(129 days)
The Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner. In addition these gloves was tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of medical glove to Permeation by Chemotherapy Drugs.
The subject device in this 510(k) Notification is Blue Nitrile Examination gloves, with claiming, this product contains Low Dermatitis Potential Claim and tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile compound, Blue color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
The provided text is related to the FDA 510(k) premarket notification for a medical device: "GEN 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for use with Chemotherapy Drugs Claims." It outlines the device's characteristics, intended use, and the tests performed to demonstrate its safety and effectiveness.
Here's an analysis of the acceptance criteria and the studies mentioned, structured as requested:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| General Physical Properties (ASTM D6319-10) | ||
| Overall Length (mm) | Min 230mm | Meets ASTM D6319-10 (Implies compliance with length requirements) |
| Width (±5mm) | Size S= 85mm, Size M= 95mm, Size L= 105mm, Size XL= 115mm | Meets ASTM D6319-10 (Implies compliance with width requirements) |
| Thickness at Palm (mm) | Min 0.05 mm | Meets ASTM D6319-10 (Implies compliance with palm thickness requirements) |
| Thickness at Finger Tip (mm) | Min 0.05 mm | Meets ASTM D6319-10 (Implies compliance with fingertip thickness requirements) |
| Tensile Strength (MPa) - Before Ageing | 14 min | 14 min (Explicitly stated in Section 7.2) / Meets ASTM D6319-10 |
| Ultimate Elongation (%) - Before Ageing | 500 min | 500 min (Explicitly stated in Section 7.2) / Meets ASTM D6319-10 |
| Tensile Strength (MPa) - After Aging | 14 min | 14 min (Explicitly stated in Section 7.2) / Meets ASTM D6319-10 |
| Ultimate Elongation (%) - After Aging | 400 min | 400 min (Explicitly stated in Section 7.2) / Meets ASTM D6319-10 |
| Freedom from Holes (ASTM D5151-06) | ||
| Pin-hole Level | AQL 2.5, Inspection Level G-1 | AQL 2.5, Inspection Level G-1 (Explicitly stated in Section 7.2) / Meets ASTM D5151-06 |
| Powder-free residue (ASTM D6124-06) | ||
| Residual Powder | ≤ 2.0 mg/glove | Meets ASTM D6124-06 (Implies compliance) |
| Biocompatibility (ISO 10993-10, ISO 10993-5) | ||
| Primary Skin Irritation | The test article was a non-irritant. | Test article is non-irritant |
| Dermal Sensitization | The test article was a non-sensitizer. | Test article is non-sensitizer |
| In vitro Cytotoxicity | The test article was not cytotoxic. | The test article was not cytotoxic (Stated in Section 9.0 Summary) |
| Resistance against Chemotherapy Drugs (ASTM D6978-05) | ||
| Carmustine (3.3mg/ml) | 22 min breakthrough | 22 min breakthrough |
| Cisplatin (1.0mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Cyclophosphamide (20mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Methotrexate (25mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Doxorubicin HCl (2.0mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Etoposide (20.0mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Fluorouracil (50.0mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Paclitaxel (6.0mg/ml) | No breakthrough up to 240 min | No breakthrough up to 240 min |
| Thiotepa (10.0mg/ml) | 47.4 min breakthrough | 47.4 min breakthrough |
| Low Dermatitis Potential Claim | ||
| Modified Draize 95 test | No clinical evidence of residual chemical additives that may induce Type IV allergy in human subjects | Test article did not induce type IV allergy to human; 205 non-sensitized adult human subjects gave all negative results. |
2. Sample Size Used for the Test Set and the Data Provenance
-
Low Dermatitis Potential Claim (Clinical Study):
- Sample Size:
- Stage 1: 30 human subjects
- Stage 2 (Total): Minimum 205 individuals (including the initial 30). The study was completed on 205 adult human subjects.
- Data Provenance: The document states the subjects "reasonably reflect the general user population in the US." No explicit country of origin for the data collection is specified beyond this, but the company is in Malaysia, suggesting the study might have been conducted there or in the US with a US-representative population. The study is prospective, as it involves human testing for irritation and sensitization.
- Sample Size:
-
Chemotherapy Drug Permeation and other ASTM/ISO tests: The document does not specify the sample size for these non-clinical tests. These tests are typically conducted on a representative number of glove samples according to the respective standard (e.g., a batch of gloves). The data provenance is industrial testing, likely from the manufacturer's quality control processes in Malaysia, as it refers to meeting ASTM standards rather than specific patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- For the clinical study (Low Dermatitis Potential Claim): The document does not explicitly state the "number of experts" or their specific "qualifications" involved in establishing the ground truth of skin irritation/sensitization. The ground truth would be based on the clinical observations of the human subjects by qualified medical professionals conducting the Modified Draize 95 test.
- For non-clinical tests (ASTM/ISO): Ground truth is established by the test methods specified in the standards themselves, not typically by human experts reviewing individual cases. The results are quantitative measurements against defined criteria.
4. Adjudication Method for the Test Set
- For the clinical study (Low Dermatitis Potential Claim): The document does not specify an adjudication method like 2+1 or 3+1. The results are presented as direct observations ("all negative results," "no clinical evidence of... Type IV allergy"). This suggests direct observation and interpretation by the clinicians conducting the study.
- For non-clinical tests: Adjudication methods like those for interpreting medical images are not applicable here. The tests involve direct measurement against a standard.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This device is a nitrile examination glove, not an AI-powered diagnostic algorithm or imaging device that would typically involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This refers to a standalone performance study of the device itself.
- Yes, for the physical and chemical properties: The ASTM and ISO standard tests (e.g., tensile strength, pinholes, chemotherapy drug permeation, biocompatibility) are standalone evaluations of the glove material and design performance. They are conducted without human intervention in the loop other than performing the test itself.
- Yes, for the clinical dermatological claim: The Modified Draize 95 Test is a standalone assessment of the glove's potential to cause skin irritation or sensitization on human subjects.
7. The Type of Ground Truth Used
- Low Dermatitis Potential Claim: Clinical observation / Outcomes data (absence of irritation/sensitization in human subjects after exposure).
- Chemotherapy Drug Resistance: Laboratory measurement against established permeation thresholds (quantitative data defined by ASTM D6978-05).
- Physical Properties (Tensile Strength, Elongation, Dimensions, Pin-holes): Laboratory measurement against specified physical standards (quantitative data defined by ASTM D6319-10 and ASTM D5151-06).
- Powder-free Residue: Laboratory measurement (quantitative data defined by ASTM D6124-06).
- Biocompatibility (Cytotoxicity): Laboratory measurement (quantitative or qualitative data defined by ISO 10993-5).
8. The Sample Size for the Training Set
The concept of a "training set" is not applicable here as this is not an AI/machine learning device. The studies described are for validation/testing of the final product.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of medical device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2020
ECO Medi Glove SDN BHD Suresh Kumar QA Manager Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, Perak, 34600 My
Re: K193121
Trade/Device Name: GEN 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for use with Chemotherapy Drugs Claims Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: December 14, 2020 Received: December 23, 2020
Dear Suresh Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elizabeth Claverie-Williams, MS Assistant Director, THT4B2: Disinfection, Reprocessing and Personal Protection DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration | Form Approved: OMB No. 0910-0120Expiration Date: 06/30/2020 |
|---|---|
| Indications for Use | See PRA Statement below |
510(k) Number (if known)
Device Name
Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with Tested for use with Chemotherapy Drugs Claims
Indications for Use (Describe)
The Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner. In addition these gloves was tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of medical glove to Permeation by Chemotherapy Drugs.
| Chemotherapy Drugs and Concentration | Minimum Breakthrough detection time in minutes(µg/cm²/minute) |
|---|---|
| 1)Carmustine (BCNU)(3.3 mg/ml)(3,300ppm) | 22 |
| 2)Cisplatin (1.0 mg/ml)(1,000ppm) | No breakthrough up to 240 min |
| 3)Cyclophosphamide (Cytoxan)(20 mg/ml)(20,000ppm) | No breakthrough up to 240 min |
| 4)Methotexale (25mg/ml)(25,000ppm) | No breakthrough up to 240 min |
| 5)Doxorubicin Hydrochloride (2.0 mg/ml)(2,000ppm) | No breakthrough up to 240 min |
| 6)Etoposide (Toposar)(20.0 mg/ml)(20,000ppm) | No breakthrough up to 240 min |
| 7)Fluorouracil (50.0 mg/ml)(50,000ppm) | No breakthrough up to 240 min |
| 8)Paclitaxel (Taxol)(6.0 mg/ml)(6,000ppm) | No breakthrough up to 240 min |
| 9)Thiotepa (10.0 mg/ml)(10,000ppm) | 47.4 |
The Maximum testing time is 240 minutes. Please note that the following drugs have low permeation time:
-
Carmustine (BCNU)(3.3mg/ml)with Permeation time of 22 minutes.
-
Thiotepa ( 10.0mg/ml)with Permeation time of 47.4 minutes.
Type of Use (Select one or both, as applicable)
| ▢ Prescription Use (Part 21 CFR 801 Subpart D) | ☒ Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| -------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|---|---|
| ---------------------- | ------------- |
{3}------------------------------------------------
510(K) Summary Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and Tested for use with Chemotherapy Drugs Claims
1.0 Submitter:
| Company Name: | ECO Medi Glove Sdn Bhd. |
|---|---|
| Company Address: | Lot 23826,Jalan Tembaga Kuning,Kamunting Raya Industrial Estate,34600,TaipingPerak,Malaysia. |
| Contact Person: | Mr Suresh Kumar |
| Telephone No: | 603-60283033 |
| Email: | qa1@riverstone.com.my |
2.0 Preparation Date: 16th March 2020
3.0 Device Identification:
Trade Name / Proprietary Name: Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for Use with Chemotherapy drugs Claims.
Device Name: Nitrile Patient Examination gloves
Device Classification Name: Patient Examination gloves (21 CFR 880.6250)
Device Class: Class I
Product Code: LZA, LZC
{4}------------------------------------------------
4.0 Predicate Device:
Class I patient Examination glove with claiming, this product contain Low Dermatitis Potential Claim and tested for use with Chemotherapy Drugs, Powder Free, LZC, which meets all the requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250.
Predicate Device: K152542, Powder Free Nitrile Examination (Blue) with Low Dermatitis Potential Claim with tested for use with Chemotherapy Drugs
5.0 Device Description:
The subject device in this 510(k) Notification is Blue Nitrile Examination gloves, with claiming, this product contains Low Dermatitis Potential Claim and tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile compound, Blue color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
6.0 Intended use of the Device:
Gen 2 patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of Medical Glove to Permeation by chemotherapy Drugs. It is for over- the-counter use.
7.0 Specification for Nitrile gloves:
7.1 Dimension and Thickness of Gloves
| Dimension | Size S | Size M | Size L | Size XL |
|---|---|---|---|---|
| Overall Length (mm) | 230min | 230min | 230min | 230min |
| Width (± 5mm) | 85 | 95 | 105 | 115 |
| Thickness at Palm (mm) | 0.05min | 0.05min | 0.05min | 0.05min |
| Thickness at Finger Tip (mm) | 0.05min | 0.05min | 0.05min | 0.05min |
{5}------------------------------------------------
7.2 Gloves Physical Properties and Holes
| Measurement | Before Ageing | After Aging at 70℃ for168 hrs @ 100℃ for 22 hrs |
|---|---|---|
| Tensile Strength (MPa) | 14min | 14 Min |
| Ultimate Elongation (%) | 500min | 400min |
| Pin-hole Level | AQL 2.5Inspection Level G-1 | AQL 2.5Inspection Level G-1 |
Gloves meet all the specification listed in ASTM D 6319-10
7.3 Technological Characteristics Comparison Table:
| Characteristics | Acceptance Criteria | Subject | Predicate | Comparison |
|---|---|---|---|---|
| Gen 2 Nitrile Examination Gloves (Blue) with Low Dermatitis Potential Claim and with tested for use with chemotherapy drugs Claim, (K193121) | Powder Free Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim and with tested for used with Chemotherapy Drugs, (K152542) | |||
| Product Code | LZA, LZC | LZA, LZC | LZA, LZC | same |
| Intended use | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the- counter use. | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over- the-counter use. | same |
| Material use | Nitrile compound | Nitrile compound | Nitrile compound | same |
| Colour | Blue | Blue | Blue | same |
| Sterility | Non sterile | Non sterile | Non sterile | same |
| Single used | Single used | Single used | Single used | same |
| Non Sterile | Non Sterile | Non Sterile | Non Sterile | same |
{6}------------------------------------------------
| Dimensions | Overall Length (mm)Min 230mm Width(±5mm)Size S= 85mmSize M =95mmSize L= 105mm SizeXL = 115mmThickness at Palm (mm) Min; 0.05mm Thickness at Finger Tip (mm)Min 0.05 mm | Meets ASTM D6319- 10 | Meets ASTM D6319-10 | same |
|---|---|---|---|---|
| Physical properties | Before Ageing Tensile Strength(MPa)= 14min UltimateElongation (%)= 500min After Aging at 70°C for168 hrs @ 100°C for 22 hrsTensile Strength (MPa)= 14min Ultimate Elongation(%)= 400min | Meets ASTMD6319-10 | Meets ASTMD6319-10 | same |
| Freedom from pinholes | AQL 2.5Inspection Level G-1 | Meets ASTM D5151-06 | Meets ASTMD5151-06 | same |
| Residual Powder | ≤ 2.0 mg/glove | Meets ASTMD6124-06 | Meets ASTMD6124-06 | same |
| Biological Evaluation onMedical Device - Part 10-Primary Skin IrritationTest | Under the conditions of thisstudy, the test article was anon- irritant. | Under the conditions of thisstudy, the testarticle was a non- irritant. | same | |
| Biological Evaluation onMedical Device - Part 10-Dermal SensitizationAssay | Under the conditions of thisstudy, the test article was anon- sensitizer. | Under the conditions of thisstudy, the test article was a non-sensitizer. | same | |
| Biological Evaluation onMedical Device – Part 5-Tests for in vitroCytotoxicity | Under the conditions of thisstudy, the test article was notcytotoxic. | NA | Different | |
| Resistance againstChemotherapy Drugs | 1) Carmustine (3.3mg/mlor 3000ppm),Breakthrough : 22 min. | 1) Carmustine (3.3mg/ml or3000ppm), Breakthrough : 20.1min. | similar | |
| 2) Cyclophosphamide(20mg/ml or 20,000ppm),Breakthrough time :>240 min. | 2)Cyclophosphamide (20mg/ml or20,000ppm),Breakthrough time :>240 min. | |||
| 3) Cisplatin (1mg/ml or1000ppm), Breakthroughtime :> 240 min | 3) Cisplatin (1mg/ml or1000ppm), Breakthrough time :>240 min | |||
| 4)Doxorubicin Hydrochloride(2.0mg/ml or 2000ppm),Breakthrough time :>240 min. | 4)Doxorubicin Hydrochloride(2.0mg/ml or 2000ppm),Breakthrough time :>240 min. | |||
| 5) Etoposide (20mg/ml or20,000ppm), Breakthroughtime :>240 min. | 5) Etoposide (20mg/ml or20,000ppm), Breakthrough time :>240 min. | |||
| 6) Flourouracil (50mg/ml or50,000), Breakthrough time :>240 min. | 6) Flourouracil (50mg/ml or50,000), Breakthrough time :>240 min. | |||
| 7) Methotexate (25mg/ml or 25,000ppm), Breakthrough time:> 240 min. | 7) Methorexate (25mg/ml or 25,000ppm), Breakthrough time :> 240 min. | |||
| 8) Paclitaxel (6mg/ml or 6,000ppm), Breakthrough time:>240 min. | 8) Paclitaxel (6mg/ml or 6,000ppm), Breakthrough time:>240 min | |||
| 9) Thiotepa (10mg/ml or 10,000ppm), Breakthrough time:47.4 min. | 9) Thiotepa (10mg/ml or 10,000ppm), Breakthrough time:50.6 min. | |||
| Low DermatitisPotential Claim | 1)ModifiedDraize 95 test | No Clinical evidencepresence of residualchemical additivesthat may induce TypeIV allergy in humansubject | No Clinicalevidence presenceof residualchemical additivesthat may induceType IV allergy inhuman subject | similar |
{7}------------------------------------------------
{8}------------------------------------------------
Resistance against Chemotherapy Drugs
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection time(Specimen 1/2/3)(Minutes) |
|---|---|
| 1)Carmustine (BCNU) (3.3mg/ml)(3,300 ppm) | 22 |
| 2) Cisplatin (1.0mg/ml)(1,000 ppm) | No breakthrough up to 240 min |
| 3)Cyclophosphamide (Cytoxan)(20mg/ml)(20,000 ppm) | No breakthrough up to 240 min |
| 4)Methotexate(25.0mg/ml)(25,000 ppm) | No breakthrough up to 240 min |
| 5)Doxorubicin Hydrochloride ( 2.0mg/ml)(2,000 ppm) | No breakthrough up to 240 min |
| 6)Etoposide (Toposar)(20.0mg/ml)(20,000 ppm) | No breakthrough up to 240 min |
| 7)Fluorouracil (50.0mg/ml)(50,000 ppm) | No breakthrough up to 240 min |
| 8)Paclitaxel (Taxol)(6.0mg/ml)(6,000 ppm) | No breakthrough up to 240 min |
| 9)Thiotepa(10.0mg/ml)(10,000 ppm) | 47.4 |
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time. Carmustine (BCNU) (3.3mg/ml) Thiotepa (10mg/ml)
{9}------------------------------------------------
8.0) Summary of Non-Clinical Testing:
| Test Method | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTM D5151-06 | Freedom fromHoles | AQL 2.5Inspection Level G-1 | Meets ASTM D6319-10 |
| ASTM D6319-10 | Dimensions,Physical Properties | ASTM D6319- 10 | Meets ASTM D6319-10 |
| ASTM D6124-06 | Powder-freeresidue | < 2.0 mg/glove | Meets ASTM D6319-10 |
| Primary skinirritation - ISO10993-10Skin Sensitization- ISO 10993-10In vitroCytotoxicity –ISO 10993-5 | Bio-compatibility | 1) The test article was a non-irritant.2) The test article was a non-sensitizer3) The test article was notcytotoxic | Test article is nonirritant and Nonsensitizer and notcytotoxic |
| ASTMD 6978-05 | Chemotherapy Drugtest | As per test report | As labeling claim |
| Modify Draize 95Test | Low DermatitisPotential Claim | Induce TypeIV allergy in human | Test article do notinduce type IVallergy to human |
9.0) Summary of Clinical Testing:
Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and tested for Use with Chemotherapy drugs were tested in accordance with Modified Draize -95 test, per FDA's guidance document "Guidance for Industry and FDA Reviewer/Staffs: Premarket Notification [510k] Submissions for testing for skin sensitization to chemical in natural Rubber Products".
The study was conducted in two stages. In the first, a population of 30 human subjects was tested to evaluate the product for the potential to cause irritation or sensitization. The second stage was initiated on a further number of subjects to a total of a minimum 205 individuals after the first stage has shown that the test product does not indicate a potential for inducing dermal irritation and does not shown sensitization capability
The study completed on 205 non sensitized adult human subjects, who reasonably reflect the general user population in the US, gave all negative results. There was no clinical evidence of the presence of residual chemical additives at the level that may induce type IV allergy in the un-sensitized general user population in the tested articles
10.0 Conclusion:
The conclusion drawn from the nonclinical and clinical test demonstrate that the subject device in 510(k) K193121, Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim and tested for Use with Chemotherapy drugs is as safe, as effective and performs as well as or better than the legally marketed predicate device (K15254)
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.