AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are powder-free, ambidextrous with beaded cuff, green colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.

The gloves are designed and manufactured in accordance with the ASTM D6319-10 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013).

AI/ML Overview

The provided text describes the acceptance criteria and the results of various tests conducted for the "Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs)". This is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, rather than a study proving the device meets acceptance criteria in a clinical context. Therefore, many of the requested elements for an AI/medical device study (like sample size for test set, data provenance, expert adjudication, MRMC studies, standalone performance, training set sample size, etc.) are not applicable to this type of submission.

However, I can extract the acceptance criteria and reported device performance from the provided biocompatibility and performance testing summaries.


1. Table of acceptance criteria and the reported device performance

Biocompatibility Testing Summary

Test StandardTest Standard DescriptionAcceptance CriteriaReported Device Performance
ISO 10993-10Primary Skin IrritationCutaneous macroscopical examinations are performed after removal of test material. The appearance of each application site and skin reaction observed are scored.No skin reactions observed. Not a skin irritant.
ISO 10993-10Dermal SensitizationChallenge sites are visually assessed for allergic reactions and the intensity of the reaction is scored.No positive allergic reactions observed. Not a skin sensitizer.
ISO 10993-5CytotoxicityTest articles scoring '0', '1', or '2' are considered 'non-cytotoxic.' Test articles scoring '3' or '4' are considered 'cytotoxic.'At the neat extraction, the test article is considered cytotoxic.
ISO 10993-11Acute Systemic ToxicityAbnormal clinical signs indicative of toxicity during the 72 hours test period are observed and reported.No signs of toxicity at any of the observation periods were observed.

Performance Testing Summary

Test StandardTest Standard DescriptionAcceptance CriteriaReported Device Performance
ASTM D6319-10Standard Specification for Nitrile Examination Gloves for Medical ApplicationSamples shall meet specified physical dimensions in accordance with the standard.All device dimensions meet the standard.
ASTM D6124-06 (Reaffirmation 2011)Standard Test Method for Residual Powder on Medical Gloves< 2mg of residual powder per glove.Complies with standard.
ASTM D5151-06 (Reapproved 2011)Standard Test Method for Detection of Holes in Medical GlovesAny glove that shows a droplet, stream or other type of water leakage shall be considered to have failed.Complies with standard.
ASTM D6978-05 (Reapproved 2013)Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsThe absorbance of test chemicals, which permeate through the test articles, is measured and reported. Note: Specific breakthrough times vary by drug.Refer to the detailed list of chemotherapy drugs and their breakthrough times (2 records below 240 mins, the rest >240 mins)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for each test mentioned (e.g., how many gloves were tested for holes, or how many samples for biocompatibility). The tests are performed according to established ASTM and ISO standards, which typically define the required sample sizes for quality control and performance evaluation.

Data provenance is not explicitly stated as "country of origin of data" but refers to tests being conducted according to international standards (ASTM, ISO). These are bench tests, not clinical studies. The data would be considered prospective in the sense that the tests were specifically performed for this submission to evaluate the device.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the tests described are laboratory/bench tests, not evaluations requiring expert interpretation of clinical data in the context of ground truth establishment. The "ground truth" here is determined by chemical and physical measurements against established standards.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The tests are objective measurements against defined standards, not requiring human adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a submission for medical gloves, not an AI-powered diagnostic or assistive device.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a submission for medical gloves.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For physical and chemical properties, the "ground truth" is based on objective measurements defined by international standards (ASTM and ISO). For example:

  • Physical dimensions are measured against the specifications in ASTM D6319-10.
  • Residual powder is measured against a limit in ASTM D6124-06.
  • Detection of holes is defined by the absence of leakage in ASTM D5151-06.
  • Chemotherapy drug permeation is measured by breakthrough time according to ASTM D6978-05.
  • Biocompatibility assessments (irritation, sensitization, cytotoxicity, systemic toxicity) are assessed by standardized laboratory protocols with established biological endpoints according to ISO 10993 series.

8. The sample size for the training set

Not applicable. This is a device (medical glove) and not an AI or machine learning algorithm that requires a training set. The device's characteristics are inherent to its manufacturing process and materials.


9. How the ground truth for the training set was established

Not applicable, as there is no training set for a medical glove.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 10, 2020

Medline Industries, Inc. Claire Pigman Associate Manager, Regulatory Affairs Three Lakes Drive Northfield, Illinois 60093

Re: K192315

Trade/Device Name: Medline Green Ambidextrous Power-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: December 6, 2019 Received: December 10, 2019

Dear Claire Pigman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192315

Device Name

Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Blenoxane (Bleomycin) 15.0 mg/ml >240 minutes Bortezomib 1.0 mg/ml >240 minutes Busulfan 6.0 mg/ml >240 minutes Carboplatin 10.0 mg./ml >240 minutes Carmustine (BCNU) 3.3 mg/ml 13.2 minutes Cetuximab (Eributux) 2.0 mg/ml >240 minutes Cisplatin 1.0 mg/ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Docetaxel 10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Epirubicin (Ellence) 2.0 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fludarabine 25.0 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Gemcitabine (Gemzar) 38.0 mg/ml >240 minutes Idarubicin 1.0 mg/ml >240 minutes Ifosfamide 50.0 mg/ml >240 minutes Irinotecan 50.0 mg/ml >240 minutes Mechlorethamine HCI 1.0 mg/ml >240 minutes Melphalan 5.0 mg/ml >240 minutes Methotrexate 25.0 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Mitoxantrone 2.0 mg/ml >240 minutes Oxaliplatin 5.0 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Paraplatin 10.0 mg/ml >240 minutes Pemetrexed 25.0 mg/ml >240 minutes Raltitrexed 0.5 mg/ml >240 minutes Rituximab 10.0 mg/ml >240 minutes Thiotepa 10.0 mg/ml 44 minutes Trisonex (Arsenic Trioxide) 1.0 mg/ml >240 minutes Vidaza (Azacitidine) 25.0 mg/ml >240 minutes Vinblastine 1.0 mg/ml >240 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes Vinorelbine 10.0 mg/ml >240 minutes

{3}------------------------------------------------

Please note that the following drug has extremely low permeation time: Carmustine (BCNU) (3.3 mg/ml) 13.2 minutes Please note that the following drug has extremely low permeation time: Thiotepa (10.0 mg/ml) 44.0 minutes

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, block letters, and to the right of the word is a white, stylized star shape. The star has four points, with the bottom point extending further down than the other three.

K192315 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92]

Submitter / 510(k) Sponsor

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592

Contact Person

Claire Pigman Associate Manager, Regulatory Affairs Phone: 847-643-4071 Email: cpigman@medline.com

Summary Preparation Date January 9, 2020

Type of 510(k) Submission Traditional

Device Name / Classification

Name of Device: Patient Examination Glove (Tested for Use with Chemotherapy Drugs) Proprietary Name: Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs) Common Name: Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LZA, LZC Classification Panel: General Hospital Regulation #: 21 CFR 880.6250

Predicate Device

Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) - K180696

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, with a stylized white starburst symbol above it. The starburst has four points, resembling a compass rose.

Device Description

There have been no changes to the Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs). The materials of construction, colorant, sizes and product specifications have not changed and are identical to what was cleared under K180696. The only change is to the Indications for Use.

As previously described in K180696, the Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are powder-free, ambidextrous with beaded cuff, green colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.

The gloves are designed and manufactured in accordance with the ASTM D6319-10 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013).

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Blenoxane (Bleomycin) 15.0 mg/ml >240 minutes Bortezomib 1.0 mg/ml >240 minutes Busulfan 6.0 mg/ml >240 minutes Carboplatin 10.0 mg./ml >240 minutes Carmustine (BCNU) 3.3 mg/ml 13.2 minutes Cetuximab (Eributux) 2.0 mg/ml >240 minutes Cisplatin 1.0 mg/ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Docetaxel 10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Epirubicin (Ellence) 2.0 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fludarabine 25.0 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Gemcitabine (Gemzar) 38.0 mg/ml >240 minutes Idarubicin 1.0 mg/ml >240 minutes Ifosfamide 50.0 mg/ml >240 minutes Irinotecan 50.0 mg/ml >240 minutes

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, block letters. A white, stylized starburst symbol is behind the text, with the word "MEDLINE" centered on the starburst.

Mechlorethamine HCI 1.0 mg/ml >240 minutes Melphalan 5.0 mg/ml >240 minutes Methotrexate 25.0 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Mitoxantrone 2.0 mg/ml >240 minutes Oxaliplatin 5.0 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Paraplatin 10.0 mg/ml >240 minutes Pemetrexed 25.0 mg/ml >240 minutes Raltitrexed 0.5 mg/ml >240 minutes Rituximab 10.0 mg/ml >240 minutes Thiotepa 10.0 mg/ml 44 minutes Trisonex (Arsenic Trioxide) 1.0 mg/ml >240 minutes Vidaza (Azacitidine) 25.0 mg/ml >240 minutes Vinblastine 1.0 mg/ml >240 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes Vinorelbine 10.0 mg/ml >240 minutes

Please note that the following drug has extremely low permeation time: Carmustine (BCNU) (3.3 mg/ml) 13.2 minutes Please note that the following drug has extremely low permeation time: Thiotepa (10.0 mg/ml) 44.0 minutes

Summary of Technological Characteristics

Device CharacteristicProposed DevicePredicate DeviceComparison Analysis
Product NameMedline Green AmbidextrousPowder-Free Nitrile ExaminationGloves with Colloidal OatmealUSP (Tested for use withChemotherapy Drugs)Medline Green AmbidextrousPowder-Free Nitrile ExaminationGloves with Colloidal OatmealUSP (Tested for use withChemotherapy Drugs)Same
510(k) ReferenceK192315K180696N/A
Product OwnerMedline Industries, Inc.Medline Industries, Inc.Same
Product CodeLZA, LZCLZA, LZCSame
Intended UseA patient examination glove is adisposable device intended formedical purposes that is worn onthe examiner's hand to preventA patient examination glove is adisposable device intended formedical purposes that is worn onthe examiner's hand to preventSame
contamination between patient and examiner.These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugscontamination between patient and examiner.These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs
Regulation Number21 CFR 880.625021 CFR 880.6250Same
MaterialsNitrileColloidal oatmealNitrileColloidal oatmealSame
ColorGreenSame
Sizesx- small, small, medium, large, x-largex-small, small, medium, large, x-largeSame
Dimensions - LengthComplies with ASTM D6319-10220mm min.Complies with ASTM D6319-10230mm min.Same
Dimensions - WidthComplies with ASTM D6319-10X-small – 70±10mmSmall – 80±10mmMedium – 95±10mmLarge – 110±10mmX-large – 120±10mmComplies with ASTM D6319-10X-small – 70±10mmSmall – 80±10mmMedium – 95±10mmLarge – 110±10mmX-large – 120±10mmSame
Dimensions - ThicknessComplies with ASTM D6319-10Palm – 0.05mm min.Finger – 0.05mm min.Complies with ASTM D6319-10Palm – 0.05mm min.Finger – 0.05mm min.Same
Physical PropertiesComplies with ASTM D6319-10 minimum:Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.Complies with ASTM D6319-10 minimum:Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min.Same
Freedom from HolesComplies with ASTM D6319-10 and ASTM D5151-06 G-1, AQL 1.5Complies with ASTM D6319-10 and ASTM D5151-06 G-1, AQL 1.5Same
Powder or Powder-FreePowder-FreePowder-FreeSame
Residual PowderComplies with ASTM D6319-10<2mg per gloveComplies with ASTM D6319-10<2mg per gloveSame
Colloidal OatmealContentMinimum 5mg per glove for all sizesSame
Contact DurationsLimited <24 hoursSame
Biocompatibility perISO 10993-1Not a skin irritantNot a skin sensitizerAt the neat extraction, the testarticle is considered cytotoxic, butthe acute systemic toxicity resultsdemonstrate the device will notcause a systemic effect.Not a skin irritantNot a skin sensitizerNot tested against cytotoxicitySame
SterilityNon-sterileNon-sterileSame
Chemotherapy DrugsTested with MinimumBreakthroughDetection Time astested per ASTMD6978Blenoxane (Bleomycin)15 mg/ml >240 min.Similar
Bortezomib 1 mg/ml>240 min.Similar
Busulfan 6 mg/ml>240 min.Similar
Carboplatin 10 mg/ml>240 min.Similar
Carmustine (BCNU) 3.3 mg/ml>240 min.Carmustine (BCNU) 3.3 mg/ml13.2 min.Same
Cetuximab (Erbitux) 2 mg/ml >240min.Similar
Cisplatin 1.0 mg/ml>240 min.Cisplatin 1.0 mg.ml>240 min.Same
Cyclophosphamide (Cytoxan) 20mg/ml>240 min.Cyclophosphamide (Cytoxan) 20.0mg/ml>240 min.Same
Dacarbazine (DTIC)10.0 mg/ml>240 min.Dacarbazine (DTIC)10.0 mg/ml>240 min.Same
Docetaxel 10.0 mg/ml>240 min.Similar
Doxorubicin Hydrochloride 2.0mg/ml >240 min.Doxorubicin Hydrochloride 2.0mg/ml >240 min.Same
Epirubicin (Ellence) 2.0 mg/ml>240 min.Similar
Etoposide (Toposar) 20.0 mg/ml>240 min.Etoposide (Toposar) 20.0 mg/ml>240 min.Same
Fludarabine 25 mg/ml>240 min.Similar
Fluorouracil 50.0 mg/ml>240 min.Fluorouracil 50.0 mg/ml>240 min.Same
Gemcitabine (Gemzar) 38 mg/ml>240 min.Similar
Idarubicin 1 mg/ml>240 min.Similar
Ifosfamide 50.0 mg/ml>240 min.Similar
Irinotecan 20.0 mg/ml>240 min.Similar
Mechlorethamine HCI 1.0 mg/ml>240 min.Similar
Melphalan 5 mg/ml>240 min.Similar
Methotrexate 25 mg/ml>240 min.Methotrexate 25 mg/ml>240 min.Same
Mitromycin C. 0.5 mg/ml >240Mitromycin C. 0.5 mg/ml >240Same
Mitoxantrone 2.0 mg/ml>240 min.Similar
Oxaliplatin 5 mg/ml>240 min.Similar
Paclitaxel (Taxol) 6.0 mg/ml>240 min.Paclitaxel (Taxol) 6.0 mg/ml>240 min.Same
Paraplatin 10 mg/ml>240 min.Similar
Pemetrexed Disodium 25 mg/ml>240 min.Similar
Raltitrexed 0.5 mg/ml>240 min.Similar
Rituximab 10 mg/ml>240 min.Similar
Thiotepa 10.0 mg/ml44.0 min.Thiotepa 10.0 mg/ml44.0 min.Same
Trisonex 0.1 mg/ml>240 min.Similar
Vidaza (5-Azacytidine) 25 mg/ml>240 minSimilar
Vinblastine 1 mg/ml>240 min.Similar
Vinorelbine 10 mg/ml>240 min.Similar
Vincristine Sulfate 1.0 mg/ml >240min.Vincristine Sulfate 1.0 mg/ml>240 min.Same

Table 1 Comparison of Proposed and Predicate Device

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is written in white, block letters on the left side of the square. To the right of the word is a white, stylized starburst symbol.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the Medline logo, which is set against a blue square background. The word "MEDLINE" is prominently displayed in white, sans-serif font, positioned to the left of a stylized white starburst symbol. The starburst has four points, resembling a compass rose, and it is oriented diagonally within the square.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image features the Medline logo against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized starburst symbol. The starburst is composed of four white, elongated triangles converging at a central point, creating a dynamic and modern design.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image features the Medline logo against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized white starburst symbol. The starburst is oriented diagonally, with its points extending beyond the text, creating a dynamic and recognizable brand mark.

Summary of Non-Clinical Testing

Biocompatibility

The biocompatibility evaluation for Medline Green Ambidextrous Powder-Free Nitrile Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) was conducted in accordance with ANSI/AAMI/ISO 10993-1:2018 Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.

The following tests were performed to evaluate the biocompatibility of Medline Green Ambidextrous Powder-Free Nitrile Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs):

TestStandardTest StandardDescriptionAcceptance CriteriaResult
ISO 10993-10Primary Skin IrritationCutaneous macroscopical examinations areperformed after removal of test material. Theappearance of each application site and skinreaction observed are scored.No skin reactionsobserved. Not a skinirritant.
ISO 10993-10Dermal SensitizationChallenge sites are visually assessed for allergicreactions and the intensity of the reaction isscored.No positive allergicreactions observed. Nota skin sensitizer
ISO 10993-5CytotoxicityTest articles scoring '0', '1', or '2' areconsidered 'non-cytotoxic.' Test articlesscoring '3' or '4' are considered 'cytotoxic.'At the neat extraction,the test article isconsidered cytotoxic.
ISO 10993-11Acute Systemic ToxicityAbnormal clinical signs indicative of toxicityduring the 72 hours test period are observed andreported.No signs of toxicity atany of the observationperiods were observed.

Table 2 Biocompatibility Testing Summary

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. Additionally, permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2013), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs. To summarize, the following tests were performed to evaluate the functional performance of the subject device:

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is written in white, block letters on the left side of the square. To the right of the word is a white, stylized star shape that is oriented vertically.

Table 3 Performance Testing Summary

Test StandardTest Standard DescriptionAcceptance CriteriaResult
ASTM D6319-10Standard Specification forNitrile Examination Gloves forMedical ApplicationSamples shall meet specifiedphysical dimensions in accordancewith the standard.All device dimensionsmeet the standard.
ASTM D6124-06(Reaffirmation2011)Standard Test Method forResidual Powder on MedicalGloves< 2mg of residual powder per glove.Complies withstandard.
ASTM D5151-06(Reapproved2011)Standard Test Method forDetection of Holes in MedicalGlovesAny glove that shows a droplet,stream or other type of waterleakage shall be considered to havefailed.Complies withstandard.
ASTM D6978-05(Reapproved2013)Standard Practice forAssessment of Resistance ofMedical Gloves to Permeationby Chemotherapy DrugsThe absorbance of test chemicals,which permeate through the testarticles, is measured and reported.Refer to Table 1 above.

Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device (K180696).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.