K Number
K192234
Device Name
ActiPatch
Date Cleared
2020-01-31

(165 days)

Product Code
Regulation Number
890.5290
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Adjunctive treatment of musculoskeletal pain

Device Description

The ActiPatch® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Device: ActiPatch® (K192234)
Intended Use: Adjunctive treatment of musculoskeletal pain


1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state "acceptance criteria" in a typical quantitative format for algorithm performance (e.g., "sensitivity must be > X%, specificity > Y%"). Instead, the substantial equivalence hinges on demonstrating that the ActiPatch is "at least as safe and effective as the predicate ActiPatch (K152432)."

Therefore, the "acceptance criteria" can be inferred from the studies' primary effectiveness endpoints, which consistently aimed to show a statistically significant reduction in pain compared to a control group (either active-treatment or placebo). The device's performance is described by achieving these significant reductions.

Acceptance Criteria (Inferred from Study Goals)Reported Device Performance (Summary of Findings)
Effectiveness (Pain Reduction):
Reduction in pain for cervical osteoarthritis (neck pain) compared to active-treatment control (NSAID)ActiPatch significantly reduced pain (measured by VAS pain) associated with COA in the device treatment group. Treatment difference between device-treatment and NSAID-treatment groups was significant (p<0.05).
Improvement in pain for osteoarthritis of the knee compared to placeboActiPatch significantly reduced pain (measured by VAS pain) associated with KOA in the device treatment group. Treatment difference between active and placebo treatment groups was significant (p<0.05) for both VAS and WOMAC scores.
Reduction in pain for plantar fasciitis (heel pain) compared to placeboActiPatch significantly reduced pain (measured by VAS pain) associated with PF in the device treatment group. Treatment difference between active and placebo treatment groups was significant (p<0.05).
Safety:
No unacceptable treatment-related adverse eventsThe studies generally mention "primary safety endpoint was all treatment-related adverse events." The conclusion states the device is "at least as safe and effective," implying the safety profile was acceptable and comparable to the predicate. Biocompatibility testing (skin sensitization, irritation, cytotoxicity) also showed no adverse reactions.
Usability (for OTC use):
Lay users understand indications, contraindications, and proper useUsability testing showed that lay users understand the indications for use, when not to use the device, how to turn it on, place it correctly, and how long to use it. (Specifically for RecoveryRx, implying a similar interface/device to ActiPatch for OTC use context).

2. Sample Sizes Used for the Test Set and Data Provenance

The "test set" in this context refers to the clinical study populations.

  • Chronic Cervical Osteoarthritis (Neck Pain) Study:
    • Sample Size: 200 intent-to-treat patients (197 completed).
    • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
  • Osteoarthritis of the Knee Study:
    • Sample Size: 66 intent-to-treat patients (60 completed).
    • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
  • Plantar Fasciitis (Heel Pain) Study:
    • Sample Size: 70 patients (all 70 completed).
    • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
  • Usability Testing:
    • Sample Size: 46 men and women.
    • Data Provenance: Not explicitly stated, but this would be a prospective study. No country of origin is mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable in the traditional sense for these types of clinical studies. The "ground truth" for pain reduction and adverse events is established by the patients themselves (self-reported pain scores like VAS, WOMAC) and clinical observation by study investigators/medical professionals for adverse events and objective measures.

  • Experts Involved: Medical professionals (physicians, nurses, study coordinators) would have been involved in diagnosing patients, conducting the study, and recording data, but they aren't "establishing ground truth" as adjudicators in the way an AI would be validated against expert labels.
  • Qualifications: Not specified for the study staff, but it's assumed they were qualified medical and research professionals appropriate for conducting clinical trials.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in diagnostic studies where multiple readers interpret images or data. This is not applicable here because the primary endpoints (pain scores, adverse events) rely on patient self-report and clinical observation, not multi-reader interpretation.

The studies were randomized and controlled, with some being double-blinded (knee osteoarthritis and plantar fasciitis studies), which serves to mitigate bias in data collection and assessment, rather than expert adjudication of an interpretive task.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done.

MRMC studies are typically used to assess the change in human reader performance when assisted by an AI. The ActiPatch is a medical device, and the studies evaluated its direct effect on patients, not its impact on human reader interpretations.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the primary clinical studies described are standalone performance studies of the device itself.

The ActiPatch device is applied directly to the patient and delivers therapy. Its effectiveness in reducing pain (the primary outcome) is measured directly, without a human intermediary interpreting its output. The studies assess the performance of the device "algorithm" (its therapeutic mechanism) by observing patient outcomes.


7. The Type of Ground Truth Used

The ground truth used in these studies is primarily:

  • Patient-reported outcomes (PROs):
    • VAS (Visual Analog Scale) pain scores: For all three pain studies.
    • WOMAC (Western Ontario and McMaster Universities Arthritis Index) scores: For the knee osteoarthritis study.
  • Clinical observation/assessment: For adverse events.
  • Patient understanding: For usability testing.

8. The Sample Size for the Training Set

Not applicable. The ActiPatch is a physical medical device (pulsed shortwave therapy) and not an AI/machine learning algorithm that requires a "training set" of data. Its mechanism of action is based on established physical principles, not learned patterns from data.


9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

January 31, 2020

BioElectronics Corporation Sree Koneru, Ph.D. VP, Product Development 4539 Metropolitan Ct Frederick, Maryland 21704

Re: K192234

Trade/Device Name: ActiPatch® Regulation Number: 21 CFR 890.5290 Regulation Name: Shortwave Diathermy Regulatory Class: Class II Product Code: PQY Dated: November 1, 2019 Received: November 4, 2019

Dear Dr. Koneru:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Vivek Pinto, Ph.D. Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192234

Device Name ActiPatch

Indications for Use (Describe) Adjunctive treatment of musculoskeletal pain

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary

1. Submitter's Name:BioElectronics Corporation
2. Address:4539 Metropolitan Court,Frederick, MD 21704, USAPhone: 301-500-2061Fax 301-874-6935
Contact Person:Sree N Koneru, Ph.D.VP, Product Development
3. Date Prepared:January 3, 2020
4. Trade Name:ActiPatch®
5. Common NameNonthermal Shortwave Therapy
6. Product Classification:21 CFR 890.5290 (b)Product Code: PQY
7. Predicate Devices:Primary Predicate: ActiPatch (K152432)
8. Description of Device:The ActiPatch® device is a pulsed shortwave therapydevice. The circuitry consists of low voltage (3 V)digital/analog electronics that control all timing functionsto produce the therapeutic radiofrequency (RF) field, wherethe antenna is placed directly above the therapeutic site.This closed loop system of the antenna, low energy signalgenerator circuit, and battery power supply, transfers theRF energy to the target tissue as a localized therapy with nofar field effects.
9. Intended Use:Adjunctive treatment of musculoskeletal pain
10. StandardsBS EN 980: 2003 Graphical Symbols For Use In TheLabeling of Medical DevicesISO 13485:2012 Medical Devices: QualityManagement SystemsISO 14971: 2012 Risk Management

{4}------------------------------------------------

IEC 60601-1:2005+A1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety and Essential Performance

IEC 60601 -1-2: 2007 Medical Electrical Equipment --Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests

EN 60601-2-3: 2014 Short-Wave Therapy Equipment

EN 60601-2-10: 2015 Safety of Nerve and Muscle Stimulators

MEDDEV 2.7.1 Rev. 4 Clinical Evaluation

MEDDEV 12.2-2 Rev. 2 Post Market Surveillance

11. Summary of technological characteristics:

The ActiPatch® device has the following technological characteristics (Table 1.) The ActiPatch operates at a carrier frequency of 27.12MHz, with a pulse frequency of 1000 Hz and a pulse width of 100 microseconds. The duty cycle is therefore 10%. The devices uses a 3v battery as the power source (CR 2032 or CR1632 or CR1620) and produces a peak spatial power density of 73 microWatts/cm2.

Table 1. Technological characteristics of the ActiPatch® Shortwave
Therapy Device
Carrier frequency27.12MHz
Peak spatial power density73 microwatts/ cm²
Pulse rate1000 pulses per second
Pulsed on duration100 micro seconds
Power sourceBattery CR2032 or CR1632 or CR1620
Antenna size12cm or 6 cm diameter
Treatment area110cm² or 30cm²
Weight9.5 grams
Operation time (lifetime ofbattery)720 hours or 360 hours or 168 hours
User ControlOn/Off switch or On-only (continuous use)switch
Recommended Treatment TimeMinimum of 12 hours per day, up to 24 hoursper day

{5}------------------------------------------------

12.Substantial Equivalence:

ActiPatch has the same intended use as the predicate device, i.e., the application of electromagnetic energy to non-thermally treat pain. The current ActiPatch's technological features are identical to those of the predicate ActiPatch device in K152432 (Table 2). No device modifications have been made to the current ActiPatch, when compared to the predicate ActiPatch device.

The predicate device is indicated for adjunctive treatment of musculoskeletal pain related to: 1) Osteoarthritis of the knee; and 2) Plantar fasciitis of the heel (K152432). The performance data submitted in the premarket notification, including the electrical safety, electromagnetic safety, biocompatibility, and clinical data described in Section 13 below, show that there is no difference in technology between the subject and predicate device, and that the subject ActiPatch is at least as safe and effective as the predicate ActiPatch.

ActiPatchPredicate ActiPatch (K152432)
Indication for UseAdjunctive treatment ofmusculoskeletal painAdjunctive treatment ofmusculoskeletal pain relatedto: (1) plantar fasciitis of theheel; and (2) osteoarthritis of theknee
TechnologyPulsed ShortwaveTherapy (Non- thermalDiathermy)Pulsed Shortwave Therapy(Non- thermal Diathermy)
Product codePQYPQY
Regulation21 CFR 890.5290(b)21 CFR 890.5290(b)
Classification NameShortwave DiathermyShortwave Diathermy
Anatomical Site ofapplicationSuperficial soft tissueSuperficial soft tissue
How the Energy iscoupledInduction CoilInduction Coil
Carrier Frequency27.1 MHz27.1 MHz

Table 2. Technological characteristics of the ActiPatch® and predicate ActiPatch®.

{6}------------------------------------------------

Pulse Duration100 micro-secs100 micro-secs
Pulse rate1000 Hz1000 Hz
Duty cycle10%10%
Power source3V DC (1 X CR2032or CR1632 or CR1620 Lithium Battery)3V DC (1 X CR2032 LithiumBattery)
Antenna Size110 cm² or 30 cm²110 cm² or 30 cm²
Average spatial powerdensity (RMS)4.4 µWatts/cm²4.4 µWatts/cm²
Specific adsorption rate(W/kg) (Peak)0.0007 W/kg0.0007 W/kg
Operation time (batterylifetime)720 hours or 360 hours or168 hours720 hours
Recommended treatmentduration (use time) basedon clinical evidenceMinimum of 12 hours perday, up to 24 hours perdayMinimum of 12 hours per day, upto 24 hours per day

13.Testing

Non-Clinical/Performance Data:

Electrical safety, electromagnetic safety, biocompatibility testing, and testing in accordance with the special controls of the October 13, 2015 Final Reclassification Order for Nonthermal Shortwave Therapy devices was performed for the ActiPatch.

The ActiPatch was tested for conformity to the following standards and was determined to conform to these standards:

  • General Safety and Requirements - Medical Equipment- IEC/EN 60601-1-2
  • General Safety and Requirements Medical Equipment- EN ● 60601-1:2006

Biocompatibility testing was conducted for the ActiPatch. The skin sensitization test performed in accordance with ISO 10993-10:2010 showed no evidence of an ActiPatch extract causing skin sensitization in guinea pigs. The skin irritation test conducted in accordance with ISO 10993-10:2010 demonstrated that gauze material saturated with extract from the ActiPatch showed no evidence of causing skin irritation in New Zealand white rabbits. The cytotoxicity test performed in accordance with ISO 10993-5:2009 showed that no observable in vitro cytotoxicity in L-

{7}------------------------------------------------

929 mouse fibroblast cells that were placed in contact with an extract prepared from ActiPatch.

The testing that was conducted in accordance with the special controls of the October 13, 2015 Final Reclassification Order demonstrated that the ActiPatch performs as intended under anticipated conditions of use. The testing determined and considered the peak output power; the pulse width; the pulse frequency; the duty cycle; the average measured output powered into the RF antenna/applicator; the specific absorption rates in a saline gel test load; the characterization of the electrical and magnetic fields in saline gel test load for each RF antenna and prescribed RF antenna orientation/position; and the characterization of the deposited energy density in saline gel test load.

Clinical Data:

The clinical data in this 510(k) includes results from three IRB approved, randomized, controlled studies. Additionally, usability testing was conducted to support the OTC use of the device.

  • A randomized, controlled trial on chronic cervical osteoarthritis (neck pain): This was a randomized, active-treatment controlled study to evaluate the safety and effectiveness of the ActiPatch medical device to reduce the pain level of patients diagnosed with cervical osteoarthritis. The active-treatment control was an NSAID of the Cox-2inhibitor family. There were 200 intent-to-treat patients, out of which 197 completed the four-week study. There were 142 women (71%) and 58 (29%) men in the study, with an average age of 45 years. The primary endpoint for efficacy was reduction in pain (VAS score) while at rest and being active, over four weeks, when compared to the beginning of the study. The results indicate that ActiPatch significantly reduced pain (measured by VAS pain) associated with COA in the device treatment group, and that the treatment differences between device-treatment and NSAID-treatment groups was significant (p<0.05).
  • A randomized controlled trial on osteoarthritis of the knee: The osteoarthritis of the knee study was a double-blinded, randomized, placebo-controlled study to evaluate the safety and effectiveness of the ActiPatch medical device to reduce the pain level of patients diagnosed with knee osteoarthritis. The placebo treatment was a device that was identical to ActiPatch but did not produce an

{8}------------------------------------------------

electromagnetic field when turned on. There were 66 intent-to-treat patients, out of which 60 patients completed the four-week study. There were 43 women (71.6%) and 17 (16.4%) men in the study, with the following average demographics at baseline: 68 years of age, BMI of 27.4 and disease duration of 12.1 years. The primary effectiveness endpoints were improvements in pain level over the four weeks as measured by the before and after VAS score and WOMAC scores, and the primary safety endpoint was all treatment-related adverse events during the study. The results indicate that ActiPatch significantly reduced pain (measured by VAS pain) associated with KOA in the device treatment group, and that the treatment differences between active and placebo treatment groups was significant (p<0.05).

  • A randomized controlled trial on plantar fasciitis (heel pain): This was a randomized, double-blinded, placebo-controlled study to evaluate the safety and effectiveness of the ActiPatch medical device to reduce the pain level of patients diagnosed with plantar fasciitis. The placebo treatment was a device that was identical to ActiPatch but did not produce an electromagnetic field when turned on. A total of 70 patients were recruited into the study, and all 70 completed the study. There were 52 women (74.4%) and 18 (25.6%) men in the study, with the following average demographics at baseline: 51.5 years of age, BMI of 31.8 and disease duration of 1.1 years. The primary effectiveness endpoint was the daily morning (AM) VAS score, and the primary safety endpoint was all treatment-related adverse events during the 7-day study. The results indicate that ActiPatch significantly reduced pain (measured by VAS pain) associated with PF in the device treatment group, and that the treatment differences between active and placebo treatment groups was significant (p<0.05).
  • Usability testing was conducted in 46 men and women over the age of 18 with a wide range of education levels who used the RecoveryRx on the knee, lower back, or shoulder. The testing showed that lay users understand the indications for use and when not to use the device. In addition, the study showed that users understand how to turn the device on, place it correctly on the right part of the body, and how long to use the device.

Conclusion: The non-clinical data, clinical data, and extensive real-world registry data demonstrate that the ActiPatch is at least as safe and effective

{9}------------------------------------------------

as the predicate device and can be used as an over-the-counter device for adjunctive treatment of musculoskeletal pain.

Note: Treatment effects of device use were clinically assessed for up to 4 weeks.

§ 890.5290 Shortwave diathermy.

(a)
Shortwave diathermy for use in applying therapeutic deep heat for selected medical conditions —(1)Identification. A shortwave diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body electromagnetic energy in the radiofrequency (RF) bands of 13.56 megahertz (MHz) or 27.12 MHz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.(2)
Classification. Class II (performance standards).(b)
Nonthermal shortwave therapy —(1)Identification. A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.(2)
Classification: Class II (special controls). The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.