(51 days)
The MSK-IMPACT assay is a qualitative in vitro diagnostic test that uses targeted next generation sequencing of formalin-fixed paraffin-embedded tumor tissue matched with normal specimens from patients with solid malignant neoplasms to detect tumor gene alterations in a broad multi gene panel. The test is intended to provide information on somatic mutations (point mutations and small insertions and deletions) and microsatellite instability for use by qualified health care professionals in accordance with professional guidelines, and is not conclusive or prescriptive for labeled use of any specific therapeutic product. MSK-IMPACT is a single-site assay performed at Memorial Sloan Kettering Cancer Center.
A description of required equipment, software, reagents, vendors, and storage conditions were provided, and are described in the product labeling (MSK-IMPACT manual). MSK assumes responsibility for the device.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text for the MSK-IMPACT assay.
Device Name: MSK-IMPACT (Integrated Mutation Profiling of Actionable Cancer Targets)
Type of Test: Next generation sequencing tumor profiling test
Purpose: Qualitative in vitro diagnostic test for detecting somatic mutations (point mutations, small insertions and deletions) and microsatellite instability (MSI) in formalin-fixed paraffin-embedded (FFPE) tumor tissue matched with normal specimens from patients with solid malignant neoplasms.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally embedded within the "Performance" section and the "Reporting" section's "Table 3. Sample Level Quality Control Metrics." The reported performance is found throughout the "Performance" section.
| Acceptance Criteria (from text) | Reported Device Performance (from text) |
|---|---|
| Specimen Requirements: | |
| Minimum Tumor Proportion: >10% of tumor cells; >20% viable tumor preferred; >25% for MSI testing. | The minimum tumor proportion required for the MSI assay was established as 25% (based on CRC specimens, assay and score reproducible to 8% tumor proportion qualitatively, but decreased trend quantitatively) (Table 1). The DNA extraction method was validated with historic data from >10,000 specimens, demonstrating invalid rates of 7.2% to 18.4%, supporting performance across FFPE tumor types (Table 5). |
| Quality Control Metrics (Table 3): | |
| Average target coverage: > 200X | For normal samples, mean coverage across all targeted exons was 571X (SD = 373X). Analysis of normal samples showed that with mean sample coverage of 571X, 98% of exons are sequenced with coverage greater than 306X (or normalized coverage >0.54), leading to a conservative threshold of 200X mean sample coverage. In silico downsampling to 203X coverage detected 94% of mutations with 10% VAF (Performance L.1.b and Table 3). |
| Coverage Uniformity: ≥ 98% target exons above 100X coverage | 99.5% of exons were sequenced to a depth of 100X or greater, and 98.6% to 250X or greater. It’s expected that 98% of exons will be sequenced to >100X coverage when mean sample coverage is 185X. (Performance L.1.b) |
| Base Quality: > 80% of bases with QS above > Q30 | Not explicitly detailed in the performance section but stated as a QC metric in Table 3. Implicitly met if overall performance is approved. |
| % Cluster passing filter (Cluster PF): > 80% | Not explicitly detailed in the performance section but stated as a QC metric in Table 3. Implicitly met if overall performance is approved. |
| % Reads passing filter (Reads PF): > 80% | Not explicitly detailed in the performance section but stated as a QC metric in Table 3. Implicitly met if overall performance is approved. |
| Hotspot Mutation calling threshold: DP ≥ 20, AD ≥ 8, VF ≥ 2% | Filtering scheme designed to reject false positives while maintaining detection capability. Example: pre-filter SNVs (hotspot) had 1 false positive, post-filter 0 (Table 4). LoD confirmation: 5 replicates for 6 SNVs at 5% MAF showed 100% positive call rates, except one replicate failing on PTEN exon 6 due to low read depth below 5% (Performance L.2.b.ii and Table 11). |
| Non-hotspot Mutation threshold: DP > 20, AD ≥ 10, VF ≥ 5% | Filtering scheme designed to reject false positives while maintaining detection capability. Example: pre-filter SNVs (non-hotspot) had 342 false positives, post-filter 0 (Table 4). LoD study showed most mutations detected at low VAFs (e.g., 2-9% in Tables 10A-J). Confirmed LoD study (Part 2) for various mutations showed 100% positive call rates for variant types except one discordant case (PTEN exon 8 deletion) at 3.6-7.9% VF (Table 11). |
| Indels: Fewer than 20% of samples in an established 'standard normal' database (This seems to be a filtering criteria for indels, not a reporting metric.) | Indels had 40,793 pre-filter false positives, reduced to 8 post-filter (Rejection Rate 0.999) (Table 4). LoD confirmation: 5 replicates for 3 deletions and 4 insertions at 5% MAF showed 100% positive call rates, except one deletion (PTEN exon 6), which also failed read depth (Performance L.2.b.ii and Table 11). |
| Positive Run Control: The difference between the observed and expected frequencies for the known mutations should be within 5%. | Mixed positive control sample with expected VFs: Results reviewed to confirm known mutations called and observed frequencies match expected values within 5% (Controls, b). |
| Negative Run Control: The correlation between expected and observed mutation frequencies should be 0.9 or higher. | Pooled negative control: Observed mutation frequencies compared against expected for 862 common SNPs; correlation expected to be 0.9 or higher (Controls, c). Figure 2 shows correlation of 0.975 (with slope 0.971 and intercept -0.004) for observed vs. expected variant frequency, establishing consistent correlation >0.9. |
| Sample-Mix up QC: Flagged if pairs of samples from the same patient with > 5% discordance and from different patients with < 5% discordance. | Pipeline computes 'percent discordance'; expected discordance between tumors and matched normal should be low (<5%), between different patients high (~25%). Samples flagged if >5% for same patient ("unexpected mismatches") or <5% for different patients ("unexpected matches") (Device Description, 4.e.i). |
| Major Contamination QC: % heterozygous sites at fingerprint SNPs < 55%; Average MAF at homozygous fingerprint SNPs < 2%. | Samples flagged if average minor allele frequency at homozygous SNP sites exceeds 2% (Device Description, 4.e.ii). |
| Criteria for calling test failure: If a sample presents with mean coverage across all exons < 50x and no mutations are detected due to the low overall coverage, the test is deemed "failed" for the sample. | Not explicitly detailed in the performance section but stated in Table 3. Implicitly met if overall performance is approved. |
| Analytical Performance (General): | |
| Precision (Within-run, Between-run, Total Variability): Using clinical samples, covering all mutation types (positive/negative), including samples near LoD. Assessed by agreement within replicates and sequencing quality metrics. | Panel-Wide Reproducibility: 69 mutations in clinical specimens and 13 in commercial cell line (total 82). All mutations showed 100% concordance except 4 in clinical specimens and 3 in commercial sample. Discordant cases were in repetitive regions, or had low frequencies near 2% (Performance L.2.a.ii and Table 7). Positive call rates varied per mutation and specimen (Table 7 and 8). Per Specimen Precision: (N=5 replicates). Overall positive call rates ranged from 80% to 100% across various specimens (Table 8). Intra-assay repeatability: All results concordant except for ARID1B exon 2 insertion and BRAF V600M point mutation (commercial control). Reference Material (NA20810): 23 replicates. Zygosity results were 100% concordant. Difference between expected and mean observed mutation frequencies was very small (absolute difference = 0.09%±0.45%), providing supplemental evidence of reproducibility (Performance L.2.a.iv). MSI Precision: 12 specimens (6 MSI-H, 6 MSS) tested with 3 inter- and 3 intra-run replicates. All samples had 100% agreement between calls (Performance L.2.a.v and Table 9). |
| Analytical Sensitivity (LoD): Defined as mutant allele fraction at which 95% of replicates are reliably detected. Confirmed with multiple replicates. | Part 1 (Dilution Series): Serial dilutions of patient samples were used to identify lowest reliable mutant fraction. Most mutations were called at lowest dilution (e.g., BRAF V600E at 2% VF, KRAS G12D at 6% VF, EGFR ins at 3% VF), except PIK3CA (PIK3CA Exon 2 (R88Q) was WT at 1:16 dilution) (Tables 10A-J). Part 2 (Confirmation): 5 replicates tested for 3 deletions, 4 insertions, and 6 SNVs at 5% minor allele frequency. All variants had 100% positive call rates except one replicate for a PTEN exon 6 deletion (mutation read depth below estimated LoD of 5%) (Performance L.2.b.ii and Table 11). LoD is stated as 2% for hotspot and 5% for non-hotspot mutations (Assay Cut-off). |
| Analytical Specificity: Maintained by paired tumor/matched normal sequencing. | Established during assay optimization; paired tumor/matched normal sequencing minimizes interference (Performance L.2.g). |
| Accuracy (Method Comparison): Using clinical specimens representing intended specimen type and range of tumor types. Specific criteria for SNV/MNVs, insertions, deletions, and MSI. | Overall Accuracy: 432 out of 433 cases (99.8% with 95% CI (98.7%, 100.0%)) successfully detected known mutations compared to orthogonal methods. One discordant case (EGFR exon 20 duplication) was identified due to filtering algorithm, which was subsequently modified. (Performance L.2.i.i) PPA by Mutation Type/Gene: SNV/MNVs showed 100% PPA for all listed genes (Table 15A). Insertions showed PPA from 93.8% (EGFR) to 100% (Table 15B). Deletions showed 100% PPA for all listed genes (Table 15C). Wildtype Calls (Supplemental Study): 95 specimens with 109 mutations and 3026 wild-type calls across 33 hotspots in 10 genes. Variant-level concordance: PPA was 100% (96.7%, 100.0% CI), NPA was 100% (99.9%, 100.0% CI) (Performance L.2.i.ii). MSI Accuracy (MSIsensor): CRC/EC (Training): Cut-off of 10 established based on concordance with MSI-PCR or MMR IHC using 138 CRC and 40 EC specimens. CRC (Validation): 135 CRC patients, 66 with both MSK-IMPACT MSI and IHC results. PPV = 92.3% (12/13, 95% CI 64.0%-99.8%), NPV = 98.1% (52/53, 95% CI 90.0%-100.0%) (Table 16). Non-CRC/EC: 119 non-CRC/EC samples assessed by MSIsensor and MSI-PCR. Excluding missing data: PPV = 93.9% (46/49, 83.1%-98.7% CI), NPV = 96.7% (58/60, 88.5%-99.6% CI). Including missing data: PPV = 78.0% (46/59, 65.3%-87.7% CI), NPV = 96.7% (58/60, 88.5%-99.6% CI). (Table 17). |
2. Sample Size Used for the Test Set and Data Provenance
-
Test Set Sample Sizes:
- Precision Studies:
- 10 samples (9 FFPE specimens, 1 commercial cell line) used for panel-wide reproducibility (Table 6).
- Well-characterized reference standard (HapMap cell line NA20810) in 23 replicates for sequencing error rates and reproducibility.
- 12 specimens (6 MSI-H, 6 MSS) for MSI precision.
- Analytical Sensitivity (LoD):
- Part 1 (Dilution Series): Patient samples (number not specified, but for 5 validation exons, implying at least 5 patients) with 5-8 serial dilutions.
- Part 2 (Confirmation): Unspecified number of samples, providing variants for 3 deletions, 4 insertions, and 6 SNVs, each tested with 5 replicates.
- MSI LoD: CRC specimens (number not specified, but 5 replicates run).
- DNA Input Assessment: Unspecified number of samples (historical data from >10,000 samples mentioned in pre-analytical performance context). Table 13 presents data by DNA input amounts but not sample count for each bin.
- Accuracy (Method Comparison):
- 267 unique mutations in 433 FFPE tumor specimens for the main comparison (Table 14).
- 95 specimens for the supplemental wildtype calls study.
- 138 colorectal cancer (CRC) and 40 endometrial carcinoma (EC) specimens (training set) for MSI cutoff establishment.
- 135 CRC patients (66 with both MSK-IMPACT and IHC) for MSI cutoff validation.
- 119 unique non-CRC and non-EC tumor-normal pair samples for MSI comparison in other cancer types.
- Precision Studies:
-
Data Provenance:
- General: The device is performed at Memorial Sloan Kettering Cancer Center (MSK), indicating the data likely originates from their patient population.
- Retrospective/Prospective:
- The pre-analytical performance (specimen invalid rates) used historical data from >10,000 specimens, implying a retrospective chart review.
- The MSI validation study (CRC patients) was a retrospective-prospective chart review.
- The clinical performance section mentions a large-scale, prospective clinical sequencing initiative using MSK-IMPACT involving >10,000 patients, whose data are publicly accessible. This cohort likely informed the broader context and understanding of the device but was not explicitly stated as the test set for the analytical validation.
- The analytical performance studies (precision, LoD, accuracy) used clinical samples/specimens, which could be retrospective or prospectively collected for the purpose of the study. The text doesn't explicitly state for each study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
The text does not specify the number of experts used to establish the ground truth for the test set, nor their specific qualifications (e.g., "radiologist with 10 years of experience").
However, it does indicate:
- For the accuracy studies, results were compared to "original results obtained with the validated orthogonal methods." This implies that the ground truth was established by these validated orthogonal methods, which are presumably performed and interpreted by qualified personnel using established clinical diagnostics.
- For MSI, the MSIsensor results were compared to "a validated MSI-PCR or MMR IHC test," a "commercially available PCR assay," or a "validated IHC panel (MLH1, MSH2, MSH6 and PMS2)." Again, this suggests ground truth from established, clinical laboratory methods.
- The "Clinical Evidence Curation" section mentions that "OncoKB undergoes periodic updates through the review of new information by a panel of experts," which informs the clinical interpretation of detected mutations. This expert panel contributes to the broader clinical context of the mutations, but not directly the ground truth for the analytical test set itself.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
The text does not describe a formal adjudication method (like 2+1 or 3+1 consensus with experts) for establishing the ground truth of the test set cases. Instead, the ground truth was derived from "validated orthogonal methods."
For example, in the accuracy study, the MSK-IMPACT results were "compared to the original results obtained with the validated orthogonal methods." This indicates that the results from the comparison methods served as the reference standard, rather than requiring an additional expert adjudication process on top of those existing validated methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not described. This document pertains to the analytical validation of a genetic sequencing assay, which inherently does not involve human readers interpreting images in a multi-reader, multi-case setup. Therefore, a comparative effectiveness study measuring human reader improvement with AI assistance (which is typical for imaging AI) is not applicable here.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the analytical performance studies (precision, analytical sensitivity, and analytical accuracy) described are all measures of the standalone performance of the MSK-IMPACT assay, which relies on its sequencing and bioinformatics pipeline without direct human-in-the-loop diagnostic interpretation to produce the raw mutation calls.
- The "Mutation calling SNVs and Indels" section and "Summary of mutation filtering scheme" (Figure 1) describe the automated pipeline for identifying mutations.
- The "Performance" section details how characteristics like precision, LoD, and accuracy were determined for the assay itself by comparing its outputs to known or established results from other validated methods. These do not involve a human interpreting the device's output to make a diagnosis within the performance evaluation but rather assess the accuracy of the device's genomic calls directly.
7. The Type of Ground Truth Used
The primary type of ground truth used was:
- Orthogonal Methods / Comparator Assays: For the accuracy studies, the MSK-IMPACT results were compared against "original results obtained with the validated orthogonal methods." This included comparison to:
- Validated orthogonal methods for SNVs and indels.
- Established MSI-PCR or MMR IHC tests for Microsatellite Instability status.
- Known Reference Material: For precision, a "well characterized reference standard (HapMap cell line NA20810)" was used, with reference genotypes obtained from the 1000 Genomes database.
- Expected Values/Dilution Series: For Limit of Detection studies, serial dilutions of patient samples with "known mutations" and "expected frequencies" were used.
Therefore, the ground truth is a combination of established methods, known reference materials, and empirically derived expected values.
8. The Sample Size for the Training Set
The document explicitly mentions training data primarily in the context of the MSI cutoff:
- MSI Cutoff Training: A "training specimen dataset consisting of 138 colorectal cancer (CRC) and 40 endometrial carcinoma (EC) specimens with matched normal and having MSI status results from a validated MSI-PCR or MMR IHC test."
For the mutation calling pipeline (SNVs and indels), the text refers to:
- Optimization of thresholds: "The threshold values for the filtering criteria were established based on paired-sample mutation analysis on replicates of normal FFPE samples, and optimized to reject all false positive SNVs and almost all false positive indel calls from the reference dataset." The size of this "reference dataset" or "replicates of normal FFPE samples" used for training/optimization of filtering thresholds is not explicitly stated as a defined "training set sample size" for the SNV/indel calling. It implies an internal dataset used during development.
9. How the Ground Truth for the Training Set Was Established
For the MSI cutoff training set:
- The ground truth was established by "validated MSI-PCR or MMR IHC test" results. These are existing, established clinical diagnostic methods for determining MSI status.
For the SNV/indel pipeline optimization/threshold establishment:
- The ground truth for optimizing filtering thresholds was based on "paired-sample mutation analysis on replicates of normal FFPE samples" and "reference dataset." This suggests that the "true" status of these calls (i.e., whether they were true positives, false positives, etc.) would have been known or definitively determined through external means (e.g., highly confident calls from a different (perhaps more laborious or deeply sequenced) method, or a known characteristic of the "normal FFPE samples"). However, the specific method for establishing this ground truth for the filtering optimization is not explicitly detailed beyond being from a "reference dataset."
{0}------------------------------------------------
EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR MSK-IMPACT (Integrated Mutation Profiling of Actionable Cancer Targets)
DECISION SUMMARY
A. DEN Number:
B. Purpose for Submission:
De novo request for evaluation of automatic class III designation for the MSK-IMPACT
C. Measurand:
Somatic single nucleotide variants, insertions, deletions, and microsatellite instability in genes in human genomic DNA obtained from formalin-fixed, paraffin-embedded tumor tissue.
Refer to Appendix 1a for complete list of hotspot mutations and Appendix 1b for complete list of genes included in this assay.
D. Type of Test:
Next generation sequencing tumor profiling test
E. Applicant:
Memorial Sloan Kettering (MSK)
F. Proprietary and Established Names:
MSK-IMPACT (Integrated Mutation Profiling of Actionable Cancer Targets)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.6080
2. Classification:
Class II
3. Product code:
PZM
{1}------------------------------------------------
4. Panel:
Pathology
H. Indications for Use:
1. Indications for Use:
The MSK-IMPACT assay is a qualitative in vitro diagnostic test that uses targeted next generation sequencing of formalin-fixed paraffin-embedded tumor tissue matched with normal specimens from patients with solid malignant neoplasms to detect tumor gene alterations in a broad multi gene panel. The test is intended to provide information on somatic mutations (point mutations and small insertions and deletions) and microsatellite instability for use by qualified health care professionals in accordance with professional guidelines, and is not conclusive or prescriptive for labeled use of any specific therapeutic product. MSK-IMPACT is a single-site assay performed at Memorial Sloan Kettering Cancer Center.
2. Special conditions for use statement(s):
For prescription use.
For in vitro diagnostic use.
3. Special instrument requirements:
Illumina HiSeq™ 2500 Sequencer (qualified by MSK)
I. Device Description:
A description of required equipment, software, reagents, vendors, and storage conditions were provided, and are described in the product labeling (MSK-IMPACT manual). MSK assumes responsibility for the device.
1. Sample Preparation:
The tumor volume and minimum tumor content needed to obtain sufficient DNA for testing to achieve the necessary quality performance are shown in the Table 1 below:
| TissueType | Volume | Minimum TumorProportion | Macrodissectionrequirements(Based on tumorproportion) | Limitations | Storage |
|---|---|---|---|---|---|
| FFPEsections | 5-20unstainedsections,10micronsthick | More than 10% of tumorcells;sections containing>20% viable tumor arepreferred.For MSI testing, >25%tumor cells. | Yes,macrodissectionto obtain non-neoplastic tissuefor analysis | Archival paraffin-embedded materialsubjected to aciddecalcification isunsuitable for analysisbecause aciddecalcification severelydamage nucleic acids. | Roomtemp |
Table 1. Specimen Handling and Processing for Validated Specimen Types
{2}------------------------------------------------
Genomic DNA is extracted from tissue specimens per protocol. DNA is quantified and concentrated if necessary. The amount of DNA required to perform the test is 100-250ng. DNA is run in singlicate. DNA shearing is conducted per protocol and a quality control check is performed. Average fragment size should be ~200bp. Sheared DNA is stored at -20°C if not proceeding directly to Library Preparation. The DNA can be stored at 37°C for 10-20 minutes, stored at 2-8°C for 24 hours, or at -20°C for longer periods.
2. Library Preparation:
Sequence libraries are prepared using KAPA Biosystems Library Preparation Reagents by first producing blunt-ended, 5'-phosphorylated fragments. To the 3' ends of the dsDNA library fragments, dAMP is added (A-tailing). Next, dsDNA adapters with 3'dTMP is ligated to the A-tailed library fragments. Library fragments with appropriate adapter sequences are amplified via ligation-mediated pre-capture PCR. A quality control check on the amplified DNA libraries is performed: Samples should be a smear; average fragment size with the peak at ~200bp; and concentration between 5-300ng/uL to ensure adequate hybridization for capture.
3. Hybrid Capture NGS:
Library capture is conducted using NimbleGen Capture reagents. Pooled sequencing libraries are hybridized to the vendor oligo pool. Capture beads are used to pull down the complex of capture oligos and genomic DNA fragments. Unbound fragments are washed away. The enriched fragment pool is amplified by ligation mediated-PCR. The success of the enrichment is measured as a quality control step: Samples should be a smear, average fragment size with the peak at ~300bp; the concentration of the amplified DNA library should be 5-45ng/uL; the LM-PCR yield should be ≥ 250ng. Reactions can be stored at 4°C until ready for purification, up to 72 hours.
4. Sequencing and Data Analysis:
Sequencing is conducted with the Illumina HiSeq2500 Sequencing Instruments and reagents and PhiX Control v3. The sequencing process uses multiple quality checks.
- a) Data Management System (DMS): Automated sample tracking and archival of runassociated metadata (barcode, run name, samples accession number, patient medical record number, source (class), specimen type, and panel version) is conducted with the following key functions: Tracking sample status through various stages of data analysis; tracking iterations of analysis applied to a given sample; recording versions of databases and algorithms used in analysis; archival of selected pipeline output files (FASTO, BAM, VCF) and sequencing run statistics (e.g., cluster density, %clusters passing filter, unassigned read indices).
- b) Demultiplexing and FASTQ generation: The analysis pipeline uses software provided by Illumina. Two FASTQ files are generated per samples corresponding to full length forward and reverse reads. Demultiplexing quality control includes quality metrics for per-base sequence quality, sequence content. GC content and sequence length distribution, relative percentages of unmatched indices.
{3}------------------------------------------------
- c) Indexing OC check: The potential for index contamination is managed by demultiplexing all sequencing reads for all possible barcodes. If the number of reads > 15,000 for any unused barcodes, then those reads are analyzed with the pipeline and the fingerprint SNPs are used to identify which of the barcodes used in the pool could be causing the appearance of extra reads.
- d) Read alignment and BAM generation: Spurious adapter sequences are trimmed prior to read alignment. Reads are aligned in paired-end mode to the hg19 b37 version of the human genome. Aligned reads are written to a Sequence Alignment Map (SAM) file, which is then converted into Binary Alignment Map (BAM) format. PCR duplicates are removed. Each base within a read is assigned a base quality score by the sequencing software, which reflects the probability an error was made with the base call. To account for systemic biases that may not accurately reflect the actual error probabilities observed empirically, the analysis pipeline uses another tool to adjust the reported quality scores based on the selected covariates. Reassigned quality scores are subject to a threshold of 20, corresponding to a 1/100 chance of error.
- e) Sample QC checks: The baits used for hybridization capture include custom intergenic and intronic probes targeting >1000 regions throughout the genome containing common single nucleotide polymorphisms (SNPs). The unique combination of SNPs specific to a given sample serves as a 'fingerprint' for the identity of the corresponding patient, and serves to identify potential sample mix-ups and contamination between samples and barcodes. OC checks involving the use of these 'fingerprint' SNPs are detailed below:
- i. Sample mix-up check: The analysis pipeline computes the 'percent discordance' between a reference and query sample, defined as the percent of homozygous sites in the reference sample that are homozygous for the alternate allele in the query sample. The expected discordance between tumors and their respective matched normal should be low (<5%). Conversely, the expected discordance between samples from different patients should be high (~ 25%). Pairs of samples from the same patient with > 5% discordance ("unexpected mismatches") and from different patients with <5% discordance ("unexpected matches") are flagged.
- ii. Sample contamination checks: Alternate alleles (percent heterozygous) at homozygous SNP sites (fingerprint SNPs) are assessed. A sample is flagged for review if the average minor allele frequency at these SNPs exceeds 2%.
- iii. Check for presence of tumor in normal: Normal samples are expected to be free of known SNVs and insertions and deletions (indels) that are commonly (somatically) recurrent in tumor samples. As a first pass check, the pipeline genotypes normal samples at several known 'hotspot' locations derived from somatic mutation catalogs. If a known tumor-specific mutation (i.e. BRAF V600E) is detected with mutation frequency > 1% in a normal sample, the normal sample is flagged for review and possible exclusion from analysis. Tumor
{4}------------------------------------------------
samples with matched normal controls excluded due to possible tumor contamination will be considered as unmatched tumor samples for subsequent analyses.
- f) Mutation calling SNVs and Indels: The analysis pipeline identifies two classes of mutations: (1) single nucleotide variants (SNVs) and (2) indels. Paired sample mutation calling is performed on tumor samples and their respective matched normal controls. In instances where a matched normal sample is unavailable, or where the matched normal sample was sequenced with low coverage (< 50X), tumor samples will be considered as unmatched samples, and will be compared against a standard, in-batch pooled FFPE normal control for mutation calling. Filtering is performed to remove low quality sequence data, sources of sequencing artifacts, and germline results.
- Analysis of pooled FFPE positive and negative controls: data from controls is i. used to confirm lack of contamination as well as analytical sensitivity.
- ii. Filters on sample coverage: A sequence coverage > 100X is required to achieve 95% power to detect mutations with underlying variant frequency of 10% or greater. To ensure that at least 98% of targeted exons meet this coverage, a per sample coverage requirement has been conservatively set at ≥ 200X. A lower coverage threshold for the matched normal is set at 50X.
- iii. Filtering for high confidence mutations: Raw SNV and indel calls are subjected to a series of filtering steps to ensure only high-confidence calls are admitted to the final step of manual review. These parameters include (1) evidence of it being a somatic mutation (i.e., ratio between mutation frequencies in the tumor and normal samples to be > 5.0); (2) whether the mutation is a known hotspot mutation (refer to Appendix 1a for details); (3) reference on in house 'standard normal' based on common artifacts; (4) technical characteristics that use coverage depth (DP), number of mutant reads (AD), mutation frequency (VF).
The filtering scheme and threshold are shown in Figure 1 below. The threshold values for the filtering criteria were established based on paired-sample mutation analysis on replicates of normal FFPE samples, and optimized to reject all false positive SNVs and almost all false positive indel calls from the reference dataset.
{5}------------------------------------------------
Image /page/5/Figure/0 description: This image shows a flowchart of the filtering process for SNVs and Indels. The raw pipeline output for SNVs is MuTect, and for Indels, it is SomaticIndelDetector. The standard filter for somatic variants is VFtumor/VFnormal ≥ 5X, AD ≥ 5, and VF ≥ 1%. After annotation with AnnoVar, a two-tiered filtering scheme is applied based on whether the variant is in a hotspot, with different DP, AD, and VF thresholds for each case, and the process ends with manual review.
Figure 1. Summary of mutation filtering scheme
- g) Mutation annotation: Predicted functional effect and clinical interpretation for each mutation is curated by automated software using information from several databases.
- h) Microsatellite Instability (MSI) status calling: The somatic MSI status is inferred by interrogating all available genomic microsatellites covered by MSK-IMPACT within tumor samples against the matched normal DNA using the program MSIsensor (Nui B et al. 2014). Essentially, the sequencing results are analyzed via MSIsensor to assess the number and length of homo-polymers / microsatellites within the targeted regions of tumor-normal sample pair. This results in a continuous rather than categorical MSI score assignment for the tumor sample. Loci are considered unstable (somatic) if k-mer distributions are significantly different between the tumor and matched normal using a standard multiple testing correction of x2 p-values. The percentage fraction of unstable sites is reported as the MSIsensor score. The assay uses a MSIsensor score threshold of 10 or greater to define MSI-H by MSIsensor.
5. Controls
- a) Matched normal control: Genomic DNA is extracted from patient-matched normal tissue (when available) or peripheral blood, for use as a matched normal control. In the event a matched normal is unavailable, or where the matched normal sample was sequenced with low coverage (<50X), tumor samples will be compared against a standard, in-batch pooled FFPE normal control for mutation calling; mutations called under these circumstances may include rare germline mutations and cannot be guaranteed to be somatic.
{6}------------------------------------------------
- b) Positive control: The positive control sample is a mixture of 3 tumor samples, each sample with a different confirmed SNV and at least one insertion or deletion, representing a range of mutation allele frequencies. Results are compared against a pooled FFPE negative control as an unmatched normal. Data generated from the mixed positive control sample are analyzed using the pipeline, and frequencies of the detected mutations are reviewed to determine if (1) the known mutations are among those called, and (2) the observed frequencies for the known mutations match their expected values within 5% of their values. The mixed FFPE positive control sample pools with expected variant frequency (VF) prior to pooling are shown in Table 2.
| Mixed Positive ID | Sample ID | VF | Known Mutation |
|---|---|---|---|
| M-1682-C3-T | 17% | KRAS Q61H | |
| M-1791-8C-T | 66% | EGFR L858R | |
| M-1913-BF | M-1754-DB-T | 61% | KITexon9ins |
| M-1671-CE-T | 25% | KITexon11del | |
| M-1693-5E-T | 24% | PIK3CA H1047R | |
| M-1914-A2 | M-1646-FC-T | 41% | BRAF V600E |
| M-1612-28-3-T | 32% | EGFR exon19 del | |
| M-1627-D9-T | 52% | NRAS Q61H | |
| M-1915-CA | M-1625-1A-2A-T | 28% | KRAS G12D |
Table 2. Positive Controls and Expected Mutation Frequencies
- c) Negative control: The negative control sample is a mixture of FFPE normal samples verified in previous reruns to be free of turnor contamination and germline copy number mutations in target genes. Polymorphisms unique to each constituent normal sample in the pool have been identified in prior analyses and the expected frequencies for each polymorphism in the pooled negative control are confirmed. The observed mutation frequencies are compared against the expected mutation frequencies for the 862 common SNPs, and the degree of concordance is measured using Pearson's correlation. The correlation between expected and observed mutation frequencies is expected to be 0.9 or higher.
- d) PCR reagent control [No Template Control (NTC)]: The NTC control should have a Qubit measurement of < 1.0ng/uL. Sequencing data from the NTC control sample will also be subjected to analysis using the pipeline, to verify that no known hotspot mutations are detected. Similar to the pooled FFPE negative control, if a hotspot mutation is detected, any samples containing that mutation in the pool will be reviewed to determine if a re-run is necessary.
6. Result Reporting:
- Oncopanel results are reported out under one of the two categories: "Cancer . Mutations with Evidence of Clinical Significance" or "Cancer Mutations with
{7}------------------------------------------------
Potential Clinical Significance". The two categories are based on the supporting level of clinical evidence. Refer to the Clinical Performance Section for more information.
- Results are reported for point mutations and small insertions and deletions in protein-● coding exons of the 468 gene panel. Refer to Appendix 1b for a list of genes.
- The MSK-IMPACT does not report mutations in 73 exons due to consistently low coverage in those exons. Refer to Appendix 1c for a list of excluded exons.
- Reporting takes in account the following quality metrics in the Table 3 below.
| QC Metrics | Acceptance Criteria |
|---|---|
| Coverage | Average target coverage > 200X |
| Coverage Uniformity | ≥ 98% target exons above 100X coverage |
| Base Quality | > 80% of bases with QS above > Q30 |
| % Cluster passing | The percent cluster passing filter (Cluster PF) > 80% |
| % Reads passingfilter | The percent reads passing filter (Reads PF) > 80% |
| Mutation Coverage (DP) ≥ 20, | |
| Hotspot Mutation*calling threshold | Number of Mutant Reads (AD) ≥ 8, |
| Mutation Frequency (VF) ≥ 2% | |
| Non-hotspotMutation**threshold | DP > 20, AD ≥ 10, VF ≥ 5% |
| Indels | Fewer than 20% of samples in an established 'standardnormal'database |
| Positive Run Control | The difference between the observed and expected frequencies for theknown mutations should be within 5%. |
| Negative RunControl | The correlation between expected and observed mutation frequenciesshould be 0.9 or higher |
| Sample-Mix up QC | Check over 1000 custom intergenic/intronic "fingerprint" SNPs.Flagged if pairs of samples from the same patient with > 5%discordance and from different patients with < 5% discordance |
| MajorContamination QC | % heterozygous sites at fingerprint SNPs < 55%; Average MAF athomozygous fingerprint SNPs < 2% |
| Criteria for callingtest failure | If a sample presents with mean coverage across all exons < 50x and nomutations are detected due to the low overall coverage, the test isdeemed "failed" for the sample. |
| Table 3. Sample Level Quality Control Metrics | ||
|---|---|---|
*Defined as Hotspot SNVs in COSMICv68, mutation hotspots reported in TCGA, reported in Cheng at. al.(Nature Biotech, 2016) and indels in selected exons of established oncogenes.
**SNVs and Indels other than the ones defined as hotspot mutations above.
{8}------------------------------------------------
J. Standard/Guidance Document Referenced (if applicable):
Not applicable
K. Test Principle:
The MSK-IMPACT assay is a custom targeted sequencing platform, utilizing solution-phase exon capture and sequencing, to detect somatic alterations (point mutations, small insertions and deletions, and microsatellite instability) in tumor specimens. The MSK-IMPACT assay involves hybridization capture and deep sequencing of all protein-coding exons of 468 cancer-associated genes. The assay uses custom DNA probes corresponding to all exons and selected introns of oncogenes and tumor suppressor genes. Probes are synthesized by a secondary manufacturer and are biotinylated to enable sequence enrichment through capture by streptavidin-conjugated beads. Probes were designed to tile the entire length of each target sequence in an overlapping fashion, typically extending 20-50 base pairs beyond the boundaries of the target. In total, the probes target approximately 1.5Mb of the human genome.
Genomic DNA is extracted from tumor and patient-matched blood/normal tissue as a normal control when available. Sequence libraries are prepared through a series of enzymatic steps including shearing of double-stranded DNA, end repair, A-base addition, ligation of barcoded sequence adaptors, and low cycle PCR amplification. Multiple barcoded sequence libraries are pooled and captured using the custom-designed biotinylated probes. Captured DNA fragments are then sequenced on an Illumina HiSeq2500 as paired-end reads. Sequence reads are then aligned to the reference human genome. By comparing the identity of bases from the tumor DNA to the matched normal DNA and the reference human genome, somatic alterations are identified in the tumor.
L. Performance:
1. Determination of pipeline thresholds:
- a) Requirements on exon coverage were established: A power analysis to compute the coverage or total number of reads needed to detect a mutation with true underlying mutation frequency 2% or greater, for varying levels of power (0.8 to 0.99), assuming a fixed alpha (Type I error rate) of 0.05 was conducted. Additionally, the 95% confidence interval ranges of observed mutation frequency as a function of coverage was also calculated. When the mutation is present at 10%, the 95% confidence interval with a coverage of 500X is expected to fall between 7.5% and 13%. When the overall coverage is 100X, the 95% CI for a mutation at 10% is estimated to fall between 5.0% and 17.6%.
To confirm these estimates, empirical data was obtained to measure the range of observed VF to expected VF using DNA from 10 normal FFPE samples from unrelated individuals which was mixed in equimolar parts so as to create a range of SNPs with expected frequencies as low as 5%. A total of 862 common SNPs were considered for this experiment.
{9}------------------------------------------------
A boxplot showing the observed mutation frequencies for the 862 common SNPs genotyped in the pooled normal sample binned by their true underlying mutation frequency is shown below. The results demonstrated that an observed VF range from 5.0% to 13.9% for a SNP with true underlying mutation frequency of 10% when the mean coverage of the sample was 480X. This range in values is roughly in line with what the theoretical statistical assessment for a coverage depth of 500X (7.5% to 13.0%). This data provided support for using a 5% as the lower limit for reporting mutations detected with true underlying frequency of 10%.
The boxplot in Figure 2 shows the correlation is 0.975, with a slope of 0.971 and intercept of -0.004. Consistent correlation is established as >0.9 as a QC metric for the whole pool analyzed.
Image /page/9/Figure/2 description: The image is a boxplot showing the relationship between true VAF (variant allele frequency) and observed VAF. The x-axis represents the true VAF, ranging from 0.05 to 1.0 in increments of 0.05. The y-axis represents the observed VAF, ranging from 0.0 to 1.0. Each boxplot shows the distribution of observed VAF values for a given true VAF, and the observed VAF generally increases as the true VAF increases.
Figure 2. Observed vs. Expected Variant Frequency
- b) Requirements on sample coverage: Ten normal (diploid) FFPE samples were profiled in duplicate using the IMPACT assay (total = 20 replicates) to generate summary statistics across all targeted exons. The mean coverage across all targeted exons for the normal samples was 571X (SD = 373X). Summary statistics were also computed on coverage values per exon normalized by per-sample coverage. There were exons that presented with consistently low coverage values. None of the exons of the genes in the clinical validation are among those with consistently low coverage. It was determined the low coverage was due to sequence similarity with other loci, and high GC content. The exons were removed from the MSK-IMPACT assay. Of the remaining exons across all genes, 99.5% were sequenced to a depth of 100X or greater while 98.6% were sequenced to a depth of 250X or greater. This analysis of normal samples indicates that with a mean sample coverage of 571X, 98% of exons are sequenced with coverage greater than 306X, or with normalized coverage greater
{10}------------------------------------------------
than 0.54. (The 'mean-normalized coverage' is the coverage of the mutation divided by the mean coverage across all exons; it serves as a measure of how deeply the validation exon was sequenced relative to the overall coverage of the sample. A mean-normalized coverage below 1 indicates the exon coverage is below average; conversely if greater than 1, it indicates above average coverage.) The data are shown in Figures 3 and Figure 4
Figure 3. Distribution of mean coverage values for targeted exons. Dashed line indicates coverage at 100X.
Image /page/10/Figure/2 description: The image is a histogram showing the frequency of coverage depth. The x-axis represents the coverage depth, ranging from 0 to 1500. The y-axis represents the frequency, with the highest frequency around a coverage depth of 500. A vertical dashed line is present at a coverage depth of approximately 100.
Figure 4. Distribution of mean coverage for targeted exons, normalized by persample coverage. Dashed line indicates 20% of mean sample coverage.
Image /page/10/Figure/4 description: The image is a histogram showing the distribution of normalized coverage depth. The x-axis represents the normalized coverage depth, ranging from 0.0 to 2.5. The y-axis represents the frequency, ranging from 0 to 400. The histogram shows a bell-shaped distribution, with a peak around a normalized coverage depth of 1.0.
Based on the calculations, 98% of exons can be expected to be sequenced to coverage greater than 100X, when mean sample coverage is 185X (0.54* 185X = 100X). (A 100X minimum coverage threshold per exon is required based on the power
{11}------------------------------------------------
calculations, which showed 100X coverage was necessary to call mutations with true underlying mutation frequency 10% or greater, with 95% power at an alpha level of 0.05).
To be conservative, a threshold of 200X on mean sample coverage is used to determine if a sample is sequenced to sufficient depth for subsequent analysis. A sample is flagged as being at increased risk of false negatives if its mean coverage is below 200X.
To provide empirical data for these requirements, MSK utilizes the pool normal sample with known expected single nucleotide mutations (n = 2436) and the underlying mutation allele fractions (MAF). In silico downsampling analysis was conducted with a pool normal mix down to 45% where the sample coverage decreased from 452X to 203X. At this coverage level, 94% of the mutations with expected underlying VAF of 10% were called.
- c) Requirements on mutation coverage, allele depth and frequency for positive calls: Permissive standard filters were used to intentionally generate false positives to identify suitable thresholds for parameters such as mutation coverage (DP), alternate allele depth (AD) and mutation frequency (VF) to optimize specificity. The following criteria allows optimal rejection of false positive SNVs (stratified by whether they are hotspots or not) and indel calls, while maintaining ability to detect true positive events with underlying frequency of 10% (5-17.6% observable). Potential strand-bias is also evaluated in the standard somatic mutation calling pipeline. An example of the number of false positive events detected pre and post filtering for coverage depth(DP), number of mutant reads (AD) and variant frequency (VF) is shown in Table 4.
| Mutations -Cosmic database | Mutations | |||
|---|---|---|---|---|
| Filter criteria | DP ≥ 20X, AD ≥ 8,VF ≥ 2% | DP ≥ 20X, AD ≥ 10,VF ≥ 5% | ||
| SNVs | Indels | SNVs | Indels | |
| Pre-filter | 1 | 24 | 342 | 40,793 |
| Post-filter | 0 | 0 | 0 | 8 |
| Rejection Rate | 1.00 | 1.00 | 1.00 | 0.999 |
| Table 4. Sample error correction by DP/AD/VF filter | |||
|---|---|---|---|
| ----------------------------------------------------- | -- | -- | -- |
-
- Pre-Analytical performance:
Minimum DNA requirements were established by measuring assay performance based on different inputs from normal blood and FFPE tumor samples. DNA samples are normalized to yield 50 - 250 ng input and maximized to 55 ul prior to shearing. The normalization and DNA quantification are performed.
- Pre-Analytical performance:
DNA extraction method was validated based on the invalid rates across multiple tumor types obtained from historical data. The data demonstrated that the DNA extraction has been optimized across tumor types to reasonably conclude that the analytical
{12}------------------------------------------------
performance presented is representative across FFPE tumor types. Table 5 shows the historical data for invalid rates from a retrospective chart review of >10,000 specimens tested with MSK-IMPACT. The range of invalid rates was 7.2% to 18.4%. The data shows that interference effects from different specimens are not significant across different tumor types supporting the performance of the pan-cancer specimen handling.
| Pre-RunInvalids | Pre-RunInvalids | Post-RunInvalids | ||||
|---|---|---|---|---|---|---|
| Tumor Type | SpecimenType | Numberof Tests | TumorInsufficient(Tumor %<20%) | DNAInsufficient(DNA yield<50ng) | SequencingFailure(Coverage<50X) | PercentInvalids |
| Non-Small CellLung Cancer | FFPE | 1995 | 53 | 208 | 75 | 16.8 |
| BreastCarcinoma | FFPE | 1588 | 41 | 126 | 97 | 16.6 |
| ColorectalCancer | FFPE | 1105 | 29 | 39 | 31 | 9.0 |
| Prostate Cancer | FFPE | 879 | 28 | 63 | 71 | 18.4 |
| Glioma | FFPE | 601 | 1 | 33 | 16 | 8.3 |
| PancreaticCancer | FFPE | 584 | 15 | 38 | 29 | 14.0 |
| Soft TissueSarcoma | FFPE | 479 | 3 | 21 | 13 | 7.7 |
| Bladder Cancer | FFPE | 480 | 12 | 20 | 25 | 9.8 |
| Melanoma | FFPE | 411 | 7 | 22 | 17 | 11.2 |
| Renal CellCarcinoma | FFPE | 403 | 12 | 15 | 16 | 10.7 |
| HepatobiliaryCancer | FFPE | 398 | 11 | 17 | 15 | 10.8 |
| EsophagogastricCarcinoma | FFPE | 374 | 5 | 12 | 16 | 8.8 |
| Germ CellTumor | FFPE | 332 | 9 | 13 | 30 | 8.1 |
| Thyroid Cancer | FFPE | 258 | 2 | 12 | 13 | 10.5 |
| Ovarian Cancer | FFPE | 244 | 4 | 8 | 8 | 8.2 |
| EndometrialCancer | FFPE | 235 | 2 | 8 | 7 | 7.2 |
| Head and NeckCarcinoma | FFPE | 208 | 8 | 8 | 6 | 10.5 |
| Cancer ofUnknownPrimary | FFPE | 224 | 15 | 15 | 10 | 17.8 |
Table 5. Specimen Invalid Rates for 17 FFPE Tumor Types
{13}------------------------------------------------
3. Analytical performance:
The hybridization-capture-based targeted re-sequencing assay is designed to detect point mutations [also referred to as single nucleotide variants (SNVs)] as well as small insertions/deletions (indels) < 30bp in length in the coding exons of 468 genes (Appendix 1b). A total of 6,357 exons are sequenced, 73 exons were excluded during assay development due to low sequence coverage and high GC content (Appendix 1c). A paired-sample analysis pipeline (tumor vs. matched normal) is used to identify somatic mutations in the targeted exons. MSK took a representative approach to validation of the SNVs and indels targeted in this panel, which is appropriate for variants of this type.1
- a) Precision Studies: The objective of the precision studies was to assess between-run and within-run precision. Extracted DNA was run once per day for 3 days using different barcodes for inter-day assessment (n=3). For one run, a sample was run in triplicates for intra-day assessment, resulting in a total of 3+1+1=5 replicates. For each replicate tested, all observed mutations were reported and assessed for precision. Details of the study are described below.
- Precision Panel: The precision of the MSK-IMPACT assay was assessed using i. 10 samples (9 FFPE specimens and one commercial cell line) to represent different tumor types, different mutation types, and the range of mutant allele frequencies. The panel included challenging specimens. The specimen panel was selected based on known mutations corresponding to "Cancer Mutations with Evidence of Clinical Significance" as well as the associated target tissue. The representative list of specimens is shown in Table 6.
| Tissue type | Mutation type | Gene/exon | cDNA change | Amino acid change | Mutationfrequency |
|---|---|---|---|---|---|
| Glioblastoma | INS | EGFRexon20 | C2290_2310dupTACGTGATGGCCAGCGTGGAC | p.Y764_D770dup | ~5% |
| CutaneousMelanoma | DNV | BRAFexon15 | c.1798_1799delinsAA | V600K | ~6.5% |
| UterineEndometrialCancer | SNV | KRASexon2 | C35G>C | G12A | ~7% |
| LungAdenocarcinoma | INS | ERBB2exon 20 | 2310_2311insGCATACGTGATG | E770_A771insAYVM | ~15% |
| LungAdenocarcinoma | SNV | EGFRexon 21 | 2573T>G | L858R | ~20% |
Table 6: Summary of the Specimens and Allele Frequencies in the Precision Studies
1 For complex structural variations, such as genomic rearrangements (fusions) and copy number variations (CNVs), the expectation is that the representative approach should be demonstrated at the gene level.
{14}------------------------------------------------
| Tissue type | Mutation type | Gene/ exon | cDNA change | Amino acid change | Mutation frequency |
|---|---|---|---|---|---|
| CRC | SNV | KRAS exon 2 | C34G>T | G12C | ~30% |
| Lung Adenocarcinoma | DEL (15bp) | EGFR exon 19 | 2236_2250delGAATTAA GAGAAGCA | E746_A750del | ~30% |
| CRC | SNV | BRAF exon 15 | c.1799T>A | V600E | ~40% |
| GIST | DEL (6bp) | Kit exon 11 | 1667_1672delAGTGGA | Q556_K558del | ~50% |
| FFPE Cell Line | DEL, SNV | Hotspot mutations in BRAF, EGFR, FLT3, GNA11, IDH1, KRAS, NRAS and PIK3CA genes | ~2%-15% |
- ii. Precision- Panel-Wide Reproducibility: The precision analysis was performed for the known mutations (as listed in Table 6), and also performed for all additional mutations identified in each specimen in any of the test replicates. A total of 69 mutations in the clinical specimens and 13 mutations in the cell line were detected for a total of 82 mutations. In addition to SNV/MNVs, there were 9 deletions and 8 insertions.
The results showed that all mutations have 100% concordance in all replicates except for 4 mutations in the clinical specimens and 3 mutations in the commercial sample. In the clinical specimen discordance was observed for an SNV (pQ64K) and a frameshift mutation (pL54fs) in AR exon1, an insertion (pA445_P446insP) ARID1B exon1; and a frameshift mutation (pT319Kfs*24) in PTEN exon 8. The discordance on AR and ARID1B mutations were due to poor mapping quality in the highly repetitive regions.
The 3 mutations from the commercial control sample that were discordant were 2 SNVs and one deletion (IDH1 exon4 R132H; BRAF exon15 V600M; EGFR exon19 E746 A750del). These 3 mutations were believed to be discordant because they have low frequencies near 2%.
The coefficient of variation (%CV) for the mutation allele frequency was also calculated for all 5 replicates. Thirty-four (45) of the 69 mutations in the clinical specimens had %CV ≤10%, 17/69 were between 10 and 20% and 7/69 were >21%. All results are summarized in Table 7. Each specimen is separated by a dark gray line. Known mutation within each specimen are in bold. Discordant cases are denoted in light grey. All runs passed the quality metrics criteria.
Table 7. Panel-wide precision summary for all 5 replicates Abbreviations: NC (normalized coverage); MAF (Mutant allele frequency)
{15}------------------------------------------------
| GeneExon | Mutation(cDNA/ProteinChanges) | NCrange | MAF range | MAFmean | MAFmedian | MAF(SD) | MAF(%CV) | Positive/TotalCalls | Positive Call Rate(two-sided 95% CI) |
|---|---|---|---|---|---|---|---|---|---|
| EGFRexon19 | c.2236_2250delGAATTAAGAGAAGCA746_750del | 0.84-1 | 0.311-0.342 | 0.323 | 0.316 | 0.013 | 4.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTENexon2 | c.T83G I28S | 0.62-0.73 | 0.502-0.569 | 0.543 | 0.544 | 0.027 | 5.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| TET2exon3 | c.C311G S104C | 1.04-1.32 | 0.085-0.103 | 0.098 | 0.102 | 0.008 | 8.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| TP53exon7 | c.C742T R248W | 0.97-1.22 | 0.648-0.664 | 0.66 | 0.663 | 0.007 | 1.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| BRAFexon15 | c.T1799A V600E | 1.26-1.44 | 0.415-0.454 | 0.431 | 0.425 | 0.015 | 3.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| BRCA2exon14 | c.A7388G N2463S | 0.84-0.96 | 0.19-0.23 | 0.209 | 0.21 | 0.015 | 7.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| BRD4exon19 | c.G3922A A1308T | 0.44-0.56 | 0.5-0.636 | 0.553 | 0.54 | 0.054 | 9.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| FBXW7exon9 | c.G1268T G423V | 0.91-1.05 | 0.369-0.418 | 0.395 | 0.391 | 0.02 | 5.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| GRIN2Aexon7 | c.C1514A A505E | 0.92-1.1 | 0.194-0.211 | 0.202 | 0.203 | 0.006 | 3.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTPRDexon12 | c.G10A V4I | 0.5-0.63 | 0.281-0.361 | 0.336 | 0.35 | 0.034 | 10.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| RUNX1exon9 | c.806-1G>A NA | 1.01-1.23 | 0.185-0.21 | 0.202 | 0.207 | 0.01 | 5.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| SPENexon12 | c.C10445TP3482L | 0.94-1.03 | 0.189-0.235 | 0.208 | 0.2 | 0.018 | 8.7% | 5/5 | 100.0% (47.8%, 100.0%) |
| SYKexon13 | c.C1768T R590W | 1.13-1.22 | 0.233-0.292 | 0.273 | 0.279 | 0.023 | 8.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| TP53exon6 | c.G610T E204X | 0.9-1.01 | 0.525-0.56 | 0.547 | 0.551 | 0.013 | 2.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| APCexon16 | c.G3856T E1286X | 0.8-1.05 | 0.326-0.39 | 0.351 | 0.349 | 0.026 | 7.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| APC exon7 | c.C646T R216X | 0.87-1.06 | 0.148-0.185 | 0.162 | 0.16 | 0.015 | 9.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| CREBBPexon29 | c.G4837AV1613M | 1-1.19 | 0.159-0.196 | 0.178 | 0.18 | 0.017 | 9.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| KRASexon2 | c.G34T G12C | 1.13-1.31 | 0.289-0.352 | 0.314 | 0.305 | 0.024 | 7.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| NOTCH1exon34 | c.7541dupCP2514fs | 1.28-1.5 | 0.144-0.211 | 0.184 | 0.189 | 0.025 | 13.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| SMAD4exon11 | c.C1333T R445X | 0.76-0.95 | 0.206-0.238 | 0.223 | 0.229 | 0.014 | 6.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| ALOX12Bexon11 | c.G1406A R469Q | 1.03-1.31 | 0.333-0.377 | 0.355 | 0.356 | 0.016 | 4.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| ARID1Bexon1 | c.1333_1334insCGC A445_P446insP | 0.2-0.2 | 0.2-0.2 | 0.2 | 0.2 | NA | NA | 1/5 | 20.0% (0.5%, 71.6%) |
| CDK8exon10 | c.C1014A D338E | 0.59-0.7 | 0.256-0.336 | 0.303 | 0.315 | 0.032 | 10.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| DNMT1exon36 | c.T4380G H1460Q | 1.18-1.51 | 0.51-0.558 | 0.534 | 0.53 | 0.017 | 3.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| ERBB2exon2 | c.G140A R47H | 1.16-1.59 | 0.596-0.712 | 0.656 | 0.666 | 0.045 | 6.9% | 5/5 | 100.0% (47.8%, 100.0%) |
| ERBB2exon20 | c.2310_2311insGCATACGTGATGE770_A771insAYVM | 1.02-1.38 | 0.142-0.199 | 0.173 | 0.171 | 0.023 | 13.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| ERCC2exon21 | c.C1904T A635V | 1.19-1.47 | 0.363-0.466 | 0.409 | 0.423 | 0.045 | 11.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| IRS1 exon1 | c.C3639A S1213R | 0.42-0.49 | 0.384-0.494 | 0.449 | 0.455 | 0.04 | 8.9% | 5/5 | 100.0% (47.8%, 100.0%) |
| MED12exon37 | c.5258_5282delCTCCTACCCTGCTAGAGCCTGAGAA A1753fs | 1.08-1.36 | 0.141-0.187 | 0.164 | 0.17 | 0.019 | 11.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| MED12exon43 | c.6339_6340insCAGCAACACCAGQ2113_Q2114insQQHQ | 0.96-1.43 | 0.37-0.422 | 0.4 | 0.399 | 0.021 | 5.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| NF1exon51 | c.C7595T A2532V | 0.92-1.04 | 0.627-0.68 | 0.664 | 0.676 | 0.022 | 3.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| NTRK1exon1 | c.G53A G18E | 0.28-0.55 | 0.6-0.668 | 0.631 | 0.63 | 0.027 | 4.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| PDGFRBexon7 | c.G946A V316M | 0.73-1.14 | 0.615-0.681 | 0.646 | 0.642 | 0.026 | 4.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3CBexon15 | c.A2150G N717S | 0.67-0.85 | 0.273-0.317 | 0.299 | 0.308 | 0.018 | 6.0% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTPRSexon32 | c.C4822TR1608W | 0.79-1.06 | 0.526-0.562 | 0.543 | 0.542 | 0.013 | 2.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| RB1 exon2 | c.138-2A>Gsplicing mutation | 0.51-0.75 | 0.231-0.345 | 0.291 | 0.284 | 0.047 | 16.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| TET1exon4 | c.G3476A R1159Q | 0.86-1.34 | 0.499-0.606 | 0.533 | 0.522 | 0.044 | 8.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| TP53exon5 | c.G524A R175H | 0.75-1.11 | 0.247-0.344 | 0.314 | 0.337 | 0.04 | 12.7% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon21 | c.T2573G L858R | 1.4-1.44 | 0.172-0.225 | 0.199 | 0.203 | 0.02 | 10.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| HNF1Aexon4 | c.C934T L312F | 0.35-0.54 | 0.033-0.077 | 0.057 | 0.059 | 0.016 | 28.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| MLL3exon42 | c.G9671A R3224H | 1.27-1.4 | 0.089-0.118 | 0.104 | 0.105 | 0.011 | 10.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| NTRK3exon14 | c.1401delC P467fs | 0.49-0.54 | 0.062-0.086 | 0.074 | 0.077 | 0.01 | 13.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| TP53exon10 | c.A1051T K351X | 0.74-0.84 | 0.075-0.116 | 0.103 | 0.108 | 0.016 | 15.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| AR exonl | c.161_171delTGCTGCTGCTGL54fs | 0.34-0.39 | 0.079-0.097 | 0.088 | 0.087 | 0.009 | 10.2% | 3/5 | 60.0% (14.7%, 94.7.0%) |
| AR exon1 | c.C190A Q64K | 0.25-0.29 | 0.134-0.135 | 0.134 | 0.134 | 0.001 | 0.7% | 2/5 | 40.0% (5.3%, 85.3%) |
| KITexon11 | c.1667_1672delAGTGGA556_558del | 1.65-1.86 | 0.554-0.595 | 0.569 | 0.566 | 0.016 | 2.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| KITexon17 | c.T2467G Y823D | 1.28-1.49 | 0.619-0.658 | 0.646 | 0.655 | 0.016 | 2.5% | 5/5 | 100.0%(47.8%, 100.0%) |
| RPS6KB2exon10 | c.G840T K280N | 0.93-1.19 | 0.435-0.473 | 0.462 | 0.468 | 0.015 | 3.2% | 5/5 | 100.0%(47.8%, 100.0%) |
| CARD11exon25 | c.3382T>Ap.V1128I | 1.34-1.58 | 0.276-0.293 | 0.284 | 0.278 | 0.009 | 3.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon20 | c.2290_2310dupTACGTGATGGCCAGCGTGGACp.Y764_D770dup | 14.36-15.46 | 0.05-0.06 | 0.055 | 0.055 | 0.004 | 7.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon7 | c.874G>Tp.V292L | 21.51-21.82 | 0.934-0.939 | 0.937 | 0.939 | 0.002 | 0.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| NOTCH3exon22 | c.3646G>Ap.A1216T | 1.35-1.52 | 0.247-0.318 | 0.281 | 0.281 | 0.026 | 9.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTENexon5 | c.395G>Cp.G132A | 0.6-0.72 | 0.605-0.667 | 0.635 | 0.631 | 0.029 | 4.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| RUNX1exon8 | c.899C>Tp.T300M | 0.81-0.92 | 0.244-0.274 | 0.26 | 0.266 | 0.015 | 5.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| STAG2exon17 | c.1544_1547delATAG p.D515Gfs*6 | 0.19-0.27 | 0.677-0.842 | 0.753 | 0.741 | 0.067 | 8.9% | 5/5 | 100.0% (47.8%, 100.0%) |
| TERTPromoter | g.1295228C>Tnon-coding | 0.55-0.67 | 0.388-0.467 | 0.421 | 0.417 | 0.033 | 7.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| AKT3exon2 | c.134T>G p.V45G | 1.14-1.36 | 0.05-0.078 | 0.066 | 0.067 | 0.012 | 18.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| BRAFexon15 | c.1798_1799delinsAA p.V600K | 1.04-1.32 | 0.065-0.095 | 0.072 | 0.067 | 0.013 | 18.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| KITexon11 | c.1735_1737delGAT p.D579del | 1.08-1.22 | 0.051-0.056 | 0.053 | 0.054 | 0.002 | 3.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| CTCFexon3 | c.610dupAp.T204Nfs*26 | 0.68-0.86 | 0.041-0.072 | 0.057 | 0.061 | 0.014 | 24.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon20 | c.2317_2319dupCAC p.H773dup | 1.15-1.19 | 0.067-0.093 | 0.078 | 0.079 | 0.011 | 14.1% | 5/5 | 100.0% (47.8%, 100.0%) |
| KDM5Cexon23 | c.3755G>Ap.R1252H | 0.88-1.17 | 0.064-0.13 | 0.088 | 0.084 | 0.026 | 29.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| KRASexon2 | c.35G>C p.G12A | 0.78-0.94 | 0.044-0.106 | 0.076 | 0.074 | 0.023 | 30.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3R1exon13 | c.1672_1683delGAAATTGACAAAp.E558_K561del | 0.43-0.52 | 0.067-0.116 | 0.085 | 0.081 | 0.019 | 22.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3R1exon9 | c.1023dupAp.E342Rfs*4 | 0.41-0.58 | 0.056-0.102 | 0.083 | 0.086 | 0.017 | 20.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3R1exon9 | c.1024G>Tp.E342* | 0.42-0.59 | 0.064-0.108 | 0.093 | 0.095 | 0.017 | 18.3% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTENexon6 | c.493-1G>Ap.X165_splice | 0.53-0.64 | 0.173-0.208 | 0.192 | 0.187 | 0.015 | 7.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| PTENexon8 | c.956_959delCTTTT p.T319Kfs*24 | 0.28-0.48 | 0.006-0.079 | 0.049 | 0.052 | 0.029 | 59.2% | 3/5 | 60.0% (14.7%, 94.7.0%) |
| SOX17exon1 | c.287C>G p.A96G | 1.16-1.51 | 0.061-0.074 | 0.069 | 0.069 | 0.005 | 7.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| BRAFexon15 | c.1798G>AV600M | 0.97-1.06 | 0.016-0.041 | 0.027 | 0.027 | 0.01 | 37.0% | 3/5 | 60.0% (14.7%, 94.7.0%) |
| BRAFexon15 | c.1799T>A V600E | 0.97-1.06 | 0.051-0.08 | 0.064 | 0.067 | 0.012 | 18.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon18 | c.2155G>AG719S | 1.23-1.33 | 0.125-0.179 | 0.158 | 0.164 | 0.022 | 13.9% | 5/5 | 100.0% (47.8%, 100.0%) |
| EGFRexon19 | c.2235_2249delGGAATTAAGAGAAGCE746_A750del | 1.01-1.19 | 0.009-0.043 | 0.023 | 0.019 | 0.013 | 56.5% | 2/5 | 40.0% (5.3%, 85.3%) |
| FLT3exon20 | c.2503G>TD835Y | 0.97-1.02 | 0.037-0.059 | 0.045 | 0.043 | 0.008 | 17.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| GNA11exon5 | c.626A>T Q209L | 1.41-1.48 | 0.036-0.054 | 0.046 | 0.044 | 0.008 | 17.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| IDH1exon4 | c.395G>A R132H | 0.5-0.53 | 0.038-0.049 | 0.035 | 0.044 | 0.020 | 57.1% | 4/5 | 80.0% (28.4%, 99.5%) |
| KRASexon2 | c.34G>A G12S | 0.9-1.03 | 0.026-0.057 | 0.041 | 0.039 | 0.011 | 26.8% | 5/5 | 100.0% (47.8%, 100.0%) |
| KRASexon2 | c.38G>A G13D | 0.91-1.06 | 0.217-0.249 | 0.231 | 0.229 | 0.012 | 5.2% | 5/5 | 100.0% (47.8%, 100.0%) |
| KRASexon4 | c.436G>A A146T | 0.82-0.88 | 0.031-0.055 | 0.042 | 0.044 | 0.009 | 21.4% | 5/5 | 100.0% (47.8%, 100.0%) |
| NRASexon3 | c.183A>T Q61H | 1.01-1.14 | 0.039-0.065 | 0.051 | 0.051 | 0.01 | 19.6% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3CAexon10 | c.1624G>AE542K | 0.67-0.87 | 0.038-0.047 | 0.042 | 0.042 | 0.004 | 9.5% | 5/5 | 100.0% (47.8%, 100.0%) |
| PIK3CAexon21 | c.3140A>GH1047R | 0.62-0.72 | 0.222-0.331 | 0.276 | 0.258 | 0.05 | 18.1% | 5/5 | 100.0% (47.8%, 100.0%) |
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
{20}------------------------------------------------
- iii. Per Specimen Precision: Results of the precision studies were combined and precision across all reportable genes was determined for each specimen. The positive call rate based on the total number of mutations along with the 2-sides 95% confidence interval were calculated. Results are summarized in Table 8.
| Specimen | Total Nouniquemutationsdetectedacrossall 5replicates* | *Positive callrateper mutation | Positive call rate*(two-sided 95% CI) | Negative call rate(two-sided 95% CI) |
|---|---|---|---|---|
| M15-22924 | 5 | 5/5 for all | 25/25100.0% (86.3%, 100.0%) | - |
| M15-3038 | 3 | 5/5 for all | 15/15100.0% (78.2%, 100.0%) | - |
| M16-19000 | 10 | 5/5 for 94/5 for 1 | 49/5098.0% (89.4%, 99.9%) | - |
| M1688-5C | 18 | 5/5 for 171/5 for 1 | 86/9095.6% (89.0%, 98.8%) | 4/580.0% (28.4%,99.5%) |
| M-1698-A9 | 5 | 5/5 for all | 25/25100.0% (86.3%, 100.0%) | - |
| M-1654-CA | 6 | 5/5 for all | 30/30100.0% (88.4%, 100.0%) | - |
| M-1612-28 | 4 | 5/5 for all | 20/20100.0% (83.2%, 100.0%) | - |
| M1648-D5 | 10 | 5/5 for all | 50/50100.0% (92.9%, 100.0%) | - |
| M-1707-12 | 5 | 5/5 for 33/5 for 1;2/5 for 1 | 20/2580.0% (59.3%, 93.2%) | 3/560.0% (14.7%,94.7%) |
| Commercialsample | 13 | 5/5 for 10;4/5 for 1 ;3/5 for 1;2/5 for 1 | 59/6590.8% (81.0%, 96.5%) | 3/560.0% (14.7%,94.7%) |
Table 8. Precision per specimen across all reportable mutations (N - 5 replicates)
*Positive call rate is calculated based on variants with majority call detected as positive #Negative call rate is calculated based on variants detected at least once, but with majority call as negative. For all other locations, the negative call rates are 100%.
The precision study was also evaluated for the intra-assay repeatability (withinrun). All results were concordant except for ARID1B exon 2 insertion from clinical specimen M-1688, and BRAF V600M point mutation in the commercial control sample as described previously. Additionally, performance with respect to quality metrics (i.e., total depth of coverage and mutant allele coverage) in all replicates was also summarized and shown to meet the pre-specified acceptance criteria (data not shown).
{21}------------------------------------------------
- iv. Precision Well-characterized reference material: The precision of MSK-IMPACT was assessed through repeated measurements of a well characterized reference standard (HapMap cell line NA20810). To determine sequencing error rates for the reference sample, DNA extracted from the HapMap cell line was included in each run tested in the accuracy study. The study investigated whether the SNPs in the targeted exons were detected at their expected frequencies. Reference genotypes for 11,767 SNPs in the targeted exons using a whole genome sequencing BAM file for NA20810, were obtained from the 1000 Genomes database. A total of 11.443 SNPs (97.2%) were homozygous for the major allele (relative to the hg19 reference genome), 212 SNPs (1.8%) were heterozygous and 112 SNPs (0.95%) were homozygous for the minor allele. The strong bias towards alleles matching the reference genome was expected, given that these SNPs occur in coding exons and there is likely strong selective pressure against deviations from the reference sequence. NA20810 was profiled with the assay multiple times across different runs, for a total of 23 replicates. Zygosity results were 100% concordant and high levels of concordance specifically, the difference between the expected and mean observed mutation frequencies was very small (absolute difference = 0.09%±0.45%). The data provide additional supplemental evidence of the reproducibility of the assay.
- v. Precision for Microsatellite Instability (MSI): Precision of the MSI calling by MSIsensor was demonstrated with a total of 12 specimens: 6 MSI-H specimens (at three MSI-score levels, 3 replicates per sample) and 6 MSS specimens. Each DNA extracted sample was tested with 3 inter- and 3 intra-run replicates. Multiple barcodes were included. All samples had 100% agreement between calls. The total number of unstable loci relative to the total number of sites surveyed along with the mean, median and standard deviation (SD) and coefficient of variance (%CV) was also presented for each specimen and score. The results supported the precision of the MSIsensor scores greater than 0.5 Results are shown in Table 9.
| N | TotalSites_range | UnstableLoci_range | Mean | Median | SD | %CV | Positive CallRate (two-sided 95% CI) |
|---|---|---|---|---|---|---|---|
| 5 | 1227-1458 | 518-650 | 43.00 | 43.00 | 1.22 | 2.8% | 100%(47.8%,100.0%) |
| 5 | 1158-1477 | 483-646 | 43.00 | 43.00 | 0.71 | 1.7% | 100%(47.8%,100.0%) |
| 5 | 1187-1429 | 500-613 | 42.00 | 42.00 | 0.71 | 1.7% | 100%(47.8%,100.0%) |
| 5 | 1287-1400 | 303-359 | 24.80 | 25.00 | 0.84 | 3.4% | 100%(47.8%,100.0%) |
| 5 | 1251-1303 | 240-318 | 23.40 | 24.00 | 2.51 | 10.7% | 100%(47.8%,100.0%) |
| 5 | 1154-1379 | 153-175 | 12.60 | 12.00 | 0.89 | 7.1% | 100%(47.8%,100.0%) |
Table 9. Precision of the MSIsensor Score Using 12 Specimens
{22}------------------------------------------------
| N | TotalSites_range | UnstableLoci_range | Mean | Median | SD | %CV | Positive CallRate (two-sided 95% CI) |
|---|---|---|---|---|---|---|---|
| 5 | 1321-1545 | 46-58 | 3.60 | 4.00 | 0.55 | 15.3% | 100%(47.8%,100.0%) |
| 5 | 1535-1604 | 44-64 | 3.40 | 3.00 | 0.55 | 16.2% | 100%(47.8%,100.0%) |
| 5 | 1411-1612 | 28-38 | 2.20 | 2.00 | 0.45 | 20.5% | 100%(47.8%,100.0%) |
| 5 | 1438-1528 | 6-9 | 0.48 | 0.50 | 0.08 | 16.7% | 100%(47.8%,100.0%) |
| 5 | 1315-1487 | 0-2 | 0.02 | 0.00 | 0.04 | 223.6% | 100%(47.8%,100.0%) |
| 5 | 1312-1532 | 0-1 | 0.01 | 0.00 | 0.03 | 223.6% | 100%(47.8%,100.0%) |
- b) Analytical Sensitivity Limit of Detection (LoD): The LoD of the IMPACT assay is defined as the mutant allele fraction at which 95% of replicates across all replicates for a variant type are reliably detected. Studies were conducted to demonstrate a putative LoD for each variant type. In the first part, a dilution series was conducted to identify the lowest reliable mutant fraction. In part 2, the putative LoD was confirmed with multiple replicates.
- Part 1: Dilution Series: The mean normalized coverage for all exons was i. determined for 10 normal FFPE specimens and the LoD was assessed with samples containing mutations in 5 validation exons (defined as representative exons harboring cancer mutations with evidence of clinical significance assessed in the accuracy study) with the lowest and highest coverage.
- . The 5 validation exons with lowest coverage correspond to 3 exons harboring SNVs, (ERBB2 exon 20 (V777L), PDGFRA exon 18 (D842V), PIK3CA exon 10 (E545K), and 2 exons harboring indels (EGFR exon 19 and KIT exon 9).
- The 5 validation exons with highest coverage correspond to 3 exons harboring SNVs (BRAF exon 15 (V600E), KRAS exon 2 (G12D) and PIK3CA exon 2 (R88Q) and 2 exons harboring indels (KIT exon 11 and EGFR exon 20).
- Part 1: Dilution Series: The mean normalized coverage for all exons was i. determined for 10 normal FFPE specimens and the LoD was assessed with samples containing mutations in 5 validation exons (defined as representative exons harboring cancer mutations with evidence of clinical significance assessed in the accuracy study) with the lowest and highest coverage.
Five to eight serial dilutions were prepared using patient samples positive for the mutations listed above, where tumor samples were either diluted with their respective matched FFPE normal sample (when available) or a previously sequenced, unmatched normal FFPE sample. One replicate at each dilution was tested and the ability to detect the mutation of interest was measured. All results were called at the lowest dilution except for PIK3CA which was called wild-type at the lowest dilution. Results are shown in Tables 10A-J.
{23}------------------------------------------------
| Table 10A. Limit of Detection -Part 1 | ||
|---|---|---|
| -- | -- | --------------------------------------- |
| SNV BRAF Exon 15 (Sample M-1648-D5-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | c.1799T>A | V600E | 1018 | 410 | 0.4 | Called |
| 1:2 | 1044 | 319 | 0.31 | Called | ||
| 1:4 | 888 | 173 | 0.19 | Called | ||
| 1:8 | 999 | 91 | 0.09 | Called | ||
| 1:16 | 783 | 26 | 0.03 | Called | ||
| 1:32 | 845 | 20 | 0.02 | Called |
Table 10B
| SNV KRAS Exon 2 (sample M-1807-ED-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNA change | AA Change | DP | AD | VF | Result |
| Neat | G12D | 907 | 405 | 0.45 | Called | |
| 1:2 | 820 | 298 | 0.36 | Called | ||
| 1:4 | c.35G>A | 400 | 97 | 0.24 | Called | |
| 1:8 | 660 | 121 | 0.18 | Called | ||
| 1:16 | ર્ભર્ટ | ਦੇ ਰੇ | 0.09 | Called | ||
| 1:32 | 632 | 41 | 0.06 | Called |
Table 10C
| SNV PIK3CA Exon 2 (Sample M-1729-E1-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | R88Q | 2029 | 629 | 0.31 | Called | |
| 1:2 | c.263G>A | 1008 | 211 | 0.21 | Called | |
| 1:4 | 1140 | 145 | 0.13 | Called | ||
| 1:8 | 997 | 62 | 0.06 | Called | ||
| 1:16 | WT |
Table 10D
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
|---|---|---|---|---|---|---|
| Neat | 2503 | 922 | 0.37 | Called | ||
| 1:2 | 1986 | 688 | 0.35 | Called | ||
| 1:4 | 1513 | 430 | 0.28 | Called | ||
| 1:8 | c.1667_1681delAGT | 556_561del | 1049 | 250 | 0.24 | Called |
| 1:16 | GGAAGGTTGTTG | 792 | 138 | 0.17 | Called | |
| 1:32 | 761 | 66 | 0.09 | Called | ||
| 1:64 | 618 | 37 | 0.06 | Called | ||
| 1:125 | 736 | 18 | 0.02 | Called |
Table 10E
| EGFR exon 20 Insertion (sample M-1674-10-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | 1484 | 400 | 0.27 | Called | ||
| 1:2 | 777 | 166 | 0.21 | Called | ||
| 1:4 | c.2308_2309insACT | D770_N771insY | 566 | 105 | 0.19 | Called |
| 1:8 | 595 | 55 | 0.09 | Called | ||
| 1:16 | 581 | 33 | 0.06 | Called | ||
| 1:32 | 608 | 21 | 0.03 | Called |
{24}------------------------------------------------
Table 10F
| SNV ERBB2 exon 20 (sample M-1801-98-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | 1471 | 408 | 0.28 | Called | ||
| 1:2 | 1482 | 240 | 0.16 | Called | ||
| 1:4 | c.2525A>T | D842V | 864 | 73 | 0.08 | Called |
| 1:8 | 903 | 38 | 0.04 | Called | ||
| 1:16 | 873 | 24 | 0.03 | Called |
Table 10G
| SNV PDGFR α Exon 18 (sample M-1670-A6-T | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | 448 | 236 | 0.53 | Called | ||
| 1:2 | 636 | 142 | 0.22 | Called | ||
| 1:4 | c.1633G>A | E545K | 962 | 95 | 0.1 | Called |
| 1:8 | 647 | 45 | 0.07 | Called | ||
| 1:16 | 707 | 16 | 0.02 | Called |
Table 10H
| SNV PK3CA exon 10 (sample M-1434-A5-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | 448 | 236 | 0.53 | Called | ||
| 1:2 | 636 | 142 | 0.22 | Called | ||
| 1:4 | c.1633G>A | E545K | 962 | 95 | 0.1 | Called |
| 1:8 | 647 | 45 | 0.07 | Called | ||
| 1:16 | 707 | 16 | 0.02 | Called |
Table 10I
| EGFR exon 19 deletion (sample M-1809-C4-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNAchange | AA Change | DP | AD | VF | Result |
| Neat | c.2236_2250delGAATTAAGAGAAGCA | 746_750del | 1278 | 790 | 0.62 | Called |
| 1:2 | 1137 | 484 | 0.43 | Called | ||
| 1:4 | 792 | 207 | 0.26 | Called | ||
| 1:8 | 666 | 94 | 0.14 | Called | ||
| 1:16 | 622 | 49 | 0.08 | Called | ||
| 1:32 | 746_750del | 499 | 17 | 0.03 | Called |
Table 10J
| Kit Exon 9 insertion (sample M-1754-DB-T) | ||||||
|---|---|---|---|---|---|---|
| Dilution | cDNA change | AA Change | DP | AD | VF | Result |
| Neat | 517 | 314 | 0.61 | Called | ||
| 1:2 | 512 | 187 | 0.37 | Called | ||
| 1:4 | c.1502_1503insTGCCTA | S501_A502insAY | 641 | 89 | 0.14 | Called |
| 1:8 | 486 | 27 | 0.06 | Called | ||
| 1:16 | 447 | 17 | 0.04 | Called | ||
| 1:32 | 521 | 14 | 0.03 | Called |
{25}------------------------------------------------
- ii. Part 2: Confirmation of the LoD. A total of 5 replicates were tested for each of the 3 deletions, 4 insertions and 6 SNVs at 5% minor allele frequency. All variants have 100% positive call rates except for one replicate for a deletion on PTEN exon 6. This replicate also failed the mutation read depth and was below the estimated LoD of 5%. The results are shown in Table 11.
| Type | Mutation | GeneExon | RangeDP | RangeAD | RangeMAF | RangeNormDP | PositiveCallRate |
|---|---|---|---|---|---|---|---|
| DEL | In_Frame_Delc.1735_1737delGATp.D579del | KIT exon11 | 509-693 | 26-38 | 0.051-0.056 | 1.08-1.22 | 100.0% |
| DEL | Frame_Shift_Delc.956_959delCTTTp.T319Kfs*24 | PTEN exon8 | 197-242 | 7-19 | 0.036-0.079 | 0.31-0.48 | 80.0% |
| DEL | In_Frame_Delc.1672_1683delGAAATTGACAAAp.E558_K561del | PIK3R1exon13 | 216-313 | 18-36 | 0.067-0.116 | 0.43-0.52 | 100.0% |
| INS | In_Frame_Insc.2317_2319dupCACp.H773dup | EGFR exon20 | 587-749 | 46-65 | 0.067-0.093 | 1.15-1.19 | 100.0% |
| INS | Frame_Shift_Insc.1023dupA p.E342Rfs*4 | PIK3R1 exon9 | 236-345 | 15-32 | 0.056-0.102 | 0.41-0.58 | 100.0% |
| INS | Frame_Shift_Insc.610dupA p.T204Nfs*26 | CTCF exon3 | 344-540 | 14-36 | 0.041-0.072 | 0.68-0.86 | 100.0% |
| INS | In_Frame_Insc.2290_2310dupTACGTGATGGCCAGCGTGGACp.Y764_D770dup | EGFR exon20 | 8601-9836 | 441-572 | 0.05-0.06 | 14.36-15.46 | 100.0% |
| SNV | Missense_Mutationc.134T>G p.V45G | AKT3 exon2 | 535-813 | 28-63 | 0.05-0.078 | 1.14-1.36 | 100.0% |
| SNV | Missense_Mutationc.1798_1799delinsAAp.V600K | BRAF exon15 | 489-747 | 33-71 | 0.065-0.095 | 1.04-1.32 | 100.0% |
| SNV | Missense Mutationc.287C>G p.A96G | SOX17 exon1 | 672-805 | 45-59 | 0.061-0.074 | 1.16-1.51 | 100.0% |
| SNV | Missense_Mutationc.35G>C p.G12A | KRAS exon2 | 445-571 | 20-55 | 0.044-0.106 | 0.78-0.94 | 100.0% |
| SNV | Missense_Mutationc.3755G>A p.R1252H | KDM5Cexon23 | 475-733 | 40-68 | 0.064-0.13 | 0.88-1.17 | 100.0% |
| SNV | Nonsense_Mutationc.1024G>T p.E342* | PIK3R1 exon9 | 242-355 | 18-37 | 0.064-0.108 | 0.42-0.59 | 100.0% |
Table 11. Limit of Detection– Part 2
{26}------------------------------------------------
- iii. Microsatellite instability (MSI): The minimum tumor proportion required to support the MSIsensor score robustness was assessed using CRC specimens. Five (5) replicates were run using multiple barcodes and runs. The data showed that qualitatively, the assay and score are reproducible to 8% tumor proportion, though a decreasing trend in the quantitative score was observed. Therefore, the minimum tumor proportion required for the assay was established as 25% with an average coverage of 200X. Separately, regardless of the tumor proportion, data showed that the score is robust across the MSIsensor score range (refer to Table 9 above and Table 12).
| Tumor Purity | Coverage | # Total site | # Unstable loci | MSIsensor Score (%) |
|---|---|---|---|---|
| Diluted to 8% | 517 | 1420 | 182 | 13 |
| Diluted to 8% | 562 | 1389 | 175 | 13 |
| Diluted to 8% | 555 | 1352 | 185 | 14 |
| Diluted to 8% | 502 | 1361 | 135 | 10 |
| Diluted to 8% | 378 | 1273 | 152 | 12 |
| Sample ID | Replicate | MSI Sensor Score |
|---|---|---|
| CRC-01 | 1 | 0.00000 |
| 2 | 0.00000 | |
| 3 | 0.00000 | |
| CRC-04 | 1 | 0.00000 |
| 2 | 0.00000 | |
| 3 | 0.00000 | |
| CRC-08 | 1 | 0.00000 |
| 2 | 0.00000 | |
| 3 | 0.00000 |
- iv. DNA-Input: The validated DNA concentration is the amount at which the average read depth over the exon regions was maintained at the criteria established (e.g., ≥20 reads per base), and have 100% positive mutation call rate. The optimized and recommended DNA concentration for the assay is 250ng. The DNA input range 50-250ng. was assessed for accuracy and sequencing failures as a function of the input DNA concentration. The results show that assay performance in terms of sequencing failures is a function of genomic DNA input values as shown in Table 13.
| DNA Input | Success | Sequencing Failure |
|---|---|---|
| 250ng | 97% | 3% |
| 201-249ng | 87% | 13% |
| 151-200ng | 87% | 13% |
| 101-150ng | 81% | 19% |
| 50-100ng | 78% | 22% |
Table 13. Sequencing Failures Relative to DNA Input
c) Linearity/assay reportable range:
Not applicable
d) Traceability (controls, calibrators, or methods):
The MSK-IMPACT is not traceable to any known standard. Controls and quality metrics are described in the device description section.
{27}------------------------------------------------
e) Stability:
Reagent stability is based on manufacturer expiration dating, and supported by MSK verification. Stability of the reagents is monitored through the use of consistent controls.
f) Expected values:
The laboratory follows protocols for the use of controls consistent with CLIA regulation. The MSK-IMPACT does not use calibrators; however, the verification of mutant allele frequency is maintained by analysis of a pooled control with expected allele frequencies.
g) Analytical specificity:
High analytical specificity is maintained by paired tumor/matched normal sequencing, and was established during assay optimization.
Interference:
The MSK-IMPACT assay pre-analytic steps are designed to minimize interference. The invalid rates in the historical testing from >10,000 samples support that any interference from any challenging tissues is minimized.
h) Assay cut-off:
The MSK-IMPACT does not report mutations below 2% for known hotspot mutations and 5% for non-hotspot mutations.
Comparison studies: i)
- i. Method comparison:
The MSK-IMPACT assay is designed to detect SNVs and small indels in 6284 exons from 468 genes. The accuracy of the MSK-IMPACT was assessed by comparison of the MSK-IMPACT result to the original results obtained with the validated orthogonal methods. Testing was conducted per protocol. A total of 267 unique mutations in 433 FFPE tumor specimens representing 48 exons in 20 genes were tested and are listed in Table 14 below.
| Gene(n=20) | #Samples(n=433) | Exon(n=48) | Type | Mutations Assessed |
|---|---|---|---|---|
| AKT | 10 | exon3 | SNV | E17K |
| ALK | 3 | exon23 | SNV | F1174V/L;S1205F |
| ALK | 4 | exon25 | SNV | R1275Q;R1260T |
| BRAF | 11 | exon11 | SNV | G466V/R;S467L;G469* |
| BRAF | 19 | exon15 | SNV | D594G;V600*;K601I |
| EGFR | 10 | exon18 | SNV | G719A/S; G724S |
| EGFR | 12 | exon19 | DEL | 745_750del; 746_748del; 746_750del;747_753del; K754fs |
| Gene(n=20) | #Samples(n=433) | Exon(n=48) | Type | Mutations Assessed |
| 10 | exon20 | SNV | T790M | |
| 16 | exon20 | INS | M766_A767insASV; V769_D770insDNP;D770_N771ins*;P772_H773ins*;H773_V774insY/H | |
| 9 | exon21 | SNV | L858R | |
| ERBB2 | 7 | exon19 | SNV | L755S;I767M;D769Y |
| 16 | exon20 | INS | E770_A771insAYVM;A771_Y772insYVMA;G776_G778ins* | |
| 3 | exon20 | SNV | V777L;G776V | |
| 7 | exon8 | SNV | S310F/Y; S305C | |
| FGFR2 | 1 | exon12 | SNV | L528H |
| 1 | exon7 | SNV | S252W | |
| 1 | exon9 | SNV | Y375C | |
| FGFR3 | 2 | exon18 | SNV | P797L |
| 1 | exon7 | SNV | A261V;A265V | |
| 5 | exon9 | SNV | F384L | |
| 1 | exon9 | INS | G370_S371insH | |
| GNA11 | 7 | exon5 | SNV | Q209L |
| GNAQ | 5 | exon5 | SNV | Q209P/L |
| GNAS | 5 | exon8 | SNV | R201C/H |
| HRAS | 3 | exon2 | SNV | G10A; G13D/V |
| 5 | exon3 | SNV | A59V; Q61R/L/K | |
| IDH1 | 8 | exon4 | SNV | R132G/C/H |
| IDH2 | 5 | exon4 | SNV | R172*;R140Q |
| 1 | exon4 | DEL | T146Lfs*15 | |
| 9 | exon11 | INS; DEL | K550fs; 552_557del; 556_558del;556_561del; 558_565del; 559_566del;P573_T574insTQLPS | |
| KIT | 9 | exon11 | SNV | V555L; W557G; V559D; D572G;L576P. |
| 6 | exon13 | SNV | V654A; K642E | |
| 5 | exon17 | SNV | D816H; D820E; N822K | |
| 10 | exon9 | INS | S501_A502insAY; A502_Y503dup | |
| KRAS | 16 | exon2 | SNV | G12*; G13D |
| 13 | exon3 | SNV | Q61* | |
| 10 | exon4 | SNV | K117N;G138E;A146* | |
| MET | 13 | exon14 | SNV | D1010*; Exon14 skipping |
| 19 | exon14 | DEL | Exon14 skipping; Other splicing defects | |
| NRAS | 4 | exon2 | SNV | G13* |
| 12 | exon3 | SNV | Q61* | |
| PDGFRA | 12 | exon18 | SNV | D842V/I |
| 1 | exon12 | SNV | V561D | |
| 4 | exon10 | SNV | E545A/K; E542K | |
| PIK3CA | 2 | exon21 | SNV | H1047R/Y |
| 1 | exon21 | INS | X1069delinsFL | |
| 8 | exon2 | SNV/MNV | F83L;R88Q;R93Q;K111E/N | |
| 2 | exon2 | DEL | E110del; 112_113del | |
| 9 | exon5 | SNV | V344M;N345I/K | |
| 9 | exon8 | SNV | E418K;C420R;P449R;E453K/Q | |
| 1 | exon8 | DEL | E453_D454del | |
| TP53 | 9 | exon4 | SNV/MNV | W53X;W91X;Q100X;G105V/C; S106R;F113C |
| 6 | exon4 | DEL | L355fs;P67fs;A84fs;109_109del;G108fs; | |
| Gene | #Samples | Exon | ||
| (n=20) | (n=433) | (n=48) | Type | Mutations Assessed |
| R110fs | ||||
| 3 | exon4 | INS | V73fs;L114fs;C124fs | |
| б | exon5 | SNV | K132Q;W146X; Y163C; R175H; R158H | |
| 3 | INS | P153fs; M160 A161insRA; | ||
| exon5 | Q167_M170dup | |||
| K132fs;A138fs;P152fs; R156fs; | ||||
| 9 | exon5 | DEL | V157_R158del; K164fs; H178fs;D184fs | |
| 2 | exon6 | SNV | R213L/X | |
| G187fs; L188fs; P191_Q192del; | ||||
| 8 | exon6 | DEL | R196_L201del; D207fs; R209fs; F212fs | |
| Y234C; Y236C; M237I; R248G/Q; | ||||
| 10 | exon7 | SNV/MNV | R249S; T256P | |
| 3 | exon7 | INS | S241dup; R249fs; T253dup | |
| S241fs; M243X; G244fs; M246X; | ||||
| 6 | exon7 | DEL | I255del; L257fs | |
| イ | exon8 | SNV/MNV | V272K; C275X; R282W; T284K | |
| 4 | exon8 | INS | C275fs; N288fs; G302fs | |
| N263 N268del; N263fs; R267fs; P278fs; | ||||
| 5 | exon8 | DEL | P301fs | |
| б | exon10 | SNV | R337L; R342X; R337C | |
| l | exon10 | INS | L344fs |
Table 14. Mutations Represented in the Accuracy Summary Per Gene
{28}------------------------------------------------
{29}------------------------------------------------
Of the 433 specimens, 418 met the criteria of ≥200X coverage, 15 samples (3.5%) failed to achieve average coverage above 200X. The known mutation associated with each sample was successfully detected in 432 out of 433 cases (99.8% with two-sided 95% CI of (98.7%, 100.0%)). One discordant case was observed in sample M-1994-BC-T, which was used for the validation of insertions in EGFR exon 20. The known mutation for this sample was a 12bp duplication which began in the intron 5' of EGFR exon 20, potentially creating an alternative splice site acceptor for the exon. This duplication event was detected by the indel calling pipeline but was incorrectly filtered out because of the calling algorithm. (The filtering algorithm was modified to improve the detection accuracy for such mutations.)
The MSK-IMPACT accuracy study included 159 unique SNV/MNVs from 20 genes (45 exons), 49 unique deletions from 6 genes (11 exons), and 39 unique insertions from 6 genes (10 exons). Performance was stratified by mutation type and gene for percent positive agreement (PPA) with 95% confidence interval (CI). Results are shown in Table 15A-C.2
2 Performance may be overestimated because specimens were selected based on the availability of results by the orthogonal methods (i.e., the specimen set may lack challenging specimens).
{30}------------------------------------------------
| Gene | Numberof exons | Number ofuniquemutations | Number ofsamples | PPA (95% CI) |
|---|---|---|---|---|
| AKT1 | 1 | 1 | 10 | 100.0% (69.2%, 100.0%) |
| ALK | 2 | 5 | 7 | 100.0% (59.0%, 100.0%) |
| BRAF | 2 | 13 | 30 | 100.0% (88.4%, 100.0%) |
| EGFR | 3 | 6 | 30 | 100.0% (88.4%, 100.0%) |
| ERBB2 | 3 | 12 | 17 | 100.0% (80.5%, 100.0%) |
| FGFR2 | 3 | 3 | 3 | 100.0% (29.2%, 100.0%) |
| FGFR3 | 3 | 3 | 8 | 100.0% (63.1%, 100.0%) |
| GNA11 | 1 | 1 | 7 | 100.0% (59.0%, 100.0%) |
| GNAQ | 1 | 2 | 5 | 100.0% (47.8%, 100.0%) |
| GNAS | 1 | 2 | 5 | 100.0% (47.8%, 100.0%) |
| HRAS | 2 | 7 | 8 | 100.0% (63.1%, 100.0%) |
| IDH1 | 1 | 3 | 8 | 100.0% (63.1%, 100.0%) |
| IDH2 | 1 | 4 | 6 | 100.0% (54.1%, 100.0%) |
| KIT | 3 | 13 | 20 | 100.0% (83.2%, 100.0%) |
| KRAS | 3 | 15 | 39 | 100.0% (91.0%, 100.0%) |
| MET | 1 | 9 | 13 | 100.0% (75.3%, 100.0%) |
| NRAS | 2 | 6 | 16 | 100.0% (79.4%, 100.0%) |
| PDGFRA | 2 | 3 | 13 | 100.0% (75.3%, 100.0%) |
| PIK3CA | 4 | 19 | 32 | 100.0% (89.1%, 100.0%) |
| TP53 | 6 | 32 | 37 | 100.0% (90.5%, 100.0%) |
Table 15A.Percent Positive Agreement for SNV/MNVs by Gene
Table 15B. Percent Positive Agreement for insertions by gene
| Gene | Numberof exons | Number ofuniquemutations | Number ofsamples | PPA (95% CI) |
|---|---|---|---|---|
| EGFR | 12 | 16 | 93.8% (69.8%, 100.0%) | |
| ERBB2 | oc | 16 | 100.0% (79.4%, 100.0%) | |
| FGFR3 | 1 | 100.0% (2.5%, 100.0%) | ||
| KIT | ಗ | 10 | 100.0% (69.2%, 100.0%) | |
| PIK3CA | 1 | 100.0% (2.5%, 100.0%) | ||
| TP53 | ર | 14 | 14 | 100.0% (76.8%, 100.0% |
| Table 15C. Percent Positive Agreement for deletions by gene | ||||
|---|---|---|---|---|
| Gene | Number ofexons | No. uniquemutations | Number ofsamples | PPA (95% CI) |
|---|---|---|---|---|
| EGFR | 1 | 6 | 12 | 100.0% (73.5%, 100.0%) |
| IDH2 | 1 | 1 | 1 | 100.0% (2.5%, 100.0%) |
| KIT | 1 | 7 | 9 | 100.0% (66.4%, 100.0%) |
| MET | 1 | 18 | 19 | 100.0% (82.4%, 100.0%) |
| PIK3CA | 2 | 3 | 3 | 100.0% (29.2%, 100.0%) |
| TP53 | 5 | 14 | 14 | 100.0% (76.8%, 100.0%) |
- ii. Supplemental Method Comparison Study for Wildtype Calls: A supplemental study was conducted to assess accuracy for 33 "hotspots" within 10 genes. A total of 95 specimens were tested and the accuracy of
{31}------------------------------------------------
MSK-IMPACT results at all 33 positions was compared to results obtained with a single orthogonal method. Within the 95 specimens, there were 109 mutations across samples and 3026 wild-type calls. Variant-level concordance (PPA and NPA) was 100% for all results with two-sided 95% confidence intervals of (96.7%, 100.0%) for mutations (PPA) and (99.9%, 100.0%) for wild-type locations (NPA).
iii. Method Comparison of the MSK-IMPACT MSIsensor:
The somatic MSI status is inferred by interrogating all available genomic microsatellites covered by MSK-IMPACT within tumor samples against the matched normal DNA using the MSIsensor program as described in the Device Description section above. An MSIsensor score assigned to each tumor sample is used to distinguish MSS from MSI-H by MSIsensor.
The cutoff was first established using a training specimen dataset consisting of 138 colorectal cancer (CRC) and 40 endometrial carcinoma (EC) specimens with matched normal and having MSI status results from a validated MSI-PCR or MMR IHC test. MSIsensor scores ranged from 0 to 47.7 for CRC and 0 to 43.7 for EC. Based on concordance to either mismatch repair immunohistochemistry (MMR IHC) for MLH1, MSH2, MSH6 and PMS2 expression, or a commercially available PCR assay that detects 5 mononucleotide microsatellite loci including MR-21, BAT-25, MONO-27, NR-24 and BAT-26, a MSIsensor cut-off of 10 was established to delineate microsatellite stable (MMS) from high microsatellite instability (MSI-H).
A separate data set was obtained to validate this cut-off. A retrospectiveprospective chart review of 135 CRC patients was conducted to identify cases that had both MSK-IMPACT MSI results and results by a validated IHC panel (MLH1, MSH2, MSH6 and PMS2). A total of 66 specimens had both sets of results. Of these, there were two discordant cases. The estimated positive predictive value (PPV) was 92.3% (12/13) with two-sided 95% confidence interval of 64.0%-99.8% and the estimated negative predictive value (NPV) was 98.1%. (52/53) with two-sided 95% confidence interval of 90.0%, 100.0%. The results are shown in Table 16 below.
| CRC/EC Concordance with IHC | MMR-D* | MMR-P* | Total | |
|---|---|---|---|---|
| MSI Sensor | MSI-H ≥ 10 | 12 | 1 | 13 |
| MSS < 10 | 1 | 52 | 53 | |
| Total | 13 | 53 | 66 | |
| PPV = 92.3% (12/13) 95% CI 64.0%-99.8% | ||||
| NPV = 98.1%. (52/53) 95% CI 90.0%, 100.0% |
Table 16. MSIsensor Results Compared to IHC MMR for CRC
*MMR-D refers to deficient in mismatch repair proteins and MMR-P indicates not deficient
{32}------------------------------------------------
To evaluate the ability of the MSIsensor to determine MSI status in cancer types other than CRC or EC cancer types, 119 unique non-CRC and non-EC tumornormal pair samples covering 25 tumor types were assessed for MSI by both MSIsensor and a validated MSI-PCR test. The results are shown in Table 17. Excluding the specimens without a MSI-PCR result from the total number of specimens analyzed, PPV is 46/49=93.9% (83.1%, 98.7%), and NPV is 58/60=96.7% (88.5%, 99.6%). When including all missing data in the analysis (i.e., consider all PCR unknown data as discordant results), the PPV=46/59=78.0% (65.3%. 87.7%). NPV= 58/60=96.7% (88.5%. 99.6%). (The MSIsensor MSI-H/MSS definition is based on genome wide analysis of over 1000 microsatellite markers and not based on the 5 or 7 MSI loci described in current clinical practice guidelines.)
| Non CRC/EC concordance with MSI-PCR | PCR Results | ||||
|---|---|---|---|---|---|
| MSI-H | MSI-L/MSS | Unknown* | Total | ||
| MSIsensor | MSI-H(≥10) | 46 | 3 | 10 | 59 |
| MSS(≥2 & <10) | 2 | 58 | 0 | 60 | |
| Total | 48 | 61 | 10 | 119 | |
| Excluding missingspecimens with 95%CI | PPV is 46/49=93.9% 95%CI (83.1%, 98.7%) | ||||
| NPV is 58/60=96.7% (88.5%, 99.6%) | |||||
| Accounting formissing specimens with95%CI | PPV=46/59=78.0% (65.3%, 87.7%) | ||||
| NPV= 58/60=96.7% (88.5%, 99.6%) |
Table 17. MSIsensor Results Compared to PCR 5 Loci MSI Panel for Other Cancer Types
- In exploratory analysis, the 10 without PCR results were all MMR-D by IHC, consistent with the MSI-H by MSIsensor findings.
MSI Supplemental Information:
The mean, median and range of MSIsensor score was determined in a large cohort of 10,900 patients with 66 different types advanced solid tumor. The MSIsensor scores ranged from 0 to 48.5, mean 1.2, median 0.4. The prevalence of MSI-H by MSIsensor was also determined, and the findings are consistent with the MSI-H prevalence as described in the literature (data not shown).
3. Clinical Performance:
MSK-IMPACT assay is a molecular profiling platform using next generation sequencing to detect somatic alterations (point mutations and small insertions and deletions and microsatellite instability) in tumor specimens using a 468 gene panel. The genes in the panel were selected for their role in cancer pathogenesis and tumor
{33}------------------------------------------------
suppression, or for clinical or mechanistic information of relevance in the management of cancer patients. The assay reports mutations under two categories: "Cancer mutations with evidence of clinical significance" and "Cancer mutations with potential clinical significance" consistent with the intended use clinical settings. Mutations with evidence of clinical significance are represented in professional guidelines as established by consensus opinion of experts in the health care community.
Clinical Evidence Curation:
MSK-IMPACT uses a clinical evidence curation resource (OncoKB) to facilitate the clinical interpretation of detected mutations. OncoKB is a knowledge base that includes biologic, clinical and therapeutic information curated from multiple information resources including professional guidelines and recommendations, therapeutic labeling, disease specific expert and advocacy group recommendations, and medical literature. OncoKB information is publicly available through an interactive web site. Classification criteria were developed by MSK to communicate the level of clinical evidence available for individual mutations in the test report. The mutations are reported under two categories (i.e., cancer mutations panel with evidence of clinical significance and cancer mutations panel with potential clinical significance) based on the pre-specified classification criteria. OncoKB undergoes periodic updates through the review of new information by a panel of experts.
4. Clinical cut-off:
Not applicable.
5. Expected values:
The prevalence of somatic mutations was explored through a large-scale, prospective clinical sequencing initiative using a comprehensive assay, MSK-IMPACT, through which tumor and matched normal sequence data from a cohort of more than 10,000 patients with advanced cancer and available pathological and clinical annotations was compiled. The prevalence of mutations and cancer type via the link to the publicly accessible data on cohort of tested patients and available pathological and clinical annotations was published by Zehir, A. et al., "Mutational landscape of metastatic cancer revealed from prospective clinical sequencing of 10,000 patients." 2017. 23(6):703-713. This information is also available at the following website. (http://www.cbioportal.org/study?id=msk impact 2017#summary)
N. Instrument Name
Illumina HiSeq 2500 (qualified by MSK)
O. System Descriptions:
1. Modes of Operation:
The Illumina HiSeq2500 is a high throughput sequencing system using Sequencing-By-Synthesis chemistry.
{34}------------------------------------------------
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes X or No
-
- Level of Concern: Moderate
-
- Specimen Handling Refer to Device Description section above.
- 5 Calibration and Quality Controls: Refer to Device Description section above.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section Above:
To support the continuous implementation of process improvements to the existing 468 gene panel, protocols with specific procedures and acceptance criteria for modifications that could be anticipated at the time of submission were provided, reviewed by FDA, and cleared as part of this marketing authorization. Future modifications by MSK for the specified types of changes below that are made in accordance with the applicable validation strategy and the pre-specified success criteria would not require a new 510(k) submission. Significant changes such as adding new genes or variant types to the panel would require a new submission with appropriate validation.
| Type of change | Validation Strategy | Pre-specified success criteria | |
|---|---|---|---|
| New pre-analytical protocol, kits orreagents | Sequence at least 10specimens withknown mutations.Measure sequencecoveragedistribution, and callsomatic mutations inall samples. | For cases sequenced to>200x, ensure that 95% ofexons are covered to 100xor more. Concordance forknown mutations shouldbe >95%. | |
| New library preparation protocol, kits,or reagents | Sequence at least 40DNA specimens(tumor / normalpairs) or three poolspreviouslysequenced by MSK-IMPACT. Measuresequence coveragedistribution, and call | For cases sequenced to>200x, ensure that 95% ofexons are covered to 100xor more. Concordance forcalling somatic mutationswith variant allele fraction>10% should be >98%. | |
| Type of change | Validation Strategy | Pre-specified successcriteria | |
| somatic mutations inall samples. | |||
| Changes to probes for alreadyanalytically validated genes | Re-capture existingsequence librariesfrom at least 3 runs(at least 40 samples)with new probes,sequence, andanalyze. | For cases sequenced to>200x, ensure that 95% ofexons in analyticallyvalidated genes arecovered to 100x or more.Concordance for callingsomatic mutations withvariant allele fraction>10% should be >98%. | |
| New sequencing instrument or reagentsusing similar chemistry and technology,and the sequence depth and read lengthare not changed from previousplatform. | Re-sequenceexisting capturedlibraries from atleast 3 runs, and callsomatic mutations inall samples. | Sequence coveragedistribution and GC biasacross targeted regionsshould be within 5% ofprior sequencing runs.Concordance for callingsomatic mutations withvariant allele fraction>10% should be >98%. | |
| Bioinformaticspipeline | Update tounderlyingannotationdatabase ortranscriptisoforms | Reanalyze FASTQfiles (rawsequencing reads)from at least 3 runs(at least 40 samples).Compare variantscalls between theclinical analysisresults and thecurrent modifiedresults | Confirm the changes donot change the variant callresults. Confirm theannotations for theunaffected transcripts donot change. Confirm theannotations for theaffected transcripts aremodified as expected. |
| Update to datamanagementsystem andsystem database | Reanalyze FASTQfiles (rawsequencing reads)from at least 3 runs(at least 40 samples)in production mode.Compare variantscalls between theclinical analysisresults and thecurrent modifiedresults | Ensure that all previouslycalled mutations arerecovered and the variantsin the database of resultsare concordant with thevariants in the pipelineoutput files | |
| Type of change | Validation Strategy | Pre-specified success criteria | |
| Modification toan existingcomponent ofthe analysispipeline (e.g.,tool oralgorithm)where theunderlyingalgorithm ormain parametersettings (e.g.minimalcoverage/VAFthreshold forSNV/indelcalling;MSIsensorscore cut-off forMSI-H calling,etc.) are notchanged. | Reanalyze FASTQfiles (rawsequencing reads)from at least 3 runs(at least 40 samples).Compare variantscalls between theclinical analysisresults and thecurrent modifiedresults | Ensure that all previouslycalled mutations arerecovered and that newlydetected mutations can beexplained by pipelinemodifications. |
{35}------------------------------------------------
{36}------------------------------------------------
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable, and the special controls for this device type.
R. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
S. Identified Risks to Health and Identified Mitigations:
| Identified Risks to Health | Identified Mitigations |
|---|---|
| Incorrect performance of the testleading to false positives, falsenegatives | General controls and special control (b)(1) |
| Incorrect interpretation of test results | General controls and special controls(b)(1)(iii)(E) and (b)(2) |
{37}------------------------------------------------
| 1. Benefit/Risk Determination | |
|---|---|
| Summary of the Benefit(s) | The MSK-IMPACT (Integrated Mutation Profiling of Actionable Cancer Targets) test provides comprehensive genomic profiling of tumor samples (point mutations, small insertions and deletions and microsatellite instability), in previously diagnosed cancer patients, for use by qualified health professionals in accordance with professional guidelines. There is probable clinical benefit of the device based on evidence from peer-reviewed clinical literature and analytical performance of the device in identifying genomic alterations. |
| Summary of the Risk(s) | Erroneous device results could adversely influence clinical interpretation and consultation for patients. The risk of an erroneous test result is mitigated by the analytical performance of this device. The accuracy of the test was demonstrated using clinical specimens covering a variety of clinically relevant variants across multiple tumor types and variant categories (i.e., point mutations, small insertions and deletions and microsatellite instability). The output of this device demonstrated a high degree of analytical concordance to comparator assays across multiple tumor types. Thus, the probable risk of this device is mitigated by the supportive analytical performance for the device, when clinical limitations and the established special controls, in combination with general controls, are considered. |
| Summary of Other Factors | Limitations statements in the test report and the established special controls, in combination with general controls, serve to mitigate the risks associated with the use of this device. |
| ConclusionsDo the probable benefits outweigh the probable risks? | The probable clinical benefits of this device, which allows for detection of somatic mutations and MSI status in patients previously diagnosed with cancer, outweigh the probable risks that are mitigated by the special controls established for this device type, in combination with general controls. |
T. Benefit/Risk Determination
U. Conclusion:
The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.6080. FDA believes that special controls, along with the applicable general controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:
| Product Code: | PZM |
|---|---|
| Device Type: | Next Generation Sequencing Based Tumor Profiling Test |
| Class: | II (special controls) |
| Regulation: | 21 CFR 866.6080 |
{38}------------------------------------------------
(a) Identification. A next generation sequencing (NGS) based tumor profiling test is a qualitative in vitro diagnostic test intended for NGS analysis of tissue specimens from malignant solid neoplasms to detect somatic mutations in a broad panel of targeted genes to aid in the management of previously diagnosed cancer patients by qualified health care professionals.
(b) Classification. Class II (special controls). A next generation sequencing based tumor profiling test must comply with the following special controls:
(1) Premarket notification submissions must include the following information:
(i) A detailed description of all somatic mutations that are intended to be detected by the test and that are adequately supported in accordance with paragraph (b)(1)(v) of this section and reported in the test results in accordance with paragraph (b)(2)(iv) of this section, including:
(A) A listing of mutations that are cancer mutations with evidence of clinical significance.
(B) As appropriate, a listing of mutations that are cancer mutations with potential clinical significance.
(ii) The indications for use must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (e.g., formalin-fixed, paraffinembedded tumor tissue).
(C) The mutation types (e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.
(D) The name of the testing facility or facilities, as applicable.
(iii) A detailed device description including the following:
(A) A description of the test in terms of genomic coverage, as follows:
( /) Tabulated summary of all mutations reported, grouped according to gene and target region within each gene, along with the specific cDNA and amino acid positions for each mutation.
(2) A description of any within-gene targeted regions that cannot be reported and the data behind such conclusion.
{39}------------------------------------------------
(B) Specifications for specimen requirements including any specimen collection devices and preservatives, specimen volume, minimum tumor content, specimen handling, DNA extraction, and criteria for DNA quality and quantity metrics that are prerequisite to performing the assay.
(C) A detailed description of all test components, reagents, instrumentation, and software required. Detailed documentation of the device software including but not limited to, software applications and hardware-based devices that incorporate software.
(D) A detailed description of the methodology and protocols for each step of the test, including description of the quality metrics, thresholds, and filters at each step of the test that are implemented for final result reporting and a description of the metrics for run-failures, specimen-failures, invalids, as applicable.
(E) A list of links provided by the device to the user or accessed by the device for internal or external information (e.g., decision rules or databases) supporting clinical significance of test results for the panel or its elements in accordance with paragraphs (b)(1)(v) and (b)(2)(vi) of this section.
(F) A description of internal and external controls that are recommended or provided and control procedures. The description must identify those control elements that are incorporated into the testing procedure.
(iv) Information demonstrating analytical validity of the device according to analytical performance characteristics, evaluated either specifically for each gene/mutation or, when clinically and practically justified, using a representative approach based on other mutations of the same type, including:
(A) Data that adequately supports the intended specimen type (e.g., formalinfixed, paraffin-embedded tumor tissue), specimen handling protocol, and nucleic acid purification for specific tumor types or for a pan-tumor claim.
(B) A summary of the empirical evidence obtained to demonstrate how the analytical quality metrics and thresholds were optimized.
(C) Device precision data using clinical samples to adequately evaluate intra-run. inter-run, and total variability. The samples must cover all mutation types tested (both positive and negative samples) and include samples near the limit of detection of the device. Precision must be assessed by agreement within replicates on the assay final result for each representative mutation, as applicable, and also supported by sequencing quality metrics for targeted regions across the panel.
(D) Description of the protocols and/or data adequately demonstrating the interchangeability of reagent lots and multiplexing barcodes.
(E) A description of the nucleic acid assay input concentration range and the
{40}------------------------------------------------
evidence to adequately support the range.
(F) A description of the data adequately supporting the limit of detection of the device
(G) A description of the data to adequately support device accuracy using clinical specimens representing the intended specimen type and range of tumor types, as applicable.
(1) Clinical specimens tested to support device accuracy must adequately represent the list of cancer mutations with evidence of clinical significance to be detected by the device.
(2) For mutations that are designated as cancer mutations with evidence of clinical significance and that are based on evidence established in the intended specimen type (e.g., tumor tissues) but for a different analyte type (e.g., protein, RNA) and/or a measurement (e.g., incorporating a score or copy number) and/or with an alternative technology (e.g., IHC, RT-qPCR, FISH), evidence of accuracy must include clinically adequate concordance between results for the mutation and the medically established biomarker test (e.g., evidence generated from an appropriately sized method comparison study using clinical specimens from the target population).
(3) For qualitative DNA mutations not described in paragraph (b)(1)(iv)(G)(2) of this section, accuracy studies must include both mutation-positive and wild-type results.
(H) Adequate device stability information.
(v) Information that adequately supports the clinical significance of the panel must include:
(A) Criteria established on what types and levels of evidence will clinically validate a mutation as a cancer mutation with evidence of clinical significance versus a cancer mutation with potential clinical significance.
(B) For representative mutations of those designated as cancer mutations with evidence of clinical significance, a description of the clinical evidence associated with such mutations, such as clinical evidence presented in professional guidelines, as appropriate, with method comparison performance data as described in paragraph (b)(1)(iv)(G) of this section.
(C) For all other mutations designated as cancer mutations with potential clinical significance, a description of the rationale for reporting.
(2) The 21 CFR 809.10 compliant labeling and any product information and test report generated, must include the following, as applicable:
{41}------------------------------------------------
(i) The intended use statement must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (e.g., formalin-fixed, paraffinembedded tumor tissue).
(C) The mutation types (e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.
(D) The name of the testing facility or facilities, as applicable.
(ii) A description of the device and summary of the results of the performance studies performed in accordance with paragraphs (b)(1)(ii), (b)(1)(iv), and (b)(1)(v) of this section.
(iii) A description of applicable test limitations, including, for device specific mutations validated with method comparison data to a medically established test in the same intended specimen type, appropriate description of the level of evidence and/or the differences between next generation sequencing results from the medically established test (e.g., as described in professional guidelines).
(iv) A listing of all somatic mutations that are intended to be detected by the device and that are reported in the test results under the following two categories or equivalent designations, as appropriate: "cancer mutations panel with evidence of clinical significance" or "cancer mutations panel with potential clinical significance."
(v) For mutations reported under the category of "cancer mutations panel with potential clinical significance," a limiting statement that states "For the mutations listed in [cancer mutations panel with potential clinical significance or equivalent designation], the clinical significance has not been demonstrated [with adequate clinical evidence (e.g., by professional guidelines) in accordance with paragraph (b)(1)(v) of this section] or with this test."
(vi) For mutations under the category of "cancer mutations panel with evidence of clinical significance," or equivalent designation, link(s) for physicians to access internal or external information concerning decision rules or conclusions about the level of evidence for clinical significance that is associated with the marker in accordance with paragraph (b)(1)(v) of this section.
{42}------------------------------------------------
Appendix 1a:
List of hotspot mutations (i.e., commonly somatically mutated in cancers) for all genes in the MSK-IMPACT panel
| Gene | Codons |
|---|---|
| ABL1 | G250, Q252, Y253, E255, T315, F317, M351, F359, H396R |
| AKT1 | E17,Q124,G171,E170 |
| AKT2 | V140 |
| ALK | K1062,D1091,C1156,M1166,I1171,F1174,L1196,A1234,F1245,I1250,R1275,Y1278 |
| APC | S1234,I1307,E1309,E1317,P1319,G1339,S1341,P1361,P1372,P1373,R1399,S1400,S1407,S1411,V1414,S1415,S1421,T1438,P1439,P1440,T1445,P1453,N1455,E1464,S1465,T1487,L1488,F1491,T1493,E1494,T1537,K1555,T1556,I1557,C1578 |
| AR | T878,T8782,Q581 |
| ARAF | S214 |
| ARID1A | D1850,G2087 |
| ARID2 | R314,S297,R285,A1773 |
| ASXL1 | Y591,E635,G645,G646,E1102D |
| ASXL2 | R591 |
| ATM | D1853,R3008,R3376,E2164 |
| ATRX | K1936,E625 |
| BARD1 | P24 |
| BCL6 | R594,R618 |
| BCOR | N1425,N14591 |
| BRAF | G464,G466,G469,Y472,N581,D594,F595,G596,L597,A598_T599,V600,V600_K601,K6010,V60010,K6010,G4694,N5810,G4660 |
| CARD11 | R170 |
| CBL | Y371,L380,C384,C404,R420Q |
| CDH1 | T263 |
| CDK4 | R24 |
| CDKN2A | S43,P48,A57,A68,D74,L78,P81,H83,D84,L97,D108,P114,H831,D1081,P1140 |
| CEBPA | P23,H24,Q83,K304_Q305,E309_T310,Q312_K313,K313_V314,K313_V314,K313,E316_L317,E316_L317insQ |
| CHEK2 | K373,K3732 |
| CIC | R215 |
| CREBBP | R1446,S1680,R14460 |
| CRLF2 | F232C |
| CSF1R | Y969C |
| CTCF | R377 |
| CTNNB1 | D32,S33,G34,I35,H36,S37,T40,T41,T42,A43,P44,S45,G48,K49,E53,K335,S376,S336,D324,T412,G349,S455,C619 |
| DICER1 | E1813 |
| DIS3 | R382,D488 |
| DNMT1 | E432 |
| DNMT3A | G543,R635,S714,F731,R882,R8820 |
| DOT1L | G1386 |
| EGFR | R108,A289,G598,R677,E709,G719,K745_E749,K745_E746,E746_A750,E746_S752,E746_T751,E746_E749,E746_T751,L747_P753,L747_A750,L747_T751,L747_S752,L747_T751,L747_E749,L747,T751,S752_I759,D761,S768,V769_D770,D770_N771,H773_V774,R776,T790,L833,H835,T847,P848,T854,L858,L861,G863,L8587,A2898,R252,R222 |
| EP300 | D1399,D13990,C1164 |
| EPHB1 | R170 |
| ERBB2 | S310,L755,D769,A775_G776,G776,V777,V842,S3108,L7553,E930,R678 |
| ERBB3 | V1043,D297,M91 |
| ERBB4 | R711 |
| ERCC2 | D312 |
| ESR1 | Y537 |
| ETV1 | R187 |
| ETV6 | R369 |
| EZH2 | Y646,R690 |
| FBXW7 | G423,R465,R479,R505,S582,R689,R4652,R5054,R4792 |
| FGFR2 | S252,P253,C382,N549,N550,K659 |
| FGFR3 | R248,S249,G370,S371,Y373,G380,A391,K650,G697,S2492,Y3730 |
| FGFR4 | V550 |
| FLT3 | D835,I836,D8358 |
| FOXL2 | C134W |
| FUBP1 | R430 |
| GATA1 | M1,S30,V74I |
| GATA2 | G320,L321,L359,R362Q |
| GNA11 | R183,Q209,R256 |
| GNAQ | R183,Q209 |
| GNAS | R201,Q227,R8448 |
| GRIN2A | R1067 |
| HIST1H3E | E74 |
| HNF1A | W206,P291,G292 |
| HRAS | G12,G13,Q61,E62,Q614,G136,G122 |
| IDH1 | G70,V71,R132,V178,R13239,P33 |
| IDH2 | R140,R172,V294,R1402,R1721 |
| IL7R | K395 |
| IRS2 | G1057 |
| JAK1 | R873 |
| JAK2 | F537_K539,H538_K539,K539,I540_E543,R541_E543,N542_E543,E543_D544,V617 |
| JAK2 | R683 |
| JAK3 | A572,A573,R657Q |
| KDR | S1100,E759 |
| KEAP1 | R470 |
| KIT | D52,D419,Y503_F504,K509,M541,K550_K558,P551_V555,P551_E554,P551_M552,Y553_K558,E554_K558,Q556_V560,W557_K558,W557,W557_V559,W557_E561,W557_V559,K558_E562,K558,K558_V560,V559,V559_V560,V559_E561,V560,E56Y570_L576,D572,L576,D579,K642,V654,T670,S715,D816,K818,D820,N822,Y823,V25, D8160 |
| KMT2C | V656 |
| KRAS | G10_A11,G12,G13,V14,L19,Q22,T58,A59,Q61,K117,A146,G1242,G133,Q619,A146 |
| LATS2 | A3243,G3630 |
| MAP2K1 | Q56,K57,D67,P124,P1240,F53,E203 |
| MAP2K4 | R134 |
| MAP3K1 | S1330,S939 |
| MAPK1 | E322 |
| MAX | R600 |
| MED12 | L36,Q43,G44,L1224,L12240 |
| MEF2B | D83V |
| MET | T1010,Y1248,Y1253,M1268,K1360 |
| MLL3 | K2797 |
| MPL | S505,W515,W515R |
| MSH6 | F1088,T1219I |
| MTOR | S22152,F1888 |
| MYC | T58 |
| MYCN | P44 |
| MYD88 | S219,S243,L265P |
| NF1 | L844 |
| NFE2L2 | D29,L30,G31,R34,E79,T80,G81,E82,E794,D294,R342 |
| NOTCH1 | L1574,L1575,V1578,L1585,L1586,F1592,L1593,L1594,R1598,R1599,L1600,L1601,L1678,L1679,Q2460,P2514,A1944 |
| NOTCH2 | E385,N463 |
| NPM1 | W288,W290 |
| NRAS | G12,G13,A18,G60,Q61,Q6193,G128,G138 |
| NTRK1 | T264 |
| PAK7 | E144 |
| PARP1 | I562 |
| PAX5 | P80R |
| PDGFRA | V561,S566_E571,N659,D842,I843_D846,D1071N |
| PIK3C2G | S670 |
| PIK3CA | R38,E81,R88,R93,G106,R108,K111,G118,V344,N345,C378,E418,C420,E453,P539,E542,E545,Q546,E547,S553,K567,H701,E726,C901,G1007,Y1021,T1025,M1043,N1044,D1045,A1046,H1047,G1049,T1052,A1066,N1068,E54534,H104715,E54217,Q5467,R887,N3453,C4209,G1187,E7265,E4535, K1113, R932, R382, R1080,E39 |
| PIK3R1 | G376,D560,N564,K567 |
| POLE | P2864,V4111 |
| PPP2R1A | P179,R182,R183,S256,W257,R258,R1832 |
| PREX2 | G233C |
| PTCH1 | P1315 |
| PTEN | K6,P38,L42,H61,Y68,Y76,Y88,H93,I101,C105,L112,H123,A126,G129,R130,C136,A151,Y155,R159,K164,G165,S170,R173,N184,E242,P246,P248,C250,K267,V290,L318,T319,T321,N323,F347,R1309,R1730,K128 |
| PTPN11 | G60,D61,E69,A72,T73,E76,S502,G503,Q510 |
| PTPRD | S431,P666 |
| RAC1 | P295 |
| RAF1 | S2570 |
| RET | E632_T636,E632_L633,C634,M918T |
| RHOA | E40,Y42 |
| RICTOR | S1101 |
| RIT1 | M90 |
| RUNX1 | L56,R107,D198,R201,R204,R162,R205 |
| SDHA | S4560,A466,R465 |
| SF3B1 | E622,R625,H662,K666,K700,K7002 |
| SMAD4 | A118,D351,R361,G386,R3619,D537,P356 |
| SMARCA | T910,G1232 |
| SMARCB | R377,A382,P383 |
| SMO | W535L |
| SPOP | F133,F1338,W131,F102 |
| SRSF2 | P95,P95_R102,P107H |
| STAG2 | R370 |
| STK11 | D194,P281,F354L |
| TET2 | C25,C262,Q764,F868,R1261,H1380,V1718L |
| TNFAIP3 | L324 |
| TP53 | E11,D49,P82,T102,G105,Y107,R110,L111,F113,K120,T125,Y126,Y126_K132,S127,P128,L130,N131,K132,M133,F134,C135,A138,K139,T140,C141,P142,V143,Q144,I145,V147,S149,P151,R152,P153,G154,T155,R156,V157,R158,A159,M160,A161 |
| I162,Y163,K164,S166,H168,M169,T170,E171,V172,V173,R174,R175,C176,P177, | |
| P177_C182,H178,H179,E180,R181,C182,D184,D186,G187,P190,P191,Q192,H193, | |
| L194,I195,R196,V197,E198,G199,N200,R202,V203,Y205,D208,R209,T211,F212,R21 | |
| ,S215,V216,V217,V218,Y220,E224,G226,S227,D228,C229,T230,I232,Y234,N235,Y2 | |
| 6, M237,C238,N239,S240,S241,C242,M243,G244,G245,M246,N247,R248,R249,P250 | |
| I251,L252,T253,I254,I255,L257,E258,D259,G262,L265,G266,R267,F270,E271,V272, | |
| R273,V274,C275,A276,C277,P278,G279,R280,D281,R282,R283,T284,E285,E286,E28 | |
| 7, N288,R290,K291,K292,E294,P300,P301,S303,K320,G334,R337,R27328,R24892, | |
| R17538,R2820,G2451,Y2202,H1938,H1797,R1583,C1763,P2783,Y1633,R2800, | |
| G2660,I1950,S2419,R2499,V1577,C2386,E2856,R3375,G2445,V1733,P1512,C2752, | |
| K1321,Y2050,V2720,C1359,D2818,E2718,V2168,M2378,Y2347,E2867,L1946, | |
| A1596,R2675,S1275,C2425,Y2364,C1414,F2704,A1613,V2743,S2153,R2132,H2142, | |
| R1101,N2390,T1550,P1520,P2500,G1050,L1300,Q136,F109 | |
| TP63 | R379 |
| TSC2 | N1515 |
| TSHR | M453,I486,L512,I568,D619,A623,L629,I630,T632,D633,D633E |
| U2AF1 | S34,Q157,S347 |
| VHL | V62,S65,S72,V74,F76,N78,S80,P81,L85,P86,L89,N90,S111,G114,H115,L118,D121, |
| VHL | L128,V130,G144,F148,I151,L153,V155,L158,E160,C162,V166,R167,L169,L184 |
| WT1 | V303,R312,A314,R394,D396,R462 |
| XP01 | E571,R749 |
{43}------------------------------------------------
{44}------------------------------------------------
{45}------------------------------------------------
{46}------------------------------------------------
{47}------------------------------------------------
| Gene Name | Transcript ID | ||
|---|---|---|---|
| ABL1 | NM_005157 | ||
| ACVR1 | NM_001111067 | ||
| AGO2 | NM_012154 | ||
| AKT1 | NM_001014431 | ||
| AKT2 | NM_001626 | ||
| AKT3 | NM_005465 | ||
| ALK | NM_004304 | ||
| ALOX12B | NM_001139 | ||
| AMER1 | NM_152424 | ||
| ANKRD11 | NM_013275 | ||
| APC | NM_000038 | ||
| AR | NM_000044 | ||
| ARAF | NM_001654 | ||
| ARID1A | NM_006015 | ||
| ARID1B | NM_020732 | ||
| ARID2 | NM_152641 | ||
| ARID5B | NM_032199 | ||
| ASXL1 | NM_015338 | ||
| ASXL2 | NM_018263 | ||
| ATM | NM_000051 | ||
| ATR | NM_001184 | ||
| ATRX | NM_000489 | ||
| AURKA | NM_003600 | ||
| AURKB | NM_004217 | ||
| AXIN1 | NM_003502 | ||
| AXIN2 | NM_004655 | ||
| AXL | NM_021913 | ||
| B2M | NM_004048 | ||
| BABAM1 | NM_001033549 | ||
| BAP1 | NM_004656 | ||
| BARD1 | NM_000465 | ||
| BBC3 | NM_001127240 | ||
| BCL10 | NM_003921 | ||
| BCL2 | NM_000633 | ||
| BCL2L1 | NM_138578 | ||
| BCL2L11 | NM_138621 | ||
| BCL6 | NM_001706 | ||
| BCOR | NM_001123385 | ||
| BIRC3 | NM_182962 | ||
| BLM | NM_000057 | ||
| BMPR1A | NM_004329 | ||
| BRAF | NM_004333 | ||
| BRCA1 | NM_007294 | ||
| BRCA2 | NM_000059 | ||
| BRD4 | NM_058243 | ||
| BRIP1 | NM_032043 | ||
| BTK | NM_000061 | ||
| CALR | NM_004343 | ||
| CARD11 | NM_032415 | ||
| CARM1 | NM_199141 | ||
| CASP8 | NM_001080125 | ||
| CBFB | NM_022845 | ||
| CBL | NM_005188 | ||
| CCND1 | NM_053056 | ||
| CCND2 | NM_001759 | ||
| CCND3 | NM_001760 | ||
| CCNE1 | NM_001238 | ||
| CD274 | CD276 | NM_014143 | NM_001024736 |
| CD79A | NM_001783 | ||
| CD79B | NM_001039933 | ||
| CDC42 | NM_001791 | ||
| CDC73 | NM_024529 |
| Appendix 1b: List of genes/transcripts included on the MSK-IMPACT panel | |
|---|---|
| ------------------------------------------------------------------------- | -- |
{48}------------------------------------------------
| CDH1 | NM_004360 | ||
|---|---|---|---|
| CDK12 | NM_016507 | ||
| CDK4 | NM_000075 | ||
| CDK6 | NM_001145306 | ||
| CDK8 | NM_001260 | ||
| CDKN1A | NM_078467 | ||
| CDKN1B | NM_004064 | ||
| CDKN2Ap14ARF | NM_058195 | ||
| CDKN2Ap16INK4A | NM_000077 | ||
| CDKN2B | NM_004936 | ||
| CDKN2C | NM_078626 | ||
| CEBPA | NM_004364 | ||
| CENPA | NM_001809 | ||
| CHEK1 | NM_001274 | ||
| CHEK2 | NM_007194 | ||
| CIC | NM_015125 | ||
| CREBBP | NM_004380 | ||
| CRKL | NM_005207 | ||
| CRLF2 | NM_022148 | ||
| CSDE1 | NM_001242891 | ||
| CSF1R | NM_005211 | ||
| CSF3R | NM_000760 | ||
| CTCF | NM_006565 | ||
| CTLA4 | NM_005214 | ||
| CTNNB1 | NM_001904 | ||
| CUL3 | NM_003590 | ||
| CXCR4 | NM_003467 | ||
| CYLD | NM_001042355 | ||
| CYSLTR2 | NM_020377 | ||
| DAXX | NM_001141970 | ||
| DCUN1D1 | NM_020640 | ||
| DDR2 | NM_006182 | ||
| DICER1 | NM_030621 | ||
| DIS3 | NM_014953 | ||
| DNAJB1 | NM_006145 | ||
| DNMT1 | NM_001379 | ||
| DNMT3A | NM_022552 | ||
| DNMT3B | NM_006892 | ||
| DOT1L | NM_032482 | ||
| DROSHA | NM_013235 | ||
| DUSP4 | NM_001394 | ||
| E2F3 | NM_001949 | ||
| EED | NM_003797 | ||
| EGFL7 | NM_201446 | ||
| EGFR | NM_005228 | ||
| EIF1AX | NM_001412 | ||
| EIF4A2 | NM_001967 | ||
| EIF4E | NM_001130678 | ||
| ELF3 | NM_004433 | ||
| EP300 | NM_001429 | ||
| EPAS1 | NM_001430 | ||
| EPCAM | NM_002354 | ||
| EPHA3 | NM_005233 | ||
| EPHA5 | NM_004439 | ||
| EPHA7 | NM_004440 | ||
| EPHB1 | NM_004441 | ||
| ERBB2 | NM_004448 | ||
| ERBB3 | NM_001982 | ||
| ERBB4 | NM_005235 | ||
| ERCC2 | NM_000400 | ||
| ERCC3 | ERCC3 | NM_000122 | NM_000122 |
| ERCC4 | NM_005236 | ||
| ERCC5 | NM_000123 | ||
| ERF | NM_006494 | ||
| ERG | NM_182918 | ||
| ERRFI1 | NM_018948 | ||
| ESR1 | NM_001122740 | ||
| ETV1 | NM_001163147 | ||
| ETV6 | NM_001987 | ||
| EZH1 | NM_001991 | ||
| EZH2 | NM_004456 | ||
| FAM175A | NM_139076 | ||
| FAM46C | NM_017709 | ||
| FAM58A | NM_152274 | ||
| FANCA | NM_000135 | ||
| FANCC | NM_000136 | ||
| FAT1 | NM_005245 | ||
| FBXW7 | NM_033632 | ||
| FGF19 | NM_005117 | ||
| FGF3 | NM_005247 | ||
| FGF4 | NM_002007 | ||
| FGFR1 | NM_001174067 | ||
| FGFR2 | NM_000141 | ||
| FGFR3 | NM_000142 | ||
| FGFR4 | NM_213647 | ||
| FH | NM_000143 | ||
| FLCN | NM_144997 | ||
| FLT1 | NM_002019 | ||
| FLT3 | NM_004119 | ||
| FLT4 | NM_182925 | ||
| FOXA1 | NM_004496 | ||
| FOXL2 | NM_023067 | ||
| FOXO1 | NM_002015 | ||
| FOXP1 | NM_001244814 | ||
| FUBP1 | NM_003902 | ||
| FYN | NM_153047 | ||
| GATA1 | NM_002049 | ||
| GATA2 | NM_032638 | ||
| GATA3 | NM_002051 | ||
| GLI1 | NM_005269 | ||
| GNA11 | NM_002067 | ||
| GNAQ | NM_002072 | ||
| GNAS | NM_000516 | ||
| GPS2 | NM_004489 | ||
| GREM1 | NM_013372 | ||
| GRIN2A | NM_001134407 | ||
| GSK3B | NM_002093 | ||
| H3F3A | NM_002107 | ||
| H3F3B | NM_005324 | ||
| H3F3C | NM_001013699 | ||
| HGF | NM_000601 | ||
| HIST1H1C | NM_005319 | ||
| HIST1H2BD | NM_021063 | ||
| HIST1H3A | NM_003529 | ||
| HIST1H3B | NM_003537 | ||
| HIST1H3C | NM_003531 | ||
| HIST1H3D | NM_003530 | ||
| HIST1H3E | NM_003532 | ||
| HIST1H3F | NM_021018 | ||
| HIST1H3G | NM_003534 | ||
| HIST1H3H | NM_003536 | ||
| HIST1H3I | NM_003533 | ||
| HIST1H3J | NM_003535 | ||
| HIST2H3C | NM_021059 | ||
| HIST2H3D | NM_001123375 | ||
| HIST3H3 | NM_003493 | ||
| HLA-A | NM_001242758 | ||
| HLA-B | HLA-B | NM_005514 | NM_005514 |
| HNF1A | NM_000545 | ||
| HOXB13 | NM_006361 | ||
| HRAS | NM_001130442 | ||
| ICOSLG | NM_015259 |
{49}------------------------------------------------
{50}------------------------------------------------
| ID3 | NM002167 |
|---|---|
| IDHI | 005896NM |
| IDH2 | 002168NM |
| IFNGR1 | 000416NM |
| IGF1 | 00111285NM |
| IGFIR | NM000875 |
| IGF2 | NM_001127598 |
| IKBKE | 014002NM |
| IKZF1 | NM006060 |
| IL10 | NM000572 |
| IL7R | NM002185 |
| INHA | NM002191 |
| INHBA | NM002192 |
| INPP4A | 001134224NM |
| INPP4BINPPL1 | 001101669NM001567NM |
| INSR | 000208NM |
| IRF4 | NM002460 |
| IRSI | NM005544 |
| IRS2 | NM_003749 |
| JAK1 | 002227NM |
| JAK2 | 004972NM |
| JAK3 | 000215NM |
| JUN | 002228NM |
| KDM5A | NM001042603 |
| KDM5C | 004187NM |
| KDM6A | NM021140 |
| KDR | 002253NM_ |
| KEAP1 | NM203500 |
| KIT | 000222NM |
| KLF4 | NM004235 |
| KMT2A | NM001197104 |
| KMT2B | _014727NM |
| KMT2C | 170606NM |
| KMT2D | 003482NM |
| KNSTRN | 033286NM |
| KRAS | NM033360 |
| LATS 1 | 004690NM |
| LATS2 | NM014572 |
| LMO1 | NM002315 |
| LYN | 002350NM006785 |
| MALT1MAP2K1 | NMNM002755 |
| MAP2K2 | NM030662 |
| MAP2K4 | NM003010 |
| MAP3K1 | 005921NM |
| MAP3K13 | NM_004721 |
| MAP3K14 | NM003954 |
| MAPK1 | 002745NM |
| MAPK3 | NM_002746 |
| MAPKAP1 | _001006617NM |
| MAX | NM002382 |
| MCL1 | NM021960 |
| MDC1 | NM_014641 |
| MDM2 | 002392NM |
| MDM4 | 002393NM |
| MED12 | NM005120 |
| MEF2B | 001145785NM |
| MENI | 000244NM |
| MET | 000245NM |
| MGA | 001164273NM |
| MITF | 198159NM |
| MLHI | NM_000249 |
| MPL | 005373NM_ |
| MRE11A | 005591NM_ |
| MSH2 | 000251NM |
{51}------------------------------------------------
| MSH3 | NM_002439 | ||
|---|---|---|---|
| MSH6 | NM_000179 | ||
| MSI1 | NM_002442 | ||
| MSI2 | NM_138962 | ||
| MST1 | NM_020998 | ||
| MST1R | NM_002447 | ||
| MTOR | NM_004958 | ||
| MUTYH | NM_001128425 | ||
| MYC | NM_002467 | ||
| MYCL1 | NM_001033082 | ||
| MYCN | NM_005378 | ||
| MYD88 | NM_002468 | ||
| MYOD1 | NM_002478 | ||
| NBN | NM_002485 | ||
| NCOA3 | NM_181659 | ||
| NCOR1 | NM_006311 | ||
| NEGR1 | NM_173808 | ||
| NF1 | NM_001042492 | ||
| NF2 | NM_000268 | ||
| NFE2L2 | NM_006164 | ||
| NFKBIA | NM_020529 | ||
| NKX2-1 | NM_001079668 | ||
| NKX3-1 | NM_006167 | ||
| NOTCH1 | NM_017617 | ||
| NOTCH2 | NM_024408 | ||
| NOTCH3 | NM_000435 | ||
| NOTCH4 | NM_004557 | ||
| NPM1 | NM_002520 | ||
| NRAS | NM_002524 | ||
| NSD1 | NM_022455 | ||
| NTHL1 | NM_002528 | ||
| NTRK1 | NM_002529 | ||
| NTRK2 | NM_006180 | ||
| NTRK3 | NM_001012338 | ||
| NUF2 | NM_031423 | ||
| NUP93 | NM_014669 | ||
| PAK1 | NM_002576 | ||
| PAK7 | NM_177990 | ||
| PALB2 | NM_024675 | ||
| PARK2 | NM_004562 | ||
| PARP1 | NM_001618 | ||
| PAX5 | NM_016734 | ||
| PBRM1 | NM_018313 | ||
| PDCD1 | NM_005018 | ||
| PDCD1LG2 | NM_025239 | ||
| PDGFRA | NM_006206 | ||
| PDGFRB | NM_002609 | ||
| PDPK1 | NM_002613 | ||
| PGR | NM_000926 | ||
| PHOX2B | NM_003924 | ||
| PIK3C2G | NM_004570 | ||
| PIK3C3 | NM_002647 | ||
| PIK3CA | NM_006218 | ||
| PIK3CB | NM_006219 | ||
| PIK3CD | NM_005026 | ||
| PIK3CG | NM_002649 | ||
| PIK3R1 | NM_181523 | ||
| PIK3R2 | NM_005027 | ||
| PIK3R3 | NM_003629 | ||
| PIM1 | PIM1 | NM_002648 | NM_002648 |
| PLCG2 | NM_002661 | ||
| PLK2 | NM_006622 | ||
| PMAIP1 | NM_021127 | ||
| PMS1 | NM_000534 | ||
| PMS2 | NM_000535 | ||
| PNRC1 | NM_006813 | ||
| POLD1 | NM 002691 | ||
| POLE | NM 006231 | ||
| PPARG | NM 015869 | ||
| PPM1D | NM_003620 | ||
| PPP2R1A | NM 014225 | ||
| PPP4R2 | NM_174907 | ||
| PPP6C | NM 002721 | ||
| PRDM1 | NM_001198 | ||
| PRDM14 | NM 024504 | ||
| PREX2 | NM 024870 | ||
| PRKAR1A | NM_212471 | ||
| PRKCI | NM 002740 | ||
| PRKD1 | NM 002742 | ||
| PTCH1 | NM 000264 | ||
| PTEN | NM 000314 | ||
| PTP4A1 | NM_003463 | ||
| PTPN11 | NM_002834 | ||
| PTPRD | NM 002839 | ||
| PTPRS | NM 002850 | ||
| PTPRT | NM_133170 | ||
| RAB35 | NM_006861 | ||
| RAC1 | NM 018890 | ||
| RAC2 | NM_002872 | ||
| RAD21 | NM 006265 | ||
| RAD50 | NM_005732 | ||
| RAD51 | NM 002875 | ||
| RAD51B | NM 133509 | ||
| RAD51C | NM_058216 | ||
| RAD51D | NM_133629 | ||
| RAD52 | NM_134424 | ||
| RAD54L | NM_001142548 | ||
| RAF1 | NM 002880 | ||
| RARA | NM 000964 | ||
| RASA1 | NM_002890 | ||
| RB1 | NM 000321 | ||
| RBM10 | NM 001204468 | ||
| RECQL | NM_032941 | ||
| RECQL4 | NM_004260 | ||
| REL | NM 002908 | ||
| RET | NM 020975 | ||
| RFWD2 | NM 022457 | ||
| RHEB | NM_005614 | ||
| RHOA | NM 001664 | ||
| RICTOR | NM 152756 | ||
| RIT1 | NM 006912 | ||
| RNF43 | NM_017763 | ||
| ROS1 | NM_002944 | ||
| RPS6KA4 | NM 003942 | ||
| RPS6KB2 | NM 003952 | ||
| RPTOR | NM 020761 | ||
| RRAGC | NM_022157 | ||
| RRAS | NM 006270 | ||
| RRAS2 | NM_012250 | ||
| RTEL1 | NM_032957 | ||
| RUNX1 | NM_001754 | ||
| RXRA | NM 002957 | ||
| RYBP | NM 012234 | ||
| SDHA | NM 004168 | ||
| SDHAF2 | NM_017841 | ||
| SDHB | NM 003000 | ||
| SDHC | RHEB | NM 003001 | NM_005614 |
| RHOA | NM_001664 | ||
| RICTOR | NM_152756 | ||
| RIT1 | NM_006912 | ||
| RNF43 | NM_017763 | ||
| ROS1 | NM_002944 | ||
| RPS6KA4 | NM_003942 | ||
| RPS6KB2 | NM_003952 | ||
| RPTOR | NM_020761 | ||
| RRAGC | NM_022157 | ||
| RRAS | NM_006270 | ||
| RRAS2 | NM_012250 | ||
| RTEL1 | NM_032957 | ||
| RUNX1 | NM_001754 | ||
| RXRA | NM_002957 | ||
| RYBP | NM_012234 | ||
| SDHA | NM_004168 | ||
| SDHAF2 | NM_017841 | ||
| SDHB | NM_003000 | ||
| SDHC | NM_003001 | ||
| SDHD | NM_003002 | ||
| SESN1 | NM_014454 | ||
| SESN2 | NM_031459 | ||
| SESN3 | NM_144665 | ||
| SETD2 | NM_014159 |
{52}------------------------------------------------
{53}------------------------------------------------
| SETD8 | NM020382 |
|---|---|
| SF3B1 | 012433NM |
| SH2B3 | 005475NM |
| SH2D1A | NM002351 |
| SHOC2 | NM007373 |
| SHOI | 018130NM |
| SLX4 | NM_032444 |
| SMAD2 | 001003652NM |
| SMAD3 | NM005902 |
| SMAD4 | 005359NM |
| SMARCA4 | 003072NM |
| SMARCB1 | NM003073 |
| SMARCD1 | 003076NM |
| SMO | NM005631 |
| SMYD3 | 001167740NM |
| SOCS 1SOSI | NM003745 |
| SOX17 | 005633NM022454NM |
| SOX2 | NM003106 |
| SOX9 | NM_000346 |
| SPEN | 015001NM |
| SPOP | 001007228NM |
| SPRED1 | NM152594 |
| SRC | NM198291 |
| SRSF2 | NM003016 |
| STAG2 | 001042749NM |
| STAT3 | 139276NM |
| STAT5A | 003152NM |
| STAT5B | NM012448 |
| STK11 | 000455NM |
| STK19 | NM004197 |
| STK40 | NM032017 |
| SUFU | NM_016169 |
| SUZIZ | 015355NM |
| SYK | 003177NM |
| TAP1 | 000593NM |
| TAP2 | _018833NM |
| TBX3 | 016569NM |
| TCEB1 | 005648NM |
| TCF3 | NM001136139 |
| TCF7L2 | 001146274NM |
| TEK | 000459NM |
| TERT | NM198253 |
| TETI | 030625NM |
| TET2 | 001127208NM |
| TGFBR1TGFBR2 | NM004612NM 001024847 |
| TMEM127 | 001193304NM |
| TMPRSS2 | 001135099NM_ |
| TNFAIP3 | NM_006290 |
| TNFRSF14 | NM_003820 |
| TOP1 | NM003286 |
| TP53 | 000546NM |
| TP53BP1 | NM 001141980 |
| TP63 | NM003722 |
| TRAF2 | NM021138 |
| TRAF7 | NM032271 |
| TSC1 | NM000368 |
| TSC2 | NM000548 |
| TSHR | 000369NM |
| U2AF1 | NM006758 |
| UPF1 | NM_002911 |
| VEGFA | NM_001171623 |
| VHL | 000551NM_ |
| VTCN1 | NM024626 |
| WHSCI | _001042424NM |
| WHSC1L1 | NM_023034 |
| WT1 | NM_024426 |
| WWTR1 | NM_001168280 |
| XIAP | NM_001167 |
| XPO1 | NM_003400 |
| XRCC2 | NM_005431 |
| YAP1 | NM_001130145 |
| YES1 | NM_005433 |
| ZFHX3 | NM_006885 |
{54}------------------------------------------------
| Appendix 1c: List of genes/exons excluded from reporting due to consistently low | |
|---|---|
| coverage. |
| Gene | Transcript ID | ChromosomeCoordinates | Exon | cDNA | Amino Acid | |
|---|---|---|---|---|---|---|
| AGO2 | NM_012154 | 8:141645584-141645605 | 1 | 1_22 | 1_8 | |
| ANKRD11 | NM_013275 | 16:89334886-89335071 | 13 | 7807_7992 | 2603_2664 | |
| CD276 | NM_001024736 | 15:73995113-73995427 | 4 | 419_733 | 140_245 | |
| CD276 | NM_001024736 | 15:73996517-73996813 | 6 | 1073_1369 | 358_457 | |
| CHEK2 | NM_007194 | 22:29085123-29085203 | 14 | 1462_1542 | 488_514 | |
| FAM58A | NM_152274 | X:152864420-152864529 | 1 | 1176_1 | 392_1 | |
| FLT3 | NM_004119 | 13:28674605-28674647 | 1 | 1_43 | 1_15 | |
| H3F3A | NM_002107 | 1:226259052-226259180 | 4 | 283_411 | 95_137 | |
| HIST2H3C | NM_021059 | 1:149812319-149812729 | 1 | 411_1 | 137_1 | |
| HIST2H3D | NM_001123375 | 1:149784826-149785236 | 1 | 1_411 | 1_137 | |
| HLA-A | NM_001242758 | 6:29911899-29912174 | 4 | 620_895 | 207_299 | |
| INSR | NM_000208 | 19:7293803-7293902 | 1 | 1_100 | 1_34 | |
| KMT2C | NM_170606 | 7:151970790-151970952 | 7 | 850_1012 | 284_338 | |
| KMT2C | NM_170606 | 7:151962123-151962294 | 8 | 1013_1184 | 338_395 | |
| KMT2C | NM_170606 | 7:151935792-151935911 | 15 | 2533_2652 | 845_884 | |
| KMT2C | NM_170606 | 7:151932902-151933018 | 16 | 2653_2769 | 885_923 | |
| KMT2C | NM_170606 | 7:151927008-151927112 | 18 | 2872_2976 | 958_992 | |
| KMT2C | NM_170606 | 7:151921520-151921701 | 19 | 2977_3158 | 993_1053 | |
| KMT2C | NM_170606 | 7:151921100-151921264 | 20 | 3159_3323 | 1053_1108 | 1053_1108 |
| KMT2C | NM_170606 | 7:151919658-151919767 | 21 | 3324_3433 | 1108_1145 | |
| KMT2C | NM_170606 | 7:151904385-151904513 | 24 | 3713_3841 | 1238_1281 | |
| MST1 | NM_020998 | 3:49726031-49726124 | 1 | 1_94 | 1_32 | |
| MST1 | NM_020998 | 3:49724380- | 6 | 608_728 | 203_243 | |
| MST1 | NM_020998 | 497245003:49724117-49724235 | 7 | 729_847 | 243_283 | |
| MST1 | NM_020998 | 3:49723746-49723914 | 8 | 848_1016 | 283_339 | |
| MST1 | NM_020998 | 3:49723495-49723625 | 9 | 1017_1147 | 339_383 | |
| MST1 | NM_020998 | 3:49722695-49722815 | 13 | 1424_1544 | 475_515 | |
| MST1 | NM_020998 | 3:49722445-49722522 | 14 | 1545_1622 | 515_541 | |
| MST1 | NM_020998 | 3:49721983-49722089 | 16 | 1770_1876 | 590_626 | |
| MST1 | NM_020998 | 3:49721747-49721886 | 17 | 1877_2016 | 626_672 | |
| MYCL1 | NM_001033082 | 1:40367480-40367560 | 1 | 1_81 | 1_27 | |
| NOTCH2 | NM_024408 | 1:120611948-120612020 | 1 | 1_73 | 1_25 | |
| NOTCH2 | NM_024408 | 1:120572529-120572610 | 2 | 74_155 | 25_52 | |
| NOTCH2 | NM_024408 | 1:120547952-120548211 | 3 | 156_415 | 52_139 | |
| NOTCH2 | NM_024408 | 1:120539620-120539955 | 4 | 416_751 | 139_251 | |
| NOTCH3 | NM_000435 | 19:15311599-15311716 | 1 | 1_118 | 1_40 | |
| PDPK1 | NM_002613 | 16:2588114-2588137 | 1 | 1_24 | 1_8 | |
| PDPK1 | NM_002613 | 16:2607704-2607964 | 2 | 25_285 | 9_95 | |
| PDPK1 | NM_002613 | 16:2611481-2611523 | 3 | 286_328 | 96_110 | |
| PDPK1 | NM_002613 | 16:2611772-2611909 | 4 | 329_466 | 110_156 | |
| PDPK1 | NM_002613 | 16:2615554-2615698 | 5 | 467_611 | 156_204 | |
| PDPK1 | NM_002613 | 16:2616357-2616454 | 6 | 612_709 | 204_237 | |
| PDPK1 | NM_002613 | 16:2627426-2627501 | 7 | 710_785 | 237_262 | |
| PDPK1 | NM_002613 | 16:2631296-2631364 | 8 | 786_854 | 262_285 | |
| PDPK1 | NM_002613 | 16:2631608-2631704 | 9 | 855_951 | 285_317 | |
| PDPK1 | NM_002613 | 16:2633413-2633586 | 10 | 952_1125 | 318_375 | |
| PIK3CA | NM_006218 | 3:178937737-178937840 | 13 | 1912_2015 | 638_672 | |
| PIK3R2 | NM_005027 | 19:18272089-18272305 | 6 | 599_815 | 200_272 | |
| PMS2 | NM_000535 | 7:6022455-6022622 | 12 | 2007_2174 | 669_725 | |
| PMS2 | NM_000535 | 7:6018227-6018327 | 13 | 2175_2275 | 725_759 | |
| PMS2 | NM_000535 | 7:6017219-6017388 | 14 | 2276_2445 | 759_815 | |
| PMS2 | NM_000535 | 7:6013030-6013173 | 15 | 2446_2589 | 816_863 | |
| PPP4R2 | NM_174907 | 3:73096337-73096507 | 3 | 117_287 | 39_96 | |
| PTEN | NM_000314 | 10:89725044-89725229 | 9 | 1027_1212 | 343_404 | |
| PTPRT | NM_133170 | 20:41818286-41818373 | 1 | 1_88 | 1_30 | |
| RECQL | NM_032941 | 12:21623128-21623280 | 16 | 1798_1950 | 600_650 | |
| RECQL4 | NM_004260 | 8:145743085-145743168 | 1 | 1_84 | 1_28 | |
| SDHA | NM_004168 | 5:254508-254621 | 14 | 1795_1908 | 599_636 | |
| SDHC | NM_003001 | 1:161332119-161332223 | 6 | 406_405 | 136_135 | |
| SDHD | NM_003002 | 11:111965529-111965694 | 4 | 315_480 | 105_160 | |
| SETD8 | NM_020382 | 12:123873980-123874101 | 2 | 11_132 | 4_44 | |
| SETD8 | NM_020382 | 12:123892040-123892250 | 8 | 849_1059 | 283_353 | |
| STAT5A | NM_003152 | 17:40452148-40452299 | 8 | 682_833 | 228_278 | |
| STAT5A | NM_003152 | 17:40452733-40452888 | 9 | 834_989 | 278_330 | |
| STAT5B | NM_012448 | 17:40371330-40371481 | 7 | 682_833 | 228_278 | |
| STAT5B | NM_012448 | 17:40370741-40370896 | 8 | 834_989 | 278_330 | |
| STK19 | NM_004197 | 6:31948781-31948826 | 8 | 1050_1095 | 350_365 | |
| SUZ12 | NM_015355 | 17:30267305-30267351 | 2 | 275_321 | 92_107 | |
| SUZ12 | NM_015355 | 17:30267441-30267505 | 3 | 322_386 | 108_129 | |
| SUZ12 | NM_015355 | 17:30274636-30274704 | 4 | 387_455 | 129_152 | |
| SUZ12 | NM_015355 | 17:30300165-30300250 | 6 | 506_591 | 169_197 | |
| SUZ12 | NM_015355 | 17:30310018-30310123 | 9 | 918_1023 | 306_341 | |
| TGFBR1 | NM_004612 | 9:101867488-101867584 | 1 | 1_97 | 1_33 |
{55}------------------------------------------------
{56}------------------------------------------------
§ 866.6080 Next generation sequencing based tumor profiling test.
(a)
Identification. A next generation sequencing (NGS) based tumor profiling test is a qualitative in vitro diagnostic test intended for NGS analysis of tissue specimens from malignant solid neoplasms to detect somatic mutations in a broad panel of targeted genes to aid in the management of previously diagnosed cancer patients by qualified health care professionals.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) A detailed description of all somatic mutations that are intended to be detected by the test and that are adequately supported in accordance with paragraph (b)(1)(v) of this section and reported in the test results in accordance with paragraph (b)(2)(iv) of this section, including:
(A) A listing of mutations that are cancer mutations with evidence of clinical significance.
(B) As appropriate, a listing of mutations that are cancer mutations with potential clinical significance.
(ii) The indications for use must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (
e.g., formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types (
e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable.
(iii) A detailed device description including the following:
(A) A description of the test in terms of genomic coverage, as follows:
(
1 ) Tabulated summary of all mutations reported, grouped according to gene and target region within each gene, along with the specific cDNA and amino acid positions for each mutation.(
2 ) A description of any within-gene targeted regions that cannot be reported and the data behind such conclusion.(B) Specifications for specimen requirements including any specimen collection devices and preservatives, specimen volume, minimum tumor content, specimen handling, DNA extraction, and criteria for DNA quality and quantity metrics that are prerequisite to performing the assay.
(C) A detailed description of all test components, reagents, instrumentation, and software required. Detailed documentation of the device software including but not limited to, software applications and hardware-based devices that incorporate software.
(D) A detailed description of the methodology and protocols for each step of the test, including description of the quality metrics, thresholds, and filters at each step of the test that are implemented for final result reporting and a description of the metrics for run-failures, specimen-failures, invalids, as applicable.
(E) A list of links provided by the device to the user or accessed by the device for internal or external information (
e.g., decision rules or databases) supporting clinical significance of test results for the panel or its elements in accordance with paragraphs (b)(1)(v) and (b)(2)(vi) of this section.(F) A description of internal and external controls that are recommended or provided and control procedures. The description must identify those control elements that are incorporated into the testing procedure.
(iv) Information demonstrating analytical validity of the device according to analytical performance characteristics, evaluated either specifically for each gene/mutation or, when clinically and practically justified, using a representative approach based on other mutations of the same type, including:
(A) Data that adequately supports the intended specimen type (
e.g., formalin-fixed, paraffin-embedded tumor tissue), specimen handling protocol, and nucleic acid purification for specific tumor types or for a pan-tumor claim.(B) A summary of the empirical evidence obtained to demonstrate how the analytical quality metrics and thresholds were optimized.
(C) Device precision data using clinical samples to adequately evaluate intra-run, inter-run, and total variability. The samples must cover all mutation types tested (both positive and negative samples) and include samples near the limit of detection of the device. Precision must be assessed by agreement within replicates on the assay final result for each representative mutation, as applicable, and also supported by sequencing quality metrics for targeted regions across the panel.
(D) Description of the protocols and/or data adequately demonstrating the interchangeability of reagent lots and multiplexing barcodes.
(E) A description of the nucleic acid assay input concentration range and the evidence to adequately support the range.
(F) A description of the data adequately supporting the limit of detection of the device.
(G) A description of the data to adequately support device accuracy using clinical specimens representing the intended specimen type and range of tumor types, as applicable.
(
1 ) Clinical specimens tested to support device accuracy must adequately represent the list of cancer mutations with evidence of clinical significance to be detected by the device.(
2 ) For mutations that are designated as cancer mutations with evidence of clinical significance and that are based on evidence established in the intended specimen type (e.g., tumor tissues) but for a different analyte type (e.g., protein, RNA) and/or a measurement (e.g., incorporating a score or copy number) and/or with an alternative technology (e.g., IHC, RT-qPCR, FISH), evidence of accuracy must include clinically adequate concordance between results for the mutation and the medically established biomarker test (e.g., evidence generated from an appropriately sized method comparison study using clinical specimens from the target population).(
3 ) For qualitative DNA mutations not described in paragraph (b)(1)(iv)(G)(2 ) of this section, accuracy studies must include both mutation-positive and wild-type results.(H) Adequate device stability information.
(v) Information that adequately supports the clinical significance of the panel must include:
(A) Criteria established on what types and levels of evidence will clinically validate a mutation as a cancer mutation with evidence of clinical significance versus a cancer mutation with potential clinical significance.
(B) For representative mutations of those designated as cancer mutations with evidence of clinical significance, a description of the clinical evidence associated with such mutations, such as clinical evidence presented in professional guidelines, as appropriate, with method comparison performance data as described in paragraph (b)(1)(iv)(G) of this section.
(C) For all other mutations designated as cancer mutations with potential clinical significance, a description of the rationale for reporting.
(2) The 21 CFR 809.10 compliant labeling and any product information and test report generated, must include the following, as applicable:
(i) The intended use statement must specify the following:
(A) The test is indicated for previously diagnosed cancer patients.
(B) The intended specimen type(s) and matrix (
e.g., formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types (
e.g., single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable.
(ii) A description of the device and summary of the results of the performance studies performed in accordance with paragraphs (b)(1)(iii), (b)(1)(iv), and (b)(1)(v) of this section.
(iii) A description of applicable test limitations, including, for device specific mutations validated with method comparison data to a medically established test in the same intended specimen type, appropriate description of the level of evidence and/or the differences between next generation sequencing results and results from the medically established test (
e.g., as described in professional guidelines).(iv) A listing of all somatic mutations that are intended to be detected by the device and that are reported in the test results under the following two categories or equivalent designations, as appropriate: “cancer mutations panel with evidence of clinical significance” or “cancer mutations panel with potential clinical significance.”
(v) For mutations reported under the category of “cancer mutations panel with potential clinical significance,” a limiting statement that states “For the mutations listed in [cancer mutations panel with potential clinical significance or equivalent designation], the clinical significance has not been demonstrated [with adequate clinical evidence (
e.g., by professional guidelines) in accordance with paragraph (b)(1)(v) of this section] or with this test.”(vi) For mutations under the category of “cancer mutations panel with evidence of clinical significance,” or equivalent designation, link(s) for physicians to access internal or external information concerning decision rules or conclusions about the level of evidence for clinical significance that is associated with the marker in accordance with paragraph (b)(1)(v) of this section.