(59 days)
No
The summary describes a mechanical microcatheter and does not mention any AI or ML components or functions.
Yes
The device is intended for the infusion of "therapeutic agents", indicating its use in delivering therapies.
Yes
The device is intended for the "infusion of diagnostic and therapeutic agents," indicating its use in diagnostic procedures.
No
The device description clearly describes a physical catheter with a lumen, guidewire compatibility, radiopaque markers, hydrophilic coating, and a luer fitting. The performance studies also focus on physical and material properties of the catheter.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for the "infusion of diagnostic and therapeutic agents" within the peripheral vasculature. This describes a device used in vivo (within the living body) for delivering substances, not for testing samples in vitro (outside the living body).
- Device Description: The description details a catheter designed for accessing and navigating blood vessels, which is an in vivo application.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), performing tests on these samples, or providing diagnostic information based on such analysis.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This microcatheter's function is to deliver substances into the body.
N/A
Intended Use / Indications for Use
The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.
Product codes
DQY, KRA
Device Description
The PG Pro Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the catheter hub is used for the attachment of accessories.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
peripheral vasculature
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Biocompatibility testing: Biocompatibility evaluation for the PG Pro Microcatheter was conducted in accordance with ISO10993-1 Biological Evaluation of Medical Devices-Part 1: Guidance on Selection of Tests. The microcatheter is considered a limited (
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2019
MicroVention, Inc., Mr. Ganesh Balachandar Regulatory Affairs Specialist 35 Enterprise Aliso Viejo, California 92656
Re: K192625
Trade/Device Name: PG Pro Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, KRA Dated: September 20, 2019 Received: September 23, 2019
Dear Mr. Balachandar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K192625
Device Name PG Pro Microcatheter
Indications for Use (Describe)
The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapentic agents.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(K) SUMMARY
SUBMITTER
MicroVention Inc. 35 Enterprise, Aliso Viejo, CA 92656 Phone: 714-247-8000 Contact Person: Ganesh Balachandar Date Prepared: 20th, September 2019
DEVICE
Name of the Device: PG Pro Microcatheter Common Name: Percutaneous Catheter Classification Name: Device, Percutaneous Catheter Regulatory Class: Class II, 21 CFR 870.1250 Product Code: DQY Additional Product Code: KRA
PREDICATE DEVICE
Headway 27 Microcatheter (K142449)
REFERENCE DEVICE
Headway 17 Microcatheter (K101542)
DEVICE DESCRIPTON
The PG Pro Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the catheter hub is used for the attachment of accessories.
INDICATION FOR USE
The PG Pro Microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.
4
TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE
Table 1: Headway Microcatheters and PG Pro Comparison table | ||||
---|---|---|---|---|
Property | Headway 17 | |||
Microcatheter | ||||
Reference Device | Headway 27 | |||
Microcatheter | ||||
Predicate Device | PG Pro Microcatheter | |||
Subject Device | Differences | |||
(K101542) | (K142449) | |||
Indication For | ||||
Use | The Headway 17 | |||
Microcatheter is | ||||
intended for general | ||||
intravascular use, | ||||
including the peripheral, | ||||
coronary and | ||||
neurovasculature for the | ||||
infusion of diagnostic | ||||
agents, such as contrast | ||||
media, and therapeutic | ||||
agents, such as occlusion | ||||
coils. | The Headway 27 | |||
Microcatheter is intended | ||||
for general intravascular | ||||
use, including the | ||||
peripheral, coronary and | ||||
neurovasculature for the | ||||
infusion of diagnostic | ||||
agents, such as contrast | ||||
media, and therapeutic | ||||
agents, such as occlusion | ||||
coils. | The PG Pro | |||
microcatheter is | ||||
intended for the | ||||
peripheral vasculature | ||||
for the infusion of | ||||
diagnostic and | ||||
therapeutic agents. | The PG Pro | |||
Microcatheter is | ||||
intended only | ||||
for the | ||||
peripheral | ||||
vasculature. | ||||
Principle of | ||||
Operation | The principle of | |||
operation incorporates | ||||
various design features | ||||
that allow it to achieve | ||||
its intended use. A semi- | ||||
rigid proximal section | ||||
allows for steerability. | ||||
Progressively softer | ||||
durometer distal | ||||
segments and flexible tip | ||||
allow for atraumatic | ||||
navigation through the | ||||
vasculature. Radiopaque | ||||
markers allow for | ||||
fluoroscopic | ||||
visualization. | ||||
Hydrophilic polymer | ||||
coating reduces friction | The principle of operation | |||
incorporates various | ||||
design features that allow | ||||
it to achieve its intended | ||||
use. A semi-rigid | ||||
proximal section allows | ||||
for steerability. | ||||
Progressively softer | ||||
durometer distal segments | ||||
and flexible tip allow for | ||||
atraumatic navigation | ||||
through the vasculature. | ||||
Radiopaque markers | ||||
allow for fluoroscopic | ||||
visualization. | ||||
Hydrophilic polymer | ||||
coating reduces friction | ||||
during navigation through | ||||
the vasculature. | The principle of | |||
operation incorporates | ||||
various design features | ||||
that allow it to achieve | ||||
its intended use. A semi- | ||||
rigid proximal section | ||||
allows for steerability. | ||||
Progressively softer | ||||
durometer distal | ||||
segments and flexible | ||||
tip allow for atraumatic | ||||
navigation through the | ||||
vasculature. Radiopaque | ||||
markers allow for | ||||
fluoroscopic | ||||
visualization. | ||||
Hydrophilic polymer | ||||
coating reduces friction | Identical | |||
Table 1: Headway Microcatheters and PG Pro Comparison table | ||||
Property | Headway 17 | |||
Microcatheter | ||||
Reference Device | ||||
(K101542) | Headway 27 | |||
Microcatheter | ||||
Predicate Device | ||||
(K142449) | PG Pro Microcatheter | |||
Subject Device | Differences | |||
during navigation | ||||
through the vasculature. | during navigation | |||
through the vasculature. | ||||
Material(s) | Catheter body: Pebax, | |||
stainless steel, Grilamid, | ||||
inner liner of PTFE and | ||||
barium sulfate | ||||
incorporated in distal | ||||
Pebax segment. | Catheter body: Pebax, | |||
stainless steel, Grilamid, | ||||
inner liner of PTFE | Catheter body: Pebax, | |||
Tungsten, Grilamid, | ||||
inner liner of PTFE | PG Pro has a | |||
Tungsten coil | ||||
versus a | ||||
Stainless-steel | ||||
coil for the | ||||
Headway 27 and | ||||
Headway 17 | ||||
Microcatheters. | ||||
Marker band: Pt/Ir | Marker band: Pt/Ir | Marker band: Pt/Ir | Identical | |
Hub: Nylon | Hub: Nylon | Hub: Nylon | Identical | |
Strain relief: Pebax | Strain relief: Pebax | Strain relief: Pebax | Identical | |
Introducer: Pebax | Introducer: Pebax | Introducer: Pebax | Identical | |
Shaping mandrel: | ||||
Stainless steel | Shaping mandrel: | |||
Stainless steel | Shaping mandrel: N/A | PG Pro does not | ||
come with a | ||||
Shaping | ||||
mandrel. | ||||
Proximal | ||||
ID/OD | ID = .0170" min | |||
OD = .031" | ID = .027" min | |||
OD = .040" | ID = .027" min | |||
OD = .038"MAX | PG Pro has a | |||
Max Proximal | ||||
O.D of .038" | ||||
which is in- | ||||
between | ||||
Headway 17 and | ||||
Table 1: Headway Microcatheters and PG Pro Comparison table | ||||
Property | Headway 17 | |||
Microcatheter | ||||
Reference Device | ||||
(K101542) | Headway 27 | |||
Microcatheter | ||||
Predicate Device | ||||
(K142449) | PG Pro Microcatheter | |||
Subject Device | Differences | |||
Distal ID/OD | ID = .0170" min | |||
OD = .022" | ID = .027" min | |||
OD = .034" | ID = .027" min | |||
OD = .038" MAX | 27 | |||
Microcatheter | ||||
proximal OD. | ||||
PG Pro has a | ||||
Max Distal O.D | ||||
of .038" | ||||
No. of Marker | ||||
bands | 2 | 2 | 2 | Identical |
Coating | Hydrophilic Coating | Hydrophilic Coating | Hydrophilic Coating | Identical |
Effective | ||||
Length | 150 cm | 150 cm and 156 cm | 140 cm and 165 cm | PG Pro comes in |
two lengths – | ||||
140 and 165 cm. | ||||
Tip | ||||
Configuration | Preshaped tips – with the | |||
option of secondary | ||||
shaping for proper | ||||
adjustment to the | ||||
anatomy prior to use. | Straight - Steam | |||
Shapeable by physician | ||||
prior to use | Straight configuration | PG Pro is | ||
available only in | ||||
the straight | ||||
configuration. | ||||
Guidewire | ||||
Compatibility | 0.014 "wires or smaller | 0.014 "wires or smaller | 0.021 "wires or smaller | PG Pro is |
compatible with | ||||
0.021" wires or | ||||
smaller | ||||
Method of | ||||
supply | Sterile and single use | Sterile and single use | Sterile and single use | Identical |
Accessories | Shaping mandrel and | |||
Introducer sheath | Shaping mandrel and | |||
Introducer sheath | Introducer sheath | PG Pro does not | ||
include shaping | ||||
mandrel |
5
K192625
Page 3 of 6
6
K192625
Page 4 of 6
7
PERFORMANCE DATA
Biocompatibility testing
Biocompatibility evaluation for the PG Pro Microcatheter was conducted in accordance with ISO10993-1 Biological Evaluation of Medical Devices-Part 1: Guidance on Selection of Tests. The microcatheter is considered a limited (