K Number
K192063
Date Cleared
2020-04-24

(267 days)

Product Code
Regulation Number
866.6080
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PGDx elio™ tissue complete assay is a qualitative in vitro diagnostic device that uses targeted next generation sequencing of DNA isolated from formalin-fixed, paraffin-embedded tumor tissue from patients with solid malignant neoplasms to detect tumor gene alterations in a broad multi-gene panel.

PGDx elio tissue complete is intended to provide tumor mutation profiling information on somatic alterations (SNVs, small insertions and deletions, one amplification and four translocations), microsatellite instability (MSI) and tumor mutation burden (TMB) for use by qualified healthcare professionals in accordance with professional guidelines in oncology for previously diagnosed cancer patients, and is not conclusive or prescriptive for labeled use of any specific therapeutic product.

Device Description

PGDx elio tissue complete is an in vitro diagnostic assay that uses NGS to detect turnor gene alterations in genomic DNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue from a variety of tumor types, using a targeted panel (505 genes). The assay takes less than 7 days from DNA to report and provides information on single nucleotide variants (SNVs) in a range of GC content and genomic contexts, insertion/ deletions (indels), 1 amplification as well as 4 translocations. It also identifies microsatellite instability based on select mononucleotide tracts and signatures of sequence mutations. The PGDx elio tissue complete assay utilizes a ~1.3 Mb region of interest (ROI) to calculate tumor mutation burden (TMB).

AI/ML Overview

The provided text describes the analytical validation studies for the PGDx elio tissue complete assay, a qualitative in vitro diagnostic device for tumor profiling.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria & Reported Device Performance

The acceptance criteria are generally established as target performance metrics (e.g., call rates, concordance rates, %CV, false positive rates). The reported device performance is the outcome of the analytical studies.

Performance Metric CategorySpecific Metric (Acceptance Criteria Implicitly Defined by Study Design/Results)Reported Device Performance
SpecificityFalse Positive Rate (SNVs and Indels)< 0.1% for variants with evidence of clinical significance; < 3.2% for hotspot SNVs (n=2/63).
False Positive Rate (MSI-H)< 1.6%
Sensitivity (LoD)SNVs (Hotspot) - Lowest MAF for $\ge$95% detection3.1% to 5.4%
SNVs (Non-hotspot) - Lowest MAF for $\ge$95% detection6.3% to 17.8%
Indels (Homopolymer context) - Lowest MAF for $\ge$95% detection13.7% to 17.5%
Indels (Non-homopolymer context) - Lowest MAF for $\ge$95% detection6.1% to 10.9%
MSI-H - LoD Tumor Purity for $\ge$95% detection18.1%
ERBB2 Amplifications - LoD Tumor Purity for $\ge$95% detection4.4%
ALK Translocations - LoD Tumor Purity for $\ge$95% detection5.6%
NTRK2 Translocations - LoD Tumor Purity for $\ge$95% detection30% (in silico suggests 3%)
NTRK3 Translocations - LoD Tumor Purity for $\ge$95% detection11.5%
RET Translocations - LoD Tumor Purity for $\ge$95% detection12.8%
TMB PerformancePrecision (CV) for TMB at $\ge$15% tumor purity3.3% to 15.9%
DNA Extraction Method CompatibilityConcordance with Method 1 (PPA)Method 2: PPA 97.8%; Method 3: PPA 97.5%
Concordance with Method 1 (NPA)Method 2: NPA 99.9%; Method 3: NPA 99.9%
DNA Input RobustnessConcordance (PPA) at varying DNA input vs. 100 ng10 ng: PPA 92.2%; 25 ng: PPA 94.8%; 50 ng: PPA 96.9%; 200 ng: PPA 97.4%
Concordance (NPA) at varying DNA input vs. 100 ngAll > 99.9%
TMB Mean Absolute Percent Error (MAPE) vs. 100 ng1.8% to 11.8%
Contamination ControlAbsence of false positives in negative samples from carryover/cross-contaminationNo positive variant results observed in known negative samples.
Exogenous InterferenceConcordance (PPA) with interfering substances vs. baseline> 97.2%
Concordance (NPA) with interfering substances vs. baseline> 99.9%
TMB MAPE with interfering substances0% to 6.0%
Overall Sample Acceptance RateFirst Pass Rate81.8% (2352/2874)
Overall Pass Rate (allowing single repeat)92.9% (2671/2874)
Accuracy (Concordance to Orthogonal Methods)PPA/NPA for various variant types (SNVs, indels, amplifications, translocations)See Table 1.10 for detailed SNV/Indel PPAs (range 80.8% - 100%) and NPAs (all 99.9% or 100%)ERBB2 Amplifications: PPA 75.0% (all cases), 87.0% (excluding borderline); NPA 96.7% (all cases), 95.9% (excluding borderline)ALK Translocations: PPA 92.9%; NPA 98.2%RET Translocations: PPA 55.6%; NPA 100%TMB: Spearman correlation coefficient 0.903 vs WESMSI: Overall PPA 98.8% (excluding failures), 94.0% (accounting for failures); NPA 99.3% (excluding failures), 77.6% (accounting for failures)
Interlaboratory ReproducibilityFirst Pass Rate90.3% (455/504)
Overall Pass Rate98.2% (495/504)
Overall Positive Call Rate (All Variants)86.2%
APA (all variants)> 92%
ANA (all variants)> 99%
TMB %CV3.5%
Lot to Lot PrecisionAPA (Variants with Evidence of Clinical Significance)96.1% to 98.7%
ANA (Variants with Evidence of Clinical Significance)99.8% to 99.9%
TMB %CV< 10% (7.1% to 9.5%)

2. Sample Size Used for the Test Set and Data Provenance

  • Specificity (LoB): 2 NIST reference standards (NA24531, NA24385) with 5 replicates each. Unique test cases from normal FFPE samples (n=63 for hotspot SNVs).
  • Sensitivity (LoD) - SNVs and Indels: 10 unique clinical cases with 10 replicates each across 2 kit lots (total 200 observations). Additional evaluations: 11 SNVs, 3 insertions, 5 deletions from 5 clinical FFPE specimens with 5 replicates per dilution level. Cell lines involved 3 samples with 10 replicates at 5 dilution levels (150 observations).
  • Sensitivity (LoD) - ERBB2, ALK, RET, NTRK2, NTRK3, MSI: 7 clinical FFPE cases (diluted) with 10 replicates per kit lot across 2 lots for translocations/amplifications. 3 MSI-H cases with 10 replicates each.
  • TMB and Tumor Purity: 8 clinical FFPE cases. Samples 1-5: 3 levels, 5 replicates each (18 total per sample). Sample 6: 5 levels, 10 replicates each (50 total). Samples 7-8: 5 levels, 5 replicates each (25 total).
  • DNA Extraction: 3 FFPE specimens and 1 cell line, extracted in duplicate by 2 operators using 3 methods (48 total extraction results), processed in duplicate (96 observations).
  • DNA Input: 4 unique FFPE samples in triplicate at 5 input levels (10, 25, 50, 100, 200 ng).
  • Sample Carryover & Cross-contamination: 29 FFPE samples (7 positive, rest negative).
  • Interfering Substances: DNA from FFPE samples tested in presence of substances (5 test cases, 8 experimental + 2 baseline conditions).
  • Endogenous Interference (Necrosis/Age): 448 samples for necrosis (0-75%), 378 for age (0-253 months) from the accuracy study enrollment (521 total).
  • Overall Clinical FFPE Sample Acceptance Rate: 2874 unique clinical cases from >40 tumor types.
  • Accuracy: 582 samples with PGDx elio tissue complete data and orthogonal data. Specifically:
    • SNVs/Indels: 582 samples.
    • ERBB2 Amplifications (FISH comparison): 147 cases (all); 120 cases (excluding borderline).
    • ALK Translocations (FISH comparison): 71 cases. Additional in silico simulation on 10 clinical samples (410 observations).
    • RET Translocations (FISH comparison): 27 cases. 3 RET translocation-positive cell lines also tested.
    • TMB: 118 cases across 8 tumor types.
    • MSI: 283 samples across 18 tumor types.
    • Wild Type Calls: 112 specimens.
  • Interlaboratory Reproducibility: 13 FFPE tissue specimens and 1 cell line (14 samples total), tested in duplicate by 2 operators on 12 sequencing runs across 3 non-consecutive days at 3 independent laboratory sites (504 total replicates).
  • Lot to Lot Precision: 5 test cases in triplicate across 3 unique kit lots (45 observations).

Data Provenance: The document does not explicitly state the country of origin for the clinical samples. However, it mentions "clinical FFPE specimens," implying real-world patient data. The studies are described in a factual manner, suggesting they are retrospective analyses of collected samples for analytical validation purposes.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The document does not specify the number or qualifications of experts involved in establishing the ground truth for the test set.

Instead, the ground truth for performance evaluation (accuracy) is established by:

  • Orthogonal methods:
    • "2 NGS targeted panels" and "PCR" (for SNVs and Indels).
    • "ERBB2 FISH," "ALK FISH," "RET FISH" (for amplifications and translocations).
    • "matched tumor-normal whole exome sequencing results" (for TMB).
    • "MSI PCR" (for MSI).
  • Validated assays/literature: For the 3 RET translocation-positive cell lines.

4. Adjudication Method for the Test Set

The document does not describe any expert adjudication process for resolving discrepancies or establishing ground truth for the test set. The ground truth is primarily based on the results from the orthogonal methods. For the in silico studies, the ground truth is derived from down-sampling "clinical samples."

5. MRMC Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned or performed. This device is a molecular diagnostic assay, not an imaging AI tool, and thus comparative effectiveness with human readers improving with AI assistance is not applicable in this context.

6. Standalone Performance

Yes, the entire document describes studies of the standalone (algorithm only, without human-in-the-loop performance) of the PGDx elio tissue complete assay. The goal is to demonstrate the analytical performance of the automated workflow, from sample preparation to data analysis and variant calling, against established ground truth methods.

7. Type of Ground Truth Used

The primary type of ground truth used is orthogonal methods, specifically:

  • Other NGS targeted panels: For SNVs and indels.
  • PCR: For certain SNVs/indels and MSI.
  • FISH (Fluorescence In-Situ Hybridization): For gene amplifications (ERBB2) and translocations (ALK, RET).
  • Whole Exome Sequencing (WES): For Tumor Mutation Burden (TMB).
  • Cell line characterization/literature: For known variants in control cell lines.

This is a form of reference standard ground truth, where the device's performance is compared against established, validated measurement techniques.

8. Sample Size for the Training Set

The document details analytical validation (testing) and reproducibility studies. It does not provide information about a "training set" size. As a molecular diagnostic assay, its development likely involves internal data for algorithm development and optimization, but specific training set sizes are not mentioned in this regulatory submission, which focuses on validation data.

9. How the Ground Truth for the Training Set Was Established

Since the document does not discuss a specific "training set" or its size, it also does not describe how ground truth for such a set was established. The focus is on the performance of the final, locked-down algorithm against independent test data with ground truth established by orthogonal methods.

{0}------------------------------------------------

April 24, 2020

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

Personal Genome Diagnostics Jennifer Dickey, Ph.D., RAC VP, Regulatory and Quality 2809 Boston Street, Suite 503 Baltimore, Maryland 21224

Re: K192063

Trade/Device Name: PGDx™ elio tissue complete Regulation Number: 21 CFR 866.6080 Regulation Name: Next generation sequencing based tumor profiling test Regulatory Class: Class II Product Code: PZM Dated: July 31, 2019 Received: August 1, 2019

Dear Jennifer Dickey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Donna Roscoe, Ph.D. Chief Division of Molecular Genetics and Pathology OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192063

Device Name PGDx elio™ tissue complete

Indications for Use (Describe)

The PGDx elio™ tissue complete assay is a qualitative in vitro diagnostic device that uses targeted next generation sequencing of DNA isolated from formalin-fixed, paraffin-embedded tumor tissue from patients with solid malignant neoplasms to detect tumor gene alterations in a broad multi-gene panel.

PGDx elio tissue complete is intended to provide tumor mutation on somatic alterations (SNVs, small insertions and deletions, one amplification and four translocations), microsatellite instability (MSI) and tumor mutation burden (TMB) for use by qualified healthcare professionals in accordance with professional guidelines in oncology for previously diagnosed cancer patients, and is not conclusive or prescriptive for labeled use of any specific therapeutic product.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, with the "D" having an orange accent. To the right of the letters, the words "Personal Genome Diagnostics" are written in a smaller, blue font. The letters PGD are large and bold, making them the most prominent part of the logo.

510(k) Summary

Submission Date: April 24, 2019

Submitter Information:

Submitted By: Personal Genome Diagnostics Inc. 2809 Boston Street, Suite 503 Baltimore, MD 21224

Contact Person: Jennifer S. Dickey PhD, RAC Vice President, Regulatory & Quality Personal Genome Diagnostics Tel: (443) 602-8833 Email: jdickey@pgdx.com

A. Proprietary and Established Names

PGDx elio™ tissue complete

B. 510(k) number

K192063

C. Measurand

Somatic single nucleotide variants, insertions and deletions, select amplifications and translocations, microsatellite instability and tumor mutation burden in human genomic DNA obtained from formalin-fixed, paraffin-embedded tumor tissue.

D. Regulatory Information

1. Regulation section

21 CFR 866.6080

2. Classifications

Class II

3. Product Code

PZM

E. Indications for Use

1. Indications for Use

The PGDx elio™ tissue complete assay is a qualitative in vitro diagnostic device that uses targeted next generation sequencing of DNA isolated from formalin-fixed, paraffin-

CONFIDENTIAL

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in blue, with a small "TM" symbol next to the word "Diagnostics".

embedded tumor tissue from patients with solid malignant neoplasms to detect tumor gene alterations in a broad multi-gene panel.

PGDx elio tissue complete is intended to provide tumor mutation profiling information on somatic alterations (SNVs, small insertions and deletions, one amplification and four translocations), microsatellite instability (MSI) and tumor mutation burden (TMB) for use by qualified healthcare professionals in accordance with professional guidelines in oncology for previously diagnosed cancer patients, and is not conclusive or prescriptive for labeled use of any specific therapeutic product.

2. Special conditions for use statement(s):

For prescription use.

For in vitro diagnostic use.

3. Special Instrument Requirements

NextSeq® 550Dx (qualified by PGDx)

F. Substantial Equivalence Information:

    1. Predicate device name(s)
      MSK-IMPACT
    1. Predicate 510(k) Number
      DEN170058

3. Comparison with predicate

PGDx elio tissue completeMSK-IMPACT(Predicate Device)
K NumberK192063DEN170058
SIMILARITIES
AssayIntended Use/ Indicationsfor UseThe PGDx elio™ tissue completeassay is a qualitative in vitrodiagnostic device that uses targetednext generation sequencing of DNAisolated from formalin-fixed,paraffin-embedded tumor tissuefrom patients with solid malignantneoplasms to detect tumor geneThe MSK-IMPACT assay is aqualitative in vitro diagnostic testthat uses targeted next generationsequencing of formalin-fixedparaffin-embedded tumor tissuematched with normal specimensfrom patients with solid malignantneoplasms to detect tumor genealterations in a broad multi genepanel. The test is intended to provide
PGDx elio tissue completeMSK-IMPACT(Predicate Device)
alterations in a broad multi-genepanel.PGDx elio tissue complete isintended to provide tumor mutationprofiling information on somaticalterations (SNVs, small insertionsand deletions, one amplificationand four translocations),microsatellite instability (MSI) andtumor mutation burden (TMB) foruse by qualified healthcareprofessionals in accordance withprofessional guidelines in oncologyfor previously diagnosed cancerpatients, and is not conclusive orprescriptive for labeled use of anyspecific therapeutic product.information on somatic mutations(point mutations and small insertionsand deletions) and microsatelliteinstability for use by qualified healthcare professionals in accordancewith professional guidelines, and isnot conclusive or prescriptive forlabeled use of any specifictherapeutic product. MSK-IMPACTis a single-site assay performed atMemorial Sloan Kettering CancerCenter.
ClassificationIISame
Product CodePZMSame
Regulation21 CFR 866.6080Same
Assay MethodQualitativeSame
Sample TypeFFPE tumor tissue from cancerpatients with solid malignantneoplasmsSame
Target PopulationPreviously diagnosed cancerpatients with solid malignantneoplasmsSame
Mode ofMeasurementPCR and Next GenerationSequencing (hybrid capturemethodology)Same
CalibratorsNoneSame
PGDx elio tissue completeMSK-IMPACT(Predicate Device)
DNA Input100 ng100-250 ng
DIFFERENCES
TestEnvironmentKitSingle-site assay (performed atMemorial Sloan Kettering CancerCenter)
ControlsPositive control, negative control,normalized to database of commongermline SNPsMatched normal, positive controland negative control
Assay TargetSNVs and indels in 505 genes MSI Amplification in ERBB2 Translocations in ALK,RET, NTRK2, and NTRK3 TMBSNVs and indels in 468 genes MSI
InstrumentNextSeq 550Dx (qualified byPGDx)HiSeq 2500 Sequencer (qualified byMSK)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Personal Genome Diagnostics, or PGDx. The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters are bolded and the text is in a sans-serif font.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "X" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in blue. The letters are in a sans-serif font.

G. Summary and Explanation

4. Product Description

PGDx elio tissue complete is an in vitro diagnostic assay that uses NGS to detect turnor gene alterations in genomic DNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue from a variety of tumor types, using a targeted panel (505 genes). The assay takes less than 7 days from DNA to report and provides information on single nucleotide variants (SNVs) in a range of GC content and genomic contexts, insertion/ deletions (indels), 1 amplification as well as 4 translocations. It also identifies microsatellite instability based on select mononucleotide tracts and signatures of sequence mutations. The PGDx elio tissue complete assay utilizes a ~1.3 Mb region of interest (ROI) to calculate tumor mutation burden (TMB). Figure 1.1 describes components of the assay. A complete list of components, equipment and materials can be found in Part 2 (User Guide) of the PGDx elio tissue complete Manual (MN-ETC-03). The panel gene list is provided in Table 1.1.

{7}------------------------------------------------

Image /page/7/Figure/3 description: The image shows the assay components of PGDx elio tissue complete. The left side of the image shows the assay components provided by PGDx, which include PGDx elio tissue complete, PGDx elio server, and PGDx elio platform. The right side of the image shows the assay components provided by the user, which include DNA extraction method, molecular laboratory equipment, NextSeq 550Dx sequencing reagents, and NextSeq 550Dx sequencer.

Figure 1.1 PGDx elio tissue complete assay components. PGDx provided components include reagent kits, software for data analysis, and a server.

Table1.1
------------
PGDx elio tissue complete Gene List‡
ABL1ABL2ACVR1ACVR1BADORA2AAKT1AKT2AKT3
ALK*ALOX12BAMER1APCARARAFARFRP1ARID1A
ARID1BARID2ARID5BASXL1ASXL2ATMATRATRX
AURKAAURKBAXIN1AXIN2AXLB2MBAP1BARD1
BBC3BCL2BCL2L1BCL2L11BCL2L2BCL6BCORBCORL1
BCRBIRC2BLMBMPR1ABRAFBRCA1BRCA2BRD4
BRIP1BTG1BTG2BTKBUB1BC11ORF30CALRCARD11
CASP8CBFBCBLCCND1CCND2CCND3CCNE1CD22
CD274CD276CD70CD79ACD79BCDC73CDH1CDK12
CDK4CDK6CDK8CDKN1ACDKN1BCDKN1CCDKN2ACDKN2B
CDKN2CCEBPACHD2CHD4CHEK1CHEK2CICCREBBP
CRKLCSF1CSF1RCSF2CSF3CSF3RCTCFCTLA4
CTNNA1CTNNB1CUL3CUL4ACXCR2CXCR4CYLDCYP17A1
DAXXDCUN1D1DDB2DDR1DDR2DICER1DIS3DNMT1
DNMT3ADNMT3BDOT1LE2F3EEDEGFL7EGFREIF1AX
EP300EPAS1EPCAMEPHA2EPHA3EPHA5EPHA7EPHB1
EPHB4ERBB2#ERBB3ERBB4ERCC1ERCC2ERCC3ERCC4
ERCC5ERCC6ERCC8ERGERRFI1ESR1ETV1ETV4
ETV5ETV6EWSR1EXT1EXT2EZH2FAM175FAM46CA
FANCAFANCBFANCCFANCD2FANCEFANCFFANCGFANCI
FANCLFANCMFASFAT1FBXW7FGF10FGF12FGF14
FGF19FGF23FGF3FGF4FGF6FGFR1FGFR2FGFR3
FGFR4FHFLCNFLT1FLT3FLT4FOXA1FOXL2
FOXP1FRS2FUBP1GABRA6GATA1GATA2GATA3GATA4
GATA6GID4GLI1GNA11GNA13GNAQGNASGPC3
GPR124GREM1GRIN2AGRM3GSK3BH3F3AH3F3BH3F3C
HDAC1HDAC2HDAC6HGFHIST1H1CHIST1H2BDHIST1H3BHNF1A
HRASHSD3B1HSP90AA1HSP90AB1ICOSLGID3IDH1IDH2
IFNGR1IGF1IGF1RIGF2IGF2RIKBKEIKZF1IL10
IL7RINHBAINPP4AINPP4BINSRIRF2IRF4IRS1
IRS2JAK1JAK2JAK3JUNKAT6AKDM5AKDM5C
KDM6AKDRKEAP1KELKITKLF4KLHL6KMT2A
KMT2CKMT2DKRASLATS1LATS2LMO1LRP1BLTK
LYNLZTR1MAFMAGI2MAML1MAP2K1MAP2K2MAP2K4
MAP3K1MAP3K13MAPK1MAXMCL1MDC1MDM2MDM4
MED12MEF2BMEN1MERTKMETMITFMKNK1MLH1
MLH3MPLMRE11AMSH2MSH3MSH6MST1RMTAP
MTORMUTYHMYBMYCMYCLMYCNMYD88MYOD1
NBNNCOA3NCOR1NF1NF2NFE2L2NFKBIANKX2-1
NKX3-1NOTCH1NOTCH2NOTCH3NOTCH4NPM1NRASNSD1
NT5C2NTRK1NTRK2*NTRK3*NUP93NUTM1PAK1PAK3
PAK7PALB2PARK2PARP1PARP2PARP3PAX5PAX8
PBRM1PDCD1PDCD1LG2PDGFRAPDGFRBPDK1PDPK1PHOX2B
PIK3C2BPIK3C2GPIK3C3PIK3CAPIK3CBPIK3CDPIK3CGPIK3R1
PIK3R2PIK3R3PIM1PLCG2PLK2PMAIP1PMS1PMS2
PNRC1POLD1POLEPOLHPOT1PPARGPPP2R1APPP2R2A
PRDM1PREX2PRKAR1APRKCIPRKDCPRSS1PRSS8PTCH1
PTENPTK2PTPN11PTPRDPTPROPTPRSPTPRTQKI
RAC1RAD21RAD50RAD51RAD51BRAD51CRAD51DRAD52
RAD54BRAD54LRAF1RANBP2RARARASA1RB1RBM10
RECQL4RELRET*RFWD2RHOARICTORRIT1RNF43
ROS1RPA1RPS6KA4RPS6KB2RPTORRUNX1RUNX1T1RYBP
SBDSSDHASDHAF2SDHBSDHCSDHDSETD2SF3B1
SGK1SH2D1ASHQ1SLIT2SLX4SMAD2SMAD3SMAD4
SMARCA4SMARCB1SMARCD1SMOSNCAIPSOCS1SOX10SOX17
SOX2SOX9SPENSPOPSPTA1SRCSTAG2STAT3
STAT4STK11STK40SUFUSUZ12SYKTAF1TBX3
TEKTERCTERTTET1TET2TGFBR1TGFBR2TIPARP
TLR4TLR7TLR8TLR9TMEM127TMPRSS2TNFAIP3TNFRSF14
TOP1TOP2ATP53TP53BP1TP63TRAF7TSC1TSC2
TSHRTYRO3U2AF1VEGFAVHLVTCN1WASWEE1
WHSC1WHSC1L1WISP3WRNWT1XIAPXPAXPC
XPO1XRCC1XRCC2XRCC3YAP1YES1ZBTB2ZNF217
ZNF703

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font. The letters "TM" are in superscript.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "x". To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size, with a trademark symbol.

: - All genes listed contribute to TMB score and SNV/indel reporting: * - Translocations reported for this gene: # - Amplifications reported for this gene

5. Sample Preparation

The PGDx elio tissue complete assay requires genomic DNA isolated from FFPE tissue specimens using commercially available DNA extraction methods. The assay is validated for use with DNA recovered from tissue with a minimum of 20% viable tumor nuclei. If less than 100% of the tissue section contains ≥20% tumor purity, the tissue should be macro-dissected to select as much viable tumor as possible and minimize the amount of adjacent non-tumor tissue. The recommended DNA input for the assay is 100 ng at a minimum concentration of 1 ng/μL; results can be obtained with inputs down to 50 ng.

6. Library Preparation

The PGDx elio tissue complete assay workflow begins with genomic DNA. Genomic DNA is quantified using a fluorometer. DNA molecules are mechanically sheared to a target size of 200 bp and subjected to a magnetic bead purification step to remove smaller fragments and perform an exchange of buffer.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for Personal Genome Diagnostics. The logo has the letters PGD in blue and the letter X in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters PGD are in a bold sans-serif font.

Fragmented DNA is end-repaired, phosphorylated, and adenylated. Indexed adapters are then ligated to the A-tailed DNA molecules. Unincorporated adapters and reagents are removed by magnetic bead purification. Adapter-ligated DNA is enriched by PCR amplification. Primer dimers and residual reagents are removed by magnetic bead purification. Library quality is assessed using a DNA fragment analyzer prior to hybrid capture.

7. Hybrid Capture NGS

The adapter-ligated library is hybridized with biotinylated RNA library baits, and targeted regions are captured using magnetic streptavidin coated beads. Captured libraries are purified to remove baits and incompletely hybridized DNA fragments. Captured libraries are enriched by PCR amplification. Primer dimers and residual reagents are removed by magnetic bead purification. Final library quality is assessed using a DNA fragment analyzer prior to sequencing.

8. Sequencing

Sample libraries are quantified and normalized into a sequencing pool of up to 15 samples and the external control. Pooled sample libraries are fluorometrically quantified, loaded on a sequencing flow cell and sequenced using a NextSeq® 550Dx instrument which has been pre-qualified by PGDx.

9. Data Analysis

Sequence data is processed using the PGDx elio platform software. The software contains a user interface that tracks sample status from sequencing through analysis and reporting. Users configure sequencing runs, and an automated pipeline of software for bioinformatic analysis identifies and reports genomic alterations. After processing, the software generates FASTQ files containing sequences and quality scores for each sample. The FASTQ files are then aligned to a reference genome to generate BAM files, which are processed for variant calling of different alteration types (SNVs, indels, amplifications, translocations, and MSI). SNVs and indels are then used to determine TMB scores reported as mutations per megabase.

10. Controls

  • i. Negative Control: A no template control (NTC) can be processed to serve as a negative control to validate the acceptability of all the test samples processed through library preparation and capture steps by testing for sample or reagent contamination. The NTC is not included on the sequencing run.
  • ii. Positive Control: An external control that is provided in the PGDx elio tissue complete assay reagent kit consists of cell line derived-DNA with multiple verified sequence mutations. The external control is processed from library preparation through sequencing to serve as an end to end control to demonstrate assay

CONFIDENTIAL

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters and text are all aligned horizontally.

performance. The external control is checked for quality during library preparation and after sequencing. Failure of the external control to meet the pre-defined quality metrics will result in all test samples on the run being reported as 'No result.'

11. Results Reporting

PGDx elio tissue complete reports SNVs and indels in protein coding regions across all genes in the panel. In addition, amplifications are reported for ERBB2 as well as translocations for ALK, RET, NTRK2, and NTRK3. The assay also reports on 2 genomic signatures, MSI and TMB.

The variants listed in the section Variants with Evidence of Clinical Significance are determined based on the selected tumor type. Only variants clinically associated with the tested tumor type will appear in the Variants with Evidence of Clinical Significance section. Any remaining detected variants will appear as the Variants with Potential Clinical Significance. A qualified healthcare professional can select the appropriate tumor type, and depending on the tumor type selected, variants will be reported as Variants with Evidence of Clinical Significance by PGDx elio tissue complete. Any variants clinically associated with tumor types other than the one selected will be reported in the section labeled 'Variants with Potential Clinical Significance. A list of all genes is provided in Appendix A.

SNVs and indels results are also presemted in terms of somatic hotspot or non-hotspot. PGDx defines hotspots as > 25 exact hits in COSMIC version 72. A lower minimum MAF is used when reporting these hotspot mutations.

Quality MetricLevel of QualificationPassing Criteria
Cluster DensityBatch-levelSequencer Cluster Density $≥$ 130
Q30 ReadsBatch-level%Q30 (Read1 and Read4) $≥$ 80%%Q30 (Read2 and Read3) $≥$ 85%
External ControlBatch-levelAll expected sequence mutations are detected and passes all other quality criteria
Percent Regions CoveredSample-level$≥$ 90% exons with > 100x Median Distinct Coverage
Percent Reads IdentifiedSample-levelPercent Reads Identified 15%-35%
Contamination QCSample-levelEstimated contamination levels < 2%

Table 1: Summary of PGDx elio tissue complete Quality Control Metrics Postsequencing

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image is a logo for Personal Genome Diagnostics, or PGDx. The letters "PGD" are in blue, and the "x" is in yellow. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller blue font. The letters are stacked on top of each other.

Quality MetricLevel ofQualificationPassing Criteria
Select SNVsand Indels withEvidence ofClinicalSignificanceAnalyte-levelMutant reads $\ge$ 4MAF > 0.4%
Hotspot SNVsand IndelsAnalyte-levelMutant reads $\ge$ 4MAF > 2%
Non-hotspotSNVsAnalyte-levelMutant reads $\ge$ 6MAF with lower bound 95% CI $\ge$ 5%
Non-hotspotIndelsAnalyte-levelMutant reads $\ge$ 6MAF > 5%
HomopolymerIndelsAnalyte-levelHomopolymer regions < 5 bp orHomopolymer regions $\ge$ 5 bp with MAF $\ge$ 12%
ERBB2AmplificationsAnalyte-levelFold change $\ge$ 2.5 in $\ge$ 25% regions covered
Translocations(ALK, NTRK2,NTRK3 andRET)Analyte-levelFusion reads $\ge$ 3

12. Analytical Studies

13. Specificity

i. Limit of Blank (LoB)

Non-cancerous FFPE tissues were assessed for analytical specificity to confirm the reporting thresholds and quality metrics minimize false positives. Two reference standards from National Institute of Standards and Technology (NIST), NA24531 and NA24385 were evaluated by PGDx elio tissue complete for variants reported at the 100ng DNA input. Specificity was observed at 100% with no unverified mutations reported across 5 replicates for each standard. Unique test cases from normal FFPE samples were processed with the recommended 100 ng DNA input across 2 different lots of the PGDx elio tissue complete assay kit. For Variants with Evidence of Clinical Significance, the rate of false positives is < 0.1% while the false positive rate for hotspot SNVs is < 3.2% (n=2/63). For MSI-H, the false positive rate is < 1.6%.

  • ii. Cross Reactivity

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "X" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller, light blue font.

An in silico cross-reactivity analysis was performed to evaluate specificity of capture baits based on sequence identity of target baits to the human reference genome. Bait sequences corresponding to gene regions of clinical significance demonstrated specificity to target regions based on mapping quality or unique sequence identity to the human reference genome. Any bait that had some sequence identity to regions other than the target region of interest was further investigated and were shown to be specific. This study demonstrated that the oligonucleotide baits in the PGDx elio tissue complete assay are specific to the target regions of DNA intended to be reported.

14. Sensitivity

  • i. Limit of Detection (LoD) - SNVs and Indels
    The LoD was assessed by variant type and is defined as the lowest MAF at which ≥95% of replicates are detected. The LoD was evaluated 2 ways, through a dilution series using cell lines to determine the LoD and by confirmation with clinical specimens. Ten unique clinical cases were selected for SNVs, insertions and deletions and diluted with normal DNA derived from FFPE tissues. Each unique clinical case was tested with 10 replicates across 2 kit lots (n=20 per specimen) for a total of 200 observations (Error! Reference source not found.). Data was aggregated across 2 reagent kit lots when possible, otherwise the lot with the higher MAF LoD is displayed.

Additional evaluations of analytical sensitivity performance used dilution series of FFPE clinical specimens. The positive call rates were assessed for a total of 11 SNVs, 3 insertions, and 5 deletions from 5 clinical FFPE specimens with 5 replicates per dilution level. A range of 5.9-12.6% MAF was observed using the lowest average MAF where the positive call rates was ≥ 95%.

Cell lines were used to establish the LoD MAF range for 451 SNVs and 31 indels across the panel. A total of 150 observations were generated (3 samples with 10 replicates at 5 dilution levels). The established ranges were then confirmed with a ≥ 95% call rate with FFPE clinical cases on a per variant level (Error! Reference source not found.) and for the entire panel (Table 1.2). A summary of the LoD by variant type across all replicates is shown in the table below.

VariantMAF RangeCell Line VariantsEvaluated toEstablish MAFRangeNumber of Variants inClinical Cases in theEstablished Range
Hotspot SNVs3.1% to 5.4%82
Non-hotspot SNVs6.3% to 17.8%443176
Indels at homopolymercontext113.7% to 17.5%109

Table 1.2 Analytical Sensitivity (LoD MAF) for Representative SNVs and Indels

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size.

Indels at non-homopolymer context6.1% to 10.9%194
---------------------------------------------------------

1 Greater than or equal to 5 bp repeat

  • ii. LoD ERBB2, ALK, RET, NTRK2, NTRK3 and MSI
    Analytical sensitivity of ERBB2, ALK, RET, NTRK3, and MSI was confirmed by testing 7 clinical FFPE cases diluted with normal FFPE DNA to achieve targeted detection levels. Each unique case was confirmed at ≥ 95% call rate at 1 tumor purity level, with 10 replicates per kit lot, across 2 unique lots for translocations and amplifications. For MSI-H, 3 cases were confirmed at 1 tumor purity level with 10 replicates each. Results summarized in Table 1.3 indicate that the assay is sensitive in detecting specific translocations, amplifications and MSI-H.

Table 1.3 Analytical Sensitivity (LoD Tumor Purity) - Translocations, Amplifications and MSI

VariantLoD Tumor Purity
MSI-H18.1%
ERBB2 amplifications4.4%
ALK translocations15.6%
NTRK2 translocations30%2
NTRK3 translocations11.5%
RET translocations12.8%

1 The enrolled ALK case was evaluated with only 17 total replicates due to insufficient DNA quantity.

2 In silico down sampling suggests LoD 3%.

iii. TMB and Tumor Purity

The minimum tumor purity requirement for input into PGDx elio tissue complete is 20%. The minimum tumor purity required for robust reporting of TMB scores by PGDx elio tissue complete was established using 8 clinical FFPE cases. Samples 1-5 were serially diluted across 3 levels with 5 replicates per level with 3 replicates (18 total), sample 6 was serially diluted across 5 levels at 10 replicates per level (50 total), and samples 7-8 were serially diluted across 5 levels with 5 replicates per level (25 total, 10 of which had a tumor purity ≥15%). The total number of replicates per sample and the %CV of all replicates with ≥15% tumor purity are shown below in Table 1.4. Together these data show PGDx elio tissue complete TMB performance across tumor purities at or above 15%. Figure 1.2 demonstrates the consistency in TMB as a function of tumor purity for each of the specimens assessed across a range of TMB Muts/Mb scores.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for PGDx, which stands for Personal Genome Diagnostics. The letters "PGD" are in blue, while the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in blue, with a small "TM" symbol next to the word "Diagnostics".

SampleReference UndilutedTMB ScoreCV of Replicates ≥ 15%Tumor PurityNumber of Replicates ≥15% Tumor Purity
133.412.5%18
224.810.5%18
350.86.3%18
464.715.9%18
531.58.0%18
6455.43.3%50
710.014.8%10
814.66.0%10

Table 1.4 TMB Precision for Samples ≥ 15% Tumor Purity

Image /page/15/Figure/5 description: The figure is a plot of mean TMB score in Muts/Mb vs tumor purity. There are 8 samples plotted on the figure, with tumor purity ranging from 0% to 80%. The mean TMB score ranges from 1 to 500 Muts/Mb. The samples are labeled as Sample 1 through Sample 8.

Figure 1.2 Linearity of TMB score with tumor purity in PGDx elio tissue complete. The tumor purity is shown on the x-axis and the mean TMB score of the replicates at a specific tumor purity is shown on the y-axis.

iv. DNA Extraction

PGDx elio tissue complete is compatible with genomic DNA extracted from FFPE samples using any appropriate commercially available FFPE extraction method. Samples (3 FFPE specimens and 1 cell line) were extracted in duplicate by 2 operators using 3 different methods, equaling 48 total samples and processed in duplicate for a total of 96 observations to assess concordance. Data for all variant types, including BRAF V600

CONFIDENTIAL

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in a smaller blue font, with a trademark symbol next to the word "Diagnostics".

SNV, ALK and NTRK3 translocations and ERBB2 amplification, were aggregated and Positive Percent Agreement (PPA) and Negative Percent (NPA) are presented in Table 1.5. Method 2 (bead-based) and Method 3 (automated) were compared to the reference Method 1 (column-based). The overall pass rate for FFPE samples was 93.1% (67/72). The %CV for TMB in assessed cases was < 12.5%. All DNA extraction methods yielded concordant results in variant calls with PGDx elio tissue complete.

DNA Extraction MethodConcordance in Variant Calls with Method 1 (n/N) (2-sided 95% CI)
Method 2PPA - 97.8% (673/688) (96.4%, 98.7%)
NPA - 99.9% (71535481/71535520) (99.9%, 100%)
Method 3PPA - 97.5% (624/640) (96.0%, 98.5%)
NPA - 99.9% (51416128/51416155) (99.9%, 100%)

Table 1.5 DNA Extraction Methods Compared to Reference

v. DNA Input

The recommended DNA input for PGDx elio tissue complete is 100 ng. To evaluate assay performance across a range of DNA inputs, 4 unique FFPE samples with known variants were prepared in triplicate at 10, 25, 50, 100, and 200 ng DNA input levels. The variant calls for these samples were compared to the respective reference DNA input of 100 ng for each case to assess concordance. Table 1.6 describes PPA and NPA for each input level where aggregated variants were analyzed, including SNVs, indels, amplifications, translocations, and MSI. For TMB, the mean absolute percent error rate of 10, 25, 50 and 200 ng DNA input compared to 100 ng were 11.8%, 3.3%, 4.4% and 1.8%, respectively. These data indicate the assay is robust around the recommended 100 ng DNA input.

DNA InputVariant Call Concordance (n/N) (2-sided 95% CI)
10 ngPPA - 92.2% (177/192) (87.5%, 95.2%)NPA - 99.9% (26825815/26825826) (99.9%, 100%)
25 ngPPA - 94.8% (182/192) (90.7%, 97.1%)NPA - 99.9% (26825815/26825826) (99.9%, 100%)
50 ngPPA - 96.9% (186/192) (93.4%, 98.6%)NPA - 99.9% (26825822/26825826) (99.9%, 100%)
200 ngPPA - 97.4% (187/192) (94.0%, 98.9%)NPA - 99.9% (26825818/26825826) (99.9%, 100%)

Table 1.6 DNA Input Compared to 100 ng Reference

vi. Sample Carryover and Cross-contamination

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image is a logo for Personal Genome Diagnostics. The logo consists of the letters "PGDx" in blue and yellow. The letters "PG" are in blue, the letter "D" is in yellow, and the letter "x" is in yellow. To the right of the letters is the text "Personal Genome Diagnostics" in blue. The text is stacked on top of each other.

Cross-contamination (contamination from one sample to another within the same batch) and sample carryover (contamination from a previous sequencing run when using the same instrument) were assessed by evaluating false positive and false negative variant calls in 29 FFPE samples. Seven of the 29 cases had known positive variants, the remaining samples were known negative samples. All FFPE samples were assessed across 2 batches to test for contamination within and between runs. In batch 1, a checkerboard pattern within a 96-well plate was created by alternating the samples with representative positive variants and known negative samples. Batch 2 contained known negative samples and was pooled and sequenced directly after completion of batch 1 sequencing. following standard instrument cleaning procedures. No positive variant results were observed in known negative samples tested. Therefore, the PGDx elio tissue complete assay workflow presents minimal risk for contamination.

15. Interference (Endogenous and Exogenous)

  • i. Interfering Substances (Exogenous)
    The impact of exogenous interfering substances on the performance of the PGDx elio tissue complete assay was assessed by processing DNA from FPPE samples tested in the presence of each interfering substance at varying amounts (Table 1.7). The samples were evaluated for concordance of variant calls when compared to samples processed without the interfering substances. Replicates for 5 test cases were analyzed for 8 experimental and 2 baseline conditions. Analysis of all variant types tested (SNVs, indels, translocations, amplifications and MSI) showed high PPA (> 97.2%) and NPA (>99.9%) for all variants. The TMB mean absolute percent error (MAPE) ranged from 0% to 6.0% across conditions. The results show minimal risk to assay performance from interfering exogenous substances.
SubstanceAmount in Excess of Standard Conditions
Proteinase K2X and 3X
Indexed adapters15% and 30%
Melanin0.2 $ \mu $ g/mL and 1.6 $ \mu $ g/mL
Ethanol2.5% and 5%
Table 1.7 Exogenous Interfering Substances Tested
---------------------------------------------------------

ii. Endogenous Interference

The impact of necrosis and FFPE block age on the performance of PGDx elio tissue complete was evaluated by assessing the first pass and overall pass rates of samples processed in the accuracy study (see Accuracy section below). Of 521 samples enrolled for accuracy, 448 were evaluated for necrosis over a range of 0-75%, and 378 were evaluated for age of block over a range of 0-253 months. The data indicated there is no

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "x" symbol. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters and text are aligned horizontally.

correlation between necrosis and pass rate. A logistic regression analysis was performed to establish the probability of sample pass/fail according to FFPE block age as shown in Figure 1.3. There is a correlation between overall pass rate and the age of block; as the age of block increases, the overall pass rate decreases. The probability of samples passing from blocks aged roughly 175 months (or 14.5 years) is approximately 75%. The results show minimal risk to assay performance from interfering endogenous factors.

Image /page/18/Figure/4 description: The image contains two scatter plots that show the probability of success as a function of FFPE Block Age. The left plot shows the probability of first pass success, while the right plot shows the probability of overall pass. In both plots, the x-axis represents the FFPE Block Age in months, and the y-axis represents the probability of success, ranging from 0 to 1. Both plots show a decreasing trend in the probability of success as the FFPE Block Age increases.

Figure 1.3 Logistic Regression Graphs of First Pass and Overall Acceptability vs FFPE Block Age. The orange line represents a regression line, and orange shading represents the 95% confidence interval. The blue dots represent individual samples assessed.

16. Assay Acceptance Rates

Multiple factors can influence overall robustness and performance of complex molecular tests, including pre-analytical factors and overall sample quality. If key in-process or automated data quality metrics are not met. PGDx elio tissue complete supports repeating samples through the workflow. Performance throughout verification and validation of the device was tracked and a summary of the rates for first pass (no repeat) and overall pass (allowing for a single repeat) are presented below.

  • i. Overall Clinical FFPE Sample Acceptance Rate
    Data were aggregated for unique clinical cases from >40 tumor types assessed during verification and validation of PGDx elio tissue complete. Resulting pass rates for clinical samples are presented in Table 1.8. The data indicate that results are obtained on a high percentage of FFPE samples after processing through the PGDx elio tissue complete assay.

Table 1.8 PGDx elio tissue complete Acceptability Rates

First Pass Rate (n/N) (2-sided 95% CI)Overall Pass Rate (n/N) (2-sided 95% CI)
81.8% (2352/2874) (80.4%, 83.2%)92.9% (2671/2874) (91.9%, 93.8%)
  • ii. Pan-Tumor Type/Tissue Comparability

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image is a logo for Personal Genome Diagnostics. The logo consists of the letters "PGDx" in blue and orange. The "x" is orange and larger than the other letters. To the right of the letters are the words "Personal Genome Diagnostics" in a smaller blue font. The letters are stacked on top of each other.

Invalid rates for the different tumor types assessed in the analytical accuracy study are provided in Table 1.9 below.

Tumor TypePassing SamplesTotal SamplesInvalid Rate (%)
Bladder6714.3
Brain10100
Breast607216.7
Colorectal91976.2
Endometrial27270
Gastric253119.4
Glioma440
Head and Neck5616.7
Lung - NOS164685.9
Melanoma34365.6
NOS1880
NSCLC185927.6
Other221224.5
Ovarian8911.1
Pediatric Glioma990
Prostate7812.5
Skin440
Triple Negative Breast11110
Total4795218.1
Table 1.9 Specimen Invalid Rates for >18 FFPE Tumor Types

1 NOS: not otherwise specified; NSCLC: non-small cell lung cancer.

20ther (n ≤ 3 cases per tumor type): cervical, cholangiocarcinoma, gallbladder, pancreatic, rhabdomyosarcoma, trachea, esophageal, fallopian tube, liver, mediastinum, peritoneal, renal, and thyroid.

17. Accuracy - Concordance to Orthogonal Methods

In order to demonstrate the accuracy of PGDx elio tissue complete as a tumor profiling device, a study was performed with 582 samples that had both PGDx elio tissue complete data and orthogonal data. Due to the rarity of specific variants in solid tumor FFPE samples, most samples selected for this study were pre-screened, resulting in enrichment of certain variants relative to real-world clinical prevalence. Data were

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image is a logo for PGDx, which stands for Personal Genome Diagnostics. The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font, also in blue.

April 24, 2020

PGDx elio™ tissue complete

aggregated at the variant level for SNVs, insertions, and deletions, gene level for amplifications and translocations, and case level for MSI and TMB.

The results summarized in Table 1.10 indicate that the assay accurately detects SNVs and indels.

VariantOrthogonalMethod(s)Performance (n/N) (2-sided 95% CI)
SNVs with Evidenceof ClinicalSignificance2 NGStargetedpanelsPPA – 97.2% (35/36) (85.8%, 99.5%)NPA - 99.9% (3994/3996) (99.8%, 99.9%)
Hotspot SNVs2 NGStargetedpanels andPCRPPA - 97.1% (132/136) (92.7%, 98.9%)NPA - 99.9% (35845/35850) (99.9%,99.9%)
Non-hotspot SNVs2 NGStargetedpanelsPPA – 85.1% (516/606) (82.1%, 87.8%)NPA - 99.9% (178513452/178513618)(99.9%, 99.9%)
SNVs with PotentialClinical Significance2 NGStargetedpanelsPPA – 86.4% (591/684) (83.6%, 88.8%)NPA - 99.9% (178513372/178513540)(99.9%, 99.9%)
Hotspot deletions2 NGStargetedpanels andPCRPPA – 100% (20/20) (84.5%, 100%)NPA – 99.9% (2064/2067) (99.6%, 99.9%)
Hotspot insertions2 NGStargetedpanelsPPA – 100% (1/1) (20.7%, 100%)NPA - 100% (2015/2015) (99.8%, 100%)
Non-hotspot indelsNGStargetedpanelPPA - 81.4% (79/97) (72.6%, 87.9%)NPA – 99.9% (67104842/67104857)(99.9%, 99.9%)
Non-hotspotinsertionsNGStargetedpanelPPA – 80.8% (21/26) (62.1%, 91.5%)NPA - 99.9% (67104926/67104928)
Non-hotspotdeletionsNGStargetedpanelPPA - 81.7% (58/71) (71.2%, 89.0%)NPA - 99.9% (67104870/67104883)
Insertions withPotential ClinicalSignificanceNGStargetedpanelPPA - 80.8% (21/26) (62.1%, 91.5%)NPA - 99.9% (67497962/67497964)(99.9%, 99.9%)
Deletions withPotential ClinicalSignificanceNGStargetedpanelPPA – 82.7% (62/75) (72.6%, 89.6%)
NPA – 99.9% (67497902/67497915)
(99.9%, 99.9%)

Table 1.10 Accuracy – SNVs and Indels

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image is the logo for Personal Genome Diagnostics. The logo has the letters PGD in blue and the letter X in orange. To the right of the letters is the text "Personal Genome Diagnostics" in blue. The TM symbol is in the bottom right corner of the text.

  • a) Accuracy for ERBB2 Amplifications
    Concordance for ERBB2 Amplifications was assessed by comparing PGDx elio tissue complete to ERBB2 FISH (Table 1.11 and Table 1.12). The PPA was 75.0% (95% CI: 62.3%, 84.5%) for all cases and 87.0% (95% CI: 74.3%, 93.9%) when excluding borderline FISH cases (defined as HER2/CEP17 ratio between 1.5 and 2.5). The NPA was 96.7% (95% CI: 90.8%, 98.9%) for all cases and 95.9% (95% CI: 88.7%, 98.6%) when excluding borderline FISH cases.

Table 1.11 Summary of Concordance between PGDx elio tissue complete and ERBB2 FISH Including Borderline Cases

ERBB2 FISH
PGDx elio tissue completeERBB2 PositiveERBB2 NegativeTotal
ERBB2 Positive42345
ERBB2 Negative1488102
Total5691147

Table 1.12 Summary of Concordance between PGDx elio tissue complete and ERBB2 FISH Excluding Borderline Cases

ERBB2 FISH
PGDx elio tissuecompleteERBB2 PositiveERBB2 NegativeTotal
ERBB2 Positive40343
ERBB2 Negative67177
Total4674120

b) Accuracy for ALK Translocations

Concordance for ALK translocations was assessed by comparing PGDx elio tissue complete to ALK FISH (Table 1.13). The PPA was 92.9% (95% CI: 68.5%, 98.7%), and NPA was 98.2% (95% CI: 90.7%, 99.7%).

Table 1.13 Summary of Concordance between PGDx elio tissue complete and ALK

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font, also in blue. The logo is simple and modern, and the use of color helps to make it stand out.

FISH

ALK FISH
PGDx elio tissuecompleteALK PositiveALK NegativeTotal
ALK Positive13114
ALK Negative15657
Total145771

ALK translocations were additionally assessed in silico due to limited availability of clinical cases close to the ALK FISH equivocal zone (10%-50% rearrangement positive nuclei). A total of 410 observations were generated for ALK by down-sampling 10 clinical samples from analytical accuracy to 4 tumor purity dilution levels with 10 replicates per level, to mimic samples in the FISH equivocal zone. For example, if the undiluted sample had a FISH score of 50% from analytical accuracy, the sample was diluted with wild type reads by a factor of 0.8 to get to a 40% positive nuclei FISH score. These data demonstrate an 88% positive call rate at 20% positive nuclei by FISH (Table 1.14).

FISH (%PositiveNuclei)FISHCallPGDx elio tissue complete Positive Call Rate (%) (n/N)(95% CI)
50 - 88+100% (10/10) (72%, 100%)
40+98% (98/100) (93%, 99%)
30+95% (95/100) (89%, 98%)
20+88% (88/100) (80%, 93%)
10-80% (80/100) (71%, 87%)

c) Accuracy for RET Translocations

Concordance for RET translocations was assessed by comparing PGDx elio tissue complete to RET FISH (Table 1.15). The PPA was 55.6% (95% CI: 26.7%, 81.1%), and NPA was 100% (95% CI: 82.4%, 100%).

Table 1.15 Summary of Concordance between PGDx elio tissue complete and RET FISH

RET FISH
PGDx elioRET PositiveRET NegativeTotal
RET Positive505
RET Negative411822

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters, the words "Personal Genome Diagnostics" are in a smaller blue font.

Total91827
------------------

1No read data supporting RET translocations was found in the raw data for the discrepant cases. In addition to the clinical samples processed to assess RET translocations, 3 RET translocation-positive cell lines were also tested with PGDx elio tissue complete. All 3 cell lines were positive for a fusion either by a validated assay performed by the cell line provider, or via literature. PGDx elio tissue complete detected all 3 fusions in these cell lines.

  • ii. Accuracy TMB
    The PGDx elio tissue complete assay reports a TMB score comprised of sequence mutations detected across the entire coding region of interest per sample. The ability of PGDx elio tissue complete to accurately identify TMB in multiple solid tissue FFPE tumor types was assessed by comparing to matched tumor-normal whole exome sequencing results. Across 8 tumor types (non-small cell lung carcinoma (NSCLC), melanoma, renal, bladder, endometrial, triple negative breast, head and neck, lung-NOS (not otherwise specified)), 118 cases were enrolled covering a dynamic range of 1.5-118.5 Muts/Mb. Of those, 31 fell below the established LoB (≤ 7.2 Muts/Mb). The Spearman correlation coefficient was used to determine the relationship between the 2 assays. Assessment of all 118 cases resulted in a Spearman correlation coefficient of 0.903. The results in Figure 1.4 show strong concordance between PGDx elio tissue complete TMB scores and tumor-normal whole exome sequencing.

Image /page/23/Figure/7 description: The image is a scatter plot that compares PGDx elio TMB (Muts/Mb) on the y-axis to Exome TMB (Muts/Mb) on the x-axis. The data points are clustered, showing a positive correlation between the two variables. A dashed line represents the regression fit, with the equation y=1.56*x + 6.09, and the Spearman correlation is 0.903, indicating a strong positive relationship.

Figure 1.4 PGDx elio tissue complete TMB score vs. Matched Tumor-Normal Exome Sequencing.

iii. Accuracy - MSI

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image is a logo for Personal Genome Diagnostics, also known as PGDx. The logo features the letters "PGD" in a dark blue color, followed by an orange "X". To the right of the letters, the words "Personal Genome Diagnostics" are stacked on top of each other in a smaller font, also in dark blue. The letters and words are all aligned horizontally.

Table 1.16 contains MSI performance for all 283 samples including PGDx elio tissue complete and PCR failures and indeterminates. This cohort is further divided into performance for colorectal (CRC) and endometrial cases (Table 1.17) and non-CRC and non-endometrial cases (Table 1.18). MSI accuracy was assessed in 18 tumor types: ampulla (1), bladder (7), breast (21), colorectal (66), endometrial (18), esophagus (1), fallopian tube (1), gall bladder (1), gastric (40), lung (39), kidney (3), omentum (1), ovarian (2), prostate (4), sarcoma (3), skin (8), thyroid (2), and cancer of unknown primary (5).

MSI PCRTotal
MSIMSSFailedIndeterminate
PGDx elio tissuecompleteMSI79101181
MSS1142052148
Failed44010054
Total84183106283
Excluding failed/indeterminatespecimens with 95%CIPPA98.8% (79/80) (93.3%, 99.8%)
NPA99.3% (142/143) (96.1%, 99.9%)
PPV98.8% (79/80) (93.3%, 99.8%)
NPV99.3% (142/143) (96.1%, 99.9%)
Accounting forfailed/ indeterminatespecimens with 95%CIPPA94.0% (79/84) (86.8%, 97.4%)
NPA77.6% (142/183) (71.0%, 83.0%)
PPV97.5% (79/81) (91.4%, 99.3%)
NPV95.9% (142/148) (91.4%, 98.1%)

Table 1.16 MSI Performance for All Cases

1This case was MSI-H by PGDx elio, and Promega PCR gave an "Indeterminate" result.

2These 5 cases did not have matching normal DNA to test via Promega PCR.

Table 1.17 MSI Performance for CRC and Endometrial Cases

MSI PCRTotal
MSIMSSFailedIndeterminate
PGDx elio tissue completeMSI5100051
MSS0330033
Failed00000
Total51330084
Excluding failed/indeterminatePPA100% (51/51) (93.0%, 100%)
NPA100% (33/33) (89.6%, 100%)

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in blue. The letters are in a sans-serif font.

specimens with 95% CIPPV100% (51/51) (93.0%, 100%)
NPV100% (33/33) (89.6%, 100%)
Accounting forfailed/indeterminatespecimens with 95%CIPPA100% (51/51) (93.0%, 100%)
NPA100% (33/33) (89.6%, 100%
PPV100% (51/51) (93.0%, 100%)
NPV100% (33/33) (89.6%, 100%)

Table 1.18 MSI Performance for Non-CRC and Non-Endometrial Cases

MSI PCRTotal
MSIMSSFailedIndeterminate
PGDx elio tissueMSI28101130
completeMSS1109052115
Failed44010054
Total33150106199
Excluding failed/indeterminatespecimens with 95%CIPPA96.6% (28/29) (82.8%, 99.4%)
NPA99.1% (109/110) (95.0%, 99.8%)
PPV96.6% (28/29) (82.8%, 99.4%)
NPV99.1% (109/110) (95.0%, 99.8%)
Accounting forfailed/ indeterminatespecimens with 95%CIPPA84.8% (28/33) (69.1%, 93.4%)
NPA72.7% (109/150) (65.0%, 79.2%)
PPV93.3% (28/30) (78.7%, 98.2%)
NPV94.8% (109/115) (89.1%, 97.6%)

1This case was MSI-H by PGDx elio, and Promega PCR gave an "Indeterminate" result.

2These 5 cases did not have matching normal DNA to test via Promega PCR.

iv. Method Comparison Study for Wild Type Calls

A study was conducted to assess accuracy for 75 hotspot loci within 20 genes. A total of 112 specimens were tested, and the accuracy of PGDx elio tissue complete at all 75 positions was compared to 2 orthogonal methods (42 samples using 1 method, and 70 using a second method). Within the 112 specimens, there were 112 mutations across samples and 8,283 wild type calls. Overall variant-level concordance (PPA and NPA) was 96.4% and 99.9% respectively with two-sided 95% confidence intervals of (91.1%, 99.0%) for mutations (PPA), and (99.9%, 99.9%) for wild type locations (NPA).

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for PGDx, which stands for Personal Genome Diagnostics. The letters "PG" are in blue, while "Dx" is in orange. To the right of the letters is the full name of the company, "Personal Genome Diagnostics TM", in a smaller font size and also in blue.

18. Reproducibility

  • i. Interlaboratory Reproducibility
    Interlaboratory reproducibility of the PGDx elio tissue complete assay was assessed across 3 different sites, using DNA extracted from 13 FFPE tissue specimens and 1 cell line. Together these 14 samples represented a range of SNVs and indels, ERBB2 amplifications, ALK, RET, and NTRK3 translocations, MSI, and TMB. Each of the 14 samples was tested in duplicate by 2 different operators on 12 sequencing runs across 3 non-consecutive days at each of the 3 independent laboratory sites using a single kit lot. The first pass rate was 90.3% (455/504) and the overall pass rate of the study was 98.2% (495/504) allowing a maximum of 1 round of repeat testing. Reproducibility was assessed 3 ways; (1) agreement for each positive variant detected across all replicates is reported (Positive call rate), and (2) Average Positive Agreement (APA) and Average Negative Agreement (ANA) and (3) modal analysis was used for per specimen reproducibility.

The positive call rate across all variants was 86.2%, while for SNVs, insertions, and deletions it was 88.8%, 82.8%, and 80.5%, respectively. Table 1.19 shows the positive call rate stratified by variant type and allele fraction.

MutationTypeMAFThresholdPositive Call Rate AmongAll Observed MutationsTotal UniqueVariants
AllMAF≥086.2% (14493/16813)474
AllMAF≥588.0% (14483/16458)464
AllMAF≥891.9% (13921/15146)427
AllMAF≥1093.1% (13404/14400)406
AllMAF≥1596.4% (12387/12846)362
All SNVsMAF≥088.4% (10549/11937)337
All SNVsMAF≥591.0% (10539/11582)327
All SNVsMAF≥895.7% (10070/10519)297
All SNVsMAF≥1097.7% (9618/9845)278
All SNVsMAF≥1597.8% (8773/8966)253
All InsertionsMAF≥082.8% (649/784)22
All InsertionsMAF≥582.8% (649/784)22
All InsertionsMAF≥886.9% (619/712)20
All InsertionsMAF≥1086.9% (619/712)20
All InsertionsMAF≥1595.9% (614/640)18
All DeletionsMAF≥080.5% (3295/4092)115
All DeletionsMAF≥580.5% (3295/4092)115
All DeletionsMAF≥882.6% (3232/3915)110

Table 1.19 Interlaboratory Reproducibility Call Rates

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image is a logo for Personal Genome Diagnostics, or PGDx. The logo has the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters and text are all aligned horizontally.

MutationTypeMAFThresholdPositive Call Rate AmongAll Observed MutationsTotal UniqueVariants
All DeletionsMAF≥1082.4% (3167/3843)108
All DeletionsMAF≥1592.6% (3000/3240)91

Agreement was also assessed by evaluating the APA and ANA, which assess the degree of agreement for variants based on the average result (Table 1.20) APA and ANA across all 3 sites were > 92% for all variant types tested.

Reproducibility was also assessed for independent variables (site, operator, day and within-run). No differences across sources of imprecision were observed.

Alteration TypeMetricOverall (CI)Alteration TypeMetricOverall (CI)
MSIAPA99.1% (98.7%,99.4%)ERBB2amplificationsAPA100% (99.3%,100%)
ANA99.3% (99.0%,99.5%)ANA100% (100%,100%)
SNVsAPA97.8% (97.7%,97.9%)ALKtranslocationsAPA98.6% (97.7%,99.1%)
ANA99.9% (99.9%,100%)ANA99.8% (99.6%,99.9%)
InsertionsAPA95.6% (95.2%,96.0%)NTRK3translocationsAPA92.7% (90.4%,94.5%)
ANA99.9% (99.9%,100%)ANA99.4% (99.2%,99.5%)
DeletionsAPA94.4% (94.2%,94.6%)RETtranslocationsAPA98.7% (97.8%,99.2%)
ANA99.9% (99.9%,100%)ANA99.8% (99.6%,99.9%)
TMB%CV3.5%

Table 1.20 Interlaboratory Reproducibility Performance

The modal positive and negative call rates for sequence mutations (SNVs and indels) in each specimen are summarized in Table 1.21.

Table 1.21 Modal Call Rates for Interlaboratory Reproducibility

SpecimennTotal UniqueMutationsDetectedAcross AllReplicatesModal Positive Call Rate1 (n/N)(two-sided 95% CI)Modal Negative Call Rate2(n/N) (two-sided 95% CI)
11099.6% (251/252) (97.8%,99.9%)97.2% (105/108) (92.2%,99.1%)
230--

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image is a logo for Personal Genome Diagnostics, or PGDx. The logo has the letters "PGD" in blue, with the "x" in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size. The letters are stacked on top of each other.

39100% (315/315) (98.8%, 100%)-
47100% (216/216) (98.3%, 100%)97.2% (35/36) (85.8%, 99.5%)
54399.5% (1462/1470) (98.9%, 99.7%)91.4% (32/35) (77.6%, 97.0%)
62098.9% (639/646) (97.8%, 99.5%)88.2% (30/34) (73.4%, 95.3%)
72697.8% (678/693) (96.5%, 98.7%)95.2% (157/165) (90.7%, 97.5%)
88196.4% (1991/2065) (95.5%, 97.1%)88.7% (683/770) (86.3%, 90.8%)
98897.8% (2710/2772) (97.1%, 98.3%)80.3% (318/396) (76.1%, 83.9%)
103099.0% (998/1008) (98.2%, 99.5%)97.2% (70/72) (90.4%, 99.2%)
119496.1% (2907/3024) (95.4%, 96.8%)83.1% (299/360) (78.8%, 86.6%)
123399.4% (1109/1116) (98.7%, 99.7%)97.2% (70/72) (90.4%, 99.2%)
139100% (216/216) (98.3%, 100%)93.5% (101/108) (87.2%, 96.8%)
142496.8% (732/756) (95.3%, 97.9%)88.0% (95/108) (80.5%, 92.8%)

1Positive call rate was calculated based on variants with majority call detected as positive. 2 Negative call rate was calculated based on variants detected at least once, but with

majority or equal call as negative. For all other locations, the negative call rates are 100%.

3 Specimen 2 was selected for presence of ALK translocation and had no detected SNVs or indels.

Precision of MSI was evaluated across 8 MSS and 6 MSI-H samples with a range of MSI scores evaluated by PGDx elio tissue complete with positive call rates provided in Table 1.22.

Table 1.22 MSI Performance in the Interlaboratory Reproducibility Study
-----------------------------------------------------------------------------------------
Case No.Modal StatusTotal ReplicatesMean MSI Score ValueMSI Score RangeSD%CVPositive Call Rate (95% CI)
1MSS3610.5(4.4, 19.1)3.735.1100% (90.4%, 100%)
2MSS3513.6(5.8, 20.4)3.626.8100% (90.1%, 100%)

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. To the right of the letters is the text "Personal Genome Diagnostics" in a smaller font size, also in blue. The logo is simple and modern, and the use of color helps to make it stand out.

CaseModalTotalMean MSIMSI ScoreSD%CVPositive Call Rate
No.StatusReplicatesScoreValueRange(95% CI)
3MSS3513.8(7.2, 20.2)3.726.8100% (90.1%,100%)
4MSS3610.5(3.9, 19.8)3.230.4100% (90.4%,100%)
5MSI-H35209.7(203.7, 216.5)3.61.7100% (90.1%,100%)
6MSS349.6(4.7, 16.0)3.031.3100% (89.9%,100%)
7MSS33-21.9(-29.9, -13.2)5.0-22.8100% (89.6%,100%)
8MSI-H35223.5(213.1, 236.3)6.02.7100% (90.1%,100%)
9MSI-H36271.6(261.6, 287.2)5.92.2100% (90.4%,100%)
10MSI-H3677.5(62.6, 102.5)6.68.5100% (90.4%,100%)
11MSI-H36219.0(212.6, 224.0)2.91.3100% (90.4%,100%)
12MSS36-56.1(-61.5, -41.5)4.1-7.3100% (90.4%,100%)
13MSS3616.4(10.4, 25.1)3.923.7100% (90.4%,100%)
14MSI-H3649.3(36.7, 61.7)6.312.894.4% (81.9%,98.5%)

Precision of TMB was evaluated across 11 samples (with TMB scores above TMB LoB of 7.2 Muts/Mb) with a range of TMB across site, operator, and day provided in Figure 1.7.

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The letters "PGD" are in blue, and the "x" is in orange. The words "Personal Genome Diagnostics" are in blue and are to the right of the letters "PGDx".

Site A Op1Site A Op2Site B Op1Site B Op2Site C Op1Site C Op2
100ooooooSample
50
25
******
10xxxxxx
Day1Day2Day3Day1Day2Day3Day1Day2Day3Day1Day2Day3Day1Day2Day3Day1Day2Day3ABCDEFGHIJK

Figure 1.5 TMB Performance in the Interlaboratory Reproducibility Study by Site, Operator, and Day

Lot to Lot Precision

Performance of PGDx elio tissue complete was assessed across 3 unique kit lots by determining concordance of variant calls in FFPE tissue samples. The 3 unique kit lots were utilized to process 5 test cases in triplicate for a total of 45 observations. All batches were sequenced on the same instrument. The overall pass rate of the study was 100% (45/45). Table 1.23 lists the APA and ANA used to assess lot to lot performance. APA for all variants is > 86%, and %CV for TMB analyses is < 10%.

Variant TypePerformanceBetween Lot 1 & 2Between Lot 1 & 3Between Lot 2 & 3
Variants withEvidence ofClinicalSignificanceAPA98.7% (93%, 99.8%)96.1% (89.2%,98.7%)97.4% (91.1%,99.3%)
Variants withEvidence ofClinicalSignificanceANA99.9% (99.6%, 100%)99.8% (99.4%,99.9%)99.9% (99.5%,100%)
MSIAPA100% (75.8%, 100%)100% (75.8%, 100%)100% (75.8%, 100%)
MSIANA100% (82.4%, 100%)100% (82.4%, 100%)100% (82.4%, 100%)

Table 1.23 Lot to Lot Precision

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows the logo for Personal Genome Diagnostics (PGDx). The logo consists of the letters "PGD" in blue, followed by an orange "X". To the right of the letters is the text "Personal Genome Diagnostics" in a smaller blue font. The letters and text are aligned horizontally.

SNVs
APA92.1% (90.8%, 93.2%)91.9% (90.6%, 93.0%)91.9% (90.7%, 93.0%)
ANA99.9% (99.9%, 100%)99.9% (99.9%, 100%)99.9% (99.9%, 100%)
Insertions
APA88.9% (80.2%, 94.0%)88.9% (80.2%, 94.0%)87.2% (78%, 92.9%)
ANA99.9% (99.9%, 100%)99.9% (99.9%, 100%)99.9% (99.9%, 100%)
Deletions
APA86.2% (82.6%, 89.1%)89.8% (86.7%, 92.2%)87.3% (83.9%, 90.0%)
ANA99.9% (99.9%, 100%)99.9% (99.9%, 100%)99.9% (99.9%, 100%)
ERBB2amplificationAPA100% (61.0%, 100%)100% (61.0%, 100%)100% (61.0%, 100%)
ANA100% (86.2%, 100%)100% (86.2%, 100%)100% (86.2%, 100%)
ALKtranslocationAPA100% (61.0%, 100%)100% (61.0%, 100%)100% (61.0%, 100%)
ANA100% (96.7%, 100%)100% (96.7%, 100%)100% (96.7%, 100%)
TMB%CV9.5%7.9%7.1%

H. Conclusions

The submitted information in this premarket notification the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs substantially equivalent to the legally marketed predicated device and supports a substantial equivalence decision

§ 866.6080 Next generation sequencing based tumor profiling test.

(a)
Identification. A next generation sequencing (NGS) based tumor profiling test is a qualitative in vitro diagnostic test intended for NGS analysis of tissue specimens from malignant solid neoplasms to detect somatic mutations in a broad panel of targeted genes to aid in the management of previously diagnosed cancer patients by qualified health care professionals.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) A detailed description of all somatic mutations that are intended to be detected by the test and that are adequately supported in accordance with paragraph (b)(1)(v) of this section and reported in the test results in accordance with paragraph (b)(2)(iv) of this section, including:
(A) A listing of mutations that are cancer mutations with evidence of clinical significance.
(B) As appropriate, a listing of mutations that are cancer mutations with potential clinical significance.
(ii) The indications for use must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (
e.g., formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types (
e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable.
(iii) A detailed device description including the following:
(A) A description of the test in terms of genomic coverage, as follows:
(
1 ) Tabulated summary of all mutations reported, grouped according to gene and target region within each gene, along with the specific cDNA and amino acid positions for each mutation.(
2 ) A description of any within-gene targeted regions that cannot be reported and the data behind such conclusion.(B) Specifications for specimen requirements including any specimen collection devices and preservatives, specimen volume, minimum tumor content, specimen handling, DNA extraction, and criteria for DNA quality and quantity metrics that are prerequisite to performing the assay.
(C) A detailed description of all test components, reagents, instrumentation, and software required. Detailed documentation of the device software including but not limited to, software applications and hardware-based devices that incorporate software.
(D) A detailed description of the methodology and protocols for each step of the test, including description of the quality metrics, thresholds, and filters at each step of the test that are implemented for final result reporting and a description of the metrics for run-failures, specimen-failures, invalids, as applicable.
(E) A list of links provided by the device to the user or accessed by the device for internal or external information (
e.g., decision rules or databases) supporting clinical significance of test results for the panel or its elements in accordance with paragraphs (b)(1)(v) and (b)(2)(vi) of this section.(F) A description of internal and external controls that are recommended or provided and control procedures. The description must identify those control elements that are incorporated into the testing procedure.
(iv) Information demonstrating analytical validity of the device according to analytical performance characteristics, evaluated either specifically for each gene/mutation or, when clinically and practically justified, using a representative approach based on other mutations of the same type, including:
(A) Data that adequately supports the intended specimen type (
e.g., formalin-fixed, paraffin-embedded tumor tissue), specimen handling protocol, and nucleic acid purification for specific tumor types or for a pan-tumor claim.(B) A summary of the empirical evidence obtained to demonstrate how the analytical quality metrics and thresholds were optimized.
(C) Device precision data using clinical samples to adequately evaluate intra-run, inter-run, and total variability. The samples must cover all mutation types tested (both positive and negative samples) and include samples near the limit of detection of the device. Precision must be assessed by agreement within replicates on the assay final result for each representative mutation, as applicable, and also supported by sequencing quality metrics for targeted regions across the panel.
(D) Description of the protocols and/or data adequately demonstrating the interchangeability of reagent lots and multiplexing barcodes.
(E) A description of the nucleic acid assay input concentration range and the evidence to adequately support the range.
(F) A description of the data adequately supporting the limit of detection of the device.
(G) A description of the data to adequately support device accuracy using clinical specimens representing the intended specimen type and range of tumor types, as applicable.
(
1 ) Clinical specimens tested to support device accuracy must adequately represent the list of cancer mutations with evidence of clinical significance to be detected by the device.(
2 ) For mutations that are designated as cancer mutations with evidence of clinical significance and that are based on evidence established in the intended specimen type (e.g., tumor tissues) but for a different analyte type (e.g., protein, RNA) and/or a measurement (e.g., incorporating a score or copy number) and/or with an alternative technology (e.g., IHC, RT-qPCR, FISH), evidence of accuracy must include clinically adequate concordance between results for the mutation and the medically established biomarker test (e.g., evidence generated from an appropriately sized method comparison study using clinical specimens from the target population).(
3 ) For qualitative DNA mutations not described in paragraph (b)(1)(iv)(G)(2 ) of this section, accuracy studies must include both mutation-positive and wild-type results.(H) Adequate device stability information.
(v) Information that adequately supports the clinical significance of the panel must include:
(A) Criteria established on what types and levels of evidence will clinically validate a mutation as a cancer mutation with evidence of clinical significance versus a cancer mutation with potential clinical significance.
(B) For representative mutations of those designated as cancer mutations with evidence of clinical significance, a description of the clinical evidence associated with such mutations, such as clinical evidence presented in professional guidelines, as appropriate, with method comparison performance data as described in paragraph (b)(1)(iv)(G) of this section.
(C) For all other mutations designated as cancer mutations with potential clinical significance, a description of the rationale for reporting.
(2) The 21 CFR 809.10 compliant labeling and any product information and test report generated, must include the following, as applicable:
(i) The intended use statement must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (
e.g., formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types (
e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable.
(ii) A description of the device and summary of the results of the performance studies performed in accordance with paragraphs (b)(1)(iii), (b)(1)(iv), and (b)(1)(v) of this section.
(iii) A description of applicable test limitations, including, for device specific mutations validated with method comparison data to a medically established test in the same intended specimen type, appropriate description of the level of evidence and/or the differences between next generation sequencing results and results from the medically established test (
e.g., as described in professional guidelines).(iv) A listing of all somatic mutations that are intended to be detected by the device and that are reported in the test results under the following two categories or equivalent designations, as appropriate: “cancer mutations panel with evidence of clinical significance” or “cancer mutations panel with potential clinical significance.”
(v) For mutations reported under the category of “cancer mutations panel with potential clinical significance,” a limiting statement that states “For the mutations listed in [cancer mutations panel with potential clinical significance or equivalent designation], the clinical significance has not been demonstrated [with adequate clinical evidence (
e.g., by professional guidelines) in accordance with paragraph (b)(1)(v) of this section] or with this test.”(vi) For mutations under the category of “cancer mutations panel with evidence of clinical significance,” or equivalent designation, link(s) for physicians to access internal or external information concerning decision rules or conclusions about the level of evidence for clinical significance that is associated with the marker in accordance with paragraph (b)(1)(v) of this section.