K Number
K191466
Date Cleared
2019-11-07

(157 days)

Product Code
Regulation Number
882.5880
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the fascia or interspinous/supraspinous ligament.

Device Description

The Stimwave Technologies Incorporated (Stimwave) SandShark Injectable Anchor (SIA) System is used to fixate the Stimwave Freedom Stimulator to surrounding tissue. The System is comprised of a carbothane anchor (SandShark Anchor) that is transferred onto the deployment handle (SandShark Injectroducer) with the Loading Rod and Loading Base. The SIA System is provided sterile. The SandShark Injectroducer is used to deploy the SandShark Anchor onto the Stimulator.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study proving device efficacy:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text describes a 510(k) submission for the SandShark Injectable Anchor (SIA) System. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through clinical trials with defined acceptance criteria for efficacy. Therefore, explicit "acceptance criteria" for performance metrics like sensitivity, specificity, or outcomes improvements are not present in this document.

Instead, the acceptance criteria are implicit in matching the predicate device's characteristics and passing established safety and performance tests.

Acceptance Criteria CategorySpecific Criteria (Implicit from text)Reported Device Performance
Intended UseSame as predicate device (K172644)Same as K172644
Technological CharacteristicsSimilar design and materials to predicate deviceMinor design update to Loading Base (hand-held vs. table-top), otherwise same as K172644
BiocompatibilityComplies with ISO 10993-1:2009Meets biological safety and compatibility requirements
SterilizationEthylene Oxide (EO) process, sterile labeling, single-useSame as K172644 (Ethylene Oxide)
Material SafetyNo negative impacts from materialsDemonstrated no negative impacts
Physical Performance (e.g., Temperature, Pressure)Functional after temperature/pressure changes as per AAMI ANSI ISO 14708-3:2008Passed temperature and atmospheric pressure change testing
Design RequirementsMeets system design requirementsComplies with all design requirements
Applicable Voluntary StandardsAdherence to relevant standards (e.g., AAMI ANSI ISO 14708-3:2008)Complies with applicable voluntary standards
Anchor DurabilityLeveraged from K172644Leveraged testing from K172644

2. Sample Size Used for the Test Set and Data Provenance

This document describes a premarket notification (510(k)) that focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study establishing efficacy in a patient population.

  • Sample Size for Test Set: Not applicable in the context of clinical testing with a "test set" of patients. The "test set" here refers to the actual device components and materials subjected to engineering and biocompatibility testing. The document does not specify exact numbers for these components but indicates that "Stimwave completed a number of tests for the SIA System."
  • Data Provenance: The testing data is from the manufacturer, Stimwave Technologies Inc. The data is non-clinical performance data and biocompatibility data. It is internal to the company's development and testing process. No country of origin for a patient data set is given as a clinical study was not performed.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • This is not applicable as the submission did not involve a test set requiring expert ground truth establishment in the clinical sense (e.g., medical image interpretation). The "ground truth" for non-clinical and biocompatibility tests would be defined by the specifications of the test methods and the pass/fail criteria of those standards.

4. Adjudication Method for the Test Set

  • Not applicable for this type of submission. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation or subjective assessments. The tests described are objective, pass/fail engineering and biological evaluations.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a physical medical device (an injectable anchor system), not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with or without AI assistance is irrelevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, a standalone algorithm performance study was not done. This is a physical medical device and does not involve an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the evaluations performed in this submission are:

  • Established industry standards and regulatory guidance: ISO 10993-1:2009 for biocompatibility, AAMI ANSI ISO 14708-3:2008 for physical performance (temperature/pressure changes), and Blue Book Memorandum G95-1.
  • Design requirements and specifications: The device was tested to verify that "the performance meets the system design requirements."
  • Predicate device characteristics: Substantial equivalence is established by comparing the device's characteristics and performance to the legally marketed predicate device (K172644). The predicate device's established safety and efficacy serve as a form of "ground truth" for equivalence.

8. The Sample Size for the Training Set

  • Not applicable. This is not a machine learning or AI-driven device, so there is no "training set" in that sense.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 7, 2019

Stimwave Technologies Inc. Elizabeth Greene Chief Compliance Officer 1310 Park Central Boulevard South Pompano Beach, Florida 33064

Re: K191466

Trade/Device Name: SandShark Injectable Anchor (SIA) System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZB Dated: August 9, 2019 Received: August 9, 2019

Dear Elizabeth Greene:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191466

Device Name SandShark Injectable Anchor (SIA) System

Indications for Use (Describe)

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the fascia or interspinous/supraspinous ligament.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words "stim" and "wave" is the word "TECHNOLOGIES" in dark blue. To the right of the word "wave" is a graphic of three curved lines that resemble radio waves.

510(k) Summary

for

SandShark Injectable Anchor (SIA) System

1. Submission Sponsor

Stimwave Technologies Incorporated 1310 Park Central Boulevard South Pompano Beach FL, 33064 USA Phone: 800.965.5134 Fax: 800.965.5134 Contact: Elizabeth Greene, Chief Compliance Officer

2. Date Prepared

May 31, 2019

3. Device Identification

Trade/Proprietary Name:SandShark Injectable Anchor System
Classification Name:Stimulator, Spinal-Cord, Implanted (Pain Relief)
Classification Regulation:882.5880
Product Code:GZB
Device Class:Class II
Classification Panel:Neurology

4. Legally Marketed Predicate Device(s)

SandShark Injectable Anchor (SIA) System (K172644)

5. Device Description

The Stimwave Technologies Incorporated (Stimwave) SandShark Injectable Anchor (SIA) System is used to fixate the Stimwave Freedom Stimulator to surrounding tissue. The System is comprised of a carbothane anchor (SandShark Anchor) that is transferred onto the deployment handle (SandShark Injectroducer) with the Loading Rod and Loading Base. The SIA System is provided sterile. The SandShark Injectroducer is used to deploy the SandShark Anchor onto the Stimulator.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Stimwave Technologies. The logo features the word "stim" in blue, followed by "wave" in green. To the right of "wave" are three concentric arcs in blue, resembling radio waves. Below the words is the word "TECHNOLOGIES" in blue, enclosed in a blue rounded rectangle.

SandShark Injectable Anchor (SIA) System

InjectroducerAn acrylonitrile butadiene styrene (ABS) handle and cannula(stainless steel 304V, KR01 Phillips K-Resin) anchor deploymentdevice that is used to secure the SandShark Anchor onto thestimulator to the fascia or interspinous/supra-spinous ligament.Identical to K172644.
SandSharkAnchorA carbothane (80A) anchor that is deployed by the Injectroducer ontothe stimulator securing the device to the fascia or interspinous/supra-spinous ligament. Four (4) SandShark Anchors are provided in theSIA System, pre-loaded onto the Loading Rod. Identical to K172644.
Loading RodAn assembly that is used with the Loading Base to transfer theSandShark Anchor onto the cannula of the Injectroducer. The handleis constructed of ABS and the rod is stainless steel 304V. Identical toK172644.
Loading BaseAn ABS base that holds the Loading Rod in place while transferringthe SandShark Anchor onto the cannula of the Injectroducer. Thedesign has been updated to be hand-held; no change in material.

6. Indication for Use Statement

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the stimulator to the fascia or interspinous/supra-spinous ligament.

7. Substantial Equivalence Discussion

The following table compares the Stimwave SIA System to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

ComparatorStimwave SandShark Injectable Anchor (SIA) System(K172644)Stimwave SandShark Injectable Anchor (SIA) System(This submission)
Product CodeGZBSame as K172644
Regulation No.882.5880Same as K172644
Regulation NameStimulator, Spinal-Cord, Implanted (Pain Relief)Same as K172644
Intended UseAccessory for securing a spinal cord stimulator tosurrounding tissueSame as K172644
Implant SiteFascia or inter-spinous/supra-spinous ligamentSame as K172644
Environmental UseHospital or Ambulatory Surgical Center OnlySame as K172644
Intended ClinicianOrthopedic, Neurosurgeon, AnesthesiologistSame as K172644
Anchor MaterialCarbothane 80ASame as K172644
Cannula MaterialStainless Steel 304VSame as K172644
Anchor Length1.5 inchSame as K172644
Anchor Outer Diameter0.087 inchSame as K172644
Loading BaseTable-top designHand-held design
Method of IntroductionPercutaneous and Anchor IncisionSame as K172644
Table 5A. Comparison of Characteristics
-----------------------------------------------

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in purple, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a rounded rectangle. To the right of the word "wave" is a blue circular wave pattern.

ComparatorStimwave SandShark Injectable Anchor (SIA) System(K172644)Stimwave SandShark Injectable Anchor (SIA) System(This submission)
Tissue ContactYesSame as K172644
SterilizationEthylene Oxide (EO)Same as K172644
LabelingLabeled as Sterile, Single Use, Prescription DeviceSame as K172644
SterileYes - ethylene oxideSame as K172644
Single-UseYesSame as K172644
Shelf Life1 yearSame as K172644
Complies with ISO 10993-1YesSame as K172644
Safety Testing PassedYesSame as K172644

(*) asterisk denotes that formulas were used for the calculations.

8. Biocompatibility Data

The materials, construction and intended use of the SIA System are identical to the predicate device, and have a long history of safety with respect to biocompatibility, thus the biological safety testing of the SIA System is leveraged from K172644. The biological safety of the SandShark Anchor was evaluated in accordance to ISO 10993-1:2009, guidance document Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, implantation (4, 8, and 13 weeks), subacute and subchronic toxicity, chronic toxicity, carcinogenicity, extractables and leechables demonstrated no negative impacts from the materials that are used in the SIA System. The SIA System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.

9. Non-Clinical Performance Data

The SIA System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The SIA System complies with all design requirements and applicable voluntary standards.

AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the SIA System was functional, receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the SIA System was functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008. The testing is leveraged from K172644 and is directly applicable for demonstration of device safety and efficacy as the packaging, mode of action, and materials remain the same. The design modification to the Loader Base does not impact the outcome of the leveraged tests for safety and effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, purple, sans-serif font, and the word "wave" is in a large, green, sans-serif font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller, purple, sans-serif font. To the right of the words is a graphic of three curved lines that resemble radio waves, also in purple.

Stimwave completed a number of tests for the SIA System that demonstrates substantial equivalence to the legally marketed predicate device. The SIA System meets all the requirements for overall design, sterilization, and biocompatibility confirms that the output meets the design inputs and specifications through leveraged or new testing described in this submission. The SIA System passed all testing stated above as shown by the acceptable results obtained. The updated design of the Loading Base does not impact device performance, and thus, demonstrates continued safety and efficacy compared to the predicate device. The inclusion of secondary contract suppliers for the SIA System demonstrates no impact to established safety and efficacy.

10. Clinical Performance Data

There was no clinical testing required to support the medical device, as the indications for use are equivalent to the legally marketed predicate device. These types of devices, including the legally marketed predicate device, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to legally marketed predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. The SIA System has the same intended use as the legally marketed predicates device. Performance testing data leveraged from K172644 verifies that the SIA System complies with all applicable voluntary standards such as AAMI ANSI ISO 14708-3. The SIA System also meets the design requirements where no applicable standard could be used. This includes leveraged anchor durability testing, as well as biocompatibility and sterilization validation of the SIA System. There were no recognized performance standards for this device. There was no clinical testing performed on this device since performance testing demonstrated similar performance as the legally marketed predicate device, and materials for the SIA System are the same as the legally marketed predicate device. The updated design of the Loading Base does not impact device performance, and thus, demonstrates continued safety and efficacy compared to the predicate. The inclusion of secondary contract suppliers for the SIA System demonstrates no impact to established safety and efficacy.

It has been shown in this 510(k) submission that the difference between the SIA System and the legally marketed predicate device do not raise any questions regarding its safety and effectiveness as compared to legally marketed predicate device. The SIA System, as designed and manufactured, is determined to be substantially equivalent to the referenced legally marketed predicate device.

§ 882.5880 Implanted spinal cord stimulator for pain relief.

(a)
Identification. An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.(b)
Classification. Class II (performance standards).