(21 days)
No
The document does not mention AI, ML, or any related technologies like deep learning or neural networks. The description focuses on the physical configuration and standard imaging capabilities.
No
The device is described as a diagnostic imaging system, intended for fluoroscopy/radiography diagnosis, and is explicitly stated as "NOT intended to be used for interventional procedure", indicating it does not provide therapy.
Yes.
The "Intended Use / Indications for Use" section explicitly states, "The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital." Additionally, the "Device Description" calls it a "universal diagnostic imaging system."
No
The device description explicitly states it is a "universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes" and mentions physical components like a "tube head" and "digital x-ray receptor panel," indicating it is a hardware device with integrated software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the equipment is for "fluoroscopy/radiography diagnosis in hospital." This involves imaging the internal structures of the body using X-rays, which is an in vivo diagnostic method (performed within the living body).
- IVD Definition: In Vitro Diagnostics (IVDs) are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. This device does not perform tests on such samples.
- Device Description: The description confirms it's an "X-ray RF system offering radiographic, fluoroscopic, and angiographic modes," all of which are imaging techniques performed on the patient directly.
The mention of "image processing" and storing images does not change its classification as an in vivo diagnostic imaging device. These are functions related to the manipulation and storage of the images acquired from the patient's body.
N/A
Intended Use / Indications for Use
The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital.
The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications.
The equipment is used for total patient population.
The equipment is NOT intended to be used for Mammography screening.
The equipment is NOT intended to be used for interventional procedure.
The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations.
Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
Product codes (comma separated list FDA assigned to the subject device)
JAA
Device Description
The Shimadzu FLUOROspeed is a universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes. As compared to our recently cleared system the SONIALVISION G4, (K190373) the positions of the tube head and the digital x-ray receptor panel have been swapped. This was done to respond to customer demand. This configuration is identical to products offered by Siemens (K173639) and GE Healthcare (K081091). The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat panel detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations. Features:
• A large-field 17-inch FPD and digital image processing enable you to acquire high quality diagnostic images. A second FPD can be added as an option. This is the wireless AeroDR SYSTEM 2 1417HQ cleared by FDA in K151465.
• The equipment employs a high-frequency inverter system with maximum frequency of 50 kHz for high-voltage generation and ensures high efficiency by low-ripple output.
• The X-ray diagnostic table holds a wide examination area and allows radiography from head to foot without the need to move the patient.
• The tabletop can be raised or lowered in the height range from 65 cm to 110 cm to help the patient get on or off and allow the operator to take an unforced pose according to required examinations. • The operator can assist the patient while operating the X-ray diagnostic table on its front control panel.
• The control panel is equipped with a 10.4-inch LCD touch panel, where the operator can view the patient information or change the fluoroscopy or radiography condition. The screen layout or display options can be switched as appropriate according to the purpose of examination. Customizable hard switches help you access to the frequently used functions with a single touch of a button.
• Low-dose exposure and high image quality are achieved at the same time by incorporating automatic selection of BH-filter (X-ray filter), wave-tail cut-off mechanism for pulsed fluoroscopy, detachable anti-scatter grid, virtual collimation, and single-mask collimator (option).
• A series of radiography conditions for examination can be registered in a program to repeat the same series of radiography.
• Image data can be stored for a prolonged period by saving it on DVD-R or CD-R disks in DICOM format.
• The equipment supports DICOM3.0 and allows smooth connection to a hospital network. DICOM Print, Media Storage, DICOM Storage, MWM/MPPS (option) and RDSR (option) are also supported
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
Not Found
Indicated Patient Age Range
Total patient population
Intended User / Care Setting
Qualified personnel, such as radiography technicians or those with equivalent qualifications. In hospital.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Summary of non-clinical testing: This device has been tested and is certified to comply with the US Radiation Safety Performance Standard. Performance and safety testing was conducted by third party NRTL certified testing laboratories and the device was found to comply with the following FDA recognized standards:
- IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 . reprint); IEC 60601-1 :2005+Corrigendum 1+Corrigendum 2+A1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, FDA 19-4
- . IEC 60601-1-3:2008+A1 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment FDA 12-269
- IEC 60601-1-6:2010+A1 General requirements for basic safety and essential performance ● Collateral standard: Usability FDA 5-89
- IEC/EN 60601-2-28:2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis FDA 12-204
- IEC 60601-2-54:2009+A1 Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, FDA 12-296
- IEC 62304:2006 Medical device software – Software life cycle processes, FDA 13-32
- IEC 62366:2007+A1 Application of usability engineering to medical devices, FDA 5-114
- EN 60601-1-2:2015, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests FDA 19-8
Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1650 Image-intensified fluoroscopic x-ray system.
(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
SHIMADZU Corporation % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
August 5, 2019
Re: K191877
Trade/Device Name: FLUOROspeed Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA Dated: July 1, 2019 Received: July 15, 2019
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
1
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
Device Name FLUOROspeed
Indications for Use (Describe)
The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital.
The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications.
The equipment is used for total patient population.
The equipment is NOT intended to be used for Mammography screening.
The equipment is NOT intended to be used for interventional procedure.
The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations.
Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary: 510(k) Number K191877
Image /page/3/Picture/1 description: The image shows the Shimadzu logo. The logo consists of a square with a white cross inside, followed by the word "SHIMADZU" in bold, black letters. The cross is centered within the square, and the wordmark is aligned to the right of the square.
Shimadzu Corporation 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone : +81-75-823-1305 Fax : +81-75-823-1377 Date Prepared: July 26, 2019 Contact: Toshio Kadowaki; Phone: +81-75-823-1920 E-mail: kadowaki@shimadzu.co.jp
1. Identification of the Device:
Trade/Device Names: FLUOROspeed Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system
2. Equivalent legally marketed device: K190373
Trade/Device Name: SONIALVISION G4 Manufacturer: Shimadzu. Regulation Number: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Common/Usual Name: Fluoroscopic x-ray system REFERENCE/SIMILARLY CONFIGURED DEVICES: K173639 (Siemens) and K081091 (GE) Trade/Device Names: Luminos (Siemens), Precision (GE) Manufacturers: Siemens and GE. Regulation Numbers: 21CFR892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB (Siemens) JAA (GE) Common/Usual Name: Fluoroscopic x-ray system
-
- Indications for Use (intended use): The equipment is intended to be used for the fluoroscopy/radiography diagnosis in hospital. The equipment must only be operated by qualified personnel, such as radiography technicians or those with equivalent qualifications. The equipment is used for total patient population. The equipment is NOT intended to be used for Mammography screening. The equipment is NOT intended to be used for interventional procedure. The equipment is used for radiographic, fluoroscopic, angiographic and pediatric examinations. Stored images in the equipment can be used for re-monitoring, image processing, storing to optical media (CD/DVD), and sending to DICOM server.
-
- Description of the Device: The Shimadzu FLUOROspeed is a universal X-ray RF system offering radiographic, fluoroscopic, and angiographic modes. As compared to our recently cleared system the SONIALVISION G4, (K190373) the positions of the tube head and the digital x-ray receptor panel have
4
been swapped. This was done to respond to customer demand. This configuration is identical to products offered by Siemens (K173639) and GE Healthcare (K081091). The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat panel detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations. Features:
• A large-field 17-inch FPD and digital image processing enable you to acquire high quality diagnostic images. A second FPD can be added as an option. This is the wireless AeroDR SYSTEM 2 1417HQ cleared by FDA in K151465.
• The equipment employs a high-frequency inverter system with maximum frequency of 50 kHz for high-voltage generation and ensures high efficiency by low-ripple output.
• The X-ray diagnostic table holds a wide examination area and allows radiography from head to foot without the need to move the patient.
• The tabletop can be raised or lowered in the height range from 65 cm to 110 cm to help the patient get on or off and allow the operator to take an unforced pose according to required examinations. • The operator can assist the patient while operating the X-ray diagnostic table on its front control
panel.
• The control panel is equipped with a 10.4-inch LCD touch panel, where the operator can view the patient information or change the fluoroscopy or radiography condition. The screen layout or display options can be switched as appropriate according to the purpose of examination. Customizable hard switches help you access to the frequently used functions with a single touch of a button.
• Low-dose exposure and high image quality are achieved at the same time by incorporating automatic selection of BH-filter (X-ray filter), wave-tail cut-off mechanism for pulsed fluoroscopy, detachable anti-scatter grid, virtual collimation, and single-mask collimator (option).
• A series of radiography conditions for examination can be registered in a program to repeat the same series of radiography.
• Image data can be stored for a prolonged period by saving it on DVD-R or CD-R disks in DICOM format.
• The equipment supports DICOM3.0 and allows smooth connection to a hospital network. DICOM Print, Media Storage, DICOM Storage, MWM/MPPS (option) and RDSR (option) are also supported
-
- Safety and Effectiveness, comparison to predicate device. The results of bench and standards testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation. Most of the components employed are identical to the predicate device.
6. Substantial Equivalence Chart: Below.
5
SONIALVISION G4 K190373 | FLUOROspeed | |
---|---|---|
Indications for | ||
Use: | The equipment is intended to be used for the | |
fluoroscopy/radiography diagnosis in hospital. | ||
The equipment must only be operated by | ||
qualified personnel, such as radiography | ||
technicians or those with equivalent | ||
qualifications. The equipment is used for total | ||
patient population. The equipment is NOT | ||
intended to be used for Mammography | ||
screening. The equipment is NOT intended to be | ||
used for interventional procedure. The | ||
equipment is used for radiographic, fluoroscopic, | ||
angiographic and pediatric examinations. Stored | ||
images in the equipment can be used for re- | ||
monitoring, image processing, storing to optical | ||
media (CD/DVD), and sending to DICOM server. | ||
The Tomosynthesis option for the SONIALVISION | ||
G4 is intended to generate tomographic images of | ||
human anatomy including chest or extremities. | ||
Tomosynthesis technique is used to produce a | ||
specific cross-sectional plane of the body by | ||
reconstruction of tomographic acquisition. | The equipment is intended to be used for the | |
fluoroscopy/radiography diagnosis in hospital. | ||
The equipment must only be operated by | ||
qualified personnel, such as radiography | ||
technicians or those with equivalent | ||
qualifications. The equipment is used for total | ||
patient population. The equipment is NOT | ||
intended to be used for Mammography | ||
screening. The equipment is NOT intended to be | ||
used for interventional procedure. The | ||
equipment is used for radiographic, | ||
fluoroscopic, angiographic and pediatric | ||
examinations. Stored images in the equipment | ||
can be used for re-monitoring, image | ||
processing, storing to optical media (CD/DVD), | ||
and sending to DICOM server.. THE SAME | ||
EXCEPT: No Tomosynthesis. Tomosynthesis is | ||
not possible because of the fixed position of | ||
the tubehead. | ||
Position of the X- | ||
ray tube head | On top, aimed down at the patient resting on | |
the table | Underneath the table aimed up through the | |
table | ||
Position of the | ||
digital x-ray | ||
receptor panel | In a slot in the table | Overhead where the tubehead used to be. |
Digital X-Ray Receptor Panel (Previously Cleared in K171194) | ||
Manufacturer | Canon Medical Equipment Group | SAME |
Model | Canon AS-10, CXDI-401RF (K171194) | |
We call it the SFD-1717BF | SAME | |
Size | 43 x 43 CM | SAME |
Pixel Size | 160 μm | SAME |
Number of Pixels | 2,688 x 2,688 | SAME |
Acquisition | ||
Modes (Binning) | Up to 3 fps (1x1) | |
Up to 15 fps (1x1) | ||
Up to 30 fps (2x2) | SAME | |
DQE @ 1 µGy in 0 | ||
lp/mm, RQA5 | 0.75 | SAME |
Spatial Resolution | ||
[MTF@2cycle/mm | ||
, RQA5] | 0.28 | SAME |
A/D Conversion | 16 Bit | SAME |
Digital Radiography | ||
Unit | DR-300 | SAME |
SONIALVISION G4 K190373 | FLUOROspeed | |
Rad/Fluoro Characteristics | ||
General | ||
Appearance | Image: SONIALVISION G4 K190373 | Image: FLUOROspeed |
Generator | 80 kW | 80 kW SAME |
Collimator | Type R-300 | SAME |
Ceiling Tube | ||
Support | CH-200 | SAME |
Radiography (Unchanged from Predicate) | ||
Tube Voltage | 40 to 150 kV | 40 to 150 kV SAME |
Tube Current | 10 to 800 mA | 10 to 800 mA SAME |
mAs | 0.5 to 800 mAs | 0.5 to 800 mAs SAME |
Time | 0.001 to 10 sec | 0.001 to 10 sec SAME |
Fluoroscopy (Unchanged from Predicate) | ||
Tube Voltage | 50 to 125 kV | 50 to 125 kV SAME |
Tube Current | Max. 20 mA | Max. 20 mA SAME |
Time | Total time display: 99 min 59 sec., | |
continuous fluoroscopy time: 10 min | Total time display: 99 min 59 sec., | |
continuous fluoroscopy time: 10 min | ||
SAME | ||
Pulsed | ||
Fluoroscopy | Max. 30 fps, grid-controlled. | Max. 30 fps, grid-controlled. SAME |
6
-
- Summary of non-clinical testing: This device has been tested and is certified to comply with the US Radiation Safety Performance Standard. Performance and safety testing was conducted by third party NRTL certified testing laboratories and the device was found to comply with the following FDA recognized standards:
- IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 . reprint); IEC 60601-1 :2005+Corrigendum 1+Corrigendum 2+A1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, FDA 19-4
- . IEC 60601-1-3:2008+A1 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment FDA 12-269
7
- IEC 60601-1-6:2010+A1 General requirements for basic safety and essential performance ● Collateral standard: Usability FDA 5-89
- IEC/EN 60601-2-28:2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis FDA 12-204
- IEC 60601-2-54:2009+A1 Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, FDA 12-296
- IEC 62304:2006 Medical device software – Software life cycle processes, FDA 13-32
- IEC 62366:2007+A1 Application of usability engineering to medical devices, FDA 5-114
- EN 60601-1-2:2015, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests FDA 19-8
We consulted the following FDA guidance documents in the development of the FLUOROspeed: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff; and Pediatric Information for X-ray Imaging Device Premarket Notifications.
-
- Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
-
- Conclusion: After analyzing standards compliance results and bench tests, it is the conclusion of Shimadzu Corporation that the MODIFIED FLUOROspeed is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.