K Number
K191877
Device Name
FLUOROspeed
Date Cleared
2019-08-05

(21 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital. The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications. The equipment is used for total patient population. The equipment is NOT intended to be used for Mammography screening. The equipment is NOT intended to be used for interventional procedure. The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations. Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.

Device Description

The Shimadzu FLUOROspeed is a universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes. As compared to our recently cleared system the SONIALVISION G4, (K190373) the positions of the tube head and the digital x-ray receptor panel have been swapped. This was done to respond to customer demand. This configuration is identical to products offered by Siemens (K173639) and GE Healthcare (K081091). The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat panel detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

AI/ML Overview

The provided text is a 510(k) summary for the SHIMADZU Corporation's FLUOROspeed device. It indicates that clinical testing was not performed to demonstrate substantial equivalence. Therefore, there is no information available regarding acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment related to clinical studies.

The submission relies on non-clinical testing and compliance with recognized standards to establish substantial equivalence to predicate devices. The "acceptance criteria" in this context are adherence to these specific standards and the predicate device's characteristics.

Here's a breakdown of the specific points based on the provided document:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are essentially the performance and safety requirements outlined in the listed IEC and EN standards, and the device's characteristics being "SAME" as the predicate device (SONIALVISION G4 K190373) for most specifications.

Acceptance Criteria (Predicate Characteristics & Standard Compliance)Reported Device Performance (FLUOROspeed)
Indications for Use:
Fluoroscopy/radiography diagnosis in hospital.SAME
Operated by qualified personnel.SAME
Total patient population.SAME
NOT for Mammography screening.SAME
NOT for interventional procedure.SAME
Used for radiographic, fluoroscopic, angiographic, pediatric exams.SAME
Stored images for re-monitoring, image processing, storing toSAME
optical media, sending to DICOM server.
Exception: Tomosynthesis option (for SONIALVISION G4)NOT possible due to fixed position of the tubehead. (This is a difference from the predicate, but the device is cleared without Tomosynthesis, making its indications for use similar to other non-Tomosynthesis predicates).
Physical Configuration:The primary difference is the swapped positions of the tube head and digital x-ray receptor panel. This is compared to products offered by Siemens (K173639) and GE Healthcare (K081091) as well, implying these configurations are acceptable.
Digital X-Ray Receptor Panel
Manufacturer: Canon Medical Equipment GroupSAME
Model: Canon AS-10, CXDI-401RF (SFD-1717BF)SAME
Size: 43 x 43 CMSAME
Pixel Size: 160 μmSAME
Number of Pixels: 2,688 x 2,688SAME
Acquisition Modes (Binning): Up to 3 fps (1x1), Up to 15 fps (1x1),SAME
Up to 30 fps (2x2)
DQE @ 1 µGy in 0 lp/mm, RQA5: 0.75SAME
Spatial Resolution [MTF@2cycle/mm, RQA5]: 0.28SAME
A/D Conversion: 16 BitSAME
Digital Radiography Unit: DR-300SAME
Rad/Fluoro Characteristics:
Generator: 80 kWSAME (80 kW)
Collimator: Type R-300SAME
Ceiling Tube Support: CH-200SAME
Radiography:
Tube Voltage: 40 to 150 kVSAME (40 to 150 kV)
Tube Current: 10 to 800 mASAME (10 to 800 mA)
mAs: 0.5 to 800 mAsSAME (0.5 to 800 mAs)
Time: 0.001 to 10 secSAME (0.001 to 10 sec)
Fluoroscopy:
Tube Voltage: 50 to 125 kVSAME (50 to 125 kV)
Tube Current: Max. 20 mASAME (Max. 20 mA)
Time: Total 99 min 59 sec, continuous 10 minSAME (Total 99 min 59 sec, continuous 10 min)
Pulsed Fluoroscopy: Max. 30 fps, grid-controlledSAME (Max. 30 fps, grid-controlled)
Compliance with recognized standards:The device was found to comply with: IEC 60601-1:2005+A1:2012, IEC 60601-1-3:2008+A1, IEC 60601-1-6:2010+A1, IEC/EN 60601-2-28:2010, IEC 60601-2-54:2009+A1, IEC 62304:2006, IEC 62366:2007+A1, EN 60601-1-2:2015/IEC 60601-1-2:2014. It is also certified to comply with the US Radiation Safety Performance Standard.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. No clinical test set. The evaluation was based on non-clinical testing and comparison to predicate device specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No clinical test set requiring expert ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No clinical effectiveness study involving human readers or AI assistance was performed. This device is an X-ray system, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an X-ray system, not an algorithm, and no standalone algorithmic performance was evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. No clinical ground truth was established as no clinical studies were conducted. The "ground truth" for this submission is established by the specifications and performance of the legally marketed predicate device and compliance with international safety and performance standards.

8. The sample size for the training set

Not applicable. There is no mention of a training set as this is a medical imaging device, not an AI/ML diagnostic algorithm.

9. How the ground truth for the training set was established

Not applicable. There is no mention of a training set or its ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

SHIMADZU Corporation % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

August 5, 2019

Re: K191877

Trade/Device Name: FLUOROspeed Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA Dated: July 1, 2019 Received: July 15, 2019

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191877

Device Name FLUOROspeed

Indications for Use (Describe)

The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital.

The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications.

The equipment is used for total patient population.

The equipment is NOT intended to be used for Mammography screening.

The equipment is NOT intended to be used for interventional procedure.

The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations.

Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: 510(k) Number K191877

Image /page/3/Picture/1 description: The image shows the Shimadzu logo. The logo consists of a square with a white cross inside, followed by the word "SHIMADZU" in bold, black letters. The cross is centered within the square, and the wordmark is aligned to the right of the square.

Shimadzu Corporation 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone : +81-75-823-1305 Fax : +81-75-823-1377 Date Prepared: July 26, 2019 Contact: Toshio Kadowaki; Phone: +81-75-823-1920 E-mail: kadowaki@shimadzu.co.jp

1. Identification of the Device:

Trade/Device Names: FLUOROspeed Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system

2. Equivalent legally marketed device: K190373

Trade/Device Name: SONIALVISION G4 Manufacturer: Shimadzu. Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system REFERENCE/SIMILARLY CONFIGURED DEVICES: K173639 (Siemens) and K081091 (GE) Trade/Device Names: Luminos (Siemens), Precision (GE) Manufacturers: Siemens and GE. Regulation Numbers: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB (Siemens) JAA (GE) Common/Usual Name: Fluoroscopic x-ray system

    1. Indications for Use (intended use): The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital. The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications. The equipment is used for total patient population. The equipment is NOT intended to be used for Mammography screening. The equipment is NOT intended to be used for interventional procedure. The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations. Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
    1. Description of the Device: The Shimadzu FLUOROspeed is a universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes. As compared to our recently cleared system the SONIALVISION G4, (K190373) the positions of the tube head and the digital x-ray receptor panel have

{4}------------------------------------------------

been swapped. This was done to respond to customer demand. This configuration is identical to products offered by Siemens (K173639) and GE Healthcare (K081091). The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat panel detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations. Features:

• A large-field 17-inch FPD and digital image processing enable you to acquire high quality diagnostic images. A second FPD can be added as an option. This is the wireless AeroDR SYSTEM 2 1417HQ cleared by FDA in K151465.

• The equipment employs a high-frequency inverter system with maximum frequency of 50 kHz for high-voltage generation and ensures high efficiency by low-ripple output.

• The X-ray diagnostic table holds a wide examination area and allows radiography from head to foot without the need to move the patient.

• The tabletop can be raised or lowered in the height range from 65 cm to 110 cm to help the patient get on or off and allow the operator to take an unforced pose according to required examinations. • The operator can assist the patient while operating the X-ray diagnostic table on its front control

panel.

• The control panel is equipped with a 10.4-inch LCD touch panel, where the operator can view the patient information or change the fluoroscopy or radiography condition. The screen layout or display options can be switched as appropriate according to the purpose of examination. Customizable hard switches help you access to the frequently used functions with a single touch of a button.

• Low-dose exposure and high image quality are achieved at the same time by incorporating automatic selection of BH-filter (X-ray filter), wave-tail cut-off mechanism for pulsed fluoroscopy, detachable anti-scatter grid, virtual collimation, and single-mask collimator (option).

• A series of radiography conditions for examination can be registered in a program to repeat the same series of radiography.

• Image data can be stored for a prolonged period by saving it on DVD-R or CD-R disks in DICOM format.

• The equipment supports DICOM3.0 and allows smooth connection to a hospital network. DICOM Print, Media Storage, DICOM Storage, MWM/MPPS (option) and RDSR (option) are also supported

    1. Safety and Effectiveness, comparison to predicate device. The results of bench and standards testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation. Most of the components employed are identical to the predicate device.

6. Substantial Equivalence Chart: Below.

{5}------------------------------------------------

SONIALVISION G4 K190373FLUOROspeed
Indications forUse:The equipment is intended to be used for thefluoroscopy/radiography diagnosis in hospital.The equipment must only be operated byqualified personnel, such as radiographytechnicians or those with equivalentqualifications. The equipment is used for totalpatient population. The equipment is NOTintended to be used for Mammographyscreening. The equipment is NOT intended to beused for interventional procedure. Theequipment is used for radiographic, fluoroscopic,angiographic and pediatric examinations. Storedimages in the equipment can be used for re-monitoring, image processing, storing to opticalmedia (CD/DVD), and sending to DICOM server.The Tomosynthesis option for the SONIALVISIONG4 is intended to generate tomographic images ofhuman anatomy including chest or extremities.Tomosynthesis technique is used to produce aspecific cross-sectional plane of the body byreconstruction of tomographic acquisition.The equipment is intended to be used for thefluoroscopy/radiography diagnosis in hospital.The equipment must only be operated byqualified personnel, such as radiographytechnicians or those with equivalentqualifications. The equipment is used for totalpatient population. The equipment is NOTintended to be used for Mammographyscreening. The equipment is NOT intended to beused for interventional procedure. Theequipment is used for radiographic,fluoroscopic, angiographic and pediatricexaminations. Stored images in the equipmentcan be used for re-monitoring, imageprocessing, storing to optical media (CD/DVD),and sending to DICOM server.. THE SAMEEXCEPT: No Tomosynthesis. Tomosynthesis isnot possible because of the fixed position ofthe tubehead.
Position of the X-ray tube headOn top, aimed down at the patient resting onthe tableUnderneath the table aimed up through thetable
Position of thedigital x-rayreceptor panelIn a slot in the tableOverhead where the tubehead used to be.
Digital X-Ray Receptor Panel (Previously Cleared in K171194)
ManufacturerCanon Medical Equipment GroupSAME
ModelCanon AS-10, CXDI-401RF (K171194)We call it the SFD-1717BFSAME
Size43 x 43 CMSAME
Pixel Size160 μmSAME
Number of Pixels2,688 x 2,688SAME
AcquisitionModes (Binning)Up to 3 fps (1x1)Up to 15 fps (1x1)Up to 30 fps (2x2)SAME
DQE @ 1 µGy in 0lp/mm, RQA50.75SAME
Spatial Resolution[MTF@2cycle/mm, RQA5]0.28SAME
A/D Conversion16 BitSAME
Digital RadiographyUnitDR-300SAME
SONIALVISION G4 K190373FLUOROspeed
Rad/Fluoro Characteristics
GeneralAppearanceImage: SONIALVISION G4 K190373Image: FLUOROspeed
Generator80 kW80 kW SAME
CollimatorType R-300SAME
Ceiling TubeSupportCH-200SAME
Radiography (Unchanged from Predicate)
Tube Voltage40 to 150 kV40 to 150 kV SAME
Tube Current10 to 800 mA10 to 800 mA SAME
mAs0.5 to 800 mAs0.5 to 800 mAs SAME
Time0.001 to 10 sec0.001 to 10 sec SAME
Fluoroscopy (Unchanged from Predicate)
Tube Voltage50 to 125 kV50 to 125 kV SAME
Tube CurrentMax. 20 mAMax. 20 mA SAME
TimeTotal time display: 99 min 59 sec.,continuous fluoroscopy time: 10 minTotal time display: 99 min 59 sec.,continuous fluoroscopy time: 10 minSAME
PulsedFluoroscopyMax. 30 fps, grid-controlled.Max. 30 fps, grid-controlled. SAME

{6}------------------------------------------------

    1. Summary of non-clinical testing: This device has been tested and is certified to comply with the US Radiation Safety Performance Standard. Performance and safety testing was conducted by third party NRTL certified testing laboratories and the device was found to comply with the following FDA recognized standards:
    • IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 . reprint); IEC 60601-1 :2005+Corrigendum 1+Corrigendum 2+A1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, FDA 19-4
    • . IEC 60601-1-3:2008+A1 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment FDA 12-269

{7}------------------------------------------------

  • IEC 60601-1-6:2010+A1 General requirements for basic safety and essential performance ● Collateral standard: Usability FDA 5-89
  • IEC/EN 60601-2-28:2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis FDA 12-204
  • IEC 60601-2-54:2009+A1 Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, FDA 12-296
  • IEC 62304:2006 Medical device software – Software life cycle processes, FDA 13-32
  • IEC 62366:2007+A1 Application of usability engineering to medical devices, FDA 5-114
  • EN 60601-1-2:2015, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests FDA 19-8

We consulted the following FDA guidance documents in the development of the FLUOROspeed: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff; and Pediatric Information for X-ray Imaging Device Premarket Notifications.

    1. Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
    1. Conclusion: After analyzing standards compliance results and bench tests, it is the conclusion of Shimadzu Corporation that the MODIFIED FLUOROspeed is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.