K Number
K192864
Date Cleared
2020-06-01

(238 days)

Product Code
Regulation Number
870.5150
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.

Device Description

The proposed UNI*FUSE™ Infusion System with Cooper Wire devices are single-lumen, nylon catheters with longitudinal slits located at 90° intervals around the distal end of the catheter for fluid delivery. An occluding ball wire (or occlusion guidewire) provides end-hole occlusion during fluid delivery. The catheters are available in 4F and 5F diameters and overall lengths of 90cm and 135cm. Additionally, they are available in multiple infusion segment lengths, including 2cm and 5cm lengths. The active infusion area can be identified under imaging by means of radiopaque markers on the catheter shaft at the distal and proximal ends of the infusion segment.

AI/ML Overview

This document is a 510(k) Summary for a medical device called the UNI*FUSE™ Infusion System with Cooper Wire. It demonstrates the substantial equivalence of the proposed device to a predicate device, which is a common pathway for FDA clearance for Class II medical devices.

The information provided does not describe an acceptance criteria table or a study that proves the device meets specific performance criteria in the way typically seen for a novel or AI-powered diagnostic device. Instead, it focuses on demonstrating substantial equivalence to a previously cleared device. This means the manufacturer is showing that their modified device is as safe and effective as a legally marketed predicate device, rather than proving performance against pre-defined quantitative acceptance criteria through a specific clinical study with human readers and ground truth, as would be the case for an AI/ML medical device.

The "studies" described here are related to proving the safety and effective use of the device when its intended use is expanded (specifically, to include the pulmonary artery vasculature). This is achieved through:

  1. Comparison to a Predicate Device: Showing that the modified device has substantially equivalent technological characteristics and performance to a legally marketed predicate device.
  2. Bench Testing and Human Factors Study: Verifying that the device can navigate, be placed, and function as intended in the newly added anatomical location (pulmonary artery).
  3. Clinical Literature Review: Evaluating existing literature to demonstrate that the expanded indication (use in the pulmonary artery) does not introduce new risks or increase existing risks compared to other anatomical locations or other types of catheters.

Therefore, providing a table of acceptance criteria and reported device performance in the context of an AI/ML study, as requested in the prompt, is not directly applicable to this document. The "performance" being evaluated here is related to the physical and functional aspects of a catheter and its safe deployment, not the diagnostic accuracy of an algorithm.

However, I can interpret the request in the spirit of the prompt by extracting the closest analogous information to "acceptance criteria" and "study results" as presented in this 510(k) summary:


Interpretation for a Non-AI Device Substantial Equivalence (as per the provided document):

The acceptance criteria for this device are not framed as quantitative metrics for diagnostic output (like sensitivity, specificity for an AI model) but rather as qualitative demonstrations of safety and efficacy compared to a predicate device and for the expanded indication. The "study" described is a combination of:

  • Comparative Assessment: Against a predicate device (Bashir™ N-X Endovascular Catheter Model 7200, K183290) and a reference device (PULSESPRAY™ and UNIFUSE™ Infusion Catheters, K163356).
  • Human Factors Study & Bench Testing: To support the expanded indication to the pulmonary artery.
  • Clinical Literature Review: To support the safety profile for the expanded indication.

1. Table of "Acceptance Criteria" (interpreted as aspects needing demonstration of substantial equivalence or safety/efficacy for new indication) and "Reported Device Performance" (interpreted as the outcome of the comparative assessment or studies):

Aspect of Evaluation (Analogous to Acceptance Criteria)Reported Device Performance / Conclusion from Study
I. Substantial Equivalence to Predicate Device (K183290)
ProCode, Regulation Number, Regulation Name, Regulatory ClassIdentical to predicate device.
Indications for Use / Intended UseSubstantially Equivalent to predicate. Both are for administration of fluids into peripheral and pulmonary artery vasculature. (Predicate: "controlled and selective infusion of physician-specified fluids"; Proposed: "administration of fluids, including thrombolytic agents and contrast media").
Operating Principle, Mechanism of Action, Patient PopulationsSame as predicate device.
Overall Design (General features)Both are single-lumen nylon catheters with longitudinal slits and an occluding ball wire.
Catheter Diameter (F)Proposed: 4F, 5F. Predicate: 7F. Substantially Equivalent (Acknowledged difference in specific dimensions, but overall function is similar).
Catheter Length (cm)Proposed: 90cm, 135cm. Predicate: 92.5cm. Substantially Equivalent.
Catheter Infusion Segment Length (cm)Proposed: 2cm, 5cm. Predicate: 12.5cm. Substantially Equivalent.
MaterialsBoth use commonly used, biocompatible materials per ISO 10993-1. Substantially Equivalent.
Performance Testing (Functional equivalence of tests)Batteries of testing for proposed and predicate device are not 1:1 identical but are substantially equivalent in that they fully test the functions of each device (e.g., flow rate, pressure, compatibility, durability, kink resistance).
II. Support for Expanded Indication (Pulmonary Artery Vasculature)
Ability to Navigate and Use in Pulmonary ArteryHuman Factors Study & In-House Bench Test: Demonstrated ability to navigate-to and use within the pulmonary artery vasculature.
Visibility under Imaging (Fluoroscopy)Human Factors Study & In-House Bench Test: Infusion segments are clearly imaged under fluoroscopy, confirming position within pulmonary artery.
Adequacy of Directions for Use (DFU)Human Factors Study: DFU provides adequately detailed instructions for accurate and reliable placement and use in pulmonary artery.
Risk Profile in Pulmonary Artery vs. Other VesselsClinical Literature Evaluation: No evidence indicating pulmonary artery is more prone to damage than other vessels for this application. Proposed Indications for Use do not increase existing risks or introduce new risks. Catheters with mechanical components may have greater risk of vessel damage compared to non-mechanical infusion catheters. The UNIFUSE™ is smaller than some other larger catheter types, suggesting a "worst-case scenario" for intrusiveness and potential damage is less with UNIFUSE™. The clinical literature supports catheter-directed therapy in the pulmonary artery vasculature as having high success rates and complications not including vessel injury/damage.

2. Sample Size and Data Provenance:

  • Test Set (for Human Factors Study): "multiple practicing physicians" (specific number not given)
  • Data Provenance: The human factors study involved "simulated-use vascular model" and practicing physicians. The clinical literature review uses publicly available articles, presumably from global or diverse sources, but the document does not specify country of origin for the underlying patient data. The studies are described as both retrospective and prospective in the text (e.g., Kuo, W. et al. – 2009 for clinical literature review).

3. Number of Experts and Qualifications:

  • Human Factors Study: "multiple practicing physicians experienced in the use and placement of infusion catheters." (Specific number and detailed qualifications/experience not specified beyond "practicing physicians"). These experts essentially acted as the "readers" evaluating the device's usability in the simulated environment.
  • Clinical Literature Review: No specific experts were used to establish "ground truth" here; rather, the company (Angiodynamics, Inc.) performed the literature review to support their conclusions on safety.

4. Adjudication Method for Test Set:

  • Human Factors Study: Not explicitly stated. The study aimed to demonstrate usability and navigability. It's implied that the observations from the "multiple practicing physicians" were collectively assessed to reach the stated conclusions about navigability, imaging clarity, and DFU adequacy. It was not a comparative diagnostic reading study.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

  • No, not applicable. This is a physical medical device (catheter), not an AI/ML diagnostic algorithm. There was no MRMC study to compare human readers with and without AI assistance, as AI is not a component of this device.

6. Standalone (Algorithm Only) Performance:

  • Not applicable. This is a physical medical device. No algorithm is being assessed.

7. Type of Ground Truth Used:

  • Human Factors Study & Bench Testing: The "ground truth" was established by the physical and simulated performance of the device itself (e.g., was it navigable, was it visible under fluoroscopy, could fluids be administered successfully). This is a functional and usability "ground truth."
  • Clinical Literature Review: The ground truth was based on the published findings and conclusions of peer-reviewed medical literature concerning catheter-directed interventions in the pulmonary artery, particularly regarding complications and safety.

8. Sample Size for Training Set:

  • Not applicable. This device is not an AI/ML algorithm that requires a training set.

9. How Ground Truth for Training Set was Established:

  • Not applicable. No training set for an AI/ML algorithm was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."

November 22, 2021

Angiodynamics, Inc. Brandon Brackett Manager, Global Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752

Re: K192864

Trade/Device Name: UNI*FUSE Infusion System with Cooper Wire Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA

Dear Brandon Brackett:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 1, 2020. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.

Sincerely,

Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2021.11.22 13:42:55 -05'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 1, 2020

Angiodynamics, Inc. Brandon Brackett Manager, Global Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752

Re: K192864

Trade/Device Name: UNI*FUSE™ Infusion System with Cooper Wire Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: April 2, 2020 Received: April 3, 2020

Dear Mr. Brackett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{2}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/5 description: The image shows the name "Gregory W. O'connell -S" in a large, bold font. The text is black and appears to be part of a document or publication. The background is a light blue color.

Digitally signed by Gregory W. O'connell -S Date: 2020.06.01 13:44:17 -04'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K192864

Device Name

UNI*FUSE™ Infusion System with Cooper Wire

Indications for Use (Describe)

The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary For the Angiodynamics, Inc. UNI*FUSE™ Infusion System with Cooper Wire

Date Prepared: June 1, 2020

Sponsor

Angiodynamics, Inc. 26 Forest Street Marlborough, MA 01752 USA

Regulatory Contact

Brandon M. Brackett, RAC

Senior Manager, Global Regulatory Affairs AngioDynamics, Inc. Telephone: 1-508-658-7984 Facsimile: 1-508-658-7976 Email: bbrackett@angiodynamics.com

Subject Device

  • Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:

Predicate Device

  • 510(k) Reference Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:

Reference Device

  • 510(k) Reference Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
    Angiodynamics, Inc. UNI*FUSE™ Infusion System with Cooper Wire Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices

K183290

Thrombolex, Inc. Bashir™ and Bashir™ N-X Endovascular Catheters Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices

K163356

Angiodynamics, Inc. PULSESPRAY™ and UNIFUSE™ Infusion Catheters Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices

{5}------------------------------------------------

Purpose

The intent of this Traditional 510(k) is to propose a modification to the UNIFUSE™ Infusion System with Cooper Wire Indications for Use statement, incorporative of a clarification being made in order to specify the pulmonary artery as a vessel indicated for device access. The UNIFUSE™ Infusion System with Cooper Wire overall design is not changing as a result of this clarification is being made in response to the clinical use of Infusion Catheters in the pulmonary artery vasculature, identified as a usage primarily via available medical literature.

Device Description

The proposed UNI*FUSE™ Infusion System with Cooper Wire devices are single-lumen, nylon catheters with longitudinal slits located at 90° intervals around the distal end of the catheter for fluid delivery. An occluding ball wire (or occlusion guidewire) provides end-hole occlusion during fluid delivery. The catheters are available in 4F and 5F diameters and overall lengths of 90cm and 135cm. Additionally, they are available in multiple infusion segment lengths, including 2cm and 5cm lengths. The active infusion area can be identified under imaging by means of radiopaque markers on the catheter shaft at the distal and proximal ends of the infusion segment. As shown in Table 4 below, all of these characteristics are substantially equivalent to those of the predicate devices. Additionally, all of these characteristics are identical when comparing the proposed to the reference devices, as shown in Table 5.

The proposed UNI*FUSE™ Infusion System with Cooper Wire devices are intended for administration of fluids such as thrombolytics and contrast media into vessels that are impacted by thrombus, including the peripheral and pulmonary artery vasculature. Given the minimal physical differences of the vessels for which the device is intended to access (e.g. diameter, structure, tortuousness), the operating principle mechanism of action, and intended use is the same independent of anatomical location, use in the pulmonary artery is equivalent to other areas of the vasculature that the device is currently indicated for.

Indications for Use/Intended Use

The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.

Comparison of Similarities and Differences in Technological Characteristics and Performance

The proposed UNI*FUSE™ Infusion System with Cooper Wire and the predicate Bashir™ N-X Endovascular Catheter Model 7200 are substantially equivalent to one another in terms of design, materials of manufacture, specifications, dimensions, Indications for Use, and sizes and/or configurations, as depicted in the comparison via Table 4 below:

Table 4: Comparison of Similarities and Differences in Technological Characteristics and Performance
Proposed UNI*FUSE™ Infusion System with Cooper Wire vs.
Predicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290)
Proposed UNI*FUSE™ Infusion System with Cooper WirePredicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290)Comparison
ProCodeKRAKRAIdentical
Regulation Number21CFR870.121021CFR870.1210Identical
Regulation NameCatheter, Continuous FlushCatheter, Continuous FlushIdentical
Regulatory ClassClass 2Class 2Identical

{6}------------------------------------------------

Table 4: Comparison of Similarities and Differences in Technological Characteristics and Performance
Proposed UNI*FUSE™ Infusion System with Cooper Wire vs.Predicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290)
Proposed UNI*FUSE™ InfusionSystem with Cooper WirePredicate Bashir™ N-XEndovascular Catheter Model7200 (K183290)Comparison
Indications forUseThe UNI*FUSE™ InfusionSystem with Cooper Wire isintended for the administrationof fluids, including thrombolyticagents and contrast media, intothe peripheral and pulmonaryartery vasculature.The Bashir™ N-X EndovascularCatheter is intended for thecontrolled and selectiveinfusion of physician-specifiedfluids into the peripheral andpulmonary artery vasculature.Substantially Equivalent
CatheterDiameter (F)4F, 5F7FSubstantially Equivalent
Catheter Length(cm)90cm, 135cm92.5cmSubstantially Equivalent
Catheter InfusionSegment Length(cm)2cm, 5cm12.5cmSubstantially Equivalent
MaterialsAll materials are commonlyused for this type of medicaldevice and are biocompatible inaccordance with ISO 10993-1.All materials are commonlyused for this type of medicaldevice and are biocompatible inaccordance with ISO 10993-1.Substantially Equivalent

Additionally, these attributes are identical when comparing the proposed UNIFUSE™ Infusion System with Cooper Wire to the reference UNIFUSE™ Infusion Catheter, as shown in Table 5 below:

Table 5: Comparison of Similarities and Differences in Technological Characteristics and Performance
Proposed UNI*FUSE™ InfusionSystem with Cooper WireReference UNI*FUSE™ InfusionCatheter (K163356)Comparison
ProCodeKRAKRAIdentical
RegulationNumber21CFR870.121021CFR870.1210Identical
Regulation NameCatheter, Continuous FlushCatheter, Continuous FlushIdentical
Regulatory ClassClass 2Class 2Identical
Indications forUseThe UNI*FUSE™ InfusionSystem with Cooper Wire isintended for the administrationof fluids, including thrombolyticagents and contrast media, intothe peripheral and pulmonaryartery vasculature.The UNI*FUSE™ InfusionSystem is intended for theadministration of fluids,including thrombolytic agentsand contrast media, into theperipheral vasculature.Identical(Aside from proposedmodification)
CatheterDiameter (F)4F, 5F4F, 5FIdentical
Catheter Length(cm)90cm, 135cm45cm, 90cm, 135cmSubstantially Equivalent
Catheter InfusionSegment Length(cm)2cm, 5cm2cm, 5cm, 10cm, 15cm, 20cm,30cm, 40cm, 50cmSubstantially Equivalent

{7}------------------------------------------------

Proposed UNI*FUSE™ InfusionSystem with Cooper WireReference UNI*FUSE™ InfusionCatheter (K163356)Comparison
MaterialsAll materials are commonlyused for this type of medicaldevice and are biocompatible inaccordance with ISO 10993-1.All materials are commonlyused for this type of medicaldevice and are biocompatible inaccordance with ISO 10993-1.Identical

Furthermore, the proposed UNIFUSE™ Infusion System with Cooper Wire devices and predicate Bashir™ N-X Endovascular Catheter Model 7200's are incorporative of the same operating principle, mechanism of action, and intended use as one another (in addition to exhibiting substantial equivalence in terms of the overall design, materials of manufacture, sizes/configurations, and Indications for Use as shown above). Lastly, there are no changes being made to the UNIFUSE™ Infusion System with Cooper Wire devices as compared to that cleared via reference 510(k) K163356. As a result, all previous biocompatibility, shelflife, performance, and other testing identified within K163356 remains applicable to the proposed version.

Comparison of Performance Data

Angiodynamics, Inc. has also compared the performance testing that the Bashir™ N-X Endovascular Catheter Model 7200 was subjected to in support of its clearance, to that which the proposed UNI*FUSE™ Infusion System with Cooper Wire has been previously subjected to. While each battery of testing is not identical in a 1:1 nature, they are substantially equivalent to one another in that they fully test the functions of each device. Please refer to Table 6, below:

Table 6: Comparison of Performance Testing
Proposed UNI*FUSE™ Infusion System with Cooper Wire vs. Bashir™ N-X Endovascular Catheter Model 7200 (K183290)
Proposed UNI*FUSE™ Infusion System with Cooper WireBashir™ N-X Endovascular Catheter Model 7200 (K183290)
-Dimensional Verification-Kink Radius
-Length Sufficiency-Trackability
-Catheter Hub-to-Catheter Shaft Connection-Advancement Force
Compatibility-Slider Actuator
-Catheter-to-Guidewire Compatibility-Catheter Retraction
-Catheter-to-Introducer Sheath Compatibility-Radial Force
-Catheter Tip Radius-Delivery Flow-Rate
-Catheter Infusion-Infusion Pressure at Various Flow Rates
-Slit Pattern Radiopacity-Infusion Pressure with Pulse Spray
-Catheter Degradation-Pressure Measurement Through Central Lumen
-Catheter Pressure-Guidewire Compatibility
-Catheter-to-Occlusion Wire Configuration (Slow Infusion-Dimensional Verification
Compatibility)-Compliance of Injection Hubs
-Catheter/Accessory Compatibility-Air Leakage
-Catheter/Fluid Compatibility-Fluid Leakage
-Catheter Hub-to-Shaft Joint Kink Resistance-Stress Cracking
-Occlusion Wire Flexibility-Resistance to Separation
-Occlusion Wire Flow-Torque Strength
-Occlusion Wire Seal-Corrosion Resistance
-Hub-to-Wire Bond/Connection-Joint Tensile Strength
-Distal Spring Tip-to-Mandrel Connection-Particulate Generation
-Occlusion Wire Withdrawal

{8}------------------------------------------------

In addition to the performance testing summarized above, Angiodynamics, Inc. has also conducted a human factors study on the proposed UNIFUSE™ Infusion System with Cooper Wire. Specifically, Angiodynamics, Inc. solicited the participation of multiple practicing physicians experienced in the use and placement of infusion catheters to evaluate the proposed UNIFUSE™ Infusion System with Cooper Wire on a simulated-use vascular model. The results of this testing demonstrated the following:

  • . The UNI*FUSE™ Infusion System with Cooper Wire configurations proposed via this submission are able to be navigated-to and used-within the pulmonary artery vasculature;
  • . The infusion segments of the UNI*FUSE™ Infusion System with Cooper Wire configurations proposed via this submission are able to be clearly imaged under fluoroscopy (and thus confirmed to reside completely within the pulmonary artery vasculature);
  • . The Directions for Use proposed for the UNI*FUSE™ Infusion System with Cooper Wire provides adequately detailed instructions in order to enable users to accurately and reliably place and use the device(s) within the pulmonary artery vasculature.

As a means of further validating the conclusions of the human factors study, Angiodynamics, Inc. also conducted an in-house "bench test" version of the study. The in-house testing corroborated those of the human factors study summarized in the dialogue above, and the cumulative results demonstrate that the proposed UNM*FUSE™ Infusion System with Cooper Wire configurations are able to navigate the vasculature, be placed and used within the pulmonary artery, and confirm their location under imaging (e.g. fluoroscopy). These determinations further justify the Indications for Use modification proposed via this submission.

Clinical Literature Evaluation and Determinations

AngioDynamics, Inc. has assessed a variety of publicly available articles and other literature to identify instances of vasculature damage/endothelial cell destruction resulting from the use of various catheters (including infusion catheters) in the pulmonary artery vasculature. While the articles have varying endpoints, each study does document the other possible effects the device may pose during use (i.e. risks, complications, and other consequences) such that the risks and complications related to given devices and/or therapies are also known. None of the risks and complications identified related to pulmonary artery (or other vessel) damage and/or injury. This leads AngioDynamics, Inc. to conclude that there is a lack of evidence indicating the pulmonary artery vasculature to be more markedly prone to damage than other vessels for this application, and that the proposed Indications for Use does not increate existing risks OR introduce new risks. Secondarily, while the articles vary in the specific types of devices being used, infusion catheters (including the proposed UNIFUSE™ Infusion System with Cooper Wire) are typically much smaller in diameter than other catheters being used, and therefore the other larger catheter types would represent a worst-case scenario in terms of hypothetical intrusiveness and potential for vessel damage. Please note: the clinical literature being cited and discussed is general in nature, and it is important to acknowledge that the subject UNIFUSE™ Infusion Catheter with Cooper Wire device was not itself evaluated in the referenced studies. Additionally, the discussion developed from the clinical literature is intended to relate only to the UNI*FUSE™ Infusion Catheter with Cooper Wire devices, and only those with infusion segment lengths that have been confirmed to be completely contained within non-peripheral pulmonary arteries in a straight configuration (i.e. 2cm and 5cm infusion segment lengths).

Summary: AngioDynamics, Inc. has identified and reviewed relevant articles that discuss catheterdirected interventions and/or therapy for pulmonary embolism, and each summary includes discussion pertaining to related complications identified. The summary begins on the following page.

{9}------------------------------------------------

"Catheter-Directed Therapy in Acute Pulmonary Embolism with Right Ventricular Dysfunction: A Promising Modality to Provide Early Hemodynamic Recovery" (Dilektasli, A.G.; et al. - 2016)} evaluates the use of catheter-directed therapy (CDT) as an alternative to systemic thrombolysis (ST) in patients. Catheter-directed therapy is a percutaneous procedure used to dissolve thrombus by administering a lytic directly into said thrombus. The primary outcomes were mortality, clinical success, and complications. The study included 15 consecutive patients who underwent immediate catheter-directed therapy. An essential conclusion of the study was related to the complications of catheter-directed therapy, specifically, "there were no technications, such as perforation of a cardiac/vascular structure, tamponade, or procedure-related death in our study." Instead, it continues by asserting that "interventionalist experience is known to influence the technical success." In this article, that conclusion is contrasted against a statement made within a referenced article titled Goldhaber's "Percutaneous Mechanical Thrombectomy for Acute Pulmonary Embolism" (Goldhaber, S. 2007)2 related to "percutaneous mechanical thrombectomy (PMT)" catheters", which states: "the percutaneous mechanical thrombectomy catheter can perforate the pulmonary artery, cause massive distal embolization, or cause hemolysis." The significance of these statements and their comparison to one another supports the following conclusion:

  • . The literature suggests that catheters that employ some type of mechanical component (or component that is in-addition-to a "typical" catheter design) may have a greater risk for the potential of vessel damage compared to catheters that do not.
    "Catheter-Directed Therapy for the Treatment of Massive Pulmonary Embolism: Systematic Review and Meta-Analysis of Modern Techniques" (Kuo, W. et al. – 2009)3 summarizes the authors' systematic review of modern techniques related to catheter-directed therapy for the treatment of massive pulmonary embolism; specifically, to evaluate the safety and effectiveness of modern catheter-directed therapy (CDT) as an alternative treatment for massive pulmonary embolism. 594 patients from 35 studies (6 prospective, 29 retrospective) were analyzed both for the clinical success rate associated with catheterdirected therapy, as well as the minor and major complications encountered during catheter-directed therapy treatment (along with the rates associated with each complication). The pooled clinical success rate from catheter-directed therapy was 86.5%, and the pooled risks of minor and major procedural complications were 7.9% and 2.4% respectively. Minor and major complications were listed by specific type of complication, and none were related to injury or damage to the pulmonary artery itself. The significance of this data is twofold:

    1. It shows that catheter-directed treatment in the pulmonary artery vasculature is a treatment that has a high success rate attributed to it; and
    1. It shows that the complications associated with catheter-based interventions in the pulmonary artery vasculature do not include – in either "minor" or "major" complication categories – damage and/or injury to the vessel itself.

"Catheter-Directed Interventions for Pulmonary Embolism" (Zarghouni, M; et al. – 2016)" is an analysis of the information and conclusions of numerous key studies related to pulmonary artery vasculature interventions, most notably (see next paqe):

4 Dilektasli, A.G.; et al. - "Catheter-Directed Therapy in Acute Pulmonary Embolism with Right Ventricular Dysfunction: A Promising Modality to Provide Early Hemodynamic Recovery" (2016)

2 Goldhaber, S. – "Percutaneous Mechanical Thrombectomy for Acute Pulmonary Embolism" (2007)

3 Kuo, W.; et al. – "Catheter-Directed Therapy for the Treatment of Massive Pulmonary Embolism: Systematic Review and Meta-Analysis of Modern Techniques" (2009)

4 Zarghouni, M.; et al. – "Catheter-Directed Interventions for Pulmonary Embolism" (2016)

{10}------------------------------------------------

  • . Piazza, et al. – "Prospective, Single-Arm, Multi-Center Trial of EkoSonic Endovascular System and Activase for Treatment of Acute Pulmonary Embolism – SEATTLE II Study" (2015)
  • . Kuo, et al. – "Pulmonary Embolism Response to Fragmentation, Embolectomy, and Catheter Thrombolysis - PERFECT Study" (2015)
  • . Kucher, et al. – "Ultrasound-Assisted Catheter Directed Thrombolysis for Acute Intermediate-Risk Pulmonary Embolism – ULTIMA Study" (2013)
  • . Meyer, et al. – "Pulmonary Embolism Thrombolysis – PEITHO Study" (2014)

The analysis outlines the number of patients involved in each study type, the endpoints of each study, the arms of each study, and the findings/complications/etc. it also analyzes various types of catheter-directed therapies in the pulmonary artery vasculature, most notably "Catheter-Directed Thrombolysis via Infusion Catheters." It describes the benefits as well as the adverse aspects of each treatment option, and identifies the complications associated with each. Of the various options identified within the literature, none of the complications identified were related to vessel damage, endothelial cell destruction, or other types of complications synonymous with pulmonary artery vasculature injury. It concludes "there has been a reemergence of interest in catheter-directed techniques. Newer guidelines employ CDI [catheter-directed intervention] in treatment protocols. CDI has become a vital to at many institutions."

510(k) Summary Conclusions

In conclusion, assessment of the similarities and differences of the proposed UNI*FUSE™ Infusion System with Cooper Wire and the predicate Bashir™ N-X Endovascular Catheter Model 7200 led Angiodynamics, Inc. to determine that the two are substantially equivalent to one another; specifically:

  • The proposed and predicate device have the identical ProCode, Regulation Number, Regulation ● Name, and Regulatory Class as one another;
  • . The proposed and predicate device have substantially equivalent Indications for Use and/or Intended Uses;
  • . The proposed and predicate devices each employ the same operating principle, mechanism of action, and are intended for the same patient populations; and,
  • . The proposed and predicate exhibit equivalent overall design, materials of manufacture, performance testing, sizes, and configurations.

Additionally, results of human factors testing on the proposed UNI*FUSE™ Infusion System with Cooper Wire provide further evidence that the devices are able to be used in the pulmonary artery vasculature. Lastly, evaluation of publicly available clinical literature leads AngioDynamics, Inc. to conclude that:

  • There is a lack of evidence indicating the pulmonary artery vasculature to be more markedly prone to damage than other vessels for this application; and,
  • . The literature suggests that catheters that employ some type of mechanical component (or component that is in-addition-to a "typical" catheter design) may have a greater risk for the potential of vessel damage compared to catheters that do not.

The sum of these evaluations and determinations lead AngioDynamics, Inc. to conclude that substantial equivalence has been demonstrated, and that the existing data, additional testing, and clinical evaluation determinations have confirmed that there are no new questions of safety or effectiveness.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).