K Number
K192819
Device Name
OEC Elite
Date Cleared
2019-11-08

(38 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, vascular, cardiac, critical care, and emergency procedures.

Device Description

The OEC Elite is a Mobile Fluoroscopic C-arm Imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-Ray images during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes.

The C-arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral, wig-wag) that allow the user to position the X-Ray image chain at various angles and distances with respect to the patient anatomy to be imaged. The C - arm is mechanically balanced allowing for ease of movement and capable of being "locked" in place using a manually activated lock. The C-Arm is comprised of the high voltage generator, software. X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector or Image Intensifier, depending on the choice of detector configuration desired.

The workstation is a stable mobile platform with an articulating arm supporting a color image, high resolution, LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems.

GE is submitting this pre-market notification for proposed labeling changes (quantitative performance claims) related to a previously-released feature, Enhanced Noise Reduction.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the GE OEC Elite mobile fluoroscopy system with "Enhanced Noise Reduction" claims. However, it does not contain the specific acceptance criteria or an explicit study proving the device meets those criteria in the format requested. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing and engineering bench testing, rather than reporting on a clinical study against predefined performance metrics.

Therefore, I cannot populate the table and answer all questions directly from the provided input. However, I can extract the information related to the non-clinical testing and the claims being made for the Enhanced Noise Reduction feature.

Here's a breakdown of what can be extracted and what information is missing:

Information that can be extracted or inferred:

  • Device Name: OEC Elite with Enhanced Noise Reduction
  • Purpose of the Submission: Proposed labeling changes (quantitative performance claims) related to the Enhanced Noise Reduction feature, demonstrating substantial equivalence to the predicate device.
  • Nature of Enhanced Noise Reduction: It's a user-selectable, augmented image processing pathway for Cardiac and Vascular acquisition profiles. It "reduces image noise in a manner characteristic of the reduction in noise resulting from an increase in photon flux" while maintaining "spatial and temporal resolution." It does not change the tube output (dose).
  • Claim: "Claims for equivalence to a higher power system without an increase in radiation dose for both cardiac and vascular applications."
  • Testing Conducted: Non-clinical testing, engineering bench testing, risk analysis, required reviews, design reviews, integration testing, performance testing, safety testing, simulated use testing. Specific mention of "image quality and dose performance using standard IQ metrics and QA phantoms" and "a wide variety of anthropomorphic phantoms."
  • Conclusion: The scientific engineering bench testing methods "demonstrate substantial equivalence." Clinical data was not required.

Missing Information (Crucial for the requested table and questions):

  • Specific Acceptance Criteria: The document mentions "quantitative performance claims" but does not detail what these exact criteria are (e.g., specific SNR improvement percentages, resolution metrics, dose reduction targets).
  • Reported Device Performance: Without explicit acceptance criteria, the "reported performance" cannot be formally assessed against them. The claim itself implies performance ("equivalence to a higher power system without an increase in radiation dose"), but specific metrics are absent.
  • Sample Size (Test Set): Not specified for any performance testing. Phantoms are mentioned.
  • Data Provenance (Test Set): Phantoms are artificial, so no country of origin or retrospective/prospective status.
  • Number of Experts for Ground Truth (Test Set): Not applicable as no human interpretation of test set images is mentioned as part of performance evaluation.
  • Qualifications of Experts for Ground Truth: Not applicable.
  • Adjudication Method: Not applicable.
  • MRMC Comparative Effectiveness Study: Explicitly stated that "clinical data is not required to demonstrate substantial equivalence." Therefore, no MRMC study with human readers comparing AI vs. without AI assistance was performed or reported.
  • Standalone Performance: The testing described is for the algorithm (Enhanced Noise Reduction) as part of the device (OEC Elite C-arm) but without human-in-the-loop performance reported.
  • Type of Ground Truth: For the "quantitative performance claims," the ground truth would typically be objective physical measurements of image quality parameters derived from phantoms.
  • Sample Size (Training Set): Not mentioned.
  • Ground Truth for Training Set: Not mentioned.

Based on the available information, here's what can be provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated/Inferred)Reported Device Performance (Summary from text)
Quantitative performance claims for Enhanced Noise Reduction related to image quality and dose.Achieved "equivalence to a higher power system without an increase in radiation dose for both cardiac and vascular applications." Reduces image noise while maintaining spatial and temporal resolution.
Compliance with IEC 60601-1 series, NEMA XR-27, 21CFR Subchapter J.Tested and compliant with all applicable standards.
No change in fundamental control mechanism, operating principle, energy type, or Intended Use.Changes described do not alter these aspects.
Substantial equivalence to predicate device (K172550, K171565).Scientific engineering bench testing demonstrated substantial equivalence, with no new safety/efficacy questions, hazards, or unexpected results.

2. Sample size used for the test set and the data provenance

  • Sample size: Not specified. Testing involved "standard IQ metrics and QA phantoms" and "a wide variety of anthropomorphic phantoms."
  • Data provenance: Not applicable in the traditional sense, as testing was performed using phantoms and engineering bench tests. This is non-clinical, in-house testing by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. The claims are based on objective, quantifiable physical measurements using phantoms, not on expert human interpretation of images for ground truth.

4. Adjudication method for the test set

  • Not applicable, as ground truth was established through physical measurements rather than human consensus or adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document explicitly states: "The new performance claims and the accumulated changes did not require clinical data in order to establish safety or efficacy." And "clinical data is not required to demonstrate substantial equivalence." Therefore, no MRMC study was performed or reported.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The "Enhanced Noise Reduction" is an algorithm (an "augmented image processing pathway"). The "additional engineering bench testing was performed to substantiate the quantitative performance claims related to Enhanced Noise Reduction" and to demonstrate "overall imaging performance... using a wide variety of anthropomorphic phantoms." This describes testing the algorithm's effect on image quality metrics without human interpretation as part of the core evaluation for these particular claims.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the non-clinical performance claims, the ground truth was established through objective physical measurements using "standard IQ metrics and QA phantoms" and "anthropomorphic phantoms." This involves measuring parameters like signal-to-noise ratio, spatial resolution, and potentially dose, against expected or ideal values from the phantoms.

8. The sample size for the training set

  • Not specified in the provided text.

9. How the ground truth for the training set was established

  • Not specified in the provided text, as details on the training set or its ground truth are absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 8, 2019

GE OEC Medical Systems, Inc. % Kenny M. Bello Regulatory Affairs Leader 384 N Wright Brothers Drive SALT LAKE CITY UT 84116

Re: K192819

Trade/Device Name: OEC Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: September 30, 2019 Received: October 1, 2019

Dear Kenny Bello:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192819

Device Name OEC Elite

Indications for Use (Describe)

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, vascular, cardiac, critical care, and emergency procedures.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo is a stylized monogram of the letters "GE" in a rounded, cursive font, enclosed within a circular shape. The color of the logo is a light blue.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) Summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:September 27, 2019
Submitter:GE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers DriveSalt Lake City, Utah 84116
Primary Contact:Kenny M BelloRegulatory Affairs LeaderTel: 801-201-3986e-mail: kenny.bello@ge.com
Secondary Contacts:Philip MalcaRegulatory Affairs DirectorGeneral Electric CompanyTel: +33 646-373-852e-mail: philip.malca@ge.comJohn JaeckleChief Regulatory Affairs StrategistGE Medical Systems, LLCTel: 262-424-9547e-mail: john.jaeckle@ge.com

PRODUCT IDENTIFICATION

Device Trade Name:OEC Elite
Regulation Name:Image-intensified Fluoroscopic x-ray system
Regulation:21CFR 892.1650
Classification:Class II
Product Code:OWB, JAA, OXO
ManufacturerGE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers Drive.Salt Lake City, Utah 84116
Manufacturing Location:GE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers Drive.Salt Lake City, Utah 84116

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of a blue circle with a stylized, cursive "GE" in white inside. The "GE" is surrounded by swirling, white shapes that resemble water droplets or stylized leaves. The blue circle has a slightly darker blue border, giving the logo a three-dimensional appearance.

Predicate Device:

Device Name:OEC Elite
510(k) number:K172550
Manufacturer:GE OEC Medical Systems, Inc
Regulation Name:Image-intensified Fluoroscopic x-ray system
Regulation:21CFR 892.1650
Classification:Class II
Product Code:OWB, JAA, OXO

Reference Device:

Device Name:OEC Elite
510(k) number:K171565
Manufacturer:GE OEC Medical Systems, Inc
Regulation Name:Image-intensified Fluoroscopic x-ray system
Regulation:21CFR 892.1650
Classification:Class II
Product Code:OWB, JAA, OXO

Marketed Device

OEC Elite with Enhanced Noise Reduction and associated claims, is a Mobile Fluoroscopic C-arm Imaging System built upon the existing technologies of the currently marketed predicate device, OEC Elite (K172550). It is of comparable type and substantially equivalent to the predicate device OEC Elite. In addition, the system has the same Intended Use and Indication for Use as the predicate device. The system is labeled as the OEC Elite.

Device Description

The OEC Elite is a Mobile Fluoroscopic C-arm Imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-Ray images during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes.

The C-arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral, wig-wag) that allow the user to position the X-Ray image chain at various angles and distances with respect to the patient anatomy to be imaged. The C - arm is mechanically balanced allowing for ease of movement and capable of being "locked" in place using a manually activated lock. The C-Arm is comprised of the high voltage generator, software. X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector or Image Intensifier, depending on the choice of detector configuration desired.

The workstation is a stable mobile platform with an articulating arm supporting a color image, high resolution, LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems.

GE is submitting this pre-market notification for proposed labeling changes (quantitative performance claims) related to a previously-released feature, Enhanced Noise Reduction.

{5}------------------------------------------------

GE Healthcare OEC Elite 510(k) Premarket Notification Submission - OEC Elite

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are three white swirls around the letters. The logo is simple and recognizable, and it is associated with a large and well-known company.

Intended Use

The OEC Elite Mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

Indications for Use

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency procedures.

Technology

In addition to the performance claims for the Enhanced Noise Reduction feature, the proposed device includes the accumulated changes to the predicate and reference devices.

The OEC Elite with Enhanced Noise Reduction and associated claims employs the same fundamental technology as that of the predicate and reference devices. The image chain including the X-ray tube, high voltage generator, collimator, X-ray filters, and detectors, remains unchanged from the predicate, OEC Elite.

The system continues to meet all applicable IEC 60601-1 series of standards, NEMA XR-27, and applicable parts of 21CFR Subchapter J. None of the accumulated changes affect the imaging characteristics and performance associated with FDA's SSXI Guidance. Incorporation of the wireless footswitch followed FDA 's wireless guidance. The new performance claims and the accumulated changes did not require clinical data in order to establish safety or efficacy.

In order for FDA review of the proposed quantitative claims, the OEC Elite's performance with the Enhanced Noise Reduction feature has been scientifically tested and substantiated and included in this submission.

Enhanced Noise Reduction is a user-selectable, augmented image processing pathway available for Cardiac and Vascular acquisition profiles. Enhanced Noise Reduction reduces image noise in a mamer characteristic of the reduction in noise resulting from an increase in photon flux. It is designed to maintain both spatial and temporal resolution. Use of Enhanced Noise Reduction does not change the tube output (dose).

Predicate Device: OEC EliteK172550 (and K171565)Proposed Device: OEC Elite withEnhanced Noise Reduction claims
X-ray InitiationandTermination> Wired footswitch> Wired hand switch> Wired footswitch> Wired hand switch> Optional wireless footswitch
CardiacImaging> Available with the 21 cm imagereceptor.> Available with the 21 and 31 cmimage receptor.
Image NoiseReduction> Image processing chainincluding noise modeling,filtering, reduction.> Image processing chain includingnoise modeling, filtering, reduction.> Enhanced Noise Reduction

The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The frame has four stylized water droplets or swirls positioned symmetrically around the circle, adding a decorative element to the design. The logo is rendered in a light blue color.

Control Panelon C-arm> Button actuated control paneland reference monitor.> Integrated touch screen controlpanel.
Motorized C-arm motionfeature> No motion control feature.> Motorized motion control
EnhancedNoiseReductionClaims> No Enhanced Noise Reductionfeature.> Claims for equivalence to a higherpower system without an increasein radiation dose for both cardiacand vascular applications.

The changes described above do not change the fundamental control mechanism, operating principle, energy type, and do not change the Intended Use of the predicate device.

Determination of Substantial Equivalence:

The device has successfully completed all testing per GE's quality system as well as addition engineering bench testing in support of this submission. The system has been tested and is compliant with the IEC 60601-1 Ed. 3 series, including IEC 60601-1-2, 60601-1-3, 60601-2-43, and 60601-2-54. All applicable 21CFR Subchapter J performance standards are met. The OEC Elite system was developed under the GE OEC Medical Systems Quality Management System, including design controls, risk management and software development life cycle processes. The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • Required Reviews
  • Design Reviews
  • · Integration testing (System verification)
  • · Performance testing (Verification)
  • · Safety testing (Verification)
  • · Simulated use testing (Validation)

The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.

Non-Clinical Testing

The verification and validation testing have been successfully completed as required by design control procedures under GE Healthcare's quality system. This includes risk management, software verification and validation testing, as well as image quality and dose performance using standard IQ metrics and QA phantoms.

Additional engineering bench testing was performed to substantiate the quantitative performance claims related to Enhanced Noise Reduction. Additionally, testing was performed to demonstrate the overall imaging performance of OEC Elite with Enhanced Noise Reduction using a wide variety of anthropomorphic phantoms.

Clinical Testing

OEC Elite with Enhanced Noise Reduction claims uses the identical imaging change, does not change the Indications for Use, and has equivalent/identical technological characteristics. This type of change supports using scientific, established, engineering-based performance testing. The system can be fully

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. There are decorative flourishes resembling water droplets or stylized leaves surrounding the circle, adding a touch of elegance to the design. The logo is simple, recognizable, and represents the well-known multinational corporation.

tested/evaluated using engineering bench testing and clinical data is not required to demonstrate substantial equivalence.

Substantial Equivalence Conclusion:

The changes associated with OEC Elite with Enhanced Noise Reduction Claims do not change the Indications for Use or Intended Use. The OEC Elite with Enhanced Noise Reduction Claims' control mechanism, operating principle, and energy type have not changed from the predicate and reference devices. In addition, the OEC Elite with Enhanced Noise Reduction Claims complies with the same international and national standards, 21 CFR performance requirements and NEMA XR standards.

It has successfully completed design controls activities, including risk management, verification and validation. The verification and validation testing as well as the additional engineering bench testing that substantiated the claims did not raise any new questions of safety or efficacy, identify any new hazards, and did not have any unexpected results.

The scientific engineering bench testing methods used to evaluate the safety and effectiveness of OEC Elite's with Enhanced Noise Reduction characteristics and proposed claims are acceptable and demonstrate substantial equivalence.

We believe the OEC Elite with Enhanced Noise Reduction Claims is of comparable type and substantially equivalent to the predicate and reference devices, OEC Elite (K172550, K171565) and therefore is safe and effective for its intended use.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.