    K Number
    K192234
    Device Name
    ActiPatch
    Date Cleared
    2020-01-31

    (165 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    PQY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive treatment of musculoskeletal pain

    Device Description

    The ActiPatch® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device: ActiPatch® (K192234)
    Intended Use: Adjunctive treatment of musculoskeletal pain


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state "acceptance criteria" in a typical quantitative format for algorithm performance (e.g., "sensitivity must be > X%, specificity > Y%"). Instead, the substantial equivalence hinges on demonstrating that the ActiPatch is "at least as safe and effective as the predicate ActiPatch (K152432)."

    Therefore, the "acceptance criteria" can be inferred from the studies' primary effectiveness endpoints, which consistently aimed to show a statistically significant reduction in pain compared to a control group (either active-treatment or placebo). The device's performance is described by achieving these significant reductions.

    Acceptance Criteria (Inferred from Study Goals)Reported Device Performance (Summary of Findings)
    Effectiveness (Pain Reduction):
    Reduction in pain for cervical osteoarthritis (neck pain) compared to active-treatment control (NSAID)ActiPatch significantly reduced pain (measured by VAS pain) associated with COA in the device treatment group. Treatment difference between device-treatment and NSAID-treatment groups was significant (p
    Ask a Question

    Ask a specific question about this device

    K Number
    K152432
    Device Name
    ActiPatch(R)
    Date Cleared
    2017-02-03

    (526 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    PQY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive treatment of musculoskeletal pain related to: (1) plantar fasciitis of the heel; and (2) osteoarthritis of the knee.

    Device Description

    The ActiPatch® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the ActiPatch® device. This type of regulatory submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel safety and effectiveness. Therefore, the information provided focuses on comparative performance and safety rather than establishing acceptance criteria as would be seen for a new medical device. The document does not explicitly state "acceptance criteria" but rather presents clinical study results as evidence of comparable safety and effectiveness to predicate devices.

    Here's an attempt to extract and organize the requested information based on the document's content:

    1. Table of Acceptance Criteria and Reported Device Performance

    As noted, the document doesn't define explicit "acceptance criteria" in the traditional sense of pre-defined thresholds. Instead, it presents clinical study results to demonstrate effectiveness and safety comparable to predicate devices. The "reported device performance" below summarizes these clinical findings.

    Criterion Type (Implicit)Acceptance Criteria (Not explicitly stated, but implied by study results)Reported Device Performance (ActiPatch®)
    Effectiveness - Knee OAImprovement in pain levels (VAS and WOMAC scores) comparable to or better than placebo/no treatment, and potentially supporting reduction in pharmacological therapy for osteoarthritis of the knee.Osteoarthritis of the knee study:
    • 36% of the treatment group reported a >30% decrease in VAS pain, compared to 9% for the placebo group.
    • 18% of the treatment group reported a >30% decrease in total WOMAC pain, compared to 3% for the placebo group.
    • 26% of the treatment group stopped pharmacological therapy, whereas 33% of the placebo group started new pharmacological therapy. |
      | Effectiveness - Plantar Fasciitis | Reduction in pain levels (daily morning VAS score) comparable to or better than placebo/no treatment for plantar fasciitis. | Plantar fasciitis study:
    • Average reported pain reduction (day 1 AM to day 7 AM VAS) for the treatment group was 40%, compared to 7% for the control group. |
      | Safety | No significant adverse events related to the device. | Osteoarthritis of the knee study: No adverse events were recorded.
      Plantar fasciitis study: Primary safety endpoint was all treatment-related adverse events; the document does not explicitly state if any were recorded but implies a positive safety profile through the "safe and effective as its predicate devices" conclusion. |
      | Usability (OTC Use) | Lay users must understand how to use the device safely and effectively for pain relief in the specified indications. | Usability testing:
    • 46 men and women (age >17, wide education levels) demonstrated understanding of indications for use, contraindications, how to turn on, correct placement, and duration of use. |
      | Non-Clinical Performance | Conformity to relevant electrical safety, electromagnetic compatibility, and biocompatibility standards. Performance as intended under anticipated conditions of use regarding output power, pulse characteristics, absorption rates, and electromagnetic fields. | Non-Clinical/Performance Data:
    • Conformed to IEC/EN 60601-1-2:2012, IEC 60601-1:2005+A1:2012, EN 60601-1:2006.
    • Biocompatibility testing (ISO 10993-10:2010 for sensitization and irritation, ISO 10993-5:2009 for cytotoxicity) showed no evidence of sensitization, irritation, or in vitro cytotoxicity.
    • Special controls testing (per October 13, 2015 Final Reclassification Order) demonstrated intended performance, characterizing peak output power, pulse characteristics, duty cycle, average output power, specific absorption rates, and electrical/magnetic fields in a saline gel test load. |

    2. Sample Size Used for the Test Set and the Data Provenance

    • Osteoarthritis of the knee study:
      • Sample size: 66 intent-to-treat patients, with 60 patients completing the four-week study.
      • Data Provenance: Not explicitly stated (e.g., country of origin, specific locations), but described as "double-blind randomized controlled study" implying a prospective design.
    • Plantar fasciitis study:
      • Sample size: 70 patients completed the study.
      • Data Provenance: Described as "double-blind, multicenter, randomized, placebo-controlled study," implying a prospective design. Country of origin not specified.
    • Usability testing:
      • Sample size: 46 men and women.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective), but the description ("These subjects demonstrated use...") implies a prospective, interactive study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" for the test set in the context of image interpretation or diagnostic accuracy studies. The clinical studies evaluated patient-reported outcomes (VAS and WOMAC scores) and adverse events, which are direct measures from the participants themselves, rather than interpretations by experts.


    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    The document describes the clinical studies as "double-blind and placebo-controlled randomized controlled trials." This implies that blinding was used for both participants and researchers/assessors regarding treatment allocation. However, it does not specify an adjudication method for outcome measures like VAS or WOMAC scores, which typically do not require such adjudication as they are direct patient responses. For adverse events, it states "No adverse events were recorded" for the knee study; the method for reviewing or adjudicating AEs in the plantar fasciitis study is not detailed beyond "primary safety endpoint was all treatment-related adverse events."


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The ActiPatch® is a physical medical device for pain relief, not an AI or imaging interpretation device that would typically involve human readers.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The ActiPatch® is a physical device, not an algorithm. Its performance is inherent in its physical and electrical characteristics and its interaction with the patient's body. The clinical studies evaluate the device's effect directly on patients.


    7. The Type of Ground Truth Used

    For the clinical studies, the "ground truth" was primarily based on patient-reported outcomes:

    • Pain levels: Measured using validated scales like the Visual Analog Scale (VAS) and Western Ontario and McMaster Universities Arthritis Index (WOMAC) scores.
    • Adverse events: Self-reported by patients or observed during the study.

    For usability testing, the "ground truth" was established by observing lay user comprehension and ability to perform tasks (turn on, place correctly, understand indications/contraindications).

    For non-clinical testing, the "ground truth" was conformance to established engineering and biological standards (e.g., IEC, ISO standards for electrical safety, biocompatibility, and specific power output measurements).


    8. The Sample Size for the Training Set

    The document does not describe the use of a "training set" as it would for an algorithm or AI model development. The clinical studies described are for evaluating the final device's performance, not for training a model.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was explicitly mentioned or used in the context of algorithm development. The document describes clinical trials and usability studies of a physical medical device.