment to accept detachable | ||||
over-denture prostheses. | Same | ||||
The principal of operation of subject | |||||
device is the same of the primary | |||||
predicate device. | |||||
Design | Coronal Geometry: To accept | ||||
Valoc Novaloc matrices | Coronal Geometry: To accept o- | ||||
ring style matrices | Coronal Geometry: To accept o- | ||||
ring style matrices | Coronal Geometry: To accept | ||||
Valoc Novaloc matrices | Equivalent | ||||
Subject devices and reference | |||||
predicates per K190040 have same | |||||
intended use and similar coronal | |||||
geometry (differing gingival heights | |||||
are not significant to performance). | |||||
Implant-to-Abutment | |||||
Interface | Internal Hex | Cone Morse | Cone Morse | Straumann BLX | Equivalent |
The implant interface is equivalent | |||||
to that of the predicate. | |||||
Indexing | Non-indexed | Non-indexed | Non-indexed | Indexed | Same |
Diameter | Platform Ø: NP - 3.5 mm | ||||
SP - 4.5 | Not Defined | Platform Ø: 3.5 to 5.0 mm | 4.5 mm | Equivalent | |
Subject devices are within the range | |||||
of reference predicate devices | |||||
diameters. | |||||
Gingival Height | NP - 1, 2, 3, 4 & 5 mm | ||||
SP - 1, 2, 3, 4 & 5 mm | 1.5, 2.5, 3.5, 4.5 & 5.5 mm | 1.5, 2.5, 3.5 ,4.5 & 5.5mm | 1.5 to 6.5 mm | Equivalent | |
Subject devices are included in the | |||||
range of the predicate devices | |||||
gingival heights. | |||||
Angulation | Straight | Straight | Straight | Straight and Angulated | Same |
Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same |
Surface Treatment | Machined and titanium nitride | ||||
coating | Machined and titanium nitride | ||||
coating | Machined and titanium nitride | ||||
coating | Machined and titanium nitride | ||||
coating | Equivalent | ||||
Subject devices and reference | |||||
predicate devices have the same | |||||
surface treatment | |||||
Single Use | Yes | Yes | Yes | Yes | Same |
COMPARISON | SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATES | EQUIVALENCE DISCUSSION | |
K193234 | |||||
NUVO IF Implant System | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K101945 | ||||
Neodent Implant System – CM | |||||
Alvim Implant System | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K173902 | ||||
Neodent Implant System - GM Line | |||||
JJGC Indústria e Comércio de | |||||
Materiais Dentários S.A. | K190040 | ||||
Straumann BLX Line Extension - | |||||
New Abutments | |||||
Institut Straumann AG | |||||
Sterilization Method | Ethylene Oxide to an SAL of | ||||
$1 \times 10^{-6}$ | Ethylene Oxide to an SAL of $1 \times 10^{-6}$ | Ethylene Oxide to an SAL of $1 \times 10^{-6}$ | Non-Sterile | Same | |
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATE | |||
COMPARISON | K193234 | ||||
NUVO IF Implant System | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | K101945 | ||||
Neodent Implant System - CM Alvim | |||||
Implant System | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | K173902 | ||||
Neodent Implant System - GM Line | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | EQUIVALENCE DISCUSSION | ||||
UCLA CoCr | UCLA CoCr IF | Not applicable | GM Exact Co-Cr Abutment for Crown | ||
Principal of operation | Abutment to facilitate the casting of base | ||||
metal and precious metal alloy prostheses. | Abutment to facilitate the casting of base | ||||
metal and precious metal alloy prostheses. | Same | ||||
The principal of operation of subject device | |||||
is the same of the reference predicate | |||||
device. | |||||
Implant-to-Abutment | |||||
Interface | Internal Hex | Cone Morse | Cone Morse | Equivalent | |
The implant interface is equivalent to that of | |||||
the predicate. | |||||
Indexing | Both Indexed and Non-indexed | Non-indexed | Equivalent | ||
The subject devices indexing is the | |||||
equivalent of reference predicate devices. | |||||
The indexed devices are presented in other | |||||
primary predicate devices. | |||||
Diameter | Platform Ø: NP – 3.5 mm | ||||
SP - 4.5 & 5.5 mm | Platform Ø: 3.5 to 6.0 mm | Equivalent | |||
Subject devices are within the range of | |||||
reference predicate devices diameters. | |||||
Gingival Height | NP - 1, 2, 3, 4 & 5 mm | ||||
SP - 1, 2, 3, 4 & 5 mm | 1 mm | Equivalent | |||
The different gingival heights do not | |||||
introduce a new worst in terms of | |||||
performance. | |||||
Angulation | Straight | Straight | Same | ||
The subject devices and the predicate | |||||
devices are straight. | |||||
Material | CoCr Alloy with POM polymer | CoCr Alloy with POM polymer | Same | ||
Subject, primary and reference predicate | |||||
devices have the same material of | |||||
construction. | |||||
Surface Treatment | Machined and anodized (only NP | ||||
platform) | Machined | Same | |||
Anodized surface treatment is presented in | |||||
other abutments of primary predicate device | |||||
K101945. | |||||
Single Use | Yes | Yes | Yes | Same | |
SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATE | |||
COMPARISON | K193234 | ||||
NUVO IF Implant System | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | K101945 | ||||
Neodent Implant System – CM Alvim | |||||
Implant System | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | K173902 | ||||
Neodent Implant System - GM Line | |||||
JJGC Indústria e Comércio de Materiais | |||||
Dentários S.A. | EQUIVALENCE DISCUSSION | ||||
Sterilization Method | Provided non-sterile. Terminally sterilized by | ||||
the user via moist heat. Moist heat cycle | |||||
parameters have been validated to an SAL of | |||||
$1 x 10^{-6}$ . | Provided non-sterile. Terminally sterilized by | ||||
the user via moist heat. Moist heat cycle | |||||
parameters have been validated to an SAL of | |||||
$1 x 10^{-6}$ . | Same | ||||
Subject and reference predicate device | |||||
utilize the same sterilization method and | |||||
minimum SAL. |
16
17
Table 6: Substantial Equivalence Comparison – Technological characteristics – Traditional Abutments (UCLA CoCr)
18
19
Table 7: Substantial Equivalence Comparison – CADCAM Abutments (Anti-Rotational Titanium Base IF)
SUBJECT DEVICES | REFERENCE PREDICATES | |||
---|---|---|---|---|
COMPARISON | K193234 | |||
NUVO IF Implant System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K150367 & K153624 | |||
Neodent Implant System | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | K163194 | |||
Neodent Implant System - GM Line | ||||
JJGC Indústria e Comércio de Materiais | ||||
Dentários S.A. | EQUIVALENCE DISCUSSION | |||
Anti-Rotational | ||||
Titanium Base IF | Anti-Rotational Titanium Base IF | Titanium Base Abutment | GM Exact Titanium Base | |
Principal of | ||||
operation | Abutment designed to accept a cement- | |||
retained prosthetic. |
Titanium Base abutments are two-piece
abutments which are composed of a
titanium base and a patient-specific
CAD/CAM top-half, that when assembled | Abutment designed to accept a cement-
retained single or multi-unit restorations, or
screw-retained single restorations.
Titanium Base abutments as two-piece
abutments which are composed of a titanium
base and a patient-specific CAD/CAM top- | Screw-retained single-unit, or Cement-
retained single or multi-unit.
Titanium Base abutments as two-piece
abutments which are composed of a titanium
base and a patient-specific CAD/CAM top-
half, that when assembled together form the | Equivalent
The principal of operation of subject device
is within the reference predicate devices. |
| | together form the final finished device. | half, that when assembled together form the
final finished device. | final finished device. | |
| Implant-to-
Abutment Interface | Internal Hex | Cone Morse | Cone Morse | Equivalent
The implant interface is equivalent to that of
the predicate. |
| Indexing | Indexed | Indexed | Indexed | Same |
| Diameter | Platform Ø: NP - 3.5
SP - 4.5 & 5.5 mm | Platform Ø: 3.5 mm, 4.5 mm | Platform Ø: 5.5 mm | Equivalent
Subject device diameters are within the range
of diameters of the predicate devices. |
| Gingival Height | NP - 0.5, 1 & 3 mm
SP - 0.5, 1 & 3 mm | 0.8, 1.5, 2.5, 3.5 & 4.5 mm | 0.8, 1.5, 2.5, 3.5, 4.5, 5.5 mm | Equivalent
Subject devices are included in the range of
the predicate device gingival heights. |
| Angulation | Straight | Straight | Straight | Same |
| CAD/CAM
Restoration
Angulation | Up to 30° | Up to 30° | Up to 30° | Same |
| CAM/CAM Material
superstructure | IPS e.max CAD
Coron (CoCr)
Zerion LT | IPS e.max CAD
Cobalt-chromium
Zirconia | IPS e.max CAD
Cobalt-chromium
Zirconia | Equivalent
The top-half material indicated for subject
device and reference predicate device are the
same. |
| Abutment Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same |
| | SUBJECT DEVICES | REFERENCE PREDICATES | | |
| | K193234 | K150367 & K153624 | K163194 | |
| COMPARISON | NUVO IF Implant System
JJGC Indústria e Comércio de Materiais
Dentários S.A. | Neodent Implant System
JJGC Indústria e Comércio de Materiais
Dentários S.A. | Neodent Implant System - GM Line
JJGC Indústria e Comércio de Materiais
Dentários S.A. | EQUIVALENCE DISCUSSION |
| Surface Treatment | Machined (SP platform)
Machined and anodized (NP platform) | Machined | | Equivalent
Anodized surface treatment is presented in
other abutments of K101945. |
| Single Use | Yes | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same |
20
21
Table 8: Substantial Equivalence Comparison – CADCAM Abutments (Rotational Titanium Base IF)
COMPARISON | SUBJECT DEVICES | REFERENCE PREDICATE | EQUIVALENCE DISCUSSION |
---|---|---|---|
K193234 | K192229 | ||
NUVO IF Implant System | Neodent Implant System - GM Titanium Base for Bridge | ||
JJGC Indústria e Comércio de Materiais Dentários S.A. | JJGC Indústria e Comércio de Materiais Dentários S.A. | ||
Rotational Titanium | |||
Base IF | Rotational Titanium Base IF | GM Titanium Base for Bridge | |
Principal of | |||
operation | Abutment designed to accept a cement or screw-retained | ||
prosthetic. | Abutment designed to accept a cement or screw-retained | ||
prosthetic. | Same | ||
Titanium Base abutments as two-piece abutments which | |||
are composed of a titanium base and a patient-specific | |||
CAD/CAM top-half, that when assembled together form | |||
the final finished device. | Titanium Base abutments as two-piece abutments which are | ||
composed of a titanium base and a patient-specific CAD/CAM | |||
top-half, that when assembled together form the final | |||
finished device. | |||
Implant-to- | |||
Abutment Interface | Internal Hex | Cone Morse | Equivalent |
The implant interface is equivalent to that of the predicate. | |||
Indexing | Non-Indexed | Non-Indexed | Same |
Diameter | Platform Ø: NP – 3.5 | ||
SP – 4.5 & 5.5 mm | Platform Ø: 3.5, 4.5, and 5.5 mm | Equivalent | |
Subject device diameters are within the range of diameters of | |||
the predicate devices or larger. Larger diameters do not | |||
represent a worst case in terms of performance. | |||
Gingival Height | NP – 0.5, 1 & 3 mm | ||
SP – 0.5, 1 & 3 mm | 0.8, 1.5, 2.5, 3.5 & 4.5 mm | Equivalent | |
Subject devices are included in the range of the predicate | |||
device gingival heights. | |||
Angulation | Straight | Straight | Same |
CAD/CAM | |||
Restoration | |||
Angulation | Up to 30° | Up to 30° | Same |
CAM/CAM material | |||
superstructure | Titanium (brand name (ticon) | Titanium (brand name ticon) | Same |
Cobalt-chromium (brand name coron) | Cobalt-chromium (brand name coron) | ||
Zirconia (brand name zerion LT) | Zirconia (brand name zerion LT) | ||
Abutment Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same |
Surface Treatment | Machined and anodized (only NP platform) | Machined | Equivalent |
Anodized surface treatment is presented in other abutments | |||
of K101945. | |||
Single Use | Yes | Yes | Same |
COMPARISON | SUBJECT DEVICES | EQUIVALENCE DISCUSSION | |
K193234 | |||
NUVO IF Implant System | |||
JJGC Indústria e Comércio de Materiais Dentários S.A. | K192229 | ||
Neodent Implant System – GM Titanium Base for Bridge | |||
JJGC Indústria e Comércio de Materiais Dentários S.A. | |||
Sterilization Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same |
SUBJECT DEVICES | REFERENCE PREDICATE | ||
COMPARISON | K193234 | K150367 | EQUIVALENCE DISCUSSION |
NUVO IF Implant System | Neodent Implant System - Titanium Base & Preface | ||
JJGC Indústria e Comércio de Materiais Dentários S.A. | JJGC Indústria e Comércio de Materiais Dentários S.A. | ||
CARES® Abutment | CARES® Abutment IF | PreFace Abutment | |
IF | |||
Principal of | Milling blank with the Implant-to-Abutment interface pre- | ||
milled at one end. Used to fabricate patient-specific | |||
abutments that are indicated with screw-retained or | |||
cement-retained single restorations | Milling blank with the Implant-to-Abutment interface pre- | ||
milled at one end. Used to fabricate patient-specific | |||
abutments that are indicated with screw-retained single | |||
restorations or cement-retained single or multi-unit | |||
restorations. | Equivalent | ||
The principal of operation of subject device is with the | |||
reference predicate device. | |||
operation | |||
Implant-to- | Internal Hex | Cone Morse | Equivalent |
The implant interface is equivalent to that of the predicate. | |||
Abutment Interface | |||
Indexing | Indexed | Indexed | Same |
Diameter | Oversize cylinder for milling 11.5 mm & 15.8 mm | ||
diameters |
Platform Ø: NP - 3.5
SP - 4.5 | Oversize cylinder for milling 11.5 mm & 15.8 mm
diameters
Platform Ø: Not defined | Same |
| Gingival Height | NP - 0.5, 1 & 3 mm
SP - 0.5, 1 & 3 mm | 0.8, 1.5, 2.5, 3.5 & 4.5 mm | Equivalent
Subject devices are included in the range of the predicate
device gingival heights. |
| Angulation | Straight | Straight | Same |
| | Angulation of milled abutment: up to 30° | Angulation of milled abutment: up to 30° | |
| Abutment Material | Titanium Alloy (Ti-6Al-4V) | Titanium Alloy (Ti-6Al-4V) | Same |
| Surface Treatment | Machined (SP platform)
Machined and anodized (NP platform) | Machined | Equivalent
Anodized surface treatment is presented in other abutment
of K101945. |
| Single Use | Yes | Yes | Same |
| Sterilization | Provided Non-Sterile | Provided Non-Sterile | Same |
| Method | Terminally sterilized by user via moist steam via
parameters validated to an SAL of 1x10-6 | Terminally sterilized by user via moist steam via parameters
validated to an SAL of 1x10-6 | Same |
22
23
Table 9: Substantial Equivalence Comparison – CADCAM Abutments (CARES® Abutment IF)
24
PERFORMANCE DATA
Dynamic fatigue test per ISO 14801 and FDA guidance entitled Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, issued on: May 12, 2004, was performed to determine the fatigue strength for the worst-case constructs assembled using the subject devices.
Sterilization of the subject endosseous dental implant devices via gamma irradiation was validated per ISO 11137-1 and ISO 11137-2.
A minimum Sterility Assurance Level (SAL) of 1 x 10 ° has been validated.
Sterilization of the subject abutments via ethylene oxide gas was validated per ISO 11135. A minimum Sterility Assurance Level (SAL) of 1 x 10 ° has been validated.
Ethylene oxide residuals have been assessed per ISO 10993-7. Residuals are within accepted limits.
Sterilization of the subject abutments via moist heat was validated per ISO 17665-2. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Biological Safety Assessment guided by ISO 10993-1 and FDA guidance entitled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" : Guidance for Industry and Food and Drug Administration Staff, issued June 16, 2016. Reference to previous biocompatibility testing is supplied as follows:
- -Cytotoxicity testing was performed per ISO 10993-5.
- -Chemical characterization was performed per ISO 10993-18.
Package transport integrity testing has been performed per ISTA 2A.
Product and package stability has been validated per ASTM F1980.
Product endotoxin testing (LAL) is performed monthly on products representative of the subject devices per ANSI/AAMI ST72.
CONCLUSION
The subject devices and the identified primary and reference predicate devices have intended use, similar designs and technological characteristics same sterilization methods and are made of the same materials. The data included in this submission demonstrate that the subject devices are substantially equivalent to the identified predicate devices.